Loading...
HomeMy WebLinkAbout#278_2016_0113_NJ_FINAL INSPECTION REPORT ROUTING SHEET To be attached to all inspection reports in-house only. Laboratory Cert. #: 278 Laboratory Name: South Cary WRF Inspection Type: Municipal Maintenance Inspector Name(s): Nick Jones, Dana Satterwhite, Gary Francies, Todd Crawford, Beth Swanson and Anna Ostendorff Inspection Date: January 13, 2016 Date Report Completed: February 11, 2016 Date Forwarded to Reviewer: February 11, 2016 Reviewed by: Beth Swanson Date Review Completed: 2/12/2016 Cover Letter to use: Insp. Initial Insp. Reg. Insp. No Finding Insp. CP Corrected Unit Supervisor/Chemist III: Todd Crawford Date Received: February 12, 2016 Date Forwarded to Linda: February 22, 2016 Date Mailed: February 23, 2016 _____________________________________________________________________ PAT MCCRORY Governor DONALD R. VAN DER VAART Secretary S. JAY ZIMMERMAN Director Water Sciences Section NC Wastewater/Groundwater Laboratory Certification Branch 1623 Mail Service Center, Raleigh, North Carolina 27699-1623 Location: 4405 Reedy Creek Road, Raleigh, North Carolina 27607 Phone: 919-733-3908 \ FAX: 919-733-6241 Internet: http://deq.nc.gov/about/divisions/water-resources/water-resources-data/water-sciences-home-page/laboratory-certification-branch February 23, 2016 278 Ms. Kelly Spainhour South Cary WRF 4900 West Lake Road Apex, NC 27502 Subject: North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) Maintenance Inspection Dear Ms. Spainhour: Enclosed is a report for the inspection performed on January 13, 2016 by Nick Jones, Dana Satterwhite, Gary Francies, Todd Crawford, Beth Swanson and Anna Ostendorff. Where Finding(s) are cited in this report, a response is required. Within thirty days of receipt, please supply this office with a written item for item description of how these Finding(s) were corrected. Please describe the steps taken to prevent recurrence and include an implementation date for each corrective action. If the Finding(s) cited in the enclosed report are not corrected, enforcement actions may be recommended. For Certification maintenance, your laboratory must continue to carry out the requirements set forth in 15A NCAC 2H .0800. Copies of the checklists completed during the inspection may be requested from this office. Thank you for your cooperation during the inspection. If you wish to obtain an electronic copy of this report by email or if you have questions or need additional information, please contact me at (919) 733-3908 ext. 251. Sincerely, Todd Crawford Technical Assistance & Compliance Specialist NC WW/GW Laboratory Certification Branch Attachment cc: Nick Jones; Master File On-Site Inspection Report LABORATORY NAME: South Cary WRF NPDES PERMIT #: NC0065102 ADDRESS: 4900 West Lake Road Apex, NC 27502 CERTIFICATE #: 278 DATE OF INSPECTION: January 13, 2016 TYPE OF INSPECTION: Municipal Maintenance AUDITOR(S): Nick Jones, Dana Satterwhite, Gary Francies, Todd Crawford, Beth Swanson and Anna Ostendorff LOCAL PERSON(S) CONTACTED: Kelly Spainhour and Stephen Avent I. INTRODUCTION: This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A NCAC 2H .0800 for the analysis of environmental samples. II. GENERAL COMMENTS: Ms. Kelly Spainhour, the laboratory Supervisor, requested this inspection as part of the laboratory’s internal quality assurance assessment. During the year leading up to the inspection, the lab worked with the Laboratory Certification Branch to revise its Standard Operating Procedures (SOPs). The laboratory has responded quickly to modify and update their operations to new requirements as they arise. The facility has all the equipment necessary to perform the analyses. The system for traceability of standard and reagent preparation is effective, thorough and easy to follow. Laboratory personnel communicate well with sample collectors and coordinate sample analyses effectively to manage workload and holding times. We would like to thank the entire staff for their professionalism. Proficiency Testing (PT) samples have been analyzed for all certified parameters for the 2015 proficiency testing calendar year and the graded results were 100% acceptable. The laboratory is reminded that all changes made in response to the Findings, Recommendations or Comments listed in this report must be incorporated in the Quality Page 2 #278 South Cary WRF Assurance (QA)/SOP document(s). The laboratory must ensure the method is being performed as stated, references to methods are accurate, and the QA and/or SOP document(s) is/are in agreement with approved practice and regulatory requirements. In some instances, the laboratory may need to create a SOP to document how new functions or policies will be implemented. Contracted analyses are performed by Environment 1, Inc. (Certification #10) and Environmental Testing Solutions, Inc. (Certification #600). Approved Procedures for the analysis of the facility’s currently certified Field parameters were provided to the laboratory before the inspection. Laboratory Fortified Matrix (LFM) and Laboratory Fortified Matrix Duplicate (LFMD) are also known a Matrix Spike (MS) and Matrix Spike Duplicate (MSD) and may be used interchangeably in this report. III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS: General Laboratory Recommendation: It is recommended that the laboratory implement a temperature grid check of all block digesters by alternating the well location of the thermometer each time samples are digested. This will document heating uniformity and consistency of all sample wells in the block. Comment: The sample preservation and storage temperature requirements may be changed in accordance with 40 CFR Part 136.3 Table II in all pertinent SOPs and in practice. The current practice is to store “at 4°C”. 40 CFR Part 136.3 Table II allows ≤6°C for most inorganic parameters and <10°C for bacterial tests, without evidence of freezing. Quality Control A. Finding: The laboratory only duplicates effluent samples. Requirement: Duplicate samples are analyzed randomly to assess precision on an ongoing basis. Ref: Standard Methods, 1020 B-2011. (8). B. Finding: The laboratory only spikes effluent samples. Requirement: To prepare an LFM, add a known concentration of analytes (ideally from a second source) to a randomly selected routine sample without increasing its volume by more than 5%. Ref: Standard Methods, 4020 B-2011. (2) (g). Recommendation: It is recommended that random selection of samples for spiking be implemented for metals analysis as well. C. Finding: Improper error correction was observed. Page 3 #278 South Cary WRF Requirement: All documentation errors must be corrected by drawing a single line through the error so that the original entry remains legible. Entries shall not be obliterated by erasures or markings. Wite-Out®, correction tape or similar products designed to obliterate documentation are not to be used. Write the correction adjacent to the error. The correction must be initialed by the responsible individual and the date of change documented. All data and log entries must be written in indelible ink. Pencil entries are not acceptable. Ref: NC WW/GW LC Policy. Comment: Writing over a number as a means of error correction was observed on the Biochemical Oxygen Demand (BOD) and Total Suspended Residue benchsheets. D. Finding: The laboratory is utilizing a thermometer with a correction factor of >1ºC from the NIST traceable thermometer. Requirement: The thermometer/meter readings must be less than or equal to 1ºC from the NIST certified or NIST traceable thermometer reading. Ref: NC WW/GW LC Policy. Comment: The thermometer in the Total Suspended Residue oven had a correction factor of -4°C posted on the thermometer. However, it was actually +4°C. So, when the oven temperature was adjusted by +4°C, it was adjusted the wrong direction. This caused the oven to be off by +8°C. Comment: The +4°C correction factor is most likely due to opening the door when taking the verification reading. The thermometer was immersed in vermiculite, not sand or oil, inside the oven, so the temperature fluctuates rapidly when the door opens. Recommendation: It is recommended that the laboratory use sand or oil to immerse the thermometer. Proficiency Testing E. Finding: The preparation of Proficiency Testing (PT) samples is not documented. Requirement: PT samples received as ampules must be diluted according to the PT provider’s instructions. The preparation of PT samples must be documented in a traceable log or other traceable format. The diluted PT sample becomes a routine environmental sample and is added to a routine sample batch for analysis. Ref: Proficiency Testing Requirements, February 20, 2012, Revision 1.2. Comment: Dating and initialing the instruction sheet for the preparation of the PT sample would satisfy the documentation requirement. F. Finding: The laboratory is not always analyzing Proficiency Testing (PT) samples in the same manner as environmental samples. Page 4 #278 South Cary WRF Requirement: All PT samples are to be analyzed and the results reported in a manner consistent with the routine analysis and reporting requirements of compliance samples and any other samples analyzed according to the requirements of 15A NCAC 2H .0800. Ref: Proficiency Testing Requirements, February 20, 2012, Revision 1.2. Comment: The laboratory occasionally analyzed PT samples in duplicate. The thought was that they would just use the PT sample as the duplicate for the analytical batch. However, duplicates are to be chosen randomly. Standard Operating Procedures Recommendation: It is recommended that the laboratory document the kPa to psi conversion factor in the Total Phosphorus SOP. Biochemical Oxygen Demand – Standard Methods, 5210 B-2001 Recommendation: It is recommended that the lab analyze a seeded blank. Comment: Data in the meter calibration area of the BOD benchsheet did not have units of measure documented for all values. The time that sample dilutions were completed was not documented. This is considered pertinent information. The North Carolina Administrative Code, 15A NCAC 2H .0805 (a) (7) (H) states: All laboratories must use printed laboratory bench worksheets that include a space to enter the signature or initials of the analyst, date of analyses, sample identification, volume of sample analyzed, value from the measurement system, factor and final value to be reported and each item must be recorded each time samples are analyzed. The date and time BOD and coliform samples are removed from the incubator must be included on the laboratory worksheet. Standard Methods 5210 B-2001. (5) (g) states: After preparing dilution, measure initial DO within 30 min. Documentation of the time that dilutions were completed and the time sample went into the incubator will bracket the initial DO reading show compliance with the method requirement. Notification of acceptable corrective action (i.e., a revised benchsheet which includes: units of measure and the time of sample dilution completion) was received by email on 2/17/2016 with an implementation date of 2/25/2016. No further response is necessary for this Finding. Comment: The laboratory was analyzing four glucose-glutamic acid (GGA) standards. The laboratory was averaging all four. However, two of them contained 1% seed and two contained 0.7%. There were not three of one kind available to average. Standard Methods 5210 B-2001. (6) (b) states: Add sufficient amounts of standard glucose- glutamic acid solution (¶ 3h) to give 3.0 mg glucose/L and 3.0 mg glutamic acid/L in each of three test bottles (20 mL GGA solution/L seeded dilution water or 6.0 mL/300- mL bottle). The resulting average BOD for the three bottles, after correction for dilution and seeding, must fall into the range of 198 ± 30.5 mg/L. Notification of acceptable corrective action (i.e., revised benchsheet including three GGA standards prepared the same) was received by email on 2/17/2016 with an implementation date of 2/17/2016. No further response is necessary for this Finding. Comment: The laboratory was not analyzing a duplicate which contained 1.0% seed, as the sample dilutions contained. A duplicate was analyzed, but contained only 0.7% Page 5 #278 South Cary WRF seed. The North Carolina Administrative Code, 15A NCAC 2H .0805 (a) (7) (C) states: Analyze five percent of all samples in duplicate to document precision. Laboratories analyzing less than 20 samples per month must analyze at least one duplicate each month samples are analyzed. Notification of acceptable corrective action (i.e., revised benchsheet including a duplicate sample seeded the same was received by email on 2/17/2016 with an implementation date of 2/17/2016. No further response is necessary for this Finding. Fecal, Coliform – Standard Methods, 9222 D-1997 Comment: The sample filtration time was not recorded on the benchsheet to show that no more than 30 minutes had passed before filters are placed into the incubator. This is considered pertinent information. The North Carolina Administrative Code, 15A NCAC 2H .0805 (a) (7) (A) states: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Standard Methods, 9222 D-1997. (2) (d) states: Place all prepared cultures in the water bath within 30 min after filtration. Notification of acceptable corrective action (i.e., revised benchsheets which include filtration time) was received by email on 1/21/2016 with an implementation date of 2/2/2016. No further response is necessary for this Finding. Comment: The laboratory was not monitoring the quality of the reagent water used in fecal coliform analysis. The NC WW/GW LC Policy states: Fecal coliform reagent water must be analyzed every twelve months to ensure suitability. At a minimum analyses must be performed for conductivity, TOC, Cd, Cr, Cu, Pb, and Zn. Recommended limits for reagent water quality are given in Standard Methods 9020 B-2005 Table 9020: II. If these limits are not met, investigate and correct or change water source. Notification of acceptable corrective action (i.e., a statement that the fecal coliform reagent water is being analyzed for conductivity by Dracor each time filter tanks are changed and TOC is being analyzed by Environment 1, Inc. on an annual basis) was received by email on 2/1/2016 and 2/17/2016 with implementation dates of 2/15/2016 and 2/17/2016, respectively. No further response is necessary for this Finding. G. Finding: Consumable materials are not properly tested. Requirement: At a minimum, make single analyses on five different water samples positive for the target organism or culture controls of known density. Ref: Standard Methods, 9020 B-2005. (5) (f) (1) (k) (2) (a). Requirement: NC WW/GW LC policy requires the testing of the following consumable materials before they can be used for sample analyses: membrane filters and/or pads (often packaged together) and media. Ref: NC WW/GW LC Policy. Please obtain a new lot (or preferably a proven lot from another laboratory) of media and perform the consumable testing on your current media and submit the results of this study with your response to this report. Recommendation: When performing comparisons of old and new lots of consumables, sample results are not consistently within the acceptable range of 20- 60 colonies. In order to more consistently meet this criterion, it is recommended that the sample be analyzed at multiple dilutions in order to establish the best volume to use, and then analyzed for comparison testing at that volume the next day. Page 6 #278 South Cary WRF Comment: Consumable testing was being conducted using only 3 positive samples. H. Finding: Plate comparison counts are not being conducted. Requirement: For routine performance evaluation, repeat counts on one or more positive samples at least monthly, record results, and compare the counts with those of other analysts testing the same samples. Replicate counts for the same analyst should agree within 5% (within analyst repeatability of counting) and those between analysts should agree within 10% (between analysts reproducibility of counting). If they do not agree, initiate investigation and any necessary corrective action. Ref: Standard Methods, 9020 B-2005. (9) (a). I. Finding: The laboratory does not have duplicate acceptance criteria for colony counts of <20 CFU/100 mL. Requirement: Each laboratory shall develop and maintain a document outlining the analytical quality control practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. The quality control document shall be available for inspection by the State Laboratory. Ref: 15A NCAC 2H .0805 (a) (7). Comment: A tiered scale for acceptance criteria is usually required. Low count acceptance criterion may be established using a colony count difference. J. Finding: The laboratory is using a partial immersion thermometer in the fecal water bath improperly. Requirement: Thermometers with no indicated depth are the total immersion type. When a partial-immersion thermometer is used, the bottom of the thermometer up to the immersion line should be exposed to the temperature being measured, with the remainder of the thermometer exposed to ambient conditions. When a total immersion thermometer is used, the bulb and the entire portion of the stem containing liquid, except for the last 1 cm, are exposed to the temperature being measured. If the thermometer is not used in this manner, the thermometer immersion is incorrect. Ref: “Selecting Alternatives to Mercury-Filled Thermometers”, http://www.epa.gov/mercury/alternatives.htm. Comment: The thermometer was touching the bottom of the bath and not immersed to the partial immersion line. Hardness, Total – Standard Methods, 2340 B-1997 K. Finding: The laboratory is referencing an unapproved method on the benchsheets. Requirement: Laboratory Procedures. Analytical methods, sample preservation, sample containers and sample holding times shall conform to those requirements found in 40 CFR-136.3. Ref: 15A NCAC 2H .0805 (a) (1). Page 7 #278 South Cary WRF Comment: The benchsheets references SM 18th ed 2340B. Dissolved, Oxygen – Standard Methods, 4500 O G-2001 L. Finding: The calibration cup has copious biological growth. Requirement: If the membrane is coated with oxygen consuming (e.g. bacteria) or oxygen producing organisms (e.g. algae), erroneous readings may occur. Ref: YSI 556 MPS Instrument Manual, Section 11.1.1.3. Chlorine, Total Residual – Standard Methods, 4500-Cl G-2000 Recommendation: It is recommended that the Total Residual Chlorine annual calibration sheet be clarified with regard to the blank type. The sheet lists only “blank”, when it is actually a reagent blank. The laboratory analysis is performed correctly. Residue, Suspended – Standard Methods, 2540 D-1997 Recommendation: It is recommended that an acceptance criterion for duplicates be split into two tiers. The method recommends that duplicates agree within 5% of their average weight. However, low concentration samples may need a separate low-level acceptance criterion which could be based upon calculated recoveries or a ± mg/L criterion. M. Finding: The daily drying time is not consistently greater than or equal to the time used for the initial verification study drying cycle. Requirement: Constant weights must be documented. The approved methods require the following: “Repeat the cycle of drying, cooling, desiccating, and weighing until a constant weight is obtained or until the weight change is less than 4% of the previous weight or 0.5 mg, whichever is less.” In lieu of this, an annual study documenting the time required to dry representative samples to a constant weight may be performed. Verify minimum daily drying time is greater than or equal to the time used for the initial verification study drying cycle. Drying cycles must be a minimum 1 hour for verification. Ref: NC WW/GW LC Policy. Comment: North Carolina allows for an annual drying study in lieu of the requirement above to repeat the drying cycle for every sample. A random full set of samples should be used for the drying study. The repeated drying time in the oven should be at least 1 hour long. The time used for the annual drying study is the minimum time that samples are to be dried until a new drying study is performed. Comment: The initial verification study was for 73 minutes. The daily drying time is typically 60 minutes. N. Finding: The benchsheet states the incorrect minimum residue yield. Requirement: Choose sample volume to yield between 2.5 and 200 mg dried residue. If volume filtered fails to meet minimum yield, increase sample volume up to 1 L. Ref: Standard Methods, 2540 D-1997 (3) (b). Page 8 #278 South Cary WRF Comment: The benchsheet states residue yield needs to be 1-200 mg. Residue, Total – Standard Methods, 2540 G-1997 O. Finding: The benchsheet references the incorrect method. Requirement: Laboratory Procedures. Analytical methods, sample preservation, sample containers and sample holding times shall conform to those requirements found in 40 CFR-136.3. Ref: 15A NCAC 2H .0805 (a) (1). Comment: The benchsheet references SM 2540 B-1997. The laboratory is not analyzing any aqueous Total Residue samples by SM 2540 B-1997 and only non- aqueous samples analyzed by SM 2540 G-1997 are documented on the sheet. P. Finding: The drying time study does not have a start time documented. This is considered pertinent data. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A) states: Requirement: Constant weights must be documented. The approved methods require the following: “Repeat the cycle of drying, cooling, desiccating, and weighing until a constant weight is obtained or until the weight change is less than 4% of the previous weight or 0.5 mg, whichever is less.” In lieu of this, an annual study documenting the time required to dry representative samples to a constant weight may be performed. Verify minimum daily drying time is greater than or equal to the time used for the initial verification study drying cycle. Drying cycles must be a minimum 1 hour for verification. Ref: NC WW/GW LC Policy. Comment The most recent annual drying time study was conducted on October 6, 2015. Phosphorus, Total – Standard Methods, 4500 P E-1999 Nitrate + Nitrite, Nitrogen – Standard Methods, 4500 NO 3 - E-2000 Recommendation: It is recommended that the laboratory not use ± 20% across the board as the acceptance criterion for back calculated calibration standards. Standard Methods states that back-calculated and true concentrations should agree within ±10%, unless different criteria are specified in an individual method. At the lower limit of the operational range, acceptance criteria are usually wider. Such criteria must be defined in the laboratory’s QA plan. Ammonia, Nitrogen – Standard Methods, 4500 NH 3 D-1997 Recommendation: It is recommended that the benchsheet be rearranged to clearly reflect the sequential order of the analysis. The benchsheet should accurately demonstrate that the sample results are appropriately bracketed by proper QC. Page 9 #278 South Cary WRF Recommendation: It is recommended that the laboratory monitor and calculate matrix spike recovery even when the parent sample has a value which is less than the reporting limit. Assume there is no contribution for the sample when calculating the spike % recovery. Nitrogen, Total Kjeldahl – Standard Methods, 4500 Norg C-1997 (Standard Methods, 4500 NH3 D-1997) Q. Finding: A calibration blank is not analyzed at the end of the sample group. Requirement: Verify calibration by periodically analyzing a calibration standard and calibration blank during a run—typically, after each batch of ten samples and at the end of the run. Ref: Standard Methods, 4020 B-2011. (2) (b). Requirement: The calibration blank and calibration verification standard (mid- range) must be analyzed initially (i.e., prior to sample analysis), after every tenth sample and at the end of each sample group to check for carry over and calibration drift. If either fall outside established quality control acceptance criteria, corrective action must be taken (e.g., repeating sample determinations since the last acceptable calibration verification, repeating the initial calibration, etc.). Ref: NC WW/GW LC Policy. R. Finding: Data that does not meet quality control requirements is not qualified on the Discharge Monitoring Report. Requirement: When quality control (QC) failures occur, the laboratory must attempt to determine the source of the problem and must apply corrective action. If data qualifiers are used to qualify samples not meeting QC requirements, the data may not be useable for the intended purposes. A notation must be made on the Discharge Monitoring Report (DMR) form, in the comment section or on a separate sheet attached to the DMR form, when any required sample quality control does not meet specified criteria, and another sample cannot be obtained. Ref: NC WW/GW LC Policy. Comment: The laboratory was not consistently transcribing data qualifiers from the contract laboratory reports to the DMR. The following error was noted: On 10/19/2015 the MS/MSD failed established QC criteria. The sample was reported on the DMR without qualification of QC failure. The same MS/MSD sample was reanalyzed on 10/22/2015 and passed established QC criteria. However, on 10/22/2015 the MS/MSD was analyzed alone without any other QC or the unspiked environmental sample. This second analysis of the MS/MSD would not be indicative of, or applicable to, the analytical conditions on 10/19/2015. Ammonia, Nitrogen – Standard Methods, 4500 NH 3 D-1997 Nitrogen, Total Kjeldahl – Standard Methods, 4500 Norg C-1997 (Standard Methods, 4500 NH3 D-1997) Comment: The laboratory uses two drops of 10N NaOH to adjust sample/ standard pH to >11. If a different volume would need to be added to samples vs. standards, then the Page 10 #278 South Cary WRF laboratory would need to use measured volumes in mL or determine the mL equivalent of a drop in order to adjust the concentration per the calculation in section 5 of the method. Ammonia, Nitrogen – Standard Methods, 4500 NH 3 D-1997 Nitrate + Nitrite, Nitrogen – Standard Methods, 4500 NO 3 - E-2000 Nitrogen, Total Kjeldahl – Standard Methods, 4500 Norg C-1997 (Standard Methods, 4500 NH3 D-1997) Phosphorus, Total – Standard Methods, 4500 P E-1999 Comment: The matrix spike stock standard concentration was not documented on the Spike Sample Identification/Description Document. This is considered pertinent information. The North Carolina Administrative Code, 15A NCAC 2H .0805 (a) (7) (A) states: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Notification of acceptable corrective action (i.e., a revised Spike Sample Identification/Description Document which includes the matrix spike stock standard concentration) was received by email on 1/22/2016 with an implementation date of 1/25/2016. No further response is necessary for this Finding. S. Finding: The laboratory is not rotating the range of the spike concentrations. Requirement: Also, rotate the range of spike concentrations to verify performance at various levels. Ref: Standard Methods, 4020 B-2011. (2) (g). Metals – EPA Method, 200.8, Rev. 5.4, 1994 Recommendation: It is recommended that the laboratory begin analyzing bottle blanks since sample bottles are cleaned and reused without an acid rinse. Comment: Metals samples were not always preserved at <2 s.u. for a minimum of 24 hours before digestion was performed. The Code of Federal Regulations, Title 40, Part 136, Vol. 77, No. 97, May 18, 2012, Table II, footnote 19 states: An aqueous sample may be collected and shipped without acid preservation. However, acid must be added at least 24 hours before analysis to dissolve any metals that adsorb to the container walls. Notification of acceptable corrective action (i.e., a statement that metals samples will sit preserved for at least 24 hours before digestion) was received by email on 1/22//2016 with an implementation date of 2/2/2016. No further response is necessary for this Finding. T. Finding: The laboratory is conducting sample digestion at 95°C. Requirement: The hot plate should be located in a fume hood and previously adjusted to provide evaporation at a temperature of approximately but no higher than 85°C. Ref: EPA Method 200.8, Rev. 5.4, (1994), Section 11.2.3. U. Finding: A calibration blank is not analyzed at the end of the sample group. Requirement: The calibration blank and calibration verification standard (mid- range) must be analyzed initially (i.e., prior to sample analysis), after every tenth sample and at the end of each sample group to check for carry over and calibration drift. If either fall outside established quality control acceptance criteria, Page 11 #278 South Cary WRF corrective action must be taken (e.g., repeating sample determinations since the last acceptable calibration verification, repeating the initial calibration, etc.). Ref: NC WW/GW LC Policy. Comment: The analyst is setting up a calibration blank at the end of each sample group. The instrument is not recognizing it as an actual blank to be qualitatively and quantitatively analyzed. It is only recognizing it as an instrument blank for cleaning purposes. Recommendation: It is recommended that the final calibration blank be identified in the software as another environmental sample. Then treat it as a blank for analysis and QC evaluation. IV. PAPER TRAIL INVESTIGATION: The paper trail consisted of comparing original records (e.g., laboratory benchsheets, logbooks, etc.) and contract laboratory reports to Discharge Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Resources. Data were reviewed for South Cary WRF (NPDES permit # NC0065102) for October, November and December, 2015. No transcription errors were detected. The facility appears to be doing a good job of accurately transcribing data. V. CONCLUSIONS: Correcting the above-cited Findings and implementing the Recommendations will help this laboratory to produce quality data and meet certification requirements. The inspector would like to thank the staff for its assistance during the inspection and data review process. Please respond to all Findings and include an implementation date for each corrective action. Report prepared by: Nick Jones Date: February 11, 2016 Report reviewed by: Beth Swanson Date: February 11, 2016