Duke Energy
Laboratory Services
13339 Hagers Ferry Road Bld. MG03A
Huntersville, NC 28078
Project ID: Ash Basin GW Assessment - Invista
Hexavalent Chromium
EPA Method 218.6
Analytical Report
23952
Element One, Inc.
6319-D Carolina Beach Rd., Wilmington, NC 28412
910-793-0128 FAX: 910-792-6853 e1lab@e1lab.com
el1
The following data for Analytical Report 23952
has been reviewed for completeness, accuracy,
adherence to method protocol,
and compliance with quality assurance guidelines.
Review by:
r"
Katie Strickland, B.S. Chemist
February 26, 2015
Report Reviewed and Finalized By:
K n Smith, Laboratory Director
February 26, 2015
e0e entOne
23952 Duke Invista Cr6 Report Packet
Page 2 of 20
elementOne
23952 Duke Invista Cr6 Report Packet
Page 3 of 20
SUMMARY OF RESULTS
elementOne
23952 Duke Invista Cr6 Report Packet
Page 4 of 20
Summary of Analysis
Summary of Method 218.6 Hexavalent Chromium Analysis
Invista/S1/OH2 – DW1 Invista/S1/OH2 – DW1 Invista/S1/OG2 – DW3
23952-1 23952-1 dup 23952-2
Element µg/L µg/L µg/L
----------------- --------------- --------------- ---------------
Chromium (VI) < 0.1 < 0.1 < 0.1
Summary of Standard Method 4500H+ pH Analysis
Invista/S1/OH2 – DW1 Invista/S1/OG2 – DW 3
23952-1 23952-2
Assay SU SU
----------------- --------------- ---------------
pH 5.45 6.49
Summary of Standard Method 2500 B Temperature Analysis
Invista/S1/OH2 – DW1 Invista/S1/OG2 – DW3
23952-1 23952-2
Assay °C °C
----------------- --------------- ---------------
Temperature 18.1 16.9
elementOne
23952 Duke Invista Cr6 Report Packet
Page 5 of 20
ANALYTICAL NARRATIVE
elementOne
23952 Duke Invista Cr6 Report Packet
Page 6 of 20
Element One Analytical Narrative
Client: Duke Energy Element One #: 23952
Client ID: Ash Basin GW Assessment - Invista Analyst: PQT, LAW
Method: EPA 218.6 Dates Received: 02/25/2015
Analytes: Cr (VI) Dates Analyzed: 02/25-26/2015
Summary of Analysis
The samples were collected by Element One, Inc. on 02/25/2015. The samples
were analyzed for pH and temperature in the field immediately following
sampling. Samples were returned to Element One, Inc. and were prepared and
analyzed for chromium (VI) according to SW -846 Method 0061 protocol.
Samples were analyzed for chromium (VI) using a Shimadzu 2010 HPLC unit
and a Timberline RDR-1 Reaction Module.
Detection Limits
The Shimadzu 2010 HPLC instrument detection limit for chromium (VI) was 0.1
µg/L.
Analysis QA/QC
Duplicate analyses relative percent difference (RPD) and sample spike recovery
data are summarized in the Quality Control Section. All QA/QC data was within
the criteria of the method.
Additional Comments
The reported results have not been corrected for any blank values or spike
recovery values. The reported results relate only to the items tested or
calibrated.
elementOne
23952 Duke Invista Cr6 Report Packet
Page 7 of 20
QUALITY CONTROL SUMMARY
elementOne
23952 Duke Invista Cr6 Report Packet
Page 8 of 20
Summary of Quality Control Data
218.6 Duplicate Analysis RPD
(QC limits: <10% for RPD)
Invista/S1/OG2 – DW 3
23952-2
Element RPD
----------------- ---------------
Chromium (VI) NA
elementOne
23952 Duke Invista Cr6 Report Packet
Page 9 of 20
SAMPLE CUSTODY
elementOne
23952 Duke Invista Cr6 Report Packet
Page 10 of 20
elementOne
23952 Duke Invista Cr6 Report Packet
Page 11 of 20
elementOne
23952 Duke Invista Cr6 Report Packet
Page 12 of 20
ANALYTICAL DATA
elementOne
23952 Duke Invista Cr6 Report Packet
Page 13 of 20
Analytical Calculations
Chromium (VI)-
Chromium (VI) (µg/L) = HPLC Results (µg/L)*Dilution
Where-
HPLC Results = Raw sample concentration (ppb)—HPLC -Data Sheet
Dilution= Diluted Volume--HPLC Run Sheet
Aliquot
elementOne
23952 Duke Invista Cr6 Report Packet
Page 14 of 20
Analytical Calculations
Spike Recovery-
Spike (%) = (Spiked Result (µg/L) – Sample Result (µg/L)) X100
Spike Amount (µg/L)
Where-
Spike Result = Raw sample concentration (ppb)--HPLC-Data Sheet
Sample Result = Raw sample concentration (ppb)--HPLC-Data Sheet
Spike Amount-- HPLC-Data Sheet
Duplicate Analysis RPD-
RPD (%) = (Duplicate Result (µg/L) - Sample Result (µg/L)) X100
Average (µg/L)
Where-
Sample Result and Duplicate Results=Raw sample concentration (ppb) -- HPLC-
Data Sheet
Average= (Duplicate + Sample Results)
2
elementOne
23952 Duke Invista Cr6 Report Packet
Page 15 of 20
elementOne
23952 Duke Invista Cr6 Report Packet
Page 16 of 20
elementOne
23952 Duke Invista Cr6 Report Packet
Page 17 of 20
elementOne
23952 Duke Invista Cr6 Report Packet
Page 18 of 20
elementOne
23952 Duke Invista Cr6 Report Packet
Page 19 of 20
elementOne
23952 Duke Invista Cr6 Report Packet
Page 20 of 20