HomeMy WebLinkAbout#5101_2022_0913_TS_FINAL
NC Department of Environmental Quality | Division of Water Resources | Laboratory Certification Branch
4405 Reedy Creek Road | 1623 Mail Service Center | Raleigh, North Carolina 27699-1623
919-733-3908
November 4, 2022
5101
Ms. Leanne Wilson
Duke Energy Carolinas - Cape Fear Plant
500 CP&L Road
Moncure, NC 27559-
Subject: North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC
WW/GW LCB) Maintenance Inspection
Dear Ms. Wilson:
Enclosed is a report for the inspection performed on September 13, 2022 by Tonja Springer. I
apologize for the delay in getting this report to you. Where Finding(s) are cited in this report, a
response is required. Within thirty days, please supply this office with a written item for item
description of how these Finding(s) were corrected. Please describe the steps taken to prevent
recurrence and include an implementation date for each corrective action. If the Finding(s) cited
in the enclosed report are not corrected, enforcement actions may be recommended. For
Certification maintenance, your laboratory must continue to carry out the requirements set forth
in 15A NCAC 02H .0800.
A copy of the laboratory’s Certified Parameter List at the time of the audit is attached. This list will
not reflect any changes made during the audit. Copies of the checklists completed during the
inspection may be requested from this office. Thank you for your cooperation during the
inspection. If you have questions or need additional information, please contact me at (919) 733-
3908 Ext. 259.
Sincerely,
Beth Swanson
Technical Assistance & Compliance Specialist
Division of Water Resources
Attachment
cc: Tonja Springer, Todd Crawford, #5101
On-Site Inspection Report
LABORATORY NAME: Duke Energy Carolinas - Cape Fear Plant
NPDES PERMIT #: NC0003433
ADDRESS: 500 CP&L Road
Moncure, NC 27559
CERTIFICATE #: 5101
DATE OF INSPECTION: September 13, 2022
TYPE OF INSPECTION: Field Industrial Maintenance
AUDITOR(S): Tonja Springer
LOCAL PERSON(S) CONTACTED:
Jonathan Stamas
I. INTRODUCTION:
This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory
Certification Branch (NC WW/GW LCB) to verify its compliance with the requirements of 15A NCAC 02H
.0800 for the analysis of compliance monitoring samples.
II. GENERAL COMMENTS:
The laboratory submitted requested documentation electronically September 8, 2022 and September 12,
2022. The inspection was performed via WebEx on September 13, 2022.
All required Proficiency Testing (PT) Samples have been analyzed for 2022 PT Calendar year and graded
results were 100% acceptable.
The laboratory is reminded that any time changes are made to laboratory procedures, QA/SOP document(s)
must be updated and relevant staff retrained. Staff must acknowledge that they have read and understand
the changes as part of the documented training program. The same requirements apply when changes are
made in response to Findings, Recommendations or Comments listed in this report, to ensure the methods
are being performed as stated, references to methods are accurate, and the QA and/or SOP document(s) is
in agreement with each approved practice, test, analysis, measurement, monitoring procedure or regulatory
requirement being used in the laboratory. In some instances, the laboratory may need to create an SOP to
document how new functions or policies will be implemented. Revisions to the SOP, based on the
Findings, Comments and Recommendations within this report must be submitted to this office by
January 31, 2023.
The laboratory is also reminded that SOPs are intended to describe procedures exactly as they are to be
performed. Use of the word “should” is not appropriate when describing requirements (e.g., Quality Control
(QC) frequency, acceptance criteria, etc.). Evaluate all SOPs for the proper use of the word “should”.
Contracted analyses are performed by Pace Analytical Services LLC - Asheville NC (Certification #40).
Page 2
#5101 – Duke Energy Carolinas – Cape Fear Plant
The Approved Procedure document for the analysis of the facility’s currently certified Field Parameter was
provided at the time of the inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Comment: No compliance samples are being analyzed for pH and Temperature. The Temperature
monitoring requirement was removed from the permit when it was renewed in 2020. Temperature was
dropped from the Certified Parameter Listing (CPL) on October 25, 2022. The laboratory has sent a written
request to make the laboratory inactive after December 31, 2022. Only pH Proficiency Testing (PT) data was
available for review.
Documentation
A. Finding: The units of measure for pH (i.e., Standard Units or S.U.) are not documented on the
benchsheet.
Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be
traceable to the associated sample analyses and shall consist of: the proper units of measure. Each
item shall be recorded each time samples are analyzed. Analyses shall conform to methodologies
found in Subparagraph (a) (1) of this Rule. Ref: 15A NCAC 02H .0805 (g) (2) (L).
Comment: The pH PT Sample result is documented on the bottom of the benchsheet and does not
include units.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing the laboratory benchsheet and the PT vendor reporting form. No
transcription errors were observed. The facility appears to be doing a good job of accurately transcribing
data.
V. CONCLUSIONS:
Correcting the above-cited Finding will help this laboratory to produce quality data and meet Certification
requirements. The inspector would like to thank the staff for their assistance during the inspection and
data review process. Please respond to all Findings and include supporting documentation,
implementation dates and steps taken to prevent recurrence for each corrective action.
Report prepared by: Tonja Springer Date: September 21, 2022
Report reviewed by: Jill Puff Date: October 25, 2022
Certificate Number:5101
Effective Date:1/1/2022
Expiration Date:12/31/2022
Lab Name:Duke Energy Carolinas - Cape Fear Plant
Address:500 CP&L Road
Moncure, NC 27559-
North Carolina Wastewater/Groundwater Laboratory Certification
Certified Parameters Listing
Date of Last Amendment:
The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below.
CERTIFIED PARAMETERS
INORGANIC
pH
SM 4500 H+B-2011 (Aqueous)
TEMPERATURE
SM 2550 B-2010 (Aqueous)
This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions
as set forth in 15A NCAC 2H.0807.