HomeMy WebLinkAbout#5127_2022_0302_JMS_FINAL NC Department of Environmental Quality | Division of Water Resources | Laboratory Certification Branch 4405 Reedy Creek Road | 1623 Mail Service Center | Raleigh, North Carolina 27699-1623 919-733-3908 April 4, 2022 5127 Mr. Trevor C. McMinn TREVCO 2020 Howard Gap Road Hendersonville, NC 28792 Subject: North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC WW/GW LCB) Maintenance Inspection Dear Mr. McMinn: Enclosed is a report for the inspection performed on March 2, 2022 by Jason Smith. Where Finding(s) are cited in this report, a response is required. Within thirty days, please supply this office with a written item for item description of how these Finding(s) were corrected. Please describe the steps taken to prevent recurrence and include an implementation date for each corrective action. If the Finding(s) cited in the enclosed report are not corrected, enforcement actions may be recommended. For Certification maintenance, your laboratory must continue to carry out the requirements set forth in 15A NCAC 02H .0800. A copy of the laboratory’s Certified Parameter List at the time of the audit is attached. This list will reflect any changes made during the audit. Copies of the checklists completed during the inspection may be requested from this office. Thank you for your cooperation during the inspection. If you have questions or need additional information, please contact me at (919) 733- 3908 ext. 251. Sincerely, Anna Ostendorff Technical Assistance & Compliance Specialist Division of Water Resources Attachment cc: Todd Crawford, Jason Smith, Master File #5127 On-Site Inspection Report LABORATORY NAME: Trevco NPDES PERMIT #: NC0086223 and NC0028975 ADDRESS: 2020 Howard Gap Road Hendersonville, NC 28792 CERTIFICATE #: 5127 DATE OF INSPECTION: March 2, 2022 TYPE OF INSPECTION: Field Commercial Maintenance AUDITOR(S): Jason Smith LOCAL PERSON(S) CONTACTED: Trevor McMinn and Heather Garren I. INTRODUCTION: This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC WW/GW LCB) to verify its compliance with the requirements of 15A NCAC 02H .0800 for the analysis of compliance monitoring samples. II. GENERAL COMMENTS: The facility is neat and well organized and has all the equipment necessary to perform the analyses. Staff was forthcoming and responded well to suggestions from the auditor. All required Proficiency Testing (PT) Samples have been analyzed for the 2022 PT Calendar Year and the graded results were 100% acceptable. The laboratory is reminded that any time changes are made to laboratory procedures, QA/SOP document(s) must be updated and relevant staff retrained. Staff must acknowledge that they have read and understand the changes as part of the documented training program. The same requirements apply when changes are made in response to Findings, Recommendations or Comments listed in this report, to ensure the methods are being performed as stated, references to methods are accurate, and the QA and/or SOP document(s) is in agreement with each approved practice, test, analysis, measurement, monitoring procedure or regulatory requirement being used in the laboratory. In some instances, the laboratory may need to create an SOP to document how new functions or policies will be implemented. Revisions to the SOPs, based on the Findings, Comments and Recommendations within this report must be submitted to this office by October 1, 2022. The laboratory is also reminded that SOPs are intended to describe procedures exactly as they are to be performed. Use of the word “should” is not appropriate when describing requirements (e.g., Quality Control (QC) frequency, acceptance criteria, etc.). Evaluate all SOPs for the proper use of the word “should”. Contracted analyses are performed by Environmental Testing Services (Certification #600). Page 2 #5127 - Trevco Approved Procedure documents for the analysis of the facility’s currently certified Field Parameters were provided at the time of the inspection. III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS: Documentation A. Finding: Error corrections are not properly performed. Requirement: All documentation errors shall be corrected by drawing a single line through the error so that the original entry remains legible. Entries shall not be obliterated by erasures or markings. Wite-Out®, correction tape, or similar products designed to obliterate documentation are not to be used; instead the correction shall be written adjacent to the error. The correction shall be initialed by the responsible individual and the date of change documented. Ref: 15A NCAC 02H .0805 (g) (1). Comment: Error corrections are not dated or initialed. B. Finding: The laboratory is not documenting all traceability information for purchased materials, reagents and standards. Requirement: 15A NCAC 02H .0805 (a)(7)(K) and (g)(7) requires laboratories to have a documented system of traceability for the purchase, preparation, and use of all chemicals, reagents, standards, and consumables. That system must include documentation of the following information: Date received, Date Opened (in use), Vendor, Lot Number, and Expiration Date (where specified). A system (e.g., traceable identifiers) must be in place that links standard/reagent preparation information to analytical batches in which the solutions are used. Documentation of solution preparation must include the analyst’s initials, date of preparation, the volume or weight of standard(s) used, the solvent and final volume of the solution. This information as well as the vendor and/or manufacturer, lot number, and expiration date must be retained for primary standards, chemicals, reagents, and materials used for a period of five years. Consumable materials such as pH buffers, lots of pre-made standards and/or media, solids and bacteria filters, etc. are included in this requirement. Ref: NC WW/GW LCB Traceability Documentation Requirements for Chemicals, Reagents, Standards and Consumables Policy. Comment: The date received is not documented. C. Finding: Chemical containers are not dated when received and when opened. Requirement: Chemical containers shall be dated when received and when opened. Ref: 15A NCAC 02H .0805 (g) (7). D. Finding: The laboratory benchsheet is lacking required documentation: the instrument identification. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the instrument identification. Each item shall be recorded each time samples are analyzed. Analyses shall conform to methodologies found in Subparagraph (a)(1) of this Rule. Ref: 15A NCAC 02H .0805 (g) (2) (C). Page 3 #5127 - Trevco E. Finding: The laboratory is not documenting the salinity value used to calibrate the Dissolved Oxygen (DO) meter. Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this Rule. Ref: 15A NCAC 02H .0805 (g) (4). Requirement: The following must be documented in indelible ink whenever sample analysis is performed: Calibration variables (temperature, elevation or barometric pressure [in mmHg], and salinity). Ref: NC WW/GW LCB Approved Procedure for the Analysis of Dissolved Oxygen (DO). F. Finding: The DO calibration log incorrectly labels the percent saturation as percent slope. Requirement: Laboratory procedures shall comply with Subparagraph (a)(1) of this Rule. Ref: 15A NCAC 02H .0805 (g) (4). Requirement: The following must be documented in indelible ink whenever sample analysis is performed: Final calibration information (final DO reading in mg/L or % saturation). Ref: NC WW/GW LCB Approved Procedure for the Analysis of Dissolved Oxygen (DO). G. Finding: The laboratory is not documenting the preparation of PT Samples. Requirement: PT Samples received as ampules are diluted according to the Accredited PT Sample Provider’s instructions. It is important to remember to document the preparation of PT Samples in a traceable log or other traceable format. The diluted PT Sample then becomes a routine Compliance Sample and is added to a routine sample batch for analysis. No documentation is needed for whole volume PT Samples which require no preparation, however the instructions must be maintained. Ref: Proficiency Testing Requirements, February 19, 2020, Revision 5, Section 3.6. Comment: Dating and initialing the instruction sheet for each prepared PT Sample would satisfy the documentation requirement. Comment: The laboratory has their contract laboratory prepare the Total Residual Chlorine PT for analysis. This requirement has been explained to the contract laboratory for future PT sample preparation. IV. PAPER TRAIL INVESTIGATION: The paper trail consisted of comparing original records (e.g., laboratory benchsheets, logbooks, etc.) and contract laboratory reports to Discharge Monitoring Reports submitted to the North Carolina Division of Water Resources. Data were reviewed for Adventure Village (NPDES permit # NC0086223) and the Saluda WWTP (NPDES permit # NC0028975) for January, May and July 2021. The following errors were noted: Date Parameter Location Value on Benchsheet Value on DMR 1/26/2021 pH Adventure Village Effluent 6.7 S.U. 6.1 S.U. 7/7/2021 Total Residual Chlorine Adventure Village Effluent 40 µg/L 45 µg/L Page 4 #5127 - Trevco 7/16/2021 Temperature Adventure Village Effluent 22 °C 26 °C To avoid questions of legality, it is recommended that you contact the appropriate Regional Office for guidance as to whether an amended DMR(s) will be required. A copy of this report will be made available to the Regional Office. V. CONCLUSIONS: Correcting the above-cited Findings will help this laboratory to produce quality data and meet Certification requirements. The inspector would like to thank the staff for their assistance during the inspection and data review process. Please respond to all Findings and include supporting documentation, implementation dates and steps taken to prevent recurrence for each corrective action. Report prepared by: Jason Smith Date: March 14, 2022 Report reviewed by: Tonja Springer Date: March 14, 2022 Certificate Number:5127 Effective Date:1/1/2022 Expiration Date:12/31/2022 Lab Name:TREVCO Address:2020 Howard Gap Road Hendersonville, NC 28792- North Carolina Wastewater/Groundwater Laboratory Certification Certified Parameters Listing Date of Last Amendment:10/26/2021 The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below. CERTIFIED PARAMETERS INORGANIC CHLORINE, TOTAL RESIDUAL SM 4500 Cl G-2011 (Aqueous) DISSOLVED OXYGEN SM 4500 O G-2016 (Aqueous) pH SM 4500 H+B-2011 (Aqueous) TEMPERATURE SM 2550 B-2010 (Aqueous) This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions as set forth in 15A NCAC 2H.0807.