INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 5257
Laboratory Name: Brunswick County Utilities, Northwest WTP
Inspection Type: Field Municipal Maintenance
Inspector Name(s): Todd Crawford
Inspection Date: May 12, 2011
Date Report Completed: May 24, 2011
Date Forwarded to Reviewer: May 24, 2011
Reviewed by: David Livingston
Date Review Completed: May 27, 2011
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding Insp. CP Corrected
Unit Supervisor: Dana Satterwhite
Date Received: June 3, 2011
Date Forwarded to Alberta: June 21, 2011
Date Mailed: June 22, 2011
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Brunswick County Utilities, Northwest WTP
NPDES PERMIT #: NC0057533
ADDRESS: P.O. Box 249
Bolivia, NC 28422
CERTIFICATE #: 5257
DATE OF INSPECTION: May 12, 2011
TYPE OF INSPECTION: Field Municipal Maintenance
AUDITOR(S): Todd Crawford
LOCAL PERSON(S) CONTACTED: David Rourk
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for
the analysis of environmental samples.
II. GENERAL COMMENTS:
The lab area was clean and well organized. All equipment necessary to perform the certified
parameters were present and appeared well maintained.
Improvements in quality control and documentation are needed.
A description of current field lab policies and technical assistance documents for the analysis of the
facility’s currently certified parameters were provided at the time of the inspection.
The requirements associated with Finding A are new since the last inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Documentation
A. Finding: The laboratory needs to increase the documentation of purchased materials and
reagents, as well as, documentation of standards and reagents prepared in the laboratory.
Requirement: All chemicals, reagents, standards and consumables used by the laboratory
must have the following information documented: Date Received, Date Opened (in use),
Vendor, Lot Number, and Expiration Date. A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the
solutions are used. Documentation of solution preparation must include the analyst’s initials,
date of preparation, the volume or weight of standard(s) used, the solvent and final volume of
the solution. This information as well as the vendor and/or manufacturer, lot number, and
expiration date must be retained for chemicals, reagents, standards and consumables used for
a period of five years. Consumable materials such as pH buffers and lots of pre-made
standards are included in this requirement. Ref: Quality Assurance Policies for Field
Laboratories.
Page 2
#5257 Brunswick County Utilities, Northwest WTP
Total Residual Chlorine – Standard Methods, 18th Edition, 4500 Cl G
B. Finding: The time of the daily calibration check standard analysis is not being documented.
Requirement: The time and date that the daily calibration check standard is analyzed must be
documented each day that sample analysis occurs. Ref: Technical Assistance for Field Analysis
of Total Residual Chlorine.
C. Finding: Values below the lowest calibration curve standard concentration are being reported
on the Discharge Monitoring Report (DMR).
Requirement: The concentrations of the calibration standards must bracket the concentrations
of the samples analyzed. One of the standards must have a concentration equal to or below the
lower reporting concentration for Total Residual Chlorine. Ref: Technical Assistance for Field
Analysis of Total Residual Chlorine.
Comment: If the laboratory chooses to have a lower reporting limit of 20 µg/L for residual
chlorine, you must analyze at least 20 µg/L or lower standard and report lower concentrations
as < 20 µg/L or < the concentration of the chosen standard.
D. Finding: The gel standard used as the daily check standard is not being analyzed on the same
meter program as the samples.
Requirement: Calibration checks must be analyzed on the same curve and/or program used
for sample analysis. Ref: Technical Assistance for Field Analysis of Total Residual Chlorine.
E. Finding: The gel standard used as the daily check standard has not been assigned a true
value on the meter program used for routine samples.
Requirement: Purchased “Gel-type” or sealed liquid ampoule standards used for daily
calibration verification must be verified initially and every 12 months thereafter, with the
standard curve. Ref: Technical Assistance for Field Analysis of Total Residual Chlorine.
Requirement: Analyze a new 5-point curve in the range from 20 – 400 µg/L (e.g., 20, 30, 50,
200, and 400 µg/L). Immediately following the new curve analysis, read the gel standard on the
same meter program and assign the observed value as the true value for that gel standard. The
assigned values will be used for the next twelve months, or until a new curve verification is
performed. The gel/liquid standard verification must be performed for each instrument on which
they are to be used. If multiple instruments and/or standard sets are used, each must have
assigned values specific for the instrument and standard set. Ref: Technical Assistance for Field
Analysis of Total Residual Chlorine. Submit all documentation with your response to this
report.
Residue, Settleable – Standard Methods, 18th Edition, 2540 F
Total Residual Chlorine – Standard Methods, 18th Edition, 4500 Cl G
F. Finding: The sampling location is not documented on the benchsheet.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified
Data must consist of date collected, time collected, sample site, sample collector, and sample
analysis time. The field benchsheets must provide a space for the signature or initials of the
analyst, and proper units of measure for all analyses. Ref: 15A NCAC 2H .0805 (g) (1).
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing field testing records and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for NPDES permit # NC0057533 for January, February and March, 2011. No transcription
errors were detected. The facility appears to be doing a good job of accurately transcribing data.
V. CONCLUSIONS:
Correcting the above-cited findings will help this lab to produce quality data and meet certification
requirements. The inspector would like to thank the staff for its assistance during the inspection and data
review process. Please respond to all findings.
Report prepared by: Todd Crawford Date: May 25, 2011
Report reviewed by: David Livingston Date: May 27, 2011