HomeMy WebLinkAbout#5274 12-Final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: #5274
Laboratory Name: Randolph County Board of Education
Inspection Type: Field Maintenance
Inspector Name(s): Jeffrey R. Adams
Inspection Date: October 30, 2012
Date Report Completed: November 1, 2012
Date Forwarded to Reviewer: November 1, 2012
Reviewed by: Chet Whiting
Date Review Completed: November 6, 2012
Cover Letter to use: Insp. Initial Insp. Reg.
Insp. No Finding Insp. CP
Corrected
Unit Supervisor: Dana Satterwhite
Date Received: November 6, 2012
Date Forwarded to Linda: November 16, 2012
Date Mailed: November 16, 2012
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Randolph County Board of Education
NPDES PERMIT #: NC0040908; NC0076287; NC0040924; NC0040975
ADDRESS: 2234-B Enterprise Street
Asheboro, NC 27205
CERTIFICATE #: 5274
DATE OF INSPECTION: October 30, 2012
TYPE OF INSPECTION: Field Maintenance
AUDITOR(S): Jeffrey R. Adams
LOCAL PERSON(S) CONTACTED: Larry Chilton and Darrell Loflin
I. INTRODUCTION:
This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater
Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A
NCAC 2H .0800 for the analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was clean and well organized. Records were filed in an orderly manner and readily
available for inspection. Proficiency Testing samples have been submitted for 2012 and were graded
acceptable.
The laboratory was given a packet containing North Carolina Laboratory Certification quality control
requirements and policies, including the Approved Procedure for Field Analysis of pH, the Approved
Procedure for Field Analysis of Dissolved Oxygen and the Approved Procedure for Field Analysis of
Temperature, during the inspection.
Contracted analyses are performed by Cameron Testing Services, Inc. (Certification #654).
The requirements associated with Findings A and B, are new policies that have been implemented by our
program since the last inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Traceability
A. Finding: The laboratory needs to increase the documentation of purchased materials and
reagents (i.e., pH buffers).
Requirement: All chemicals, reagents, standards and consumables used by the laboratory
must have the following information documented: Date Received, Date Opened (in use),
Vendor, Lot Number, and Expiration Date. A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the
solutions are used. Documentation of solution preparation must include the analyst’s initials,
Page 2
#5274 Randolph County Board of Education
date of preparation, the volume or weight of standard(s) used, the solvent and final volume of
the solution. This information as well as the vendor and/or manufacturer, lot number, and
expiration date must be retained for chemicals, reagents, standards and consumables used for
a period of five years. Consumable materials such as pH buffers and lots of pre-made
standards are included in this requirement. Ref: Quality Assurance Policies for Field
Laboratories.
Proficiency Testing
B. Finding: The laboratory is not documenting Proficiency Testing (PT) sample analyses in the
same manner as environmental samples (i.e., on a laboratory benchsheet).
Requirement: All PT sample analyses must be recorded in the daily analysis records as for any
environmental sample. This serves as the permanent laboratory record. Ref: Proficiency Testing
Requirements, February 20, 2012, Revision 1.2.
Comment: The laboratory only retained a copy of the data reporting form used to submit
proficiency testing results to the vendor.
Temperature – Standard Methods, 2550 B-2000
Comment: The temperature sensors on the pH and DO meters used to obtain reported temperature
values has not been checked every 12 months against a National Institute of Standards and
Technology (NIST) traceable thermometer. The Approved Procedure for Field Analysis of Temperature
states: Thermometers and temperature measuring devices, used to measure temperature for
compliance monitoring, must be checked every 12 months against a NIST traceable thermometer. The
thermometer/meter readings must be less than or equal to 1ºC from the NIST traceable thermometer
reading. The documentation must include the serial number of the NIST traceable thermometer that
was used in the comparison. Document any correction that applies on both the thermometer/meter and
on a separate sheet to be filed. The laboratory had the City of Asheboro perform the NIST temperature
sensor checks on both the DO and pH meters and a copies of the results were submitted to the auditor
on October 31, 2012. No further response is necessary for this finding.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing field testing records and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for Tabernacle Elementary School (NPDES permit #NC0040908), Farmer Elementary School
(NPDES permit #NC0076287), Seagrove Elementary School (NPDES permit #NC0040924) and
Coleridge Elementary School (NPDES permit #NC0040975) for March, April and May, 2012. The
following error was detected:
Farmer Elementary School
Date Parameter Location Value on Benchsheet
*Contract Data
Value on DMR
3/13/12 Ammonia Nitrogen Effluent *0.1 mg/L No Value
Page 3
#5274 Randolph County Board of Education
In order to avoid a possible monitoring frequency violation, it is recommended that you contact the
appropriate Regional Office for Guidance as to whether an amended Discharge Monitoring Report will
be required. A copy of this report will be forwarded to the Regional Office.
V. CONCLUSIONS:
Correcting the above-cited findings will help this lab to produce quality data and meet certification
requirements. The inspector would like to thank the staff for its assistance during the ins pection and
data review process. Please respond to all findings.
Report prepared by: Jeffrey R. Adams Date: November 1, 2012
Report reviewed by: Chet Whiting Date: November 6, 2012