HomeMy WebLinkAbout#5069 11-final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: #5069
Laboratory Name: Town of Smithfield Water Treatment Plant
Inspection Type: Field Maintenance
Inspector Name(s): Jeffrey R. Adams
Inspection Date: September 28, 2011
Date Report Completed: October 5, 2011
Date Forwarded to Reviewer: October 5, 2011
Reviewed by: Jason Smith
Date Review Completed: October 13, 2011
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding Insp. CP __ Corrected
Unit Supervisor: Dana Satterwhite
Date Received: October 14, 2011
Date Forwarded to Linda October 21, 2011
Date Mailed: October 21, 2011
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Town of Smithfield Water Treatment Plant
NPDES PERMIT #: NC0083348
ADDRESS: P.O. Box 761
Smithfield, NC 27577
CERTIFICATE #: 5069
DATE OF INSPECTION: September 28, 2011
TYPE OF INSPECTION: Field Maintenance
AUDITOR(S): Jeffrey R. Adams
LOCAL PERSON(S) CONTACTED: Judy Bateman and Ray Peal
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for
the analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was clean and well organized. Records were well organized and maintained according to
the appropriate retention schedule, however, some quality control procedures need to be implemented.
Proficiency testing samples have been analyzed for all certified parameters for the 2011 proficiency
testing calendar year and the graded results were 100% acceptable.
The laboratory was given a packet containing North Carolina Laboratory Certification quality control
requirements and policies during the inspection.
Findings A and B are new policies that have been implemented since the last inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Traceability
A. Finding: The laboratory needs to increase the documentation of purchased materials and
reagents, as well as documentation of standards and reagents prepared in the laboratory.
Requirement: All chemicals, reagents, standards and consumables used by the laboratory
must have the following information documented: Date Received, Date Opened (in use),
Vendor, Lot Number, and Expiration Date. A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the
solutions are used. Documentation of solution preparation must include the analyst’s initials,
date of preparation, the volume or weight of standard(s) used, the solvent and final volume of
the solution. This information as well as the vendor and/or manufacturer, lot number, and
expiration date must be retained for chemicals, reagents, standards and consumables used for
a period of five years. Consumable materials such as pH buffers and lots of pre-made
Page 2
#5069 Town of Smithfield Water Treatment Plant
standards are included in this requirement. Ref: Quality Assurance Policies for Field
Laboratories.
Pipettors
B. Finding: The auto-pipettors have not been calibrated annually as required.
Requirement: Mechanical volumetric liquid-dispensing devices (e.g., fixed and adjustable auto-
pipettors, bottle-top dispensers, etc.) must be calibrated at least every twelve months and
documented. Each liquid-dispensing device must meet the manufacturer’s statement of
accuracy. For variable volume devices used at more than one setting, check the accuracy at
the maximum, middle and minimum values. Testing at more than three volumes is optional.
When a device capable of variable settings is dedicated to dispense a single specific volume,
calibration is required at that setting only. Ref: Quality Assurance Policies for Field
Laboratories. A copy of the pipettor calibrations must be submitted with the response to
this report. (Refer to the attachment titled, Calibration of Mechanical Volumetric
Liquid-Dispensing Devices Procedure for an example procedure for calibration of liquid
dispensing devices).
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing laboratory benchsheets and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for the Town of Smithfield (NPDES permit #NC0083348) for the following months: June, July
and August, 2011. No transcription errors were detected. It appears the laboratory is doing a good job
of accurately transcribing data.
V. CONCLUSIONS:
Correcting the above-cited findings will help this lab to produce quality data and meet certification
requirements. The inspector would like to thank the staff for its assistance during the inspection and
data review process. Please respond to all findings.
Report prepared by: Jeffrey R. Adams Date: October 5, 2011
Report reviewed by: Jason Smith Date: October 13, 2011
Calibration of Mechanical Volumetric Liquid-Dispensing Devices Procedure
A laboratory may use the following method for the gravimetric determination of measurement error of
mechanical volumetric liquid-dispensing devices (e.g., fixed and adjustable auto-pipettors, bottle-top
dispensers, etc.) used to dispense volumes greater than or equal to 100 μl.
Apparatus and Equipment
Weighing vessel (of sufficient capacity to hold all volumes dispensed for the test and a 3:1 height-to-
diameter ratio is recommended to minimize evaporation)
Reagent water (e.g., distilled or deionized water)
Analytical balance (accurate to 0.0001 g for 100 μl to 10 ml, and accurate to 1 mg for 10 ml to 200 ml
dispensed volumes)
Calibrated thermometer (graduated, at a minimum, in 0.5 ºC increments)
Documentation Requirements
Date
Analyst's signature or initials
Device identification (if there is no unique identification such as serial number, assign a laboratory
identification)
Temperature of the reagent water
Nominal or test volume
Weights of dispensed volumes
Procedure
1. Allow apparatus and reagent water to equilibrate to room temperature.
2. Check the calibration of the balance.
3. Wet the liquid-dispensing device and lubricate the piston insuring smooth operation by pipetting
several volumes to waste.
4. Weigh the vessel and tare the balance.
5. Dispense a minimum of five aliquots of water into the vessel, weighing and recording the mass
for each replicate.
6. Calculate the following:
m = mean mass = total of individual weight measurements ÷
total number of measurements
Z = Z correction factor obtained from Table 1 using the temperature of the test liquid and
current barometric pressure.
V = volume equivalent (or corrected volume) of mass measured = m x Z
RE = relative error (%) = [Vm - Vo] x 100
Vo
where:
Vm = corrected mean volume = total of corrected volumes ÷
total number of dispensed volumes
Vo = nominal or test volume
S = standard deviation = For each replicate, subtract the mean volume from the corrected
volume, then square the value. Sum the squared values of all replicates, and then divide the
sum by n-1 (the number of measurements minus 1). Take the square root of this value to get
the standard deviation.
CV = coefficient of variation (%) = s x 100
Vm
To simplify this process, spreadsheets for calculating and documenting the relative error and
coefficient of variation can be found on the NC Wastewater/Groundwater Laboratory Certification (NC
WW/GW LC) website at http://h2o.enr.state.nc.us/lab/download.htm.
The Relative Error (a measure of accuracy) and Coefficient of Variation (a measure of precision) must
not be greater than the manufacturer's specifications. If the results are greater than the manufacturer's
specification, repeat the procedure. If the results remain greater, follow the manufacturer's instructions
for maintenance and adjustment and repeat the procedure. If repeated calibration attempts do not yield
results within the manufacturer's specifications, the liquid-dispensing device must be repaired or
replaced.
Note: Microliter pipettors may need to be tested under more rigorously controlled conditions or sent to
a calibration service for testing. Since most analytical balances used in environmental laboratories read
to only four decimal places, it may be necessary, when checking microliter pipettors (less than 1000
μl), to include the weight of the weighing vessel in each replicate and subtract it and subsequent
aliquots out manually in order to have a measurable mass for the purpose of this verification. Contact
your auditor if you have questions.
Rev. 03/2009
Table 1. Z Correction Factor Table.
Temperat
ure C
Air pressure, kPa (mbar)
80
(800)
85 (850) 90 (900) 95 (950) 100
(1000)
101.3
(1013)
105
(1050)
15.0 1.0017 1.0018 1.0019 1.0019 1.0020 1.0020 1.0020
15.5 1.0018 1.0019 1.0019 1.0020 1.0020 1.0020 1.0021
16.0 1.0019 1.0020 1.0020 1.0021 1.0021 1.0021 1.0022
16.5 1.0020 1.0020 1.0021 1.0021 1.0022 1.0022 1.0022
17.0 1.0021 1.0021 1.0022 1.0022 1.0023 1.0023 1.0023
17.5 1.0022 1.0022 1.0023 1.0023 1.0024 1.0024 1.0024
18.0 1.0022 1.0023 1.0023 1.0024 1.0025 1.0025 1.0025
18.5 1.0023 1.0024 1.0024 1.0025 1.0025 1.0026 1.0026
19.0 1.0024 1.0025 1.0025 1.0026 1.0026 1.0027 1.0027
19.5 1.0025 1.0026 1.0026 1.0027 1.0027 1.0028 1.0028
20.0 1.0026 1.0027 1.0027 1.0028 1.0028 1.0029 1.0029
20.5 1.0027 1.0028 1.0028 1.0029 1.0029 1.0030 1.0030
21.0 1.0028 1.0029 1.0029 1.0030 1.0031 1.0031 1.0031
21.5 1.0030 1.0030 1.0031 1.0031 1.0032 1.0032 1.0032
22.0 1.0031 1.0031 1.0032 1.0032 1.0033 1.0033 1.0033
22.5 1.0032 1.0032 1.0033 1.0033 1.0034 1.0034 1.0034
23.0 1.0033 1.0033 1.0034 1.0034 1.0035 1.0035 1.0036
23.5 1.0034 1.0035 1.0035 1.0036 1.0036 1.0036 1.0037
24.0 1.0035 1.0036 1.0036 1.0037 1.0037 1.0038 1.0038
24.5 1.0037 1.0037 1.0038 1.0038 1.0039 1.0039 1.0039
25.0 1.0038 1.0038 1.0039 1.0039 1.0040 1.0040 1.0040
25.5 1.0039 1.0040 1.0040 1.0041 1.0041 1.0041 1.0042
26.0 1.0040 1.0041 1.0041 1.0042 1.0042 1.0043 1.0043
26.5 1.0042 1.0042 1.0043 1.0043 1.0044 1.0044 1.0044
27.0 1.0043 1.0044 1.0044 1.0045 1.0045 1.0045 1.0046
27.5 1.0045 1.0045 1.0046 1.0046 1.0047 1.0047 1.0047
28.0 1.0046 1.0046 1.0047 1.0047 1.0048 1.0048 1.0048
28.5 1.0047 1.0048 1.0048 1.0049 1.0049 1.0050 1.0050
29.0 1.0049 1.0049 1.0050 1.0050 1.0051 1.0051 1.0051
29.5 1.0050 1.0051 1.0051 1.0052 1.0052 1.0052 1.0053
30.0 1.0052 1.0052 1.0053 1.0053 1.0054 1.0054 1.0054
Z correction factors are for distilled water as a function of liquid test temperature and air
pressure.
Z values are in microliters per milligram.
Reprinted from ISO 8655-6:2002(E).
If your laboratory does not have a mercury barometer, the true local barometric pressure for locations throughout
North Carolina may be obtained from the North Carolina State University, State Climate Office of North Carolina
website at http://www.nc-climate.ncsu.edu/.