HomeMy WebLinkAbout#5542 08-Final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 5542
Laboratory Name: Normandeau Associates
Inspection Type: Initial Field Inspection
Inspector Name(s): Mr. Ramon Cook
Inspection Date: September 17, 2008
Date Report Completed: September 22, 2008
Date Forwarded to Reviewer: September 22, 2008
Reviewed by: David Livingston
Date Review Completed: October 14, 2008
Cover Letter to use: X Insp. Initial Insp. Reg. Insp. No Finding Insp. CP
Unit Supervisor: Dana Satterwhite
Date Received: 10/17/2008
Date Forwarded to Alberta: 10/30/2008
Date Mailed:
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Normandeau Associates
ADDRESS: 1019B Cherokee Avenue
Aiken, SC 29801
CERTIFICATE #: 5542
DATE OF INSPECTION: September 17, 2008
TYPE OF INSPECTION: Field Initial
AUDITOR(S): Ramon Cook
LOCAL PERSON(S) CONTACTED: Jeff Wollis
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for the
analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory is spacious and well equipped. All facilities and instrumentation are well maintained.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Temperature – Standard Methods, 18th Edition, 2550 B
Dissolved Oxygen – Standard Methods, 18th Edition, 4500-O G
Conductivity – Standard methods, 18th Edition, 2510B
pH – Standard Methods, 18th Edition, 4500 H+ B
A. Finding: The Temperature correction factors were not documented on any of the
thermometers/meters in use.
Requirement: All thermometers and temperature sensing devices on meters must be
calibrated against a NIST certified or NIST traceable thermometer annually (every 12 months)
and proper corrections made and documented. The thermometer/meter reading must be less
than or equal to 1ºC from the NIST certified or NIST traceable thermometer reading to be
acceptable. Document the serial number and manufacturer of the NIST thermometer that was
used in the comparison. Document any correction that applies (even if zero) on both the
thermometer/meter and on a separate sheet to be filed. Ref: Technical Assistance for Field
Analysis of Dissolved Oxygen, Rev. 07/2008. Ref: Technical Assistance for Field Analysis of
Conductivity, Rev. 09/2007. Ref: Technical Assistance for Field Analysis of pH, Rev. 10/2008.
Ref: Technical Assistance for Field Analysis of Temperature, Rev. 07/2008.
Conductivity- Standard methods, 18th Edition, 2510 B
B. Finding: The Automatic Temperature Compensator (ATC) has not been verified.
Page 2
# 5542 Normandeau Associates
Requirement: The ATC must be verified annually (i.e., every twelve months) by analyzing a
buffer at 25°C (the temperature that conductivity values are compensated to) and a
temperature(s) that bracket the temperature ranges of the samples to be analyzed. This may
require the analysis of a third temperature reading that is > 25°C. Ref: North Carolina
Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) Policy, 02/2006.
C. Finding: The meter calibration time is not documented each day of use as required.
Requirement: The following criteria must be documented whenever sample analysis is
performed: date and time of sample collection, date and time of sample analysis,
collector/analyst’s name, sample site/sample identification, value of standard used for
calibration, instrument calibration time, concentration and obtained value of the check standard,
reported values for all samples in μmhos/cm at 25ºC or corrected to 25ºC, The temperature
correction (even if it is zero) must be posted on the meter as well as in hard copy format (to be
retained for 5 years). Ref: NC WW/GW LC Technical Assistance for Field Analysis of
Conductivity Samples, Rev. 09/2007.
D. Finding: The meter calibration is not verified each analysis day.
Requirement: Analyze and document a calibration check standard. (It is recommended that this
standard be at a lower range). The value obtained from the calibration check standard must
read within ± 10% of the true value of the check standard. If the value is outside of the ± 10%
range, corrective action must be taken. Ref: Technical Assistance for Field Analysis of
Conductivity Samples, Rev. 09/07
pH – Standard Methods, 18th Edition, 4500 H+ B
E. Finding: The Automatic Temperature Compensator (ATC) has not been verified.
Requirement: The ATC must be verified annually (i.e., every twelve months) by analyzing a
buffer at 25°C (the temperature that pH values are compensated to) and a temperature(s) that
bracket the temperature ranges of the samples to be analyzed. This may require the analysis of
a third temperature reading that is > 25°C. Ref: North Carolina Wastewater/Groundwater
Laboratory Certification (NC WW/GW LC) Policy Statement 09/2007.
F. Finding: The lab is not analyzing a third buffer to verify pH instrument calibration.
Requirement: The laboratory must calibrate the pH meter with at least two buffers and read
a third buffer below pH 10, approximately 3 pH units different from the second buffer and
the reading should be within 0.1 pH units from the third buffer. If the meter response shows
a difference greater than 0.1 pH unit from expected value, look for trouble with the
electrodes or potentiometer. Ref: Standard Methods 18th Edition 4500-H+ B. (4) (a).
G. Finding: The time of sample analysis is not documented.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data
must consist of date collected, time collected, samples site, sample collector, and sample
analysis time. The field bench sheets must provide a space for the signature of the analyst, and
proper units of measure for all analyses. Ref: 15A NCAC 2H .0805 (g) (1).
Page 3
# 5542 Normandeau Associates
Temperature – Standard Methods, 18th Edition, 2550 B
H. Finding: The NIST traceable thermometer that was used to check the temperature sensor was
expired. The recalibration date was March 2001.
Requirement: NIST traceable thermometers used for temperature measurements must be
recalibrated in accordance with the manufacturer’s recalibration date. If no recalibration date is
given, the NIST traceable thermometer must be recalibrated annually Ref: Technical Assistance
for Field Analysis of Temperature, Rev. 07/2008. Please submit a copy of the temperature
sensor check performed with an appropriately calibrated NIST thermometer and the NIST
thermometer calibration documentation along with the report reply.
IV. PAPER TRAIL INVESTIGATION:
A review of data was conducted. This consisted of comparing benchsheets to field logs and data
submitted to clients. No transcription errors were detected. The facility appears to be doing a good job
of accurately transcribing data.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendations will help this lab to produce
quality data and meet certification requirements. The inspector would like to thank the staff for its
assistance during the inspection and data review process. Please respond to all findings.
Report prepared by: Ramon L. Cook Date: September 22, 2008
Report reviewed by: David Livingston Date: October 14, 2008