5041
Town of Weldon WWTP
Field Maintenance
Anna Ostendorff, Gary Francies, and Todd Crawford
February 3, 2016
March 3, 2016
March 3, 2016
Tonia Springer
March 3, 2016
❑ Insp. Initial ❑ Insp. Reg.
❑ Insp. No Finding ❑ Insp. CP
® Corrected ❑ Insp. Reg. Delay
Gary Francies
March 3, 2016
March 4, 2016
March 4, 2016
Water Resources
ENVfRONMENTAL QUALITY
March 7, 2016
5041
Mr. Donald Crowder
Town of Weldon WWTP
PO Box 551
Weldon, NC 27890
PAT MCCRORY
DONALD R. VAN DER VAART
S. JAY ZIMMERMAN
SUBJECT: North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC)
Maintenance Inspection
Dear Mr. Crowder:
Enclosed is a report for the inspection performed on February 3, 2016 by Anna Ostendorff, Gary
Francies and Todd Crawford. Since the Finding(s) cited during the inspection were all corrected prior to
the completion of the enclosed report, a response is not required. The staff is commended for taking the
initiative in correcting the Findings in such a timely manner. For Certification maintenance, your
laboratory must continue to carry out the requirements set forth in 15A NCAC 2H .0800.
Copies of the checklists completed during the inspection may be requested from this office. Thank you
for your cooperation during the inspection. If you wish to obtain an electronic copy of this report by
email, or if you have questions or need additional information, please contact me at (828) 296-4677.
Sincerely,
Gary Francies, Technical Assistance/Compliance Specialist
Division of Water Resources
Attachment
cc: master file
Anna Ostendorff
On -Site Inspection Report
LABORATORY NAME:
NPDES PERMIT #:
ADDRESS:
CERTIFICATE M
DATE OF INSPECTION:
TYPE OF INSPECTION:
AUDITOR(S):
LOCAL PERSON(S) CONTACTED:
INTRODUCTION:
Town of Weldon WWTP
NCO025721
PO Box 551
Weldon, NC 27890
5041
February 3, 2016
Field Maintenance
Anna Ostendorff, Gary Francies, and Todd Crawford
Donald Crowder, Bill Hux, and Bobby Lyons
This laboratory was inspected by a representative of the North Carolina Wastewater/
Groundwater Laboratory Certification (NC WW/GW LC) program to verify its compliance
with the requirements of 15A NCAC 2H .0800 for the analysis of environmental samples.
GENERAL COMMENTS:
The facility has all the equipment necessary to perform the analyses. The system for
traceability of standard and reagent preparation is effective, thorough and easy to follow.
The Laboratory has done a very good job of keeping up with and implementing regulatory
changes since the last inspection.
Proficiency Testing (PT) samples have been analyzed for all certified parameters for the
2015 proficiency testing calendar year and the graded results were 100% acceptable.
Contracted analyses are performed by Meritech, Inc. (Certification #165).
Current Quality Assurance Policies for Field Laboratories and Approved Procedures for
the analysis of the facility's currently certified parameters were provided at the time of
the inspection.
FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Documentation
Recommendation: It is recommended that the ditto marks be eliminated from the
standard true value column of the Total Residual Chlorine (TRC) benchsheet. This could
be replaced with a down arrow to prevent possible issues with data defensibility and/or
transcription errors.
Comment: The laboratory benchsheet was lacking pertinent data: Instrument
identification. The NC WW/GW LC Approved Procedure for the Analysis of Total Residual
Chlorine, NC WW/GW LC Approved Procedure for the Analysis of pH, and NC WW/GW
LC Approved Procedure for the Analysis of Temperature documents state: The following
must be documented in indelible ink whenever sample analysis is performed: Instrument
Page 2
#5041 Town of Weldon
Identification. Acceptable corrective action (instrument ID was added to the TRC and
pH/Temperature benchsheet master copies) was performed by the laboratory and
observed by the inspector to be adequate during the inspection. No further response is
necessary for this Finding.
pH — Standard Methods, 4500 H+ B-2000
Comment: An Automatic Temperature Compensation (ATC) check on the pH electrode is
no longer required.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing original records (e.g., laboratory benchsheets,
logbooks, etc.) and contract lab reports to Discharge Monitoring Reports (DMRs)
submitted to the North Carolina Division of Water Resources. Data were reviewed for
February, June and October, 2015. The following errors were noted:
Date
Parameter
Location
Value on
Benchsheet
Value on DMR
2/6/15
pH
Effluent
6.08 S.U.
6.8 S.U.
6/3/15
Temperature
Influent
21.1 °C
22.1 °C
6/24/15
pH
Effluent
6.8 S.U.
6.5 S.U.
6/26/15
Temperature
Effluent
25.7°C
23.7°C
6/29/15
Temperature
Influent
24.5°C
25.1 °C
To avoid questions of legality, it is recommended that you contact the appropriate
Regional Office for guidance as to whether an amended DMR will be required. A copy
of this report will be made available to the Regional Office.
V. CONCLUSIONS:
All Findings noted during the inspection were adequately addressed prior to the
completion of this report. The inspector would like to thank the staff for its assistance
during the inspection and data review process. No response is required.
Report prepared by: Anna Ostendorff Date: March 3, 2016
Report reviewed by: Tonja Springer Date: March 3, 2016