Loading...
HomeMy WebLinkAbout#5364_2017_0928_AO_FINALINSPECTION REPORT ROUTING SHEET To be attached to all inspection reports in-house only. Laboratory Cert. #: 5364 Laboratory Name: Town of Severn WWTP Inspection Type: Field Municipal Maintenance Inspector Name(s): Anna Ostendorff Inspection Date: September 28, 2017 Date Forwarded for Initial November 2, 2017 Review: Initial Review by: Jason Smith Date Initial Review November 2, 2017 Completed: ❑ Insp. Initial ❑ Insp. Reg Cover Letter to use: ❑Insp. No Finding ❑Corrected ❑Insp. CP ®Insp. Reg. Delay (to use: rt click, properties, check) Unit Supervisor/Chemist III: Dana Satterwhite Date Received: November 3, 2017 Date Forwarded to Admin.: November 14, 2017 Date Mailed: November 15, 2017 Special Mailing Instructions: Anna — send copy to RRO regarding Finding G. W aterResources ENVIRONMENTAL. QUALITY November 15, 2017 5364 Mr. M.E. Lassiter Town of Severn WWTP P.O. Box 401 Severn, NC 27877 ROY COOPER MICHEAL S. REGAN LINDA CULPEPPER Subject: North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) Maintenance Inspection Dear Mr. Lassiter: 11Ni11'iw ;>f i'h' 10" Enclosed is a report for the inspection performed on September 28, 2017 by Anna Ostendorff. I apologize for the delay in getting this report to you, Where Finding(s) are cited in this report, a response is required. Within thirty days of receipt, please supply this office with a written item for item description of how these Finding(s) were corrected. Please describe the steps taken to prevent recurrence and include an implementation date for each corrective action. If the Finding(s) cited in the enclosed report are not corrected, enforcement actions may be recommended. For certification maintenance, your laboratory must continue to carry out the requirements set forth in 15A NCAC 2H .0800. A copy of the laboratory's Certified Parameter List at the time of the audit is attached. This list will reflect any changes made during the audit. Copies of the checklists completed during the inspection may be requested from this office. Thank you for your cooperation during the inspection. If you wish to obtain an electronic copy of this report by email or if you have questions or need additional information, please contact me at (828) 296-4677. Sincerely, Dana Satterwhite, Environmental Program Supervisor Division of Water Resources Attachment cc: Anna Ostendorff, Master File #5364 Water Sciences Section NC WastewateriGroundwater Laboratory Certification Branch 1623 Mail Service Center, Raleigh, North Carolina 27699-1623 Locaticn: 4405 Reedy Creek Road, Raleigh, North Carolina 27607 Phone: 919-733-390V FAX: 919-733-6241 Internet: http://deQ.nc.gov/aboutid!visions/water-resourcestwater-resources•data/water-sciences-home- age/labo€atory.certification-branch LABORATORY NAME: NPDES PERMIT #: ADDRESS: CERTIFICATE #: DATE OF INSPECTION: TYPE OF INSPECTION: AUDITOR(S): LOCAL PERSON(S) CONTACTED: I. INTRODUCTION: Town of Severn WWTP WQ0001868 314 Main Street Severn, NC 27877 5364 September 28, 2017 Field Municipal Maintenance Anna Ostendorff Marshall Lassiter and Troy Holloman This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A NCAC 2H .0800 for the analysis of environmental samples. II. GENERAL COMMENTS: Staff were forthcoming and were receptive to adopting the necessary changes in response to the inspection. The laboratory regularly collaborates with neighboring municipalities on issues pertaining to laboratory certification and wastewater testing and analysis to the benefit of everyone. All required Proficiency Testing (PT) Samples have been analyzed for the 2017 PT Calendar Year and the graded results were 100% acceptable. Requirements that reference 15A NCAC 2H .0805 (g) (1), stating "Data pertinent to each analysis must be maintained for five years. Certified Data must consist of date collected, time collected, sample site, sample collector, and sample analysis time. The field benchsheets must provide a space for the signature or initials of the analyst and proper units of measure for all analyses", are intended to be a requirement to document all pertinent information for historical reconstruction of data. It is not intended to imply that existing records are not adequately maintained unless the Finding speaks directly to that. Contracted analyses are performed by Environment 1, Inc. (Certification # 10). Quality Assurance Policies for Field Laboratories and Approved Procedure documents for the analysis of the facility's currently certified Field Parameters were provided at the time of the inspection. III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS: Page 2 #5364 Town of Severn VWVfP Proficiency Testing A. Finding: The laboratory is not documenting PT Sample analyses in the same manner as routine Compliance Samples. Requirement: As specified in 15 NCAC 2H .0800, in order to meet the minimum standards for Certification, laboratories must use acceptable analytical methods. The acceptable methods are those defined or referenced in the current State and federal regulations for the environmental matrix being tested. All samples, (including PT Samples) that are, or that may, be used for Certification purposes, must be analyzed using approved methods only. All PT Samples are to be analyzed and the results reported in a manner consistent with the routine analysis and reporting requirements of Compliance Samples. Laboratories must document any exceptions. All PT Sample analyses must be recorded in the daily analysis records as for any Compliance Sample. This serves as the permanent laboratory record. Ref: Proficiency Testing Requirements, May 31, 2017, Revision 2.0. Comment: The analysis of PT Samples is designed to evaluate the entire process used to routinely report environmental analytical results; therefore, PT Samples must be analyzed and the process documented in the same manner as Compliance Samples. Documentation B. Finding: The laboratory needs to increase the traceability documentation of purchased materials and reagents. Requirement: All chemicals, reagents, standards and consumables used by the laboratory must have the following information documented: Date Received, Date Opened (in use), Vendor, Lot Number, and Expiration Date. This information must be retained for chemicals, reagents, standards and consumables used for a period of five years. Consumable materials such as pH buffers and lots of pre -made standards are included in this requirement. Ref: Quality Assurance Policies for Field Laboratories. Requirement: Data pertinent to each analysis must be maintained for five years. Ref: 15A NCAC 2H .0805 (g) (1). Comment: An example traceability log was provided by the auditor during the inspection and may be used to fulfill this requirement. C. Finding: The analyst is not signing or initialing laboratory benchsheets indicating who collected and analyzed the sample. Requirement: Data pertinent to each analysis must be maintained for five years. Certified data must consist of date collected, time collected, samples site, sample collector, and sample analysis time. The field bench sheets must provide a space for the signature of the analyst, and proper units of measure for all analyses. Ref: 15A NCAC 2H .0805 (g) (1). Requirement: The following must be documented in indelible ink whenever sample analysis is performed: Collector's/analyst's name or initials. Ref: NC WW/GW LC Approved Procedure for the Analysis of pH and NC WW/GW LC Approved Procedure for the Analysis of Temperature. Page 3 #5364 Town of Severn WWTP D. Finding: The laboratory benchsheet was lacking pertinent data: Instrument identification. This is considered pertinent data. Requirement: Data pertinent to each analysis must be maintained for five years. Certified data must consist of date collected, time collected, samples site, sample collector, and sample analysis time. The field bench sheets must provide a space for the signature of the analyst, and proper units of measure for all analyses. Ref: 15A NCAC 2H .0805 (g) (1). Requirement: The following must be documented in indelible ink whenever sample analysis is performed: Instrument identification. Ref: NC WW/GW LC Approved Procedure for the Analysis of pH and NC WW/GW LC Approved Procedure for the Analysis of Temperature. E. Finding: The units of measure for pH (i.e., Standard Units or S.U.) are not documented on the benchsheets. Requirement: Data pertinent to each analysis must be maintained for five years. Certified data must consist of date collected, time collected, samples site, sample collector, and sample analysis time. The field bench sheets must provide a space for the signature of the analyst, and proper units of measure for all analyses. Ref: 15A NCAC 2H .0805 (g) (1). F. Finding: The pH calibration check standard was not recorded to one decimal point. Requirement: Use a pH meter accurate and reproducible to 0.1 pH unit (as demonstrated daily by acceptable performance of a check standard buffer) with a range of 0 to 14 and equipped with temperature -compensation adjustment. The meter must be calibrated with at least two buffers. In addition to the calibration buffers, the meter calibration must be verified with a third standard buffer solution. The calibration and check standard buffers must bracket the range of the samples being analyzed. The check standard buffer must read within ±0.1 S.U. to be acceptable. If the meter verification does not read within ±0.1 S.U., the meter must be recalibrated before any samples are analyzed. Ref: NC WW/GW LC Approved Procedure for the Analysis of pH. Comment: The analyst travels to several monitoring wells and performs the analysis on site. The pH meter calibration is checked with a 7.0 S.U. buffer after each analysis and before traveling to the next location. The documented value of the check standard does not include any decimal values and is recorded as 7. The analyst is aware the check standard must measure within 0.1 S.U. of the true value. However, it cannot be determined from the records if the check buffer measured within the required range of 6.9 to 7.1 S.U. pH — Standard Methods, 4500 H+ B-2000 (Aqueous) G. Finding: The laboratory was reporting non -compliant data for pH on the Non -Discharge Monitoring Report (NDMR). Requirement: Analyze within 15 minutes. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 82, No. 165, August 28, 2017; Table 11. Page 4 #5364 Town of Severn WWTP Comment: The laboratory was analyzing pH on -site, but reporting the pH value provided by the contract laboratory on the NDMR, which has a statement that the value is "not to be used for reporting". The hold time for pH was not met for the analyses performed by the contract laboratory. The laboratory must use the value from the on -site analysis for reporting purposes. Recommendation: To avoid questions of legality, it is recommended that you contact the appropriate Regional Office for guidance as to whether an amended eDMR(s)/ DMR(s) will be required. A copy of this report will be made available to the Regional Office. IV. PAPER TRAIL INVESTIGATION:. The paper trail consisted of comparing original records (e.g., laboratory benchsheets, logbooks, etc.) and contract lab reports to NDMRs submitted to the North Carolina Division of Water Resources. Data were reviewed for the Town of Severn WWTP (Water Quality permit # WQ0001868) for November and August 2016 and February 2017. No transcription errors were observed. The facility appears to be doing a good job of accurately transcribing data. V. CONCLUSIONS: Correcting the above -cited Findings will help this laboratory to produce quality data and meet Certification requirements. The inspector would like to thank the staff for their assistance during the inspection and data review process. Please respond to all Findings and include supporting documentation, implementation dates and steps taken to prevent recurrence for each corrective action. Report prepared by: Anna Ostendorff Date: November 2, 2017 Report reviewed by: Jason Smith Date: November 2, 2017 E a E m Q Z m p m W c m J ��- m 0 U A=CL a) U W W p N�N LL ui H W Q a 0 w_ LL w W U d c a`) () w O 0 0 O O N W m d' U = � Z O Qt' v W Z a fn w 0 O a) a a O O O N m O Lo LO N