January 25, 2022
5353
Mr. Neil Carroll
Town of Roseboro
P.O. Box 848
Roseboro, NC 28382-
Subject: North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC
WW/GW LCB) Maintenance Inspection
Dear Mr. Carroll:
Enclosed is a report for the inspection performed on November 22, 2021 by Tonja Springer. I
apologize for the delay in getting this report to you. Where Finding(s) are cited in this report, a
response is required. Within thirty days, please supply this office with a written item for item
description of how these Finding(s) were corrected. Please describe the steps taken to prevent
recurrence and include an implementation date for each corrective action. If the Finding(s) cited
in the enclosed report are not corrected, enforcement actions may be recommended. For
Certification maintenance, your laboratory must continue to carry out the requirements set forth
in 15A NCAC 02H .0800.
A copy of the laboratory’s Certified Parameter List at the time of the audit is attached. This list will
reflect any changes made during the audit. Copies of the checklists completed during the
inspection may be requested from this office. Thank you for your cooperation during the
inspection. If you have questions or need additional information, please contact me at (919) 733-
3908 Ext. 259.
Sincerely,
Beth Swanson
Technical Assistance & Compliance Specialist
Division of Water Resources
Attachment
cc: Tonja Springer, Todd Crawford, #5353
On-Site Inspection Report
LABORATORY NAME: Town of Roseboro
NPDES PERMIT #: NC0026816
ADDRESS: 10160 Roseboro Hwy.
Roseboro, NC 28382
CERTIFICATE #: 5353
DATE OF INSPECTION: November 22, 2021
TYPE OF INSPECTION: Field Municipal Maintenance
AUDITOR(S): Tonja Springer
LOCAL PERSON(S) CONTACTED:
Neil Carrol and John Puglisi
I. INTRODUCTION:
This laboratory was inspected by representatives of the North Carolina Wastewater/Groundwater Laboratory
Certification Branch (NC WW/GW LCB) to verify its compliance with the requirements of 15A NCAC 02H
.0800 for the analysis of compliance monitoring samples.
II. GENERAL COMMENTS:
The facility is neat and well organized and has all the equipment necessary to perform the analyses. Staff
were forthcoming and responded well to suggestions from the auditor.
The laboratory does not have Quality Assurance (QA) and/or Standard Operating Procedure (SOP)
document(s) in place for currently certified parameter methods. These documents must be submitted for
review as specified in Finding A.
The laboratory is reminded that any time changes are made to laboratory procedures, the laboratory must
update the QA/SOP document(s) and inform relevant staff. Any changes made in response to the pre-
audit review or to Findings, Recommendations or Comments listed in this report must be incorporated to
ensure the method is being performed as stated, references to methods are accurate, and the QA and/or
SOP document(s) is in agreement with each approved practice, test, analysis, measurement, monitoring
procedure or regulatory requirement being used in the laboratory. In some instances, the laboratory may
need to create an SOP to document how new functions or policies will be implemented.
The laboratory is also reminded that SOPs are required to be reviewed at least every two years and are
intended to describe procedures exactly as they are to be performed. Use of the word “should” is not
appropriate when describing requirements (e.g., Quality Control (QC) frequency, acceptance criteria, etc.).
Evaluate all SOPs for the proper use of the word “should”.
Contracted analyses are performed by Environmental Chemists, Inc. (EnviroChem) (Certification # 94).
Page 2
#5353 Town of Roseboro
Approved Procedure documents for the analysis of the facility’s currently certified Field Parameters were
provided at the time of the inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
General Laboratory
A. Finding: SOPs have not been developed for all the methods included on the laboratory’s
Certified Parameters Listing (CPL).
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this Rule. A
copy of each analytical method or Approved Procedure and Standard Operating Procedure
shall be available to each analyst and available for review upon request by the State
Laboratory. Standard Operating Procedure documentation shall state the effective date of the
document and shall be reviewed every two years and updated if changes in procedures are
made. Each laboratory shall have a formal process to track and document review dates and
any revisions made in all Standard Operating Procedure documents. Supporting Records
shall be maintained as evidence that these practices are implemented. Ref: 15A NCAC 02H
.0805 (g) (4).
Comment: An SOP describes the method in such detail that an experienced analyst
unfamiliar with the method can obtain acceptable results and meet documentation
requirements. SOPs must be developed which describe in detail how a method is performed
in that particular laboratory. Templates are available on our website for all Field Parameters
to aid in constructing SOPs. The Dissolved Oxygen (DO) and Temperature SOPs must be
submitted for review by May 31, 2022.
Comment: The SOP for DO only includes directions for calibrating the meter. The laboratory
does not have an SOP for Temperature.
Documentation
B. Finding: Error corrections are not always properly performed.
Requirement: All documentation errors shall be corrected by drawing a single line through
the error so that the original entry remains legible. Entries shall not be obliterated by erasures
or markings. Wite-Out®, correction tape, or similar products designed to obliterate
documentation are not to be used; instead the correction shall be written adjacent to the error.
The correction shall be initialed by the responsible individual and the date of change
documented. Ref: 15A NCAC 02H .0805 (g) (1).
Comment: Some instances of overwriting were observed on the following dates: October 7,
2020, November 13, 2020, December 2, 2020, December 30, 2020 and January 20, 2021.
C. Finding: Documentation of the calibration variables for the DO meter does not include all
pertinent data.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this Rule. Ref:
15A NCAC 02H .0805 (g) (4).
Requirement: The following must be documented in indelible ink whenever sample analysis
is performed: Calibration variables (temperature, elevation or barometric pressure [in mmHg],
Page 3
#5353 Town of Roseboro
and salinity). Ref: NC WW/GW LCB Approved Procedure for the Analysis of Dissolved
Oxygen (DO).
Comment: Only Temperature is being documented for the calibration. The benchsheet has
a column labeled “Adjusted Air Calibration” in which Temperature is being documented. The
other calibration variables (e.g., salinity, elevation or barometric pressure) must also be
documented. These variables may have been programmed in the meter initially.
Comment: Per NC WW/GW LC Branch policy, facilities may use a default value of zero
ppt for Salinity when calibrating the DO meter unless it is known or suspected that the
Salinity value of the samples being analyzed is > 9 ppt. In those situations, actual Salinity
values must be used. Regardless of which value is used, it must be documented.
D. Finding: Documentation of the compliance temperature-measuring device calibration
verification does not include the stated accuracy of the Reference Temperature-Measuring
Device used for the comparison.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this Rule. Ref:
15A NCAC 02H .0805 (g) (4).
Requirement: All compliance temperature-measuring devices without a valid NIST certificate
must be checked initially and every 12 months against an NIST traceable temperature-
measuring device and the process documented. Documentation must include the serial
number of the device being checked. The serial number, stated accuracy and expiration date
of the NIST traceable temperature-measuring device used in the comparison must also be
documented. Verification data must be kept on file and be available for inspection for 5 years.
(NOTE: Vendors or other Certified laboratories may provide assistance in meeting this
requirement. When a vendor or other Certified laboratory provides this assistance, they must
provide a copy of their NIST Certificate or the serial number, accuracy and calibration
expiration date.) Ref: NC WW/GW LCB Approved Procedure for the Analysis of Temperature.
Requirement: All NIST traceable temperature-measuring devices must have a stated
accuracy of at least ± 0.5°C and be within their expiration date. Ref: NC WW/GW LCB
Approved Procedure for the Analysis of Temperature.
Comment: This Finding applies to the verification performed by the Environmental Chemists,
Inc. (EnviroChem) (Certification # 94).
E. Finding: The laboratory benchsheet is lacking required documentation: the method or
Standard Operating Procedure reference.
Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall
be traceable to the associated sample analyses and shall consist of: the method or Standard
Operating Procedure. Each item shall be recorded each time sample are analyzed. Analyses
shall conform to methodologies found in Subparagraph (a)(1) of this Rule. Ref: 15A NCAC
02H .0805 (g) (2) (A).
Comment: The method reference is not documented on the benchsheet for Temperature.
Quality Control
Page 4
#5353 Town of Roseboro
F. Finding: Temperature sensor check readings for the YSI Model 550 A DO meter used to
obtain compliance Temperature values varied more than 0.5°C from the National Institute of
Standards and Technology (NIST) traceable thermometer readings.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this Rule. Ref:
15A NCAC 02H .0805 (g) (4).
Requirement: To check a compliance temperature-measuring device, compare readings at
two temperatures that bracket the range of compliance samples routinely analyzed against
an NIST traceable temperature-measuring device and record all four readings. The readings
from both devices must agree within 0.5ºC. If they do not, the device may not be used for
temperature compliance monitoring. Ref: NC WW/GW LCB Approved Procedure for the
Analysis of Temperature. Please submit a copy of the temperature-measuring device
verification with the response to this report.
Dissolved Oxygen – Standard Methods, 4500 O G-2016 (Aqueous)
G. Finding: The meter is not being calibrated according to the manufacturer’s instructions.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this Rule. Ref:
15A NCAC 02H .0805 (g) (4).
Requirement: Instruments are to be calibrated according to the manufacturer’s
calibration procedure prior to analysis of samples each day compliance monitoring is
performed. Ref: NC WW/GW LCB Approved Procedure for the Analysis of Dissolved
Oxygen (DO).
Requirement: Calibration In % Saturation mode 1. Ensure that the sponge inside the
instrument’s calibration chamber is moist. Insert the probe into the calibration chamber. 2.
Power the instrument on and allow readings to stabilize. This may take 5 to 15 minutes,
depending on the age of the instrument and condition of the probe. 3. Press and release both
the UP ARROW and DOWN ARROW keys at the same time to enter the calibration menu. 4.
Press the Mode key until “%” is displayed on the right side of the screen for oxygen units.
Press ENTER. 5. The LCD will prompt you to enter the local altitude in hundreds of feet. Use
the arrow keys to increase or decrease the altitude. When the proper altitude appears on the
LCD, press the ENTER key. EXAMPLE: Entering the number 12 here indicates 1200 feet. 6.
CAL will now display in the lower left corner of the screen, the calibration value in the lower
right corner and the current DO reading (before calibration) will be the main display. Once the
current DO reading is stable, press the ENTER button. 7. The LCD will prompt you to enter
the approximately salinity of the water you are about to analyze. You can enter any number
from 0 to 70 parts per thousand (PPT) of salinity. Use the arrow keys to increase or decrease
the salinity setting. When the correct salinity appears on the LCD press ENTER key. The
instrument will return to normal operation. Ref: YSI 550A Operator’s Manual.
Comment: Per NC WW/GW LC Branch policy, facilities may use a default value of zero
ppt for Salinity when calibrating the DO meter unless it is known or suspected that the
Salinity value of the samples being analyzed is > 9 ppt. In those situations, actual
Salinity values must be used. Regardless of which value is used, it must be documented.
Comment: For calibration in % Saturation mode, the approximate altitude (in feet) of the
region where you are located is required. This information can be obtained over the internet
or from a local airport or weather station. To convert from meters to feet, divide by 0.3048.
Page 5
#5353 Town of Roseboro
Comment: Switch the meter back to mg/L after the % Saturation calibration.
Comment: The following calibration steps are being followed by the laboratory, which do not
match either the mg/L or % Saturation calibration instructions from the instrument manual 1)
Make sure the sponge is wet. 2) Insert probe into chamber. 3) Turn instrument on and allow
to stabilize. (may take 5-15 minutes). Note temperature displayed. 4) Press and release both
the UP and DOWN arrow keys at the same time and release. This enters calibration menu.
“CAL” should be displayed. 5) Press mode until mg/L is displayed on the right side of screen.
Press enter. (arrow pointing to left). 6) Using temperature noted from 3 above, refer to
DO/Temp chart, adjust readings with UP and DOWN arrows until correct reading is displayed.
Press enter twice.
Comment: Using mg/L mode during calibration requires a solution of known concentration.
If wishing to calibrate in mg/L mode, these instructions are also found in the YSI 550A
Operator’s Manual.
H. Finding: The laboratory is not calibrating the DO meter prior to sample analysis at each
sample site or performing a Post-Analysis Calibration Verification when analyses are
performed at multiple sample sites.
Requirement: Laboratory procedures shall comply with Subparagraph (a) (1) of this Rule.
Ref: 15A NCAC 02H .0805 (g) (4).
Requirement: When performing analyses at multiple sample sites, a post-analysis calibration
verification Standard must be analyzed at the end of the run for all types of DO probes, unless
the meter is recalibrated at each sample site. Below is a procedure for verifying the calibration
of a DO probe.
1) Follow the manufacturer’s instructions for meter operation.
2) Place probe in a plastic bag, the probe storage cup, the storage well of the meter
(each containing a wet sponge), or a BOD bottle partially filled with water. Allow
appropriate instrument warm up time.
3) Read DO and temperature.
4) Check the reading vs. the solubility table below and apply appropriate
atmospheric (barometric) pressure or altitude correction factor.
5) Calculated DO value must verify meter reading within ± 0.5 mg/L (do NOT
calculate and apply a correction factor to calculated theoretical DO).
Ref: NC WW/GW LCB Approved Procedure for the Analysis of Dissolved Oxygen (DO).
Comment: This is applicable since the laboratory is analyzing upstream and downstream
samples.
Temperature – Standard Methods, 2550 B-2010 (Aqueous)
Recommendation: It is recommended that Temperature results be reported on the DMR in
whole numbers, as stated in the “Precision in Discharge Monitoring Reports” document issued
by the Division of Water Resources.
Reporting
I. Finding: Data qualifiers from the contract laboratory reports are not being transferred to the
DMR.
Page 6
#5353 Town of Roseboro
Requirement: Reported data associated with quality control failures, improper sample
collection, holding time exceedances, or improper preservation shall be qualified as such.
Ref: 15A NCAC 02H .0805 (e) (5).
Comment: See the table in the Paper Trail Investigation section.
Recommendation: To avoid questions of legality, it is recommended that you contact the
appropriate Regional Office for guidance as to whether an amended DMR(s) will be
required. A copy of this report will be made available to the Regional Office.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing original records (e.g., laboratory benchsheets, logbooks, etc.) and
contract laboratory reports to DMRs submitted to the North Carolina Division of Water Resources. Data
were reviewed for Town of Roseboro (NPDES permit # NC0026816) for November 2020, March 2021 and
June 2021. The following errors were noted:
Sample
Collection
Date
Parameter
Location
Contract Lab Report Qualifier
6/30/2021 BOD Effluent Sample estimated. Did not meet quality control
requirements.
6/30/2021 BOD Influent
Blank above acceptance limit.
Upstream results for Temperature (17.4 °C), DO (6.20 mg/L) and Downstream results for Temperature
(11.7 °C), DO (6.04 mg/L) were reported on November 11, 2020 instead of November 13, 2020.
Upstream results for Temperature (7.2°C) and DO (10.76 mg/L) and Downstream results for Temperature
(7.6°C) and DO (10 mg/L) were reported on November 18, 2020 instead of November 19, 2020.
V. CONCLUSIONS:
Correcting the above-cited Findings and implementing the Recommendation(s) will help this laboratory to
produce quality data and meet Certification requirements. The inspector would like to thank the staff for
their assistance during the inspection and data review process. Please respond to all Findings and
include supporting documentation, implementation dates and steps taken to prevent recurrence
for each corrective action.
Report prepared by: Tonja Springer Date: December 9, 2021
Report reviewed by: Tom Halvosa Date: December 10, 2021
Certificate Number:5353
Effective Date:1/1/2021
Expiration Date:12/31/2021
Lab Name:Town of Roseboro
Address:10160 Roseboro Hwy.
Roseboro, NC 28382-
North Carolina Wastewater/Groundwater Laboratory Certification
Certified Parameters Listing
Date of Last Amendment:11/1/2021
The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below.
CERTIFIED PARAMETERS
INORGANIC
DISSOLVED OXYGEN
SM 4500 O G-2016 (Aqueous)
TEMPERATURE
SM 2550 B-2010 (Aqueous)
This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions
as set forth in 15A NCAC 2H.0807.