HomeMy WebLinkAbout#5571_2022_0303_TS_FINALNC Department of Environmental Quality | Division of Water Resources | Laboratory Certification Branch
4405 Reedy Creek Road | 1623 Mail Service Center | Raleigh, North Carolina 27699-1623
919-733-3908
May 12, 2022
5571
Mr. Ervin Glenn Holland
Holland Environmental
408 South Bentonville Road
Four Oaks, NC 27524-
Subject: North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC
WW/GW LCB) Maintenance Inspection
Dear Mr. Holland:
Enclosed is a report for the inspection performed on March 3, 2022 by Tonja Springer. I apologize
for the delay in getting this report to you. Where Finding(s) are cited in this report, a response is
required. Within thirty days, please supply this office with a written item for item description of
how these Finding(s) were corrected. Please describe the steps taken to prevent recurrence and
include an implementation date for each corrective action. If the Finding(s) cited in the enclosed
report are not corrected, enforcement actions may be recommended. For Certification
maintenance, your laboratory must continue to carry out the requirements set forth in 15A NCAC
02H .0800.
A copy of the laboratory’s Certified Parameter List at the time of the audit is attached. This list will
reflect any changes made during the audit. Copies of the checklists completed during the
inspection may be requested from this office. Thank you for your cooperation during the
inspection. If you have questions or need additional information, please contact me at (919) 733-
3908 Ext. 259.
Sincerely,
Beth Swanson
Technical Assistance & Compliance Specialist
Division of Water Resources
Attachment
cc: Tonja Springer, Todd Crawford, #5571
On-Site Inspection Report
LABORATORY NAME: Holland Environmental
NPDES PERMIT #: NC0026662, NC0038938, NC0072877
ADDRESS: 408 South Bentonville Road
Four Oaks, NC 27524
CERTIFICATE #: 5571
DATE OF INSPECTION: March 3, 2022
TYPE OF INSPECTION: Field Commercial Maintenance
AUDITOR(S): Tonja Springer
LOCAL PERSON(S) CONTACTED:
Glenn Holland
I. INTRODUCTION:
This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater
Laboratory Certification Branch (NC WW/GW LCB) to verify its compliance with the requirements of
15A NCAC 02H .0800 for the analysis of compliance monitoring samples.
II. GENERAL COMMENTS:
Staff was forthcoming and responded well to suggestions from the auditor. Laboratory
documentation, however, was incomplete and inaccurate and additional quality control procedures
need to be implemented. The inspector is concerned about the number of transcription errors and
record omissions detected during the data review process. Many of the data points recorded on the
benchsheet did not compare to the values reported on the Non-Discharge Monitoring Reports
(NDMRs). These observations are documented in the Paper Trail Investigation section of this report.
All required Proficiency Testing (PT) Samples for the 2022 PT Calendar Year have not yet been
analyzed. The laboratory is reminded that results must be received by this office directly from the vendor
by September 30, 2022.
The laboratory is reminded that any time changes are made to laboratory procedures, QA/SOP
document(s) must be updated and relevant staff retrained. Staff must acknowledge that they have read
and understand the changes as part of the documented training program. The same requirements apply
when changes are made in response to Findings, Recommendations or Comments listed in this report,
to ensure the methods are being performed as stated, references to methods are accurate, and the QA
and/or SOP document(s) is in agreement with each approved practice, test, analysis, measurement,
monitoring procedure or regulatory requirement being used in the laboratory. In some instances, the
laboratory may need to create an SOP to document how new functions or policies will be implemented.
Revisions to the SOPs, based on the Findings, Comments and Recommendations within this
report must be submitted to this office by November 1, 2022.
Page 2
#5571 Holland Environmental
The laboratory is also reminded that SOPs are required to be reviewed at least every two years and
are intended to describe procedures exactly as they are to be performed. Use of the word “should” is
not appropriate when describing requirements (e.g., Quality Control (QC) frequency, acceptance
criteria, etc.). Evaluate all SOPs for the proper use of the word “should”.
Contracted analyses are performed by Environmental Chemists, Inc. (Certification # 94).
Approved Procedure documents for the analysis of the facility’s currently certified Field Parameters
were provided at the time of the inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Documentation
A. Finding: Error corrections are not properly performed.
Requirement: All documentation errors shall be corrected by drawing a single line
through the error so that the original entry remains legible. Entries shall not be obliterated
by erasures or markings. Wite-Out®, correction tape, or similar products designed to
obliterate documentation are not to be used; instead the correction shall be written
adjacent to the error. The correction shall be initialed by the responsible individual and
the date of change documented. Ref: 15A NCAC 02H .0805 (g) (1).
Comment: Several instances were noted of overwriting and error corrections lacking
the date and initials of the individual responsible for the change.
B. Finding: The laboratory is not documenting all traceability information for purchased
materials, reagents and standards.
Requirement: 15A NCAC 02H .0805 (a)(7)(K) and (g)(7) requires laboratories to have
a documented system of traceability for the purchase, preparation, and use of all
chemicals, reagents, standards, and consumables. That system must include
documentation of the following information: Date received, Date Opened (in use),
Vendor, Lot Number, and Expiration Date (where specified). A system (e.g., traceable
identifiers) must be in place that links standard/reagent preparation information to
analytical batches in which the solutions are used. Documentation of solution
preparation must include the analyst’s initials, date of preparation, the volume or weight
of standard(s) used, the solvent and final volume of the solution. This information as well
as the vendor and/or manufacturer, lot number, and expiration date must be retained for
primary standards, chemicals, reagents, and materials used for a period of five years.
Consumable materials such as pH buffers, lots of pre-made standards and/or media,
solids and bacteria filters, etc. are included in this requirement. Ref: NC WW/GW LCB
Traceability Documentation Requirements for Chemicals, Reagents, Standards and
Consumables Policy.
Comment: Buffer lot numbers and expiration dates are documented on the pH
benchsheet. There is no traceability information documented on the benchsheet for
Total Residual Chlorine (TRC) DPD reagents or Conductivity standards.
Comment: Dates received and opened are written on the DPD reagent packets and pH
buffer bottles. While this can provide a traceability link to analyses by looking at the dates
that the chemicals were in use, that link is lost once the boxes and bottles are discarded.
Conductivity standards were not available for review at the time of inspection.
Page 3
#5571 Holland Environmental
Comment: An example reagent/standard receipt log was given at the time of
inspection.
C. Finding: All original records are not being maintained for five years.
Requirement: All analytical records, including original observations and information
necessary to facilitate historical reconstruction of the calculated results, shall be
maintained for five years. Ref: 15A NCAC 02H .0805 (g) (1).
Comment: The testing data is initially written on a piece of paper, transported back to
the laboratory, and transferred to the operator’s logbook and/or DMR. This original
paperwork is then discarded.
Comment: This Finding applies to the pH, DO and Temperature samples for permit #
NC0038938 Johnston County Schools.
Comment: Data omission, as noted in the Paper Trail section of this report, can be
perceived by third parties as falsification of data.
D. Finding: Sample collection and sample analysis time are not consistently documented.
Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data
shall be traceable to the associated sample analyses and shall consist of: the date and
time of sample collection; the time of sample analyses (when required to document a
required holding time or when time-critical steps are imposed by the method, a federal
regulation, or this Rule). Ref: 15A NCAC 02H .0805 (g) (2) (F) and (H).
Requirement: Alternatively, one time may be documented for collection and analysis
with the notation that samples are measured in situ or immediately at the sampling site
(i.e., immediately following collection at a location as near to the collection point as
possible). Ref: NC WW/GW LCB Approved Procedure for the Analysis of Temperature
and NC WW/GW LC Approved Procedure for the Analysis of Dissolved Oxygen (DO).
Comment: This Finding applies to DO, pH and TRC. Compliance samples are not
currently being analyzed for Conductivity and Settleable Residue. Any benchsheets for
future compliance sample data must adhere to the same Requirements. The sample
analysis time would be documented for PT samples as well.
Comment: In data reviewed for permit # NC0026662 Effluent, analyst initials are
documented in the space for DO sample collection/analysis time.
Comment: In data reviewed for permit # NC0038938 Effluent, sample
collection/analysis time is not documented for Temperature, DO and TRC.
Comment: In data reviewed for permit # NC0072877, sample collection/analysis time
is not documented for Temperature. Only one time is documented for Upstream and
Downstream samples without a notation that samples are measured in situ or
immediately at the sample site.
E. Finding: The laboratory benchsheet is lacking required documentation: the method
reference and the proper units of measure.
Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data
shall be traceable to the associated sample analyses and shall consist of: the method
Page 4
#5571 Holland Environmental
or Standard Operating Procedure; the proper units of measure. Each item shall be
recorded each time samples are analyzed. Analyses shall conform to methodologies
found in Subparagraph (a)(1) of this Rule. Ref: 15A NCAC 02H .0805 (g) (2) (A) and
(L).
Comment: The DO method is documented as SM 4500 O G without the reference year.
The pH, TRC, Settleable Residue and Conductivity method references include the
reference year (i.e., 2011) documented twice.
Comment: The PT Sample benchsheet reviewed for June 2021 lacked units of
measure for Conductivity and Settleable Residue.
F. Finding: The laboratory benchsheet is lacking required documentation for TRC: Daily
Check Standard analysis time(s).
Requirement: The State Laboratory may develop Approved Procedures for Field
Parameters based upon the methods in any of the sources referenced in Parts(a)(1)(A)
through (F) of this Rule. Ref: 15A NCAC 02H .0805 (a) (1) (F).
Requirement: The following must be documented in indelible ink whenever sample
analysis is performed: Daily Check Standard analysis date and time(s). Ref: NC
WW/GW LCB Approved Procedure for the Analysis of Total Residual Chlorine (DPD
Colorimetric by SM 4500 Cl G-2011).
Comment: There is a space to document the Daily Check Standard analysis time, but
the time is not being documented.
G. Finding: The laboratory benchsheet is lacking required documentation for Dissolved
Oxygen: meter calibration time.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this
Rule. Ref:15A NCAC 02H .0805 (g) (4).
Requirement: The following must be documented in indelible ink whenever sample
analysis is performed: Meter calibration and meter calibration time(s). Ref: NC
WW/GW LCB Approved Procedure for the Analysis of Dissolved Oxygen (DO).
H. Finding: The laboratory is not documenting the barometric pressure and salinity values
used to calibrate the DO meter.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this
Rule. Ref: 15A NCAC 02H .0805 (g) (4).
Requirement: The following must be documented in indelible ink whenever sample
analysis is performed: Calibration variables (temperature, elevation or barometric
pressure [in mmHg], and salinity). Ref: NC WW/GW LCB Approved Procedure for the
Analysis of Dissolved Oxygen (DO).
Comment: Per NC WW/GW LC Branch policy, facilities may use the Salinity default
value of zero when calibrating the DO meter unless it is known or suspected that the
Salinity value of the samples being analyzed is > 9 ppt. In those situations, actual
Salinity values must be used. Regardless of which value is used, it must be
documented. If the salinity is always the same a blanket statement can be added to
the benchsheet.
Page 5
#5571 Holland Environmental
Comment: Temperature is the only calibration variable documented.
I. Finding: The laboratory is not documenting the true value and percent recovery of QC
standards on laboratory benchsheets.
Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data
shall be traceable to the associated sample analyses and shall consist of: the quality
control assessments. Each item shall be recorded each time samples are analyzed. Ref:
15A NCAC 02H .0805 (g) (2) (O).
Requirement: The following must be documented in indelible ink whenever sample
analysis is performed: True value of the calibration verification check standard. Ref: NC
WW/GW LCB Approved Procedure for the Analysis of Specific Conductance
(Conductivity).
Requirement: The following must be documented in indelible ink whenever sample
analysis is performed: True value of the Daily Check Standard. Ref: NC WW/GW LC
Approved Procedure for the Analysis of Total Residual Chlorine (DPD Colorimetric by
SM 4500 Cl G-2011).
Comment: This Finding applies to TRC and Conductivity. The true value of QC
standards and percent recovery obtained must be documented on the benchsheet and
evaluated against established acceptance criteria to demonstrate that the analyst was
aware of any out-of-control situation. The corrective actions taken must be documented.
Any samples not meeting the acceptance criteria must be reanalyzed, if possible. If this
is not possible, the data must be flagged on the laboratory reports as all QC
requirements not met.
Chlorine, Total Residual – Standard Methods, 4500 Cl G-2011 (Aqueous)
Comment: TRC is only reported for permit # NC0038938.
Recommendation: The laboratory currently verifies the calibration curve using 10, 20, 50, 100,
300 and 400 µg/L standards. It is recommended that the laboratory verify 20, 40, 50, 200 and
400 µg/L standards. This places emphasis on the portion of the curve between the permit limit
and compliance limit, which is the range of most samples.
J. Finding: The three values used to assign the true value to the gel-type standard are not
documented.
Requirement: All analytical records, including original observations and information
necessary to facilitate historical reconstruction of the calculated results, shall be
maintained for five years. All analytical data and records pertinent to each certified
analysis shall be available for inspection upon request. Ref: 15A NCAC 02H .0805 (g)
(1).
Requirement: To assign a true value to the gel-type or sealed liquid standard: 1. Zero
the instrument with the calibration blank. 2. Read and record gel standard values. 3.
Repeat steps 1 and 2 at least two more times. 4. Assign the average value as the true
value. Ref: NC WW/GW LCB Approved Procedure for the Analysis of Total Residual
Chlorine (DPD Colorimetric by SM 4500 Cl G-2011).
Page 6
#5571 Holland Environmental
K. Finding: The laboratory is not analyzing a Method Blank when using laboratory-
prepared standards.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this
Rule. Ref: 15A NCAC 02H .0805 (g) (4).
Requirement: Method Blanks would be required when using laboratory-prepared
standards [including Proficiency Testing (PT) Samples] and anytime sample dilutions
are performed. Ref: NC WW/GW LC Approved Procedure for the Analysis of Total
Residual Chlorine (DPD Colorimetric by SM 4500 Cl G-2011).
Requirement: The Method Blank is deionized or distilled water from the same source
used to prepare the calibration verification standards or the PT Sample, and is analyzed
like a sample (i.e., with DPD/buffer added). The concentration of the Method Blank must
not exceed 50% of the reporting limit (i.e., the lowest calibration verification standard
concentration) or corrective action must be taken. Ref: NC WW/GW LC Approved
Procedure for the Analysis of Total Residual Chlorine (DPD Colorimetric by SM 4500
Cl G-2011).
Comment: The prepared PT Sample is considered a laboratory-prepared standard.
Comment: The “Chlorine Meter Calibration Check & Result Calculation” portion of the
Johnston County Schools benchsheet from February 2021 does not appear to have
data documented correctly. The Check Standard and Reagent Blank values are
identical.
Comment: On the 2021 TRC PT Sample benchsheet, the daily check standard result
(i.e., 184 µg/L) is documented in the space designated for the reagent blank.
Recommendation: The cell designated to document the method blank, per the
definition found in the Approved Procedure, is labeled as a reagent blank. It is
recommended to change this on the benchsheet to a method blank.
Conductivity – Standard Methods, 2510 B-2011 (Aqueous)
Comment: The laboratory does not currently analyze compliance samples for Conductivity.
Proficiency Testing data was made available for review during the audit.
L. Finding: The Automatic Temperature Compensator (ATC) was not verified prior to initial
use and every 12 months thereafter.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this
Rule. Ref: 15A NCAC 02H .0805 (g) (4).
Requirement: The Automatic Temperature Compensator (ATC) must be verified prior
to initial use and annually (i.e., 12 months) thereafter at two temperatures by analyzing
a standard or sample at 25°C (i.e., the temperature to which conductivity values are
reported) and a temperature(s) that brackets the temperature ranges of the
environmental samples routinely analyzed. This may require the analysis of a third
temperature reading that is > 25°C. As the temperature increases or decreases, the
value of the conductivity standard or sample must be within ±10% of the true value of
the standard or ±10% of the value of the sample at 25°C. If not, corrective action must
be taken. Ref: NC WW/GW LCB Approved Procedure for the Analysis of Specific
Conductance (Conductivity).
Page 7
#5571 Holland Environmental
Comment: Anticipated temperatures can be obtained from a review of the DMRs from
the peak summer and winter months. Historical data should provide a reasonably
accurate estimation of ranges that will bracket the expected sample temperatures.
Dissolved Oxygen – Standard Methods, 4500 O G-2016 (Aqueous)
M. Finding: When analyses are performed at multiple sample sites, the laboratory is not
calibrating prior to sample analysis at each sample site or performing a Post-Analysis
Calibration Verification.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this
Rule. Ref: 15A NCAC 02H .0805 (g) (4).
Requirement: When performing analyses at multiple sample sites, the meter must be
calibrated at each sample site prior to analysis or a post-analysis calibration verification
must be performed at the end of the run, regardless of meter type. The calculated
theoretical DO value must verify the meter reading within ±0.5 mg/L. If the meter
verification does not read within ±0.5 mg/L of the theoretical DO, corrective action must
be taken. If the meter is not calibrated at each sample site, it is recommended that a
mid-day calibration be performed when samples are extended over an extended period
of time. Ref: NC WW/GW LCB Approved Procedure for the Analysis of Dissolved
Oxygen (DO).
Comment: If the meter is recalibrated at each sample location before sample analysis
a post-analysis calibration verification would not be required.
Comment: This is applicable since the laboratory is analyzing upstream and
downstream samples.
.
pH – Standard Methods, 4500 H+ B-2011 (Aqueous)
N. Finding: Values were reported that exceed the method specified accuracy of 0.1 units.
Requirement: By careful use of a laboratory pH meter with good electrodes, a precision
of ±0.02 unit and an accuracy of ±0.05 unit can be achieved. However, ± 0.1 pH unit
represents the limit of accuracy under normal conditions, especially for measurement of
water and poorly buffered solutions. For this reason, report pH values to the nearest 0.1
pH unit. Ref: Standard Methods, 4500 H+ B-2011. (6).
Comment: pH is being reported to 2 decimal places on the Discharge Monitoring
Report (DMR) for Permit # NC0038938 and # NC0072877.
Comment: Per PT Vendor instructions, the PT Sample results should be reported to
two decimal places, which is an exception to the requirement for compliance samples.
Proficiency Testing
O. Finding: The laboratory is not documenting the preparation of PT Samples.
Requirement: PT Samples received as ampules are diluted according to the
Accredited PT Sample Provider’s instructions. It is important to remember to document
the preparation of PT Samples in a traceable log or other traceable format. The diluted
PT Sample then becomes a routine Compliance Sample and is added to a routine
Page 8
#5571 Holland Environmental
sample batch for analysis. No documentation is needed for whole volume PT Samples
which require no preparation, however the instructions must be maintained. Ref:
Proficiency Testing Requirements, February 19, 2020, Revision 5, Section 3.6.
Comment: Dating and initialing the instruction sheet for each prepared PT Sample
would satisfy the documentation requirement.
Temperature – Standard Methods, 2550 B-2010 (Aqueous)
Recommendation: It is recommended that Temperature results be reported on the DMR
in whole numbers, as stated in the “Precision in Discharge Monitoring Reports” document
issued by the Division of Water Resources.
P. Finding: The temperature sensor on the pH meter used to obtain reported temperature
values has not been checked against a Reference Temperature-Measuring Device
every 12 months.
Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this
Rule. Ref: 15A NCAC 02H .0805 (g) (4).
Requirement: All compliance temperature-measuring devices without an NIST
traceable certificate, or with an expired NIST traceable certificate, must be verified
against a Reference Temperature-Measuring Device and the process documented
initially and every 12 months. Verification documentation must include the serial
number of the device being checked. The serial number, stated accuracy and
expiration date of the Reference Temperature-Measuring Device used in the
comparison must also be documented. Verification data must be kept on file and be
available for inspection for 5 years. (Note: International Organization for
Standardization (ISO) 17025 compliant vendors or other Certified laboratories may
provide assistance in meeting this requirement. When an ISO compliant vendor
provides this assistance, they must provide the serial number, accuracy and
calibration date for the Reference Temperature-Measuring Device used for the
verification. When a Certified laboratory provides this service, they must provide a
copy of the NIST traceable certificate of the Reference Temperature-Measuring
Device used for the verification). Ref: NC WW/GW LCB Approved Procedure for
the Analysis of Temperature.
Comment: The most recent temperature sensor verification was performed January
2021.
Reporting
Q. Finding: Data qualifiers from the contract laboratory reports are not being transferred
to the DMR.
Requirement: Each certified Field Laboratory shall be in accordance with Paragraph
(e) of this Rule. Ref: 15A NCAC 02H .0805 (g) (17).
Requirement: Reported data associated with quality control failures, improper sample
collection, holding time exceedances, or improper preservation shall be qualified as
such. Ref: 15A NCAC 02H .0805 (e) (5).
Comment: See the table in the Paper Trail Investigation section.
Page 9
#5571 Holland Environmental
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing original records (e.g., laboratory benchsheets, logbooks, etc.)
and contract laboratory reports to DMRs submitted to the North Carolina Division of Water
Resources. Data were reviewed for Town of Princeton (NPDES permit # NC0026662) for November
and December 2021 and January and February 2022; Johnston County Schools (NPDES permit #
NC0038938) for November and December 2021 and January 2022; Town of Newton Grove (NPDES
permit # NC0072877) for November 2021 and February 2022. The following errors were noted:
Date Parameter Location
Value on
Benchsheet
Value on DMR
12/28/2021 pH Effluent
Town of Princeton 6.17 S.U. No value reported
12/29/2021 pH Effluent
Town of Princeton 5.62 S.U. No value reported
12/31/2021 pH Effluent
Town of Princeton 6.25 S.U. No value reported
12/30/2021 pH Effluent
Town of Princeton 6.02 S.U. No value reported
1/1/2022 TSS Influent 2
Town of Princeton *510 mg/L No value reported
1/1/2022 TSS Effluent 2
Town of Princeton *10.00 mg/L No value reported
1/3/2022 DO Effluent
Town of Princeton 4.58 mg/L 6.35 mg/L
1/5/2022 pH Effluent
Town of Princeton 6.01 S.U. No value reported
1/6/2022 DO Effluent
Town of Princeton 9.09 mg/L 9.07 mg/L
1/6/2022 pH Effluent
Town of Princeton 6.86 S.U. No value reported
1/7/2022 DO Effluent
Town of Princeton 9.62 mg/L 9.52 mg/L
1/7/2022 pH Effluent
Town of Princeton 6.65 S.U. No value reported
1/8/2022 TSS Effluent 2
Town of Princeton *12.5 mg/L 12.8 mg/L
1/11/2022 pH Effluent
Town of Princeton 6.74 S.U. 6.23 S.U.
1/11/2022 BOD Influent
Town of Princeton *49 mg/L 316 mg/L
1/11/2022 TSS Influent
Town of Princeton *316 mg/L 49 mg/L
1/12/2022 pH Effluent
Town of Princeton 6.51 S.U. No value reported
1/12/2022 Temperature Effluent
Town of Princeton 4.7 °C 4.9 °C
1/12/2022 DO Effluent
Town of Princeton 14.61 mg/L 11.3 mg/L
1/13/2022 pH Effluent
Town of Princeton 6.88 S.U. No value reported
Page 10
#5571 Holland Environmental
1/14/2022 pH Effluent
Town of Princeton 6.71 S.U. No value reported
1/19/2022 pH Effluent
Town of Princeton 6.55 S.U. No value reported
1/20/2022 pH Effluent
Town of Princeton 6.93 S.U. No value reported
1/20/2022 TSS Influent
Town of Princeton *23.0 mg/L 230 mg/L
1/21/2022 pH Effluent
Town of Princeton 6.47 S.U. No value reported
1/24/2022 pH Effluent
Town of Princeton 6.70 S.U. No value reported
1/26/2022 pH Effluent
Town of Princeton 5.35 S.U. No value reported
1/26/2022 Temperature Upstream
Town of Princeton 4.4 °C 4.5 °C
1/27/2022 pH Effluent
Town of Princeton 6.24 S.U. No value reported
1/28/2022 pH Effluent
Town of Princeton 5.36 S.U. No value reported
11/1/2021 BOD Outfall 001
Johnston County Schools *< 2.0 mg/L No value reported
11/2/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 17.1 °C
11/3/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 17 °C
11/4/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 16.3 °C
11/5/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 15.8 °C
11/8/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 15.2 °C
11/9/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 14.9 °C
11/10/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 15.5 °C
11/11/2021 Total Nitrogen Outfall 001
Johnston County Schools *39.6 mg/L No value reported
11/12/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 15.3 °C
11/14/2021 NO3+NO2
Nitrogen
Outfall
Johnston County Schools *29.4 mg/L No value reported
11/15/2021 Temperature Effluent
Johnston County Schools 15.9 °C 15.2 °C
11/16/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 14.9 °C
11/17/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 14.6 °C
11/18/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 14.3 °C
11/19/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 15.8 °C
Page 11
#5571 Holland Environmental
11/19/2021 Total Nitrogen Effluent
Johnston County Schools *31.2 mg/L No value reported
11/22/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 16 °C
11/23/2021 Total Nitrogen Effluent
Johnston County Schools *30.0 mg/L No value reported
11/24/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 15.4 °C
11/25/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 16 °C
11/26/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 15.5 °C
11/29/2021 Total Nitrogen Outfall 001
Johnston County Schools
* 45.8 mg/L No value reported
11/29/2021 Total
Phosphorus
Outfall 001
Johnston County Schools
*8.05 mg/L No value reported
11/29/2021 Fecal Coliform Outfall 001
Johnston County Schools
*<1 MPN/100 mL No value reported
11/30/2021 Temperature Effluent
Johnston County Schools
No value on
benchsheet 15.2 °C
12/9/2021 Total Nitrogen Outfall 001
Johnston County Schools *39.1 mg/L No value reported
12/13/2021 pH Outfall 001
Johnston County Schools 6.8 S.U. 6.9 S.U.
1/6/2022 Total Nitrogen Outfall 001
Johnston County Schools *46.1 mg/L No value reported
1/20/2022 Total Nitrogen Outfall 001
Johnston County Schools *42.3 mg/L No value reported
1/24/2022 Total Nitrogen Outfall 001
Johnston County Schools *36.1 mg/L No value reported
11/13/2021 Temperature Effluent
Town of Newton Grove 17.9 °C 16 °C
11/13/2021 pH Effluent
Town of Newton Grove 7.7 S.U. 7.4 S.U.
11/13/2021 DO Effluent
Town of Newton Grove 8.7 mg/L 7.87 mg/L
11/13/2021 Temperature Upstream
Town of Newton Grove 14.4 °C 12.3 °C
11/13/2021 DO Upstream
Town of Newton Grove 4.5 mg/L 3.0 mg/L
11/13/2021 Temperature Downstream
Town of Newton Grove 13.9 °C 13.4 °C
11/13/2021 DO Downstream
Town of Newton Grove 5.2 mg/L 5.8 mg/L
11/20/2021 DO Upstream
Town of Newton Grove 3.1 mg/L 3.2 mg/L
11/27/2021 Temperature Downstream
Town of Newton Grove 14.9 °C 15.1°C
11/27/2021 DO Downstream
Town of Newton Grove 3.8 mg/L 3.7 mg/L
11/29/2021 Temperature Effluent
Town of Newton Grove
No value on
benchsheet 13.1 °C
Page 12
#5571 Holland Environmental
11/29/2021 pH Effluent
Town of Newton Grove
No value on
benchsheet 7.6 S.U.
11/29/2021 DO Effluent
Town of Newton Grove
No value on
benchsheet 7.8 mg/L
* Contract Laboratory Data
Upstream (Town of Princeton) result reported for Temperature (12.5 °C) was reported on
December 30, 2021, instead of December 29, 2021.
Effluent (Town of Princeton) results for pH (6.23 S.U.) and DO (11.49 mg/L) were reported on
January 11, 2022, instead of January 10, 2022.
Effluent (Town of Princeton) result for pH (6.19 S.U.) was reported on February 3, 2022 instead of
February 2, 2022.
Upstream (Town of Princeton) results reported for DO (15.2 mg/L) and Temperature (4.0 °C) were
reported on February 3, 2022, instead of February 2, 2022.
On the Town of Princeton benchsheet for January 2022, “2021” is written in for January 1, 3, 4, 5,
6, 7 and 8.
Comment: The following tables are in reference to Finding Q.
Sample
Collection
Date
Parameter
Location
Contract Lab Report Qualifier
1/11/2022
1/20/2022 TSS Influent
Town of Princeton
Filtration stopped at 10 min. per method. Result is
valid.
2/1/2022 BOD
Influent
Town of Princeton
Sample estimated. Did not meet quality control
requirements. Blank = 0.25 mg/L. Above
acceptance limit 0.2 mg/L
2/1/2022 BOD
Effluent
Town of Princeton
Sample estimated. Did not meet quality control
requirements. Blank = 0.25 mg/L. Above
acceptance limit 0.2 mg/L
Sample
Collection Date
Parameter
Location
Contract Lab Report Qualifier
1/24/2022
2/1/2022
Ammonia
Nitrogen
Outfall 001
Johnston County
Schools
Ammonia and TKN are within QC recovery limit.
2/14/2022
BOD
Outfall 002
Johnston County
Schools
Sample estimated. Did not meet quality control
requirements. Blank = 0.31 mg/L. Above
acceptance limit 0.2 mg/L
Page 13
#5571 Holland Environmental
To avoid questions of legality and possible monitoring frequency violations, it is recommended that
you contact the appropriate Regional Office for guidance as to whether an amended DMR(s) will be
required. A copy of this report will be made available to the Regional Office.
Comment: Data omission can be perceived by third parties as falsification of data.
V. CONCLUSIONS:
Correcting the above-cited Findings and implementing the Recommendation(s) will help this
laboratory to produce quality data and meet Certification requirements. The inspector would like to
thank the staff for their assistance during the inspection and data review process. Please respond
to all Findings and include supporting documentation, implementation dates and steps taken
to prevent recurrence for each corrective action.
Report prepared by: Tonja Springer Date: April 7, 2022
Report reviewed by: Jill Puff Date: April 12, 2022
Certificate Number:5571
Effective Date:1/1/2022
Expiration Date:12/31/2022
Lab Name:Holland Environmental
Address:408 South Bentonville Road
Four Oaks, NC 27524-
North Carolina Wastewater/Groundwater Laboratory Certification
Certified Parameters Listing
Date of Last Amendment:9/1/2021
The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below.
CERTIFIED PARAMETERS
INORGANIC
CHLORINE, TOTAL RESIDUAL
SM 4500 Cl G-2011 (Aqueous)
CONDUCTIVITY
SM 2510 B-2011 (Aqueous)
DISSOLVED OXYGEN
SM 4500 O G-2016 (Aqueous)
pH
SM 4500 H+B-2011 (Aqueous)
RESIDUE, SETTLEABLE
SM 2540 F-2015 (Aqueous)
TEMPERATURE
SM 2550 B-2010 (Aqueous)
This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions
as set forth in 15A NCAC 2H.0807.