INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 5425
Laboratory Name: SS MHP Laboratory
Inspection Type: Field Maintenance
Inspector Name(s): Jeffrey R. Adams
Inspection Date: November 16, 2009
Date Report Completed: December 8, 2009
Date Forwarded to Reviewer: December 14, 2009
Reviewed by: Todd Crawford
Date Review Completed: December 16, 2009 __________________
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding Insp. CP
Unit Supervisor: Dana Satterwhite
Date Received: December 18, 2009
Date Forwarded to Alberta: December 22, 2009
Date Mailed:
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: SS MHP
NPDES PERMIT #: NC0038300
ADDRESS: 1808 Pinecrest Street
Burlington, NC 27215
CERTIFICATE #: 5425
DATE OF INSPECTION: November 16, 2009
TYPE OF INSPECTION: Field Maintenance
AUDITOR(S): Jeffrey R. Adams
LOCAL PERSON(S) CONTACTED: James Edwards
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for the
analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was neat and all equipment was clean and appeared well maintained. Records were well
maintained and organized, however, some quality control procedures need to be implemented.
The laboratory was given copies of the Technical Assistance Guidance Documents for all field parameters
during the inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Dissolved Oxygen – Standard Methods, 18th Edition, 4500 O G
Recommendation: The sponge being used in the probe storage well was dirty. Consult the operators’
manual for cleaning and maintaining the probe.
A. Finding: The proper units of measure (mg/L) are not being documented on the benchsheet.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data
must consist of date collected, time collected, sample site, sample collector, and sample
analysis time. The field bench sheets must provide a space for the signature or initials of the
analyst, and proper units of measure for all analyses. Ref: 15A NCAC 02H .0805 (g) (1).
pH – Standard Methods, 18th Edition, 4500 H+ B
B. Finding: The proper units of measure (S.U.) are not being documented on the benchsheet.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data
must consist of date collected, time collected, sample site, sample collector, and sample
analysis time. The field bench sheets must provide a space for the signature or initials of the
analyst, and proper units of measure for all analyses. Ref: 15A NCAC 02H .0805 (g) (1).
Page 2
#5425 SS MHP
Total Residual Chlorine – Standard Methods, 18th Edition, 4500 Cl G
C. Finding: The 200 µg/L Gel® Standard abeing used as a daily calibration verification standard
is not being verified with the standard curve annually.
Requirement: Purchased “Gel-type” standards may be used for daily calibration verification only.
The "Gel-type" standards must be verified every 12 months with the standard curve. Ref:
Technical Assistance for Field Analysis of Total Residual Chlorine.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing contract lab reports to Discharge Monitoring Reports (DMRs)
submitted to the North Carolina Division of Water Quality. Data were reviewed for NPDES permit
#NC0038300 for July, August and September, 2009. No transcription errors were detected. The facility
appears to be doing a good job transcribing data.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendation will help this lab to produce
quality data and meet certification requirements. The inspector would like to thank the staff for its assistance
in the inspection and data review process. Please respond to all findings.
Report prepared by: Jeffrey R. Adams Date: December 8, 2009
Report reviewed by: Todd Crawford Date: December 16, 2009