NC Department of Environmental Quality | Division of Water Resources | Laboratory Certification Branch
4405 Reedy Creek Road | 1623 Mail Service Center | Raleigh, North Carolina 27699-1623
919-733-3908
October 4, 2023
5779
Mr. Brian Doliber
NC DOT Ferry Division
159 Lucinda Lane
Powells Point, NC 27966
Subject: North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC
WW/GW LCB) Maintenance Inspection
Dear Mr. Doliber:
Enclosed is a report for the inspection performed on September 14, 2023 by Tom Halvosa. Where
Finding(s) are cited in this report, a response is required. Within thirty days, please supply this
office with a written item for item description of how these Finding(s) were corrected. Please
describe the steps taken to prevent recurrence and include an implementation date for each
corrective action. If the Finding(s) cited in the enclosed report are not corrected, enforcement
actions may be recommended. For Certification maintenance, your laboratory must continue to
carry out the requirements set forth in 15A NCAC 02H .0800.
A copy of the laboratory’s Certified Parameter List at the time of the audit is attached. This list will
not reflect any changes made during the audit. Copies of the checklists completed during the
inspection may be requested from this office. Thank you for your cooperation during the
inspection. If you have questions or need additional information, please contact me at (919) 733-
3908 Ext. 259.
Sincerely,
Beth Swanson
Technical Assistance & Compliance Specialist
Division of Water Resources
Attachment
cc: Todd Crawford, Tom Halvosa, #5779
On-Site Inspection Report
LABORATORY NAME: NC DOT Ferry Division
NPDES PERMIT #: NC0056065
WATER QUALITY PERMIT #: WQ0000265 and WQ0012696
ADDRESS: 8550 Shipyard Rd.
Manns Harbor, NC 27953
CERTIFICATE #: 5779
DATE OF INSPECTION: September 14, 2023
TYPE OF INSPECTION: Field Initial
AUDITOR: Tom Halvosa
LOCAL PERSON(S) CONTACTED:
David Pharr, Bill Neeland and Amanda Grenier
I. INTRODUCTION:
This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory
Certification Branch (NC WW/GW LCB) to verify its compliance with the requirements of 15A NCAC 02H
.0800 for the analysis of compliance monitoring samples.
II. GENERAL COMMENTS:
Staff were forthcoming and responded well to suggestions from the auditor.
All required Proficiency Testing (PT) Samples have been analyzed and the laboratory has fulfilled its PT
requirements for the 2023 PT Calendar Year.
Any time changes are made to laboratory procedures, Quality Assurance (QA) and/or Standard Operating
Procedure (SOP) document(s) must be updated and relevant staff retrained. Staff must acknowledge that
they have read and understand the changes as part of the documented training program. The same
requirements apply when changes are made in response to Findings, Recommendations or Comments
listed in this report, to ensure the methods are being performed as stated, references to methods are
accurate, and the QA and/or SOP document(s) is in agreement with each approved practice, test, analysis,
measurement, monitoring procedure or regulatory requirement being used in the laboratory. In some
instances, the laboratory may need to create an SOP to document how new functions or policies will be
implemented. Revisions to the SOPs, based on the Findings, Comments and Recommendations
within this report must be submitted to this office with the report response.
The laboratory is reminded that SOPs are intended to describe procedures exactly as they are to be
performed. Use of the word “should” is not appropriate when describing requirements (e.g., Quality Control
(QC) frequency, acceptance criteria, etc.). Evaluate all SOPs for the proper use of the word “should”.
Page 2
#5779 NC DOT Ferry Division
Contracted analyses are performed by Environmental Chemists, Inc. (Certification # 94).
Approved Procedure documents for the analysis of the facility’s currently certified Field Parameters were
provided at the time of the inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Documentation
A. Finding: Error corrections are not always properly performed.
Requirement: All documentation errors shall be corrected by drawing a single line through
the error so that the original entry remains legible. Entries shall not be obliterated by
erasures or markings. Wite-Out®, correction tape, or similar products designed to obliterate
documentation are not to be used; instead the correction shall be written adjacent to the
error. The correction shall be initialed by the responsible individual and the date of change
documented. Ref: 15A NCAC 02H .0805 (g) (1).
Comment: There were several instances of incorrect error corrections such as write-
overs and no date of change or initials of the responsible individual.
B. Finding: Chemical containers are not consistently dated when received and when opened.
Requirement: Chemical containers shall be dated when received and when opened. Ref:
15A NCAC 02H .0805 (g) (7).
Temperature – Standard Methods, 2550 B-2010 (Aqueous)
Comment: Temperature is being reported to 0.1°C on the Discharge Monitoring Report (DMR).
Recommendation: Unless greater precision is required by the permit or data receiving agency,
it is recommended that all temperatures reported for compliance monitoring, be reported in whole
numbers as recommended by the DWR’s Precision in Discharge Monitoring Reports document.
Reporting
C. Finding: Data qualifiers from the contract laboratory reports are not being transferred to the
DMR.
Requirement: Each certified Field Laboratory shall be in accordance with Paragraph (e)
of this Rule. Ref: 15A NCAC 02H .0805 (g) (17).
Requirement: Reported data associated with quality control failures, improper sample
collection, holding time exceedances, or improper preservation shall be qualified as such.
Ref: 15A NCAC 02H .0805 (e) (5).
Comment: During data review, one contract lab report included a qualifier for the BOD
analysis performed August 2, 2023. The laboratory did not transfer this qualifier to the DMR.
Recommendation: To avoid questions of legality, it is recommended that you contact the
appropriate Regional Office for guidance as to whether an amended DMR(s) will be required.
A copy of this report will be made available to the Regional Office.
Page 3
#5779 NC DOT Ferry Division
IV. CONCLUSIONS:
Correcting the above-cited Findings and implementing the Recommendations will help this
laboratory to produce quality data and meet Certification requirements. The inspector would like
to thank the staff for their assistance during the inspection and data review process. Please
respond to all Findings and include supporting documentation, implementation dates and
steps taken to prevent recurrence for each corrective action.
Report prepared by: Tom Halvosa Date: September 18, 2023
Report reviewed by: Tonja Springer Date: September 19, 2023
Certificate Number:5779
Effective Date:3/24/2023
Expiration Date:12/31/2023 3
Lab Name:NC DOT Ferry Division
Address:8550 Shipyard Rd.
Manns Harbor, NC 27953
North Carolina Wastewater/Groundwater Laboratory Certification
Certified Parameters Listing
Date of Last Amendment:
The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below.
CERTIFIED PARAMETERS
INORGANIC
CHLORINE, TOTAL RESIDUAL
SM 4500 Cl G-2011 (Aqueous)
CONDUCTIVITY
SM 2510 B-2011 (Aqueous)
DISSOLVED OXYGEN
SM 4500 O H-2016 (Aqueous)
pH
SM 4500 H+B-2011 (Aqueous)
TEMPERATURE
SM 2550 B-2010 (Aqueous)
This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions
as set forth in 15A NCAC 2H.0807.