HomeMy WebLinkAbout#5572-final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 5572
Laboratory Name: Camden county Wastewater Treatment Plant
Inspection Type: Field Initial
Inspector Name(s): Todd Crawford
Inspection Date: February 25, 2010
Date Report Completed: March 1, 2010
Date Forwarded to Reviewer: March 1, 2010
Reviewed by: Jason Smith
Date Review Completed: March 8, 2010
Cover Letter to use: X Insp. Initial Insp. Reg. Insp. No Finding Insp. CP
Unit Supervisor: Dana Satterwhite
Date Received: March 9, 2010
Date Forwarded to Alberta: March 15, 2010
Date Mailed: March 15, 2010
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Camden County Wastewater Treatment Plant
PERMIT #: WQ0029894
ADDRESS: 129 Keeter Barn Rd
South Mills, NC 27976
CERTIFICATE #: 5572
DATE OF INSPECTION: February 25, 2010
TYPE OF INSPECTION: Field Initial
AUDITOR(S): Todd Crawford
LOCAL PERSON(S) CONTACTED: Tony Conant
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for
the analysis of environmental samples.
II. GENERAL COMMENTS:
This was the first inspection for this facility. The lab was very clean and well organized. All equipment
necessary to perform the certified parameters were present and appeared well maintained.
A description of current field lab policies and technical assistance documents for the analysis of the
facility’s currently certified parameters were provided at the time of the inspection.
The facility is certified for Dissolved Oxygen and Temperature; however, reporting of these parameters
is not required by their permit.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
pH – Standard Methods, 20th Edition, 4500 H+ B
Comment: The facility is currently documenting a “duplicate” sample analysis for pH. The pH analysis
is being performed in-situ, which would not allow for a true duplicate. In addition, duplicate analysis of
pH samples is not required for Field Labs.
Settleable Solids – Standard Methods, 20th Edition, 2540 F
A. Finding: The sample collection time is not being documented.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data
must consist of date collected, time collected, sample site, sample collector, and sample
analysis time. The field benchsheets must provide a space for the signature or initials of the
analyst, and proper units of measure for all analyses. Ref: 15A NCAC 2H .0805 (g) (1).
Page 2
#5572 Camden County Wastewater Treatment Plant
Documentation
B. Finding: Error corrections are not being done in an approved manner.
Requirement: All documentation errors must be corrected by drawing a single line through the
error so that the original entry remains legible. Entries shall not be obliterated by erasures or
markings. Wite-Out®, correction tape or similar products designed to obliterate documentation
are not to be used. Write the correction adjacent to the error. The correction must be initialed
by the responsible individual and the date of change documented. All data and log entries must
be written in indelible ink. Pencil entries are not acceptable. Ref: North Carolina
W astewater/Groundwater Laboratory Certification document, Quality Assurance Policies for
Field Laboratories.
C. Finding: Preparation and/or purchase of standards, reagents and consumables are not being
documented in such a way as to provide traceability.
Requirement: All chemicals, reagents, standards and consumables used by the laboratory
must have the following information documented: Date Received, Date Opened (in use),
Vendor, Lot Number, and Expiration Date. A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the
solutions are used. Documentation of solution preparation must include the analyst’s initials,
date of preparation, the volume or weight of standard(s) used, the solvent and final volume of
the solution. This information as well as the vendor, manufacturer, lot number, and expiration
date must be retained for chemicals, reagents, standards and consumables used for a period of
five years. Consumable materials such as pH buffers and lots of pre-made standards are
included in this requirement. Ref: North Carolina W astewater/ Groundwater Laboratory
Certification document, Quality Assurance Policies for Field Laboratories.
D. Finding: The sample collection sites are not being documented on the benchsheets.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data
must consist of date collected, time collected, sample site, sample collector, and sample
analysis time. The field bench sheets must provide a space for the signature or initials of the
analyst, and proper units of measure for all analyses. Ref: 15A NCAC 02H .0805 (g) (1).
E. Finding: The proper units of measure are not being documented on the benchsheets.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data
must consist of date collected, time collected, sample site, sample collector, and sample
analysis time. The field bench sheets must provide a space for the signature or initials of the
analyst, and proper units of measure for all analyses. Ref: 15A NCAC 02H .0805 (g) (1).
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing field testing records and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for permit # WQ0029894 for October, November and December, 2009. The following errors
were detected:
Page 2
#5572 Camden County Wastewater Treatment Plant
The monitoring report column headings for TOC, Chloride, TDS, Nitrate and Turbidity did not
contain any concentration units.
Results reported as less than the detection limit are not being treated as zero when calculating
the arithmetic mean.
In addition to those stated above, the following transcription error was noted:
Date Parameter Location Value on Benchsheet Value on DMR
12/02/09 pH Effluent 6.7 S.U. 6.4 S.U.
In order to avoid questions of legality, it is recommended that you contact the appropriate Regional Office
for guidance as to whether an amended Monitoring Reports will be required. A copy of this report will be
forwarded to the Regional Office.
V. CONCLUSIONS:
Correcting the above-cited findings will help this lab to produce quality data and meet certification
requirements. The inspector would like to thank the staff for its honesty and assistance during the
inspection and subsequent data discovery and review process. Please respond to all findings.
Report prepared by: Todd Crawford Date: March 1, 2010
Report reviewed by: Jason Smith Date: March 8, 2010