HomeMy WebLinkAbout#5390 12-Final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 5390
Laboratory Name: LTH3 Operations
Inspection Type: Field Maintenance
Inspector Name(s): Jeffrey R. Adams
Inspection Date: July 30, 2012
Date Report Completed: August 10, 2012
Date Forwarded to Reviewer: August 10, 2012
Reviewed by: Chet Whiting
Date Review Completed: August 13, 2012
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding Insp. CP __ Corrected
Unit Supervisor: Dana Satterwhite
Date Received: August 15, 2012
Date Forwarded to Linda: August 27, 2012
Date Mailed: August 27, 2012
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: LTH3 Operations
NPDES PERMIT #: NC0037869; NC0056731; NC0042803
ADDRESS: 4620 Reigal W ood Road
Durham, NC 27712
CERTIFICATE #: 5390
DATE OF INSPECTION: July 30, 2012
TYPE OF INSPECTION: Field Maintenance
AUDITOR(S): Jeffrey R. Adams
LOCAL PERSON(S) CONTACTED: Thomas Harden
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for
the analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was clean and records were well organized; however, some quality control procedures
need to be implemented. Proficiency Testing (PT) samples have not been analyzed for all certified
parameters for the 2012 calendar year. The laboratory obtained unacceptable PT results for Total
Residual Chlorine. A remedial PT is due by September 21, 2012.
The laboratory was given a packet containing North Carolina Laboratory Certification quality control
requirements and policies during the inspection.
The requirements associated with Findings A, B, C and F are new policies that have been implemented
by our program since the last inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Traceability
A. Finding: The laboratory needs to increase the documentation of purchased materials and
reagents.
Requirement: All chemicals, reagents, standards and consumables used by the laboratory
must have the following information documented: Date Received, Date Opened (in use),
Vendor, Lot Number, and Expiration Date. A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the
solutions are used. Documentation of solution preparation must include the analyst’s initials,
date of preparation, the volume or weight of standard(s) used, the solvent and final volume of
the solution. This information as well as the vendor and/or manufacturer, lot number, and
expiration date must be retained for chemicals, reagents, standards and consumables used for
a period of five years. Consumable materials such as pH buffers and lots of pre-made
Page 2
#5390 LTH3 Operations
standards are included in this requirement. Ref: Quality Assurance Policies for Field
Laboratories.
Comment: The laboratory had no documentation showing when pH buffers and Total Residual
Chlorine DPD reagents were received and opened (or put in use).
Proficiency Testing
B. Finding: The preparation of Proficiency Testing (PT) samples is not documented.
Requirement: PT samples received as ampules must be diluted according to the PT provider’s
instructions. The preparation of PT samples must be documented in a traceable log or other
traceable format. The diluted PT sample becomes a routine environmental sample and is added
to a routine sample batch for analysis. Ref: Proficiency Testing Requirements, February 20,
2012, Revision 1.2.
C. Finding: The laboratory is not documenting Proficiency Testing (PT) sample analyses in the
same manner as environmental samples (i.e., on a laboratory benchsheet or field notebook).
Requirement: All PT sample analyses must be recorded in the daily analysis records as for any
environmental sample. This serves as the permanent laboratory record. Ref: Proficiency Testing
Requirements, February 20, 2012, Revision 1.2.
Comment: There was no supporting documentation of the PT analyses. Only a copy of the
electronic data submittal to the PT vendor was retained.
Dissolved Oxygen – Standard Methods, 18th Edition, 4500 O G
D. Finding: The DO probe was being stored in a dirty storage well.
Requirement: A best effort must be made to perform analyses in a manner where possible
sources of contamination or error will not be introduced. Ref: Quality Assurance Policies for Field
Laboratories.
Comment: The instrument has a convenient calibration/storage chamber built into the instrument’s
side. This chamber provides an ideal storage area for the probe during transport and extended non-
use. If you look into the chamber, you will notice a small round sponge in the bottom. Carefully add
3 to 6 drops of clean water into the sponge. Turn the instrument over and allow any excess water to
drain out of the chamber. The wet sponge creates a 100% water saturated environment for the
probe which is ideal for dissolved oxygen calibration. To keep the probe from drying out, store the
probe in the calibration/storage chamber with the wet sponge. Ref: YSI Operators Manual Sections
3.4 and 4.5 (8). YSI 1/07.
Comment: If the membrane is coated with oxygen consuming (e.g., bacteria) or oxygen evolving
organisms (e.g., algae), erroneous readings may occur. Ref: YSI Operators Manual Section 4.5 (2).
YSI 1/07.
Recommendation: It is recommended that a light colored sponge be used in the storage chamber
to discern cleanliness.
Total Residual Chlorine – Standard Methods, 18th Edition, 4500 Cl G
Page 3
#5390 LTH3 Operations
Comment: The laboratory is using an instrument that is no longer in production, and the manufacturer is
no longer in business. This means that technical support is no longer offered. It is recommended that the
laboratory consider obtaining another instrument for analyzing Total Residual Chlorine of its effluent.
E. Finding: The annual calibration curve verification does not bracket the concentration of the
annual Proficiency Testing (PT) samples.
Requirement: The concentrations of the calibration standards must bracket the concentrations
of the samples analyzed. One of the standards must have a concentration equal to or below the
lower reporting concentration for Total Residual Chlorine. Ref: Approved Procedure for the Field
Analysis of Total Residual Chlorine.
Recommendation: It is recommended that the laboratory verify the internal calibration using
the concentrations: 15, 20, 50, 200 and 400 µg/L. This will verify the analytical range used to
measure Proficiency Testing (PT) samples as well as environmental samples.
Comment: The assigned value of the PT sample was 232 µg/L. The laboratory’s curve
consisted of the following concentrations: 15, 25, 50, 100 and 200 µg/L.
F. Finding: The laboratory is reporting the Total Residual Chlorine values on the DMR as < 50
µg/L.
Requirement: Facilities will still be required to report actual results on their monthly DMR
submittals, but for compliance purposes all TRC values below 50µg/L will be treated as zero.
For example, if the facility has a limit of 17 µg/L and reports a TRC value of 40 µg/L on the
DMR, this value will be considered complaint under this new policy. Ref: Total Residual
Chlorine Compliance Memo dated May 1, 2008 to NPDES Permittees from NC DWQ Point
Source Branch Supervisor, Matt Matthews. A copy of this memo is enclosed with this
report.
Comment: The instrument was verified to 15 µg/L so actual values will be reported down to
that level. Any values below that concentration will be reported as <15 µg/L.
Temperature – Standard Methods, 18th Edition, 2550 B
G. Finding: The temperature sensors on the pH and DO meters used to obtain reported
temperature values has not been checked every 12 months against a NIST thermometer.
Requirement: Thermometers and temperature measuring devices, used to measure
temperature for compliance monitoring, must be checked every 12 months against a NIST
traceable thermometer. The thermometer/meter readings must be less than or equal to 1ºC
from the NIST traceable thermometer reading. The documentation must include the serial
number of the NIST traceable thermometer that was used in the comparison. Document any
correction that applies on both the thermometer/meter and on a separate sheet to be filed. Ref:
Approved Procedure for Field Analysis of Temperature. Please submit a copy of the sensor
checks with the response to this report.
H. Finding: The laboratory’s National Institute of Standards and Technology (NIST) traceable
thermometer (H-B S/N #0559506) calibration expired on June 16, 2011.
Requirement: NIST traceable thermometers used to verify the calibration of other
thermometers or temperature sensors (i.e., limited use only) must be recalibrated in
Page 4
#5390 LTH3 Operations
accordance with the manufacturer’s recalibration date. If no recalibration date is given, the
NIST traceable thermometer must be recalibrated every 5 years. Ref: Approved Procedure for
Field Analysis of Temperature. Please submit documentation of the calibration verification
with the response to this report.
Comment: This thermometer is used to verify the calibration of the temperature sensors used
to report temperature for compliance monitoring.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing field testing records and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for LTH3 Operations for Arbor Hill MHP (NPDES permit #NC0037869); Grande Oaks
(NPDES permit NC0056731 and Birchwood MH & RV Park (NPDES #NC0042803) for March, April and
May, 2012. The following error was noted:
Grande Oaks WWTP NPDES #NC0056731
Date Parameter Location Value on Benchsheet
*Contract Data
Value on DMR
5/3/12 BOD Effluent * 2.5 mg/L < 2.5 mg/L
In order to avoid questions of legality, it is recommended that you contact the appropriate Regional
Office for guidance as to whether an amended Discharge Monitoring Report will be required. A copy of
this report will be made available to the Regional Office.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendations will help this lab to
produce quality data and meet certification requirements. The inspector would like to thank the staff for
its assistance during the inspection and data review process. Please respond to all findings.
Report prepared by: Jeffrey R. Adams Date: August 10, 2012
Report reviewed by: Chet Whiting Date: August 13, 2012