UVF Trilogy Analytical Report
Samples Collected: December 2, 2021
Samples Analyzed: December 10, 2021
Site Name: Speedway #8654
Site Address: 5068 NC Highway 33 East, Greenville, NC
GRI Project Number: 3389
December 13, 2021
Joanna Alley, E.I.T.
Operator
Technical Procedures
Calibration
The UFV Trilogy has standard calibrations stored for the following analysis:
GRO
DRO
Total PAH
Target PAH
Heavy PAH
Prior to use, calibration standards are used to perform a 5-point verification on the factory stored
calibrations for GRO and DRO. The calibration standards used are as follows:
GRO
0.5 ppm
1 ppm
3 ppm
5 ppm
10 ppm
DRO
0.1 ppm
0.5 ppm
1 ppm
3 ppm
5 ppm
Operation
Sample Extraction
5mg of soil are weighed into a clean extraction jar
Methanol is added to the extraction jar to create a 2x or 4x dilution
o 2x Dilution: 10 mL of methanol
o 4x Dilution: 20 mL of methanol
The extraction jar is shaken for several minutes to thoroughly mix the soil with solvent
(methanol)
The mixture is allowed to settle for several minutes
A clean syringe is used to draw liquid from the extraction jar
A clean filter is attached to the end of the syringe and the liquid dispensed through the filter
into a clean test tube
This liquid is referred to as the extraction
Dilution Preparation
An adjustable micropipette is used to precisely measure out the amount of extract needed for
the dilution
The measured out amount of the extract is added to a clean test tube and methanol is added
dilute the extract. Examples of dilutions are shown below:
Extract Amount
500 μL
200 μL
100 μL
50 μL
25 μL
Add Solvent to the
5 mL line
5 mL line
5 mL line
5 mL line
5 mL line
2x Extract
Dilution Factor
20x
50x
100x
200x
400x
4x Extract
Dilution Factor
40x
100x
200x
400x
800x
Technical Procedures
Sample Analyzation
The dilution is shaken to thoroughly mix, then poured into a clean glass cuvette
The desired module is inserted into the UVF Trilogy and the corresponding factory calibration is
selected
“Measure Fluorescence” is pressed on the touchscreen and the reading is recorded
The reading should be between the concentrations of the calibration standards (GRO: 0.5-10
ppm; DRO 0.1-5 ppm)
o Reading below the calibration standard range are recorded as below the detection limit
Detection Limit = (lower range of calibration standards range) x dilution factor
(i.e. a GRO sample with a 100x dilution factor has a detection limit of 50 ppm)
o Readings above the calibration standard range should be further diluted and reanalyzed
o Negative readings are an indication of quenching (see below)
The reading is multiplied by the dilution factor to calculate the final concentration
Check suspected contaminated samples for quenching by running additional dilutions
o According to the siteLAB Quick Reference Guide:
“Quenching” can occur when the detector is swamped by too many hydrocarbons or
organic interferences, producing low or negative concentrations. Test the sample at
multiple dilutions to confirm results are linear and accurate.”
QA/QC
Method Blank
A method blank of methanol will be analyzed following calibration and following the last sample
analyzed each day
Duplicates
One duplicate will be analyzed for every 20 samples
The duplicate will be prepared from the extract of the sample and diluted to the same factor as
the sample analyzed
Project Name:Date Collected:
Project Number:Date Analysed:
Project Manager:Operator:
Sample ID Matrix Dilution
Factor x
GRO
Output
(ppm)
=GRO Value
(mg/Kg)
Dilution
Factor x
DRO
Output
(ppm)
=DRO Value
(mg/Kg)
DSL-2'Soil 100 9.63 963 100 8.91 891
Dup DSL-2'Soil 100 10.00 1000 100 9.39 939
UVF Analytical Report
siteLAB UVF Trilogy
12/10/2021
12/2/2021
Joanna Alley
Speedway #8654
3389
Nicole Long
Site:
Speedway #8654
Report Date:
12/13/2021
Page:
1 of 1