Loading...
HomeMy WebLinkAboutNCD980557656_19970509_NC State University (Lot 86 Farm Unit 1)_FRCBERCLA SAP QAPP_Quality Assurance Project Plan Volume II of II (Revision 0)-OCR! ~· I I ~ GE I Consultants, Inc. I I I I I I I I I I I I II II II I I QUALITY ASSURANCE PROJECT PLAN LOT 86 SUPERFUND SITE RALEIGH, NORTH CAROLINA Volume II ofll Submitted to: North Carolina State University Environment, Health & Safety Center Campus Box 8007 Raleigh, North Carolina 27695 7721 Six Forks Road, Suite 136 Raleigh, North Carolina 27615 (919) 676-0665 Revision 0 May 9, 1997 Project 97191 I I I I I I I I I I I I I I 'I I I APPENDIX A Laboratory Quality Assurance Plan Pace Analytical Services, Inc. I I I I I I I I I I I I I I I I I I I Prepared by Pace Analytical Services, Inc. 1710 Douglas Drive North Minneapolis, Minnesota 55422 (612)544-5543 LABORATORY QUALITY ASSURANCE PLAN Pace Approval / -1/'.2 .;,_ /7.S Date Date: 12/22/95 Section 1.0 Revision 0.01 Page 1 of 4 All rights reserved. No part of this manual may be reproduced or used in any form or by any means -graphic, elec!ronic, meci1anic:al, including photocopying, recording, taping, or information storage and retrieval systems -without permission of the publisher. Copy __ issued to: Name Affiliation Date This is/ is not a controlled document (control status must be circled before release). An E,ual Opportunity Employer :lqapreva\.secti .doc 1 . 1 FOREWORD Date: 12/22/95 Section 1.0 Revision 0.01 Page 2 of 4 Pace Analytical Services, Inc. is a privately held, full service environmental testing firm operating a system of 7 laboratories and multiple service centers nationwide. The Pace Minnesota laboratory was established in Minneapolis, Minnesota, on. August 3, 1978. Since its inception, Pace has provided analytical services for clients ranging from federal and municipal government to industrial firms and private consulting groups. As an independently owned environmental laboratory company, with Pace, there is never the question of conflict of interest. Since the foundation of its first laboratory, Pace has always retained quality as its primary objective. Pace Analytical Services, Inc. offers extensive services, including: bioassay for aquatic toxicity, air toxics, explosives, field services and mobile laboratory capabilities. The Pace system offers extensive capacity, and the ability to transfer work within the integrated system of laboratories assures that tum-around times are met. And, perhaps most importantly, geographic expansion has brought to Pace many valued and dedicated employees, with diverse interests and areas of expertise. There are nearly 400 people who contribute daily to the success of Pace. Over the years, Pace has developed and continues to develop a strict system of QNQC protocols, originally modeled after the USEPA Contract Laboratory Program (CLP) requirements. In addition, Pace has developed an advanced data management system, which is highly efficient and allows for flexible data reporting. Together, the two systems insure data reliability and timeliness. The advances have not been limited to the company itself. Pace and Pace employees have been instrumental in the development of the environmental testing industry. Pace employees are among the founders and board members of the industry's two major associations: the American Council of Independent Laboratories (ACIL}, and the International Association of Environmental Testing Laboratories (IAETL). Pace employees have delivered papers and published articles on laboratory management, Quality, and, most recently, on the newly developed model laboratory contract. Today, Pace is not only keeping stride with the evolving industry, but is actively engaged in that evolution. Pace is operating a high productivity environmental testing laboratory in northern California near San Francisco. The laboratory was designed with process efficiency and quality as its major objectives. The results in enhancements to analytical, quality, and data management systems will be replicated in other Pace laboratories. The strength of our company comes from how we are organized. We understand how important it is to develop long-term, on-going communication with our clients. Wrth the client at the center, we have an integrated local support team which revolves around the client. The national system provides the local team with additional capacity, specialty services, and additional experts in all areas of the business, in order to ensure that requirements are met. :lqapreva\sect1 .doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 1.2 1.3 Date: 12/22/95 Section 1.0 Revision 0.01 Page 3 of 4 Our goal is to continuously combine our expertise in the laboratory with customized solutions to meet the specific needs of our clients. By providing the right chemistry and the right solution, Pace has become known as a leader in the industry with satisfied, long-term clientele. CORPORATE PHILOSOPHY The criteria for selecting an analytical laboratory have changed significantly in recent years. Increased environmental liabilities have altered the attitudes of users and providers of laboratory services. Quality is now the primary criterion. Our philosophy at Pace, as it always has been, is to provide clients with the standards of service they require and deserve. It is a philosophy dedicated to providing: Uncompromising Quality Service Responsive to Clients' Needs A Single Source of Comprehensive Services Since the company's inception, Pace professionals have worked diligently to meet these goals. Our continued commitment to these standards remains the top priority at Pace. THE MISSION OF PACE ANALYTICAL SERVICES, INC. To be your Preferred Choice for Environmental Analytical Services in the Laboratory and in the Field For our clients: by consistently meeting our commitments by delivering responsive service, on time, with high value by assuring data quality and technical excellence For our employees: by offering equal opportunity for professional development by providing stimulating, participative, and safe work environments by valuing personal worth; encouraging excellence through recognition and reward For our shareholders: by generating a return on investment which meets our obligations and promotes company objectives For our suppliers: by offering long term relationships to those who support Pace's quality and business objectives :lqapreva\seci1.doc 1.4 For our communities: by being environmentally responsible and a good corporate citizen Date: 12/22/95 Section 1.0 Revision 0.01 Page 4 of 4 Pace strives to be the preferred choice for all its stakeholders, by providing quality services with the highest level of professional and ethical standards. CODE OF ETHICS In carrying out its Corporate Mission, Pace requires its employees to abide by the highest professional, ethical standards. Employees will conduct their tasks according to the highest professional, technical and ethical standards applicable· to their area of expertise. As such, Pace requires a commitment from all staff to abide by the principles set forth by the Company. This applies to all procedures, documented and undocumented, executed by employees. The following information summarizes the essential standards of ethical behavior required of Pace employees. Simply stated, Pace's fundamental ethical principles are as follows: • Each Pace emp.loyee is responsible for the propriety and consequences of his or her actions. • Each Pace employee must conduct all aspects of Company business in an ethical and strictly legal manner, and must obey the laws of the United States and of all localities, states and nations where Pace does business or seeks to do business. • Employee conduct on behalf of the Company with clients, suppliers, the public and one another must reflect the highest standards of honesty, integrity and fairness. Strict adherence by each Pace employee to this Code and to the Standards of Conduct is essential to the continued vitality of Pace. Therefore, compliance with and effective enforcement of the Code and Standards are key responsibilities of Pace management and will be addressed as elements of each employee's regular performance evaluation. Failure to comply with the Code or Standards will result in disciplinary action up to and including termination and referral for civil or criminal prosecution where appropriate. An employee will be notified of an infraction and given an opportunity to explain as prescribed under current disciplinary procedures. :lqapreva\sect~ .doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Section No. 1.0 2.0 :\lq aoreva \se ct2 .doc 2.0 TABLE OF CONTENTS Section Name No. of Pages Title Page 4 1.1 Forward 1.2 Corporate Philosophy 1.3 The Mission of Pace Analytical Services, Inc. 1.4 Code of Ethics Date: 12/22/95 Section 2.0 Revision 0.01 Page 1 of 9 Revision Date 12/22/95 Revision Number 0.01 Table of Contents 9 12/22/95 0.01 2.1 List of Tables 2.2 5.1 Quality Control Objectives: Methods 801 OB & 8020A 5.2 Quality Control Objectives: Purgeable Petroleum Hydrocarbon Analysis 5.3 Quality Control Objectives: Extractable Petroleum Hydrocarbon Analysis 5.4 Quality Control Objectives: Methods 8080A & CLP SOW 5.5 Quality Control Objectives: Methods 8240B & CLP SOW 5.6 Quality Control Objectives: Methods 8270B & CLP SOW 5.7 Quality Control Objectives: Metals by SW846 & CLP SOW 5.8 Quality Control Objectives: General Chemistry 6.1 Sampling and Preservation Requirements - Water 6.2 Sampling and Preservation Requirements - Soil 6.3 Sampling and Preservation Requirements - Air 8.1 Summary of Calibration Requirements 8.2 Summary of Routine Calibration Requirements 9.1 Analytical Protocols 9.2 List of Analy1ical Methods 9.3 BFB Key Ions & Ion Abundance Criteria 9.4 DFTPP Key Ions & Ion Abundance Criteria 11.1 Summary of Calibration and Quality Control Procedures 13.1 Scheduled Maintenance Procedures and Representative Spare Parts for Major Instrumentation List of Figures 4.1 Pace Analytical Services, Inc. Organizational Structure 4.2 Pace Analytical Services, Inc. Laboratory Organizational Structure 6.1 Presampling Communication, Sample Collection and Holding Schematic Section No. 3.0 4.0 No. of Section Name Pages 7 .1 Chain of Custody 7.2 Sample I.D. and Condition Form 7.3 Discrepancy Report Form 7.4 Internal Chain of Custody 7 .5 Client Sample Return Letter 7.6 Sample Disposition Form 7.7 Hazardous Sample Disposal Option Form 10.1 Analytical Data Review Process 11.1 Spike Recovery Control Chart 11.2 RPO (Duplicate) Control Chart 15 .1 Corrective Action Form 2.3 LQAP Distribution List Date: 12/22/95 Section 2.0 Revision 0.01 Page 2 of 9 Revision Date Revision Number Introduction 12 12/22/95 0.01 3.1 Program Objectives 3.2 Statement of Policy 3.3 Purpose and Scope 3.4 Quality Assurance Documents 3.4.1 QA Manual 3.4.2 Standard Operating Procedures Manuals 3.4.3 Project QA Manuals 3.4.4 Document Control, Distribution and Revision 3.5 Terms and Definitions Laboratory Organization and Responsibility 10 12/22/95 0.01 4.1 Laboratory Organization 4.2 Description of Responsibilities 4.2.1 Corporate Quality Assurance Officer 4.2.2 Laboratory General Manager 4.2.3 Laboratory Project Manager 4.2.4 Quality Assurance Officer 4.2.5 Operations Manager 4.2.5 Supervisors 4.2.7 Analysts 4.2.8 Sample Custodian 4.3 Training and Orientation :\lq a oreva \sect2. doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Section No. 5.0 6.0 7.0 : \lq a orev a \se ct2. doc 4.4 4.5 Section Name Laboratory Safety Security and Confidentiality Quality Assurance Objectives 5. 1 Level of QA Effort 5.2 Accuracy and Precision 5.3 Completeness 5.3.1 Random Error 5.4 Representativeness 5.5 Comparability 5.6 Traceability No. of Pages 12 5. 7 Quality Assurance Project Plan Exceptions 5.8 Personnel Quality Objective Sampling Procedures 8 6.1 Introduction 6.2 Sampling Services 6.2.1 Ground Water Monitoring 6.2.2 Waste Water Monitoring 6.2.3 Hazardous Waste Sampling 6.2.4 Flow Monitoring 6.2.5 Soil & Soil Gas Sampling 6.2.6 PCB Services 6.2. 7 Ambient Air Monitoring and Stack Emission Testing 6.3 Field Support 6.4 Preservation 6.5 Sample Bottles 6.6 Sample Receipt Schedule Sample Custody 20 7 .1 Sample Receipt 7.2 Chain-of-Custody 7 .3 Sample Verification 7.4 Sample Log-In 7.5 When Samples are Received with no Paperwork 7.6 Responsibilities for Sample Log-In 7.7 Sample Storage 7.8 Sample/Data Access and Internal Chain-of-Custody 7.9 Subcontracting Analytical Services 7.10 Sample Disposal 7.11 Excess Sample Disposition Date: 12/22/95 Section 2.0 Revision 0.01 Page 3 of 9 Revision Date Revision Number 12/22/95 0.00 12/22/95 0.00 12/22/95 0.01 Section No. 8.0 9.0 :\lqaoreva\sect2.doc Section Name Calibration Procedures and Frequency 8.1 Standards and Traceability 8.2 General Calibration Procedures 8.2.1 Analytical Balances 8.2.2 Thermometers 8.2.3 pH/Electrometers 8.2.4 Spectrophotometers 8.3 GC/MS Calibration Procedures 8.4 Non GC/MS Chromatography Calibration Procedures 8.5 Calibration of ICPs and AAs No. of Pages 9 Date: 12/22/95 Section 2.0 Revision 0.01 Page 4 of 9 Revision Date 12/22/95 Revision Number 0.00 Analytical Procedures 31 12/22/95 0.00 9.1 Analytical Methods· 9.2 Sample Preparation Methods 9.3 9.2.1 Digestion of Aqueous Samples for Metals -Method 3005A 9.2.2 Digestion of Aqueous Samples for Metals -Method 3010A and the CLP sow 9.2.3 Digestion of Aqueous Samples for Metals -Method 3020A and the CLP sow 9.2.4 Digestion of Solid Samples for Metals - Method 3050A and the CLP SOW 9.2.5 Separatory Funnel Extraction -Method 35108 9.2.6 Continuous Liquid/Liquid Extraction - Method 3520A 9.2.7 Soxhlet Extraction -Method 35408 9.2.8 Sonication Extraction -Method 3550A 9.2.9 Waste Dilution -Method 3580A 9.2.10 Purge-and-Trap Sample Introduction - Method 5030A 9.2.11 Extraction Procedure Toxicity Test (EP-TOX) - Method 131 0A 9.2.12 Toxicity Characteristic Leaching Procedure (TCLP) -Method 1311 9.2.13 California Assessment Manual Waste Extraction Test (CAM WET) Calibration and Analysis Procedures for Organics 9.3.1 Halogenated Volatile Organics - Method 80108 I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Section No. I :\lqaoreva\sect2.doc No. of Section Name Pages 9.3.2 Aromatic Volatile Organics -Method 8020A 9.3.3 Organochlorine Pesticides and PCBs - Method 8080A and the CLP SOW 9.3.4 Volatile Organics -Method 8240B and the CLP SOW 9.3.5 Semivolatile Organics -Method 8270A and the CLP SOW 9.3.6 Purgeable Petroleum Hydrocarbons 9.3. 7 Extractable Petroleum Hydrocarbon 9.4 Representative Calibration and Analysis Procedures for lnorganics 9.4.1 Metals by ICPS -Method 601 DA and the CLP SOW 9.4.2 Metals by GFM-Methods 7060A, 7421, 7740, 7841 and the CLP SOW 9.4.3 Mercury by CVM -Methods 7470, 7471A and the CLP SOW 9.4.4 Total and Amenable Cyanide - Methods 9010N9012 and the CLP SOW 9.4.5 Anions -Method 300.0 9.4.6 pH -Method 150.1 9.4.7 Non-Filterable Residue -Method 160.1 9.4.8 Filterable Residue -Method 160.2 9.4.9 Nitrate-Nitrite -Method 353.2 9.4.10 Total Organic Carbon (TOC) - Methods 9060 and 415.1 9.4.11 Oil and Grease -Methods 9070/9071 and413.1 9.4.12 Oil and Grease-Method 413.2 9.4.13 Total Recoverable Petroleum Hydrocarbons (TRPH) - Method 418.1 9.5 Method Validation 9.6 Method Detection Limits 9.7 Compliance 9. 7 .1 Definition 9. 7 .2 Understanding the Regulatory Framework 9.7.3 Commitment 9.7.4 Resolving Compliance Contradictions and Hierarchies 9.7.5 Disclosure of Noncompliance Date: 12/22/95 Section 2.0 Revision 0.01 Page 5 of9 Revision Date Revision Number Section No. 10.0 11.0 12.0 :\lqaoreva\sect2.doc Section Name Data Reduction, Validation and Reporting 10.1 Data Reduction 10.2 Data Validation 10.3 Data Report 10.4 Data Archive 10.5 Response to Inquiries Quality Control Procedures 11.1 Organic Analysis 11.2 Metals Analysis 11.2.1 Accuracy 11.2.2 Precision 11.2.3 Limits 11.3 Standards 11.4 Method Detection Limit 11 . 5 Control Charts _11.5.1 Field Blanks 11.5.2 Trip Blanks 11.5.3 Equipment Rinsate Blanks 11.5.4 Matrix Spike/Matrix Spike Duplicate Samples 11.6 Laboratory Control Samples 11.6.1 GC Methods 11.6.2 GC/MS Methods 11.6.3 Metals Analysis 11.6.4 Cyanide Analysis 11.6.5 Anion Analysis 11.6.6 Fluoride Analysis 11.6.7 Total Organic Carbon Analysis 11.6.8 Oil and Grease Analysis No. of Pages 6 41 Date: 12/22/95 Section 2.0 Revision 0.01 Page 6 of 9 Revision Date 12/22/95 Revision Number 0.00 12/22/95 0.00 11.6.9 Total Recoverable Petroleum Hydrocarbons (TRPH) Analysis 11.6. 10 California Assessment Manual Waste Ex1raction Test (CAM WEntEx1raction Procedure Toxicity Test Method (EP-Tox)/Toxicity Characteristic Leaching Procedure (TCLP) Quality Assurance Audits and Performance 7 Evaluations 12.1 Internal Audits 12.1.1 Quality Assurance Auditor 12.1.2 Scope and Frequency of Internal Audits 12.1.3 Internal Audit Report and Corrective Action Plans 12.2 External Audits 12/22/95 0.01 I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Section No. 13.0 14.0 15.0 :\lqaoreva\sect2.doc Section Name 12.3 Total Quality System Audit No. of Pages 12.4 Performance Evaluation Audits 12.4.1 Pace PE Samples 12.4.2 EPA WP and WS Studies 12.4.3 Other PE Studies Preventive Maintenance 6 13. 1 Maintenance Responsibilities 13.2 Maintenance Schedules 13.2.1 Preventive Maintenance -GC/MS 13.2.2 Preventive Maintenance -GC 13.2.3 13.2.4 Preventive Maintenance -ICP Preventive Maintenance -AA Graphite Furnace 13.2.5 Preventive Maintenance -Mercury Analyzer 13.2.6 Preventive Maintenance -General Laboratory Areas 13.3 13.4 13.2. 7 Preventive Maintenance - T h e r m o m e t e r s , Refrigerators, Ovens and Balances Maintenance Documentation Spare Parts Assessment of Precision, Accuracy, Completeness 14.1 Precision 14.2 Accuracy 14.3 Control Charts 14.3.1 Warning Limits 14.3.2 Control Limits 14.3.3 Utilization of Acceptance Limits 14.4 Representativeness 14.5 14.6 Completeness Comparability 7 Date: 12/22/95 Section 2.0 Revision 0.01 Page 7 of 9 Revision Date Revision Number 12/22/95 0.01 12/22/95 0.00 Corrective Action 13 12/22/95 0.00 15.1 Non-Conformance Memo 15.2 Out of Control Events 15.2.1 Volatile Organic Analyses 15.2.2 Semivolalile Organic Analyses 15.2.3 Gas Chromatography Analyses 15.2.4 Metals Analyses 15.3 Out-of-Statistical-Control Blank Spike Control Chart Data 15.3.1 Out-of-Control Blank Spike Recovery Data 15.3.2 Out-of-Statistical-Control Conditions Section No. 16.0 17.0 :\fqaoreva\sect2.doc 15.4 No. of Section Name Pages 15.3.3 Corrective Action for Out-of-Statistical Control Conditions Unusual Occurrences Quality Assurance Reports to Management 2 16.1 Quality Assurance Auditor 16.2 Quality Assurance Officer 16.3 Management Review of the Quality Assurance Program 16.4 Quarterly Quality Reports to Management Summary of Revisions 1 17.1 Revision Designation 17.2 Summary of Revisions Date, 12/22/95 Section 2.0 Revision 0.01 Page 8 of 9 Revision Date 12/22/95 12/22/95 Revision Number 0.00 0.00 I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 2.3 I :\lqaoreva\sect2.doc LQAP DISTRIBUTION LIST Date: 12/22/95 Section 2.0 Revision 0.01 Page 9 of 9 Individual copy distribution of this document may originate at any Pace facility and from the Corporate Office. When a copy of the Quality Assurance Plan is released, a designation is made on the cover as to whether the document is a controlled copy. Recipients of controlled copies will automatically be issued an updated version whenever revisions are made to the existing document. _Each Pace location which distributes copies of this plan shall maintain a record of the name of the individual receiving the document, their affiliation, the number of the copy issued and the control status (i.e., controlled vs. uncontrolled). I I I I I I I I I I I I I I I I I I I 3.1 3.2 3.0 PROGRAM OBJECTIVES INTRODUCTION Date: 12/22/95 Section 3.0 Revision 0.01 Page 1 of 12 The Pace Laboratory Quality Assurance Plan (LQAP) presents in specific terms the policies, organization, functions, and specific quality assurance (QA) and quality control (QC) requirements designed to achieve the data quality goals required for clients of Pace Analytical Services, Inc .. The U.S. Environmental Protection Agency's (U.S. EPA) QA policy requires a written and approved Quality Assurance Project Plan (QAPP) for every monitoring and measurement project mandated or supported by the U.S. EPA through regulations, contracts, or other formalized means not currently covered by regulation. Guidelines followed in the preparation of this plan are set forth in the document entitled "EPA Requirement for Quality Assurance Project Plans for Environmental Data Operations", EPA QA/R-5, Draft Interim Final (August 1994). Other documents that have been referenced for this plan include U.S. EPA Region IX Guidance for Preparing Quality Assurance Project Plans for Superfund Remedial Projects (September 1989); U.S. EPA's Guidance on Remedial Investigations Under CERCLA (June 1985); Guidance on Feasibility Studies Under CERCLA (June 1985); Compendium of Superfund Field Operations Methods (September 1987); Data Quality Objectives for Remedial Response Activities (March 1987); and Guidelines for Assessing and Reporting Data Quality for Environmental Measurements (January 1983). This detailed plan has been prepared for use by contractors who perform environmental services to ensure that the laboratory produces data that are scientifically valid and defensible. The establishment and documentation of these procedures will also ensure that the data are collected, reviewed, and analyzed in a consistent manner. STATEMENT OF POLICY Pace Analytical Services, Inc. is committed to providing the highest quality product to our clients. The validity and reliability of the data generated are ensured by the adherence to rigorous quality assurance/quality control (QA/QC) protocols and a Total Quality Management (TQM) system. Pace emphasizes the application of sound QA/QC principles beginning with the initial planning of the project, through all the field and laboratory activities, and ultimately to the generation of the final report. The principles of concise data quality objectives, representativeness, completeness, comparability, precision and accuracy are applied. The major elements of the overall Laboratory Quality Assurance Program at Pace are summarized below: The use of appropriate methodologies by technically competent, well-trained personnel with state-of-the-art instrumentation and equipment. : \lqapreva\sect3.doc 3.3 Date: 12/22/95 Section 3.0 Revision 0.01 . Page 2 of 12 Adherence to well-defined standard operating procedures with emphasis on good laboratory and measurement practices. Analysis and assessment of quality control samples including (but not limited to) matrix spike samples, matrix spike duplicate samples, duplicate samples, blanks and independent laboratory control standards. Successful participation in external quality evaluation programs. Accreditation by state, federal, and other applicable agencies for the work performed. Internal and external auditing to ensure compliance to protocols and provide assessment of the analytical methods. Pace is committed to providing the resources, including facilities, equipment and personnel, to ensure the adherence to these rigorous quality assurance/quality control protocols. Pace's quality assurance policy is based on the definition of quality as conformance to requirements; and further, on the premise that the requirements are governed by Company policies, government regulations and standard operating procedures. This commitment recognizes the need for data to be representative of the environmental conditions under consideration, and for data to be generated within a system of functions that is designed to meet applicable regulatory compliance criteria. To this end, Pace has developed a Quality Assurance (QA) Plan and maintains an ongoing QA Program. Our Quality Assurance Program contains provisions for establishing, maintaining and executing protocols which lead to results of known, appropriate and acceptable quality; documentation of these activities is an integral part of the QA program. No other concern will be permitted to interfere with the quality of data Pace provides to clients. This manual describes the set of policies and principles which guide day-to-day operations. Specific protocols are included by reference and are contained in a series of volumes cited.in Section 9.0 of this document. This document describes ongoing laboratory operations for routine analyses performed at Pace. As such, the material contained within is subject to change. Changes may be based on specific project requirements or procedural system modifications geared towards operational process and quality improvements. At a minimum, this document is reviewed and updated on a yearly basis. PURPOSE AND SCOPE This manual details the quality assurance program in effect at all Pace Analytical Laboratories. It is meant to be a teaching tool and source of information for laboratory personnel. The Manual is divided into logical sections, each dealing with a different phase : \lqa p rev a \sect3. doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 3.4 Date: 12/22/95 Section 3.0 Revision 0.01 Page 3 of 12 of laboratory operation, yet all sections overiap and function together to form a complete quality assurance program. The Manual is based on Good Laboratory Practices, common sense, and industry-accepted standard analytical practices .. The Manual must be read and understood by all laboratory personnel as part of their training program. The Manual should also be referred to regularly as a source of information. A system of continuous updating is built into the Manual to allow it to change as laboratory conditions change or as new regulations are promulgated. This manual is a controlled document, which means that its identity, development, distribution, and status must be known and traceable at all times. All Pace laboratory personnel have access to a controlled copy. Whenever a technician or analyst is in doubt as to proper procedures in a specific circumstance, the Manual should be consulted. Omissions or errors should be immediately reported to the Quality Assurance Officer, for corrective action. IT IS THE RESPONSIBILITY OF EACH LABORATORY WORKER TO ENSURE THAT THE PROVISIONS OF THIS MANUAL ARE FOLLOWED. Disagreement with specific requirements or knowledge of changes causing deviation from the procedures should be discussed with the immediate supervisor before further work is completed. Laboratory personnel are encouraged to comment on the Manual and make recommendations for more efficient procedures. · The latest revision of each section of the Manual is the applicable rule. Therefore, revisions will be announced to all laboratory personnel. An uncontrolled copy of the Manual is offered to clients and regulatory agencies as the definitive quality assurance program used at Pace. QUALITY ASSURANCE DOCUMENTS 3.4.1 QA Manual This document describes management policies related to operation of the analytical laboratories. It provides overall guidance regarding acceptable practices and discusses each element of the Quality Assurance Program. It functions as the Project QA Manual where no other Quality Assurance Project Plan, Statement of Work or other contractually mandated project plan has been specified. Adherence to the practices described in this manual is required of all employees. This manual may be revised and/or superseded only with the written authority of Vice President of Quality/Technical Director. Copies of this manual are controlled and distribution is administered by the Corporate Quality Office. 3.4.2 Standard Operating Procedures Manuals All procedures related to sample collection, storage, preparation, analysis, disposal, data validation, data reporting and employee training and safety shall be contained in written Standard Operating Procedures (SOPs). Each SOP shall :\Jqapreva\sect3.doc · Date: 12/22/95 Section 3.0 Revision 0.01 Page 4 of 12 contain the elements outlined in the Pace Corporate SOP ALL-P-001-A, Guidance Document for Pace Analytical Services, Inc. for the Preparation of Standard Operating Procedure Documents. All sections shall be structured in a step-wise manner using numbered sections. All record-keeping requirements shall be described at each step in the SOP. Examples of forms used shall be included as tables or figures and referenced within the text. Preparation of SOPs which have company-wide application will be the responsibility of the Corporate QAO. Analytical and evidentiary SOPs which are unique to an operating facility shall be prepared by designated personnel (e.g., analytical-section supervisor; evidentiary-laboratory QAO). SOPs shall be assigned a number from the Inventory List for SOPs maintained by the Corporate Quality Office or the Quality Assurance Department of the individual lab, as applicable. This number shall become part of the document control number when the SOP is accepted for implementation by Pace management. Laboratory SOPs shall be reviewed and approved by the relevant Section Supervisor (and Operations Manager for all SOPs related to analytical procedures) and the QA Officer, and submitted by the QA Department to the Operations Manager and the General Manager for approval prior to implementation. 3.4.3 Project QA Manuals· Project QA Manuals shall be implemented as required. These shall include such documents as Quality Assurance Project Plans (QAPPs). For those projects which require specific QNQC criteria, a QAPP which has been approved by a regulatory agency, usually the EPA, is provided to Pace by the client. Often the analytical section of a QAPP is written by Pace for the client. In this instance, the QAPP is reviewed and approved by the appropriate Pace Quality Assurance Officer and the Pace Operations Manager. 3.4.4 Document Control, Distribution and Revision In order that this document achieve the goals outlined in Section 3.2, it is necessary that each Pace laboratory employee be familiar with the current provisions of this document. It is also necessary that this document represent a consensus among Pace management and operational personnel as to the quality level desired and the means to that end. Prior to its publication as a controlled document, this manual must be approved by the Vice President of Quality/Technical Director. To obtain such approval, the document proceeds through an iterative process of review and revision, involving the affected managers and their designated representatives. The signature page at the beginning of the manual represents acceptance. :\lq a prev a \sect3. doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Date: 12/22/95 Section 3.0 Revision 0.01 Page 5 of 12 Each time a revision is made to this manual, it must also be approved. The Vice President of QualityfTechnical Director must approve each revision. 3.5 TERMS AND DEFINITIONS Accuracy: Aliquot· Analyte: Batch· Blind Sample: CRDL CRQL Calibration- Calibration Check: Comparability: Completeness· :\lqapreva\sect3.doc The degree of agreement between a measured value and the true or expected value. A measured portion of a sample taken for analysis. The specific entity an analysis seeks to determine. A grouping of no more than twenty samples of similar matrix which are prepared and/or analyzed together with the same method and the same lots of reagents within the same time frame. A sample may be analyzed in a different analytical batch than the one with which it was prepared. A blank is an artificial sample designed to detect and/or monitor the contribution of analyte and non-analyte contamination, instrumental background and sample processing to the measurement system. A sample submitted for analysis whose composition is known to the submitter but unknown to the analyst. Contract required detection limit. Contract required quantitation limit. The process of establishing the relationship between instrument response and known, traceable quantities of analytes of interest. Verification of the ratio of instrument response to analyte amount, a calibration check, is done by analyzing for analyte standards in an appropriate solvent. Calibration check solutions are made from a stock solution which is different from the stock used to prepare standards. Comparability is a qualitative parameter expressing the confidence with which one data set can be compared to another. Comparable data are produced through the use of standardized procedures and techniques. Measure of the amount of valid data obtained from a measurement system compared to the amount that was expected to be obtained under correct normal conditions. The equation for completeness is: # of data points obtained X 100 = % completeness # of data points expected Continuing Calibration: Control Chart: Control Limit: Detection Lim.it Dry Weight Duplicate Analysis· · Duplicate Sample: Environmental Sample: :\lqapreva\sect3.doc Date: 12/22/95 Section 3.0 Revision 0.01 Page 6 of 12 The process of analyzing standards periodically to verify the maintenance of calibration of the analytical system. A graphical plot of test results with respect to time or sequence of measurement, together with limits within which they are expected to lie when the system is in a state of statistical control. A range within which specified measurement results must fall to signify compliance. Control limits may be mandatory, requiring corrective action if exceeded, or advisory, requiring that nonconforming data be investigated and flagged. The minimum concentration ·of a substance that can be measured and reported with 99% confidence that the analyte concentration is greater than zero. The weight of a sample based on percent solids. The weight after drying in an oven. A second measurement made on the same sample extract or digestate to assist in the evaluation of precision of analysis. A second aliquot of the same sample that is treated the same as the original sample in order to determine the precision of the method. An environmental sample or field sample is a representative sample of any material (aqueous, nonaqueous, or multimedia) collected from any source for which determination of composition or contamination is requested or required. Environmental samples can generally be classified as follows: Surface Water and Ground Water Drinking Water • Delivered (treated or untreated) water designated as potable water. Water/Wastewater • Raw source waters for public drinking water supplies, ground waters, municipal influents/effiuents, and industrial influents/effiuents. Sludge • Municipal sludges and industrial sludges. I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I. I I Equipment llifil)k Fjeld Blank· Field Sample· Holding Time: Homogeneity· Instrument Detection Limit- :\lqapreva\sect3.doc Date: 12/22/95 Section 3.0 Revision 0.01 Page 7 of 12 Soil Predominately inorganic matter ranging in classification from sands to clays. Waste -Aqueous and nonaqueous liquid wastes, chemical solids, and industrial liquid and solid wastes. Special type of field blank used primarily as a check on equipment decontamination procedures. After decontamination, the sampling equipment is rinsed with DI water and the water collected for analysis. A quality control sample that is used to assess the contamination effects on accuracy due to the combined activities of sampling and analysis. Typically, it is composed of analyte free matrix (e.g., deionized water) prov_ided by the laboratory. A portion of material received by the laboratory to be analyzed, that is contained in single or multiple containers and identified by a unique field ID number. The elapsed time expressed in days from the date of sample collection by the field personnel until the date of its processing/analysis. For the Contract Laboratory Program, holding times start at the Verified Time of Sample Receipt by the laboratory. Holding time requirements are dictated by the method or QAPP. The degree to which a property or substance is evenly distributed throughout a material. The minimum concentration of a substance that can be measured and reported on a specific analytical instrument with 99% confidence that the analyte concentration is greater than zero. The instrument detection limit is determined by replicate analyses of a standard solution prepared at the instrument. The instrument detection limit is generally more sensitive than the method detection limit because its determination does not include sample preparation steps. Initial Calibration- Internal Standards· Laboratory Control Sample: LIMS· J.Qt MB.Q; Matrix: Matrix Spike: :\lqapreva\sect3.doc Date: 12/22/95 Section 3.0 Revision 0.01 Page 8 of 12 The process of analyzing standards, prepared at specified concentrations, to define the quantitative response, linearity and dynamic range of the instrument to the analytes of interest. Initial calibration is performed whenever the results of a continuing calibration do not confonm to the re-quirements of the method in use or at a frequency specified in the method. Analytes added to every standard, blank, job control sample, matrix spike, matrix spike duplicate, and sample at a known concentration, prior to analysis for the purpose of adjusting the response factor used in quantitating target analytes. Internal standards are used as the basis for quantitation of the target compounds, and are generally applicable to or-ganic analyses. A control sample of known composition spiked with a known concentration of analytes of interest. Aqueous and solid laboratory control samples are analyzed using the same preparation, reagents, and analytical methods employed for field samples. Laboratory Information Management System, the Pace company-wide LIMS, has been identified as Environmental Project Information Controller (EPIC). A quantity of bulk material of similar composition processed or manufactured at the same time. Method Requirements Documents are written guidelines which outline a consistent definition of work perfonmance for basic method compliance. The documents serve to interpret and define the subjective (vague) portions of the EPA's method for company-wide application. Each MRD is intended to establish a company-wide, baseline level of consistency for a single regulatory-derived method. The predominant material of which the sample to be analyzed is composed. Aliquot of sample fortified (spiked) with known quantities of specified target compounds or analytes and subjected to the entire sample preparation and analysis procedure in order to assess the appropriateness of the method for the sample matrix by measuring recovery. I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Matrix Spike Duplicate: Method Blank· Method Detection Limit· Pace Reporting Limit· Performance Audit or Evaluation: Precision: Protocol: !:QI.: :\lqapreva\sect3.doc Date: 12/22/95 Section 3.0 Revision 0.01 Page 9 of 12 A second aliquot of the sample that is treated the same as the original matrix spike sample. The relative percent difference between the matrix spike and matrix spike duplicate is calculated and used to assess analytical precision. An analytical control consisting of a blank matrix containing all reagents, internal standards .and surrogate standards, that is carried through the entire analytical procedure. The method blank is used to define the level of laboratory background and contamination, and to demonstrate that this level does not exceed acceptance limits. Acceptable levels of contamination are defined by project specific data quality objectives: The minimum concentration of a substance that can be measured and reported with 99% confidence that the ana\yte concentration is greater than zero. Method Detection Limits are determined using replicate spike samples prepared by the lab and taken through all preparation and analysis steps of the method. The method detection limit is calculated using the appropriate Student's !-parameter times the standard deviation of a series of spiked samples. PRLs were developed in conjunction with analysis codes (A-codes) for the EPIC LIMS. PRLs create uniformity across the company by establishing a standardized reporting limit by method to be utilized by all Pace laboratories. The PRL has been defined as the highest statistically derived· MDL value for a particular method found at any of the Pace laboratory operations which are performing the method. A process to evaluate the proficiency of an analyst or laboratory by evaluation of the results obtained on test materials in either a known, single or double-blind fashion. The measurement of agreement of a set of replicate results among themselves without any prior information as to the true result. Precision is assessed by means of duplicate/replicate sample analysis. A stated plan that clearly defines the objectives, methods and procedures for accomplishing a task. The practical quantitation limit (PQL) is the lowest level that can be reliably achieved within specified limits of precision and accuracy during routing laboratory operating conditions. QAPP: Quality Assurance: Quality Control: Reagent Grade: Replicate · Samples: Reporting Limit. Rounding ~ :\lqapreva\sect3.doc Date: 12/22/95 Section 3.0 Revision 0.01 Page 10 of 12 A Quality Assurance Project Plan or QAPP is a project specific document that describes the policies, organization, objectives, functional activities, and specific QA and QC activities designed to achieve the data quality goals of a specific project. A system of policies and procedures whose purpose is to ensure, confirm and document that the product or service rendered fulfills the requirements of Pace and its client. Quality Assurance includes quality planning, quality control, quality assessment (auditing), quality reporting and corrective action. A system of checks and corrective measures, integrated with the activities that directly generate the product or service, that serves to monitor and adjust the process to maintain conformance to predetermined requirements. Analytical reagent (AR) grade, ACS reagent grade, and reagent grade are synonymous terms for reagents which conform to the current specifications of the Committee on Analytical Reagents of the American Chemical Society. A second, separate sample collected at the same time, from the same place, for the same analysis, as the original sample in order to determine precision between the two samples. The level at which method, permit, regulatory and client specific objectives are met. The reporting limit may never be lower than the statistically determined MDL, but may be higher based on any of the above considerations. Reporting limits are corrected for sample · amounts, including the dry weight of solids, unless otherwise specified. Reporting limits are often set according to action or cleanup levels for a particular site or project which have been established in accordance with Data Quality Objectives (DQOs) under which the analytical work is to be processed. If the figure following those to be retained is less than 5, the figure is dropped, and the retained figures are kept unchanged. As an example, 11.443 is rounded to 11.44. If the figure following those to be retained is greater than 5, the figure is dropped, and the last retained figure is raised by 1. As an example, 11.446 is rounded to 11.45. If the figure following those to be retained is 5, and if there are no figures other than zeros I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Sample Delivery Group <SPGJ: Sensitivity· Split Sample· Standard: Standard Blank: Standard Curve: Standard Operating Procedure: : \lq a p rev a \sect3. doc Date: 12/22/95 Section 3.0 Revision 0.01 Page 11 of 12 beyond the five, the figure 5 is dropped, and the last-place figure retained is increased by one if it is an odd number or it is kept unchanged if an even number. As an example, 11.435 is rounded to 11.44, while 11.425 is rounded off to 11.42. If a series of multiple operations is to be performed (add, subtract, divide, multiply), all figures are carried through the calculations. Then the final answer is rounded to the proper number of significant figures. A unit within a single project that is used to identify a group of samples for delivery. An SDG is a group of 20 or fewer field samples within a project, received over a period of up to 14 calendar days. Data from all samples in an SDG are reported concurrently. A Sample Delivery Group is generally defined by one of the following, whichever occurs first: • All samples within a project; or • Every set of 20 field samples within a project; or • All samples received within a 14-day calendar period Samples may be assigned to Sample Delivery Groups by matrix (i.e., all soil samples in one SDG, all water samples in another), at the discretion of the laboratory. Clients may establish different SDG classifications to meet project specific requirements. Capability of methodology or instrumentation to discriminate between samples having differing concentrations or containing differing amounts of an analyte. A portion or subsample of a total sample obtained in such a manner that is not believed to differ significantly from _other portions of the same sample. A substance or material, the properties of which are known with sufficient accuracy, to permit its use to evaluate the same property in a sample. A calibration standard consisting of the same solvent/reagent matrix used to prepare the calibration standards without the analytes. It is used to construct the calibration curve by establishing instrument background. A standard curve is a curve which plots concentrations of known analyte standard versus the instrument response to the analyte. A procedure adopted for repetitive use when performing specific measurement or sampling operation. It may be an industry accepted standard method or one developed by the user. Surrogates: Systems Audit: Traceability: Trip Blank: Validation: Warning Limits: :\lqapreva\sect3.doc Date: 12/22/95 Section 3.0 Revision 0.01 Page 12 of 12 When employed, these are compounds added to every blank, sample, matrix spike, matrix spike duplicate, lab control sample, and standard prior to any processing or preparation; used to evaluate analytical efficiency by measuring recovery. Surrogate compounds are not .expected to be detected in environmental media, but are similar to the analytes of interest. Surrogates are generally utilized for organic analyses. An on-site inspection or assessment of a laboratory's quality control system. The ability to trace the source and accuracy of a material (i.e. standard) to a recognized primary reference source such as the National Institute of Standards and Technology (NIST) or USEPA. Also, the ability to independently reconstruct and review all aspects of the measurement system through available laboratory notebooks and documentation and reach the same results. This blank is used to detect sample contamination from the container and preservative during transport and storage of the sample. A cleaned sample container is filled with laboratory pure water: any preservative used in the sample is added; and then the blank is stored, shipped, and analyzed with its group of samples. The process by which a sample, measurement, method, or piece of data is deemed useful for a specified purpose as based upon the DQOs established for quality control measurements such as accuracy, precision, representativeness, and completeness. The limits (typically 2 standard deviations either side of the mean) shown on a control chart within which most results are expected to lie (within a 95% probability) while the system remains in a state of statistical control. I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 4.0 QA ORGANIZATION AND PERSONNEL I Date: 12/22/95 Section 4.0 Revision 0.01 Page 1 of 10 Pace Analytical Services, Inc. is a privately held, full service environmental testing firm operating an integrated system of seven laboratori~s. plus multiple service centers nationwide. Each laboratory within the system is set up as an !individual entity with local management, but all share common systems and receive support from the corporate office. The chief function of the corporate office is to assist the system laboratories. The corporate office centralizes company wide accounting, business development, fin~ncial management, human resources development, information systems, marketing and quality activities. The organizational structure of the corporation is provided in Figure 4.1. · For efficient laboratory operation, it is important that all laboratory employees understand the operational structure, specific areas of respohsibility and lines of authority within the organization. It is equally important for laboratory personlnel to understand that the structures of the Quality Organization may be separate from other ikboratory operations but that the quality function is totally integrated into every aspect of laboratory operation. All laboratory personnel are responsible for knowing and following pro'per methods and standard operating procedures; recording quality control information require'd by those procedures in the proper location; and suspending analyses when quality control cri/eria are not met. The organizational structure of a Pace Anal~ical Services, Inc. analytical chemistry laboratory is provided in Figure 4.2. The laboratory is managed by the General Manager. The Client Services (Sample and Project Management) and Quality Assurance Groups report directly to the General Manager. I Under the direction of the Laboratory Operations Manager, the technical staff of the laboratory is generally organized into the following function~! groups: Sample Preparation -Organic Sample Preparation -Metals Wet Chemistry Metals Analysis GCAnalysis GC/MS Volatiles Analysis GC/MS Semivolatiles Analysis Reporting/Data Validation In some laboratory operations the Laboratory Operations Manager position may not exist; in such a case, the responsibilities of the position a1re distributed between the Organic and Inorganic Department Managers. Each group is headed by a Group Leader or Section Supervisor who is responsible for operations on a daily basit Environmental chemists, analysts, laboratory technicians and laboratory assistants report to \he Group Supervisors. :\lqapreva\sect4 .doc 4.1 LABORATORY ORGANIZATION Date: 12/22/95 Section 4.0 Revision 0.01 Page 2 of 10 It is the individual responsibility of each analyst and technician to perform their assigned tasks according to the applicable SOPs, QA Project Plans, Study Protocols, and Work Plans. This responsibility includes performing quality control analyses as specified in the method SOP and entering the QC data in the appropriate method control file system. The analyst shall report out-of-control results to the Group Supervisor/Leader. Group Supervisors/Leaders shall ensure that analysts and technicians are instructed in the requirements of the Pace Laboratory QA Manual, site-specific QA Project Plans, SOPs, Protocols, and Work Plans for the analytical method or other procedure. Group Supervisors/Leaders shall review sample QC data at frequent intervals designed to ensure that QC analyses are being performed at the required frequency, that data are documented in the method control file system and that established corrective action procedures for out-of-control situations are followed and the results documented. It is the responsibility of the Group Supervisor/Leader to ensure that data have been validated and reported· to the Operations Manager. Group Supervisors/Leaders shall report to the appropriate Manager. The Operations Manager shall take overall responsibility for technical conduct, evaluation and reporting of all analytical tasks associated with each study. The Operations Manager ensures that approved procedures are documented and followed, that all data are recorded and verified and that all deviations from approved procedures are documented. The Operations Manager shall ensure that Group Supervisors/Leaders are instructed in the requirements of the Pace Laboratory QA Manual, study-specific QA Project Plans, SOPs, Protocols, and Work Plans. The Operations Manager provides guidance and assistance in the development of laboratory quality control procedures; approves quality · control limits for methods; works with supervisors to bring out-of-control methods back to within established acceptance limits; and assists supervisors in correcting analytical problems revealed in QA audits. The Operations Manager shall report to the General Manager. The Quality Assurance Department, under the direction of the Quality Assurance Officer, shall be responsible for conducting systems audits and inspections for compliance with this manual, SOPs and QA Project Plans or other project-specific protocols, maintaining the archives, maintaining historical files of all QA documents, reviewing QC charts, documenting findings and corrective actions, and reporting findings to management. The Quality Assurance Officer shall report directly to the General Manager of the Pace facility. The Pace General Manager shall designate and replace if necessary, the Operations Manager, and is responsible for managing all activities related to laboratory services, including the Quality Assurance Program. The Pace General Manager shall ensure that there is a Quality Assurance Department, that personnel and other resources are adequate, that personnel have been informed of their responsibilities, that deficiencies are reported to the Operations Manager and that corrective actions are taken and documented. Any significant changes to written SOPs shall be authorized in writing by :\lqapreva\sect4.doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 4.2 Date: 12/22/95 Section 4.0 Revision 0.01 Page3of10 either the General Manager or the Operations Manager and the Quality Assurance Officer of the Pace location. DESCRIPTION OF RESPONSIBILITIES Individuals involved with implementing procedures outlined in the LQAP have the following quality related duties and responsibilities: 4.2.1 Pace's Vice President of Quality is responsible for assisting in the development, implementation and monitoring of quality programs for the company. Responsibilities include: 1. Review and direct implementation of appropriate analytical Standard Operating Procedures. 2. Formulate and implement analytical product deliverables. 3. Provide technical direction to laboratories regarding existing and new analytical operations. 4. Provide leadership and direction to the laboratory Quality Assurance Officer. 5. Perform laboratory and project specific audits. 6. Assist in development, implementation, and monitoring of appropriate training programs. 4.2.2 The Pace Laboratory General Manager is responsible for overall laboratory operations. Specific responsibilities that relate to quality assurance are: 1. Implement the QA Program within the specific laboratory. 2. Regularly determine the effectiveness of the QA program. 3. Supervise quality control activities. 4. Approve laboratory-specific attachments to the QA manual and project-specific Quality Assurance Project Plans. 5. Recommend changes in the QA Program to the laboratory Quality Assurance Officer. 6. Maintain a current distribution list for QAPPs and generic LQAP. 7. Approval oversight for all reports issued by the laboratories. 8. Serve as the focal point for the reporting and disposition of all nonconformances. 9. Maintain a current laboratory organization chart. 4.2.3 The Pace Laboratory Proiect Manager is the lead person within the laboratory for direct oversight of all aspects of a specific project. Specifically, some of the project manager's responsibilities are: 1. Establishing direct dialogue with the client pertaining to project requirements, including methodology, TAT, technical information, etc. 2. Arranging bottle orders and shipment of sample kits to client. :\lqapreva\sect4 .doc 3. 4. 5. 6. 7. 8. Date: 12/22/95 Section 4.0 Revision 0.01 Page 4 of 10 Verifying log-in information relative to project requirements and field sample chain of custodies. Interfacing with laboratory operations staff to update and set job priorities. Updating clients on job status. Providing verbal and facsimile results to clients. Assisting laboratory staff with report preparation. Working with clients, laboratory staff, and other appropriate Pace staff to develop project statements of work or resolve problems of data quality, turnaround, or completeness. 4.2.4 The Pace laboratory Quality Assurance Officer (QAO) reviews all aspects of QA/QC for the laboratory. The duties of the laboratory QAO are to: 1. Assist the project manager in specifying QA/QC procedures to be used during the project. 2. Execute QC procedures and techniques to ensure that the laboratory achieves established standards of quality. 3. Evaluate data quality and maintain records on related QC charts and other pertinent information. 4. Monitor laboratory activities to determine conformance with authorized QA policy, and to implement appropriate steps to ensure adherence to QA programs. 5. Coordinate with the client's representative concerning external audits. 6. Review performance evaluation results. 7. Assist in development and implementation of appropriate training programs. 4.2.5 The Operations Manager oversees day-to-day production and quality activities of both inorganics and organics laboratory section providing wet chemistry, metals prep, metals, pesticide/PCB, volatiles and semivolatiles analyses. The specific duties of the Operations Manager are: :\lqapreva\sect4 .doc 1. Provide supervision of laboratory operations. 2. Implement the laboratory quality assurance plan. 3. Ensure proper scheduling and execution of testing programs. 4. Ensure that quality assurance and quality control criteria of analytical methods and projects are satisfied. 5. Assess data quality and take corrective action when necessary. 6. Notify the project team of specific laboratory nonconformances and changes. 7. Approve and release technical and data management reports. 8. Ensure that analysts and technicians maintain sample custody in the laboratory. 9. Approve project specific laboratory quality assurance plans. 10. Coordinate management of projects through technical supervisors. I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Date: 12/22/95 Section 4.0 Revision 0.01 Page 5 of 10 4.2.6 Group Supervisors/Leaders affect data quality by fulfilling responsibilities to: 1. Serve as the lead analyst within the specific sections. 2. Lead the training of analysts in laboratory operations and analytical procedures. 3. Organize and schedule analyses with consideration for sample holding times. 4. Implement data verification procedures. 5. Assign duties to analysts as data validators. 6. Prepare data summaries for review by the Laboratory Operations Manager. 7. Evaluate instrument performance and supervise instrument calibration and preventive maintenance programs. 8. Report noncompliance situations in regard to the project to the Laboratory Managers or laboratory Quality Assurance Officer, as appropriate. 4.2.7 Analysts are responsible for tasks identified in the scope of work. They perform the laboratory technical activities within these tasks. The duties of analysts are to: 1. Assist in planning for each phase of their tasks and in defining objectives and activities. 2. Respond to work plan revisions related to their tasks. 3. Advise the project manager of progress, needs, and potential problems of their tasks. 4. Train and qualify alternate analysts in specified laboratory QC and analytical procedures. 5. Verify that laboratory QC and analytical procedures are being followed as specified. 6. Review sample QC data at least daily. This includes examination of raw data such as chromatograms (and checking of calculations for a minimum of 10% of the samples analyzed) as well as an inspection of reduced data, calibration curves, and laboratory notebooks. 7. Inform project managers if the daily review indicates a decline in data quality and implement corrective action. 4.2.8 The Sample Custodian serves as sample coordinator for the entire laboratory. :\lqapreva\sect4 .doc Responsibilities are to: 1. Sign for incoming field samples and verify the data entered on the chain-of-custody forms. 2. Enter the sample information into the computerized Laboratory Information Management System for tracking and reporting. 3. Generate computerized sample analysis and data entry form_s (SADEF). 4.3 4.4 Date: 12/22/95 Section 4.0 Revision 0.01 Page 6 of 10 4. Transfer samples and tracking forms to laboratory project analysts. TRAINING AND ORIENTATION Each new permanent employee receives a four part orientation: a human resources orientation, a safety department orientation, a quality assurance department orientation, and a supervisory orientation. The human resources orientation involves matters of immediate personal concern such as benefits, salary, and company policies. The safety department orientation is an in-depth examination of the Pace Chemical Hygiene Plan and safety program, which are consistent with the requirements of OSHA's Hazard Communication Program (29 CFR 1910.1200). The Quality Assurance orientation provides the new employee with information on the Pace QA program through a brief introduction to the QA manual and SOPs, acceptable record keeping practices, and the individual's responsibility with respect to the quality assurance program. The new employee's Group Supervisor provides the employee with a basic understanding of the · role of the laboratory within the structure of Pace, Inc. and the basic elements of that individual's position within the laboratory. Temporary employees receive the same orientation as permanent staff with the exception of the Human Resources orientation. The training of a new employee concentrates on his/her scientific background and work experience to provide the employee with a level of competence so that the individual will be able to function within the defined responsibilities of his/her position ASAP. Training is a process used to assist laboratory personnel in their professional development. The training techniques utilized include: On-the-job training Lectures Programmed learning Conferences and seminars Short courses Specialized training by instrument manufacturers • Participation in check-sample or proficiency sample programs. Group Supervisors shall be responsible for providing documentation of training and proficiency for each employee under their· supervision. The Training Documentation File indicates what procedures (SOPs) a technician is capable of performing either independently or only with supervision. The files shall also include examples demonstrating performance of passing QC samples. The Group Supervisor is responsible for keeping a training documentation file for each person under their supervision which is updated and current. The QA department shall maintain a file for each technical employee. These files shall include a current curriculum vitae or resume. LABORATORY SAFETY Sample receiving areas and laboratories shall be equipped with suitable hoods, respirators, protective clothing and eye wear, gloves, barrier creams and or other :\lqapreva\sect4.doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 4.5 Date: 12/22/95 Section 4.0 Revision 0.01 Page 7 of 10 measures to prevent or m1rnm1ze staff contact with hazardous substances. Safety equipment such as eyewash stations, drench showers, spill adsorbents and neutralizers, fire extinguishers, first aid materials, and breathing oxygen shall be available. As a matter of policy, Pace shall not accept known initiator explosives, known dioxin-contaminated materials or unusual biohazard materials except where a specific Pace facility has been designed to safely handle high hazard samples. Pace shall accept nitroaromatics and nitroamines providing that the client makes provisions for disposal of samples with a positive explosive identification. A• laboratory staff member shall be designated as Safety Manager by the General Manager. The Safety Manager prepares and maintains safety-related SOPs, conducts safety and occupational health orientation, training and review sessions as required, and maintains up to date familiarity with safety and occupational health issues pertinent to the laboratory. The Safety Manager prepares and maintains educational programs as required to comply with state and federal "right to know" legislation. The Safety Manager or his designee shall conduct an orientation session with each new staff member to familiarize him/her with routine and emergency safety procedures and equipment. Eye protection and a lab coat shall be issued to the employee. A respirator will be issued, as required, after respiratory protection training. A tour of the laboratory shall be conducted. During the tour, needs for eye, skin, and respiratory protection shall be discussed as well as the use of safety glasses, face shields, goggles, partial and full-face respirators, ventilated work areas, fume hoods, gloves, barrier creams, and Tyvek coveralls. The location of eye wash stations, drench showers, fire extinguishers, and first aid equipment shall be shown to the employee and their use shall be described or demonstrated. Fire and spill notification, emergency procedures, and evacuation stations shall be taught during this . session. The orientation concludes with an introduction to potential chemical hazards and the Material Safety Data Sheets (MSDS). MSDS shall be made available for review. Employees shall be responsible for their own safety. The Operations Manager and Group Supervisors may require that certain levels of protective equipment be worn when in their judgment it is appropriate. Failure of an employee to wear required protective equipment will result in immediate disciplinary action. SECURITY AND CONFIDENTIALITY Three tiers of security shall be maintained within Pace for the purpose of controlling external influences on samples, analytical processes, and data. These security procedures help ensure the completeness, representativeness, accuracy, and precision of analytical results. :\Jqapreva\sect4.doc Date: 12/22/95 Section 4.0 Revision 0. O 1 Page 8 of 10 The first tier of security maintained shall be controlled access to laboratory buildings. Exterior doors to laboratory buildings shall remain either locked or continuously monitored by a Pace staff member. Keyless door-lock combinations (and computer access codes/logins) shall be changed every time an employee terminates employment at Pace. Posted signs shall direct visitors to the reception office and mark all other areas as off limits to unauthorized personnel. All visitors to the facilities must sign the Visitor's Logbook maintained by the receptionist. All visitors shall be accompanied by a staff member during the duration of their stay on the premises. The staff member shall escort the visitor back to the reception area at the end of his/her visit where he/she shall sign out in the Visitor's Logbook. Prior to departure of the last staff member at the close of each· day, all windows shall be locked and all doors checked and locked by the last staff member. The second security level shall be within the facility and may be designated as required by the Operations Manager in consultation with the General Manager. Individual Operations Manager or Group Supervisors may close specific areas under their responsibility to entry by unauthorized persons. A list of authorized persons shall be prepared and signed by the General Manager. "Closed Areas" shall be designated by prominent postings at all points of access. The final tier of security shall be comprised of specific secure areas for sample, data and client report storage which shall be lockable within the facilities, and to which access shall be limited to specific individuals or their designees. Security of sample storage areas shall be the responsibility of the Sample Manager. Security of samples and data during analysis and data reduction shall be the responsibility of Group Supervisors and Operations Manager. Security of client report archives shall be the responsibility of the Quality Assurance Officer or an appropriate designee. These secure areas will be locked whenever these individuals or their designees are not present in the facility. Designated laboratory sample storage locations are designed to limit access to authorized personnel only, and provisions for lock and key access shall be provided. No samples are to be removed without authorization, which consists of having a work list requesting analysis on an aliquot. No samples are to be removed without filling out the associated chain-of-custody records. Standard business practices of confidentiality shall apply to all documents and information regarding client analyses. Specific protocols for handling confidential documents are described in Pace SOPs. Additional protocols for internal identification of samples and data by number only shall be implemented as required under contract-specific Quality Assurance Project Plans. :\lqapreva\sect4.doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I :\lqapreva\sect4.doc Figure 4.1 Date: 12/22/95 Section 4.0 Revision 0.01 Page 9 of 10 Pace Analytical Services, Inc. Organizational Structure Chairman of the Board I PresidenUCEct I I Corporate Offic: I Laboratories I I Quality Assuran ie Human Quality & Office r- Resources >--Technical Oirecti n Sales & Information Inorganic Organic Marketing r-r-Systems Laboratory ,--Laboratory Finance Field Marketing & -Services ,-,-Client Servicei Support la{ Accountin Services - Client Services :\lqapreva\sect4 .doc Figure 4.2 Pace Analytical Services, Inc. Laboratory Organizational Structure General Manager QA Officer Operations Manager Group Services Analysts and Technicians Field Services Date: 12/22/95 Section 4.0 Revision 0.01 Page 10 of 10 Support Services I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 5.0 QUALITY ASSURANCE OBJECTIVES Date: 12/22/95 Section 5.0 Revision 0.01 Page 1 of 12 The purpose of this Quality Assurance Plan is to define procedures for the documentation, evaluation, validation, and reporting of data. The objective is to provide a uniform basis for sampling, sample handling, instrument maintenance and calibration, methods control, performance evaluation and analytical data generation and reporting. Specific procedures to be used for sampling, chain of custody, calibration of field instruments (pH, conductivity meters, etc.), laboratory analysis, reporting, internal quality control, audits, preventive maintenance, and corrective actions are described in specific sections of this plan. This section addresses the objectives of precision, accuracy, representativeness, completeness, and comparability (PARCC) which are used to assess whether the data meet the established DQOs (Data Quality Objectives), that are based upon the intended end use of the data. The quality assurance objective of the laboratories is to provide data of known and documented quality. Data quality is assessed by precision, accuracy, representativeness, completeness, and comparability. The QA protocols used in the laboratories for the majority of analyses performed are taken from the following sources: EPA Contract Laboratory Program's Statement of Work (Organics and lnorganics), 40 CFR 136 methodologies, and SW 846 methodologies which contain detailed descriptions of the quality control measures routinely employed by Pace Analytical Services, Inc .. As stated, the objective of the Quality Assurance Program for the laboratory is to provide data of known quality. To accomplish this, Pace will: 5.1 • Maintain an effective, on-going QNQC program that measures and verifies laboratory performance. • Provide a quality organization independent of the pressures of project performance with the responsibility and authority for auditing and recommending corrective action. • Provide a quality organization with clear paths of communication with management. • Provide sufficient flexibility to allow controlled changes in routine methodology to meet client specific data requirements contained in project-specific quality plans. • Recognize as soon as possible and provide correction for any factors which adversely affect data quality. • Monitor operational performance of the laboratory on a routine basis and provide corrective action as needed. • Maintain complete records of sample submittal, raw data, laboratory performance, and completed analyses to support reported data. LEVEL OF QA EFFORT The reliability of data generated in the laboratory will be evaluated at the 99% confidence level (mean +/-3 standard deviations) for control and at the 95% confidence level (mean +/-2 standard deviations) for warnjng. Precision of analyses will be evaluated using sample duplicates and matrix spike duplicates. Analytical accuracy will be monitored using recovery of analytes from surrogate spikes, matrix spikes, EPA reference check standards (when available) and/or lab control samples, and Performance Evaluation (PE) samples. :\lqapreva\sectS.doc 5.2 PRECISION AND ACCURACY Date: 12/22/95 Section 5.0 Revision 0.01 Page 2 of 12 Precision measures the reproducibility of repetitive measurements. It is strictly defined as the degree of mutual agreement among independent measurements. as the result of repeated application of the same process under similar conditions. Analytical precision is a measurement of the variability associated with duplicate (two) or replicate (more than two) analyses of the same sample in the laboratory and is determined by analysis of laboratory duplicates. Total precision is a measurement of the variability associated with the entire sampling and analysis process. It is determined by analysis of duplicate or replicate field samples and incorporates variability introduced by both the laboratory and field operations. Precision data must be interpreted by taking into consideration these possible sources of variability. Duplicate (two) samples or spiked samples are analyzed to assess field and analytical precision as required under certain programs (e.g., Air Force tasks), and the results are assessed using the relative percent difference (RPO) between duplicate measurements. Precision objectives are presented for each analytical method in the corresponding Pace Standard Operating Procedure (SOP). Accuracy is a statistical measurement of correctness and includes components of random error (variability due to imprecision) and systematic error. It therefore reflects the total error associated with a measurement. A measurement is accurate when the value reported does not differ from the true value or known concentration of the spike or standard. Analytical accuracy is typically measured by determining the percent recovery of known target analytes that are spiked into a field sample (a surrogate or matrix spike) or reagent water (laboratory control sample [LCS] or QC check sample). Surrogate compound recovery is reported and is used to assess method performance for each sample analyzed for volatile and semivolatile organic compounds. The stated accuracy objectives apply to spiking levels at least five times the method detection limi_ts (MDLs) or background concentration. Both accuracy and precision are calculated for analytical batches, and the associated sample results must be interpreted by considering these specific measures. Calculation of precision and accuracy to measurement sample results is discussed in the QC section of each SOP. The QA objectives for precision and accuracy are to achieve the QC acceptance criteria specified in the proposed analytical procedures. For the organic and inorganic procedures, the precision and accuracy guideline requirements are specified in the individual methods. · Field blanks and duplicates are collected and analyzed to assess field sampling activities. The results check procedural contamination and/or ambient conditions at the site. Due to the extensive number of organic parameters and potential matrices, the development of precision and accuracy objectives and control limits for every matrix is difficult. This is typically done with (1) matrix spike and matrix spike duplicate compounds which are added to selected samples before extraction and analysis, and/or :\lqapreva\sect5.doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 5.3 Date: 12/22/95 Section 5.0 Revision 0.01 Page 3 of 12 (2) surrogate spike compounds which are added to every sample, before extraction and analysis. Although the surrogate and matrix spike analyses do not provide statistically valid statements about precision and accuracy for every compound in a sample, they do give the data reviewer enough information to make judgments about precision and accuracy on a sample-by-sample basis. Inorganic precision and accuracy data are determined by using duplicate or matrix spike duplicate samples {precision}, matrix spike and laboratory control samples (accuracy). The following procedure is used: For a duplicate (or matrix spike duplicate) sample analysis, at least one duplicate (or MSD) sample is analyzed per sample matrix type (e.g., water, soil) and concentration (e.g., low, medium) per batch of samples or for each 20 samples received, whichever is more frequent, or as specified by state/project requirements. Samples identified as field blanks can NOT be used for duplicate (or MSD) samples analyses. If two analytical methods are used to obtain the reported values for the same element for a batch of samples (i.e., ICP, GFAA}, duplicate samples will be run by each method. The relative percent difference (RPD) for each component is calculated for later use during data assessment. The QC limits for accuracy and precision are developed based upon laboratory derived data. When applicable, interlaboratory control limits established by the EPA CLP are used to judge acceptability of data generated by the laboratories. Where EPA acceptability criteria does not exist for a given method being utilized for the first time, the laboratories will establish control limits derived from a minimum of four data points. Until verified by a statistically significant data population, the control limits will be considered as advisory limits only and will not automatically initiate a rerun or reanalysis criteria if they are not met. Representative QC objectives for selected organic parameters are listed in Tables 5.1 to 5.6. Similarly representative QC objectives -for selected metal and inorganic parameters are listed in Tables 5.7 to 5.8. Generally, QC acceptance limits are laboratory specific, having been statistically derived from an individual laboratory's data. QC objectives for a specific laboratory will be included in a project specific QAPP or for general information as a facility specific addendum to this document. COMPLETENESS Completeness is a measure of all information necessary for a valid scientific study. For completeness, it is expected that the methodology proposed for chemical characterization of the samples collected will provide data meeting QC acceptance criteria following standard laboratory data review and validation for at least 95% of all samples collected. Completeness may also be defined as a comparison of the number of tests successfully completed (with acceptable QC) to the number of tests requested. Discrepancy reports are completed to provide explanation when QC criteria are not met. Every attempt will be made to generate completely valid data. However, it is recognized that some samples will exhibit highly contaminated matrices necessitating multiple :\lqapreva\sectS.doc 5.4 5.5 Date: 12/22/95 Section 5.0 Revision 0.01 Page 4 of 12 analyses and/or extensive dilutions. As a result of these atypical applications, recoveries and MDLs or Rls, as applicable, may be deemed questionable based on internal QC results by the ex1ernal data validation process. The objective will be to have 95% completeness on samples unaffected by matrix interferences. For uncontaminated background samples and first time samples not showing interferences, completeness should be 100% with a mandatory requirement for reanalysis of these critical samples if objective is not met. 5.3.1 Random Error EPA has established (preamble to 40 CFR Part 136, Vol. 49, No. 209, October 26, 1984) that there is a 5% probability that the results obtained for any one analyte will exceed the control limits established for the test due to random error. As the number of compounds measured increases in a given sample, the probability for statistical error also increases. The number of compounds present in numerous EPA methods (e.g., GC/MS methods 82408 and 82708, and metals included in ICP method 6010A) increases the probability that one or more analytes will not meet acceptance criteria to significantly more than the 5% per analyte frequency. The number of target analy1es included in these tests can be used to show that a minimum of four to seven target analy1es will exceed the control limits established for these methods due to the statistical probability for random error. The establishment of QC criteria that are not consistent with the measurement of the quality objectives for which they are intended should be discouraged. REPRESENTATIVENESS Representativeness is a qualitative element that is related to the ability to collect a sample that reflects the characteristics of that part of the environment that is to be assessed. Sample representativeness is dependent on the sampling techniques used and is considered individually for each project. It is specifically addressed in the work plan. Representativeness is a measure of how closely the measured results reflect the actual concentration or distribution of the chemical compounds in the sample. Sample handling protocols (e.g., collection, storage, preservation and transportation) have been developed to preserve the representativeness of the samples. Proper documentation will establish that protocols have been followed and sample identification and integrity assured. Every attempt will be made to ensure that the aliquots taken for analysis are homogeneous and representative of the samples received. COMPARABILITY Comparability is also considered during preparation of a site specific work plan. The objective of comparability is to ensure that results of similar activities conducted by different parties are comparable. This often involves the use of two independent laboratories on a project or site, whereby the second laboratory is used to confirm a pre- :\lqapreva\sect5.doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 5.6 5.7 Date: 12/22/95 Section 5.0 Revision 0.01 Page 5 of 12 established percentage of sample analyses. Pace uses EPA-approved or-other methods and procedures to ensure comparability with data from previous or following studies. Pace participates in external and interlaboratory performance evaluation (PE) studies as additional means of establishing comparability in the laboratory. TRACEABILITY Traceability is the extent to which results can be substantiated by hard-copy documentation. Traceability documentation exists in two forms: that which links final numerical results to authoritative measurement standards, and that which explicitly describes the history of each sample from collection to analysis. Refer to the sections on sample custody and records management for more specifics on Pace procedures. QUALITY ASSURANCE PROJECT PLAN EXCEPTIONS Due to the unknown nature of environmental samples prior to analysis, Pace has minimal control over analytical and quality control complications which arise from unique sample matrix conditions. These conditions may include such items as: highly concentrated samples containing target compounds of interest and/or non-target components; extremes in sample pH, viscosity, and solubility; and high organic content (both natural and synthetic). Each of these conditions presents a variety of challenges to the laboratory. Most often these extremes in sample matrix composition necessitate the laboratory to employ dilution techniques in order to change the sample state into one which can be analyzed by the desired protocol. Unfortunately, dilution techniques raise reporting limits (Rls) and often adversely impact the surrogate standard and matrix spiking acceptance criteria. The laboratory has the responsibility to clearly identify cases where matrix interferences preclude the generation of "compliant" data. This is done by demonstrating through reproducibility (i.e., reanalysis of the affected sample) that the quality control measurement failure resulted from unique sample matrix conditions beyond the control of laboratory, and not as a result of laboratory error. For example, in situations where the surrogate standard recoveries fall outside of control limits, samples are re-extracted and/or re-analyzed. Similar "non-compliant" results in the reanalysis indicate that it is something inherent to the sample which prevented the laboratory from reporting results deemed method compliant under data validation criteria. Analytical projects containing particularly "dirty" samples (i.e., highly contaminated) will often fail to meet pre-established QA completeness goals (set forth in the QAPP) when prior site history does not reveal the potential for excessive values. Again, while the laboratory perfomns all analytical testing by the prescribed protocols, the results obtained may not meet validation criteria as a result of elevated Rls or the frequency at which surrogate and matrix spikes failed to meet acceptance limits. In cases where the laboratory is unable to meet QC criteria because of sample matrix complications beyond their control, results which are fiagged "qualified" or "rejected" by data validation guidelines are often still "useable" by the end user of the data. :\lqapreva\sect5.doc 5.8 Date: 12/22/95 Section 5.0 Revision 0.01 Page 6 of 12 · Pace is committed to adhering to method requirements and program quality control applications as established by our client and will work rigorously to provide data which is of the highest quality possible. However, the uncertainties associated with environmental samples do not allow Pace to assume responsibility for conditions beyond our reasonable control which directly impact the "validity" versus the usability of the associated analytical data generated. PERSONNEL QUALITY OBJECTIVES Pace is committed to the philosophy that quality operations result from quality planning, design, and work performance by skilled operational personnel. Pace's policy is to perform its varied types of technical work in accordance with standard quality assurance practices such as Good Laboratory Practices (GLP) and the EPA Contract Laboratory Program (CLP),· as well as other appropriate regulatory agency guidelines and requirements. Each laboratory within Pace has a Quality Assurance Officer responsible for maintenance of standard operating procedures, laboratory audits, performance evaluations, federal and state certifications and quality assurance documentation. Each laboratory worker is responsible for checking standard operating procedures when necessary; following these procedures during routine analyses; recording quality control information required by those procedures in the proper location, and taking appropriate corrective action including suspending analyses when quality control criteria are not met. :\lqapreva\sectS.doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Table 5.1 Date: 12/22/95 Section 5.0 Revision 0.01 Page 7 of 12 Representative Spike Recovery Acceptance Criteria for. Volatile Organic Analysis by Methods 80108 and 8020A MS¾R LCS¾R (SW-846) (Statistical)* Analyte Water/Soil Water SQl! Method 8010A 1, 1-Dichloroethene 28-167 63-145 47-137 Chloroform 49-133 70-137 66-140 Carbon tetrachloride 43-143 72-138 63-143 1,2-Dichloroethane 51-147 74-138 60-157 Trichloroethene 35-146 75-147 63-152 Tetrachloroethene 26-162 79-134 72-138 Chlorobenzene 38-150 76-126 65-136 1,4-Dichlorobenzene 42-143 70-123 64-127 Method 8020 Benzene 39-150 74-135 43-156 Chlorobenzene 55-135 74-130 80-126 1,4-Dichlorobenzene 42-143 70-125 75-120 ¾R (Statistical)* Surrogate Water SQl! · Bromochloromethane 54-115 60-109 1,4-Bromofluorobenzene 70-125 68-112 • Statistically derived acceptance limits will vary by individual laboratory operation. :\lqapreva\sect5.doc Table 5.2 Date: 12/22/95 Section 5.0 Revision 0.01 Page 8 of 12 Representative Spike Recovery Acceptance Criteria for Purgeable Petroleum Hydrocarbon Analysis (California LUFT Method) Analyte Benzene Toluene Ethylbenzene Xylene Surrogate Bromofluorobenzene MS¾R (SW-846) Water/Soil LCS¾R (Statistical)* Wru.er Soil 39-150 46-148 32-160 32-160 80-110 80-110 83-113 83-113 ¾R (Statistical)* Wfiler S.oi.l 70-113 51-120 Table 5.3 49-103 49-103 52-106 56-104 Representative Spike Recovery Acceptance Criteria for Extractable Petroleum Hydrocarbon Analysis (California LUFT Method) Analyte Diesel Surrogate 2-Fluorobiphenyl o-Terphenyl MS¾R (Advisory) Water/Soil 50-150 LCS¾R (Statistical)* Ware! Soil 62-122 57-123 ¾R (Statistical)* Wi!.te..r S.oi.l 53-131 41-149 38-128 37-169 • Statistically derived acceptance limits will vary by individual laboratory operatio_n. :\lqapreva\sect5.doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Table 5.4 Date: 12/22/95 Section 5.0 Revision 0.01 Page 9 of 12 Representative Spike Recovery Acceptance Criteria for Pesticides/PCB Analysis by Method 8080A and CLP SOW MS¾R (SW-846) Water/Soil MS¾R (CLP) LCS¾R (Statistical)* Analyte Water 5_oj_) Ws!.te.r 5_oj_) Aldrin gamma-BHC DDT Dieldrin Endrin Heptachlor Surrogate Tetrachloro-m-xylene Decachlorobiphenyl 42-122 32-127 25-160 36-146 30-147 34-111 40-120 56-123 38-127 52-126 56-121 40-131 ¾Recovery (SW-846 Statistical)* Water Soil 31-121 29-153 32-108 56-125 Table 5.5 34-132 46-127 23-134 31-134 42-139 35-130 32-128 33-135 39-135 37-139 42-138 34-130 ¾Recovery (CLP Advisory) Water Soil 60-150 60-150 60-150 60-150 Representative Spike Recovery Acceptance Criteria for Volatile Organic Analysis by Method 82408 and CLP SOW 32-104 31-103 23-125 36-108 37-115 34-106 Analyte MS¾R (SW-846) Water/Soil MS¾R (CLP) LCS¾R (Statistical)* Ws!.te.r 5_oj_) Water 5_o.i.! 1, 1-Dichloroethene Trichloroethene Chlorobenzene Toluene Benzene Surrogate 59-155 71-157 37-160 47-150 37-151 Toluene-d8 4-Bromofluorobenzene 1,2-Dichloroethane 61-145 71-120 75-130 76-125 76-127 ¾Recovery (SW-846) Water Soil 88-110 86-115 76-114 81-117 74-121 70-121 59-172 62-137 60-133 59-139 66-142 63-123 77-119 81-117 79-115 83-119 ¾Recovery (CLP) Watfil: Soil 88-110 86-115 76-114 84-138 59-113 70-121 • Statistically derived acceptance limits will vary by individual laboratory operation. :\lqapreva\sect5.doC 54-126 77-113 85-115 84-114 83-119 Table 5.6 Date: 12/22195 Section 5.0 Revision 0.01 Page10of12 Representative Spike Recovery Acceptance Criteria for Semivolatile Organic Analysis by Method 8270B and CLP SOW MS¾R MS¾R LCS¾R (SW-846) (CLP) (Statistical)* Analyte Water/Soil ~ w ~ fuill 1 , 2, 4-Trichlorobenzene 44-142 39-98 38-107 53-95 29-103 Acenaphthene 47-145 46-118 31-137 65-101 45-98 2,4-Dinitrotoluene 39-139 24-96 28-89 63-96 46-92 Pyrene 52-115 26-127 35-142 56-111 49-100 N-Nitroso-di-n-propylamine D-230 41-116 41-126 67-103 40-101 1.4-Dichlorobenzene 20-124 36-97 28-104 48-84 22-95 Pentachlorophenol 14-176 9-103 17-109 40-139 43-123 Phenol 5-112 12-110 26-90 51-105 34-102 2-Chlorophenol 23-134 27-123 25-102 54-106 36-100 4-Chloro-3-methylphenol 22-147 23-97 26-103 64-106 47-99 4-Nitrophenol D-132 10-80 11-114 40-130 49-108 %Recovery %Recovery (SW-846) (CLP) Surrogate ~ S.Qil ~ Soil 2-Fluorophenol 21-100 25-121 21-110 25-121 Phenol-d5 10-94 24-113 10-110 24-113 Nitrobenzene-d5 35-114 23-120 35-114 23-120 2-Fluorobiphenyl 43-116 30-115 43-116 30-115 2,4,6-Tribromophenol 10-123 19-122 10-123 19-122 Terphenyl-d,. 33-141 18-137 33-141 18-137 * Statistically derived acceptance limits will vary by individual laboratory operation. :\lqapreva\sectS.doc I I I I I I I I I I I I I I I I I I I Date: 12/22/95 I Section 5.0 Revision 0.01 Page 11 of12 I Table 5.7 I Representative Spike Recovery Acceptance Criteria for Metals Analysis by SW-846 and CLP SOW I LCS¾R I (Statistical)* earn meter MetbQd Analyte MS¾R ~ ~ Metals-I CPS 6010A/CLP Aluminum 75-125 85-114 51-149 Antimony 75-125 66-135 17-183 I Arsenic 75-125 89-116 48-152 Barium 75-125 87-109 68-132 Beryllium 75-125 75-111 63-137 I Boron 75-125 75-116 Cadmium 75-125 82-123 59-141 Calcium 75-125 88-113 67-133 I Chromium 75-125 88-116 60-140 Cobalt 75-125 90-113 63-137 Copper 75-125 90-112 61-139 I Iron 75-125 86-125 62-138 Lead 75-125 88-116 55-145 Magnesium 75-125 89-111 62-138 I Manganese 75-125 91-112 68-132 Molybdenum 75-125 88-113 61-139 Nickel 75-125 90-119 59-141 I Potassium 75-125 80-115 64-136 Selenium 75-125 85-115 50-150 Silver 75-125 79-120 43-157 I Sodium 75-125 86-112 52-148 Strontium 75-125 80-120 Thallium 75-125 91-120 48-152 I Tin 75-125 80-120 Titanium 75-125 80-120 Vanadium 75-125 90-113 66-134 I Zinc. 75-125 93-121 55-145 Metals-GFM 7041/CLP Antimony 75-125 80-120 17-183 I 7060A/CLP Arsenic 75-125 74-126 48-152 7421/CLP Lead 75-125 74-127 55-145 7740/CLP Selenium 75-125 68-119 50-150 I 7841/CLP Thallium 75-125 79-130 48-152 7470A/CLP Mercury 75-125 75-118 52-148 I Other Metals 7196A Chromium(VI) 75-125 87-116 86-115 CADHS Organic lead 75-125 60-145 71-128 • Statistically derived acceptance limits will vary by individual laboratory operation . !I :\lqapreva\sectS.doc Example Table 5.8 Date: 12/22/95 Section 5.0 Revision 0.01 Page 12 of 12 Representative Spike Recovery Acceptance Criteria for General Chemistry Analyses Analyte MS%R Cyanide 75-125 Total alkalinity (titration} 75-125 Total alkalinity (Automated) 75-125 Bromide (IC) 75-125 Chloride (IC) 75-125 Chloride (Automated) 75-125 Fluoride (ISE) 75-125 MBAS (colorimetric) 75-125 Nitrate (IC) 75-125 Nitrate (Automated) 75-125 Nitrite (IC) 75-125 Nitrite (Automated) 75-125 Nitrate/Nitrite (Automated) 75-125 Oil & grease 75-125 Total phenolics (Automated) 75-125 a-Phosphate (IC) 75-125 a-Phosphate (Automated) 75-125 Total phosphate (Automated)75-125 Sulfate (IC) 75-125 TKN 75-125 TRPH 75-125 LCS¾R (Statistical}* Water Soil 76-125 85-117 78-110 86-108 81-112 79-127 84-122 79-114 81-110 83-119 79-116 92-110 85-117 63-121 67-127 77-115 75-125 85-114 86-113 75-125 69-125 41-159 • Statistically derived acceptance limits will vary by individual laboratory operation. :\lqapreva\sectS.doc I I I I I I I I I I I I I I I I I I I E I I I I I I I I I I I I I 6.1 6.2 6.0 SAMPLING PROCEDURES INTRODUCTION Date: 12/22/95 Section 6.0 Revision 0. O 1 Page 1 of 9 Obtaining representative samples and maintaining their integrity are critical parts of any monitoring program. Analytical methods have been standardized but the results of analyses are only as good as the sampling and the sample preservation methods. Defining the magnitude and the nature of an environmental problem requires collecting representative samples for laboratory analysis and data evaluation. The careful collection of samples is key to obtaining an accurate assessment of the site's environmental impact, and to developing the appropriate remedial solution. Defining in detail the numerous available sampling procedures and their associated quality elements applicable to environmental testing is beyond the scope of this document. Quality elements required to meet the DQOs for a given sampling event must be contained in a project specific sampling plan or within an overall site work plan. The plans should present the best approved techniques currently available for sampling and sample preservation. In sampling, the objective is to remove a small portion of an environment that is representative of the entire body. Once the sample is taken, the constituents of the · sample must stay in the same condition as when collected. The length of time that these constituents will remain stable is related to their character and the preservation method used. Since preservation methods relate to the parameters to be analyzed, these techniques are classified by parameter. SAMPLING SERVICES. Various Pace locations provide a variety of sampling services. A well-defined communication mechanism is critical to obtaining samples which are representative of site conditions. Figure 6.1 lists minimum elements which must be established, communicated, and followed during each phase of the sampling and analysis project. Listed below are the types of sampling events for which Pace can provide services. 6.2.1 Ground Water Monitoring Collection and analysis of grourid water samples from sanitary landfills, Superfund sites, abandoned hazardous waste dumps and spill sites involves the use of an extensive array of state-of-the-art sampling equipment with the ability to pre-pump and sample wells of all sizes, to depths of more than 200 feet. 6.2.2 Waste Water Monitoring Collection of samples for routine wastewater monitoring; special compliance monitoring for metals, cyanide and total toxic organics (TTO); NPDES permit application monitoring; and more, as required by local, state of federal agencies. :\lqapreva\sect6.doc ; ,- 6.3 6.2.3 Hazardous Waste Sampling Date: 12/22/95 Section 6.0 Revision 0.01 Page 2 of 9 I I Pace can conduct an inventory, collect and analyze samples, and repackage and II properly label hazardous waste for shipment. U 6.2.4 Flow Monitoring Monitoring flow in most types of discharges through installation of wiers or flumes and also determine flow using fluormetric dye tracing methods. 6.2.5 Soil & Soil Gas Sampling I I Technicians statistically develop soil sampling programs to identify pollution I problems. Typically this involves collecting soil samples for volatile and semi- volatile organics, inorganics, hazardous waste constituents and most other I chemical parameters of concern. Also collecting and analyzing soil gas samples. 6.2.6 PCB Services Collection of samples from transformers, electrical switches and capacitors to be - tested for the presence of P~Bs. Handling of PCB-related spills includes the collection and analysis of wipe, swab or soil samples. 6.2.7 Ambient Air Monitoring and Stack Emission Testing Pace has experience in providing sampling as specified in the Toxic Organic (TO method series) protocols for ambient air monitoring, along with NIOSH and AIHA specified applications. FuH capability stack sampling and testing (e.g., VOST, impinger, etc.) is av_ailable for process optimization and emission monitoring. FIELD SUPPORT I I I I I Pace provides shipping containers, custody documents, custody seals, sample bottles, I labels, chemical preservatives for water samples, "blue ice" packs to maintain thermal preservation, and trip and field blanks to support field sampling events. Tables 6.1, 6.2 and 6.3 list sample container types, preservatives and holding times. Certain Pace I locations can provide pick up and delivery services to their clients. Upon receipt of the field samples at the laboratory, Pace ensures that sample bottles I are maintained according to preservation requirements and that sample storage conditions do not contribute to the presence of test analytes in the samples. I I :\lq apreva \sect6. doc I D I I I I I I I I I I I 6.4 6.5 6.6 Pace Shipping Containers Date: 12/22/95 Section 6.0 Revision 0.01 Page 3 of 9 Pace typically uses commercial coolers for the transport of environmental samples from the field to the laboratory. Chain-of-custody seals and forms, employed for each cooler packed at Pace ensure complete documentation and provide evidence of unbroken custody of the cooler contents. Coolers meet or exceed all protocol requirements (i.e., DOT, USEPA, ASTM) for shipping. Coolers are prepared at the laboratory to provide the client with all of the sample containers needed for the analyses required by a project. PRESERVATION Pace provides the required chemical preservatives for water samples and "blue ice" packs, for thermal preservation when the samples are shipped back to the lab. High quality, reagent grade chemical preservatives are used. The ice packs are supplied pre-frozen or at ambient temperatures based upon the client's needs. It is the responsibility of those collecting the samples to properly use these materials and ensure that proper preservation techniques are performed and preservative requirements are met. Upon receipt of samples at the laboratory, the temperature of each cooler is measured. and recorded on the chain of custody documents. Similarly, the pH of bottles to which chemical preservative was added is measured (with the exception of sample collected for volatile organic compounds), and the check recorded. A disposable pipette is used to remove an aliquot of the sample for the pH check. When deviations from the required chemical or thermal preservation are noted, the project manager is notified, and clients may become involved in determining a course of action to follow. Water samples for GC and GC/MS volatile aromatics determinations are monitored for pH just prior to analysis, at which time the pH of each individual sample bottle used is checked. The portion of sample used for the analy1ical determination is removed from the vial prior to checking the sample's pH. Sample pH measurements are recorded on laboratory chronicles as they are taken. SAMPLE CONTAINERS Pace provides precleaned sample containers in the shipping containers for sample collection. Used sample bottles are never used by the laboratory. Vendor prepared (certified contaminant-free) containers can be provided as projects necessitate. SAMPLE RECEIPT SCHEDULE Samples are normally delivered to the Pace facility during normal business hours within one day following field sampling unless different arrangements are made in advance with an authorized Pace representative. Shipping containers received at the laboratory on business holidays, weekends or after normal work hours will be placed in the walk-in refrigerator and opened on the next regular business day unless prior arrangements are made in advance for that day's receipt and log-in. :\lqapreva\sect6.doc Client Appro-.e safTl)ting schedule. o-.ersee project plan ~ Conm.micate safll)ling requirerrents, rate of sample de~-.ery, etc. M:>nitoc corrm.inication betv.een tab & fiek1 personnel, remlin Informed on project rela~ issues :\lqapreva\sect6.doc Figure 6.1 Date: 12/22/95 Section 6.0 Revision 0.01 Page 4 of9 Pre-Sampling Communication Pace Sampling Personnel or Client Pace Laboratory Personnel Performing Sampling I Estabish Available I Capacity & Resources Sample Collection & Holding Pre-cleaned san-ple bottles presen.ed, labelled. packaged & sealed '°' shiprreot to site. Bottle shipping manifest form initiated and inluded t I Sclfl1)1e Bottle coolers I I Sarrples coDected I shipped to site I I presen.ed & packaged + I Chain of Custxty Initiated j + I Receipt In laboratory I I Shipment of satll)les I I to laboraby + J eon,,1ete Chain of Custody j I E>camine sartl)le for condition I t Project Mgr. comrunicates I Estabish Available discrepancies on COC: broken I Capacity & Resources sa"l)te bottles, etc. ' I SarrpleStorage I t I Re\itew hok::ling times, I prioritize in sal11'.)le stream I I I I I I I I I I I I I I I I I I D I I I I I I I I I I I I I TABLE 6.1 NAME Inorganic Tests: Acidity Alkalinity Ammonia Biochemical oxygen demand Bromide Biochemical oxygen demand, carbonaceous Chemical oxygen demand Chloride Chlorine, total residual Color Cyanide, total amenable to chlorination Fluoride Hardness Hydrogen ion (pH) Kjeldahl and organic nitrogen Metals: Chromium VI Mercury (SW846) Mercury (CLP, 200 series) Metals, except chromiumVI and mercury Nitrate N itra te-n itrite Nitrite Oil and grease Organic carbon Orthophosphate Phenols Phosphorus ( elemental) Phosphorus, total Residue, total Residue, filterable Residue, nonfilterable (TSS) Residue, Settleable Residue, volatile Silica :\lqapreva\sect6.doc Date: 12/22/95 Section 6.0 Revision 0.01 Page 5 of 9 List of Containers, Preservatives and Holding Times for Inorganic and Organic Analyses of Aqueous Samples: CONTAINER' PRESERVATION" MAXIMUM HOLDING TIME" Cool, 4°C P,G 14 days P,G Cool. 4°C 14 days P,G cool, 4°c, H,so, to pH<2 28 days P,G Cool, 4°C 48 hours P,G None Required 28 days P,G . Cool, 4°C 28 days P,G Cool. 4°C, H,so, to pH<2 28 days P,G None Required 28 days P,G None Required Analyze immediately P,G Cool, 4°C 48 hours P,G Cool, 4°C, NaOH to pH>12 14 days 0.6g ascorbic acid4 p None Required 28 days P,G HNO3, to pH<2, H2SO, to 6 months pH<2 P,G None Required Analyze immediately P,G Cool, 4°C, H2SO, to pH<2 28 days P,G Cool, 4°C 24 hours P,G HNO3 to pH<2 38 days in glass P,G HNO3 to pH<2 13 days in plastic 28 days P,G HNO3 to pH<2 6 months P,G Cool, 4°C 48 hours P,G Cool, 4°c, H,so, to pH<2 28 days P,G Cool, 4°C 48 hours G Cool, 4°C, H,so, to pH<2 28 days P,G Cool, 4°c, H2SO, to pH<2 28 days P,G Filter immediately, 48 hours Cool, 4°C G only Cool, 4°C, H2SO4 to pH<2 28 days G Cool, 4°C 48 hours P,G Cool, 4°C, H2SO, to pH<2 28 days P,G Cool, 4°C 7 days P,G Cool, 4°C 7 days P.G Cool, 4°c 7 days P,G Cool, 4°C 48 hours P,G Cool, 4°C 7 days p Cool, 4°C 28 days Date: 12/22/95 Section 6.0 Revision 0.01 Page 6 of 9 TABLE 6.1 (cont.) List of Containers, Preservatives and Holding Times for Inorganic and Organic Analyses of Aqueous Samples· NAME CONTAINER' PRESERVATION' MAXIMUM HOLDING TIME' /norganics Continued: Specific conductance P,G Cool, 4°C 28 days Sulfate P,G Cool, 4°C 28 days Sulfide P,G Cool, 4°C, add zinc 7 days acetate & sodium hydroxide to pH>9 Sulfite P,G None Required Analyze immediately Surfactants P,G Cool. 4°C 48 hours Turbidity P,G Cool, 4°C 48 hours Organic Tests: Oil and Grease G Cool, 4°C, HCI or H2SO4 28 days to pH<2 Organic carbon, Total P,G Cool, 4°C. HCI or H,so, 28 days (TOC) to pH<2 Purgeable Halocarbons G,Teflon-lined Cool, 4°C. 0.008% 14 days septum Na,s,o,' Purgeable Aromatic G,Teflon-lined Cool, 4°C, 0.008% 14 days Hydrocarbons septum Na S O 4 HCl5'6 2 2 J I Acrolein and acrylonitrile G,Teflon-lined Cool, 4°C, 0.008% 14 days septum Na2S203 • .A~just pH to 4-5 Phenols G, Teflon-lined cap Cool, 4 C, 0.008% 7 days until extraction, 40 days after Na,s,o, • extraction Benzidines G,Teflon-lined cap Cool, 4°C. 0.008% 7 days until extraction, 40 days after Na,s,oi' extraction Phthalate esters G,Teflon-lined cap Cool,4 7 days until extraction, 40 days after extraction Nitrosamines G,Teflon-lined cap Cool, 4°C, store in dark, 7 days until extraction, 40 days after 0.008% Naas,o,' extraction PCBs G,Teflon-lined cap Cool, 4 C 7 days until extraction, 40 days after extraction Nitroaromatics and cyclic G, Teflon-lined cap Cool, 4°C, store in dark, 7 days until extraction, 40 days after ketones 0.008% Na,s,o,' extraction Poly nuclear aromatic G,Teflon-lined cap Cool, 4°C, 0.008% 7 days until extraction, 40 days after hydrocarbons Nao'.s20,' extraction Haloethers G,Teflon-lined cap Cool, 4 C, 0.008% 7 days until extraction, 40 days after . . extraction Na,s,o, Chlorinated Hydrocarbons_ G,Teflon-lined cap Cool, 4°C, HCI or H2SO4 7 days until extraction, 40 days after extraction Dioxins and Furans G,Teflon-lined cap Cool, 4°C, 0.008% 7 days until extraction, 40 days after Na,s,o,' extraction Total organic halides (TOX) G, Teflon-lined cap Cool, 4°C, HCI or H,so, 28 days Pesticides G,Teflon-lined cap to pH <2 Cool, 4°C 7 days until extraction, 40 days after pH 5-9 extraction :\lqapreva\sect6.doc I I I I D I I I I I I I I I I I I I I D I I I I I I I I I I I I Table Footnotes: 1 Polyethylene (P) or glass (G) 2 Sample preservation should be performed immediately upon sample collection. 3 Holding times are based upon from time of sample collection. ' Should only be used in the presence of residual chlorine. 5 Free chlorine must be removed prior to addition of HCI by the appropriate addition of No2S20 3 6 Sample receiving no pH adjustment must be analyzed within seven days of sampling. :\lqapreva\sect6.doc Date: 12/22/95 Section 6.0 Revision 0.01 Page 7 of 9 Table 6.2 NAME Semivolatile Organics/Organoch/orine Pesticides/PCBs and Herbicides Concentrated waste samples Liquid samples, no residual Chlorine present Residual Chloride, present Soil/sediments and sludges Volatile Organics Concentrated waste samples Liquid samples. no residual Chlorine present Residual Chlorine, present Acrolein & Acrylonitrile Soil/sediments and sludges Date: 12/22/95 Section 6.0 Revision 0.01 Page 8 of9 Required Containers, Preservation Techniques, and Holding Times for Aqueous, Non-Aqueous, Soil or Solid Matrices (as specified in SW-846): CONTAINER PRESERVATION MAXIMUM HOLDING TIME 8 oz. wide mouth None 14 days until extraction, 40 days glass w/Tefion liner after extraction 1 gal. or 2 1/2 gal. Cool, 4°C Samples must be extracted within 7 amber glass w/Tefion days & extracts analyzed within 40 liner days 1 gal. or 2 1/2 gal. Add 3ml 10% sodium Samples must be extracted with in 7 amber glass w/T efion thiosulfate days & extracts analyzed within 40 liner days 8 oz. wide mouth Cool, 4°C 14 days until extraction, extracts glass w/Tefionlliner analyzed within 40 days: 8 oz. wide mouth None 14 days glass w/Tefionlliner 3x40 ml vials Cool, 4°C2 14 days w/Tefion lined septum caps 3x40 ml vials Collect sample in a 4 oz. 14 days w/Tefion lined septum soil VOA container which caps has been pre-preserved w/4 drops of 10% sodium thiosulfate. Gently mix sample & transfer to a 40ml VOA vial2. Cool to 4°C 3x40 ml vials Adjust to pH 4-5, Cool to 14 days w/Tefion lined septum 4°c caps Cool to 4°C 4 oz. (120ml), wide 14 days mouth glass w/Teflon liner or wide mouth glass container sealed w/a septum ' Holdrng trmes are based upon from trme of sample collectron. 2 Adjust pH<2 w/H2SO4, HCI or solid NaHSO4 :\lqapreva\sect6.doc I I I I I I I H I I I I I I I I I I D D I I I I I I I I I I ' Table 6.3 TEST TO1 TO2 VOST TO4 TO1O TO11 TO13 Date: 12/22/95 Section 6.0 Revision 0.01 Page 9 of9 Required Containers, Preservation, and Technical Hold Times for Air Methods: MEDIA PRESERVATION MAXIMUM HOLDING TIME ' Tenaxtubes Freezer -20°C 14 days Carbo Sieve Cool to 4°C 14 days Tenaz/Tenaz-Cool to 4°C 14 days charcoal Puf 3" long, 60mm Freezer -10°C or below Extracted 7 days after collection diameter Puf 10cm long, 20mm Cool to 4°C Extracted 7 days after collection diameter Absorbent cartridge Cool to 4°C 30 days Puf XAD/XAD Extracted 7 days after collection Holding times are based upon from time of sample collection. : \lqapreva \sect6 .doc D E I I I I I I I I I I I I 7.1 7.2 7.0 SAMPLE CUSTODY SAMPLE RECEIPT Date: 12/22/95 Section 7.0 Revision 0.01 Page 1 of 19 Sample shipments are received at the sample receiving area. Sample custodians verify the number of shipping containers received against the numbers listed on the shipping manifesVchain-of-custody. Any damage to the shipping containers or other discrepancy observed is noted on the chain-of-custody before signing it or on the sample receiving non-conformance report. A copy is filed for future reference. When practical, the external chain-of-custody must be signed by the carrier for relinquishment of samples and signed by sample custodian personnel for sample receipt. The actual chain-of-custody may be supplied by Pace (Figure 7.1 ), or may be the client's own form. The chain-of-custody remains in the project file at all times. CHAIN OF CUSTODY Chain-of-Custody encompasses three major elements: field sampling, laboratory analysis and final data file. A Chain-of-Custody (COC) document may be the means in some types of legal proceedings by which evidence of custody of samples from time of receipt to completion of analysis is proven in the courts. Pace has implemented standard operating procedures to ensure that sample custody objectives of traceability and responsibility are achieved for every project. This section covers quality related activities from the receipt of samples at the laboratory through the issuance of final analytical data and the storage of data in its final data file. The National Enforcement Investigations Center (NEIC) of EPA defines evidence of custody in the following manner: 1. 2. 3. 4. It is in your actual possession, or It is in your view, after being in your physical possession, or It was in your possession and then you locked or sealed it up to prevent tampering, or It is in a secure area. Samples may be physical evidence and should be handled according to certain procedural safeguards. Field personnel or Client representatives complete a Chain-of-Custody Form for all samples. Samples are received by the laboratory accompanied by these forms. The sampler should provide the following information: • . Client project name Project location • Field sample number/identification Date and time sampled Sample type :\lqapreva\sect7 .doc 7.3 Preservative Analysis requested Sampler signature Signature of person relinquishing samples Date and time relinquished Sampler remarks Custody Seal Number (if applicable) Date: 12/22/95 Section 7.0 Revision 0.01 Page 2 of 19 I I The record is filled out completely and legibly. Errors are corrected by drawing a single line I through and initialing and dating the error. The correct information is then recorded with indelible ink. All transfers of samples except to and from commercial couriers must be II recorded on the Chain-of-Custody via the "relinquished" and "received by" sections. All II information except signatures may be printed. SAMPLE VERIFICATION 7.3.1 Upon arrival of a sample shipment, sample control personnel perform sample inspection. Pace's Sample I.D. and Condition Sheet or equivalent (Figure 7.2) serves as a check-off list of procedures to follow and as documentation of the • following: 1. Presence/absence of custody seals or tapes of the shipping containers and the condition of the seals (i.e., intact, broken). 2. Presence/absence of chain-of-custody; (if present, is it complete?) 3. 4. 5. Presence/absence of sample tags; (if present, an~ they removable?) Agreement/non-agreement between the sample tags, chain-of-custody, and any client documentation. Condition of the samples when received, including: • Sample temperature • Intact, broken/leaking • Headspace in VOA vials • Sample holding time • Sample pH when required If discrepancies are found, the Pace project manager is contacted immediately (verbally and by using a Discrepancy Report Form or equivalent (Figure 7.3). If the project manager is not available, the QAO is contacted for further directions. A copy of the Discrepancy Report Form is attached to the project data package. :\lqapreva\sect7 .doc I I I I I I I I I I I I D D I I I I I I I I I I I I I 7.4 Date: 12/22/95 Section 7.0 Revision 0.01 Page 3 of 19 SAMPLE LOG-IN 7.4.1 General Policies a. b. C. d. e: f. Upon completing sample receipUcustody procedures, all sample and analysis data must be complete and documented on the chain of custody or accompanying forms for input into the Laboratory Information Management System (LIMS). Sample.and analysis data must include: 1. 2. 3. 4. 5. 6. 7. Client name and contact Client number Pace project number Pace project manager Sample descriptions Due date List of analyses requested Sample and requested analyses data are input into the LIMS. All samples received are logged into the LIMS on the day of receipt. A Sample and Analysis Data Entry Form (SADEF) or equivalent (Figure 7.5) is generated immediately by the LIMS. Distribution of SADEE: To the Pace Project Manager with a photocopy of the chain-of-custody form. (Include a copy of the Discrepancy Report if applicable). To the QC project file with the original chain of custody. Photocopy to the Organic or Inorganic Department Manager as it applies for RUSH samples. To the client. SADEF is to be reviewed against the chain of custody. Sample containers are labeled with the corresponding sample number and the stamped date of receipt. g. Samples are ready for storage. :\lqapreva\seci7 .doc 7.5 7.6 7.7 Date: 12/22/95 Section 7.0 Revision 0.01 Page 4 of 19 WHEN SAMPLES ARE RECEIVED WITH NO PAPERWORK 7.5.1. If delivered by a client: Client is asked if previous arrangements were made for analysis (and with whom). The client completes a chain of custody and/or request for analysis, relinquishes samples to sample custodian personnel, and is given a copy of the COC. 7.5.2. If received by courier or shipping the following ordered steps are taken: 1. 2. 3. 4. 5. 6. Routine Client File is checked Anticipate Sample Alert File is checked Sampling Kit Request File is checked Pace key client contact is consulted QC department manager is consulted to determine the designated Pace project manager Information is requested from the Pace project manager 7.5.3. If analysis information cannot be determined on the day of sample receipt, sample data entry personnel proceed to assign sample numbers and put samples on hold. · Follow-up with project manager occurs until the analyses are determined and samples can be properly logged in. RESPONSIBILITIES FOR SAMPLE LOG IN 7.6.1. Sample Custodian • • Has the primary responsibility of ensuring that sample information is input into the LIMS as described in the SOP. Has the responsibility to make recommendations to the QC manager for revising the SOP. 7.6.2. Sample Management Officer • Has the overall responsibility for ensuring that this procedure is implemented for all samples received into the laboratory. • Has overall responsibility for ensuring that samples are logged in correctly (given that appropriate information has been supplied). SAMPLE STORAGE 7.7.1. General Procedures Samples are stored immediately upon receipt to prevent sample degradation. :\lqapreva\sect7 .doc I I I I I I I I I I I I I I I I I I I D I I I I I I I I I I I I I Date: 12/22/95 Section 7.0 Revision 0.01 Page 5 of 19 7.7.2. Refrigerated Storage Area Maintenanc_e All refrigerated storage areas are maintained at 4°C (+/-2°C). The temperature is monitored and recorded each work day (certain programs may require more frequent monitoring; e.g., twice daily). If the temperature fails outside the limits of 2°-6°C, corrective action is to be taken as follows and appropriately documented. 1. Temperature is monitored at 60 minute intervals with the refrigerator door closed. 2. QAO is notified if the problem persists longer than one hour. 3. Samples are relocated to a proper storage environment if temperature cannot be maintained after corrective actions are implemented. 7.7.3. Routine Sample Storage 1. General Samples Samples within each project are stored in sample number order. Waters and soils are generally stored on labeled separate shelves and in separate refrigerated units. 7. 7.4. Specific Procedures 1. Volatiles 2. 3. :\lqapreva\sect7 .doc Samples within a project are stored in numerical order in vial containers. The holders are then stored where space permits in one of the designated volatile organic refrigerated storage areas. Semi-Volatiles Samples within a project are stored in numerical order in a designated, refrigerated storage area. Hazardous Materials Pure product or potentially heavily contaminated samples are tagged as "hazardous" and stored within a secured area, separate from other samples. This area is used only for hazardous samples and is labeled per Occupational Health and Safety Administration (OHSA) requirements. 7.8 4. Date: 12/22/95 Section 7.0 Revision 0.01 Page 6 of 19 Special Projects • Volatiles Samples within a project are stored in sample number order in vial containers. The holders are then stored as space permits in the Special Project Volatiles (VOA) refrigerated storage area. 7.7.5. Responsibilities for Sample Storage 1. Sample Management Officer has direct responsibility for ensuring that the Standard Operating Procedure (SOP) is followed, samples are stored properly upon receipt, and refrigerated storage area temperatures are maintained. 2. Sample custodians are responsible for storing all samples upon receipt into the appropriate storage area, maintaining high level security for those samples under custody, and for keeping a current custody sample inventory. 3. Sample management personnel have the responsibility of daily sample storage area maintenance, disposal of old samples, and providing space for incoming samples in routine storage areas. 4. Assigned individuals are responsible for maintaining and documenting: (a) refrigerated storage area temperatures, and (b) corrective actions. SAMPLE/DATA ACCESS AND INTERNAL CHAIN-OF-CUSTODY 7.8.1. General Policies and Procedures Pace has implemented standard operating procedures to assure the integrity of samples and data so that they are not degraded or disclosed to unauthorized personnel. In order to ensure that this policy is maintained, the laboratory facilities are operated under controlled access. Only employees are allowed into the laboratory facilities; visitors must register at the front desk. - Samples are removed from their proper location by designated personnel and returned to the storage area immediately after the required sample quantity has been taken. This procedure minimizes unnecessary time spent searching for samples and helps prevent matrix degradation from prolonged exposure to room temperature. After the final report is sent and clients are allowed adequate time to review the results, the samples are properly discarded or returned to the client. Upon client request, additional and more rigorous chain-of-custody protocols for samples and data can be implemented. For samples involving a high degree of confidentiality or potential litigation, Pace has developed extensive sample and data handling protocols to ensure the scientific and legal defensibility of the :\lqapreva\.sect7 .doc I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 7.9 Date: 12/22/95 Section 7.0 Revision 0.01 Page 7 of 19 report submitted. These protocols include those specified by the USEPA Contract Laboratory Program. Analysts and technicians follow strict internal chain-of-custody procedures to further ensure the validity of all data. All samples are signed out in a sample custody log book when they are removed for analysis. The sample ID, date, time, analyst, and lab of analysis is recorded in the sample custody log (Figure 7.4) or equivalent. Samples are signed back in noting date, time, and storage location, upon return. 7.8.2. Responsibilities for SOP Compliance 1. The QAO has the overall responsibility for ensuring that the SOP is implemented and followed. 2. 3. Sample custodian personnel have the responsibility for ensuring that the SOP is properly followed, and to notify the QAO of problems. All employees checking out samples are required to follow procedures. SUBCONTRACTING ANALYTICAL SERVICES Every effort is made to perform chemical analyses for Pace clients within a Pace laboratory. There are, however, instances where subcontracting of analytical services is necessary. Should subcontracting be necessary, samples are generally placed at other labs within the Pace integrated system of laboratories if at all possible. Currently, the following analyses are processed by specialty laboratories within Pace: • Air Analyses • Bioassay • Explosives When subcontracting becomes necessary, a preliminary verbal communication with an appropriate laboratory is undertaken. Work performed under specific protocols may involve special consideration, for instance, work involving NFESC samples may be subcontracted only to NFESC approved laboratories. The contact and preliminary arrangements and terms of agreement are made between the Pace Project Manager and the appropriate subcontract laboratory personnel (i.e., Laboratory Manager, customer services contact, or the appropriate laboratory section manager). The specific terms of the subcontract laboratory agreement should include (when applicable): • Method (EPA or otherwise) of analysis • Number and type of samples expected • Project specific QNQC requirements • Deliverables required • Applicable laboratory certification status • Price per analysis • Tum around time requirements :\lqapreva\sect7 .doc Date: 12/22/95 Section 7.0 Revision 0.01 Page 8 of 19 Chain-of-Custody forms must be generated for samples which require subcontracting to other laboratories. The sample management personnel repackage the samples for shipment, create a transfer chain-of-custody form and record the following information: • Pace Laboratory Number • Matrix • Requested analysis • Special instructions (quick tum around, required detection or reporting limits, · unusual information known about the samples or analytical procedure). • Signature in "Relinquished By" All subcontracted sample data reports are sent to the Pace Project Manager. The Project Manager sends the report to the appropriate Pace laboratory manager for review. Any Pace work sent to other labs within the Pace network is handled as subcontracted work. All of the conditions and considerations noted in Section 7 .10 and 7 .11 apply. 7.10 SAMPLE DISPOSAL After completion of sample analysis and submission of the analytical report, unused portions of samples are retained by the laboratory for a minimum of 2 weeks. After 2 weeks, samples will be disposed of according to the nature of the samples. The Hazardous Waste Manager receives a copy of the data report and uses that information to select the appropriate waste stream for the samples. The samples are considered hazardous waste, then they will be disposed of by state and federally licensed hazardous waste disposal finns. Upon disposal of samples, a computer spreadsheet is maintained by the Hazardous Waste Manager listing the sample number, inherent waste stream and date disposed. This data file is updated on a weekly basis and is kept on file by the Hazardous Waste Manager and Sample Management. 7.11 EXCESS SAMPLE DISPOSITION Samples not consumed during the analyses are returned to the client or disposed of by Pace. It is the project manager's responsibility to ensure that proper disposal has taken place. If the sample is detennined to be non-hazardous by the project manager, it may then be disposed of by Pace via a non-manifested process. 7.1.1. Notification of Sample Return :\lqapreva\sect7 .doc The project manager and client receive written notification at the time of project initiation in the following manner: 1. The project proposal states the following paragraph in its Conditions and Tenns Statement: I I I I I I I I I I I I I I I I I I I 0 D D E I I I I I I I I I I 2. 3. Date: 12/22/95 Section 7.0 Revision 0.01 Page 9 of 19 "Pace Analytical Services, lnc.'s Standard Operating Procedure is to return all samples of hazardous materials to the client at project completion, and Pace Analytical Services, Inc. reserves the right to return or dispose of all samples at its discretion unless contractually agreed otheiwise." The Sample and Analysis Data Entry Form (or equivalent) states the following: "Pace Analytical Services, Inc. reserves the right to return all samples at our discretion." This form is printed by the LIMS at sample check-in. The Sample and Analysis Data Entry Form cover letter contains the following paragraph: 1. "Pace Analytical Services, lnc.'s Standard Operating Procedure is to return all samples of hazardous materials or wastes to the client at project completion. Pace Analytical Services, Inc. reserves the · right to return or dispose of all samples at our discretion" (Figure 7.5). This is a pre-printed cover letter that accompanies the Sample and Analysis Data Entry Form. 4. The Sample and Analysis Data Entry Form and cover letter (or equivalents) are generally sent to the client by the project manager depending upon project requirements. 7.11.2 Sample Return and Disposal If samples are to be returned to the client or held longer than 60 days, a sample disposition form is generated. Otheiwise, samples are disposed of a minimum of two weeks after project completion. · a. The example Sample Disposition Form (Figure 7.6) contains the following information: 1. 2. 3. 4. 5. Client name, address, and contact Pace project number Client project identification number Pace sample identification number Pace project manager name This form may vary by location 7.11.3 Procedure for Use of the Sample Disposition /SD} Form /or equivalent) :\lqapreva\sect7 .doc 1. 2. :\lqapreva\sect7 .doc Date: 12/22/95 Section 7.0 Revision 0.01 Page 10 of 19 The project manager separates the sample disposition form from the report package, signs the form, and routes it to the sample custodian. If the project requires, the hazardous waste manager may hold the form for a required amount of time before return or disposal. It is important that this form be used and not discarded. It is part of the internal Chain of Custody and is filed with the project report The hazardous waste manager or designee will use action codes such as: 1 = Return to client C = Clean 2 = In house disposal D = Dirty As a general rule, soil samples will be returned and water samples will be disposed of in-house. Water samples which are highly contaminated will be returned. If a sample has an extremely high level of contamination, the contaminant will be noted by the project manager on the SD form. For In-House Sample Disposal All preserved -Non-hazardous-Neutralize/sink Hazardous-Toxic waste Unpreserved water-Non-hazardous-Sink Hazardous-Toxic waste Soil/Sludge-Non-hazardous-Refuse Disposal Hazardous-Toxic waste All VOA's-Non-hazardous-Neutralize/sink Hazardous-Toxic waste AH Extracted samples CAM Extracts -Non-hazardous -Neutralize/sink Hazardous -Acid metals waste Other Extracts -Hazardous waste Liquid/Unknown Miscellaneous -Project manager specify Subsequent to receipt of the Sample Disposition Form by the sample custodian, samples will be removed from storage using the information provided on the form. If the Sample Disposition Form indicates "disposed," the Sample Custodian will remove samples from storage and place them at a sample disposal station for proper disposal. The process of disposal is performed by the I I I I I I I I I I I I I I I I I I D D 0 • I I I I I I I I I I I 3. Date: 12/22/95 Section 7.0 Revision 0.01 Page 11 of 19 sample custodian or appropriate laboratory staff. The Sample Disposition Form is signed and dated by the sample custodian then routed to the project manager for filing with other project information. If samples are to be returned, the Sample Custodian removes them from storage, initials and dates the Sample Disposition Form. The samples, Sample Disposition Form, and a copy of the client's chain of custody are then delivered to the shipping clerk by the sample custodian for return to the client. Upon receipt of the samples and Sample Disposition Form, the shipping clerk signs and dates the form. The Sample Disposition Form is copied and the original form with the samples is returned to the client, along with a copy of the client's chain of custody. A copy of the Sample Disposition Form and the original chain of custody is routed to the file clerk for filing with other project information (QC file). The shipping clerk labels the box with an appropriate hazard label and ships the samples back to the client. 4. Sample Disposition Forms are filed in project files. 7.11.4 Hazardous Material/Waste Sample Disposition Option :\lqapreva\sect7 .doc The preferred method for disposition of hazardous samples is to return the excess · sample to the client. It may not be feasible to return samples in all cases or the client may require Pace to dispose of excess samples. Pace will dispose of excess samples when required and will charge a disposal fee to recover costs for management and disposal. Procedure for Disposal Option for Excess Hazardous Material/Waste Samples: 1. When analyses are complete, the project manager indicates disposal as the option on the Sample Disposition Form and completes and attaches Hazardous Sample Disposal Option Form (Figure 7.7) or equivalent. An entry must be made in all fields of this form as it will determine the basis for lab packing and disposal. 2. The project manager routes the Disposal Option Form to sample check in. 3. 4. The project manager is responsible for billing the client for disposal. The sample custodian is responsible for maintaining a file of Disposal Option Forms for all samples awaiting disposal. Hazardous material/waste samples are stored in a safe manner and segregated by compatibility groups as indicated by the hazardous waste disposar SOP. 5. :\lqapreva\sect7 .doc Date: 12/22/95 Section 7.0 Revision 0.01 Page12of19 The hazardous waste manager is responsible for reviewing accumulated samples awaiting disposal and initiating the disposal process when warranted. The Field Services, Inorganic, Organic, and Environmental Services Departments cooperate and participate in the disposal process. (For compatibility and compositing, see the Hazardous Waste Disposal SOP.) I I I I I I I I I I I I I I I I I I I -- Clienc Address Phone Sampled By {PAINT): Samplo:tf Sign.ah.lie 2 3 4 5 6 7 8 Additional Commonls --- -- 0 Oa11 Sampled w > ~ w ~ w ~ ~ z ~ - Ao To: P.O., I BIiiing Relerence P,ojoct Name I No. 0 cS i" < z 0 z > - - I!!!!!!!-I!!!!!!! I!!!!!!! 3?3R75 CHAIN-OF-CUSTODY RECORD Analytical Request P.ac.o C~en1 No. PiaCI ProjKt Man.age, REMARKS SEE .REVERSE SIDE FOR INSTRUCTIONS I!!!!!! -u ., n (1) )> :, ., -<" ~-n ., en (1) < ff (1) Ill 'TI -(0· ~ :, C: ., p ~ 3 (") .... ""£. :::c )> :.... (1) z 6 .,, h C en --t 0 0 -< .,, 0 ;u l!: -u ;o (/) 0 Ill (ti ct> ru t0<n~ CD c,;· 5· ~ 5' :, _.. w ::J .....,~ 0 a· N :: a a t::! U) _.. ~ Client: _______ _ Project No.: ______ _ Date Received: ____ _ FIGURE 7.2 SAMPLE I.D. AND CONDITION FORM (Format may vary by location) Date: 12/22/95 Section 7.0 Revision 0.01 Page 14 of 19 SAMPLE CONDITION UPON RECEIPT CHECKLIST I I I I I I Complete checklist (A) during sample receipt. If any items are marked "NO," complete section (B) of this form. Otherwise, go to record samples. I (A) (B) 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. Are there custody seals or tapes on the shipping container? Are custody seals on the shipping container intact? Is there a completed Chain-Of-Custody (C-0-C)? Do the numbers of samples received and the sample matrices agree with C-O-C? Are there tags attached to each sample? Are sample tags, sample containers and C-O-C all in agreement? Is the C-O-C complete with requested analyses? Are the samples preserved correctly? Is there enough sample to do all analyses? Do the samples have the proper temperature? Are the sample containers intact (e.g., not broken, leaking)? Are VOA vials head-space free? Are all samples within the holding times for requested analyses? _ Is pH recorded for non-VOA's? YES l'1.Q Explain "NO" item here:. ______________________ _ Send a copy of this form to Project Manager with Discrepancy Report Form. Copy of both forms remain in the QC file. Custodian Signature: ___________ _ :\lqapreva\sect7 .doc I I g I I I I I I I I I ' D D D I E m I I I I I I I I I I I I FIGURE 7.3 Pace Analytical Services, Inc. DISCREPANCY REPORT FORM (Format may vary by location) Urgency Level: 1 (_) Requires immediate attention 2(_) Requires attention today 3(_) Requires attention this week Date: 12/22/95 Section 7.0 Revision 0.01 Page 15 of 19 Initiator _______________ Client: ___________ _ Date: _______ _ Project# ______ _ Sample(s) # _____________ _ Discrepancy (if more space needed, use the back of this form): ___________ _ To QC Manager. Date: _________________ _ Client Notified? YES ( ) NO ( ) Date & Time: _______________ _ Project Manager Notified? YES ( ) NO ( ) Date & Time:. _____________ _ QC Response:. ____________________________ _ Project Manager Response: _______________________ _ Cause and Resolution (proposed or carried out): Completed by:. ___________ _ Manage~s Initials: PM Signature: ________ Date:. _________________ _ QC Signature: ________ Date:. _________________ _ cc: Project File :\lqapreva\se_ct7 .doc Project Date Created Initials RELEASED SDG CLOSED A B C D E SAMPLES OUT Initials Datemme :\lqapreva\sect7 .doc F G H I J FIGURE 7.4 Pace Analytical Services, Inc. Internal Chain of Custody (Format may vary by location) Date: 12/22/95 Section 7.0 Revision 0.01 Page 16 of 19 Fractions Available Potentially Radioactive? r Yes r No SAMPLES K p u L Q V M R w N s X 0 T y Which Where is it IN Any Totally Comments Fraction Going? Initials Consumed? Datemme I Page: I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I May 24, 1995 Dear Valued Client: FIGURE 7.5 Pace Client Letter SAMPLE Date: 12/22/95 Section 7.0 Revision 0. O 1 Page 17 of 19 A new policy has been implemented in the Sample Receiving Department of Pace Analytical Services, Inc. We hope that this policy will be helpful to you. Upon acceptance of samples into the laboratory, the Sample Custodian completes a Sample and Analysis Data Entry Form. This form is designed to accommodate a short description of the samples received (sample name and/or sample reference), the type of container, and a list of the analyses requested to be performed on each sample. A copy of this form will be sent to the Client. Enclosed is a copy of the Sample and Analysis Data Entry Fonm relevant to the samples we recently received from you. Please compare the information on the fonm to assure that it is consistent with your request. If there is any inconsistency or if you have any questions on your project, please call the Pace Contact indicated on the Sample and Analysis Data Entry Form. The Pace Contact has primary responsibility for monitoring the progress of your project through the laboratory. It is also part .of Pace Analy1ical Services, lnc.'s Standard Operating Procedure to return all samples pertaining to the infonmation attached that are hazardous materials or hazardous wastes to the client at project completion. Pace Analy1ical Services, Inc. reserves the right to return or dispose of all samples at our discretion. We have implemented this procedure to better serve our clients, and would appreciate any comments you may have. Sincerely, :\lqapreva\sect7 .doc Date removed: ----Initials: ______ _ Date shipped: ___ _ Initials: ______ _ FIGURE 7.6 SAMPLE DISPOSITION FORM (Format may vary by location) RE: Client Project ID: ______________ _ Pace Project No.: _______________ _ Sample ID Dear Date: 12/22/95 Section 7.0 Revision 0.01 Page 18 of 19 All requested analyses of the samples for the above referenced project have been completed. Enclosed are the remaining portions of the samples which are being returned to you for final disposition. If you have any questions, please call me. Sincerely, Project Manager :\lqapreva\sect7 .doc I I I I H D I I I I I I I I I I I I I E D I I I I I I I I I I I I I Pace Project # Project Manager Pull Sample Date Sample# Matrix Remarks: 1 = Return to Client 2 = In House Disposal FIGURE 7.7 HAZARDOUS SAMPLE DISPOSAL OPTION FORM (Format may vary by location) Location Disposal Method C = Clean D = Dirty Removed from· Refrigerator (initial/date) Date: 12/22/95 Section 7.0 Revision 0.01 Page 19 of 19 Charge Returned to Client (initial/date) ________________ _ Disposed of Samples (initial/date _______________ _ :\lqapreva\sed7.doc I I I I I I I I I I Date: 12/22/95 Section 8.0 Revision 0.01 Page 1 of 9 8.0 CALIBRATION PROCEDURES AND FREQUENCY All instruments and equipment used in the laboratory must follow a well defined calibration routine. Calibration may be accomplished by laboratory personnel using certified standard reference materials traceable to the National Institute of Standards and Technology (NIST) or EPA certified materials or by external· standardizing bodies or commercial standard manufacturers. The discussion presented here is general in nature because the requirements for calibration are instrument (or equipment) and method specific. Details of calibrations can be found in Pace Standard Operating Procedures, analytical methods, and instrument operations manuals. In addition to the summary calibration information pertaining to general analysis categories contained in the following subsections, Tables 8.1 and 8.2 list detailed calibration information for representative methods and applications most frequently performed at Pace. 8.1 STANDARDS AND TRACEABILITY Analytical standards are prepared from pure compounds or are purchased as neat chemicals or diluted standard solutions from reputable vendors. They are used to prepare· serial dilutions from which calibration and spiking standards are prepared. Each laboratory section is responsible for the preparation, storage and disposal of its standards. The preparation information is recorded into section specific Standards Notebooks in order to document traceability of prepared standards to their source material(s). Each standard is given an internal identification number. The preparation of all stock standards shall be documented in a Standards Notebook which is used to record the date of preparation, analyst's initials, the source of the reference material, standard components, amounts used, final volume, final concentration(s), solvent used, expiration date of prepared standard, and the serial reference number of that stock solution. All standards shall be labeled with the standard serial reference number and expiration date (small glass ampules), and if space permits, with the name, concentration, date of preparation and initials of preparer. All diluted working standards not consumed during an analytical session shall be labeled fully, including the serial reference number of any stock standard used in its preparation. If no expiration date has been assigned by the manufacturer, then an expiration date of one year from the date of preparation ( or the date first opened in the case of sealed ampules) is reported unless degradation prior to this date is observed. To help determine if a standard has degraded, one must note inconsistencies. For instance, very poor recoveries from newly prepared quality control spikes or abnormally low instrument response to a specific standard are indications of possible standard degradation. However, for some standards, degradation is more easily noted. For instance, DDT breaks down to form DOD and ODE. Here one can visually note, on a chromatogram, the degradation of DDT by the increased concentrations of ODD and ODE. If degradation is observed before the default expiration date, it should be noted in the Standard Notebook for that standard and the standard removed from service. :\lqapreva\sect8.doc 8.2 Date: 12/22/95 Section 8.0 Revision 0.01 Page 2 of 9 Before any set of standards can be utilized in a calibration curve they must be verified either externally by the standard supplier or internally by a secondary source process: • • Analysis of qualified QC Check Sample (e.g., A2LA approved), or Analysis of an independently prepared check standard prepared either from a . different manufactured lot for the same vendor supplying the calibration standard or from a second supplier. GENERAL CALIBRATION PROCEDURES Calibration standards for each parameter are chosen to bracket the expected concentrations of those parameters in the sample and to operate within the linear response range of the instrument. Samples that fall outside the calibration range are diluted until bracketed by the calibration standards. A low level standard is routinely analyzed to verify the reporting limit. Calibration standards are prepared typically at a minimum of three concentration levels, usually chosen at two to five times, five to ten times, and up to twenty times the estimated method detection limit plus a calibration blank, with the exception of most organic analyses which do not require a calibration blank. Either an internal standard or external standard quantification technique can be utilized. The reporting limit is verified by analysis of a standard at the reporting limit. Calibration standards are prepared from materials of the highest available purity. To establish instrument calibration, working standards are prepared from more concentrated working stock solutions. All organic standards are refrigerated or frozen. Inorganic standards are refrigerated as necessary. Standard preparation information is recorded within each laboratory section in designated Standards Notebook. Instrumental responses to calibration standards for each parameter are subjected to an appropriate statistical test of fitness (least squares linear regression, quadratic equation, or relative standard deviation of response factors) or as required by the method or QAPP. The calibration must reflect an acceptable correlation of data points or linearity to be acceptable. Point-to-point curve fitting shall not be used for establishing initial calibration correlation acceptance. In cases where the calibration data are outside these criteria, the analyst must rerun the calibration standards (meeting the same criteria) and/or prepare a new curve, changing instrumental conditions as necessary. For analyses which are performed frequently and for which substantial calibration data is available, a complete recalibration is not required each time an analysis is performed providing that the following criterion is met: one calibration standard is analyzed at the beginning of the analysis which may vary from the expected response (based on the most recent initial calibration curve) by~ 25% difference or as specified by the method, SOP or QAPP. If this criterion is not met, a complete recalibration is necessary. :Uqaprevalsect8.doc I I I I I I I a I I I I I I I I I I I I I I I I I I I I I I I I I Date: 12/22/95 Section 8.0 Revision 0.01 Page 3 of 9 During the course of analysis, calibration standards are routinely analyzed to ensure that the instrumental response has not changed. The continuing calibration criteria stipulated in each method or SOP are used by the analyst to determine whether the instrument must be recalibrated or the instrument conditions further optimized. The accuracy of prepared standards is periodically checked by comparison with a standard from an independent source. Certain equipment such as balances, pH meters, and turbidity meters are normally calibrated with NIST traceable standard reference material. 8.2.1 Analytical Balances Every 12 months, calibration of the entire analytical range shall be checked by a qualified service technician. The calibration of each balance is checked each day the balance is used with weights traceable to NIST. Calibration weights are ASTM Class 1 (replaces Class S designation) and are recertified every two years. If balances are calibrated by an external agency, verification of their weights shall be provided. All information pertaining to balance maintenance and calibration is· recorded in the individual balance logbook and/or is maintained on file in the QA department. 8.2.2 Thermometers Certified, or reference, thermometers are maintained for checking calibration of working thermometers. Reference thermometers are provided with NIST traceability for initial calibration and are recertified every year with equipment directly traceable to the NIST. Working thermometers are compared with the reference thermometers every 12 months; digital working thermometers are verified for accuracy on a quarterly . frequency. Each thermometer is tagged and individually numbered. In addition, working thermometers are visually inspected by laboratory personnel prior to use. Calibration temperatures and acceptance criteria are based upon the working range of the thermometer and the accuracy required for its use. Laboratory . thermometer inventory and calibration data are maintained in the QA department or designated area. 8.2.3 pH/Electrometers The meter is calibrated before use each day, and once after each four hours of continuous use, using fresh buffer solutions. 8.2.4 Spectrophotometers :\lqapreva\sectB.doc 8.3 Date: 12/22/95 Section 8.0 Revision 0.01 Page 4 of9 During use, spectrophotometer performance is checked at established frequencies in analysis sequences against initial calibration verification (ICV) and continuing calibration verification (CCV) standards. The instrument operating capability is also evaluated annually (more frequently if required by certification agencies} by qualified laboratory personnel or by an outside instrument maintenance service. GC/MS CALIBRATION PROCEDURES The minimum operations necessary to satisfy analytical requirements associated with the determination of organic compounds in water and soil/sediment samples are listed below. The following operations should be performed routinely in the laboratory: • Documentation of GC/MS mass calibration and abundance pattern • Documentation of GC/MS response factor stability • Internal standard response and retention time Prior to initiating data collection, it is necessary to establish that a given GC/MS meets the standard mass spectral abundance criteria. This is accomplished through the analysis of • decafluorotriphenylphosphine (DFTPP} for base/neutral and acid (BNA} compounds or p- bromofluorobenzene (BFB} for volatile compounds. Each GC/MS system used for analysis of volatile or semivolatile organic compounds must be tuned to meet method or program specific ion abundance criteria before analysis of standards, blanks, or samples can proceed. Prior to the analysis of samples and after tuning criteria have been met, the GC/MS system must be initially calibrated with a minimum of five concentrations of each compound being analyzed to determine the linearity of response. USEPA criteria specify both the concentration levels for initial calibration and the specific internal standard to be used on a compound-by-compound basis for quantitation. The response factor (RF} for each compound at each concentration level is calculated using the following Equation 8.1: RF= Where:· * (8.1) As Cx = area of the characteristic ion for the compound to be measured = area of the characteristic ion for the specific internal standards = concentration of the internal standard (mg/ml) = concentration of the compound to be measured (ng/ul} Using the RF from the initial calibration, the percent relative standard deviation (¾RSD} for compounds identified as Calibration Check Compounds (CCCs} is calculated using Equation 8.2: :\lqapreva\sect8.doc I I I I I I I I I I I I I I I I I I D I I I I I I I I I I I I I Where: s %RSD = X 100 X RSD = relative standard deviation (8.2) Date: 12/22/95 Section 8.0 Revision 0.01 Page 5 of 9 s = standard deviation of initial five response factors (per compound). x = mean of initial five response factors (per compound). The % RSD for each individual CCC must be ~ than 30% or as specified by the method. This criterion must be met for the initial calibration to be valid. A calibration check standard containing all compounds of interest as well as all required surrogates, is performed each day of analysis. The RF data from the standard is compared each day against the average RF from the initial calibration for a specific instrument. If the response to a calibration check standard differs from the initial calibration by more than ±20% or as specified by the method, then investigation and corrective action must be performed, including a complete recalibration if necessary. 8.4 NON GC/MS CHROMATOGRAPHY CALIBRATION PROCEDURES · Initially, a three or five point calibration curve, consisting of all compounds of interest (plus a calibration blank for certain analyses such as VOCs), is established to define the usable range of the instrument. Calibration may be accomplished as best-fit line, quadratic equation, or average RF. The curve is determined to be linear if the correlation coefficient is .:". 0.995. Linearity may also be determined using response factors. Response factors are calculated for each compound at each concentration level. These RFs will be averaged to generate the mean RF for each compound over the range of the standard curve. The curve is determined to be linear if the RSD of the response factors is <20%. The mean response factor will be used to calculate the sample concentration of the compound of interest. When sample responses exceed the range of the standard curve, the sample will be diluted to fall within the range of the standard curve and be reanalyzed. The results of the daily GC standardization will be tabulated and filed with the corresponding sample analyses. Daily full calibration is not necessary if a calibration check standard validates the initial calibration curve. If the response to a calibration check standard differs from the initial calibration by more than ±15% for any analyte being quantitated or as specified by the method, then investigation and corrective action will be performed, including complete recalibration, if necessary. Continuing Calibration is checked as described in Pace SOPs or methods. 8.5 Calibration of Inductively Coupled Argon Plasma Spectrometer (ICP) and Atomic Absorption Spectrophotometer (AAS) The ICP and AAS are standardized for the metal of interest by the analysis of a set of calibration standards prepared by diluting a stock solution of known concentration. Working standards are prepared by dilution of the stock standard. For the AAS, the concentration of the calibration standards is chosen so as to cover the working range of :\lqapreva\sectB.doc Date: 12122/95 Section 8.0 Revision 0.01 Page6of9 the instrument. For ICP, a standard is analyzed as a sample to determine the upper limit of the calibration. Subsequently, all sample measurements are performed within this working range. After the working standards have been prepared, they are analyzed on the ICP or AAS and the instrument response is calibrated to provide a direct readout in concentration. The calibration is accomplished by entering the metal concentration equivalent to the readout in absorbance units (or emission intensity) during analysis of the working standards. After the initial calibration, the analysis of the working standards is repeated during sample analysis to standardize instrument response during analysis and to confirm the calibration settings. A typical analysis sequence is presented below. Working standards are prepared by dilution of a stock standard solution of the metal of interest. A calibration curve within the working range of the instrument is established by . analysis of three to five working standards. An independent standard is analyzed to confirm the calibration settings. If the calibration settings are not confirmed, the instrument is recalibrated. The samples are analyzed for the metal of interest. During sample analysis, a check standard is analyzed to monitor instrument. stability. If the analysis indicates that instrument calibration has changed by more than .±10% for ICP or more than ±20% for AAS, the instrument is recalibrated and the analysis is repeated. Following completion of the sample analyses, the check standard is reanalyzed to confirm calibration settings. If calibration settings are confirmed, the analysis is completed. However, if the calibration settings are not confirmed, the problem is corrected, and the analyses are repeated. Written records of all calibrations shall be filed with the raw data. :\lqapreva\sect8.doc I I I I I n I I I I I I I I I I I I I Instrument GC/MS (8270B) GC/MS (8240B) Gas Chromatograph (8080A) :\lqapreva\sect8.doc TABLE 8.1 Summary of Calibration Requirements Calibration Standards Used, Initial and Pally Minimum Tune: DFTPP Initial: 5 level (20,50,80, 120,160, ppb) Daily: 1 level (50) (every 12 hours) Tune: BFB Initial: 5 level (10,20,50, 100,200 ppb) Daily: 1 level (50) (every 12 hours) Initial: 5 level (cone. based upon instr. response) Mid level DDT/Endrin standard Acceptance Limits Meets criteria RSD <30% for RFs of CCCs RF ~ 0.050 (SPCC) % Difference <20% of the average five- point RF (CCC) Meets Method Criteria RSD <30% for RFs of CCCs RF~ 0.300 (0.250 for bromoform) (SPCC) % Difference <20% of the average five-point RF (CCC) Std curve or calibration factor (CF) if% RSD .'.': 20 DDT/Endrin breakdown < 20% Date: 5/22/95 Section 8.0 Revision 0.00 Page 7 of 9 Corrective Actions Re-tune instrument Repeat DFTPP analysis Repeat Calibration Evaluate system Repeat Calibration Evaluate system Take corrective action Repeat Calib. Check; see Lab Supervisor Re-tune instrument Repeat BFB Analysis Repeal Calibration Evaluate System Repeal Calibration Evaluate System Take Corrective Action Repeal Calib. Check; See Lab Supervisor Make new standards or establish new calibration curve. Rerun standard once Perform col. maint. Instrument Gas Chromatograph (herbicides) Inductively Coupled Plasma Emission Spectrometer TABLE 8.1 Summary of Calibration Requirements (continued) Calibration Standards Used, Initial and Pally Minimum Daily: 1 level of check standard (midrange) Std check every 10 samples Initial: 5 levels ( cone. based upon instr. response) Daily: 1 level of check standard (midrange) Std check every 1 0 samples Initial: high standard + blank Daily: instrument check standard and calibration std. & blank every 1 0 samples Acceptance Limits CF < ± 15% of initial calibration CF± 15% of daily calibration (< ± 20% for confirmation). Retention times within retention time windows.(For methods using retention time windows.) Standard curve or calibration factor (CF) if% RSD < 20 CF < ± < 15% of initial calibration CF± 15% of daily calibration (< ± 20% for confirmation). Retention times within retention time windows. (For methods using retention time windows.) ICV: < 90-110% CCV: < 80-120% Date: 5/22/95 Section 8.0 Revision 0.00 Page 8 of 9 Corrective Actions Repeat initial calibration Reanalyze samples that were analyzed after standard that failed criteria and before next standard that passes criteria Make new standards or establish new calibration curve Rerun samples that were analyzed between standards failing criteria Recalibrate. Repeat twice; if outside control limit, then recalibrate making new stds if necessary :\lq a preva \sects. doc ----------limliil liaa ---== 111111::1 li:ill'll:I -- iiiii --- Instrument Atomic Absorption Spectrophoto- meter pH Meter UV-Visible Spectra- photometer Instrument Analytical Balance Thermometers :\lqapreva\sect8.doc - - -------l!!l!!!!!!!!!I l!!l!!!!!!!!!I l!!l!!!!!!!!!I TABLE 8.1 Summary of Calibration Requirements {continued) Calibration Standards Used, Initial and Daily Minimum Acceptance Limits Initial: 5 levels + blank Linear regression correlation Daily: 1 check standard coefficient :::_0.995; (midrange) & blank ICV: 90-110% per 10 samples CCV: 80-120% Daily: 2 levels +0.05 pH unit Bracket sample range Initial: 5 levels + blank Linear regression correla- Daily: check standard tion coefficient > 0.995; ICV: 90-110% TABLE 8.2 Summary of Routine Calibration Requirements Calibration Standards Used, Initial and Daily Minimum Daily: Sensitivity (with ASTM Class "1" weight) Annually; Calibrate in constant temperature baths at two temperatures against precision thermometers certified against an NIST thermometer Acceptance Limits ±0.001 gm (varies by method) ±0.1 to ±_0.5 C (depending upon method) . Date: 5/22/95 Section 8.0 Revision 0.00 Page 9 of 9 Corrective Actjons Make new standards or establish new calibration curve Clean or replace electrode; recalibrate Recalibrate services Correctjve Actions Adjust sensitivity. re-level Tag and remove from service, replace l!!l!!!!!!!!!I l!!!!!I !!!!I I I I I I I I I I I I I I II I II 9.0 ANALYTICAL PROCEDURES Date: 2/28/97 Section 9.0 Revision 2.0 Page 1 of 27 Pace Analytical laboratories are capable of analyzing the full range of environmental samples from all media, including surface and groundwater, soil, sediment, tissue, and waste. Refer to Table 9.2 for a representative listing of specific Pace Analytical capabilities. Methodologies are employed with guidance from agencies such as EPA, ASTM, USGS, NIOSH and, in certain instances, state regulatory agencies. In some situations, Pace Analytical develops and validates methodologies which are more applicable to a specific problem or objective. Analytical procedures are detailed descriptions of any and all processing, preparation and analysis of samples In the laboratory. In some instances, data format, presentation and delivery are also described. All analytical procedures shall be conducted in strict adherence with written Standard Operating Procedures manuals which have been reviewed and approved by the Laboratory Operations Manager, the Pace QA Officer and the Pace General Manager. Documents from which SOPs are developed include the references listed in Table 9.1. Additional SOPs may be adapted from other sources or generated in-house as project needs require. 9.1 ANALYTICAL METHODS Numerous sources of information are available to offer guidance in analytical methods. Selection of the appropriate method is dependent upon data usage and the regulatory requirements during the analysis. Table 9.1 describes the analytical references routinely used by Pace Analytical. Pace Analytical may modify existing methods based on the following considerations: 1) In order to meet project specific objectives; 2) in order to incorporate modifications or improvements in analytical technology; 3) In order to comply with changing regulations and requirements: 4) in order to address unusual matrices not covered in available methods. Pace Analytical will make every effort to disclose to its clients any instances in which modified methods are being used in the analysis of samples. The following subsections contain method synopses for representative methods most frequently performed at Pace Analytical laboratories. For clarity purposes, c1;rtain method summaries also contain calibration criteria, several of which have been previously detailed in Section 8. 9.2 SAMPLE PREPARATION METHODS 9.2.1 Digestion of Aoueous samples for Metals -Method 3005A This method is an acid digestion procedure used for the preparation of water samples for metals analysis. The digested samples can· be analyzed for dissolved and total recoverable metals by flame (FAA) or furnace (GFM) atomic absorption spectrophotometry or by inductively coupled argon plasma emission spectroscopy (!CPS). Method .3005A may be used to prepare samples for analysis of the following metals: :llqaprevalsect9.do<: Aluminum Antimony Arsenic Barium Beryllium Cadmium Calcium Chromium Cobalt Copper Iron Lead Magnesium Manganese Molybdenum Nickel Potassium Selenium Silver Sodium Thallium Vanadium Zinc Date: 2128/97 Section 9.0 Revision 2.0 Page 2 of 27 For the analysis of total recoverable metals, the entire sample is acidified at the time of collection with nitric acid (HN03). Sample preparation involves heating the sample with acid and concentrating to a specified volume. The sample is not allowed to boil because some of the elements are in a volatile state and may be easily lost. The digestate is then filtered (if necessary) and diluted to the desired concentration for analysis. For the analysis of dissolved metals, the samples are filtered through a 0.45-um filter immediately upon collection and prior to acidification with nitric acid. In the lab, the sample is heated with acid and the volume is reduced. The digestate is filtered again (if necessary) and diluted to volume. 9.2.2 Digestion of Aqueous Samples to(Metals • Method 3QJOA and the CLP sow These methods describe the preparation of aqueous sampjes for total metals determination by flame atomic absorption spectrophotometry (FM) and by inductively coupled argon plasma emission spectroscopy (ICPS). By method 301 OA. samples are vigorously digested with nitric acid. By CLP protocol. samples are digested with a mixture of nitric acid and hydrochloric acid. 9.2.3 Digestion of Aqueous Samples tor Metals -Method 3020A and the CLP sow These methods describe the preparation of aqueous samples for total metals determination by graphite furnace atomic absorption spectroscopy (GFM). By method 3020A, samples are vigorously digested with nitric acid. By CLP protocol, samples are digested wlth a mixture of nitric acid and hydrogen peroxide. 9.2.4 Digestion of So/id samples for Metals -Method 3050A and the CLP sow These methods are applicable to the preparation of sediment, sludge, and soil samples for metals determination by FAA or GFM or by ICPS. One gram of solid sample is digested with nitric acid and hydrogen peroxide. The digestate is then refluxed with nitric or hydrochloric acid, depending on the analysis performed. When using hydrochloric acid as the final refluxing acid, the digestates may not be boiled because antimony is in a volatile state and may be easily lost. A separate sample aliquot is dried to determine the percent moisture in the sample. 9.2.5 separatocy Extraction -Method 351 PB Method 3~1 OB is designed to quantitatively extract nonvolatile and sernivolatile organic compounds from liquid samples using separatory funnel techniques. :~qepreva\sect9.doc I n I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I II Date: 2/28/97 Section 9.0 Revision 2.0 Page 3 of 27 The sample and extracting solvent must be immiscible in order to yield recovery of target compounds. Subsequent cleanup and detection methods are described in the organic analytical method that will be used to analyze the extract. Samples are pH-adjusted and serially extracted by vigorous shaking for 1-2 minutes with the appropriate solvent for the analytical method. Samples are extracted three times, the combined extracts are dried with anhydrous sodium sulfate and concentrated in a Kuderna-Danish apparatus. 9.2.6 continuous Uguld/Uguid Extraction -Method 3520B Method 3520B is designed to quantitatively extract nonvolatile and semivolatile organic compounds from liquid samples using continuous liquid-liquid extractors. The sample and extracting solvent must be immiscible in order to yield recovery of target compounds. Subsequent cleanup and detection methods are described in the organic analytical method that will be used to analyze the extract. Samples are pH-adjusted and extracted with the appropriate solvent for the analytical method. Samples are extracted for 18 to 24 hours, the extracts are dried with anhydrous sodium sulfate, and then concentrated in a Kudema-Danish apparatus. 9.2.7 Soxh!et Extraction -Method 35408 Method 3540B is a procedure for extracting nonvolatile and semivolatile organic compounds from solids such as soils and sludges. The So:xhlet extraction process ensures intimate contact of the sample matrix with the extraction solvent. Extraction Is accomplished by mixing the solid sample with anhydrous sodium sulfate, placing it in an extraction thimble or between two plugs of glass wool, and extracting it with an appropriate solvent in the Soxhlet extractor for 18 to 24 hours. The extract is dried and concentrated and then treated using a cleanup method or analyzed directly _by the appropriate measurement technique. 9.2.B Sonicatioo Extraction -Method 3550A Method 3550A is a procedure for extracting nonvolatile and semivolatile organic compoundl> from solids such as soils and sludges. The Sonication process ensures intimate contact of the sample matrix with the extraction solvent. A weighed sample of the solid waste is mixed with sodium sulfate, then dispersed into the solvent using sonication. The extract is dried with anhydrous sodium sulfate and concentrated with a Kuderna-Danish apparatus. The resulting solution may then be cleaned up or analyzed directly using the appropriate technique. 9.2.9 Waste Pllutioo -Method 3580A Method 35B0A is a technique for solvent dilution of non-aqueous waste samples prior to sample cleanup and/or analysis. It is designed for wastes that may contain organic constituents at concentrations greater than 20,000 ug/kg and that are soluble in the dilution solvent. :~qaprevalsect9.doc 9,3 9.2.1 o Purge-and-Trai;, Sample Introduction -Method 5030A Date: 2/28/97 Section 9.0 Revision 2.0 Page 4 of 27 Method 5030A is used to determine the concentration of volatile organic compounds in a variety of liquid and solid waste matrices using a purge and trap gas chromatographic procedure. The success of this method depends on the level of interferences in the sample. Results may vary due to the large variability and complexity of various matrices. Inert gas is bubbled through a 5-ml or 25-mL aqueous sample aliquot at ambient temperature to transfer the volatile components to the vapor phase. The vapor is swept to a sorbent column where the volatile components are trapped. After purging is completed, the sorbent column is flash heated and backflushed with inert gas to desorb and transfer the volatile components onto the head of a GC column. The column is heated to elute the volatile components. which are detected by the appropriate detector for the analytical method used. n B I I g I I Solid samples may be analyzed using one of two techniques. For Low level soil I analysis, 5 g of solid sample is dispersed into 5 ml of Contaminant-free laboratory water and the sample is purged as described above. This technique is referred to as the direct purge method. For medium level soil analysis, an aliquot of solid I sample is dispersed in methanol to dissolve the volatile constituents and a portion of the methanol extract is combined with contaminant-free laboratory water and purged as described above. I 9.2.11 Extraction Procedure To:r;icity Test IEP-Tox} -·Method 1310A This method is used to determine whether a waste exhibits the characteristics of I extraction procedure (EP) toxicity. If a representative sample of the waste contains >0.5% solids, the solid phase of the sample is ground to pass a 9,5 mm sieve and extracted with deionized (DI) water that is pH adjusted with acetic acid. Wastes I containing <0,5% solid material are extracted and analyzed as a single phase. 9.2.12 Toxicity Characteristic Leaching Procedure CTCLP} -Method 1311 This method is used to determine whether a waste exhibits toxicity leaching characteristics. The procedure includes a leaching extraction for semivolatile compounds and metals and a zero-headspace extraction for volatile compounds. 9. 2.13 California Assessment Manual Waste Extraction Test /CAM WED This waste extraction test, described in the California Administrative Code, Title 22, Article 11, Section 66700, can be used to determine the amount of extractable substance in a waste or other material. CALIBRATION AND ANALYSIS PROCEDURES FOR ORGANICS I I I I 9.3.1 tJalogenated Volatile Organics -Method so, PB I Halogenated volatile organics in water and soil samples are analyzed using I :\lqaprevalsect9.doc I I I I I I I I I I I I I I I I Date: 2/28/97 Section 9.0 Revision 2.0 Page 5 of27 method 801 OB, which is a gas chromatography (GC) method using purge and trap sample introduction (method 5030A). An inert gas is bubbled through a water matrix to transfer the volatile halocarbons from the liquid to the vapor phase. Volatile halocarbons are collected on a sorbent trap, then flash thermally desorbed and transferred to a GC column. Target analytes are detected with an electrolytic conductivity detector (ELCO). Soil samples may be heat purged directly in reagent water or are extracted with methanol; if extracted in methanol an aliquot of sample extract is added to blank reagent water for purge and trap GC analysis. Positive results are confirmed by GC analysis using a second GC column of dissimilar phase. When second column analysis is performed, peak retention times (RTs) on both columns must match expected RTs within the calculated RT windows. Also, calculated quantitations from each column should be in agreement with one another (generally they should match within a factor of two) for the presence of an analyte to be considered confirmed. Cafjbration • Calibration standards are prepared and analyzed at five i:oncentration levels. A linear calibration curve not forced through the origin is developed for each analyte of interest. This function is used for the calibration curve if the correlation coefficient (r) for that analyte is ~0.995, otherwise, a curve function is used that meets this criterion. Each working day, the calibration is verified with the analysis of a continuing calibration standard at the beginning and end of the run sequence and after every 10 analyses. The calibration factor for each analyte to be quantitated must not exceed a 15% difference when compared to the initial standard of the analysis sequence, When this criterion is exceeded. inspect the GC system to determine the cause and perform whatever maintenance is necessary before recalibrating and proceeding with sample analysis. All samples that were injected after the standard exceeding the criterion must be reinjected to avoid errors in quantitation, if the initial analysis indicated the presence of the specific target analytes that exceeded the criterion. 9.3.2 Aromatic VOiatiie Organics -Method B02QA Aromatic VQlatile organics in water and soil samples are analyzed using method 8020A, which Is a gas chrornatography (GC) method using purge and trap sarnple introduction (method 5030A). An inert gas is bubbled through a water matrix to transfer volatile aromatic hydrocarbons from the liquid to the vapor phase. Volatile aromatics are collected on a sorbent trap, then flash thermally cJesorbed and transferred to a GC column. Target analytes are detected using a photoionization detector (PIO). Soil samples may be heat purged directly in reagent water or are extracted with methanol; if extracted with methanol an aliquot of sample extract is added to blank reagent water for purge and trap GC analysis. Positive results are confirmed by GC analysis using a second GC column of dissimilar phase. When second column analysis is performed, peak RTs on both columns must match expected RTs within the calculated RT windows. Also, calculated quantilations from each column should be in agreement with one :\lqapreva\sectS.doc .:.::. . Date: 2/28/97 Section 9.0 Revision 2.0 Page 6 of27 another (generally they should match within a factor of two) for the presence of an analyte to pe considered confirmed. Calibration -Calibration standards are prepared and analyzed at five concentration levels. A linear calibration curve not forced through the origin is developed for each analyte of interest. This function is used for the calibration curve if r ~0.995 for that analyte; otherwise, a curve function is used that meets this criterion. Each working day, the calibration is verified with the analysis of a continuing calibration standard at the beginning and end of the run sequence and after every 1 O · analyses. For each analyte of interest, the %D of the response in the continuing calibration standard must agree with the expected response by ~15% in order for the run sequence to continue. 9.3.3 Organochlorine Pesticides and PCBs -Method BOBOA and the CLP sow Organochlorine pesticides and PCBs are analyzed by gas chromatography following either method 8080A or the CLP Organic SOW. Each of these analyses involves solvent extraction of the sample followed by analysis by gas chromato- graphy with electron capture detection (GC-ECD). Positive results are confirmed using a second GC column of dissimilar phase. For an analyte to be considered confinned, the peak RTs on both columns must match the expected RTs. Also, the calculated quantitations between the two columns should be in agreement with one another (generally they should match within a . factor of two) for the presence of the analyte to be considered confirmed by method BOBOA. For analysis by CLP protocol, the results are flagged with a "P" if the two quantitations differ by more than 25%. In addition, the breakdown of 4,4'- DDT and endrin is monitored. If the breakdown cif either of these compounds is found to exceed 20%, the analytical sequence must be discontinued. For analysis by CLP protocol. the combined breakdown must also not exceed 30%. Calibraljon for Method BOSOA -Calibration standards are prepared and analyzed at five concentration levels. A linear calibration curve not forced through the origin is developed for each analyte of interest. This function is used for the calibration curve if r ~0.995 for that analyte; otherwise, a curve function is used that meets this criterion. Each working day, the calibration is verified with the analysis of a continuing calibration standard at the beginning and end of the run sequence and after every 1 O analyses. For each analyte of interest, the %D of the response in the continuing calibration standard must agree with the expected response by !:15% in order for the run sequence to continue. Calibration foe the CLP SOW -Calibration and analysis are performed in strict accordance with the CLP Organic SOW. 9.3.4 \/Qlatile Organics -Method 8240B and the CLP sow :\lqapreva\sect9.doc Samples may be analyzed for volatile organics by gas chromatography/mass spectrometry (GC/MS) following the procedure described in method 8240B or the CLP Organic SOW. Analyte identification and quantitation are a=mplished using I R I H I I I I I I I I I I I I I I I E u I I I I I I I I I I I I I I II Date: 2/28197 Section 9.0 Revision 2.0 Page 7 of 27 response factors and retention times generated from a five-point calibration curve, relative to the closest eluting internal standard. The three internal standards used for these methods are: • • • Bromochloromethane 1,4,Difluorobenzene Chlorobenzene-d5 If requested by the client, non-target analytes are reported as tentatively identified compounds (TICs), when an acceptable match is obtained between the spectrum of the analyte and a spectrum found by library search. Unidentified TICs are labeled "unknown". The TICs are quantitated using response factors of 1 relative to the nearest eluting internal standards. Instrument Performance Check -The mass spectrometer is tuned daily and after every 12 hours of operation to yield an acceptable spectrum for p-bromofluorobenzene (BFB). Relative ion abundance criteria for BFB are given in Table 9.3. CaUbratjon for Method 8240B -· After passing the instrument perfon11ance check criteria and prior to analyzing samples, a 5-point initial calibration is performed. From that calibration, the calibration check compounds (CCCs) must meet the RSD criteria of !:30% and the system performance check compounds (SPCCs) must meet the minimum RRF criteria given in the method. A continuing calibration standard is analyzed after every 12 hours of operation. In that standard, the CCC compounds must meet the %D criteria of !:20% and SPCC compounds must meet the minimum RRF criteria listed in the method. Caljbrajjon for the CLP SOW -After passing the instrument performance check criteria and prior to analyzing samples, a 5-point initial calibration is performed. From that calibration, the compounds listed in Table 2 of Exhibit D, Section IV r,./OA) of the CLP sow must meet the RSD criteria of !:20,5% and the minimum RRF criteria listed in the method. A continuing calibration standard is analyzed after every 12 hours of operation. In that standard, the Table 2 compounds must meet the %D criteria of !:25% and the minimum RRF criteria listed in the method. 9.3.5 Semivo!atile organics -Method 8270B and the CLP sow Semivolatile extracts are analyzed by gas chromatography/mass spectrometry following method 8270B or the CLP Organic SOW. All samples are prepared following extraction methods described in the applicable protocol. Identification and quantitation is performed using response factors and retention times generated from a five-point calibration curve, relative to the closest eluting of six internal standards, The six internal standards are: • 1,4-Dichlorobenzene-cL. • Naphthalene-de • Acenaphthene-d10 • Phenanthrene-d10 :~qaprevalsoct9.doe • • Chrysene-d12 Perylene-d12 Date: 2128197 Section 9.0 Revision 2.0 Page 8 of27 If requested by the client, non-target analytes are reported as tentatively identified compounds (TICs), when an acceptable match Is obtained between the spectrum of the analyte and a spectrum found by library search. Unidentified TICs are labeled "unknown". The TICs are quantitated using response factors of 1 relative to the nearest eluting internal standards. Instrument Performance Check • The mass spectrometer is tuned daily and after every 12 hours of operation to give an acceptable spectrum for DFTPP. DFTPP ion abundance criteria are given in Table 9-4. Caljbratjon for Method 8270B -After passing the instrument performance check criteria and prior to analyzing samples, a 5-point initial calibration is performed. From that calibration, the CCC compounds must meet the RSD criteria of :::30% and the SPCC compounds must meet the minimum RRF criteria given in the method. A continuing calibration standard is analyzed after every 12 hours of operation. In that standard, the CCC compounds must meet the %D criteria of ::,20% end SPCC compounds must meet the minimum RRF criteria listed in the method, Calibration for the CLP SOW -After passing the instrument performance check criteria and prior to analyzing samples, a 5-point initial calibration is performed. From that calibration, the compounds listed in Table 5 of Exhibit D, Section IV (SV) of the CLP SOW must meet the RSD criteria of .:;20.5% and the minimum RRF criteria listed in the method. A continuing calibration standard is analyzed after every 12 hours of operation. In that standard, the Table 5 compounds must meet the ¾D criteria of ::,25% and the minimum RRF criteria listed in the method. I I I B I D I I I I I I 9.3.6 purgeab\e Petroleum Hydrocarbons I Gasoline and volatile aromatic compounds, including benzene, toluene, ethylbenzene, and the xylenes (BTEX), are analyzed by a modified method 8015A using the direct purge technique described above for method 5030A. Analysis is I performed on a GC equipped with a photoionization detector (PIO) and a flame ionization detector (FID) connected in series. If BTEX compounds are found without the associated presence of gasoline, confirmation analysis is performed 1 with a second GC column of dissimilar phase and retention characteristics in accordance with the requirements of method 8020A. Caljbratjon -Calibration standards are prepared and analyzed at five concentration levels. A linear calibration curve not forced through the origin is developed for each analyte of interest. This function is used for the calibration curve if r ~0.995 for that analyte; otherwise, a curve function is used that meets this criterion. Each working day, the calibration is verified with the analysis of a continuing calibration standard at the beginning and end of the run sequence and after every 1 O analyses. For each analyte of interest, the %D of the response in the continuing :~qapreva\Gocl9.doc I I I I I I I I I I I I I I I ,I I I I 9.4 Date: 2/26/97 Section 9.0 Revision 2.0 Page 9 of 27 calibration standard must agree with the expected response by ~15% in order for the run sequence to continue. 9.3.7 Extractable Petroleum Hydrocarbons Aqueous samples analyzed for diesel, kerosene, jet fuel, and motor oil are prepared using method 35108 (separatory funnel liquid/liquid extraction) or method 35208 (continuous liquid/liquid extraction). Solid samples are prepared using method 3540B (Soxhlet extraction), method 3550 (sonication extraction), or wrist action shaker extraction (California LUFT method). One liter of water or 30 g of soil/sludge are extracted and concentrated to a volume of 1 ml. Analysis is performed by a modified method 8015A on a GC equipped with a capillary or megabore column and an FID detector. Cafjbratjon -Calibration standards a~prepared and analyzed at five concentration levels. A linear calibration curve not forced through the origin is developed for each analyte of interest. This function is used for the calibration curve if r ?.0.995 for that analyte; otherwise, a curve furaction is used that meets this criterion. Each working day, the calibration is verified with the analysis of a continuing calibration standard at the beginning and end. of the run sequence and after every 1 O analyses. For each analyte of interest. the %D of the response in the continuing calibration standard must agree with the expected response by ~15% in order for the run sequence to continue. REPRESENTATIVE CALIBRATION AND AN!A.LYSIS PROCEDURES FOR INORGANICS 9.4.1 Metals by !CPS -Method 601 OA and the CLP sow These methods describe the simultaneous or sequential determination of metal elements using !CPS. The method measures element-emitted light by optical spectromet,y. Samples are nebu\ized and the resulting aerosol is passed through a plasma torch, Element-specific atomic-line emission spectra are prnduced which are dispersed by a grating spectrometer and monitored for intensity by photomultiplier tubes. Calibration -n,e calibration procedl!lres for !CPS are detailed in method 6010A and the CLP lnorganics SOW. Prior :10 the analysis of samples, an initial multipoint calibration is analyzed for all elements of interest. The initial calibration is checked with an initial calibration verification. standard (ICV). For each element, the ICV responses must agree with the initia'I calibration within ±10% for the calibration to be verified. Following the ICV and after the analysis of every 10 samples, a continuing calibration verification standard (CCV) is analyzed. The response for each element in the CCV must agree within :!:20% of the expected value for the analysis to continue. 9.4.2 Metals by GFM -Methods 70601\ 7421, 7740, 7841 and the CLP sow - Graphite furnace atomic absorption (GFAA) techniques may be used for the determination of arsenic, lead, selenium, thallium, and other mete1ls depending upon the sensitivity required. Following sample digestion, an aliquot of sample is :~qaprevalsect9.doc Date: 2128/97 Section 9.0 Revision 2.0 Page 10 of27 placed in a graphite tube in the furnace, evaporated to dryness, charred, and atomized, The sample is placed in the light path of an atomic absorption spectrophotometer. The absorption cf light by the atomized metal is measured with a photomultiplier tube. Ca!ibratjon -Calibration procedures for the GFAA analyses are detailed in the respective methods in SW-846 and the CLP SOW. For the element of interest. a m1,1ltipoint initial calibration is performed. The calibration correlation coefficient must be ?,0.995 to be acceptable. The initial calibration is verified by the analysis of an !CV standard prepared from a source independent of the calibration standards. The response of the ICV must agree with the expected response within ±10% in order for the calibration to be verified. A CCV check standard is analyzed following the analysis of the !CV and after the analysis of every 10 samples. The response of the CCV must agree with the expected value within ±20¾ in order for the analysis to contin\.le. 9.4.3 Mercury by CVAA -Methods 7470. 7471A and the CLP sow· Cold-vapor atomic absorption (CVAA) techniques are used for the determination of mercury. Sample preparation is specified in the method. Following dissolution, mercury in the sample is reduced to the elemental state and aerated from solution in a closed system. The mercury vapor passes through a cell positioned in the light path of an atomic absorption spectrophotometer. Ca/lbratjon • The calibration procedure is detailed in SW-846 and the CLP SOW. Prior to the analysis of samples, a multipoint Initial calibration is performed. The calibration correlation coefficient must be ?,0. 995 to be acceptable. The initial calibration is verified by the analysis of an ICV standard prepared from a source independent of the calibration standards. The response of the ICV must agree with the expected response within ±20¾ In order for the calibration to be verified. A CCV check standard is analyzed following the analysis of the ICV and after the analysis of every 1 O samples. The response of the CCV must agree with the expected value within ±20% in order for the analysis to continue. 9.4.4 Total and Amenable Cyanide -Method so1owso12 and the CLP sow These methods are 1,1sed to determine the concentration of inorganic cyanide in aqueous or solid samples. These methods are 1,1sed to determine values for both total cyanide and cyanide amenable to chlorination. Cyanide, as hydrocyanic acid (HCN). is released by refluxing the sample with strong acid and distilling the HCN into an absorber-scrubber containing sodium hydroxide solution. The cyanide ion In the absorbing solution is then determined by UV spectrophotometry. Calibration -Prior to sample analysis, a multipoint initial calibration is performed, The calibration correlation coefficient must be ?,0. 995 to be acceptable. The initial calibration is verified by the analysis of an !CV standard prepared from a source independent of the calibration standards. The response of the ICV must agree with the expected response within ±10% for the calibration to be verified. A CCV check standard Is analyzed following the analysis of the !CV and after the analysis :llqaprevo\se<:t9.doc u I D I I I I I I I I I I I I I I I I I I D D I -m I I I I I I I I I I Date: 2128/97 Se,ction 9,0 Revision 2.0 P<1ge 11 of 27 of every 10 samples, The response of the CCV must agree with the expected value within ±10% for analysis to continue. 9.4.5 Anions -Method 300.o Method 300.0 may be used to analyze anions. including chloride. nitrite. nitrate. o- phosphate. bromide, and sulfate. in aqueous samples by ion chromatography (IC). A volume of sample Is injected into the ion chromatograph. The anions of interest are separated and measured using a chromatography system consisting of a guard column, separator column, suppressor device and conductivity detector. Samples must be refrigerated at 4 °C and analyzed within 48 hours of sample collection if nitrate, nitrite, and/or o-phosphate are analyzed. or within 28 days of sample collection if chloride, bromide and/or sulfate are analyzed. Calibration -Prior to sample analysis. a multipoint initial calibration is analyzed. The calibration correlation coefficient must be :c:_0.995 to be acceptable. The initial calibration is verified by the analysis of an !CV standard prepared from a source independent of the calibration standards. The response of the !CV must agree with the expected response within :t.10% for the calibration to be verified. A CCV check standard is analyzed following the analysis of the !CV and after the analysis of every 1 n samples. The response of the CCV must agree with the expected value within ±15% for analysis to continue. 9.4.6 pH· Methods 1501. 9040. & 9045A Methods 150.1 and 9040 are used to measure the pH of aqueous and multiphase samples where the aqueous phase constitutes at least 20% of the total sample volume. The pH of the sample is determined electrometrically using either a glass electrode in combination with a reference potential or a combination electrode. Method 9045A is used to determine the pH in soil samples. CaJjbratjon -The pH meter is calibrated with three standard buffer s1Jlutions. The reading must be within +0.05 to ±0.1 pH units (depending upon the instrument) of the true value of each buffer solution. 9.4,7 Non-Filterable Residue -Method 160.1 This method is applicable to drinking, surface and saline waters, and domestic and industrial wastes. A well mixed sample is filtered through a glass fiber filter. The residue that passes through the filter is dried and measured gravimetrically. CaHbra!jon -The analytical balance must be checked each day of use with ASTM Class 1 weights. Balance readings must read within ±D.001 g of the true weight. 9.4.8 Filterable Residue -Method 1so,2 This method is applicable to drinking, surface and saline waters. and domestic and industrial wastes. A well mixed sample is filtered through a glass fiber filter. The residue on the filter is dried and measured gravimetrically. :\lqapreva\sect8.doc ·. ·-···, ·., Date: 2/28/97 Section 9.0 Revision 2.0 Page 12 of27 Caljbratlon -The analytical balance must be checked each day of use with Class S weights. Balance readings must read within ±0.001 g of the true weight. 9.4.9 Nitrate-Nitrite -Method 353.2 Method 353.2 is used to determine the concentrations of nitrate and nitrite in aqueous samples. Nitrite concentration is determined by diazotization with sulfanilamide and complexation with N-(1-naphthyl)-ethylenediamine dihydrochloride to form a highly colored azo dye which is measured colorimetrically. Combined nitrate-nitrite concentration is determined by first carrying out a copper-cadmium reduction step. A filtered sample is passed through a column containing granulated copper and cadmium to reduce nitrate to nitrite. Nitrate concentration is determined from the difference of the nitrate- reduced nitrite value and the nitrite value. Samples must be preserved with sulfuric acid to pH -9 and refrigerated at 4°C. If analysis for N02 or N03 only is desired. no preservative should be used. Calibration -Prior to sample analysis, a multipoint initial calibration is analyzed. The calibration correlation coefficient must be ~0.995 to be acceptable. The initial calibration Is verified by the analysis of an ICV standard prepared from a source independent of the calibration standards. The response of the ICV must agree with the expected response within ±15% for the calibration to be verified. A CCV check standard is analyzed following the analysis of the ICV and after the analysis of every 10 samples. The response of the CCV must agree with the expected value within ±15% for analysis to continue. 9.4.10 Total Organic Carbon rroc1 -Methods soso and 415,1 Methods 9060 and 415.1 are used to determine the concentration or total organic carbon In samples. TOC Is analyZed by combustion of organic material in the sample to carbon dioxide, followed by infared ( IR) detection of the carbon dioxide. Caljbration -The instrument is calibrated by analyzing four replicates of a single concentration standard. The initial calibration is verified by the analysis of an ICV standard prepared from a source independent of the calibration standards. The response of the ICV must agree with the expected response within ±15% for the calibration to be verified. A CCV check standard is analyzed following the analysis of the ICV and after the analysis of every .15 samples. The response of the CCV must agree with the expected value within ±20% for analysis to continue. 9.4.11 Oil and Grease -Methods 9070/9071A and 413,1 Methods 9070 and 413.1 are used to determine the concentration of oil and grease in waters and wastes. The aqueous sample is acidified with HCI to pH <2 and extracted with Freon•TF (1,1,2-trichloro-1,2,2-trifluoroethane) in a separatory funnel. Sample extracts are evaporated to dryness and measured gravimetrically on an analytical balance. Method 9071A is used to prepare solid samples for gravimetric analysis of oil and grease. By this method, solid samples are Soxhlet extracted with Freon-TF and the extracts are evaporated to dryness and measured gravimetrically on an analytical balance. :~qapreva\sect9.doc I I I I I I I R I I I I I I I I I I I I a g 0 D R I E I I I I I I I I Date: 2/28/97 Section 9.0 Revision 2.0 Page 13 of27 Calibration - A balance calibration check is performed at the beginning and end of each analytical sequence with 1 g and 100 g ASTM Class 1 weights. Measurements must agree to within ,±0.001 g of the true weight. 9.4.12 Oil and Grease -Method 413.2 This method is used to determine the concentration of oil and grease in waters and wastes. Samples are acidified with HCI to pH <2 and extracted with Freon-TF in a separatory funnel. Sample extracts are analyzed by infrared (IR) spectrophotometry. Calibration -Prior to sample analysis, a multipoint initial calibration is performed. The calibration correlation coefficient (r) must be ::.0.995 for the calibration to be acceptable. A continuing calibration standard is analyzed after the analysis of every 1 O samples. The continuing calibration must agree with the initial calibration within ±20%. 9.4.13 Total Recoverable Petroleum Hydrocarbons CTBPH) -Method 41 B,1 This method is used to determine the concentration of total petroleum hydrocarbons in waters and wastes. The sample is acidified with HCI to pH <2 and extracted with Freon-TF in a separatory funnel. Extracts are shaken with silica gel to remove interferences, then the extracts are analyzed by infrared (IR) spectrophotometry. Ca/jbratjon -Prior to sample analysis. a multipoint initial calibration is performed. The calibration correlation coefficient (r) must be ::.0. 995 for the calibration to be acceptable. A continuing calibration standard is analyzed after the analysis of every 10 samples. The continuing calibration must agree with the initial calibration within ,±20%. 9.5 METHODVALIDATION 9.6 When non-promulgated methods (i.e. methods other than EPA, NIOSH, ASTM, AOAC, etc.) are required for specific projects or analytes of interest, or when the laboratory develops a method, the laboratory establishes the validity of the method prior to applying it to client samples. Method validity is established .by meeting certain criteria for precision and accuracy as established by the data quality objectives specified by the end user of the data. METHOD DETECTION LIMITS Method detection limit studies are performed for each method in use at least annually and after any procedural or configurational change. · Method detection limits are determined at Pace Analytical for analyses done on samples originating under Safe Drinking Water Act (SDWA) and Clean Water Act (CWA) provisions by using replicate spiked analyte-free water samples. A minimum of seven replicates of a sample spiked for the purpose are processed through the entire analytical ;llqaprava\sect9.doc Date: 2/28/97 Section 9.0 Revision 2.0 Page 14 of27 method. The concentration of the detection limit sample should be between 2 and 5 times the anticipated detection limit. The laboratory calculates the detection limit as the Student's t(n-1) value (e.g., t value= 3.143 for seven replicate determination) times the standard deviation of 7 replicate spiked sample measurements. The reader is referred to 40 CFR Part 136, Appendix B for further discussion. For samples which are analyzed by methodology approved under the Resource, Conservation and Recovery Act (RCRA}, the MDL Is determined by multiplying the appropriate one-sided 99% I-statistic by the standard deviation obtained from a minimum of three analyses of a matrix spike containing the analyte of interest at a concentration three to live times the estimated MDL, where the !-statistic is obtained from standard references. Estimate the MDL by obtaining the concentration value that corresponds to: a) an instrument signal/noise ratio within the range of 2.5 to 5.0, or b) the region of the standard curve where there is a significant change in sensitivity (i.e., a break in the slope of the standard curve). The reader is referred to SW-846, Third Edition, Chapter One, Volume 1A for further discussion. IT IS IMPERATIVE TO NOTE THAT METHOD DETECTION LIMITS ARE HIGHLY MATRIX DEPENDENT. LIMITS DETERMINED BY PACE ANALYTICAL MAY NOT BE ACHIEVABLE IN ALL MATRICES. 9.7 COMPLIANCE 9. 7 .1 Definition -Compliance is the proper execution of recognized, documented procedures which are either approved or required, Adherence to these procedures is required in order to provide data products acceptable to a regulatory body of competent jurisdiction in a specific regulatory context. Compliance is separate from, but not inconsistent with, technical scientific quality. Pace Analytical accepts compliance as part of the Pace Analytical corporate definition of quality; "Quality is the fulfillment of expectations and needs in all activities, demonstrated by the satisfaction of those we serve." Pace Analytical understands that the expectations of our clients commonly include the assumption that data and reports will satisfy a regulatory purpose and will be found acceptable and compliant with regulatory requirements for the performance of tests and generation of data. 9.7.2 Understanding the Regulatory Framework -Compliance is not likely to be achieved in the absence of an understanding of the regulatory framework. Pace Analytical will attempt to ascertain, prior to beginning a project, what regulatory jurisdiction (USEPA, NJDEPE, etc.) pertains to a project; within the regulatory jurisdiction, what body of regulation is meant to be satisfied (RCRA, SDWA, 21 E, etc.); and finally, within this context, what protocols are required/expected (CLP, AFCEE, NFESC, ASP, etc.). Pace Analytical will work with its clients to come to a mutual understanding of all requirements. :~q•preva\soct9.doc I I I I I I I I I n I I I I I I I I I I I I I D D D I I I I I I I I Date: 2/28/97 Section 9.0 Revision 2.0 Page 15 of 27 9.7.3 Commitment -Experience has shown that the complexity of 1mvironmental regulations and their overlapping Jurisdiction can result in conflicting DQOs to be established for a project or site by local, state and federal regulatory agencies. As a result of these types of complicating factors, clients and intermediaries working on their behalf may, but often do not, fully understand their compliance needs. Clients may sometimes fail to communicate their compliance requirements to Pace Analytical. Nevertheless, Pace Analytical Services, Inc., in defining quality as in 9.7.1 above, has accepted much responsibility for compliance. Pace Analytical makes the following commitments to its clients: Pace Analytical will proactlvely attempt to identify and understand the regulatory context of clients' needs. Pace Analytical will strive to be expert in understanding and executing the regulatory requirements for compliance. Pace Analytical will idenUfy and disclose to clients instances of non- compliance in a forthright fashion. 9.7.4 Beso\yjng Compliance Contradjctjons and Hierarchies -It is a common occurrence that multiple regulatory jurisdictions overlap in a specific case. This causes uncertainty or even contradictions to arise in a work plan. Pace Analytical will make every effort to detect such inconsistencies, and will communicate them to clients so that an informed decision can be made by the client regarding execution of the project. Similarly, methods and protocols will often be prescribed in a scope of work or QAPP which either will not achieve stated or implied DQOs or which are in conflict with the regulatory requirements. Pace Analytical will attempt to detect these inconsistencies, and upon detection, disclose same to our client. Pace Analytical voluntarily accepts a responsibility to provide advice to clients, however, the primary responsibility forthis issue remains with the client. 9. 7.5 Disclosure of Noncompliance -As stated previously, it is Pace Analytical policy to disclose in a forthright manner any detected noncompliance that may effect the usability of data produced by Pace Analytical. It is not within our expertise to predict the manner in which a specific regulator or regulatory body will interpret the rules governing analysis; therefore, Pace Analytical is unable to guarantee compliance. It is Pace Analytical policy that our responsibility begins with a bona fide and competent attempt to evaluate potential compliance issues and ends with disclosure of any findings that may be useful to our client in their making the final judgment. :llqapreval<ect9.doo • • • • • • • • • :\lqapreva\sect9.doc TABLE 9.1 ANALYTICAL PROTOCOLS Date: 2/28/97 Section 9.0 Revision 2.0 Page 16 of 27 ', ::..·: "Guidelines Establishing Test Procedures for the Analysis of Pollutants Under the Clean Wat~r Act." Federal Register, 40 CFR Part 136, October 26, 1984. "Test Methods for Evaluating Solid Wastes: Physical/Chemical Methods," SW- 846. 2nd edition, 1982 (revised 1984), 3rd edition and 1st Update, Update II and IIA, 1994, Office of Solid Waste and Emergency Response, U.S. EPA. "Methods for Chemical Analysis of Water and Wastes", EPA 600/4-79-020, 1979 Revised 1983, U.S. EPA U.S. EPA Contract Laboratory Program Statement of Work for Organic Analysis, SOW 2188, OLM01 .8, 8/91, OLM02.0, and OLM03.0. U.S. EPA Contract Laboratory Program Statement of Work for Inorganic Analysis, SOW No. 788, ILM01 .0. 3/90 through ILM03.0. "Standard Methods for the Examination of Water and Wastewater'', 15th, 16th, 17th and 18th editions, 1980, 1985, 1969, 1992. APHA-AWJVA-WPCF. "Annual Book of ASTM Standards", Section 4: Construction, Volume 04.04: Soil and Rock; Building Stones. American Society for Testing and Materials, 1987. "Annual Book of ASTM Standards", Section 11: Water and Environmental Technology, American Society for Testing and Materials, 1987. "NIOSH Manual of Analytical Methods", Third Edition, 1984, U.S. Department of Health and Human Services, National Institute for Occupational Safety and Health. "Methods for the Determination of Organic Compounds in Finished Drinking Water and Raw Source Water'', U.S. EPA, Environmental Monitoring and Support Laboratory -Cincinnati (September 1986). New York Stale Department of Environmental Conservation. Analytical Services Protocol, September, 1989 (revised December 1991 ). I I I I I I I I I I I I I I I I I I I I I I I g u D D m I I I I I I I I Table 9.2 UST OF ANALYTICAL METHODS 1 Organic Analyses Halogenated Volatile Organics Non-Halogenated Volatile Organics Purgeable Aromatics and Unsaturated Organics Acrolein and Acrylonitrile GC GC GC GC Organochlorine Pesticides and GC Polychlorinated Biphenyls Polynuclear Aromatic GC Hydrocarbons Chlorinated Hydrocarbons GC Base/Neutrals & Acids GC/MS Organophosphorus Pesticides GC Chlorinated Herbicides GC Volatile Organic Compounds GC/MS Fuel Hydrocarbons and BTEX GC or IR Alachlor, Atrazine GC Chlordane, Heptachlor, GC Heptachlor Epoxide, Lindane, Methoxychlor Carbofuran HPLC Endothall Total Petroleum Hydrocarbons :\lqopreva\sect9.doc GC IR 503/5022 508 1613 525 515.1 524.2 531.1 548 601 602 603 608/608.1 608.2 612 625 614/622 615/608.1 608.2 624 602/418.1 619/645 608/617 418, 1 Date: 2/28/97 Section 9.0 Revision 2.0 Page 17 of 27 8010A 8015A 8020NB021A 8030A --- 8080A Mod 8080 8270B 8120A 8270B 8140 8150B 8240B/ 8260A 8015A 8080N8140 8080A 2. Inorganic Analyses Alkalinity Biochemical Oxygen Boron Bromide Chemical Oxygen Demand Chloride Chloride Chlorine, Residual Color Cyanide, Total Amenable Fluoride, Total Fluoride, Dissolved Hardness, Total :\lqapreva\sect9.doc Table 9.2 (cont) Potentiometric Titration 5-Day, 20°c ICP Ion Chromatography Dichromate Reflux (High) Dichromate Reflux (Low) Mercuric Nitrate Ion Chromatography Titration Colorimetric Visual Comparison Pyrtdlne-Barbituric Acid, Colorimetric Chlorination- Colorimetric Distillation- Electrode Electrode Ion Chromatography EDTA Titration Calculation 2320 5220 5220 4500-CI" 4500-CI" 4500-Cr 2120 4500-CN" 4500-CN. 4500-F 4500-F 2340 . 310.1 405.1 200,7 300.0 410.1 410.2 325.3 300.0 330.5 110.2 110.3 335.2 335.1 340.2 340.2 300.0 130.2 Date: 2/28197 Section 9.0 Revision 2.0 Page 18 of27 D1067 D1252 D512 D512 D2036 D2036 D1179 01179 D1179 01126 6010 9252 9010A 9012 9010A 9012 I I I I I -. I I I I n I I I I I I I I I I I n D D D I I I I I I I Hardness, Calcium Hardness, Calcium Calculation lodine(ide) Nitrogen, Ammonia Kjeldahl Nitrate Nitrite Nitrite Organic Oil and Grease pH (Hydrogen Ion) Phenol Phosphorus Total Ortho Silica, Dissolved I :\lqaprevalsect9.doc Table 9.2 (cont.) EDTA Titration Ion Chromatography Distillation Titration Potentiometric Digestion/Distillation Automated Cadmium Brucine Sulfate Ion Chromatography Automated Cadmium Ion Chromatography Spectrophotometric Kjeldahl-NH3 Kjeldahl- Potentiometric Soxhlet Partition-Gravimetric IR Electrode Distil)ation-Extraction Colorimetric Persulfate Digestion Ascorbic Acid Reduc. Ascorbic Acid Reduc. Molybdosilicate ICP 3111 4500-NH3 4500-N organic 4500-N03 4500-N03 4500-N02 4500-N organic 55208 55208 4500-H+ 4500-P 4500-P 4500-S1 242.1 200.7 300.0 350.2 350.3 351.3 353.2 352.1 300.0 353.2 300.0 354.1 351.3 351.4 413.1 413.2 150, 1 420.1 365.2 365.2 370.1 200.7 Date: 2/28/97 Section 9.0 Revision 2.0 Page 19 of27 D511 D3590 D3867 D091 D3867 D3590 D1293 D1783 D515 D515 D859 9200 9070 9071 9040A 9065 Table 9.2 (cont.) Specific Conductance Meter 2510 Sulfate Ion Chromatography 4500-S04 2" Turbidimetric Sulfide Colorimetric 4500°S2" Titration 4soo-s2• Sulfite Titration 4500-sot Surfactants Methylene Blue 5540 (MBAS) Turbidity Meter 2130 :~qaprevalsect9.doc 120.1 300.0 375.2 375.4 376.2 377.1 425.1 180.1 Date: 2/28/97 Section 9.0 Revision 2.0 Page 20 of 27 D1339 D2330 D1889 9030A I I I I I I I I I I I I I I I I I g n D E I I I I I I I I Aluminum Antimony Arsenic Barium Beryllium Cadmium Calcium Chromium, Total (Hexavalent) Cobalt Copper Iron :~qapreva\soct9.doc Table 9.2 (cont.) AA-Direct Aspiration 3111 AA-Furnace 3113 ICP-AES AA-Direct Aspiration 3113 AA-Furnace 3113 ICP-AES AA-Furnace 3313 ICP-AES AA-Direct Aspiration 3110 AA-Furnace 3113 ICP-AES AA-Direct Aspiration 3110 AA-Furnace 3113 ICP-AES AA-Direct Aspiration 3110 AA-Furnace 3113 ICP-AES AA-Direct Aspiration 3110 AA-Furnace 3500-Ca ICP-AES AA-Direct Aspiration 3110 AA-Furnace 3113 ICP-AES Colorimetric 3500-Cr MIBK Extraction AA-Direct Aspiration 3110 AA-Furnace 3113 ICP-AES AA-Direct Aspiration 3110 AA-Furnace 3113 ICP-AES I AA-Di~ect 3110 AA-Furnace 3113 ICP-A$S ! Date: 2/28/97 Section 9.0 Revision 2.0 Page 21 of 27 202.1 202.2 200.7 204,1 204.2 200.7 206.2 200.7 208.1 208.2 200.7 210.1 210.2 200.7 213.1 213.2 200.7 215.1 215.2 200.7 218.1 218.2 200.7 219.1 219.2 200.7 220.1 220.2 200.7 236.1 236.2 200.7 7020 60·10A ' 7040 7041 6010A 7060A 6010A 7080A 7081 6010A 7090 7091 6010A 7130 7131A 6010A 7140 6010A 7190 7191 6010A 7196A 7197 7200 7201 6010A 7210 7211 6010A 7380 7381 6010A Lithium Magnesium Manganese Mercury Molybdenum Nickel Potassium Selenium Silica Silver Sodium Strontium :\fqapreva\sed9.doc Table 9.2 (cont.) AA-Direct Aspiration AA-Direct Aspiration ICP-AES AA-Direct Aspiration AA-Furnace ICP-AES AA-Cold Vapor Automated Cold Vapor AA-Direct Aspiration AA-Furnace AA-Direct Aspiration AA-Furnace ICP-AES AA-Direct Aspiration ICP-AES AA-Furnace ICP-AES ICP-AES AA-Direct Aspiration AA-Furnace ICP-AES AA-Direct Aspiration ICP-AES AA-Direct Aspiration ICP-AES 3500-Li 3111 3111 3113 3112 3111 3113 3111 3113 3111 3113 303A 3113 3111 3111 Date: 2/28/97 Section 9.0 Revision 2.0 Page 22 of 27 242.1 200.7 243.1 243.2 200.7 245.1 245.2 246.1 246.2 249.1 249.2 200.7 258.1 270.2 200.7 200.7 272.1 272.2 200.7 273.1 200.7 200.7 6010A 7450 6010A 7460 7461 6010A 7470A 7471A 7480 7481 7520 6010A 7610 6010A 7741A 7740 6010A 6010A 7760A 7761 6010A 7770 6010A 7780 6010A I I I I I I I I I I I I I I I I I I I I I a D u D E I I I I I I I I Thallium Tin Titanium Vanadium Zinc :llqaproval.sect9.doc Table 9.2 (conl) AA-Direct Aspiration AA-Furnace ICP-AES AA-Direct Aspiration AA-Furnace ICP-AES AA-Direct Aspiration AA-Furnace ICP-AES AA-Direct Aspiration AA-Furnace ICP-AES AA-Direct Aspiration AA-Furnace ICP-AES 3111 3113 3111 3113 3111 3113 3111 3113 3111 3113 Date: 2/28/97 Section 9.0 Revision 2.0 Page 23 of 27 279.1 279.2 200,7 282.1 282.2 200.7 283,1 283,2 200.7 286.1 286,2 200.7 289,'I 289,;! 200.7 7840 7841 6010A 7870 6010A 6010A 7910 7911 6010A 7950 7951 6010A 3. Wastes & Oil Analysis %Ash Density Flash Point Closed Cup Free Liquids Leach Test. EP Toxicity Sulfide, Total Reactive pH Specific Conductance Specific Gravity Cyanide, Total Amenable Reactive TCLP :~qapreva\sect9.doc Gravimetric Gravimetric TAG Paint Filter Extraction ntration Titration Electrode Meter Mass & Displacemen t Distill -Color Chlorination- Colorimetric Purge-Color Leach Table 9.2 (cont) 2540 2710 2710 D93-80 261.23 261.23 40CFR268 Date: 2128/97 Section 9.0 Revision 2.0 Page 24 of27 1010 9095 1310A 9030A Chapter 7-7.3.4.2 9040A 9050/9045B 9010A 9010A Chapter 7-7.3.3.2 1311 I I I I I I I I I I n I I I I I I I I I I I I n g D D I I I I I I I I Table 9.2 (cont.) 6. Ust of sample Preparation Methods 1311 TCLP 1312 Synthetic precipitation leaching procedure 3015 Microwave dig. aqueous 3051 Microwave dig. sludges, oil soil 3510 Separatory Funnel Liquid -Extraction 3520 Continuous Liquid -Extraction 3540 Soxhlet Extraction 3541 Automatic soxhlet extraction 3550 Sonication Extraction 3640 Gel Permeation Chromatography 3580 Waste Dilution 3630 Silica gel 3660 Sulfur clean up 5050 Bomb combustion. method for T. Halides 5080 Purge and Trap Date: 2/28/97 Section 9.0 Revision 2.0 Page 25 of 27 3005 Acid Digestion of Waters for Total Recoverable or Dissolved Metals for Analysis by Flame AA or ICP 3010 Acid Digestion of Aqueous Samples and Extracts for Total Metals for Analysis by Flame AA or ICP 3020 Acid Digestion of Aqueous Samples and Extracts for Total Metals by Furnace AA 3050 Acid Digestion of Sediments, Soils, and Sludges 7. Screening Methods 3810 Headspace 3820 Hexadecane extraction and screening of purgeable organics 8. Qllm 40 CFR 261 40 CFR 261 40 CFR 261 40 CPR 261 NIOSH 0600 NIOSH 0500 NIOSH 7500 Characteristic of lgnitability Characteristic of Corrosivity Characteristic of Reactivity TCLP Nuisance Dust, Respirable Nuisance Dust, Total Respirable Silica (XRD) I :~qapreva\seci9.doc Table 9.2 References Date: 2/28/97 Section 9.0 Revision 2.0 Page 26 of27 1. Handbook for Analytical Quality Control in water and Wastewater Laboratories. u.s. EPA · 600/4-79-019, March, 1979. 2. Federal Register, 40 CFR Part 136, October 26, 1984. 3, Test Methods for Evaluating Solid Waste. Physical/Chemical Methods. SW-846. 3rd Edition & Final Updates One and Two, U.S. EPA, revised Sept., 1994. 4. Quality Assurance of Chemical Measurements, Taylor, John K.; Lewis Publishers. Inc. 1987. 5. standard Methods for the Examination of Water and wastewater, APHA, AWWA, WPCF: 18th Edition, 1992. 6. N!OSH Manual of Analytical Methods, U.S. Department of Health, Education, and Welfare; Second Edition. 1977. 7. Methods for Non-conventional Pesticides Chemicals Analysis of Industrial and Municipal Wastewater, Test Methods, EPA-440/1-83/079-C. 8. Methods for Chemical Analysis of Water and Wastes. EPA-600/4-79--020, 1983, 9. The Determination of Inorganic Anions In Water by Ion Chromatography -Method 300.0 Test Method, EPA-600/4-84-017. March, 1984. 10. Environmental Measurements Laboratory <EMU Procedures Manual-HASL-300, us DOE, February, 1992. 11. Requirements for Quality Control of Analytical Pata. HAZWRAP, DOE/HWP-65/R1, July. 1990. 12. Requirements for Quality Control of Analytical Data for the Environmental Restoration Program. Martin Marietta, ES/ER/TM-16. December. 1992. 13. Quality Assurance Manual for Industrial Hygiene Chemistry. AIHA, 1988. :~qapreva\sect9.doc I I I I I I I I I D I I I I I I I I I I I I I D D D I I I I I I I Date: 2/28/97 Section 9.0 Revision 2.0 Page 27 of 27 Table 9.3 BFB Key Ions and Ion Abundance Criteria Mass SW-846, Method 8240B 50 15-40% of mass 95 75 30-60% of mass 95 95 base peak, 100% of rel. abundance 96 5-9% of mass 95 173 less than 2% of mass 174 174 greater than 50% of mass 95 175 5-9% of mass 174 176 95-101 % of mass 174 177 5-9% of mass 176 CLP Statement of Work -VOA 8-40% of mass 95 30-66% of mass 95 base peak, 100% rel. abundance 5-9% of mass 95 less than 2% of mass 17'4 50-120% of mass 95 4-9% of mass 17 4 93-101% of mass 174 5-9% of mass 176 Table 9.4 DFTPP Key Ions and Ion Abundance Criteria Mass SW-846, Method 8270B 51 30-60% of mass 198 68 <2% of mass 69 69 N/A 70 <2% of mass 69 127 40-60% of mass 198 197 <1% of mass 198 198 base peak, 100% rel. abundance 199 5-9% of mass 198 275 10-30% of mass 198 365 <1 % of mass 198 441 Present but less than mass 443 442 <40% of mass 198 443 17-23% of mass 442 CLP Statement of Work -SV 30-80% of mass 198 <2% of mass 69 Present <2% of mass 69 25-75% of mass 198 <1% of mass 198 base peak, 100% rel. abundance 5-9% of mass 198 10-30% of mass 198 <D.75 of mass 198 Present but less than mass 443 40-11-% of mass 198 15-24% of mass 442 I :~qaprevalseot9.doe I I g g H n E I I I I I I I I 10.0 DATA REDUCTION, VALIDATION AND REPORTING Date: 12/22/95 Section 1 o. O Revision 0.01 Page 1 of 8 Data reduction, validation and reporting describes the processes that result in the delivery of quantitative analytical data to the data user. These processes include calculation of raw data into final concentration units, reviewing results for accuracy and assembly of the technical report contents for delivery to the data user. All analytical data generated within the Pace laboratories undergo a well-defined, well-documented multi-tier review process before being reported to the client. The following describes procedures employed at Pace for translating raw analytical data into accurate, finished sample reports and data storage. Figure 10.2 shows schematically the sample flow through the laboratory. while Figure 10.3 shows the parallel flow of information concerning the sample analysis and reporting. 10.1 DATA REDUCTION When primary analytical data, otherwise known as "raw data." are manually generated, the data are recorded either in bound logbooks with prenumbered pages or on preprinted forms. Records of analysis indicate the method used, raw data, calculations, and final· results. Entries are made in black ink and are initialed and dated by the individual who makes the entry. It is acceptable to initial and date once for an entire page. Errors are corrected by drawing a single line through the entry; this change is initialed and dated by the individual who makes the change. Raw data may not be obscured in any way. The use of white-out is prohibited on all raw data. including instrumental hardcopy. All data generated by Pace are reviewed by designated, trained personnel. The analysts who acquire the data are responsible for initial on-line checks for compliance to the analytical requirements. After a sample batch is acquired, the data review procedure includes data interpretation and quantitation, inspection of quality control data against criteria, data reduction, narrative or comments writing, and ensuring that the data package includes all required analytical and quality control results, raw data an_d laboratory chronicles. The analyst who completes the analysis assembles all relevant raw data and results together with chromatograms, strip chart recordings, instrument settings and other information essential to data interpretation. For data which are reduced by manual calculations, the calculations are documented in a laboratory notebook or on an analyst's · worksheet. The results are transferred to a standardized laboratory reporting form which has been approved by the appropriate Group Supervisor and Laboratory Operations Manager. Reporting forms include at a minimum the sample identification number, the date analyzed. the result expressed per unit volume, the method reference and the analyst's initials. From the reporting forms, the results are entered into the LIMS. 10.2 DATA VALIDATION Data validation is the process of examining data and accepting or rejecting it based on pre-defined criteria. Pace data review personnel use the following .. criteria to validate laboratory data: :\Jqapreva\sect1 O.doc • Use of approved analytical procedures. • Use of properly operating and calibrated instrumentation. Date: 12/22/95 Section 1 O. O Revision 0.01 Page 2 of 8 • Precision and accuracy comparable to that achieved in similar analytical programs. Analysts performing the analysis and subsequent data reduction have the primary responsibility for the quality of the data produced. The primary analyst initiates the data validation process by reviewing and accepting the data, provided QC criteria have been met for the samples being reported. Data review checklists may be used to document the data review process. The completed data package is then sent to the Group Supervisor or designated reviewer. The Group Supervisor provides a technical assessment of the data package and technical review for accuracy according to methods employed and laboratory protocols. This involves a quality control audit for use of the proper methodology and detection limits, compliance to quality control protocol and criteria, presence and completeness of required deliverables, and accuracy of calculations and data quantitation. Group Supervisors also review analyst generated calculations. For data which are reduced via computer, calculations are checked by the analyst (or· designee) assigned to this task at a frequency designed to assure that the data manipulations are valid. This data validation step is documented by the analysts' initials on the hardcopy of the raw data. The results are either manually transferred to a standard reporting form or reported via computer generation of forms. Once the data have been technically reviewed and approved, authorization for release of the data from the analytical section is indicated by initialing and dating the data review checklist or otherwise initialing and dating the data. The Group Supervisor drafts any narrative comments if required by the Quality Assurance Project Plan, and forwards the report and the data package to the Reporting Department. Each data package is reviewed by designated reporting personnel to ensure compliance with client orders by reviewing on-line input in the Pace computer tracking system. The laboratory data are assembled in the client's technical reports. Reports are reviewed for completion prior to copying and binding. Figure 10.1 provides a summary listing of staff responsibilities concerning data generation, review, validation and reporting. The Reporting Department assembles the data with other data from the sample set, generates the final report, checks for transcription errors, and provides the final report to either the Laboratory Operations Manager, Project Manager, or an appropriate designee for final signature. The Operations or Project Manager examines the report for method appropriateness, detection limits and whether or not QC criteria were satisfied. Any deviations from the referenced methods are checked for documentation and validity, and QC corrective actions are reviewed for successful resolution. The Operations or Project Manager or an appropriate designee signs the completed report prior to its release to the client. :\lqapreva\sect1 O.doc I I I I I I I I I I I I I I I I I I I I n D H I I I I I I I I Date: 12/22/95 Section 10.0 Revision 0.01 Page 3 of 8 The Operations Manager may delegate the final review and signing of reports as necessary. Use of checklists ensure that all data are systematically handled and no steps are omitted. Checklists are reviewed and are retained and accessible should they need to be referenced at a later date. The data and deliverables are checked and signed during processing procedures and then systematically filed by reference identification numbers. 10.3 DATA REPORTING All data segments pertaining to a particular Pace Laboratory Number are channeled to the Reporting Department for assembly into the final report format and generation of the analy1ical narrative. All points mentioned during technical and QC review are included in the narrative if it is deemed to impact the quality of the data. The final report is given to either the Laboratory Operations Manager, Project Manager or an appropriate designee for final review and release. After verifying the report's completeness and accuracy, the Operations or Project Manager signs the cover letter or authorization line within report indicating acceptance of the report. Technical reports are prepared to include the components or level of deliverables that are requested by clients for samples or projects, or contractually required. The standard Pace commercial report to the client consists of the following sections: 1) A cover letter 2) A technical narrative (when necessary) 3) Sample receipt condition report (information may be included on C.O.C. form) 4) Sample ID table 5) Sample results 6) Chain-of-Custody forms The narrative briefly describes the condition of the samples upon receipt, sample holding time performance, instrument calibration information, and the quality control results. Any discrepancies discovered and matrix problems encountered are also addressed in this section. The sample results are tabulated by sample number and parameter. Pace number, client identification, and dates of sample preparation and analysis are presented along with the observed concentrations for each parameter analyzed and corresponding reporting limits. Pace prepares technical reports that include full data deliverables for validation purposes, and lesser, abbreviated reports. Full deliverables include all raw and processed data applicable to the analyses performed. Pace prepares single sample technical reports or multi-sample report packages. The multi-sample technical reports contain results for a sample delivery group (SDG) or other client or laboratory defined sample set. Pace recommends multi-sample reports when full deliverables packages are required. :\lqapreva\sed1 O.doc Date: 12/22/95 Section 10.0 Revision 0.01 Page 4 of 8 The Pace laboratory prepares electronic data deliverables (EDD) as required for contracts and upon client request. 10.4 DATA ARCHIVE Each data report which supports the analytical process for all samples received by the laboratory is thoroughly reviewed for completeness and accuracy. After the technical review it is routed to the Reporting Department for assembling the final report for submission ·to the client. The report is approved, signed, and submitted. Sufficient records are retained to recreate analytical events at the laboratory. Pace will retain analytical data for five years and financial data for three years relating to services performed following the transmittal of the final report to the client. Certain contractual arrangements or regulatory requirements for specific projects may shorten or extend the records retention period stated here. Records are catalogued and maintained in limited access areas. Data archive and storage is managed by designated individuals who control the access to stored information. One copy of the report remains with all the raw data which is stored in the data archives under the control of the QA Department or other designated group. All information retained at the Pace facility is stored in secured areas. The Data Archivist has oversight responsibility for the data archive ensuring the continued integrity of all documentation generated in support of laboratory analyses. All hard-copy information is stored on-site at the laboratory or off-site at a commercial document storage facility equipped with a professional security system. All electronic data is stored on-site at the laboratory or off-site at a commercial document storage facility equipped with a professional security system and a controlled environment suitable for storage of magnetic media. The archive room is a secure storage area· with limited access to non-authorized personnel. Sign-out procedures are in place where every document removed from the archive room must be signed out by authorized personnel. A copy of the report or summary of samples classified as hazardous is forwarded to the Hazardous Waste Coordinator or designee for use in characterizing the samples for ultimate disposal. Pace reserves the right to transfer hard-copy information onto microfilm or write- protected electronic media. Pace reserves the right to store information in hard-copy files, on magnetic media and/or microfilm. The information is retained and accessible for a minimum of seven years unless otherwise specified through a client specific contract. 10.5 RESPONSE TO INQUIRIES The Pace laboratory which conducted analyses for the client recognizes the importance of its timely response to inquiries regarding the laboratory's work for samples and :\lqapreva\sect1 O.doc I I I I I I I I I I B I I I I I I I I I I u D D E I I I I I I I Date: 12/22/95 Section 10.0 Revision 0. 0 1 Page 5 of 8 projects. The laboratory will respond to inquiries as rapidly as possible as part of its corrective action plan. The Pace laboratory which originally received samples from the client should be considered the primary contact for all data inquiries when subcontract or other Pace laboratories are used for analyses. :\lqapreva\sect1 O.doc FIGURE 10.1 Date: 12/22/95 Section 10.0 Revision 0.01 Page 6 of 8 Analytical Data Review Process, Pace Analytical Services, Inc. Analyst Supervisor Manager Data Review/Reporting Quality Assurance Office Project Manag_er Responsibilities Sample analysis LIMS* entry and generation Data review -1st level (bench) Control charting -real time Narrative notes Discrepancy initiation Provide copies of log books, as necessary Oversee daily analytical activities Review control chart comments daily LIMS data entry and validating Draft and review of narrative Supervise contractual and technical compliance Discrepancy review Review quality control daily (calibrations, etc.) Sign-off case narrative Ensure program compliance Review discrepancies requiring manager resolution Technical conference calls with client Ensure technical validity of data Generate forms package Final data review and validation Prepare package and paginate Electronic deliverables generation Maintain data package files 10 percent contractual compliance review (data packages) C_ustody when required; Calculations; -Methods criteria; QC criteria; Forms; and Control charting. Review narratives for accuracy Review packages for completeness and quality Cover letter Collate organic and inorganic packages Client/laboratory liaison Deliver package to client Note: *Laboratory Information Management System :\lqapreva\sect1 O.doc I I I I I I I I I I I I I I I I I I I I I I I D n n D E m • I I I I I I I I :\lqapreva\sect10.doc FIGURE 10.2 Laboratory Sample Flow Schematic Samples 4 --- Samples In ~~ ...---------- Storage (Cooler) ~ ~ Sample Prep. I Check-In J Sample Analysis ~---'-----~.--'---~ Data Production I Control Ch rts I _,,,om D:::noy ac ""z___/ I a . ~ Data A/9 Entered Into LOMS • • \ Data Validation I Quality Assurance Officer I Mll!yst/SuperviSOI' Valida!H Both QC & Sample Resits Report Generation Report Review Project Manager/Dept Manager Review/Sign ~---'---~....-----...i-------~ Report Sent to Client j Samples Returned to Client "--....1 Sample Disposal I Date: 12/22/95 Section 10.0 Revision 0.01 Page 7 of 8 :\lqapreva\seci10.doc FIGURE 10.3 Information Flow Schematic Samples4' '-.... ~--~ Real . I Check-In I ~ /I ~ Samples In g_ ~ ~-·-~ Project Manaoe, Storage (Cooler) ~ NIGHTL y Clienl File Modified U>MS To Pel'fom, Analyt,is Subsets Outstanding Analysis Sutm,ts Permanent Rll<Xlfd Bench Analyst (available Maf\llQ8fflenl ·-..:a A)~.SIS RESULTS TolFrom Storage L . t ~ RelUlts Entry ,.-:-... _.,_--· VIDEO DISPLAY Validation List Outstanding Analysis Another Bench Analyst (available on request• 2ncl shift) Subsets Management Valicsate Results ~ RES UL TS (NigN!y OnlrJ _...,. N-/Changed A<:coun1ing Batch LOMS (NoontNighl) v""'"" Lid Cloud _, DISK Projects ~ -------.._I ,_""' I ·-lt•m• =11"1-.oidnQ """""' ,_ ........ -Nllt,tkll ,.,_., o.ta Pl'Odudlon Refl(lrtS control Ub lnwldng Clltrt OP Rcpon, Ult, ln-.oidng Cltrt. (Nlglllyont,) Con1rOI (Nlgf'CIJ~ I Date: 12/22/95 I Section 10.0 Revision 0.01 Page 8 of 8 I I I I I I I I I I I I I I I I I I I I I a 0 D u I I I I I I I I 11.0 QUALITY CONTROL PROCEDURES Date: 12/22/95 Section 11 Revision 0.01 Page 1 of . .42 A quality control (QC) program is a systematic process that controls the validity of analytical results by measuring the accuracy and precision of each method and matrix, developing expected control limits, using these limits to detect errors or out-of-control events, and requiring corrective action measures to prevent or minimize the recurrence of these, events. QC procedures are implemented to ensure that sample data meet the quality objectives of the laboratory and the client. An effective QC program must be able to control the quality of the data through the monitoring of QC indicators. Criteria frequently applied to environmental QC data include measurements of accuracy and precision. Precision measures the randomness associated with an analytical measurement and reflects the inherent variability in that measurement system. Accuracy reflects the degree to which the measured value approximates the actual or "true" value for a given parameter and reflects the influence of systematic biases in the measurement. Thus, the "quality" of QC data can be said io be a measure of both the randomness and biases in a specific measurement system. This section addresses the specific QC procedures applied to representative analytical methods performed at Pace. Table 11.1 presents method-specific information about QC procedures, acceptance criteria, and required corrective actions for the various analysis types. 11.1 ACCURACY AND PRECISION MEASUREMENT CONVENTIONS The results of quality control samples created in the laboratory represent estimates of accuracy and precision for the preparation and analysis steps of sample handling. This section describes the quality control information provided by each of these analytical measurements. Information on the procedures to follow in preparation of the samples or spiking solutions is described for each method and matrix in the respective method Standard Operating Procedure. Method Blank A method blank is a volume of deionized and/or distilled laboratory water for water samples, or a purified solid matrix for soil/sediment samples, carried through the entire analytical procedure. The volume or weight of the blank must be approximately equal to the sample volume or weight processed. Analysis of the blank verifies that method interferences caused by contaminants in solvents, reagents, glassware, and other sample processing hardware are known and minimized. Optimally, a method blank should contain no greater than five times (5X) the method detection limit, or reporting limit where applicable, for common laboratory solvents and phthalate esters; less than the detection (or reporting) limit for all other parameters unless otherwise specified in the method or project QA plan. Results of method blank analyses are maintained with other QC data in the respective laboratories. If requested by the client, this data will be included in the report. :\lqapreva\sect11.doc Date: 12/22/95 Section 11 Revision 0.01 Page 2 of 42 Method Blank Frequency Organics: The laboratory shall prepare and analyze QM laboratory reagent blank (method blank) for each group of samples of a similar matrix (for water or soil samples), extracted by a similar method (separatory funnel, continuous liquid-liquid extraction, or sonication), and a similar concentration level (when low vs. medium level analyses are available) for: • every 20 samples, or • whenever samples are extracted -whichever is more frequent. lnorganics: At least QM preparation blank (method blank), consisting of blank reagent water processed through each sample preparation and analysis procedure, shall be prepared and analyzed with every group of 20 samples, or with each batch (a group of samples prepared at the same time, e.g. daily) of samples digested, extracted, prepared or directly analyzed, whichever is more frequent. Accuracy Measurements Laboratory Control Samples (LCS) consist of aliquots of laboratory blank matrices · (water, sand, etc.) spiked with analytes of interest. LCSs for methods with extensive lists of analytes that may interfere with one another may include a limited number of analytes, but the analytes included must be representative of as many analytes as is practical. In the case of metals analysis, all analytes of interest must be included. Laboratory pure water is used to prepare most LCSs for methods for analysis of water. Highly characterized solids, where available, are used for LCSs for methods for analysis of solids. Where no such solid LCS is available, spiked laboratory pure water or spiked reagent blanks may be substituted. LCSs provide an estimate of bias based on recovery of the compounds from a clean, control matrix. They provide evidence that the laboratory is performing the method within accepted guidelines without potential non-matrix interferences. They are prepared at a rate of one per batch of twenty or fewer samples. For tests that are performed infrequently, an LCS shall be analyzed at least monthly if the number of samples is less than 20. This monthly requirement shall NOT apply to low-volume tests for which state certification is not sought, or for tests expected to be performed solely as part of a special project, or for tests involving study specific matrices other than water, soil, sludges and oils. Matrix Spikes/Matrix Spike Duplicates are similar to Laboratory Control Samples except the analytes used for spiking are added to a second and third separate aliquot from the same container of selected client samples in a batch of analyses. They :\lqapreva\sect11.doc I I I I I I I I I I I I I I I I I I I I D I I I I I I I I I Date: 12/22195 Section 11 Hevision 0.01 Page 3 of 42 enable one to assess sample matrix effects and field conditions. MS/MSDs are routinely prepared at a frequency of 5% (one set per twenty samples) when adequate sample volume is provided or once every 14 days, which ever is more frequent. An LCS/LCSD pair shall be substituted when sufficient sample volume is not available to prepare an MS/MSD sample set. Surrogates provide an estimate of bias based on recovery of chemically similar compounds which are not expected to be in the sample, to the compounds of interest for each sample, incorporating sample matrix effects and field conditions. Surrogates are added to all samples analyzed by GC/MS and certain GC analyses prior to sample preparation. An Internal standard is an analyte that has the same characteristics as the surrogate, but is added to each sample in a batch, just prior to analysis and is used for quantitation. It corrects for bias or change in instrument performance from sample to sample, incorporating matrix effects associated with the analytical process only. Accuracy is expressed as % Recovery. For LCSs, Surrogate, and Blank Spike samples, percent recovery (%R) is calculated as: Where: %R =(SR/ SA) x 100 SR is the concentration determined SA is the concentration spiked For the matrix spike samples, the percent recovery is calculated as: Where: %R = (SSR-SR)/SA x 100 SSR is the spiked sample determined result SR is the original sample determined result SA is the amount of spike added (expected) Precision Measurements A Sample Duplicate is a sample that has been homogenized and split into two equal portions before the method sample preparation process. It measures sample precision associated with the preparation through analysis and is prepared and analyzed at a rate of one per batch or one per twenty samples or once every 14 days whichever is greater in the inorganic laboratories. For organic analyses the MS/MSDs fulfill this function and provide a measure of overall precision. The comparison of the values determined for a sample and its duplicate (MS/MSD) is expressed as relative percent difference (RPO). This calculation is as follows: :\lqapreva\sect11.doc IS-DI RPD = ---x 100 [(S+D)/2] Date: 12/22/95 Section 11 Revision 0.01 Page 4 of 42 The vertical bars in the above equation indicate the absolute value of the difference, hence RPO is always expressed as a positive value. 11.2 CONTROL CHARTS Control charts are quality control tools which graphically display the QC parameters over time. Accuracy (Figure 11.1) and precision (Figure 11.2) control charts are generally maintained for each method; however, for certain methods tabulated control limits are used to monitor acceptability of quality control measurements. Each chart can be broken into three parts: sample identification, sample response/calculation, and graphic representation (the plot). 11.2.1 11.2.2 Accuracy Accuracy charts are maintained for Surrogate and Laboratory Control Sample recovery. Each sample is identified by the date it was analyzed and its Pace sample number. The percent recovery is plotted onto the graph where: The x-axis is the sample ID.; and • The y-axis is the range of percent recoveries. Precision In cases where precIsIon charts are maintained, the relative percent difference is plotted on the graph where: The median, zero, represents 0% difference • The x-axis is the number of data points per chart; and • The y-axis is the range of relative percent differences. Both samples are identified by the date(s) analyzed and their Pace number. 11.2.3 · Limits :\lqapreva\sect11.doc Both upper and lower warning limits and upper and lower control limits are established to interpret performance. Warning limits express a _narrower confidence interval and are used to warn the analyst or supervisor of possible system inconsistencies or failures, before an out-of-control event I I I I I I I I R I a I I I I I I I I I I a n D D I I I I I I I I I I Date: 12/22/95 Section 11 Revision 0.01 Page 5 ot-12 occurs. Control limits express the outer limits of accepted method variability. Control limits and warning limits are reviewed periodically against performance. Based on statistical considerations, an evaluation is made to determine whether the control limits need to be revised. warning Limits When not othe_rwise mandated by the method, Pace adopts warning limits to be the mean +/-2 standard deviations or a 95% confidence interval, where: 1 n Mean X = L Xi n I= 1 Standard Deviation n n z: X; 2 -(Z:X,)2/n i=1 i=1 s = or :\lqapreva\sect11.doc n -1 In this equation, control Limits n -1 = population size = ith observation in the sample = sample mean Unless otherwise stipulated in a particular method or program, acceptance limits (control limits) will be statistically derived from laboratory generated data. Control limits will be based upon +/-3s (i.e., 99% confidence interval) from the mean and warning limits established at +/-2s from the mean (i.e., 95% confidence interval). All data used to generate these limits will undergo a Dixon Outlier test to reject outlier data points. The use of hardcoded limits nonstatistically derived from current laborato,y generated data (e.g., CLP limits) shall be limited to programs which specifically permit their application. Control limits shall be updated annually at a minimum and at a maximum interval of once every 20 data points generated. At a minimum, tabulated control limits shall be available and followed by all analysts performing the associated test. However, control charts are the preferred mechanism for monitoring quality control measurements on a real 11.3 Date: 12/22/95 Section 11 Revision 0.01 Page 6 of 42 time basis. Where interlaboratory expected ranges have been determined, Pace's goal is for their control limits to fall within these multi-laboratory expected ranges for that method. Suspicious/Out-of-Control Events Plotting and connecting successive data points on control charts enables the laboratory to detect many types of suspicious and out-of-control situations. These events can be caught by monitoring the following: outliers (suspicious and out-of-control), runs (suspicious), trends (suspicious), and periodicity (suspicious). Excursions There are two types of excursions: any particular point that falls outside the control limits or any point that falls outside the warning limits. A point that falls outside the control limits is classified as an out-of-control event; a point that falls outside the warning limits is classified as a suspicious event. Runs I I I I I I I I I A run is defined as a series of points that line up on one side of the central I line (the mean). Any run that has a length of seven points is indicative of a potential abnormality in the process, a suspicious event. A run can suggest 1 several potential problems such as a leak in the system, elevated contamination, or incorrect dilutions of standards. Trends I A trend is defined as a series of points that are marked by an unbroken rise II or fall. Any trend with a length of five points (may vary up to seven points) II is classified as a suspicious event. A trend may indicate a change in instrument sensitivity due to a dirty source or injection port or standard degradation, to name a few. I Periodicity Periodicity is a term used to describe a recurring pattern of change over equal intervals. This occurrence may be of any length or amplitude; thus, · careful observation of the control chart is necessary. QC BATCH DEFINITION Organics: The laboratory will perform: I • One spiked sample analysis (matrix spike), and :\lqapreva\sect11.doc I I I I I I g I D D n I I I I I I I I I I Date: 12/22/95 Section 11 Revision 0.01 Page 7 of 42 • One duplicate spiked sample analysis (matrix spike duplicate) for each group of samples of a similar matrix (for water or for soil samples) and concentration level (when low vs. medium level analyses are available) for: every 20 samples, or .• each 14 calendar day period during which field samples were received (said period beginning with the receipt of the first sample by the laboratory), whichever is more frequent. • One spiked laboratory control sample (LCS) must be processed each time a group of 20 samples or less are extracted, prepared, or directly analyzed (Note: an organic LCS is processed at the same frequency as a laboratory reagent blank (method or prep blank). lnorganics: The laboratory will perform: • One spiked sample analysis (matrix spike), and • One straight sample duplicate analysis {duplicate) for each group of samples of a similar matrix (for water or for soil samples) and concentration level (when low vs. medium level analyses are available) for: every 20 samples, OR each 14 calendar day period during which field samples were received (said period beginning with the receipt of the first sample by the laboratory), whichever is more frequent. .• One spiked laboratory control sample (LCS) must be processed each time a group of 20 samples or less are digested, extracted, prepared, or directly analyzed (Note: an inorganic LCS is processed at the same frequency as a laboratory reagent blank (method or prep blank). The frequencies listed above for matrix spiking applications are to be followed regardless of whether or not clients have committed to "paying for QC." Laboratory operations must make a conscious effort to periodically request or collect as part of a field sampling event (i.e., sampling performed by Pace personnel) sufficient quantities of samples for those analyses (e.g., method 418.1 TPH) that are routinely analyzed under QC batches that do not contain matrix spike applications because "insufficient sample vol4me was received." Samples selected for QC which contain limited volume or quantity should be evaluated based upon the type of analysis to be performed to :\lqapreva\sect11.doc r Date: 12/22/95 Section 11 Revision 0.01 Page 8 of 42 establish whether modifications can be made to allow for using smaller initial sample size than normally applied. For example, most organic extraction procedures require that a 1 liter sample size be analyzed for .aqueous samples. However, often only a single 1 liter sample bottle remains following the extraction of the safT]ple thus leaving insufficient volume to perform both an MS and MSD on the sample because each normally require a full liter of sample. In most cases, the usability of the QC data for assessing the accuracy and precision of the analysis is not adversely impacted if the remaining liter of sample is split into two 500 ml aliquots, spiked and carried through the procedure. While following this modification would result in a two fold increase in MDLs, since the spiked compounds are present at concentrations which are close to the midpoint of the calibration curve, the elevated MDLs will have no effect on determining recovery and RPO values. In this example, a further method modification of concentrating the final volume of the MS/MSD extracts to 0.5 ml (versus the normal 1.0 ml final volume) would provide the 1000 fold concentration requirement of the method. Any options, such as the example given above, contemplated for use to overcome limited sample size when applying QC applications must be discussed with clients prior to their implementation. Finally, for methods in which no similar technical justification can be made for decreasing the initial sample size or changing the analysis · process, when none of the associated samples in a QC batch contain sufficient sample volume or quantity to permit matrix spiking to be performed, laboratory control samples should be analyzed in duplicate (LCS/LCSD) to afford assessment of both the accuracy and precision of the test. 11.4 UTILIZATION OF QUALITY CONTROL DATA The purpose for preparing and analyzing quality control samples is to demonstrate, through the known entities, how accurate and precise the investigative sample data are. Table 11-1 summarizes the quality control assessment criteria by matrix for the most commonly used methods by Pace. Different criteria may be dictated by different methods or by project QA plans. 11.5 SAMPLING QUALITY CONTROL I I I I I I I I B I I I I Quality ·control is an integral part of sample collection as well as laboratory operations. I Sample collection protocols must include checks to ensure that the sample collected is representative of the site from which it was collected and free from collection-related contamination or biases. Although different laboratory procedures will be used to I analyze for the various parameters of interest, certain general QC procedures are applicable to most sampling methods. QC procedures frequently applied in the field are described below. The analysis types and frequency of collection for each of these field I QC samples are detailed in each project's sampling and analysis plan (SAP). 11.5.1 Fjeld Blanks -Field blanks are QC samples consisting of blank water that are prepared in the field. This type of QC sample serves to check for potential contamination that may be present in the environment where field samples are collected. :\lqapreva\sect11.doc I I I I I I n n D D I I I I I I I I I I 11.5.2 11.5.3 11.5.4 Date: 12/22/95 Section 11 Revision 0.01 Page 9 of-42 Trip Blanks -Trip blanks are similar to field blanks except that they are prepared in the laboratory before the sampling event. These blank samples accompany the other sample containers to the field and then accompany the collected samples back to the lab. Trip blanks serve to check for potential contamination that samples and sample containers may be exposed to during transportation to and from the field. Equipment Rinsate Blanks -These field QC samples consist of rinsates of the equipment used to collect field samples using blank water provided by the laboratory. Equipment rinsate blanks serve to check the adequacy of equipment cleaning between successive sample collections. · Inadequate cleaning of sample collection equipment after the collection of a sample could result in the contamination of the next sample collected. Matrix Spike/Matrix Spike Duplicate Samples -At a minimum frequency of one set per 20 samples, split sample volumes are collected to be used for matrix spike (MS) and matrix spike duplicate (MSD) analyses by the · laboratory. 11.6 LABORATORY QUALITY CONTROL In addition to the sampling-related QC procedures described above, additional QC procedures are performed in the laboratory as part of routine analytical protocol. These procedures are described below for representative analytical methods. 11.6.1 '.\lqapreva\sect11.doc GC Methods Analytical quality control procedures for GC analyses are described in Method 8000A of SW-846, 3rd Edition, Final Update 1 and 2, and the EPA CLP Organic SOW. They include the following: • Initial demonstration of proficiency • Retention time window determination • Surrogate spiked sample analysis • Method blank analysis • Matrix spike/matrix spike duplicate analysis • Laboratory control sample analysis The application of each of these analyses is described below. Initial Demonstration of Proficiency {SW-846) -Before sample analysis can begin, the laboratory must perform a one-time demonstration of the ability to generate data with acceptable accuracy and precision. This is accomplished by analyzing four aliquots of a QC check sample by the same procedure used to analyze samples. The calculated average recovery and standard deviation :\lqapreva\sect11.doc Date: 12/22/95 Section 11 Revision 0.01 Page 10 of42 for each analyte of interest are compared to acceptance criteria provided in the specific SW-846 method. If the calculated accuracy and precision data are within acceptance limits, analysis of samples may proceed. If not, remedial action must be taken to improve system performance and the proficiency test must be repeated. Retention Time Window Determination -Retention time (RT) windows are calculated for each target analyte peak(s) and for each GC column used for sample analysis. To establish RT windows, the laboratory measures the RTs of each analyte peak (or of each selected peak for multi-component analytes) from three analyses of the continuing calibration standard over a 72-hour period. The RT window is determined as .±3 times the standard deviation of the three measured RTs. Daily RT windows are established for each analyte peak using the RT in the daily calibration verification standard as the centerpoint of the window determined above. In successive continuing calibration standards, the RT of each analyte peak must fall within the prescribed RT window for the analysis sequence to continue. RT windows must be recalculated whenever a new GC column is installed. Surrogate Spiked Sample Analysis -Surrogates are compounds that have similar chemical properties to analytes of interest except that they are not expected to occur naturally in environmental samples. The use of surrogate compounds may be project dependent and limited by the ability to select a suitable surrogate for a particular analytical method. Representative surrogate compounds and surrogate recovery acceptance limits for GC methods are given in Section 5, Tables 5.1 to 5.4. For these methods, corrective action must be taken if the surrogate spike recoveries in any analysis fall outside the prescribed acceptance limits. Corrective actions include: • • • • Checking for errors in the calculation or preparation of the surrogate or standard solutions. Checking instrument performance . Recalculating the data and/or reanalyzing the sample or extract if any of the above checks reveal a problem. Re-extracting and reanalyzing the sample if none of the above are determined to be the problem. Method Blank Analysis -For analysis by purge-and-trap methods, a method blank must be analyzed each day of analysis. For extraction methods (including methanol extraction of volatiles for purge-and-trap analysis), at least one method blank must be extracted and analyzed for each batch or sub-batch of samples extracted to demonstrate that both the extraction and analytical systems are free from contamination. Blank samples are carried through all stages of sample preparation and analysis. Lack of contamination is demonstrated if no target analytes are present at concentrations at or I I I I I I I I I I I I I I I I I I I I n D n I I I I I I I I I I Date: 12/22/95 Section 11 Revision 0.01 Page 11 o_f 42 above their reporting limits (or contract required quantitation limits (CRQLs) for CLP). Holding Blanks -Holding or refrigerator blanks are prepared in the laboratory and stored in the refrigerators where VOA samples reside. Holding blanks are analyzed each week and are used to monitor the potential of laboratory contamination. Matrix Spike/Matrix Spike Duplicate Analyses -At a minimum frequency of one set per batch of up to 20 samples of similar matrix, replicate aliquots of one of the samples are spiked with a mix of target analytes and the resulting matrix spike (MS) and matrix spike duplicate (MSD) samples are analyzed to evaluate the percent recovery of the spiked compounds. Representative limits for percent recovery are shown in Section 5, Tables 5.1 to 5.4. Recovery data falling outside the acceptance limits may indicate a problem in sample preparation or in the analytical system, or may be due to sample matrix interference. Analysis of laboratory control samples (LCSs) in conjunction with MS/MSD samples aids in determining whether or not the · problem is sample matrix related. Acceptable recoveries of LCS spike analytes indicate that the analytical system is in control and that problems with associated recoveries in the MS and MSD samples are likely due to sample matrix interference. Laboratory Control Sample Analysis -The laboratory control sample (LCS) consists of a subset of target analytes of interest (typically the same as in the MS and MSD samples) spiked at concentrations in the mid-calibration range. The LCS is prepared along with the samples for analysis and is used to verify that the analytical system is in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in method/instrument performance. 11.6.2 GC/MS Methods :\Jqapreva\seci11.doc Analytical quality control procedures for GC/MS analyses are described in methods 8000A, 8240B, and 8270B in SW-846 or in the EPA CLP Organic SOW. They include the following: • Initial demonstration of proficiency • Mass spectrometer sensitivity check • Daily GC/MS performance test • Surrogate spiked sample analysis • Method blank analysis • Matrix spike/matrix spike duplicate analysis • Laboratory control sample analysis The application of each of these analyses is described below. ;\lqapreva\sect11.doc Date: 12/22/95 Section 11 Revision 0.01 Page 12 of42 I nitjal Demonstration of Proficiency (SW-846) -Before sample analysis can begin, the laboratory must perform a one-time demonstration of the ability to generate data with acceptable accuracy and precision. This is accomplished by analyzing four aliquots of a QC check sample by the same procedure used to analyze samples. The calculated average recovery and standard deviation for each analyte of interest are compared to acceptance criteria provided in the specific SW-846 method. If the calculated accuracy and precision data are within acceptance limits, analysis of samples may proceed. If not, remedial action must be taken to improve system performance and the proficiency test must be repeated. Mass Spectrometer Sensitivity Check -If the extracted ion current profile (EICP) area for any internal standard changes by more than a factor of two compared to the daily calibration verification standard, the mass spectrometer must be inspected for malfunctions and corrective action taken. Samples analyzed while the system was malfunctioning must be reanalyzed. Daily GC/MS Performance Tests -Each day that analyses are performed, the GC/MS system must be checked using bromofluorobenzene (BFB) for volatiles analysis or decafluorotriphenylphosphine (DFTPP) for semivolatiles analysis. The acceptance criteria presented in Section 9 must be met prior to performing any sample analyses. If all criteria are not met, the instrument must be retuned and the test repeated until all criteria are met. Surrogate Spjked Sample Analysis -All samples are spiked with surrogate standards as described in the specific methods in SW-846 and the CLP SOW. The surrogate compounds and representative surrogate recovery acceptance lim~s for GC/MS methods are shown in Tables 5.5 and 5.6. If the surrogate spike recovery in any analytical run is not within limits, the following steps must be taken: • Check for errors in the calculation or preparation of the surrogate or standard solutions. • Check instrument performance. • Recalculate the data and/or reanalyze the sample or extract if any of the above checks reveal a problem. • Re-extract and reanalyze the sample if none of the above are determined to be the problem. Method Blank Analysis -For volatiles analysis by GC/MS, a method blank must be analyzed within each 12 hour run sequence. For semivolatiles analysis, at least one method blank must be prepared and analyzed for each batch or sub-batch of samples extracted, to demonstrate that both the extraction and analytical systems are free from contamination. Blank samples are carried through all stages of sample preparation and analysis. I I I I I I I I I I I I I I I I I I I I I I 0 D E I I I I I I I I I I I 11.6.3 :\lqapreva\sect11.doc Date: 12/22/95 Section 11 Revision 0.01 Page 13 of 42 Lack of contamination is demonstrated if no target analytes (with the exception of common laboratory solvents) are present at concentrations at or above their reporting limits (CRQLs for CLP). For volatile analyses, common laboratory contaminants, such as methylene chloride, acetone, 2-butanone, and toluene, must not exceed five times the CRQL for CLP or five times the reporting limit for SW-846. For semivolatile analyses, the concentrations of the most commonly encountered laboratory contaminants, phthalate esters, must not exceed five times the reporting limit or CRQL. Holding Blanks -Holding or refrigerator blanks are prepared in the laboratory and stored in the refrigerators where VOA samples reside. Holding blanks are analyzed each week and are used to monitor the potential of laboratory contamination. Matrix Spike/Matrix Spike Duplicate Analysis - A minimum of one set of matrix spike (MS) and matrix spike duplicate (MSD) samples is prepared for each analytical batch of up to 20 samples of similar matrix. Acceptance limits for percent recovery are shown in Section 5, Tables 5.5 and 5.6. Recovery data· falling outside the prescribed acceptance limits may indicate a problem in sample preparation or the analytical system, or may be due to sample matrix interference. Analysis of laboratory control samples (LCSs) in conjunction with MS/MSD samples aids in determining whether or not the problem is sample matrix related. Acceptable recoveries of LCS spike compounds indicate that the analytical system is in control and that problems with associated recoveries in the MS and MSD samples are likely due to sample matrix interference. Laboratory Control Sample Analysis -The laboratory control sample (LCS) consists of a subset of the target analytes (typically the same as in the MS and MSD samples) spiked at concentrations in the mid-calibration range. The LCS is prepared along with the samples for analysis and is used to verify that the analytical system is in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in meth_od/instrument performance. Metals Analysis The quality control procedures applied to metals analysis by ICPS are described in SW-846 Method 6010A and in the CLP Inorganic SOW. Quality control procedures for atomic absorption analyses are described in SW-846 Method 7000 series and the CLP Inorganic SOW. These procedures include the analysis of: • An initial calibration blank • A continuing calibration blank • A preparation blank :\lqapreva\sect11.doc • MS/MSD/duplicate samples • An instrument check standard • A laboratory control sample • An interference check standard Each of these analyses is described below. Date: 12/22/95 Section 11 Revision 0.01 Page 14 of 42 Initial Calibration Blank Analysis -Following the analysis of the initial calibration verification standard (ICV) and prior to the analysis of samples, an initial calibration blank (ICB) is analyzed to demonstrate that the analytical system is free from contamination. This blank analysis must be free from all elements of interest at or above the reporting limits (CRDLs for analysis by CLP protocol), or the instrument must be recalibrated before sample analysis may begin. Continuing Calibration Blank Analysis -Following the analysis of each continuing calibration standard (CCV) in an analytical sequence, a continuing · calibration blank (CCB) is analyzed to demonstrate that the analytical system · is free from contamination throughout the course of that sequence. This blank analysis must be free from all elements of interest at or above the reporting limits (CRDLs for analysis by CLP protocol), or sample analysis must be discontinued and the previous 10 samples must be reanalyzed under a new calibration. Preparaljon Blank Analysis - A preparation blank, containing all of the reagents and volumes used in the processing of samples and carried through the complete preparation and analysis procedure, is analyzed at a minimum frequency of one per sample batch or sub-batch.· The preparation blank is analyzed to demonstrate that the sample preparation procedure is free from contamination. This blank must be free of all elements of interest at or above the reporting limits (CRDLs for analysis by CLP protocol), or the entire sub-batch of samples must be reprepared and reanalyzed. Laboratory Control Sample Analysis -Control samples may be obtained from commercial vendors or prepared from suitable reference materials, but must be prepared independently from the calibration standards. The LCS is prepared along with the samples for analysis and is used to verify that the analytical system is in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in method/instrument performance. Matrix Spike/Matrix Spike Duplicate or Sample Duplicate Analysis -For each analytical batch and sample matrix type, a matrix spike sample and either a matrix spike duplicate sample or duplicate matrix sample are analyzed at a minimum frequency of one set per batch. Matrix spike recoveries should fall within 75-125% (or within lab derived limits if applicable) (or within lab I I I I I I I u I I I I I I I I I I I u D I I I I I I I I I I I 11.6.4 :\lqapreva\sect11.doc Date: 12/22/95 Section 11 Hevision 0.01 Page 15 _cif 42 derived limits of applicable of the spike concentration for water and soil matrices. If the spike is not recovered within the specified limits, the data should be flagged as suspect due to sample matrix effects. Depending upon the project, provisions should be established to determine when the method of standard addition (MSA) should be employed to compensate for matrix effects. Interference Check Standard Analysis OCPS\ -The inte,ference check standard is analyzed at the beginning and end of the analytical sequence and at intervals during the sequence. This standard contains the analytes of interest at minimal concentrations and by known concentration of interfering elements. If results exceed ±20% of the expected value, the instrument must be recalibrated before sample analysis may proceed. Cyanide Analysis Inorganic cyanide is detenmined colorimetrically by method 9010A, method 9012 or the CLP Inorganic SOW. Quality control procedures for this analysis include the analysis of: • A preparation blank • A laboratory control sample • A matrix spike sample • A matrix spike duplicate or sample duplicate Each of these analyses is described below. Preparation Blank Analysis - A preparation blank, containing all of the reagents and volumes used in the processing of samples and carried through the complete preparation and analysis procedure, is analyzed at a minimum frequency of one per sample batch or sub-batch. The preparation blank is analyzed to demonstrate that the sample preparation procedure is free from contamination. This blank must be free of cyanide at or above the reporting limit (CRDL for analysis by CLP protocol), or the entire sub-batch of samples must be reprepared and reanalyzed. Laboratory Control Sample Analysis -Control samples may be obtained from commercial vendors or prepared from suitable reference materials, but must be prepared independently from the calibration standards. The LCS is prepared along with the samples for analysis and is used to verify that the sample preparation and analysis steps are in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in method/instrument performance. Matrix Spike/Matrix Spike Duplicate or Sample Duplicate Analysii, -For each analytical batch and sample matrix type, a matrix spike sample and either a matrix spike duplicate sample or duplicate matrix sarriple are analyzed at a 11.6.5 :\Jqapreva\sect11.doc Date: 12/22/95 Section 11 Revision 0.01 Page 16 of 42 minimum frequency of one set per batch. Matrix spike recoveries should fall within 75-125% (or within lab derived limits if applicable) (or within lab derived limits if applicable) of the spike concentration for water and soil matrices. If the spike is not recovered within the specified limits, the data should be flagged as suspect due to sample matrix effects. Anion Analysis Anions, including chloride, nitrite, nitrate, a-phosphate, bromide, and sulfate, may be analyzed by ion chromatography as described by Method 300.0. Quality control procedures for this method include the analysis of: • A preparation blank • A laboratory control sample • A matrix spike sample • A matrix spike duplicate or sample duplicate Each of these analyses is described below. Preparation Blank Analysis - A preparation blank, containing all of the reagents and volumes used in the processing of samples and carried through the complete preparation and analysis procedure, is analyzed at a minimum frequency of one per sample batch or sub-batch. The preparation blank is analyzed to demonstrate that the sample preparation procedure is free from contamination. This blank must be free of the anions being measured at or above the reporting limits, or the entire sub-batch of samples must be reprepared and reanalyzed. Laboratory Control Sample Analysis -Control samples may be obtained from a commercial vendor or prepared from suitable reference materials, but must be prepared independently from · the calibration standards. The LCS is prepared along with the samples for analysis and is used to verify that the sample preparation and analysis steps are in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in method/instrument performance. Matrix Spike/Matrix Spike Duplicate or Sample Duplicate Analysis -For each analytical batch and sample matrix type, a matrix spike sample and either a matrix spike duplicate sample or duplicate matrix sample are analyzed at a · minimum frequency of one set per batch. Matrix spike recoveries should fall within 75-125% (or within lab derived limits if applicable) (or within lab derived limits if applicable) of the spike concentration for water and soil matrices. If the spike is not recovered within the specified limits, the data should be flagged as suspect due to sample matrix effects. Duplicate analyses should agree within 20% RPO. I I I I I I I I I I I I I I I I I I a D D D I I I I I I I I I I I I I 11.6.6 11.6.7 :\lqapreva\sect11.doc Fluoride Analysis Date: 12/22/95 Section 11 Revision 0.01 Page 17 of 42 Fluoride is determined potentiometrically by method 340.2. Quality control procedures include the analysis of: • A preparation blank • A laboratory control sample • A matrix spike sample • A matrix spike duplicate or sample duplicate Each of these analyses is described below. Preparation Blank Analysis -A preparation blank, containing all of the reagents and volumes used in the processing of samples and carried through the complete preparation and analysis procedure, is analyzed al a minimum frequency of one per sample batch or sub-batch. The preparation blank is analyzed to demonstrate that the sample preparation procedure is free from contamination.· This blank must .be free of fluoride at or above the reporting · limit, or the entire sub-batch of samples must be reprepared and reanalyzed. Laboratory Control Sample Analysis -Control samples may be obtained from commercial vendors or prepared from suitable reference materials, but must be prepared independently from the calibration standards. The LCS is prepared along with the samples for analysis and is used to verify that the sample preparation and analysis steps are in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in method/instrument performance. Matrix Spike/Matrix Spike Duplicate or Sample Duplicate Analysis -For each analytical batch and sample matrix type, a matrix spike sample and either a matrix spike duplicate sample or duplicate matrix sample are analyzed at a minimum frequency of one set per batch. Matrix spike recoveries should fall within 75-125% (or within lab derived limits if applicable) of the spike concentration for water and soil matrices. If the spike is not recovered within the specified limits, the data are flagged as suspect due to sample matrix effects. Duplicate analyses should agree within 20% RPO. Tota\ Organic Carbon Analysis Combustion of organic carbon and detection of carbon spectrometry is performed by method 9060 or 415.1. procedures include the following analyses: • A preparation blank • A laboratory control sample • A matrix spike sample dioxide by IR Quality control • A matrix spike duplicate or sample duplicate Each of these analyses is described below. Date: 12/22/95 Section 11 Revision 0.01 Page 18 of 42 Preparation Blank Analysis - A preparation blank, containing all of the reagents and volumes used in the processing of samples and carried through the complete preparation and analysis procedure, is analyzed at a minimum frequency of one per sample batch or sub-batch. The preparation blank is analyzed to demonstrate that the sample preparation procedure is free from contamination. This blank must be free of organic carbon at or above the reporting limit, or the entire sub-batch of samples must be reprepared and reanalyzed. Laboratory Control Sample Analysis -Control samples may be obtained from a commercial vendor or prepared from suitable reference materials, but must be prepared independently from the calibration standards. The LCS is prepared along with the samples for analysis and is used to verify that the sample preparation and analysis steps are in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in method/instrument performance. Matrix Spike/Matrix Spike Duplicate or Sample Duplicate Analysis -For each analytical batch and sample matrix type, a matrix spike sample and either a matrix spike duplicate sample or duplicate matrix sample are analyzed at a minimum frequency of one set per batch. Matrix spike recoveries should fall within 75-125% (or within lab derived limits if applicable) of the spike concentration for water and soil matrices. If the spike is not recovered within the specified limits, the data are flagged as suspect due to sample matrix effects. Duplicate analyses should agree within 20% RPO. · 11.6.8 OH and Grease Analysis :\lqapreva\sect11.doc Total oil and grease is determined gravimetrically by methods 9070/9071A and 413.1 and spectrophotometrically (IR) by method 413.2. Quality control procedures include the following analyses: • A preparation blank • A laboratory control sample • A matrix spike sample • A matrix spike duplicate sample Each of these analyses is described below. Preparation Blank Analysis - A preparation blank, containing all of the reagents and volumes used in the processing of samples and carried through the complete preparation and analysis procedure, is analyzed at a minimum I I I I I I I I I I I I I I I I I I I g n R D m I I I I I I I I I I I I 11.6.9 :\lqapreva\sect1, .doc Date: 12/22/95 Section 11 REwision 0.01 Page 19 of42 frequency of one per sample batch or sub-batch. The preparation blank is analyzed to demonstrate that the sample preparation procedure is free from contamination. This blank must be free of oil and grease at or above the reporting limit, or the entire sub-batch of samples must be r,3prepared and reanalyzed. Laboratory Control Sample Analysis -Control samples may be obtained from a commercial vendor or prepared from suitable reference materials, but must be prepared independently from the calibration standards. The LCS is prepared along with the samples for analysis and is used to verify that the sample preparation and analysis steps are in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in method/instrument performance. Matrix Spike/Matrix Spike Duplicate Analysis -For each analytical batch and sample matrix type, a matrix spike sample and a matrix spike duplicate sample are analyzed at a minimum frequency of one set per batch. Matrix spike recoveries should fall within 75-125% (or within lab derived limits if· applicable) of the spike concentration for water and soil matrices. If the spike is not recovered within the specified limits, the data are flagged as suspect due to sample matrix effects. Duplicate analyses should agree within 20% RPO. Total Recoverable Petroleum Hydrocarbons (TRPHl Analysis TRPH is determined spectrophotometrically (JR) by method 418.1. Quality control procedures include the following analyses: • A preparation blank • A laboratory control sample • A matrix spike sample • A matrix spike duplicate sample Each of these analyses is described below. Preparation Blank Analysis - A preparation blank, containing all of the reagents and volumes used in the processing of samples and carried through the complete preparation and analysis procedure, is analyzed at a minimum frequency of one per sample batch or sub-batch. The preparation blank is analyzed to demonstrate that the sample preparation procedure is free from contamination. This blank must be free of TRPH at or above the reporting limit, or the entire sub-batch of samples must be reprepared and reanalyzed. Laboratory Control Sample Analysis -Control samples may be obtained from·. commercial vendors or prepared from suitable reference materials, but must be prepared independently from the calibration standards. The LCS is Date: 12/22/95 Section 11 Revision 0.01 Page 20 of 42 prepared along with the samples for analysis and is used to verify that the sample preparation and analysis steps are in control. LCS recovery data are plotted on blank spike control charts to monitor the analytical system for trends or events that indicate a change in method/instrument performance. Matrix Spike/Matrix Spike Duplicate Analysis -For each analytical batch and sample matrix type, a matrix spike sample and a matrix spike duplicate sample are analyzed at a minimum frequency of one set per batch. Matrix spike recoveries should fall within 75-125% (or within lab derived limits if applicable) of the spike concentration for water and soil matrices. If the spike is not recovered within the specified limits, the data are flagged as suspect due to sample matrix effects. Duplicate analyses should agree within 20% RPO. 11.6.1 o California Assessment Manual Waste Extraction Test /CAM WED/Extraction Procedure Toxicity Test Method /EP-Tox}/Toxicity Characteristic Leaching Procedure <ICLPl Waste extraction is performed according to the procedure described in the appropriate waste extraction regulation. Quality control procedures include the following analyses: • A preparation blank • A duplicate sample Each of these analyses is described below. Preparation Blank Analysis -A minimum of one method blank per sample batch of up to 20 samples is analyzed to demonstrate the absence of contamination above reporting limits. Duplicate Sample Extraction -A duplicate sample extraction is performed with each batch of up to 20 samples. Results of analyses of the duplicate extracts are used to estimate overall measurement variability and generally should agree within 20% RPD. 11.7 STANDARDS The term standard shall apply to any analyte solution of known concentration which is traceable to a certified reference material. This includes calibration standards, spiking solutions and laboratory control samples. Upon receipt, all purchased standard reference materials (neat and stock solutions) are recorded into section-specific standards logbooks. Standard logbook entries include Pace unique ID, name of the neat compound or solution, manufacturer, manufacturer's lot number, certified purity, and expiration date. Subsequent :\Jqapreva\sect11.doc I I I I I I I D I I I I I I I I I I I I n D 0 I I I I I I I I I I I I I I Date: 12/22/95 Section 11 Revision 0.01 Page 21 of 42 preparations of stock, intermediate, and working solutions are also documented in the standards logbooks. These entries must include all discrete measurements· made during preparation, sources of materials, solvent(s} and a Pace ID numb,~r. The standard vial should have a reference label affixed containing the following information (if sPace permits): Standard ID number Name of standard Preparation date Preparer's initials Solvent Preservation, if applicable Expiration date The Standard Operating Procedure (#MN-P-004-B) "Standards Traceability in Laboratory and Field" contains further instructions for assigning unique ID numbers, shelf life of standards, and good laboratory practices. All primary reference standard and standard solutions are purchased from reliable commercial sources. Standards traceable to NIST are preferred; however, ASTM or equivalent specifications are acceptable. Certification records of all standards received are retained. Second source reference standards and standard solutions are purchased from a different supplier than ·the primary standard or from a different manufactured lot from the same vendor. If a second supplier is not available, the second source standard can be prepared from a different lot number of the same composition from the same supplier. Newly prepared standard solutions (surrogate, internal, calibration, spiking) are verified against another known standard prepared from another source prior to utilization. The verification data is maintained on file in the respective area. In place of performing in-house standard verification, laboratories can purchase verified second source standards from a vendor which supplies a data package demonstrating verification. 11.8 SOLVENT LOT/ACID LOT VERIFICATION All laboratory extraction solvents utilized are at least Pesticide Grade or better. Prior to accepting a solvent lot from the supplier, a quantity of the solvent is analyzed either by Pace or by the primary vendor under a National Qualified Materials (NQM) program to access the purity. The NQM verification program is administered by the Pace Corporate office for use by all laboratory locations. If the lot is determined to meet purity standards/requirements, the lot is sequestered for Pace laboratories. The quality of organic extraction solvents is constantly monitored through the .analysis of method blanks. :\lqapreva\sect11.doc Date: 12/22/95 Section 11 Revision 0.01 Page 22 of 42 Acid lots are verified for purity prior to utilization for digestion of samples. If the acid lot is determined to meet purity criteria, the acid may be used for sample preparation digestion. :\lqapreva\sect11.doc I I I I I I B I I I I I I I I I I I I . - - - - - - - - -1111111 _ . - - - -l!!!!!!!I l!!!!!!I l!!!!!!I l!!!!!!I l!!!!I ------------------------------------· >- (5 > 0 u ltd sw·-. 8021 · Bl~N7.ENJ~ G /0:1. /95 SOJI, 130.---------------------~ 120 ··-·····---------------------a-m-lll'a-m-m-a-•-•-a-•-•-•-•-•---~-----~ 110 ··--·-·------------------- 100 ··-·-··------------------- 90 ·--·*··•.2K. ... 2K-:-·* .. •-~~ -*---*:::7.~.-)K __ 2~ .. ..2K._~D.~ ... 2K . ..i.K ... .ZK __ lK ... 2f~ .. .2t{ __ )K .. ~ ... >< >< ><➔<->«-><-><➔< >< >< ><-><·➔<->< >< )( >< >< ><->< so1--------------~-----1 70 -······--------------------1 60 1--------- 50 ·----··-------------- ~o ··-·--·-------------------'---1 30 ············------------------- 20 ·················----- 10 -l~~i~l~~~I---,--1·1~ ~I -.-1 -,~,,-.1~1~1-.I I I I 1 3 5 7 9 11 13 15 17 19 · 2 4 6 8 1 0 1 2 1 ~ 16 18 20 DATA POINTS . --- UCL 11,J. --1-- UWL AVG 92 -El- LWL )( 112.166 107.005 oa.636 66,271 61.0BB - - - 03/22 WATER SPll<E LIMITS-ACCURACY BENZENE Olfl◄ 03/17 O]fJO 0-1,01 O•UO) 0◄/05 na 50 McHUl"" 06.636 so .. 6.183 cv~ 6.36% Min-65.100 Maxa 101.000 0-1/07 - - - - -----liilllll - ,3 osn ,2.osn -2.0SO ·3.0S0 11111111 llllll 1111!1 l!!!l!I - - - - - --- l!!!!!!!!!I l!!!!!!!!!I I!!!!! !!!!!I !!Im == == llliil liilEI liiillii liiiia. SW-0021 J3l~NZENE G/01/95 SOIT_J so~----------------~ '10 ···········--------------- 20 -----·---·li11-liil-611-M-liill-■-lll-lla-l!iil-llil ......... . 10 0-1 I I I I I I I I I ·1 3 5 7 9 2 ,,. 6 8 10 DAT/\ POINTS -1111- UCI_ 20 -,-- UWL 15 /\VG 1'I "Tl m· C ., (I) ...... ...... i-, "1J :u CJ) 0 i.,) ro n, Cl> ~ s. n ro ILi ¥?. 0 . , , l..l n:::, _. ln :J _., ~ Q, p -~ i.o r-> _ _. lD ; u, 11.◄31 8.047 3,970 -0.088 -3.◄72 03/17 -- OJ/21 03/22 WATER SPll<E LIMITS-PRECISION BENZENE - OJ/25 03/27 01/30 0◄/01 0-UOJ 0,&/05 0◄/07 nlil 25 Muan• 3.078 sn .. ~.◄8-t CV• 62.◄2'!. Min,. 0.000 Max-" 13.72~ ---liillliil liiilil -Ill.Iii 1\81 13.05() •2 OSCl "Tl in' C .. CD ~ Mdan ~ i-., " a :, r+ - •2.0Sfl ·3 OS() 0.,,01 0.,,01 ---- - -- ANALYTICAL METHOD 8010B 8020A :\lqapreva\sectl 11 - - APPLICABLE PARAMETER Purgeable Halocarbons Purgeable Aromatics - --!!!!!!!I !!!!!!I I!!!!! I!!!!! 11::1::11 TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Initial 5-point calibration Continuing calibration Method blank Surrogate spikes Matrix spike & Matrix spike duplicate Laboratory control sample (LCS) Second column confirmation Initial 5-point calibration Contim.iing calibration Method blank FREQUENCY As needed - Refer to method Daily & every 10 samples 1 per batch & sub-batch Every analysis 1 set per batch 1 per batch & sub-batch 100% for positive results :::. reporting limit As needed Refer to method Daily & every 1 0 samples 1 per batch & sub-batch ACCEPTANCE CRITERIA r :::.0.995 %D .:,15% (except gases and 2-CEVE) All analytes < reporting limit Bromochloromethane (See Table 5.1) Method 8010B limits (See Table 5.1) Statistical limits (See Table 5.1) Qualitative confirmation r :::_0.995 %D ~15% All analytes < reporting limit lilillil liiliiil Date: 12/22/95 Section 11 Revision 0.01 Page 27 of 42 liiiliil CORRECTIVE ACTION Repeat calibration 1. Repeat test 2. Recalibrate iliil Clean system & reanalyze sub- batch Reanalyze sample Narrate in report Correct problem & reanalyze sub-batch N/A Repeal calibration 1. Repeat test 2. Recalibrate Clean system & reanalyze sub-batch - ANALYTICAL METHOD 8080A CLP SOW :\lqaprevaUectl 11 -- APPLICABLE PARAMETER Organochlorine Pesticides and PCBs -11111 TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Surrogate spikes Matrix spike & Matrix spike duplicate Laboratory control sample (LCS) Second column Initial calibration: 3-point (CLP) 5-point (8080A) Continuing calibration Breakdown check (Endrin and DDT) Method blank Surrogate spikes FREQUENCY Every analysis 1 set per batch 1 per batch and 100% for positive results confirmation As needed Refer to method Every 12 hours (CLP) Daily & every 10 samples (8080A) Every 12 hours (CLP) Daily (8080A) 1 per batch & sub-batch All analyses 11111111 r:::.lm ACCEPTANCE CRITERIA 1,4-Bromofluorobenzene (See Table 5.1) Method 8020A limits (See Table 5.1) Statistical limits (See Table 5.1) Qualitative confirmation :!. reporting limit r ?.0.995 (See method for CLP) ¾D .:,25% (CLP) ¾D .:,15% (8080) :::::20% for each compound (& .:,30% total for CLP) All analytes < CRQL (CLP) < Reporting limits (8080A) Tetrachloro-m-xylene Decachlorobiphenyl (See Table 5.4) - Date: 12/22/95 Section 11 Revision 0.01 Page 28 of 42 CORRECTIVE ACTION Reanalyze sample Narrate in report Correct problem and renanalyze sub-batch NIA Repeat calibration 1. Repeat test 2. Recalibrate 1. Correct problem 2. Recalibrate Re-extract & reanalyze sub-batch Flag data (CLP) Reanalyze sample (8080A) ----- - - ANALYTICAL METHOD California LUFT Manual :\lqapreva\sea 111 -- APPLICABLE PARAMETER. Purgeable Petroleum Hydrocarbons - - l!!!!!I l!!!!l!!!!I l!!!!!I l!!!!!!I == TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL ACCEPTANCE CHECK FREQUENCY CRITERIA Matrix spike & 1 set per batch Method limits Matrix spike duplicate (See Table 5.4) Laboratory control 1 per batch & sub-batch Statistical limits sample (LCS) (See Table 5.4) Second column 100% for positive results qualitative confirmation confirmation ~ reporting limit Initial 5-point As needed r :::_0.995 calibration Refer to method Continuing calibration Daily & every 10 samples %05:15% Method blank 1 per batch & sub-batch All analytes < reporting limit Surrogate spikes Every analysis 1,4-Bromofluorobenzene (See Table 5.2) Matrix spike & 1 set per batch Method 8020A limits Matrix spike duplicate (See Table 5.2) Laboratory control 1 per batch & sub-batch Statistical limits sample (LCS) (See Table 5.2) Second column 100% for BTEX results Qualitative confirmation confirmation :::. reporting limit if no gasoline present Date: 12/22/95 Section 11 Revision 0.01 Page 29 of 42 ail CORRECTIVE ACTION Flag data (CLP) Narrate in report (8080A) Re-extract & reanalyze sub-batch NIA Repeat calibration 1. Repeat test 2. Recalibrate Clean system and reanalyze sub-batch Reanalyze sample Narrate in report Correct problem and reanalyze sub-batch N/A liillii . ANALYTICAL METHOD California LUFT Manual 82408 CLP SOW .\lqap,eva\seci 111 - APPLICABLE PARAMETER Extractable Petroleum Hydrocarbons Volalile Organics byGC/MS liiiililil liiBlil TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Initial 5-point calibration Continuing calibration Method blank Matrix spike & Matrix spike duplicate Laboratory conlrol sample (LCS) Mass scale calibration using PFTBA Mass spectral ion intensity check using BFB Initial 5-point calibration -ail FREQUENCY As needed Refer to melhod Daily & every 1 0 samples 1 per batch & sub-batch 1 set per batch 1 per batch & sub-batch Daily Every 12 hours As needed Refer to method 1111111 ACCEPTANCE CRITERIA r "'.0.995 %0 ~15% All analy1es < reporting limil Statistical limits (See Table 5.3) Statistical limits (See Table 5.3) N/A Refer to method RSD :,20.5% & min. RRF for Table 2 compounds (CLP), RSD :,30% for CCC & min. RRF for SPCC compounds (8240B) - Date: 12/22/95 Section 11 Revision 0.01 Page 30 of 42 CORRECTIVE ACTION Repeal calibration 1. Repeat test 2. Recalibrate Re-extract & reanalyze sub-batch Narrate in report Re-extract & reanalyze sub-batch NIA 1. Retune instrument 2. Repeal BFB analysis Repeat calibration ----- - - - - - - 11!!1 l!!!!!!!I! !!!!!!I l!!!!!I 1!!!!11 TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES ANALYTICAL APPLICABLE QUALITY CONTROL ACCEPTANCE METHOD PARAMETER CHECK FREQUENCY CRITERIA Continuing calibration Every 12 hours %0 ~25% & min. RRF for Table 2 compounds (CLP), %0 ~20% for CCC & min. RRF for SPCC compounds (82408) Method blank Every 12 hours All analytes < CRQL, < 5X CRQL for common solvents (CLP); < Reporting limit, < 5X reporting limit for common solvents (82408) Surrogate spikes Every analysis Method limits (See Table 5.5) Internal standard Every analysis -50% to +100% area count of continuing calibration RT shift <30 sec. for ISs in daily std. Matrix spike & 1 set per batch Method limils Matrix spike duplicate {See Table 5.5) Laboratory control 1 per batch & sub-batch Statistical limits sample (LCS) (See Table 5.5) 82708 Semivolatile Mass scale calibration Daily N/A CLP SOW Organics using PFT8A by GC/MS :\lqapre~a\secl 111 a.a liiiiiii Date: 12/22/95 Section 11 Revision 0.01 Page 31 of 42 liilil CORRECTIVE ACTION 1. Repeat test 2. Recalibrate Clean system & reanalyze sub-batch Reanalyze sample Reanalyze sample Narrate in report liill Correct problem and reanalyze sub-batch N/A ANALYTICAL METHOD :\lqapreva\sedl 11 - APPLICABLE PARAMETER -liilii TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Mass spectral ion inlensity check using OFTPP Initial 5-point calibration Continuing calibration Method blank Surrogate spikes Internal standard llllill FREQUENCY Every 12 hours As needed Refer to method Every 12 hours 1 per batch & sub-batch Every analysis Every analysis ACCEPTANCE CRITERIA Refer to method RSO :,20.5% & min. RRF for Table 5 compounds (CLP). RSO :,30% for CCC & min. RRF for SPCC compounds (82708) %0 :,25% & min. RRF for Table 5 compounds (CLP), %0 :,20% for CCC & min. RRF for SPCC compounds (82708) All analytes < CRQL, < 5X CRQL for phthalates (CLP); < Reporting limit, < 5X reporting limit for phthalates (82708) Method limits (See Table 5.6) •50% to +100% area count of continuing calibration RT shift <30 sec. for ISs in daily std. - Date: 12/22/95 Section 11 Revision 0.01 Page 32 of 42 CORRECTIVE ACTION 1. Retune instrument 2. Repeat OFTPP analysis Repeat calibration 1. Repeal test 2. Recalibrate Re-extract & reanalyze sub-batch Re-extract & reanalyze sample Reanalyze sample ----- -- ANALYTICAL METHOD 6010A CLP SOW :\lqapreva\seci 111 -- APPLICABLE PARAMETER Trace Metals By ICPS - - I!!!!! l!!m == TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL ACCEPTANCE CHECK FREQUENCY CRITERIA Matrix spike & 1 set per batch Method limils Matrix spike duplicate (See Table 5.6) Laboratory control 1 per batch & sub-batch Statistical limits sample (LCS) (See Table 5.6) Initial calibration Daily Refer to method Initial calibration Daily 90-110% of true value verification (ICV) Initial calibration Daily All elements_<; CRDL (CLP) blank (ICB) < Reporting limit (601 0A) Interference check Beginning & end of 80-120% of true value (A, AB) run sequence CRDL check (CLP) Beginning & end of NIA run sequence Continuing calibration Every 10 samples 90-110% of true value verification (CCV) Continuing calibration Every 10 samples All elements _<; CRDL (CLP) blank (CCB) < Reporting limit (601 0A) Method blank 1 per batch & sub-batch All elements .'.: CRDL (CLP) < Reporting limit (601 0A) liiliiiil Eliiiiil Date: 12/22/95 Section 11 Revision 0.01 Page 33 of 42 CORRECTIVE ACTION Narrale in report Re-extract & reanalyze sub-batch Refer to method Recalibrate Recalibrate Recalibrate (Initial) Reanalyze samples (Final) NIA 1. Recalibrate 2. Reanalyze samples 1. Recalibrate 2. Reanalyze samples Redigest & reanalyze sub-batch iai TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES Date: 12122195 Section 11 Revision 0.01 Page 34 of 42 ANALYTICAL METHOD 7041 (Sb) 7060A (As) 7740 (Se) 7421 (Pb) 7841 (Tl) 7470A(Hg) 7471A (Hg) CLP SOW :\lqapreva\sect11t APPLICABLE PARAMETER Trace metals byGFAA and CVAA QUALITY CONTROL CHECK Laboratory control sample (LCS) Replicate exposures Matrix spike Matrix spike duplicate (6010A) Duplicate analysis (CLP) Serial dilution Initial multipoint calibration Initial calibration verification (ICV) Initial calibration blank (ICB) CRA standard (CLP) Continuing calibration verification (CCV) FREQUENCY 1 per batch & sub-batch Every analysis 1 per batch 1 per batch 1 per batch 1 per batch Daily Daily Daily Daily Every 10 samples --liiiilil li!iil -.. .. -== ACCEPTANCE CORRECTIVE CRITERIA ACTION 80.120% recovery (CLP) Redigest & reanalyze Statistical limits sub-batch (See Table 5. 7, 601 0A) RSD:5.20% Reanalyze sample 75-125% recovery Narrate in report 75-125% recovery Narrate in report RPD :5.10% Flag data 90-110% of undiluted value Flag data (CLP) r ;!0.995 Repeat calibration 90.110% (GFAA) 80.120% (CVAA) Recalibrate 5. CRDL (CLP) Recalibrate < Reporting limit (7000) NIA NIA 90.110% (GFAA) 1. Recalibrate 80-120% (CV AA) 2. Reanalyze samples ------- - - -- ANALYTICAL METHOD 7196A :\lqapceva\s&cl I It -- APPLICABLE PARAMETER Hexavalent Chromium - - l!!l!!I l!!!!!!!!I !!!!!I l!!!l!I I!!!!!!! TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL ACCEPTANCE CHECK FREQUENCY CRITERIA Continuing calibration Every 10 samples :;.CRDL (CLP) blank (CCB) < Reporting limit (7000) Method blank 1 per batch & sub-batch :;_ CRDL (CLP) < Reporting limit (7000) Laboratory control 1 per batch & sub-batch 80-120% recovery (CLP) sample (LCS) Statistical limits . (See Table 5.7, 7000) Matrix spike 1 per batch 75-125% recovery Matrix spike duplicate 1 per batch 75-125% recovery (7000) Duplicate analysis (CLP) 1 per batch RPD:;.20% Analytical spike (GFAA) Every sample Refer lo method (CLP) Initial multipoint Daily r ::_0.995 calibration Initial calibration Daily 90-110% of true value verification (ICV) Initial calibration Daily < Reporting limit blank (ICB) liiiiiil Date: 12/22195 Section 11 Revision 0.01 Page 35 of 42 -- CORRECTIVE ACTION 1. Recalibrate 2. Reanalyze samples Redigest & reanalyze sub-batch Redigesl & reanalyze sub-batch Flag data (CLP) Narrate in report (7000) Narrate in report Flag data Refer lo method Repeal calibration Recalibrate Recalibrate - ANALYTICAL METHOD 9010N9012 CLP SOW :\lqapreva\sea11 I - APPLICABLE PARAMETER Cyanide illill TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Continuing calibration verification (CCV) Continuing calibration blank (CCB) Method blank Malrix spike & Matrix spike duplicate Laboratory control sample (LCS) lnilial multipoint calibration Initial calibration verification (ICV) Initial calibration blank (ICB) Continuing calibration verification (CCV) Method blank 1111111 FREQUENCY Every 10 samples Every 10 samples 1 per batch & sub-batch 1 set per batch 1 per batch & sub-batch Daily Daily Daily Every 10 samples 1 per batch & sub-batch mm ACCEPTANCE CRITERIA 90-110% of true value < Reporting limit < Reporting limit 75,.125% recovery Statistical limits (See Table 5.8) r ,:0.995 85,.115% of true value :; CRDL (CLP) < Reporting limit (9010N9012) 85,.115% of true value :; CRDL (CLP) < Reporting limit (9010N9012) --- Date: 12/22/95 Section 11 Revision 0.01 Page 36 of 42 CORRECTIVE ACTION 1. Recalibrate 2. Reanalyze samples 1. Recalibrate 2. Reanalyze samples Reprep & reanalyze sub-batch Narrate in report Reanalyze sub-batch Repeat calibration Recalibrate Recalibrate 1. Recalibrate 2. Reanalyze samples Reprep & reanalyze sub-batch ----- - - ANALYTICAL METHOD 300.0 :\lqaprevalsect 111 -- APPLICABLE PARAMETER Anions by Ion Chromatography --l!!!!!!!I I!!!!! I!!!! . !!!!I . -i==s == TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Laboratory control sample (LCS) Matrix spike Matrix spike duplicate (9010A/9012) Duplicate analysis (CLP) Initial multipoint calibration Initial calibration verification (ICV) Initial calibration blank (ICB) Continuing calibration verification (CCV) Continuing calibration biank (CCB) Method blank FREQUENCY 1 per batch & sub-batch 1 per batch 1 per batch 1 per batch Daily Daily Daily Every 10 samples Every 10 samples 1 per batch & sub-batch ACCEPTANCE CRITERIA 80-120% recovery (CLP) Statistical limits (See Table 5.8) 75-125% recovery 75-125% recovery RPD.'.,20% r ,:0.995 90-110% of true value All analytes < reporting limit 85-115% of true value All analytes < reporting limit All analytes < reporting limit aiiil Date: 12/22/95 Section 11 Revision 0.01 Page 37 of 42 lililil CORRECTIVE ACTION Reprep & reanalyze sub-batch Flag data (CLP) Narrate in report (9012) Narrate in report Flag data Repeat calibration Recalibrate Recalibrate 1. Recalibrate 2. Reanalyze samples 1. Recalibrate 2. Reanalyze samples Reanalyze sub-batch iiil _ lliiii_ . ANALYTICAL METHOD 120.1 130.2 150.1 :\lqapreva\sect11t - APPLICABLE PARAMETER Conductance Hardness pH liiilliill lillil TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL ACCEPTANCE CHECK FREQUENCY CRITERIA Matrix spike & 1 sel per batch 75-125% recovery Matrix spike duplicate s20%RPD Laboratory control 1 per batch & sub-batch Statistical limits sample (LCS) (See Table 5.8) Multipoint calibration Daily 98-102% of true value for each ·standard Calibration verification Daily 90-110% recovery Duplicate analysis 1 per batch RPD~20% Laboratory control 1 per batch & sub-batch Statistical limits sample (LCS) (See Table 5.8) Method blank 1 per batch & sub-batch < Reporting limit Duplicate analysis 1 per batch RPD~20% 2 to 3-point calibration Daily ±0.05 to> :t0.1 pH units Calibration verification Daily ±0.05 to> :t0.1 pH units Duplicate analysis 1 per batch ±0.05 to> :t0.1 pH units iiiliil -Elllllil ma -- Date: 12/22/95 Section 11 Revision 0.01 Page 38 of 42 CORRECTIVE ACTION Narrate in report Reanalyze sub-batch Repeat calibration 1 . Re peat check 2. Recalibrate 3. Reanalyze batch Narrate in report 1. Restandardize titrant 2. Reanalyze sub-batch Reanalyze batch Narrate in report Repeat calibration 1. Recalibrate 2. Reanalyze batch Narrate in report ----- - - ANALYTICAL METHOD 160.1 160.2 310.1 340.2 :\Jqapreva\sect11l -- APPLICABLE PARAMETER Total Dissolved Solids Total Suspended Solids Carbonate/ Bicarbonate Fluoride - --l!l!!!!I ~ !!!!!! !!!!!I _ ll!!S == TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Method blank Duplicate analysis Method blank Duplicate analysis 3-point calibration Calibration verification Method blank Duplicate analysis Matrix spike & Matrix spike duplicate Multipoint calibration Calibration verification Method blank FREQUENCY 1 per batch & sub-batch 1 per batch 1 per batch & sub-batch 1 per batch Daily Daily 1 per batch & sub-batch 1 per batch 1 set per batch Daily Daily 1 per batch & sub-batch ACCEPTANCE CRITERIA ±0.5 mg RPD:::20% :t0.5 mg RPO :::20% ±0.05 to ±0.1 pH units 85-115% recovery < Reporting limit RPO :::20% 75-125% recovery r ~0.995 90-110% recovery < Reporting limit l:iiiiiil liiiiiiiii Date: 12/22/95 Section 11 Revision 0.01 Page 39 of 42 liiliii CORRECTIVE ACTION Reprep and reanalyze batch Narrate in report Reprep and reanalyze batch Narrate in report Repeat calibration 1. Recalibrate 2. Reanalyze batch Reanalyze batch Narrate in report Narrate in report Repeat calibration Recalibrate Reanalyze batch ANALYTICAL METHOD 350.3 :\lqapreva\sect 111 -- APPLICABLE PARAMETER Ammonia lilllil TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Matrix spike & Matrix spike duplicate or Duplicate Laboratory control · sample (LCS) Multipoint calibration Calibration verification Method blank Matrix spike & Matrix spike duplicate or Duplicate Laboratory control sample (LCS) FREQUENCY 1 set per batch 1 per batch & sub-batch Daily 1 per batch 1 per batch & sub-batch 1 set per batch 1 per batch & sub-batch ACCEPTANCE CRITERIA 75-125% recovery RPO $20% Statistical limils (See Table 5.8) r::.0.995 85-115% recovery < Reporting limit 75-125% recovery RPO $20% Statistical limits (See Table 5.8) -- Date: 12/22/95 Section 11 Revision 0.01 Page 40 of 42 CORRECTIVE ACTION Narrate in report Reanalyze batch Repeat calibration 1. Recalibrate 2. Reanalyze batch Reanalyze batch Narrate in report Reanalyze batch ----- -- ANALYTICAL METHOD 9060 415.1 9070/9071 413.1 413.2 .\lqapreva\sed111 - - APPLICABLE PARAMETER Total Organic Carbon Oil & Grease Gravimetric Oil & Grease by IR - - I!!!!!! !!!!! !!!!!!I == == TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Initial calibration (4 replicates of 1 point) Calibration verification Method Blank Duplicate analysis Matrix spike Laboratory control sample (LCS) Balance calibralion check at 1 g & 100 g Method blank Matrix spike & Matrix spike duplicate or Duplicate Laboratory control sample (LCS) Initial 5-point calibration FREQUENCY Daily Daily 1 per batch & sub-batch 1 per batch 1 set per balch 1 per batch & sub-balch Beginning & end of analytical sequence 1 per batch & sub-batch 1 set per batch 1 per batch & sub-batch Daily ACCEPTANCE CRITERIA 1125-1875 average raw instrument reading 85-115% of true value < Reporting limit RPO :,50% 75-125% recovery Statistical limits (See Table 5.8) ±0.001 g of true weight < Reporting limil 75-125% recovery RPO "20% Statistical limits (See Table 5.8) r ,:0.995 lliililil liiliii lliiil Date: 12/22/95 Section 11 Revision 0.01 Page 41 of42 CORRECTIVE ACTION Reprep standard & recalibrate Recalibrate Reanalyze batch Narrate in report Narrate in report Reanalyze balch Recalibrate balance 1. Redesiccate & reweigh samples 2. Re-extract sub-batch Narrate in report Re.extract & reanalyze sub-batch Repeat calibration ANALYTICAL METHOD 418.1 :\lqapreva\sect 111 -- APPLICABLE PARAMETER Total Petroleum Hydrocarbons bylR liiiilii TABLE 11.1 SUMMARY OF CALIBRATION AND QUALITY CONTROL PROCEDURES QUALITY CONTROL CHECK Continuing calibration Method blank Matrix spike & Matrix spike duplicate or Duplicate laboratory control sample (lCS) Initial 5-point calibration Continuing calibration Method blank Matrix spike & Matrix spike duplicate or Duplicate laboratory control sample (lCS) lili'II - FREQUENCY Every 10 samples 1 per batch & sub-batch 1 set per batch 1 per batch & sub-batch Daily Every 1 D samples 1 per batch & sub-batch 1 set per batch 1 per batch & sub-batch ACCEPTANCE CRITERIA 80-120% of true value < Reporting limit 75-125% recovery RPO !>20% Statistical limits (See Table 5.8) r ~0.995 80-120% of true value < Reporting limit 75-125% recovery RPO !>20% Statistical limits (See Table 5.8) -- Date: 12/22/95 Section 11 Revision 0.01 Page 42 of 42 CORRECTIVE ACTION 1. Repeat test 2. Recalibrate 1. Clean system & recheck 2. Re-extract & reanalyze sub-batch Narrate in report Re-extract & reanalyze sub-batch Repeat calibration 1. Repeat test 2. Recalibrate 1. Clean system & recheck 2. Re-extract & reanalyze Narrate in report Re-extract & reanalyze sub-batch ----- I I m I I D D D I I I I I I I I 12.0 QUALITY ASSURANCE AUDITS AND PERFORMANCE EVALUATIONS Date: 12/22/95 Section 12 Revision 0.01 Page 1 of 7 Integral to Pace's quality assurance program is a program of internal audits designed to provide feedback about the effectiveness and completeness of the various quality control and quality assurance systems in the laboratory. This section describes the types of audits conducted at Pace and discusses the roles and_responsibilities of Pace personnel related to these audits. 12.1 INTERNAL AUDITS 12.1.1 12.1.2 :\lqapreva\sect 12.doc Quality Assurance Auditor The QA auditor is responsible for designing and/or perfonming QA performance and systems audits. Since QA audits represent an independent assessment of laboratory functions, the auditor must be functionally independent from laboratory operations to ensure objectivity. However, the auditor must be familiar enough with the objectives, principles, and procedures of laboratory operations to be able to perform a thorough and effective evaluation. The auditor's ability to identify components of systems that are critical to overall data quality is especially important. Internal audits are typically conducted by the laboratory QA Officer, who may be assisted by other laboratory personnel. The QA Officer reports directly to the Laboratory General Manager, and therefore is independent of laboratory operations. The QA Officer evaluates audit observations and verifies the completion of corrective actions. Scope and Frequency of Internal Audits Internal systems audits are conducted at a minimum frequency of one per quarter. The scope of these audits may include the examination of the operations of a specific analy1ical department or may focus on the evaluation of a specific quality-related system as applied throughout the laboratory. · Examples of system-wide elements which can be audited include: • Standard operating procedures, including system of review, issue, filing, maintenance, training, understanding, documentation of deviations and implementation of SOPs. • Adherence to standard operating procedures, the QAP and regulations. Personnel and training files, including job descriptions, resumes, documented training and training file maintenance. General laboratory safety, including appropriate clothing, waste disposal, health and safety plan review, obvious safety concerns. Labeling of reagents, solutions, standards, and associated documentation. Equipment and instrumentation documentation, calibration/ maintenance records, operating manuals. • • Date: 12/22/95 Section 12 Revision 0.01 Page 2 of 7 Sample handling, storage and disposal including storage locations, security, tracking/chain-of-custody, disposal practices and records, labeling and retention. Documentation of sample analysis, methodologies, quality control requirements. Documentation of discrepancy reports and corrective action . General procedures for data security, review, documentation, reporting and archiving. When the operations of a specific department are evaluated, a number of functions are reviewed, such as: Documentation of technical training and analyst proficiency • Method detection limit studies Internal chain-of-custody documentation Nonconformance documentation Documentation of standard preparations Instrument maintenance documentation • Standard operating procedures Control charts Documentation of sample preparation and analysis Documentation of data review As required on specific projects, internal audits are performed to ensure laboratory conformance to site workp/ans, sampling and analysis plans, QAPP, etc. Project audits can include review of the following items of concern: Sample log-in and chain-of-custody records Sample storage procedures and records Sample preparation and analysis procedures • Method validation (where applicable) Conformance to QAPP Control charts (if applicable) Precision and accuracy assessment Method blanks, reagent blanks, duplicates, check samples, fortifications, surrogates, etc. Calibration • Data packages • Analyst qualifications Data validation and reporting I I I I I I I I I I I I I I I I 12.1.3 Internal Audit Reports and Corrective Action Plans I A full description of the audit, including the identification of the department or operation audited, the date(s) on which the audit was conducted, the specific systems examined, and the observations made during the course of the audit, are I summarized in an internal audit report. Although other personnel may assist with :\lqapreva\sect12.doc I I I I n n u H D E I I I I I I I Date: 12/22/95 Section 12 Revision 0.01 Page 3 of 7 the performance of the audit, the QA Officer writes and issues the internal audit report. The QA Officer identifies which audit observations are deficiencies that require corrective action. Once completed, the internal audit report is issued jointly to the laboratory general manager and the manager(s)/supervisor(s) of the audited departrnent(s) or operation(s). The responsible manager(s)/supervisor(s) respond with a plan to correct all of the deficiencies cited by the due date specified in the audit report. Each response must include timetables for completion of all proposed corrective actions. The QA Officer reviews and accepts the audit responses. If !he response is accepted, the QA Officer uses the action plan(s) and timetable(s) as a guideline for verifying completion of the corrective action(s). If the QA Officer determines that the audit response does not adequately address the correction of cited deficiencies, the response will be returned for modification. To complete the audit process, the QA Officer performs a re-examination of the areas where deficiencies were found to verify that all proposed corrective actions· have been implemented. An audit deficiency is considered closed once implementation of the necessary corrective action has been verified. If corrective action cannot be verified, the associated deficiency remains open until that action is completed. 12.2 EXTERNAL AUDITS Pace is audited as required by regulatory agencies to maintain laboratory certifications, and by various commercial clients. External audits include those by state laboratory certification agencies, USEPA, Army Environmental Center (AEC), Army Corps of Engineers, and other appropriate federal, state and private agencies (e.g., MITRE). Audit teams external to the company review the laboratory to assess the existence of systems, implementation of the systems, and degree of technical expertise. QA staff host the audit team and collect notes during the audit process. These notes are communicated to the General Manager, Laboratory Manager and the supervisor. Generally, the auditors will prepare a formalized audit report listing deficiencies observed and follow-up requirements for the laboratory. In some cases, in lieu of an official report, items of concern are discussed during a debriefing convened at the end of the on-site review process. · The laboratory staff and supervisors develop corrective action plans to address any deficiencies with the guidance of the QAO. The laboratory manager provides the necessary resources for staff to develop and implement the corrective action plans. The QAO collates this information and provides a written report to the audit team. The re port contains the corrective action plan and expected completion dates for each element of the plan. QAO staff follow-up with the laboratory staff to ensure that corrective actions are implemented. :\lqapreva\sect12.doc Date: 12/22/95 Section 12 Revision 0.01 Page 4 of 7 12.3 TOTAL QUALITY SYSTEM AUDIT The Corporate Quality Office coordinates on-site audits of each laboratory facility. The corporate audit is conducted by the Vice President of Quality with the assistance of Pace QAO's. This audit is designed to evaluate all aspects of facility operations and is not limited to only laboratory operations. Audits may either be system related or technical in nature, depending on the type of information needed for making quality improvements. Assessment of quality/technical practices within laboratory operations involves three types of review. 1. 2. 3. Documentation -On-going monitoring of quality issues in laboratory offices requires that the corporate quality office receive the following hardcopy information on an "as released' basis, supplied as part of the Quarterly Quality Report to management. · PE scores (e.g., WP, WS, CLP, COE, AEC, EML, EMSL, NIOSH, etc.) Certification/parameter list approvals • External audit reports and responses lnternai audit reports • Copies of all newly developed SOPs (also critical for the MRD program) Periodic data validation report reviews Pre-audit Documentation • Verify methods capability matrix infonmation ( completed for each method routinely perfonmed in the laboratory) Sample report for each reporting deliverable level used by the facility Facility and equipment inventory Current organization chart LIMS sample management reports Facility Audit Process • .. • Checklist approach with minimal textual report Objective scoring system based on checklist results Technical review based upon compliance to MRDs and associated SOPs, published methods, federal program requirements "GLP' documentation trail (retrace the path of a sample through the laboratory) Non-analytical documentation (e.g., training logs, discrepancy/corrective action reports, preventative maintenance, standards traceability, etc.) Review of QA procedures (relative to the Pace Generic Quality Assurance Plan) Credential/certification verification Instrumentation and facility utilization (e.g., automation, workflow) :\lqapreva\sect12.doc I I I I I I I I I I n I I I I I I I I I I I I D n D I I I I I I I I Date: 12/22/95 Section 12 Revision 0.01 Page 5 of 7 Health and safety, waste disposal programs (reviewed in conjunction with technical audit) Debriefing meeting Issue final audit report and establish corrective action schedule 12.4 PERFORMANCE EVALUATION AUDITS 12.4.1 Pace PE Samples Double blind performance evaluation (PE) samples are submitted periodically to all Pace laboratories to evaluate all areas of the laboratory. The program is identified as the Pace lnterlaboratory Testing Survey (PITS). Results from internal PE sample analyses are processed by the analy1ical departments and reported to the responsible Project Manager. The Project Manager issues a standard analytical report to a fictitious client (PE vendor) in a manner identical to that done for all other client work orders. Evaluated PE results are given to the laboratory for review. For parameters where the reported results fall within the defined acceptance limits, no further action is required. For parameters where the reported results fall outside· acceptance limits, the responsible Department Manager/Supervisor must investigate in an effort to find the root cause of each problem. For each missed quantitation, the Manager/Supervisor summarizes the findings of their investigation in a PE investigation report. Each report must include a description of the corrective action(s) that will be taken to prevent recurrence of the problem. Completed reports are passed to the QA Officer for review, then forwarded to Pace's corporate quality office. The internal auditing process is. used to verify implementation of corrective actions. 12.4.2 EPA WP and ws PE Studies . Pace labs routinely participate in EPA's Water Pollution (WP) and Water Supply (WS) round-robin PE studies. Each of these studies is conducted twice per year. Generally, all analy1ical sections of the laboratory participate in the WP and WS PE -studies. Examples of parameters analyzed under each study are listed below. Satisfactory perfonnance on these studies is essential as they are a fundamental requirement of state accreditation programs. ;\lqapreva\sect12.doc Date: 12/22/95 Section 12 Revision 0.01 Page 6 of 7 • EPA Performance Evaluations -Water Supply -Semiannual (April and September) • Trace Metals Nitrate/Nitrite/Fluoride Insecticides Herbicides PAHs Adipate/Phthalates Trihalomethanes (THMs) Volatile Organic Compounds Turbidity Total Filterable Residue Calcium (as CaCO3) pH Alkalinity Corrosivity Sodium Sulfate Total Cyanide EPA Performance Evaluations -Water Pollution -Semiannual (February and August) · Trace Metals Minerals Nutrients Demand PCBs PCBs in Oil Pesticides Volatile Halocarbons Volatile Aromatics Total Cyanide Non-Filterable Residue Oil and Grease Total Phenolics For PE quantitations evaluated as being outside the acceptance ranges, investigations must be performed by the responsible Manager(s)/Supervisor(s) and reports must be completed as described above for the Pace PE samples. The reports are reviewed by the QA Officer, then forwarded to the various state accreditation agencies for their review. :\lqapreva\sect12.doc I I I I I I I I I I I I I I I I I I I a I n ' ~ I E I I I I I I I 12.4.3 Other PE Studies Date: 12/22/95 Section 12 R.evision 0.01 Page 7 of7 Other PE samples may be performed by Pace in conjunction with a specific program or contract. Examples include: PE samples distributed by the US Army Corps of Engineers, Missouri River Division, as part of their laboratory evaluation process Quarterly blind (QBs) PE samples distributed to laboratories participating in EPA's Contract Laboratory Program In addition, clients may arrange for PE samples to be analyzed, either as part of a laboratory evaluation process or as a periodic performance check. :\lqapreva\sect12.doc I I I R u u D I I I I I I I I I 13.0 PREVENTIVE MAINTENANCE Date: 12/22/95 Section 13 R,evision 0.01 Page 1 of 6 The objective of Pace's preventive maintenance program is to establish a system of instrument care that prevents the loss of analytical quality control and results in a minimum of lost productivity due to instrument failure. This program includes a system for documenting all routine and non-routine instrument maintenance and repairs. Pace maintains service contracts for mo_st major analytical equipment including chromatographic instruments. balances, atomic absorption. and inductively coupled plasma instruments. All equipment and instruments generating analytical results have calibration and maintenance records. 13.1 MAINTENANCE RESPONSIBILITIES The Laboratory Operations Manager and Supervisors are responsible for providing technical leadership to all staff involved with chemical analysis. This leadership role includes serving as a technical resource to help solve equipment and method problems, evaluating and recommending investments in new technologies, improvin~1 efficiency, and coordinating instrument repair and maintenance. The primary responsibility for the maintenance of instruments and equipment rests with each analytical Department Manager/Supervisor. The Department Manager/Supervisor is further responsible for developing procedures and schedules for maintaining each major instrument or piece of equipment and for delegating specific maintenance responsibilities to department staff. 13.2 MAINTENANCE SCHEDULES The effectiveness of the maintenance program relies heavily on adherence to prescribed schedules for maintaining each instrument or piece of equipment. A schedule is established for all routine maintenance. Other maintenance activities may also be identified as requiring attention on an as-needed basis. Manufacturers' recommendations provide· the primary basis for developing these schedules, and manufacturers' service contracts provide primary maintenance for some major instruments. To minimize downtime and interruption of analytical work, preventive maintenance is routinely performed on each analytical instrument. SOPs are written for each instrument that cover basic operation and maintenance procedures. The following are brief summaries of maintenance for each type of major instrumentation. This information is also listed by major instrumentation system in Table 13.1. 13.2.1 Preventive Maintenance -GC/MS :\lqapreva\sect13.doc Regularly performed maintenance includes, but is not limited to the following for GC/MS instrumentation: 13.2.2 13.2.3 :\lqapreva\sect13.doc Hard tune with calibration gas (pftba) Date: 12/22/95 Section 13 Revision 0.01 Page 2 of6 Removal of 2-3 inches from the injection end of the capillary columns Replacement of 2-3 inches of column packing from the injection end of packed columns Injection port liner replacement Replace injection port septum Clean ion source as needed Check vacuum pump oil level Check carrier gas tanks Replace or recondition vent traps Preventive Maintenance -GC I I I I I I I Regularly performed maintenance includes, but is not limited to the following 1 for extractable GC instrumentation: Removal of 5-10 inches of guard column (if applicable) and 2-3 inches from 1 the injection end of the capillary columns Replacement of 2-3 inches of column packing from the injection end of · packed columns Injection port liner and RP seal replacement I Replacement of septum Fill solvent rinse bottles in auto sampler Check carrier and support gases I NRC wipe test ECO Regularly performed maintenance includes, but is not limited to, the following for volatile organics GC instrumentation: Clean and bake sparge tubes • Replace trap as needed Check carrier and support gases • Replace transfer line as needed • Replace nickel tube as needed • Clean or replace PIO as needed Preventive Maintenance -ICP Check liquid argon tank level • Change pump tubing • Clean nebulizer and spray chamber as needed • Replace and realign plasma torch when required Check cooling system water level • Empty waste reservoir when full I I I I I I I I I I I a. I n D H I E m I I I I I I I 13.2.4 13.2.5 13.2.6 13.2.7 :\lqapreva\sect13.doc Preventive Maintenance -AA Graphite Furnace Check and align source lamps Date: 12/22/95 Section 13 Revision 0.01 Page 3 of6 Clean and inspect graphite tube, replacing when surface appears excessively burnt or cracked Clean and inspect contact ring, replacing when excessively worn Clean mirrors for optical sensor and sample compartment windows Check autosampler injector alignment and deposition Preventive Maintenance -Mercury Analyzer Check and align source lamp • Remove and clean sample cell and connecting tubes • Check sparger for proper operation Clean sample compartment windows Preventive Maintenance -General Laboratory Areas Calibrate automatic pipets and burets monthly Clean, check, calibrate to manufacturers specifications all pH, DO, conductivity and, turbidity meters, and spectrophotometers annually • General housekeeping: keep counter tops, hoods, and floors clean and keep safety equipment accessible Check airflow in hoods once a quarter Preventive Maintenance Thermometers, Refrigerators, Ovens and Balances: Laboratory thermometers are calibrated against NIST traceable thermometers annually. Digital thermometers are calibrated quarterly. 1he results are recorded in a logbook specific to that purpose. Correction factors are recorded on the thermometer tags, along with the unique thermomet,3r identification number and calibration date, and are used by Pace personnel to correct actual temperature measurements. The correction factor is applied to each reading until the thermometer is calibrated again. Use of thermometers with a correction of> 5° C is avoided. Pace minimizes the need to apply corrections by utilizing the correct media, thermometers and procedures during calibration. Refrigerators, freezers and ovens are monitored once or twice daily or ·as used, dependent upon the function of the unit. Logbooks are maintained for documentation of readings and corrective actions. If a unit fails acceptance criteria, monitoring is continued until the temperature stabilizes within the range or appropriate corrective actions are taken. Monitoring occurs at one hour intervals for a . maximum four hour period; if the reading following the temperature control adjustment is out, the unit is considered "out of order", and Date: 12/22/95 Section 13 Revision 0.01 Page 4 of6 is emptied and serviced. It is not put back into service until shown to be stable at the required temperature range. Analytical balances are calibrated annually (or more frequently if required under a specific program) by an outside service. A dated sticker, certifying the calibration, is placed on each balance. Records for balance calibration/servicing are maintained in Pace QA files. Multi and single point calibration checks are regularly performed to ensure the accuracy of each balance. The results are recorded in dedicated logbooks that are maintained at each balance location. Balances that do not satisfy specifications are taken out of service for replacement or repair. ASTM Class "1" weights must be verified/calibrated every two years. 13:3 MAINTENANCE DOCUMENTATION All routine and non-routine instrument maintenance is documented in maintenance I I I I I I I I logbooks assigned to each instrument. To provide a clear and complete history of repairs I and maintenance associated with each instrument, each maintenance entry must include the following elements: 1. 2. 3. An explanation of the reason for the maintenance or repair, e.g., was this action taken to fix a problem or was it part of routine instrument maintenance A full description of the maintenance or repair actions taken A description of how the analyst demonstrated that the analytical system was operating in control after completion of the maintenance actions, but before the resumption of sample analysis When maintenance is performed to repair an instrument problem, the entry should include a description of the symptoms or problem that precipitated the maintenance actions. Depending on the initial problem, demonstration of return to control may be satisfied by the successful analysis of a reagent blank or continuing calibration standard. The entry must include a summary of the results of that analysis and a verification by the analyst that the instrument has been returned to an in-control status. In addition, each entry must include the initials of the analyst making the entry, the dates the maintenance actions were performed, and the date the entry was made in the maintenance logbook, if different from the date(s) of the maintenance. 13.4 SPARE PARTS Along with the development of maintenance schedules, an adequate inventory of spare parts is required to minimize equipment downtime. This inventory should emphasize those parts and supplies that: are subject to frequent failure, :\lqapreva\sect13.doc I I I I I I I I I I I I a g D D D I I I I I I I I have limited useful lifetimes, or cannot be obtained in a timely manner should failure occur. Date: 12/22/95 Section 13 Revision 0.01 Page 5 of6 Department Manager/Supervisors are responsible for maintaining an adequate inventory of necessary spare parts for all major instruments and equipment items. Examples of spare parts maintained for major instrumentation systems are listed in Table 13.1. :\lqapreva\sect13.doc TABLE 13.1 Date: 12/22/95 Section 13 Revision 0.01 Page 6 of 6 Scheduled Maintenance Procedures and Representative Spare Parts for Major Instrumentation INSTRUMENT MAINTENANCE PROCEDURE SPARE PARTS Gas Chromatography 1. Change septa and inserts as 1. Syringes Mass Spectrometry needed 2. Septa and inserts (GC/MS) 2. Clip column 3. GC columns 3. Replace pump oils as needed 4. Various electronic components 4. Change gas line dryers as 5. Plumbing supplies -tube needed fittings 5. Clean source as needed 6. Replace electron multiplier as needed Gas Chromatography (GC) 1. Change septa and inserts as 1. Syringes needed 2. Septa and inserts 2. Clip column 3. GC columns 3. Change gas line dryers as 4. Various electronic components needed 5. Plumbing supplies -tube 4. Leak check when installing fittings new analytical column 5. Check inlet system for residue buildup periodically Purge and Trap Sample 1. Replace trap as needed 1. Spare traps Concentrator 2. Decontaminate system as 2. Various electronic components required by blank analysis and circuit boards 3. Leak check system 3. Plumbing supplies -tube 4. Measure flowrates for each fittings sparging position monthly Inductively Coupled Argon 1. Clean torch assembly and 1. Spare torch and mixing Plasma Spectrometer (ICP) mixing chamber when dis-chamber colored or after 8 hours of 2. Spare coil running high dissolved solids 3. Plumbing supplies -tube samples fittings Graphite Furnace Atomic 1. Change graphite contact rings 1. Contact rings Absorption as needed 2. Graphite cups and electrodes Spectrophotometer 2. . Clean quartz windows as 3. Autosampler tubing needed Hg Analyzer 1. Clean tubing and quartz cell 1. Quartz cells as needed 2. Aspirator 2. Clean aspirator as needed 3. Plumbing supplies 3. Replace drying tube media daily :\lqapreva\sect13.doc , I I I I I I I I I I I I I I I I I I I I n g u H u I I I I I I I I 14.0 DATA QUALITY ASSESSMENT Date: 12/22/95 Section 14.0 Revision 0.01 Page 1 of 7 Data quality assessment requires the review of quality control samples for precision, accuracy, representativeness, completeness, and comparability. Precision and accuracy data are used to determine the acceptability of analytical results. Standard operating practices require the use of a minimum of 20 tabulated precision or accuracy data points to prepared quality control charts. However, preliminary control limits can be established using as few as four data points. The Shewhart technique is the statistical method used to construct the charts. These quality control charts provide a quick visual means for monitoring the daily performance of the laboratory and identifying nonconformance trends. For every batch of samples analyzed, a series of quality control samples are analyzed to assess the precision, accuracy and validity of the analysis. These data are reviewed before release of the data. All QC data are stored at Pace and are useable for determination of method precision and accuracy. Pace makes every effort to meet or exceed the accuracy and precision data as defined within specific methodologies. However, for actual matrices these data may not be comparable. To estimate accuracy, spiked blank samples (laboratory control samples) and matrix spike sample recoveries are evaluated. This allows for the determination of both method and actual sample batch accuracy. Precision is measured and monitored in two ways: using range control for duplicate pairs and relative percent differences. Pace uses the formulas presented in Standard Methods, SW 846 and the USEPA Quality Assurance handbooks for calculations of precision and accuracy. This section illustrates calculations for determining data quality in terms of precision and accuracy. In addition to calculations concerning precision and accuracy, those which pertain to representativeness, completeness, and comparability are used to ascertain the level at which DQOs have been satisfied. Calculations for these other data quality indicators are included as well. 14.1 PRECISION Precision is the degree to which the measurement is reproducible. Precision can be assessed by duplicate measurements of a laboratory control sample or an environmental sample. The precision of laboratory analytical data can be expressed using one of several statistical determinations, including: 1-standard deviation, 2-range, 3-relative standard deviation, also known as the coefficient of variation, and 4-relative percent difference. Standard deviation is a measure of the variance of individual observations from the mean. It is usually denoted as "s" and is defined as: n n EX/-(D</tn i=1 i=1 s = or n -1 :\lqapreva\sect14.doc In this equation, n X; X = population size = ith observation in the sample = sample mean Date: 12/22/95 Section 14.0 Revision 0.01 Page 2 of? Standard deviation can be used to determine variation among several RPO values for duplicate pairs and establish statistical limits for duplicate RPO. Range control may also be used. Range is the largest observation in a data set minus the smallest observation in the data set, often denoted as "R". Where: R= A-B X= A..±-6 R X A B n n = Range of a pair of results = Average of a pair of results = Duplicate value 1 = Duplicate value 2 =: 2 (represents a single duplicate pair) To graphically represent the data of numerous duplicate pairs on control _charts, the following calculations are performed using statistical numbers. Where: X= the sum of X / n R = the sum of R / n X R X R n = Grand Mean = Average Range = Average of a pair of results = Range of a pair of results = 2 (represents a single duplicate pair) Control limits for ranges (R -bar chart): CL= 3.27 (R) WL = R + 2/3 (3.27 R-R) Where: R = Average Range CL = Control Limit WL = Warning Limit To determine if the proper range control chart is being used for evaluation of a duplicate pair of results, the X control chart may be used. :\lqapreva\sect14.doc I I I I I I I I I I I I I I I I I I I I I I g n 0 H D I I I I I I I I Control limits for averages (X -bar chart): Where: UCL = X + 1.88 (R) LCL = X-1.88 (R) UWL = X + 2/3 (1.88 R) LWL = X -2/3 (1.88 R) X R UCL LCL UWL LWL = Grand Mean = Mean Range = Upper Control Limit = Lower Control Limit = Upper Warning Limit = Lower Warning Limit Date: 12/22/95 Section 14.0 Revision 0.01 Page 3 of. 7_. Relative standard deviation (RSD), or coefficient of variation (CV), is a commonly used measure of variability that is adjusted for the magnitude of the values in the sample: RSD = Standard Deviation x 100%. Mean RSD is used most often when the size of the standard deviation changes with the size of the mean. Individual measurements of RSD or CV can be combined (pooled) to give an overall measure of variability for a given type of analysis or measurement: Pooled CV= In this equation, n 2 ~ DF1 i=1 Xi = CV of data set i DF1 = degrees of freedom from data set i (k; -1) n = number of data sets k; = number of data points in set i = data set 1, 2, 3 ... n Relative percent difference (RPO) is another commonly used measure of variability that is adjusted for the magnitude of the measured values. It is used only when the sample :\lqa preva~ect14 .doc Date: 12/22/95 Section 14.0 Revision 0.01 Page 4 of? contains two observations. The results of the duplicate analyses are computed and the absolute RPO is calculated using the following equation: Where: RFD= Ix, -x,I xlOO x1 +x2 2 = Relative Percent Difference = first sample value (original) = second sample value (duplicate) For duplicate results RPO is directly related to RSD by: RPO= )2RSD The RPDs are tabulated, the average RPO and standard deviation are calculated, and a control chart constructed. Formulas for control limit are: 14.2 ACCURACY UCL = B.E.Q+ 35 UWL = RPO +25 LCL always equals 0 Accuracy measures the degree of difference between observed and true values. The actual test result is compared to the theoretical result of 100% recovery and the percent recovery calculated. The accuracy of sample data can be assessed using the laboratory control spike, the environmental sample spiked with target analytes (matrix spike), or surrogate standards. Accuracy data are evaluated against established control limits. The percent recovery is computed using the following equation: Where:· M B T M-B %R=--xl00 T = Measured concentration of analyte in spiked sample = Background concentration of unspiked sample = Target value (known concentration of analyte in spike) The percent recovery data for a compound or parameter are tabulated, the average percent recovery and standard deviation calculated, and a control chart constructed. The accuracy of laboratory analytical data can also be presented in terms of: 1-percent relative error, and 2-confidence intervals at the 95% level. :\lqapreva\sect14.doc I I I I I I I I n I I I I I I I I I I I a n n D E E I I I I I I I I I I I Date: 12/22/95 Section 14.0 Revision 0.01 Page 5 of 7 Percent Relative Error = Measured Value -Actual Value x 100% Actual Value 95% Confidence Interval = In this equation, X s n a· X + = sample mean = sample standard deviation = sample size = risk level (0.025 for 95% confidence interval) = value of the tabulated student's "t" distribution for n-degrees of freedom and risk level a Percent recovery is related to percent relative error by: % Recovery = % Relative Error + 100 The correlation coefficient, "r'' is used to determine the acceptability of multi-point initial calibration data. The correlation coefficient value reflects the degree of fit of the calibration data with a linear or other curve function and is calculated as: n(Exy) -(Ex) (Ey) r = In this equation, x = concentration of the standard y = instrument response (peak area) n = number of calibration points (x,y data pairs) 14.3 CONTROL CHARTS Once a minimum of 20 QC data points are tabulated, an accuracy control c:hart can be constructed as follows: Compute the mean value of the tabulated points. Where: :\lqapreva\sect14.doc p P; n "p P=L., I n = Mean Value = Sample result = Total number of results in data set Date: 12/22/95 Section 14.0 Revision 0.01 Page 6 of? Using the mean (P), compute the standard deviation (SD) of the data set. Where: SD X p n SD= = = = = I(x-P)' (n-1) Standard deviation Sample result Mean value Total number of results in data set I I I I I I Employing the mean and standard deviation of the data set, determine the upper and I lower warning and control limits are determined as described in 14.3.1 and 14.3.2. 14.3.1 14.3.2 Warning Limits I Warning limits represent the 95% confidence interval and are equal to the ii mean value for the control sample plus or minus two standard deviations ti (2SD). Exceeding these limits is warning that the analy1ical system may be approaching an out-of-control situation and should be inspected for possible sources of error. The warning limits are calculated with the following equation: I Where: Control Limits WL= P±2SD WL p SD = = Warning limits Mean Standard deviation Control limits represent the 99% confidence interval and are equal to the mean value of the control sample, plus or minus three standard deviations (3SD). Exceeding these limits indicates that the analy1ical system is out-of-control. Control limits are calculated using the following equation: Where: CL= P±3SD CL p SD = = = Control limits Mean Standard deviation I I 14.3.3 Utilization of Acceptance Limits I I I I I I I :\lqapreva\sect14.doc Once the warning and control limits are established, a control chart is constructed. To verify the control chart, the initial data points are checked against the newly generated limits for statistical outliers. Subsequently generated QC sample I I I n D D I I I I I I I I I I I Date: 12/22/95 Section 14.0 Revision 0.01 Page 7 of 7 data are then plotted on the chart. The plotted points must fall within the control limits for the result to be accepted and the associated sample data validated. Outliers are evaluated for corrective action measures. Control charts are prepared for each required parameter, the limits updated at least annually and graphs produced for identification of trends. · The laboratory must also review control charts for out-of-statistical control trends. Any of the following trends is considered an out-of-control trend occurrence. • Any three consecutive points are outside warning limits. • Any seven consecutive points are on the same side of the mean of the central line. • Any six consecutive points are such that each point is larger (smaller) than its immediate predecessor. • Any obvious cyclic pattern in the points. 14.4 REPRESENTATIVENESS Representativeness is a qualitative element related to the ability to collect a sample that reflects the characteristics of that part of the environment being assessed. Sample representativeness is dependent on the sampling techniques used and is considered individually for each project site. It is specifically addressed in each work plan. 14.5 COMPLETENESS Completeness is a measure of the amount of valid data obtained from a measurement system compared with the amount that was expected to be obtained unde,r normal conditions. It is expected that laboratories should provide data meeting QC acceptance criteria for 95% or more of the requested determinations. It is necessary for data users to identify any sample types which require 100% completeness. The mathematical formula is as follows: Where: 14.6 COMPARABILITY C=½xIOO C V T = = = Percent completeness Number of measurements jud!Jed valid Total number of measurements Comparability expresses the confidence with which one data set can be compared to another data set measuring the same property. For example, the use of EPA approved methods and procedures ensure comparability with other data from previous of following studies using the same methods. :\lqapreva\sect14.doc I I g u 0 D D I I I I I I I I I I 15.0 CORRECTIVE ACTION Date: 2/28/97 Section 15 Revision 0. 02 Page 1 of 13 This section describes the quality assurance system at Pace Analytical, established to address the documentation and correction of problems encountered during sample handling and analysis. In many instances. the accurate and timely communication of sample handling and analysis problems to laboratory Managers/Supervisors, Project Managers, and/or the QA Officer can mean the difference between a situation that is corrected with little or no impact to sample data quality and one that results in resampling. When a problem arises, action to correct the problem must be taken promptly. For example, laboratory Managers/Supervisors must initiate corrective action whenever QC results for control parameters fail to stay within acceptance limits. In many situations, input from the client is important in deciding how problems are to be resolved. When errors, deficiencies, unusual occurrences, or out-of control situations exist, the QA program provides systematic procedures, called "corrective actions", to resolve problems and restore proper functioning to the analytical system. Within Pace Analytical, a distinction is made between "out-of-control events" and "unusual occurrences" for the purposes of requiring corrective actions. An out-of-control event is any event which is beyond the acceptance limits established for laboratory operation by Pace Analytical SOPs, EPA methods, or client specific contracts or protocols. This can be due to data which are outside of the accepted bounds for accuracy and/or precision, method contamination, improper instrument calibration or maintenance, or deviations from the contract or SOP detected by a QA audit. An unusual occurrence is a situation in which the analytical system is, strictly speaking, compliant with the protocol or SOP and therefore in control but an atypical or undesirable incident has occurred which warrants further investigation. Such an occurrence could be a holding blank which is contaminated or differences in the pattern of non-spiked target compounds between a spiked and unspiked aliquot of a sample used as the matrix spike. Both out-of-control events and unusual occurrences are formally documented. Within Pace Analytical, the formal documentation report is identified under the following designations: Corrective Action Report (CAR); Non-conformance Memo (NCM); or Discrepancy Report (DR). Each of these reports serves the same purpose (because of this, the names are used inter- changeably) of documenting whenever either type of event is noted. A representative report form is illustrated in Figure 15.1. 15.1 NON-CONFORMANCE MEMO The primary tool for documenting deficiencies and problems is the nonconformance memo (NCM). NCMs may be initiated by an analyst, laboratory Manager/Supervisor, Project Manager, or other laboratory personnel. The NCM is used to document a specific problem or deficiency noted during sample handling or analysis. Depending on the specific problem or deficiency, corrective actions may be taken by the Analyst, laboratory :\lqapreva\sect15.doc Date: 2/28/97 Section 15 Revision 0.02 Page 2 of 13 Manager/Supervisor, Project Manager, or other laboratory personnel. Since problems encountered with sample analysis often have the potential to impact data quality, appropriate corrective action is frequently determined in communications between the laboratory Project Manager and the client. Each NCM requires the initials of the person documenting the problem as well as those of any person documenting additional information or corrective action. Each NCM is then reviewed and initialed by the department Manager/Supervisor. If the deficiency or problem impacts client sample data, the department Manager/Supervisor passes the NCM directly to the appropriate Project Manager for their review and followup. If the nonconformance does not directly impact sample data quality or integrity, the department Manager/Supervisor passes the NCM to the QA Officer. The NCM must describe the actions taken at each step of the review process. Where an NCM documents an analytical event that is judged to be out-of-control, evidence of return to control must also be documented. Once documentation of the problem, corrective action, and return to control is complete, the NCM is forwarded to the QA Officer for QA review. QA review is documented with the QA Officer's initials and the completed original NCM is either passed back to the appropriate Project Manager to be filed in the client project file or is filed in the QA files, depending on the nature of the nonconformance. 15.2 OUT OF CONTROL EVENTS Out-of-control events associated with the statistical analysis and review of data are straight forward to identify. The Analyst generating the data is responsible for checking the results against the established limits. Any deviations are immediately addressed. If data are outside accepted limits, the Analyst immediately notifies the responsible Section Supervisor. If the situation can not be corrected to prevent an out-of-control condition, the Section Supervisor shall notify the Operations Manager and the Quality Assurance Officer. The Operations Manager and Group Supervisors are respo.nsible for identifying the source of the problem and initiating corrective action. Completion of corrective action should be evidenced by the return of data to prescribed acceptable limits. Events which do not cause an immediate obvious effect on data quality are more difficult to identify. Such events could be samples stored at an incorrect temperature or held beyond prescribed holding times, or improper maintenance of records. Everyone in the laboratory is responsible for reporting "system" problems. Analysts should report out-of- control events to their Group Supervisor, who should then in turn report the situation to the Operations Manager. Corrective action is again the responsibility of the Operations Manager and the Group Supervisors. They shall review and approve the action taken. If an out-of-control event does occur during analysis, for instance an LCS recovery falls outside the expected range, the analyst must describe on the corrective action report the event, the investigative and corrective actions taken, the cause of the event, and notify the QA Officer. In some cases, investigation of an out-of-control event will reveal no problems. In such cases, only the event and the investigative action is recorded. :\lqapreva\sect15.doc I I I I I I I I I I I I I I I I I I I I I n D I I I I I I I I I I I I I Date: 2/28/97 Section 15 Revision 0.02 Page 3 cl' 13 The investigative action taken is somewhat dependent on the analysis and the event. However, listed below is a progression of steps which may be taken to find the cause of an out-of-control event: Check calculations to ensure there are no errors • Check standard and spiking solutions for degradation or contamination Check instrument performance If the problem is with the standards or instrument performance, the analyst must recalibrate or retune the instrument before reanalyzing the sample extracts affected. If the out-of-control condition is still not remediated, the samples may require reextraction and reanalysis or data qualification. It is occasionally necessary to qualify data when the accompanying quality control data are not within established performance criteria. The qualifying of data alert the data end user to the fact that the analysis was, to some degree, flawed and that the precision and accuracy of the data produced may not fulfill the data quality objectives (DC!Os) for that particular project. Based on the project DQOs, analytical data with qualifiers may not be appropriate for the intended use. 15.2.1 Volatile Organic Analyses :\lqapreva\sect15.doc Method Blanks If target compounds are detected in the method blank above the detection limit (or reporting limit if different from the detection limit) (above 5 times the detection and/or reporting limit for methylene chloride, acetone, toluene, and 2-butanone) the corrective action consists of checking the calculations, reanalyzing the blank, qualifying the associated sample data, and investigating the source of the problem to implement corrective action for .the future. When target list compounds are detected in a method blank, the following condition applies: When any target compound is detected in a method blank above the action levels listed earlier, but not in associated samples, then no qualifier is applied. Surrogates The % recovery of each surrogate is calculated for each sample, blank, and LCS. Corrective action is taken whenever one (or more) surrogate recovery is outside the acceptance criteria. The following corrective actions are taken when required as stated above: Check calculations to assure there are no errors; :\lqapreva\sect15.doc • • • • Date: 2/28/97 Section 15 Revision 0.02 Page 4 of 13 Check internal standard and surrogate solutions for degradation, contamination, etc., and check instrument performance; If instrument failure is indicated, reanalyze the sample; If a method blank surrogate is outside of acceptance criteria, then the problem must be corrected before proceeding with sample analysis. This may include reanalysis, reextraction or recalibration; If the surrogate could. not be measured because the sample required a dilution, no corrective action is required. The recovery of the surrogate is recorded with the note "surrogate diluted out". If all QC associated with the sample is within acceptance limits (the method blank surrogate recovery and LCS spike recovery}, the problem may be attributed to a matrix effect. To identify the matrix as the problem, reanalyze the sample. Samples exhibiting a matrix effect will be qualified and discussed in the report narrative. Laboratory Control Samples The % recovery of each spiked analyte in the Laboratory Control Sample (LCS) is calculated. Corrective action is taken whenever the % recovery is outside the established acceptance criteria for that analyte. The following corrective actions are taken when required as stated above: • • Check calculations to assure there are no errors; Check internal standard and spiking standard solutions for degradation, contamination, etc., and check instrument performance; Reanalyze samples associated with a failed LCS, if available; If that does not correct the problem, then the data is reported and a qualifying statement included in the report narrative. For Matrix Spike and Matrix Spike Duplicates, if all QC associated with a sample is within acceptance limits (method blank and LCS spike recoveries), the problem may be attributed to a matrix effect. Samples exhibiting a matrix effect will be qualified and discussed in the report narrative as appropriate. Calibration For an initial 5 point calibration curve to be valid, the % relative standard deviation of the individual relative response factors (RRF) for the Calibration Check Compounds (CCC) shall be less than or equal to 30%. If this criteria is not met, then the calibration curve shall be reanalyzed. I I I I I I I I I I I I I I I I I I I I I n n I I I I I I I I I I I I I 15.2.2 I :\lqapreva\sect15.doc Date: 21;18197 Section 15 Revision 0.02 Page 5 ol' 13 For continuing calibration checks to be valid, the relative response factor for each of the System Performance Check Compounds (SPCC) should be at least 0.300 (0.250 for Bromoform) and the RRF for each of the CCC should be .:::20% different from the average RRF from the initial calibration. If these criteria are not met, then the following corrective actions should be taken: Check internal standard and standard solutions for degradation, contamination, etc., Check instrument for contamination at the injection port inlet and front end of the column; If no source of the problem is identified, then a complete 5 point initial calibration must be performed. The SPCC and CCC for Volatiles are: ~ Chloromethane 1, 1-Dichloroethane Bromoforrn 1, 1,2,2-Tetrachloroethane Chlorobenzene Semivolatile Organic Analyses Method Blanks CCC Vinyl Chloride 1, 1-Dichloroethene Chloroform 1,2-Dichloropropane Toluene Ethylbenzene If target compounds are detected in the method blank above the detection limit (or reporting limit if different from the detection limit) (above 5 times the detection limit and/or reporting for phthalate esters) the corrective action consists of the following: • Checking the calculations; • • Reanalyzing the blank; Flagging the associated sample data; Investigating the source of the problem to implement corrective action for the future. :\lqapreva\sect15.doc Date: 2/28/97 Section 15 Revision 0.02 Page 6 of 13 When target list compounds are detected in a method blank, the following data condition applies: • When any target compound is detected in a method blank above the action levels listed earlier but not in associated samples, then no flag is applied. Surrogates The % recovery of each surrogate is calculated for each sample, blank, and standard. Corrective action is taken whenever one (or more) surrogate recovery from either the base/neutral or acid fraction is outside the acceptance criteria. The following corrective actions are taken when required as stated above: Check calculations to assure there are no errors; Check internal standard and surrogate solutions for degradation, contamination, etc., and check instrument performance; If instrument failure is indicated, reanalyze the sample; If more than one method blank surrogate is outside of acceptance criteria or if one surrogate yields less than 10% recovery, then the problem must be corrected before proceeding with sample analysis. This may include reanalysis, reextraction or recalibration; • If the surrogate could not be measured because the sample required a dilution, no corrective action is required. The recovery of the surrogate is recorded with the note "surrogate diluted out"; If all QC associated with the sample is within acceptance limits {the method blank surrogate recovery and LCS spike recovery), the problem may be attributed to a matrix effect. If any one surrogate yields less than 10% recovery or if more than one surrogate in a fraction fails, reanalyze the sample to demonstrate matrix interference. Samples exhibiting a matrix effect will be qualified and discussed in the report narrative. Laboratory Control Samples The % recovery of each spiked analyte in the Laboratory Control Sample is calculated. Corrective action is taken whenever recovery is outside the acceptance criteria. The following corrective action is taken when required as stated above: Check calculations to assure there are no errors; Check internal standard and spiking standards solutions for degradation, contamination, etc., and check instrument performance; I I I I D I I I I I I I I I I I I I I I n D m I I I I I I I I I I I I I I :\lqapreva\sect 15.doc Reanalyze all associated samples, if available; Date: 2/28/97 Section 15 Revision 0.02 Page 7 of 13 If that does not correct the problem, then the data is reported and a qualifying statement regarding the laboratory control sample is included in the report narrative. For Matrix Spike and Matrix Spike Duplicates, if all QC associated with a sample is within acceptance limits (the method blank and LCS/LCS dup spike recovery), the problem may be attributed to a matrix effect. Samples exhibiting a matrix effect will be qualified arid discussed in the report narrative as appropriate. Calibration For an initial 5 point calibration curve to be valid, the % relative standard deviation of the individual relative response factors (RRF) for the Calibration Check Compounds (CCC) should be less than or equal to 30%. If this criteria is not met, then the calibration curve should be reanalyzed. For continuing calibration checks to be valid, the relative response factor for each of the System Performance Check Compounds (SPCC) should be at least 0.050 and the RRF for each of the CCC should be :o. 20% different from the average RRF from the initial calibration. If these criteria are not met, then the following corrective actions should be taken: Check internal standard and standard solutions for degradation, contamination, etc., Check instrument for contamination at the injection port inlet and front end of the column; If no source of the problem is identified, then a complete 5 point initial calibration must be performed. The SPCC and CCC for Semivolatiles are: .secc n-Nitroso-di-n-propylamine Hexachlorocyclopentadiene 2,4-Dinitrophenol 4-Nitrophenol Acenaphthene 1,4-Dichlorobenzene Hexachlorobutadiene n-Nitroso-di-phenylamine Di-n-octylphthalate Fluoranthene Benzo(a)pyrene Date: 2/28/97 Section 15 Revision 0.02 Page 8 of 13 4-Chloro-3-methylphenol 2,4-Dichlorophenol 2-Nitrophenol Phenol Pentachlorophenol 2,4,6-Trichlorophenol 15.2.3 Gas Chromatography Analyses Method Blanks :\lqapreva\sect15.doc If target compounds are detected in the method blank above the detection limit (or reporting limit if different from the detection limit) the corrective action consists of the following: • Checking the calculations; Reanalyzing the blank; Flagging the associated sample data; Investigating the source of the problem to implement corrective action for the future. When target compounds are detected in a method blank, the following conditions apply: When any target compound is detected in a method blank above the action levels listed earlier, but not in associated samples, then no flag is applied. Surrogates The % recovery of each surrogate is calculated for each sample, blank, and standard. Corrective action is taken whenever one (or more) surrogate recovery is outside the acceptance criteria. The following corrective action is taken when required as stated above: • Check calculations to assure there are no errors; Check standard and surrogate solutions for degradation, contamination, etc., and check instrument performance; If instrument failure is indicated, reanalyze the sample; I I I I I I I I I I I I I I I I I I I n D m I I I I I I I I I I I I I I I I :\lqapreva\seci15.doc Date: 2/28/97 Section 1S Revision 0.02 Page 9 of 13 If the surrogate could not be measured because the sample required a dilution, no corrective action is required. The recovery of the surrogate is recorded with the note "surrogate diluted out"; If all QC associated with the sample is within acceptance limits (the method blank surrogate recovery and LCS spike recovery), the problem may be attributed to a matrix effect. Samples exhibiting a matrix effect will be qualified and discussed in the report narrative. Laboratory Control Samples The % recovery of each spiked analyte in the Laboratory Control Sample is calculated and corrective action is taken whenever recovery is outside the acceptance criteria. The following corrective action is taken when required as stated above: Check calculations to assure there are no errors; Check standard and spiking standard solutions for degradation, contamination, etc., and check instrument performance; • If that does not correct the problem, then the data is reported and a qualifying statement regarding the laboratory control sample is included in the report narrative. Calibration For an initial 5 point calibration curve to be valid, the responses for each compound should be linear over the calibration range (generally linearity is defined as having an RSD :s; 20%). If this criteria is not met, then the calibration curve should be reanalyzed. For continuing calibration checks to be valid, the % difference in the calibration factor for each compound in calibration should not exceed 15% from the initial calibration. If these criteria are not met, then the following corrective actions should be taken: · • Check standard solutions for degradation, contamination, etc., • Check instrument for contamination at the injection port inlet and front end of the column; If no source of the problem is identified, then a complete 5 point initial calibration must be performed. Date: 2/28/97 Section 15 Revision 0.02 Page 10 of 13 15.2.4 Metals Analyses Method Blanks :\lqapreva\sect15.doc If target analytes are detected in the method blank above the reporting limit the corrective action consists of the following: • • • Checking the calculations; Reanalyzing the blank; Investigating the source if the problem to implement corrective action for the future; Redigesting and reanalyzing the associated samples if the analyte concentration in the samples is less than 5 times the blank concentration and greater than the reporting limit. Data that cannot be regenerated acceptably is flagged as non-compliant. When target analytes are detected in a method blank, the following data condition applies: When any target analyte is detected in a method blank above the action levels listed earlier but not in associated samples, then no flag is applied. Laboratory Control Samples The % recovery of each spiked analyte in the Laboratory Control Sample is calculated. Corrective action is taken whenever recovery is outside the acceptance criteria. The following corrective action is taken when required as stated above: • Check calculations to assure there are no errors; Check standard and spiking standard solutions for degradation, contamination, etc., check instrument performance; Redigest and reanalyze samples if there is no indication of failure in any of the above; • If that does not correct the problem, then the data is reported and a qualifying statement regarding the laboratory control sample is included in the report narrative. For Matrix Spike, Matrix Spike Duplicates and Sample Duplicates, if all QC associated with a sample is within acceptance limits (method blank and LCS spike recoveries), the problem may be attributed to a matrix effect. Samples I I D I I I I I I I I I I I I I I I I D m I I I I I I I I I I I I I I I I I 15.3 Date: 2/28/97 Section 15 Revision 0.02 Page 11 of 13 exhibiting a matrix effect will be qualified and discussed in the report narrative as appropriate. An exception to this criteria is allowed for matrix spike samples when the sample concentration exceeds the spike concentration by a factor or 4 or more. In that instance, the data is reported unqualified. Calibration For an initial and continuing instrument calibration to be valid, the responses for each analyte must be linear over the calibration range and the accuracy of calibration, as determined by analysis of an independent check standard, must be within ±10% of the true value for ICP analysis and within ±20% for cold vapor and graphite furnace AA analyses. If these criteria are not met, then the following corrective actions taken: Check standard solutions for degradation, contamination, etc., Check instrument for contamination, incorrect operating conditions, etc.; If no source of the problem is identified, then a complete instrument calibration must be performed. OUT-OF-STATISTICAL-CONTROL BLANK SPIKE CONTROL CHART DATA In accordance with the requirements of certain federal programs (Navy's NFESC [formerly NEESA]. HAZWRAP), blank spike control charts are maintained for all analyses performed for that program. Control chart data consist of recovery values derived from blank spike (LCS) data. Each control chart has five lines representing a statistical analysis of a set of percent recoveries from previously analyzed LCS samples: 1. 2. 3. 4. 5. arithmetic mean upper warning limit (+2 standard deviations from the mean) lower warning limit (-2 standard deviations from the mean) upper control limit (+3 standard deviations from the mean) lower control limit (-3 standard deviations from the mean) Plotted blank spike recovery data are evaluated against these limits to monitor for out-of- control and out-of-statistical-control conditions. 15.3.1 Out-of-Control Blank Spike Recovery Pata For a specific parameter and matrix, the laboratory process is considered out of control if any one blank spike recovery value is outside the control limits on the respective blank spike control chart. Except where LCS control limits are defined by the client, excursions of the blank spike control limits equate to LCS recoverie_s falling out of control. :\lqapreva\sect15.doc Date: 2/28/97 Section 15 Revision 0.02 Page12of13 15.3.2 Out-of-Statistical-Control Conditions For a specific parameter and matrix, the laboratory process is considered out of statistical control whenever one or more of the conditions described below is demonstrated by control chart monitoring: 1. Any three consecutive points are outside the warning limits 2. Any eight consecutive points are on the same side of the centerline 3. Any six consecutive points are such that the value of each is larger ( or smaller) than its immediate predecessor 4. Any obvious cyclic pattern is seen in the points The blank spike control charts serve as a mechanism both to note excursions from the prescribed control limits and for recognizing trends that may represent a degradation of an analytical system's quality control. They are valuable as an early warning indicator that corrective action is needed to prevent more serious loss of quality control. 15.3.3 Corrective Action for Out-of-Statistical-Control Conditions In the course of plotting blank spike recovery data, if a condition is observed that meets one or more of the criteria described above as an out-of-statistical-control condition, corrective action must be taken. An NCM must be initiated that identifies the trend causing the out-of-statistical-control condition. Where known, the NCM should describe the root cause of the trend or excursion and the actions taken to prevent recurrence. For some statistical trends, such as eight or more points on one side of the mean, the initial required corrective action may be no more than to continue monitoring future blank spike recovery d_ata. If, however, the out-of-statistical- control condition persists, or if the condition recurs repeatedly, further corrective action, such as instrument maintenance or recalibration, must be performed before sample analysis may resume. 15.4 UNUSUAL OCCURRENCES Whereas out-of-control events involve occurrences outside of pre-established acceptance windows, unusual occurrences are more subjective and involve incidents which may be compliant with the assessment criteria but still warrant investigation. Many of these investigations will be the result of the professional judgement of the analyst, auditor or data reviewer that the analysis was not typical or reasonable. Another example of this type of investigation is an inquiry or questioning of data received from a client or from the results of performance evaluation samples. :\lqapreva\sect15.doc I R I I I I I I I I I I I I I I I I I D I I I I I I I I I I I I I I I I I FIGURE 15.1 Example PACE CORRECTIVE ACTION REPORT Revision 2 -3/94 Date: 2/28/97 Section 15 Revision 0.02 Page13of13 CLP_ NEESA_ SW846_ NPDES_ Drinking water_ AFCEE_ ACOE_ Sample ID Number(s) Involved and QC batch#.: _____________________ _ Type of Event: ·· Out-of Control Event ·· Attachments ·· Unusual Occurrence Disposition of Samples: 1) Description of Event: .. 1. LCS Failure Contamination ·· 2. Blank Contamination Failure ·· 3. Poor Precision Reprepped ·· Reanalyzed ·· Narrated ·· 4. Prep Error ·· 7. Detection Limit .. 5. Hold Time ·· 8. Calibration Failure ·· 6. Login Error ·· 9. Retention Time Window ·· 10. Linearity ·· 11. Sample ·· 12. Surrogate ··13_ Malrix Spike Failure ··14.Other (Describe Below) 2) Discussion of Known or Suspecled Cause.: ________________________ _ 3) Corrective Action(s) Taken (include date, person and action): __________________ _ 4) Return to Control:--------------------------------- Initial and Date below please 1) Reported by_~--2) Corrective Action Taken By· .. __ _,__ 3) Corrective Action Approved. __ ~ 4) Documented Return to Control 5) Supervisor, __ .L__ 6) QA __ / :\lqapreva\sect15.doc I I I I I I I I I I I I I I I I I Date: 12/22/95 Section 16 Revision 0.01 Page 1 of 2 16.0 QUALITY ASSURANCE REPORTS TO MANAGEMENT The objective of the Pace quality assurance program is to ensure that an operational system is in place which enables management to determine the quality of all data produced within the laboratory system. An essential component of the system is the communication pathways and feedback mechanisms which ensure that management obtains quality information promptly and consistently. To achieve·this objective, Pace employs informal and formal reporting processes to ensure that management is informed of operational quality. This information enables Pace to take corrective action promptly when required. Reporting occurs at the following frequency. Daily meetings to discuss possible quality assurance problems and proposed solutions. Weekly meeting with upper management to discuss laboratory performance, upcoming audits, certification programs, and past audit performances. Quarterly written status reports to ·upper management; inclusion of all qualify assurance concerns and pertinent laboratory issues. As required, internal departmental audit reports with observations and suggested corrective action procedures. The Quality Assurance Officer and Quality Assurance Auditor are responsible for preparing reports to management indicating effectiveness of the laboratory Quality Assurance Program. 16.1 QUALllY ASSURANCE AUDITOR Results of audits perfonmed by the QA staff are detailed in fonmal, written audit reports. These reports are distributed to the audited personnel, section supervisor, Laboratory Operations Manager, QA Officer, and General Manager for review and appropriate action. These and other QA-related reports are distributed as produced, with no set scl1edule. Auditor reports will include, but not be limited to: • • • Results of internal laboratory review activities Results of internal data review activities Results of Proficiency Evaluation studies Results of state certification applications Summary of holding time exceedence and data qualification Method detection limit study status To demonstrate management review, the audit report will contain a page which will be signed and dated by the QA Officer and General Manager acknowledging that they have received the report and have reviewed its contents, and taken the necessary action dictated by their position. :ILOAPREVA\SECT15.DOC 16.2 QUALITY ASSURANCE OFFICER Date: 12/22/95 Section 16 Revision 0.01 Page 2 of 2 I I I The Quality Assurance Officer will issue a report of QA activities and findings on a regular I basis to the General Manager. The status report will include: Results of internal systems or performance audits I Corrective Action recommendations Discussion of QA issues raised by laboratory users Results of third party or external audits I Status of laboratory certifications Other significant events Perfonmance Evaluation Sample Results I 16.3 MANAGEMENT REVIEW OF THE QUALITY ASSURANCE PROGRAM Review of the appropriateness and adequacy of the Quality Assurance Program is ongoing. At anytime, any laboratory employee, through the Laboratory Operations Manager, may present recommended changes to the Quality Assurance Officer. During system audits, the Quality Assurance Program should be discussed. The audif report will document recommendations made by either the Laboratory Operations Manager or the auditor for revision. 16.4 QUARTERLY QUALITY REPORTS TO MANAGEMENT Quarterly reports are provided by the Quality Assurance Office staff to the Corporate Quality Office and the General Manager. The report summarizes quality assurance activities including details of corrective actions recommended or implemented, internal and external audit results, status of performance evaluation samples, certification status, and the status of internal procedure (evidentiary and technical) documents. :\LQAPREVA\SECT15.DOC I I I I I I I I I I I I I I I I I I I I I I I I I I I I I 17.0 SUMMARY OF REVISIONS Date: 12/22/95 Section 17. 0 Revision 0.01 Page 1 of 1 On November 10, 1995, a business transaction was consummated which created a new company, Pace Analytical Services. Inc., consisting of seven laboratories of PACE Incorporated. At the time of the transaction, in order to ensure an efficient transition to forming the new company, quality systems (e.g., QA plan, SOPs, etc.) which were previously in place at each of the laboratories remained intact. This document describes a new consolidated quality assurance program including (as appropriate) key elements of each facility's previous quality control practices. As such, this document has been designated as Revision 0.01, representing the first revision of the former QA plan of PACE Incorporated. 17.1 REVISION DESIGNATION Revision numbers are designated by an integer followed by two decimal places. At a minimum, this document will be reviewed in its entirety on an annual basis. Annual, document-wide review/revisions are tracked by incrementing the integer by one (e.g., Rev. 0.00 issued 5/22/95 will be revised no later than 5/22/96 and designated as Rev.· 1.00). The decimal places are used to track individual section or page revisions which occur in the interim between annual revisions. For example, when a section is initially revised independently of any other, the revision number indicated in the header will be listed as _.01 while the unrevised sections would remain at _.00. Subsequent revisions of the same section or page would incrementally increase the decimal designation (e.g., 0.02, 0.03, 0.04, etc.) The cover page revision number will always reflect the total number of individual revisions performed between annual revisions. At the time of the annual revision, all individual section and/or page revision numbers are returned to an initial _.00 designation, which initiates a new overall revision number. 17.2 SUMMARY OF REVISIONS This section will provide a historical chronology of all future revisions. The listing will contain the date of the revision, the section or page revised and the new revision number. :ILQAPREVAISECT17.DOC D I I I I I I I I I I I I I I I ,1 I I I Pace Analytical Services, Inc. Minnesota Laboratory Method Detection Limits, Pace Reporting Limits and Quality Control Limits D m I I I I I I I I I I I I I I I PACE Analytical Services, Inc. Minnesota Laboratory Method. 160.1 160.2 160.3 180.1 300.0 300.0 300.0 300.0 310.1 325.2 340.2 350.3 351.4 353.2 353.2 353.2 354.1 365.2 365.2 375.4 376.1 410.2 410.4 413.1 5310 . 5210B SM3500-Cr D 200.7 General Chemistry MDLs and PRLs WATER Component MDL (mg/L) Total Dissolved Solids NA Total Suspended Solids NA Total Solids NA Turbidity NA Fluoride 0.013 Chloride 0.011 Bromide 0.035 Sulfate 0.028 Alkalinity 1.5 Chloride 0.36 Fluoride 0.0066 Ammonia Nitrogen 0.056 Total Kjeldahl Nitrogen 0.057 Nitrate as Nitrogen 0.035 Nitrite as Nitrogen 0.035 Nitrate + Nitrite as Nitrogen 0.035 Nitrite as Nitrogen 0.035 Phosphorus 0.012 Orthophosphate-P 0.0073 Sulfate 1.1 Sulfide 0.1 Chemical Oxygen Demand 1.5 Chemical Oxygen Demand 4.1 Oil & Grease 2.1 Total Organic Carbon 0.28 BOD 5-Day NA Hexavalent Chromium 0.0013 Hardness (sum Mg & Ca C03) NA •·•···•·•/Method) •/· C::omponent''.,r· ........ -··. :-:::,.-:;/ ., ... ··.·-:.,,,.·.··:·:·.::·,:<',··:-:=.;·::r:::, I<.••·MDL••(ug/L)f 335.2 Cyanide 2.1 420.1 Phenol, Total 7.3 Page 1 ·.:, PRL (ing/L) 10 10 10 0.5 0.1 1 1 1 5 1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.05 0.02 5 10 10 50 5 1 6 0.01 1 §\ if ~i:(1igi()i< 10 50 3/96 Pace Analytical Services, Inc. Minnesota Laboratory 1::/i Method No. 160.1 160.2 ,160:3 160.4 180.1 300.0 335.2 420.1 310.1 325.2 340.2 350.2 351.3 353.2 365.2 365.2 370.1 375.4 · 376.2 410.2 410.4 413.1 418.1 425.1 5210B 5210B 5310 Sum of Ca & Mg as CO3 ... ,,· · ..... •<:\::.·· Analyte/ Component Filterable Residue Nonfilterable Residue Total Residue· Volatile Residue Turbidity Bromide Chloride Fluoride Sulfate Cyanide Phenols, Total Alkalinity Chloride Fluoride Ammonia Nitrogen Total Kjeldahl Nitrogen Nitrate+Nitrite Nitrogen Nitrate Nitrogen Nitrite Nitrogen Phosphorus Orthophosphate-P Silica Sulfate Sulfide COD COD Oil and Grease TRPH MBAS BOD 5-Day CBOD 5-Day TOC Total Hardness 21nterim limits until statistically derived limits determined ------- General Chemistry Quality Control Limits · ... ;Ac_curacy:,,. precision-'.' MS/MSD Recov. (%) MS/MSD RPD (¾) Water Solid Water Solid NA NA 20' NA NA NA 20' NA NA NA 20' NA NA NA 20' NA NA NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125! NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' NA 20" NA 75-125' NA 20' NA 75-125' NA 20' NA 75-1252 NA 20' NA 75-1252 NA 20' NA 75-125' NA 20' NA 75-125' NA 20' NA 75-125' 75-125' 20' 20' 75-1252 NA 20' NA 75-125' NA 20 1 NA 75-125' NA 20 2 NA 75-1252 NA 20 2 75-125' NA 20' NA Page 1 ---- I : . Accuracy Precision LCS Recovery('%) LCS/LCSD RPD (%) Water Solid Water Solid 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20" NA 80-1202 NA 20' NA 80-120' NA 20-, NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20..., NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20' NA 80-120' NA 20 2 NA 80-120' NA 20 2 NA 80-120' NA 20 2 NA 80-120' NA 20 2 NA 80-120' NA 20 2 NA 80-120' NA 20' NA 80-120' NA 20 2 NA 80-120' NA 20 2 NA 80-120' 80-120' 20 2 20 2 80-120' NA 20 2 NA 80-1202 NA 20 2 NA 80-1202 NA 20 2 NA 80-1202 84-119 20..., 17 80-120' NA 20' NA Rev.2 8/96 -----liiillll ------------------~ Pace Analytical Services, Inc. Minnesota Laboratory ICP-AES Water Method , Analyte i•'• ·.·,. < \ MDL (ug/L) 6010 Aluminum 18 Antimony 21 Arsenic 79 Barium 3.1 Beryllium 1.2 Boron 17 Cadmium 3.2 Calcium 83 Chromium 4.7 Cobalt 5.3 Copper 2.5 Iron 11 Lead 37 Magnesium 21 Manganese 2.2 Molybdenum 10 Nickel 15 Potassium 420 Selenium 93 Silver 2.3 Sodium 69 Thallium 110 Vanadium 4.9 Zinc 5.1 Metals MDLs and PRLs PRL (ug/L) ··Method· 50 6010 50 90 5.0 5.0 100 5.0 200 5.0 . Method 10 7041 5.0 7060 25 7091 40 7131 30 7191 5.0 7201 20 7211 20 7421 2000 249.2 100 7740 5.0 7761 100 7841 220 10 20 IV!ethod 7470 Page 1 ICP-AES SuperTrace Water Arialyte · MDL (ug/L) · PRL (ug/L) Antimony 2.7 5.0 Arsenic 3.7 5.0 Cadmium 0.36 0.50 Lead 1.4 3.0 Selenium 2.8 5.0 Thallium 1.5 5.0 GFAA Water Analyte MDL (ug/L) PRL (ug/L) Antimony 2.4 5.0 Arsenic 2.1 5.0 Beryllium 0.058 0.20 Cadmium 0.042 0.10 Chromium 1.0 1.0 Cobalt 1.0 1.0 Copper 1.1 2.0 Lead 0.92 3.0 Nickel 2.5 5.0 Selenium 1.3 5.0 Silver 0.45 0.50 Thallium 2.8 5.0 CVAAWATER Analyte MDL (ug/L) · PRL (ug/L) Mercury 0.030 0.20 Rev.2 8/96 I I I I I I I I I I I I I I 'I I I I I APPENDIXB Standard Operating Procedures GEi Consultants, Inc. I I .•<>--. I I I I I I I I (_ I 'I I I I I I 1 · . "-. ,' ·..,,.~·. I SOP No. RE-OCH Page 1 of 3 Objective FIELD NOTEBOOK Revision No. 0 Date: 2s6-95 The field notebook is intended to serve as a record of significant field activities performed or observed by GEL The field notebook will serve as a factual basis for preparing field observation reports. if required, and reports to clients and regulatory agencies .. Procedure l. Use a separate bound notebook for each site/location/project number. 2. Write neatly using black or blue ink (or note if field conditions (i.e., cold or wet weather require use of pencil). 3. Write the project name, project number, and book number (i.e., I of 3) on the front cover. On the inside cover, identify the project name, project number, and "Return Book to" GEI's address. 4. Number all of the pages of the field book starting with the first entry. 5. Record activities as they occur. 6. Neatly cross out mistakes using a single line and initial them. Erasures are not permitted. 7. Sign or initial and date the bottom of every page with an entry. Cross out unused portions of a page. 8. Record the following information upon each arrival at the site: a) Date/time/weather/project number b) GEi personnel c) Purpose of visit/daily objectives 9. Record conversations with: [Reco=endation -If possible, record telephone numbers of individual contacts for the site in the field notebook.} a) Contractors b) Clients c) Visitors (include complete names, title, affiliations, whenever possible). d) GEi office staff SOP No. RE-0()1 Page 2 of 3 Landowners (site or abutters) Revision No. 0 Date: 2-6-95 e) f) Note time of arrival and departure of individuals visiting the site. I D I 10. Examples of the field information to be recorded includes and time of occurrences. I a) b) c) d) e) f) g) h) i) j) General site work activities Subcontractor's progress Type and quantity of monitoring well construction materials used Use of field data sheets (i.e., boring logs, monitoring well sampling logs, etc.) Ambient air monitoring data Locations of sampling points Surveying data (including sketches with north arrows) Changes in weather Rationale for critical field decisions Recommendations made to the client representative and GEI PM. 11. Record the following information upon departure: a) Include a site sketch of conditions at the end of the day. b) Time ·c) Summarize work completed/work remaining d) Place a diagonal line though and sign portions of pages not used or skipped. Precautions • Only record facts. Do not record opinions. • Do not fail to record an observation because it does not appear to be relevant at that time. • Identify conditions or events which could effect[rmpede your ability to observe conditions. • Do not use spiral notebooks because pages can be easily removed. References 1. ASFE Model Daily Field Report (1991), ASFE, Inc. Attachment Example Field Notebook I I I I I I I I I I I I I I I ---------- --.k~:}£-ii Ye . . L• ~pl~-!--<' \'Ylc;,11 t,., '· () ~' , I. !l:':i;y1f''"".,·1 ""'.J c[:1,,..,,i cJ 01 j'),,/i<,:. _Q (lt1(r/ c,700 0 ?(iO i:0:1 Lr.:·\\:. d ,! • ",1 "( l. cl Bl i.=., . .. . __ .. _ . . _ . Len f, 11 ! 1 r J Cc ! • l h, 5 , d, I Sq p. /.,;., ___ , e~~'l ,5. frt· ~''.>"''.':i c::!. . .j'' :Sf2/.:J- _________ .S[?frn ~•rnr/ r. -·----· ........ ,;:<,"('le --H_,...1_ ~.+-e~l--_-_·-y_f:!)__ µ ! . i '" I J l r I:· P' :;! :, .,I .) I I I I I I I I I I I I I I I I I - I SOP No. RE-002 Page 1 of 6 Objective FIELD OBSERVATION REPORT Revision No. 0 Date: 2-6-95 To accurately summarize the activities, observations, and decisions made during the day's field work. The daily field observation report may serve as a permanent record of the day's activity for the Project Manager, IHC, and client. Procedures At the close of the day's field work, a Field Observation Report must be prepared by the individual responsible for the field notebook. This report must be completed before leaving work for the day. Contents of the report should include, at a minimum, the following information: 1. A record of person(s) present at the site, time of arrival, departure times (e.g., GEi, contractor(s), client, etc.). 2. A record of the daily objective(s) and the activities performed (e.g., drilled five borings in the overburden). 3. A summary of deviation(s) from the field plan or objectives. 4. A summary of field decision(s) made, who made it/them, and the basis for such decision(s). 5. A diagram, sketch, and/or map showing the location and extent of the work or other significant observation(s) made during the day. 6. Any recommendations that may result from field observations and any actions that resulted from those recommendations 7. A summary listing and field sketch showing location(s) of field activity 8. Submit a draft report to the PM/IHC for review and editing related to the clarity and conciseness of the report. Complete any editorial changes, sign, date, and submit the report to PM/IHC for approval/signature. Field Observations Reports should be written neatly. They need not be typed unless specifically requested by the PM. SOP No. RE-002 Page 2 of 6 Precautions: • Not all projects require daily field observation reports. Revision No. 0 Date: 2-6-95 • The Field Observation Report should be based sole! y upon factual information not opinions. Any speculation should be clearly noted in the report as such. The Field Observation Report should never be released to anyone other than the PM/IHC prior to review and sign-off unless explicitly authorized by the PM/IHC. References I. GEI Technical Manual, dated July, 1987 Attachment Example Field Observation Report. I R I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I ,I I I I FIELD OBSERVATION REPORT Date 0] /J 819? PROJECT ABC Industries Buildinq Demolition Report No. 8 CLIENT Checkmore Development Project No. 99709 CONTRACTOR Demolition Brothers Page _-±-of . TIME OF ARRIVAL '"" " DEPARTURE 1 ISA IS WEATHER C:••-----~ ~" l:' PERSONS CONTACTEDlAFFILIATION GEi REPRESENTATIVES Reporter: Gerry E. Inline Tom Tuttle/ Demolition Inc. Joe Robinson/ Checkmore Development Gary Bark/ Checkmore Development Fran Garman I Checkmore Development OBSERVATIONS 1. Pur12ose of Site Visit: The purpose of the site visit was to observe the excavation of No. 2 fuel oil contaminated soil from beneath and adjacent to the southwest corner of the former ABC Industries (ABC) building and to collect confirmatory soil samples from the bottom and sidewalls of the excavation for DEF-certified laboratory testing. on March 5, 1991, a release of No. 2 fuel oil was observed upon removal of an underground storage tank (UST) for the ABC. contaminated soil beneath and adjacent to southwest corner of the ABC was left in place during the initial clean up in the spring of 1991. The ABC was demolished during December 1991 and January 1992 to allow for the construction of Parking Deck D5. The figure on page 5 shows the location of the former UST, the ABC building and the limits of contaminated soil removed in the spring of 1993. 2 • Demolition of ABC: Wreckem Inc. demolished and removed the majority of the ABC building before I arrived on site. The basement wall and footing in the southwest corner of the building (in the vicinity of the contaminated soil) was left in place. 3 • Removal of Conci;:ete Floor Slab: Demolition Inc. removed the floor slab in the southwest corner of the ABC building. Concrete which was oil on plastic. stained was stockpiled q) GEI Consultants. Inc. By-J:/ <a:::,,, £ · 1,::2l&£ ·M App'd , FIELD OBSERVATION REPORT Date Ql/78(92 PROJECT ABC Building Demolition Report No. R __.,__ __ _ Project No. 93295 CONTRACTOR Demolition Brothers, Inc. Page 2 of 4. Excavation and Stockpiling of Soil: Demo!ition Brocners excavated soil with a Komatsu excavator and stockpiled the soil on site, north of the Mall building adjacent to Thurber Construction's field trailer. I collected samples of the excavated soil for jar headspace screening for volatile organic compounds (VOCs). I screened the samples using Massachusetts Department of Environmental Procection's (DEF) Jar Headspace Method (DEF Policy #WSC- 400-89, Attachment I) using an organic vapor meter (OVM). Demolition Bros. stockpiled the excavated soil on plastic in three separate stockpiles based on the following criteria: Stockpile 1: Stockpile 2: Stockpile 3: voe levels greater than 100 parts per million (ppm) and/or visible oil staining. voe levels between 50 and 100 ppm without visible oil staining. voe levels less than 50 ppm without visible oil staining. The table on Page 3 shows the results of the jar headspace screening. The figure on Page 6 shows the excavation limits and the soil sample locations. 4 ~l===============s"'=========================I I D I I I I I I I I I I I I I I I I iL~Q)~~G~E~I~C~o~ru~u~lra~n~~=-~ln~c~.-LB~d:ft.~~-~W½~"~{=~='~===~A=pp~•ct~======~·~· E I FIELD OBSERVATION REPORT Date 0l('.18('.92 I PROJECT ABC Industries Building Demolition Report No. 8 CLIENT Checkmore Development Company Project No. 99709 I CONTRACTOR Demolition Brothers, Inc. . Page 3 of 4 I I Sample OVM Location Sample Number Reading2 co=ents Number1 (ppm) l 99709-14N-45E-1613 0 I 2 99709-0N-29E-160.5 19 3 99709-4N-35E-159 352 I 4 99709-3S-35E-157 475 5 99709-l0N-27E-158 406 I 6 99709-3N-34E-156.5 16 at groundwater table 7 99709-15N-31E-158 91 I 8 99709-13N-36E-156.5 28 at groundwater table 9 99709-12N-40E-158 389 I 10 99709-3S-55E-161 86 adjacent to concrete floor slab 11 99709-10N-35E-157 67 I 12 99709-10N-35E-156.5 37 at groundwater table Tested for TPlf' I Notes: l. Sample locations shown on page 4. 2. DEP Jar Headspace Method #WSC-400-89, Attachment I. I 3. Sample No. 99709-14N-45E-161 is 24 ft. north and 55 ft. east of the center of Parking Deck D footing Dl6/17 at elevation 161. 4. Sample delivered to a DEF-certified laboratory for I TPH testing. I I ::l I ~ ::, ~ ~ -I "' ~ e ~ :!. cp GEl Consultants, Inc. B~ f'. ,(~,M..,App'd ! m FIELD OBSERVATION REPORT ABC Industries Building Demolition PROJECT __________________ _ Checkmore Development company CLIENT ___________________ _ CONTRACTOR Demoltion Brothers, Inc. r 'I '5 L1 :1 I >I I 3 " :, ,. ~ -4 r '-I I· ' ( 7 •• , , .,,. ,, ✓ .,,. ., \ "\ f,S-Z.5E 1 .J ? ' ' i c; c.,,1..c: ,J \ \ . , ,· , , , , WALL. ~\MIT s o,o ~.<£.AJAi'I:) N . '"'"\,\\1 s ) IO fwo g_ SLA~ Q_i M<lVAL • :..~c.An~...i <Jf S..:,,L SAr'W 1..-:; ,-.x:,. I. Date 01/18/94 Report No. 8 Project No. 99709 Page 4 of I ,, ., '-" , -=-::c., @ L.-lCAT1J...i ur :..:Ni=-,V-.--'°1.-..,'1..-1 .S:.:,le, SA"'-'\..:. ~u. l2.. • 4 I n I I I I I I I I I I I I I I I I :6=====~=========91 e cb r-rt r-___ .,JM_., l~r avr/:::to,,,,..,.,, l~ ),J.,,,,.,,_App'd I I I I I I I I I I I I I I I I I I I SOP No. SS-005 Page 1 of 2 Revision No. 0 Date: 2-6-95 MONITORING WELL DEVELOPMENT Objective The objective of this SOP is to standardize the development of monitoring wells for environmental investigations. The purposes of developing a monitoring well are to remove fluids introduced during drilling and to maximize the movement of water into the well by reducing the fines which may be trapped in the sand pack around the screen, in order to reduce siltation during sampling. Procedure 1. Decontaminate all development equipment prior to use (see Equipment Decontamination SOP). 2. 3. 4. 5. 6. 7. Calculate or estimate the amount of water introduced to the borehole during drilling. At a minimum, this is the amount of water that must be removed during development In addition, compute the volume of water in the monitoring well. Place a 12-volt submersible pump ( or grundfos pump) attached to a power source into the borehole or use a manual sampling device such as a bailer or wate,rra pump. Collect a sample of the standing water in the borehole and record the physical properties (color, turbidity, etc.). Then, at a mioim11m, remove the greater of the following two amounts of ground water: a) ten well volumes b) the amount of water introduced during drilling Pump the ground water into a 5-gallon pail so that the volumetric flow rate and water volume from the pump or bailer can be calculated. Monitor the ground water level in the well as the water is being pumped to determine if the pumping rate is sufficient to create a drawdown in the well. The "over- pumping" development method requires that the well be developed/stressed at a faster rate than the well would normally be pumped or bailed for sample collection. Collect ground water samples every few well volumes during the pumping and record the physical properties (color, turbidity, etc.) (see respective SOPs). Once half the desired volume of water has been pumped, request that the drillers attach surge blocks on a rod into the well. This surge block apparatus may be operated by hand or be attached to the drill stem on the drill rig and is operated by the driller. Slowly move the surge block up and down in the upper portion of the well. Start at a slow pace and progress to a faster surging action further clown the well screen. SOP No. SS-005 Page 2 of 2 Revision No. 0 Date: 2-6-95 n u 8. During surging, remove the surge block periodically to pump more water from the I well to remove accumulated fines. 9. Monitor the turbidity and color of the water during this procedure. The well is considered fully-developed when all of the following criteria have been met: a) the volume of water introduced during drilling has been removed b) the water removed from the well is relatively free of fine-grained materials. 10. Record the final amount of water removed and the physical properties (color, turbidity) of the well water. Precautions At all times, follow safety procedures as defined in the site-specific Health & Safety Plan. I I I I I • Always remove ground water with fines from the well before surging, as these fines 1 may be forced into the well screen otherwise by the surging action. • If the ground water in the monitoring well is contaminated, the water removed during well development will need to placed in a properly-labelled drum (see IDW SOP). I • If the soils around the well screen are composed of fine-grained silts. and clays, over- pumping and mechanical surging is not reco=ended since these more vigorous techniques can cause mixing of the fines into the filter pack. To develop these wells, use of a bailer is recommended. References 1. Standard Practice for Design and Installation of Ground Water Monitoring Wells in Aquifers (October 1990), American Society for Testing and Materials [ASTM] D5092- 90. 2. Nielsen, D.M. (1993), "Correct Well Design Improves Monitoring," Environmental Protection, July. pp. 38-49. 3. "The Methods & Mechanics of Well Development, Part 2 of 5," National Drillers Buyers Guide, March 1993, p. 17. 4. Standard References for Monitoring Wells (April 1991), Commonwealth of Massachusetts Department of Environmental Protection, WSC-310-91. I I I I I I I I I I I I I I I I I I I I I I I I I I I SOP No. SA-002 Page 1 of 4 Objective GROUND WATER SAMPLING Revision No. 0 Date: 2-6-95 To outline a method to collect ground water samples which accurately and precisely represent the aquifer conditions. Procedures 1. Record all activities in the field notebook (see Field Notebook SOP) and on the Ground Water Sampling Form (attached). Use a separate form for each sampling location and evenL 2. Calibrate PID, pH/temperature, and specific conductance (SC) meters (see respective SOPs). 3. Purge Well a) Unlock well cap and, if required by the Health & Safety Plan. measure organic vapor concentration in the well pipe (see H&S Monitoring for VOCs SOP). b) Measure (water level probe and/or oil water interface probe) depth to water and overall well depth to the nearest 0.01 foot from the top of casing and the highest point (or "V" notch) on the PVC. If the top of casing cannot be used, note the reference location. Mark the datum point with an indelible marker and note reference location in field book. c) Calculate one well volume. V = 7t r2 h or use conversions on ground water sampling form. d) Purge well to remove stagnant water from sampling zone. Generally, water should be drawn down from the upper portion of the water column in high formations and the bottom of the water column in low yield formations. Record the volume purged. e) Monitor and record the ground water parameters (pH, temperature, and SC) from the first bailer and after each well volume (see SOPs). f) Purging is complete when one of the following has been achieved: i) A minimmn of three well volumes have been purged or on the reference ground water parameter has stabilized. (±0.2 for pH, ± 0.1 for temp., ±<).1 for SC); or SOP No. SA-002 Page 2 of 4 ii) The well has been pumped/purged dry. g) Dispose of purge water according to the field plan. Revision No. 0 Date: 2-6-95 h) Collect samples as soon as the volume of water is sufficient to fill the intended sample containers. 4. Collect Samples a) Begin sampling at the least contamioatP.d well and work toward the most contaminated well. b) To achieve comparable results, use a similar method to collect all samples whenever possible. c) Lower sampling device slowly into ground water to the middle of the screened interval or water depth. Determine sampling location based upon project specifics. d) Fill sample containers directly from the sampling device in order of decreasing volatility (see Sampling Handling SOP). e) Remove sampling device and decontamioatP. (see Equipment Decontamination SOP). f) Store samples in cooler between 2°C and 6°C for transport to the laboratory. g) Secure well cap. Precautions • At all times, follow safety procedures as defined in the site-specific Health & Safety Plan. • Prior to departure for the field, obtain available on information on well construction for use in field investigation (i.e., screen and riser material, well diameter and depth, and screened interval). • When purging a well with a pump, the pump may require lowering to the middle of the screened interval to obtain a reasonable flow on the basis of operational requirements of the pump. If possible, the pumping rate should match the natural aquifer yield rate. • Measure ground water elevation twice. I n I I I I I I I I I I I I I I I I I I I I I I I I I •• I I I I I I I I I I SOP No. SA-002 Page 3 of 4 Revision No. 0 Date: 2-6-95 • Be aware of any preservatives in the sample bottles and handle with care, in accordance with the Health & Safety Plan. References 1. Characterization of Hazardous Waste Sites -A Methods Manual: Volume II (December 1984). U.S. EPA (EPA-600/4-84-076). 2. Draft Ground Water Monitoring Guidance for Owners and Operators of Interim Status Facilities (1983), U.S. EPA. 3. Procedures Manual for Groundwater Monitoring at Solid Waste Disposal Facilities (December 1980), U.S. EPA, (EPA-530/SW-611). 4. Test Methods for Evaluating Solid Waste (1986). U.S. EPA (SW-846). 5. Barcelona, Michael J.; Gibb, James. P.; and Miller, Robin A., A Guide to the Selection of Materials for Monitoring Well Construction and Ground Water Sampling (August 1983), Illinois State Water Survey Contract Report (ISWS) #327 (EPA Contract No. EPA CR-809966-01). 6. Standard Reference for Monitoring Wells (April 19, 1991), Massachusetts DEP, DEP Publication #WSC-310-91. 7. Groundwater Sampling Procedures Guidelines (1987), Wisconsin Department of Natural Resources (WR-153-87). ;-J\ Project I Date Well ID I Start nme _.-:: Weather I I End nme ' Sampling Personnel ' Well Construction Water Level Well Volume Well Diameter Well Depth 7 Conversion : Roadbox Water Depth . Diam. Factor • Top of Casing Water Height = linl lnaVftl Top of Screen Conversion Factor X 1.5 0.09 Bottom of Screen One Well Volume = 2 0.16 .· . Bottom of Well 4 0.65 Place 1 for 1 sl step, 2 for 2nd step, etc. Volume Purged I I 6 1.5 • Decontamination Calibration Dates Eauioment Used Pur□e Sample Ethyl Alcohol PID Teflon Bailer Deionized Water S.C. Meter PVC Bailer Alconox & DI H,0 pH/Temp. Submersible Pump Tap Water Other: Peristaltic Pump Solvent Tubing Other: Other: ~ Check the equipment which applies. : D -dedicated equipment. Field Analysis Data VOA Screening (ppm as Benzene): Ambient Air I I Well Mouth • Well Purae Data Volume Evacuated gal gal gal gal gal Temperature (C) : pH (std units) Specific Cond. (umho/cm) WellVOAs (ppm) EH (mV) Dissolved Oxygen (ppm) Other: ( ) ·• ·. Samples Collected for Laboratory Analysts ' Samnles Collected Sample Bottle ID VOA8240 VOA 8010 Time Sampled I I .-TPH ABN Color I PCB ALK/COD/BOD Turbidity I Metals TYee of filter membrane ' Other: Filtered in Field ? I ~es I no I I ; ! Additional Observations: i : I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I •• I I I I I I I I SOP No. SA-005 Page 1 of 2 Objective SAMPLE HANDLING Revision No. 0 Date: 2-6-95 Sampling Handling involves the collection and shipping of environmental samples to a laboratory for chemical analysis. The overall objective of Sample Handling is to ensure: • samples are properly labelled and documented; • samples are proper! y preserved; • samples are properly packaged; and • samples are proper! y transported to laboratories. Procedures l. Label all laboratory glassware prior to collecting samples. The label should have an adhesive and be placed on the jar or bottle, not on the cap. 2. Record the following information on the label and in the field notebook (See Field Notebook SOP): project number, sample identification (i.e., MW201 or SS-2), date and time of collection, samplers initials, and preservative, if present 3. At each sampling location, samples must be collected in order of volatility, most volatile first Samples collected for volatile analysis must be placed in sample containers immediate! y upon retrieval of the sample. 4. Aqueous samples for volatile analysis must be collected without air bubbles. Soil samples for volatile analysis should be compacted to eliminate as much heads pace as possible. Other laboratory glassware should also be filled when possible. 5. If compositing of samples is performed in the field, specify basis for composite (i.e., volume, weight, spoon recovery, etc.) and record procedure for composing sample in the field book. 6. Once samples have been collected, place samples in a cooler with ice or a blue pack and start the chain-of-custody (See Chain-of-Custody SOP). See comment No. 1. 7. For shipping, individually wrap each sample bottle with "bubble packing" or suitable packing material and place the wrapped bottles in the cooler with sufficient packing material between samples to avoid breakage. 8. Place a layer of packing material above and below the sample bottles. Place blue ice packs or ice bags on top of the packing material. Fill the remaining space in th,e cooler with packing material to eliminate the possibility of vertical movement of samples. SOP No. SA-005 Page 2 of2 Revision No. 0 Date: 2-6-95 9. Fill out the appropriate shipping or courier forms and attach to the top of the cooler. If necessary, place the proper shipping labels on the cooler. Place a custody seal on the cooler. Precautions • At all times, follow safety procedures as defined in the site-specific Health & Safety Plan. • Field personnel must be aware of analyses which have short holding times and schedule sampling events and shipping accordingly. Shipment of samples for analyses with short holding times must be planned in advance. • In general, glassware for aqueous samples contains preservatives, i.e., HN03 or HO. When collecting the sample, take care not to overfill the container, thus flushing the preservative out of the bottle. • Never composite samples for VOCs in the field. Collect individual aliquots and direct laboratory to perform compositing. • Collection of aqueous samples should not be performed over the opening of a monitoring well. Preservatives from overfilling, a marker pen or other objects could fall into the well. • If the recharge volume for a monitoring well is low, completely fill all volatile vials . and then collect the minimum sample volume required for each remaining analysis. • During subsurface soil sampling, if the recovery from the split-spoon sample is inadequate, if appropriate, resample the bottom of the borehole to obtain proper sample volume. • Laboratories will homogenize and test the contents of the sample container, unless directed otherwise. Samples should not contain rocks, twigs, leaves, etc. unless these materials are of interest References l. Manual of Ground-Water Quality Sampling Procedures (September 1981), U.S. EPA Office of Research and Development (EPA-600/2-81-160). 2. Soil Sampling Quality Assurance User's Guide {March 1984), U.S. EPA Office of Research and Development Environmental Monitoring Systems Laboratory, Cooperative Agreement CR 810550-01 (EPA-600/4-84-043). 3. Standard References for Monitoring Wells (January 1991), Massachusetts Department of Environmental Protection, DEP Publication# WSC-310-91. D I I I I I I I I I I I I I I I I I E I I I I I I I I I I I I I I I I .I I STANDARD OPERATING PROCEDURE -SAMPLE CUSTODY - PREPARED: March 15. 1991 REVISION: O ~ ~:.-::==- PREPARED BY~~~ 1_ , DATE: APPROVED BY: ~~..!(..,-,.,~ DATE: 1. PURPOSE Revision_0_ Date March 15, 1991 Due to the potential evidentiary nature of samples collected as part of environmental investigations at GEI Consultants. Inc., (GEI) possession must be traceable from the time of collection until disposal or archival of the samples. This SOP details the procedures and responsibilities for the GEi staff for maintaining and documenting custody of the samples from the time of collection until shipment to the laboratory or disposal. I D I I I I I I I I I I I I I I I I D I I I I I I I I I I I I I I I I .I -2- 2. POUCY Revision_Q_ Date March 15, 1991 It is GEi's policy that all samples collected as pan of an environmental investigation be maintained under custody. A sample is under custody if: 1. It is in your possession, or 2. It is in your view, after being in your possession, or 3. It was in your possession and then you locked it up to prevent tampering, or 4. It is in a designated secure area. -3- 3. PROCEDURE Revision.JL Date March 15, 1991 The custody procedures defined are for field and GEi storage activities until the time of sample shipment to a laboratory or disposal. 3.1 Field Custody Samples collected in the field are the responsibility of each sampler until the samples are transferred to a person designated as the field sample custodian. Custody, as defined above, must be maintained by the sampler. The sampler is personally responsible for: 1. Labeling each sample in waterproof ink; 2. Recording the information on the label in a field notebook; 3. Proper sample preservation and storage of the sample until transferring the sample to the field sample custodian; and 4. Maintaining sample custody. The field sample custodian, who may be the same person as the sampler, is personally responsible for: 1. Completion of the chain-of-custody form (Fig. l); 2. Checking and maintaining proper sample preservation and storage; 3. Maintaining sample custody; 4. Properly packaging the samples for transfer to the GEI facility or the laboratory for analysis; 5. Signing and dating the chain-of-custody records when the samples are relinquished and including the original chain-of-custody in the shipping container; 6. Maintaining a copy of the chain-of-custody form(s); 7. Sealing the shipping containers with chain-of-custody seals prior to shipment; and 8. Arranging shipment or delivery of the samples. I D I I I I I I I I I I I I I I I I I D D fl D m I I I I I I I I I I I I -4- 3.2 Custody at the GEi Facility Revision_O_ Date Marcin 15. 1991 Samples may be stored at GEI following field activities prior to shipment to a laboratory. In addition, samples for screening may also be delivered 10 the GEI laboratory. 3.2.1 Sample Storage Samples may be delivered to the GEI facility by the field sample custodian/sampler, courier service, or other delivery service. Sample receipt and storage at the GEI facility is controlled by a limited number of GEI employees, each of whom may be designated as facility sample custodian. Control is established by possession of keys that permit access to the sample s10rage refrigerawrs. The employees with keys are listed in Table 1. The field sample custodian, who may be the same person as the sampler, is personally responsible for: 1. Relinquishing the samples under chain-of-custody to one of the GEI employees listed on Table l; and 2. Signing and dating the chain-of-custody forms. In the event that none of the facility sample custodians are available at the time of sample delivery, the shipping container must be stored in a secure area with custody seals to prevent tampering until such time that a facility sample custodian can be notified. The facility sample custodian is personally responsible for: 1. Verifying that the information on the sample labels matches that on the chain-of-custody record; 2. Noting any discrepancies between the sample labels and the chain-of-custody on the chain-of-custody form; 3. Signing the chain-of-custody form, thus establishing cuswdy of the samples; 4. Logging the samples into the GEI sample refrigerator log (Fig. 2); and 5. Placing the samples in the refrigerator and locking the refrigerator. -5- Revision_Q_ Date March 15, 1991 In addi!icn, when samples are requested by another GEI employee for shipment to an outside laboratory or disposal, the facility sample custodian is personally responsible for: 1. Completing the GEI sample refrigerator log; 2. Having the person receiving the samples sign the log; and 3. Relocking the refrigerator. 3.2.2 GEi Screening GEI operates a sample screening program. If samples are delivered by the field sample custodian for sample screening, they should be delivered to the sample screening laboratory. The field sample custodian, who may be the same person as the sampler, is personally responsible for: 1. Relinquishing the samples under chain-of-custody to one of the GEi employees listed on Table 1; and 2. Signing and dating the chain-of-custody forms. In the event that none of the screening laboratory personnel are available at the time of sample delivery, the shipping container must be stored in a secure area with custody seals to prevent tampering until such time that the screening laborat0ry personnel can be notified. The screening laboratory personnel are personally responsible for: 1. Verifying that the information on the sample labels matches that on the chain-of-custody record; 2. Noting any discrepancies between the sample labels and the chain-of-custody on the chain-of-custody form; 3. Signing the chain-of-custody form, thus establishing custody of the samples; 4. Logging the samples into the GEI screening refrigerator log (Fig. 2); and 5. Placing the samples in the refrigerator and locking the refrigerator. I I I I I n I I I I I I I I I I I I I D fl B n D D D D D u 0 0 n D H I n n -6- 4. STORAGE AND DISPOSAL Revision_!)_ Date March 15. 1991 Samples will be stored in refrigerator's maintained at 4 • C (.±.2 C). The temperature will be recorded on a daily basis on all working days. These records will be maintained and available for review upon request. Samples will be stored for 30 days from sample receipt. At that time, the facility sample custodian will notify the project manager that the storage period has expired and will request direction for final sample disposal. Samples will be disposed of at the direction of the project manager and the disposal will be recorded in the GEi sample refrigerator log. TOTAL P.08 \ I I I I I I ,j I I 1 , I I I I I I I I I I I I APPENDIXC Standard Operating Procedure No. 7 Turner Hart & Hickman, P.C. I u D D n I I I I I I I I I I I I No. 7 HYDROGEN, GC/RGD, FIELD LABORATORY 1.0 SCOPE AND APPLICATION I. I This procedure is used to identify and determine the concentration of hydrogen in groundwater samples 2.0 SUMMARY OF METHOD 2.1 Samples are introduced into the GC using direct injection of the gas stripped from a continuous flow of ground water. Detection is achieved by a reducing gas detector (RGD). 3.0 INTERFERENCES 3. I Impurities in the gases, either those used in the GC system or other gases stripped from the ground water with the hydrogen, account for the majority of contamination problems. The analytical system must be demonstrated to be free from contamination under the conditions of the analysis by running laboratory reagent blanks. The use of non-TFE plastic coating, non-TFE thread sealants, or flow controllers with rubber components should be avoided. 3 .2 Samples can be contaminated by cross-contamination between sampling locations. Sampling equipment must be throughly decontaminated between locations. 4.0 APPARATUS AND MATERIALS 4.1 Gas chromatograph 4. I. I Gas Chromatograph -Analytical system complete with gas chromatograph suitable for gas sampling and/or on-column injections and all required accessori.es, including detectors, column supplies, recorder, gases, and syringes. A data system for measuring peak heights and/or peak areas is recommended. 4.1.2 Columns -the following column is recommended, however, alternative columns may be employed as long as 60"/o resolution is demonstrated for target analytes: J&W Scientific GS-Molesieve Capillary Column, 0.53 mm I.D. x 30 m. 4.1.3 Detector -Reducing gas detector (RGD). 4 .2 Syringes - A variety of Luerlok glass hypodermic syringes, gas-tight, with and without shutoff valves. 27 5.0 REAGENTS 5.1 Ultra-pure gases shall be used in all tests. Unless otherwise indicated, it is intended that all reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available. Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high purity to permit its use without_ lessening the accuracy of the determination. 5.2 Organic-free reagent water -All references to water in this method refer to organic-free reagent water, as defined in Chapter One ofSW-846. 5.3 Methanol, CHJ0H. Pesticide quality or equivalent. Store away from other solvents. 5 .4 Stock standards -Stock solutions will be purchased as certified solutions or gases. 5.5 Secondary dilution standards -Using stock standards, prepare secondary dilution standards. The secondary dilution standards should be prepared at concentrations such that the calibration standards prepared in Section 5.6 will bracket the working range of the analytical system. Secondary standards should be checked frequently for signs of degradation, especially just prior to preparing calibration standards from them. 5.6 Calibration standards -Calibration standards at three (for screening level data) to five (for definitive level data) concentrations are prepared by inert gas dilution of purchased gas standards. One of the concentrations should be at a concentration near, but below, the expected reporting detection limit. The remaining concentrations should correspond to the expected range of concentrations found in real samples or should define the working range of the GC. 6.0 SAMPLE COLLECTION, PRESERVATION,AND HANDLING 6.1 Sample collection is accomplished using a proprietary gas stripping system. A standard gas sampling bulb is configured such that a continuous stream of water is stripped for 15 minutes at a rate of500 ml/minute (other combinations of flow rate and time are possible). A gas tight syringe is used to collect a sample of the stripped.gas once an equilibrium condition is reached in the bulb. Samples are collected in duplicate. Generally, field analyses will be performed within one hour of sample collection but storage for as long as 24 hours is possible in a cool, dark, and dry location. 7.0 PROCEDURE 7. I Samples are introduced into the gas chromatograph in the same manner as calibration standards. 7.2 Follow the instrument manufacturers instructions for initial setup and testing. Establish initial instrument operating conditions based on those instructions. Run a series of test samples and review the data for adequate response and resolution. Adjust gas flow rates, temperature 28 I I I I I I I I I I I I I I I u n D D E E I I I I I I I I I I I I programs, or other instrument operating parameters as required to establish mid level calibration standard response at approximately 50% of scale and resolution between all analytes of interest of at least 60%. 7.3 Calibration 7.3. I Prepare calibration standards at three to five concentrations (refer to QAPjP for guidance) by introducing an inert gas through a dilution nozzle. One of the external standards should be at a concentration near, but below, the reporting detection limit. The other concentrations should correspond to the expected range of concentrations found inreal samples or should define the working range of the detector. 7.3.2 Inject each calibration standard using the technique that will be used to introduce the actual samples into the gas chromatograph (e.g. direct injection, sample loop, etc.). Tabulate peak height or area responses against the mass injected. The results can be used to prepare a calibration curve for each analyte. Alternatively, for samples that are introduced into the gas chromatograph using a syringe, the ratio of the response to the amount injected, defined as the calibration factor (CF), can be calculated for each analyte at each standard concentration. If the percent relative standard deviation (¾RSD) of the calibration factor is less than the requirements given in the QAPjP corresponding to the application over the working range, linearity through the origin will be assumed, and the average calibration factor can be used in place of a calibration curve. Total Area of Peak Calibration factor = Mass injected (in nano grams) 7.3.3 The working calibration curve or calibration factor must be verified on each working day by the injection of one or more calibration standards. The frequency of verification is established in the QAPjP. If the response for any analyte varies from the predicted response by more than the specification of the QAPjP, a new calibration curve must be prepared for that analyte. RI -R2 Percent Difference = X 100 RI where: RI = Calibration Factor from first analysis. R2 = Calibration Factor from succeeding analyses. 7.4 Gas chromatographic analysis 7.4. I Introduce samples into the gas chromatograph using a technique consistent with the initial calibration as described above. 29 7.5 Retention time windows 7.5.1 Using the data from the initial calibration of the instrument, establish retention time windows as follows. 7. 5 .1.2 Calculate the standard deviation of the absolute retention times for each component. 7.5.1.2 Plus or minus three times the standard deviation of the absolute retention times for each standard will be used to define the retention time window; however, the experience of the analyst should weigh heavily in the interpretation of chromatograms. 7.5.1.3 In those cases where the standard deviation for a particular standard is zero, the laboratory must substitute the standard deviation of a close eluting, similar compound to develop a valid retention time window. 7. 5. 3 The laboratory must calculate retention time windows for each standard on each GC column and whenever a new GC column is installed. The data must be retained by the laboratory. 7.6 Start each day of analysis with the injection of a media blank. Analyze a calibration verification standard, establish daily retention time windows, and verify that retention times and response are consistent. Retention times should not shift more than 2. 0% for packed columns, 0.3% for narrow bore capillary columns, or 1.5% for wide bore capillary columns. Response must meet the criteria given in section 7.3.3. If these conditions are not met, the analyst should review instrument operating conditions and rerun the standard once. If appropriate retention times and responses are not observed, a new initial calibration must be performed. 7 .6.1 Tentative identification of an analyte occurs when a peak from a sample extract falls within the daily retention time window. Confirmation will be performed on an alternate column only if required by the QAPjP. 7. 7 Record the sample volume purged or injected and the resulting peak sizes (in area units or peak heights). 7.7.1 Calculate raw analytical results by directly reading the·result from the calibration curve, or by multiplying the measured response times the calibration factor established in section · 7 .3 .3. The raw analytical result must be modified to produce a reportable result by means of Henry's law. Concentration ofX in air (or gas) (ug/1) = {[A(x)/(CF)]/[Vg]} • 0.81 (Henry's constant for hydrogen) where: 30 I I I I I R I I I I I I I I I I I I I D D I I I I I I I I I I I I CF= Calibration factor from section 7.3.2. V g = Volume of gas presented to the GC in liters, adjusted for dilution. 7. 9 If analytical interferences are suspected, or for the purpose of confirmation, analysis using the second GC column is possible. 7. IO If the response for a peak is off-scale, prepare a dilution of the sample. The dilution must be performed on an aliquot of the sample which has been properly sealed and stornd prior to use. 8.0 QUALITY CONTROL 8.1 Quality assurance requirements are presented in the QAPjP. 9.0 METHOD PERFORMANCE 9.1 Each analyst that performs the method must demonstrate the ability to meet the precision and accuracy requirements of the QAPjP prior to performing these analyses. At a minimum, duplicate spikes must produce percent recoveries and relative percent differences that meet the precision and accuracy goals as defined in the QAPjP. 10.0 REFERENCES IO. I SW-846 Methods 8000A, 8010B. 31 I I I D D D D I I I I APPENDIX D I Standard Operating Procedures Pace Analytical Services, Inc. I I I I I I I I I 0 R E m m m I I I I I I I I I I I I I Pace Analytical Pace Analytical Services. Inc. 9800 !(incey Avenuf:, Suite 100 Huntersville, NC 28078 Tel: 704-875-9092 Fax:"704-875-9091 STANDARD OPERATING PROCEDURE The Determination of Volatile Organic Compounds by Gas Chromatography/ Ma:is Spectrometry Reference Method: SW846 Method 8260A SOP NUMBER: EFFECTIVE DATE: SUPERSEDES: APPROVAL -S'D{i,,-,_~(,~ Operations Manager Quality Assurance Officer ( CA:./ General Manager CONTROLL.EO COPY NC!-O-017 COPY NO._ 3 April 15, I 997 Draft SOP Date /rlp-97 Date The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NCIO0l7 Apr. I 5, l 997 2 of 19 I. PURPOSE -The purpose of this Standard Operating Procedure (SOP) is to determine the concentration of volatile organic compounds in a variety of solid waste matrices by purge-and-trap GC/MS following SW-846 Method 8260A. This method is applicable to nearly all types of samples, regardless of water content. Common matrices are ground water, wastewater and aqueous sludges. II. SCOPE AND APPLICATION A. C. D. Analytes -Volatile organic compounds that have boiling points below 200°C and that are insoluble or slightly soluble in water. Volatile water-soluble compounds may be included in this analytical technique. However, for the more soluble compounds, quantitation limits are approximately ten times higher because of poor purging efficiency. (i.e. ketones) See list of individual compounds in calibration standards in section XIII. Summary -The volatile compounds are introduced into the gas chromatography by the purge-and-trap method. Purged sample components are trapped in a tube· of sorbent materials. When purging is complete, the sorbent tube is heated and backflushed with helium to desorb trapped components. The analytes are desorbed onto a capillary column. The column is temperature programmed to separate the analytes which are then detected with a mass spectrometer. Interferences -Interfering contamination may occur when a sample containing low concentrations of volatile organic compounds is analyzed immediately after or on an autosampler port which previously contained a sample containing high concentrations of volatile organic compounds. The preventive technique is rinsing the purging apparatus and sample syringes with two portions of organic free water between samples. Analyze one or more method blanks to check for cross contamination prior to sample analysis. Special precautions must be taken to analyze for methylene chloride. The analytical and sample storage area should be isolated from atmospheric sources of methylene chloride. Chromatography gas lines should be constructed from stainless steel or copper tubing. Laboratory clothing previously exposed to methylene chloride fumes during extraction procedures can contribute to sample contamination. Hazards and Precautions The toxicity and carcinogenicity of standards and reagents used in this method have not been precisely defined. Each chemical compound should be treated as a potential health hazard. Reduce exposure by the use of gloves, lab coats and I I I I I I I I I I I I I I I I I I I I u I m w I 0 D D I I I I I I I I I The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017 Apr. 15, 1997 3 of 19 safety glasses. Material Safety Data Sheets(MSDSs) are on file in the library and available to all personnel. Standard solutions should be prepared in a hood. ill. RESPONSIBILITIES IV. A. B. Analysts I. Analysts are responsible for adherence to the SOP. 2. All laboratory personnel are responsible for notifying their supervisor of any required revisions to the SOP. Operations Manager I. 2. The Operations Manager is responsible for ensuring adherence to this SOP. The Operations Manager is responsible for performing an annual review of the SOP. C. Quality Assurance Officer (QAO) D. I. The QAO is responsible for conducting laboratory audits to monitor adherence to this SOP. Results of the audit will be reported to the Operations Manager, General Manager and Corporate Quality. 2. The QAO is responsible for coordinating annual reviews of this SOP with the operations manager and the general manager. 3. The QAO is responsible for ensuring that all revisions to the SOP are implemented. 4. The QAO is responsible for monitoring distribution of and maintaining document control for this SOP. General Manager (GM) I. The GM is responsible for the overall implementation of and adherence to this SOP. 2. The GM is responsible for reviewing of this SOP with the QAO and the operations manager REVISIONS/REVIEWS A. This SOP will be reviewed on an annual basis at a minimum by the operations manager, general manager and the quality assurance officer. The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017 Apr. 15, I 997 4 of 19 B. The revised SOP will be distributed to all appropriate personnel and the superseded version replaced. V. DISTRIBUTION VI. Distribution of this SOP will be determined by the QAO. APPARATUS AND MATERIALS A. Reagents, Glassware and Materials B. I. Organic-free reagent water(OFW) 2. Methanol -purge and trap quality. (P&T MeOH) 3. 5 rnL volumetric flasks 4. 1 rnL amber vials 5. 4 rnL amber vials with Teflon lined screw caps 6. Teflon minninert valves for lrnL and 4rnL amber vials 7. 5 rnL Hamilton gas-tight syringe with shut-off valve 8. Gas-tight syringes: I rnL, S0uL, 25uL and I 0uL 9. Dispensing bottle for OFW 10. Teflon dispensing bottle for Methanol Instrumentation 1. Purge and Trap: Automatic Liquid Sampler -Tekmar Model ALS-2016 Liquid Sample Concentrator -Tekmar Model LSC-2000 Purge and Trap parameters: Trap: Supelco -Vocarb 3000 Standby Temp.: 39°C Purge: IO min. at 40 rnL/min. Dry Purge: 6 min. Desorb preheat: 245°C Desorb: 2.0 min. @ 250°C Bake: 8 min. @ 260°C Auto Drain: OFF Bake Gas Bypass: ON BGB Delay: 120 seconds Valve: 130°c Line: 130°C Mount: 4o•c I I I I n I I I I I I I I I I I I I I D u u I I I I I I I I I I I I I I The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NCIO017 -. Apr. 15, 1997 5 of 19 C. 2. 3. 2016 Valve: 2016 Line: 13o•c 13o•c GAS CHROMATOGRAPH -Hewlett Packard 5890 Column: HP-VOC 90m x 0.53mm id x 3.0um Carrier gas: Helium GC Temperature Programs: BFB VOA Initial Temp: 100°c 40°c Hold: 1 min 5 min Ramp: 6°C/min 8 °C/min Final Temp: 200°c 200°c Hold: 0 min 4min Run Time: 1433 min 29 min Mass Spectrometer -Hewlett Packard 5970 Mass Selective Detector (MSD) GC/MS Temperatures: Injection Port: 250°C Detector B(Inter.): 300°C Detector A(Jet): 220°C Mass Spec. Parameters: Scan Range: 3 5-300 Multiplier Voltage: Variable Threshold: 500 Sampling #: 3 AID Samples: 8 Standards: All standards are prepared using purge and trap methanol and stored in amber vials with Teflon lined screw caps or rnininert valves at 4°C. Gas standards must be replaced one month after the ampule is cracked. All other standards must be replaced 6 months after the ampule is cracked. Replace sooner if the standards show signs of degradation. As each standard, from the vendor is opened, record all pertinent information in the stock standard logbook. Record all standard preparations in the working standard logbook. I. Stock Solutions and Neat Standards: {As listed or equivalent source.) (For list of individual compounds in each standard, see section XII.) Supelco 8260 IS Mix@ 2000ug/mL Cat.# 4-8958 Supelco 8260 SS Mix@ 2000ug/mL Cat.# 4-8959 Supelco VOC Mix 1 @ 2000ug/mL Cat.# 4-8775 Supelco VOC Mix 2@ 2000ug/mL Cat.# 4-8777 Supelco VOC Mix 3 @ 2000ug/mL Cat.# 4-8779 Supelco VOC Mix 4@ 2000ug/mL Cat.# 4-8786 The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017 Apr. 15, 1997 6 of 19 2. 3. Supelco VOC Mix 5 @ 2000ug/mL Supelco VOC Mix 6 @ 2000ug/mL Supelco 8240B Cal. Std. Mix 2 @ 2000ug/mL Aldrich: 2-chloroethylvinyl ether (neat) Aldrich: vinyl acetate (neat) Aldrich: Acrolein (neat) Aldrich: Acrylonitrile (neat) Aldrich: trans-I, 4-dichloro-2-butene ( neai) Aldrich: carbon disulfide (neat) Aldrich: iodomethane (neat) Adrich: Freon I 13(112-trichlorotrifluoroethane) OC Standards(Altemate source): Cat.# 4-8797 Cat.# 4-8799 Cat.# 4-7364 Cat.# 10,998-3 Cat.# V 150-3 Cat.# 11,022-1 Cat.# 11,021-3 Cat.# 32,451-5 Cat.# 34,227-0 Cat.# I 850-7 Cat.# 17,282-0 Cat.# M-502A-R-I OX Cat.# M-502B-l OX AccuStandard M-502A-R-l0X@2.0mg/mL AccuStandard M-502B-l0X @ 2.0mg/mL AccuStandard M-8260ADD-l OX@ 2.0mg/mL Aldrich: Acrolein (neat) Cat.# M-8260-ADD-l0X Cat.# 11,022-1 Aldrich: Acrylonitrile (neat) Aldrich: trans-1,4-dichloro-2-butene (neat) Adrich: Freon I 13(112-trichlorotrifluoroethane) Internal Standard (IS) @ 25ug/mL Cat.# 11,021-3 Cat.# 32,451-5 Cat.# 17,282-0 -Halffill a 5mL volumetric flask with P&T MeOH and add: -62.5uL Supelco 8260 IS Mix@ 2000ug/mL -Dilute to 5mL with P&T MeOH. Stopper and invert 3 times to mix. -Transfer to a 4mL amber vial with a Teflon minninert valve. Internal/ Surrogate Standard Mix (IS/SS) @ 25ug/mL -Halffill a 5mL volumetric flask with P&T MeOH and add the following: -62.5uL Supelco 8260 IS Mix@ 2000ug/mL -62.5uL Supelco 8260 SS Spike Mix@ 2000ug/mL ·c Dilute to 5mL with P&T MeOH. Stopper and invert 3 times to mix. -Transfer to a 4mL amber vial with a Teflon minninert valve. 4. · Misc. 8260 Stds. (from neats)@2000/4000/10,000/20,000ug/mL -Half fill a 5mL volumetric flask with P&T MeOH and add the following: Density Final Cone. 9.5uL 2-chloroethylvinyl ether 21. 4uL vinyl acetate @ l.048mg/uL 2000ug/mL @ 0.934mg/uL 4000ug/mL 62uL Acrylonitrile @0.806mg/uL 10,000ug/mL l 20uL Acrolein @ 0.839mg/uL 20,000ug/mL I I I I I I D I I I I I I I I I I I I D I I I I I I I I I I I I I I I The Determination of Volatile Organics by GC/MS Method 8260A NC!-0-017 File Name: Date: Page: NC!O0!}. Apr. 15, 1997. 7 of 19 8.4uL trans-1,4-Dichloro-2-butene 7.9uL carbon disulfide 4.4uL iodomethane 6.4uL Freon 113 @ l.183mg/uL @ l.266mg/uL @ 2.280nig/uL @ l.57mg/uL 2000ug/rnL 2000ug/rnL 2000ug/mL 2000ug/rnL -Dilute to SmL with P&T MeOH. Stopper and invert 3 times to mix. -Transfer to a 4mL amber vial with Teflon lined screw cap. 5. 8260 Calibration Stock Solution@ 100/200#/500*/l000**ug/L -Transfer 350uL P&T MeOH to a lmL amber vial and add the following: -SOuL Supelco 8260 SS Mix@ 2000ug/mL -SOuL Supelco VOC Mix l @ 2000ug/mL -S0uL Supelco VOC Mix 2 @ 2000ug/mL -S0uL Slipelco VOC Mix 3 @ 2000ug/mL -S0uL Supelco VOC Mix 4 @ 2000ug/mL -S0uL Supelco VOC Mix 5 @ 2000ug/mL -50uL Supelco VOC Mix 6 @ 2000ug/mL -250uL Supelco 8240B Cal. Std. Mix 2 @ 2000ug/mL -50uL Misc. 8260 Stds. @2000/4000#/10,000*/20,000**ug/mL -Cap and invert 3 times to mix. (Final Volume = 1 mL) (# vinyl acetate, • isobutanol, acetone, 2-butanone, 2-hexanone, 4-methyl-2- pentanone and acetonitrile, • • acrolein and acrylonitrile) 6. 8260 Calibration Dilution@ 10/20#/50*/l00**mg/mL -Transfer 900uL P&T MeOH to a lmL amber vial and add: - 1 0OuL Calibration Stock Solution -Cap and invert 3 times to mix. (Final Volume = l mL) (# vinyl acetate, • isobutanol, acetone, 2-butanone, 2-hexanone, 4-methyl-2- pentanone and acetonitrile, •• acrolein and acrylonitrile) 7. Misc. 8260 OC Stock Std. (from neats)@ 2000/20,000ug/mL. -Half fill a SmL volumetric flask with P&T MeOH and add the following: Density Final Cone. 124uL Acrylonitrile @0.806 20,000ug/mL 120uL Acrolein @ 0.839mg/uL 20,000ug/.mL 8.4uL trans-l,4-Dichloro-2-butene @ l.183mg/uL 2000ug/mL 6.4uL Freon 113 @ l.57mg/uL 2000ug/mL -Dilute to 5mL with P&T MeOH. Stopper and invert 3 times to mix. -Transfer to a 4mL amber vial with Teflon lined screw cap. The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017 Apr. 15, 1997 8 of 19 8. 8260 LCS Solution@ 10/20*/l OO**ug/rnL -Half fill a SrnL volumetric flask with P&T MeOH and add: -25uL AccuStandard M502AR-I0X @2000ugirnL -25uL AccuStandard M502B-10X @2000ugirnL -50uL AccuStandard M8260 ADD-I OX @2000ugirnL -25uL Misc. 8260 QC Stds. @ 2000/20,000ugirnL -Dilute to SrnL with P&T MeOH. Stopper and invert 3 times to mix. -Transfer to a 4ml amber vial with a Teflon minninert valve. (* isobutanol acetone, 2-butanone, 2-hexanone, 4-methyl-2- pentanone and acetonitrile, • • acrolein and acrylonitrile) 9. 8260 Matrix Spike Solution@50/I00*/500**uglrnL -Halffill a 5rnL volumetric flask with P&T MeOH and add: -62.5uL AccuStandard M502AR-I0X @2000ugirnL -62.5uL AccuStandard M502B-1 OX @ 2000ugirnL -125uL AccuStandard M8260 ADD-l0X @2000ugirnL -62.5uL Misc. 8260 QC Stds. @2000/20,000ugirnL -Dilute to 5ml with P&T MeOH. Stopper and invert 3 times to mix. -Transfer to a 4ml amber vial with a Teflon minninert valve. (* isobutanol, acetone, 2-butanone, 2-hexanone, 4-methyl-2-pentanone, and acetonitrile, •• acrolein and acrylonitrile) VII. TUNING I CALIBRATION A. Tune Standard -50ng 4-Bromofluorobenzene(BFB) -Analyzed daily, BFB must pass the key m/z abundance criteria as defined below prior to sample analysis. Tune period = l 2hours. BFB Key m/z abundance criteria from table 4 of Method 8260A Mass 50 75 95 96 173 174 175 176 177 m/z Abundance criteria 15 to 40% of mass 95 30 to 60% of mass 95 · base peak, 100% relative abundance 5 to 9"/o of mass 95 less than 2% of mass 17 4 greater than 50% of mass 95 5 to 9"/o of mass 174 greater than 95% but less than 101% of mass I 74 5 to 9"/o of mass 176 I I I I H I u I I I I I I I I I I I I I D I m I I I I I I I I I I I I I I The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017, .. Apr. 15, 1997 . 9 of 19 B. Initial Calibration -The method requires calibration with a minimum of five concentration levels. The lowest calibration standard must be at the reporting limit as required by the state ofNorth Carolina. The other levels should correspond to the range of concentrations expected to be found in real samples. The instrument data system calculates a response factor (RF) for each analyte. In addition the data system generates a calibration report displaying the average response factor and percent relative standard deviation for each analyte in the series of calibration standards. Instrument calculations are listed below. Response Factor (RF) = (A,C;,) I (A;,C.) A, = Area of characteristic ion for the compound being measured. A;, = Area of characteristic ion for the specific internal standard. C;, = Concentration of the specific internal standard. C. = Concentration of the compound being measured. Average RF (RF) = sum of all RF's(across the calibration range) for "X'' # of calibration levels Percent Relative Standard Deviation (¾RSD) = SD ---X 100 mean The percent relative standard deviation(¾RSD) for the calibration check compounds (CCC's) must be <30% and the average minimum response factors for the system performance check compounds (SPCC's) are listed below. If the% RSD of any compound is greater than I 5%, construct calibration curves of area ratio (AJA;,) versus concentration using first or higher order regression fit of the five calibration points. SPCC's {minimum RF) Chloromethane (0. I) I, I 0Dichloroethane (0.1) Bromoform (>O. I) I, 1,2,2-Tetrachloroethane (0.3) Chlorobenzene (0.3) CCC's {¾RSD <30%) I, 1-Dichloroethene Chloroform 1,2-Dichloropropane Toluene Ethylbenzene Vinyl Chloride The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017 Apr. 15, 1997 10 of 19 C. Continuing Calibration -The Calibration curve must be verified each day by analysis of a mid-range calibration standard. We currently analyze the standard at SOugiL. The percent difference from the calibration for the calibration check compounds(CCC's) must be <20% and the minimum response factors for the system performance check compounds (SPCC's) must meet the criteria listed below. Internal Standard areas must be within-SO% and +100% from the last daily standard. Internal standard retention times must be within 30 seconds from the last daily standard. SPCC's (minimum RF) Chloromethane (0.1). 1, 1-Dichloroethane (0. 1) Bromoform (>0.1) 1, 1,2,2-Tetrachloroethane (0.3) Chlorobenzene (0.3) CCC's (¾Difference <20%) 1, 1-Dichloroethene Chloroform 1,2-Dichloropropane Toluene Ethylbenzene Vmyl Chloride % Difference = RF (from initial calibration) -RF (from continuing calibration) RF (from initial calibration) VIII. PROCEDUDRE: A B. C. All samples and standards must be allowed to warm to ambient temperature before analysis. Verify and document that all samples are properly preserved. After a portion of sample is poured into a SmL syringe (Step D-7), test the remaning sample with pH paper and KI starch paper. The pH paper should read <2 and there should be no residual chlorine detected on the KI starch paper (paper turns blue in the presence of chlorine). Document results in the injection log. Set up the purge-and-trap system and the GC/MS as outlined in section IV-B. The GC/MS system must be hardware-tuned using perfluorotributylarnine(PFTBA). In the manual tune program of the instrument, adjust the MS parameters to obtain the following mass ratios. Mass 69 131 219 Relative Abundance 100%, 30-40% 35-45% I I I R I I I I I I I I I I I I I I D D D I m I I I I I I I I I I I I I I The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017 .. Apr. 15, 1997 11 of 19 D. E. F. For best results, the abundance of219 should be slightly higher than the 131. Mass fragments associated with air and water (m/z 18, 28 & 32) should be less than 10% of the 69 ion. Print out a tune report displaying the profile scan and the spectrum scan. Clean the auto sampler ports. When the Tekmar is in the purge ready mode, press the [F2] key to get to the autosampler screen. Note the current ALS position. Press the [DRAIN] key. L 2. 3. 4. Open the valve on the current ALS port. Water will be forced out the top of the port. Catch the waste water in a beaker. Rinse the port two times with SmL of OFW, discard rinse water. If the port is contaminated (contained analytes above the calibration range), rinse with SmL ofMeOH prior to rinsing with water. Press the [STEP] key to advance to the next port. Repeat steps 1-3 for each position. When all ports are clean, press the [DRAIN] key to turn off the drain. Generate a sequence to run a batch of samples. The typical batch should include the following: BFB Tune Standard Calibration Standards or Continuing Calibration Standard Method Blank Laboratory Control Sample 20 samples Matrix Spike/Matrix Spike Duplicate Duplicate (Sample Duplicate or MSD every 10 samples) Note: The BFB tuning criteria and GC/MS calibration verification criteria must be met before analyzing samples. See section VII. Load the autosampler with standards and samples for the batch created above. I. BFB -SmLOFW -2uL IS/SS mix@ 25ugimL. -Fill a SmL gas tight syringe with SmL OFW and add 2uL IS/SS mix. Load onto the appropriate ALS position. The Determination of Volatile Organics by GC/MS Method 8260A NCI-O-017 File Name: Date: Page: NC1O017 Apr. 15, 1997 12 of 19 Calabration Standard I 2 3 4 ' 5 6 2. Initial Cahbration Standards -Prepare six calibration standards by filling 5 mL gas-tight syringes with OFW and add the appropriate amount of calibration solution and IS as indicated below. Load onto the autosampler. Vol. of Stock Cone.of Stock Final Final Solvent Solution Solution Volume Cone. (ug/L) (uL) (ug/mL) (mL) 2.5uL Cal. Di!. 10/20#/50•/ 100 .. 5 5/10#/25•150•• OFW IOuL IS 25 50 5.0uL Cal. Di!. 10/20#/50•/ 10/20#/50•/ 100•• 5 100•• OFW IOuL IS 25 50 lOuL Cal. Di!. 10/20#/50•/ 20/40#/JOO•t 100•• 5 200•• OFW IOuL IS 25 50 2.5uL Cal. Stock 100/200#/500•/ 50/100#/250•/ 1000•• 5 500 .. OFW lOuL IS 25 50 5.0uL Cal. Stock 100/200#/500•/ 100/200#/500*/ 1000•• 5 1000•• OFW lOuL IS 25 50 lOuL Cal. Stock 100/200#/500*/ 200/400#/1000*/ 1000•• 5 2000•• OFW IOuL IS 25 50 (# vinyl acetate, • isobutanol, acetone, 2-butanone, 2-hexanone, 4-methyl-2-pentanone and acetonitrile, • • acrolein and acrylonitrile) 3. 4. 5. Continuing Calibration Standard -Mid-range calibration Standard Final Cone. -5mLOFW -2.5uL Cal. Stock -lOuL IS@25ug/mL Method Blank -5mLOFW -1 OuL IS/SS @ 25ug/mL Laboratory Control Sample -5mLOFW -1 OuL LCS Solution -1 OuL IS/SS @ 25ug/mL 50/100#/200*/500**ug/L 50ug/L Final Cone. 50ug/L Final Cone. 20/40*/200**ug/L 50ug/L I I I I I I I I I I m I I I I I I I I D I I I I I I I I I I I I I I I The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NCIO0I7 Apr. 15, 1997 13 of 19 IX. 6. 7. (• isobutanol, acetone, 2-butanone, 2-hexanone, 4-methyl-2- pentanone and acetonitrile, • • acrolein and acrylonitrile) Matrix Spike/Matrix Spike Duplicate Final Cone. -5mL Sample -!OuL Matrix Spike Solution 50/I0o•t5oo••ug/L -I OuL IS/SS mix @ 25ugimL 50ug/L (• isobutano~ acetone, 2°butanone, 2-hexanone, 4-methyl-2- pentanone and acetonitrile, • • acrolein and acrylonitrile) Samples -5mL Sample Final Cone. -1 OuL IS/SS mix @ 25ugimL 50ugiL Sample Dilution I :2 1:5 I: IO 1:20 1:25 1:50 I: 100 1:1000 Volume of Sample to dilute to 5mL with OFW 2.5mL I.0mL 0.5mL 250uL 200uL I0OuL 50uL 5uL G. Analyze all standards, QC samples and samples from above a=rding to the instrument parameters as outlined in section IV. All QC samples must pass QC requirements as outlined in section VlI. CALCULATIONS (Quantitative Analysis): A. Concentration (ug/L) -The instrument will automatically calculate the concentration of each analyte in terms of ugiL. This is our final reporting unit. Therefore, no further calculations are necessary unless a dilution of the sample has been analyzed. ug/L (as calculated by the instrument) = (A,C;,) I (A.,RF) A. = Area .of characteristic ion for the compound being measured. A;, = Area of characteristic ion for the specific internal standard. C;. = Concentration of the specific internal standard. RF = Average response factor from initial calibration The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017 Apr. 15, 1997 14 of 19 Reported Concentration = ug/L x Dilution Factor B. Surrogate Recovery: C. D. % Recovery = LCS Recovery: % Recovery = Cone. Reported( ug/L) True Value ( 50 ug/L) Cone. Reported (ug/L) True Value (ug/L) X 100 X 100 Matrix Spike Recovery: % Recovery = Cone. of spiked sample -Cone. of sample True Value Spiked X 100 E. Duplicate Results -Relatvie Percent Difference (RPD): RPO = Difference x 100 Mean = (Cone. I -Cone. 2) x I 00 (Cone. I + Cone. 2) / 2 X. QUALITATIVE ANALYSIS (Data Interpretaion): A. B. Retention Time Comparison: The sample component relative retention tirne(RRT) must compare within+/-.0.06 RRT units of the RRT of the component in the continuing calibration standard. For reference, the standard must be run within the same 12 hours as the sample. Mass Spectrum Comparison: 1. All ions present in the standard mass spectra at a relative intensity greater than 10"/o (most abundant ion in the spectrum equals 100"/o) must be present in the sample spectrum). 2. The relative intensities of the ions greater than 10"/o must agree within+/-20% between the standard and sample spectra. XL QUALITY CONTROL A. Method Blank -To be analyzed daily with each batch of samples. The method blank must not contain any target analytes at or above the reporting limit. I I I B n I I I I I I I I I I I I I I D I I I I I I I I I I I I I I I I The Detennination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017- Apr. 15, 1_997 15 ofl9 B. C. D. E. F. G. Laboratory Control Sample(LCS) -Analyze one LCS daily. Spike 5mL ofOFW with all compounds being reported resulting in a final concentration of20ug!L for each analyte. Spiked compounds must pass the QC acceptance criteria as outlined in Table 6 of Method 8240(see LCS form in Appendix I). In-house limits to be determined. Matrix Spike/ Matix Spike Duplicate (MS/MSD) -Analyze one MS/MSD pair every 20 samples. Spike 5mL of a sample with all compounds such that the resulting concentration is 50ug!L for each analyte. The QC limits below were taken from the Contract Laboratory Program (CLP) -Statement ofWork 3/90. QC Limits ¾Recovery RPD I, 1-Dichloroethene 61-145 14 Benzene 76-127 14 Trichloroethene 71-120 11 Toluene 76-125 13 Chlorobenzene 75-130 13 Note: The state of NC requires that we spike the MS/MSD with all compounds. Duplicates -The state ofNC requires that we analyze one MSD or sample duplicate every IO samples. Surrogate Standards -To be added to all samples, spikes, control samples and method blanks, prior to purge-and-trap, to monitor method accuracy. Spike each 5mL portion of sample with surrogates such that the resulting concentration is 50ug!L for each surrogate. Surrogate recoveries must be within the following limits as outlined in table 9 of Method 8260A Dibromofluoromethane Toluene-d8 4-Bromofluorobenzene ¾Recovery 86-118 88-110 86-115 Internal Standards -Internal standard areas must be within a factor of two (-50% to +100%) compared to the associated continuing calibration standard. Internal standard retention times must be within 30 seconds from the retention times in the associated continuing calibration standard. Precision and Accuracy Study (P&A) -Prior to analysis of samples each analyst must establish the ability to generate acceptable accuracy and precision. Analyze 4 replicates of a QC standard @ 20 ug!L. Calculate the average recovery in ug!L and the standard The Determination of Volatile Organics by GC/MS Method 8260A NC!-O-017 File Name: Date: Page: NC!O0I7 Apr. 15, 1997 16 of 19 H. L J. K deviation of the the recovery in ug/L, for each analyte using the four results. Compare results to tables 7 and 8 in the method. Similar results should be obtained. Method Detection Limit (MDL) -A MDL study must be conducted annually. Analyze 7 replicates of a QC standard @ 5ug/L. Calculate a concentration for each analyte. Calculate the mean and standard deviation(S) of measurements for each analyte. Calculate the MDL as follows: MDL= t(n-1, a= .99) x S t(n-1, a = . 99) is the statistical t-value appropriate for the number of samples used to detennine the standard deviation at the 99 percent confidence level. Pace Reporting Limit (PRL) -Sug/L for most compounds. Sample Preservation: HCl to pH <2. Cool, 4°C. Sample Holding Time -14 days. XII. LIST OF STANDARDS: A. B. C. Internal Standard: Supelco 8260 IS Mix @ 2000ug/mL pentafluorobenzene I, 4-difluorobenzene I, 4-dichlorobenzene-d4 chlorobenzene-d5 Surrogate Standard: Supelco 8260 SS Mix @ 2000ug/mL I dibromofluoromethane toluene-d8 4-bromofluorobenzene Calibration Standards: Supelco VOC Mix I @ 2000ug/mL propylbenzene 1,2-dichlorobenzene I, 4-dichlorobenzene o-xylene 2-chlorotoluene tert-butylbenzene chlorobenzene 1,3-dichlorobenzene isopropylbenzene p-xylene sec-butyl benzene 4-chlorotoluene Cat.# 4-8958 Cat.# 4-8959 Cat.# 4-8775 I I I R e I I I I I I I I I I I I I I D D I I I I I I I I I I I I I I The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: NC1O017 Apr. 15, 1991 I 7 of 19 Date: Page: Supelco VOC Mix 2@2000ug/mL I ,3, 5-trimethylbenzene 1,2, 4-trichlorobenzene styrene p-isopropyltoluene bromobenzene 1,2, 4-trimethylbenzene toluene naphthalene ethylbenzene m-xylene n-butylbenzene 1,2,3-trichlorobenzene benzene Supelco VOC Mix 3 @ 2000ug/mL Cat.# 4-8777 Cat.# 4-8 779 1,2-dichloroethane I, I ,2,2-tetrachloroethane cis-1,3-dichloropropene 1,2-dibromoethane(ED B) trichloroethylene I, I, 1,2-tetrachloroethane I ,3-dichloropropane I, 1,2-trichloroethane 1,2-dichloropropane trans-1,3-dichloropropene hexachlorobutadiene 1,2-dibromo-3-chloropropane(DBCP) I, 1-dichloropropene 1,2,3_ -trichloropropane Supelco VOC Mix 4 @ 2000ug/mL I, I, I-trichloroethane bromoform bromochloromethane 2,2-dichloropropane chloroform I, 1-dichloroethane tetrachloroethylene dibromomethane carbon tetrachloride Cat.# 4-8786 Supelco VOC Mix 5 @ 2000ug/mL Cat.# 4-8797 I, 1-dichloroethylene methylene chloride trans-1,2-dichloroethylene cis-1,2-dichloroethylene bromodichloromethane dibromochloromethane Supelco VOC Mix 6 @ 2000ug/mL Cat.# 4-8799 vinyl chloride chloromethane bromomethane chloroethane trichlorofluoromethane dichlorodifluoromethane Supelco 8240B Cal. Std. Mix 2 @ 2000ug/mL isobutanol acetone 2-butanone 4-methyl-2-pentanone acrylonitrile 2-hexanone acetonitrile Cat.# 4-73 64 The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 File Name: Date: Page: NC1O017 Apr. 15, 1997 18 of 19 D. Aldrich: 2-chloroethylvinyl ether (neat) Aldrich: vinyl acetate (neat) Aldrich: Acrolein (neat) Aldrich: Acrylonitrile (neat) Aldrich: trans-1.4-dichloro-2-butene (neat) Aldrich: carbon disulfide (neat) Aldrich: iodomethane (neat) Aldrich: Freon I 13 (I 12-trichlorofluoroethane) QC Standards(Alternate source): Cat.# 10.998-3 Cat.# Vl50-3 Cat.# I I. 022-I Cat.# I 1.021-3 Cat.# 32.451-5 Cat.# 34.227-0 Cat.# I850-7 Cat. #17,282-0 AccuStandard M-502A-R-I0X@ 2.0mg/mL Cat.# M-502A-R-I OX benzene bromobenzene bromochloromethane bromodichloromethane bro mo form n-butylbenzene sec-butylbenzene tert-butylbenzene carbon tetrachloride chlorobenzene chloroform 2-chlorotoluene 4-chlorotoluene dibromochloromethane 1.2-dibromo-3-chloro-propane(DBCP) I .2-dibromoethane(ED B) 1,2-dichlorobenzene I. 4-dichlorobenzene I ,2-dichloroethane cis-I ,2-dichloroethene 1,2-dichloropropane 2,2-dichloropropane cis-1,3-dichloropropene ethylbenzene isopropylbenzene methylene chloride n-propylbenzene 1, 1, 1,2-tetrachloroethane tetrachloroethene dibromomethane I ,3-dichlorobenzene I, 1-dichloroethane I, 1-dichloroethene trans-1.2-dichloroethene 1,3-dichloropropane I, 1-dichloropropane trans-1,3-dichloropropene hexachlorobutadiene p-isopropyltoluene naphthalene styr~ne 1, 1,2,2-tetrachloroethane toluene I I B H I I I I I I I I I I I I I I I D ' I I I I I I I I I I I I I I I I I I The Determination of Volatile Organics by GC/MS Method 8260A NCl-O-017 1,2,3-trichlorobenzene 1,3, 5-trimethylbenzene m-xylene File Name: Date: Page: 1,2, 4-trimethylbenzene o-xylene p-xylene NC1O017 Apr. 15, 1997 19 of 19 Accustandard M-502B-10X @2.0mg/mL Cat.# M-502B-10X bromomethane chloroethane chloromethane trichlorofluoromethane dichlorodifluoromethane vinyl chloride Accustandard M-8260ADD-l0X@2.0mg/mL Cat.# M-8260-ADD-l0X acetone 2-butanone carbon disulfide 2-chloroethylvinyl ether 2-hexanone iodomethane 4-methyl-2-pentnaone vinyl acetate Aldrich: Acrolein {neat) Cat.# 11,022-1 Aldrich: Acrylonitrile (neat) Cat.# 11,021-3 Aldrich trans-1,4-dichlioro-2-butene {neat) Cat.# 32,451-5 Aldrich: Freon I 13 (112-trichlorofluoroethane) Cat. # 17,282-0 XIII. APPENDIX 1: GC/MS Forms attached. D I I I I I I I I I I I I I I DATE: VOLATILE GC/MS DAILY QC CHECKLIST SW846 -Method 8260A {Revision 1, September 1994) --------ANALYST: _____ _ 9FB TUNE REPORT (Present and criteria from Table 4 is met) _INITIAL CALIBRATION REPORT (A minimum of 5 standards are present and the following criteria is met) SPCC's lminim\lm RF! CCC's I%RSD < 30% *l Chloromethane (0.1) 1, 1 -Dichloroethene 1, 1-Dichloroethane (0.1) Chloroform Bromoform ( > 0. 1 I 1,2-Dichloropropane 1, 1,2,2-Tetrachloroethane (0.3) Toluene Chlorobenzene (0.3) Ethylbenzene Vinyl Chloride *Note: If the %RSD of any compound is greater than 15%, construct calibration curves of area ratio(Area of compound/ Area IS) versus concentration using first or higher order regression fit of the five calibration points. _CONTINUING CALIBRATION REPORT (A midpoint calibration standard is present and the following criteria is met) SPCC's {minimum RF! Chloromethane (0. 1) 1, 1-Dichloroethane (0. 11 Bromoform ( > 0. 1) 1, 1 ,2,2-Tetrachloruethane (0.31 Chlorobenzene (0.3) CCC's 1%RSD <20%•1 1, 1 -Dichloroethene Chloroform 1 ,2-Dichloropropane Toluene Ethyl benzene Vinyl Chloride -Internal standard areas must be within 50% from:the last daily standard. -Internal standard retention times must be within 30sec. from the last daily std. •Note: If the CCC's are not analytes required by the permit, then all required analytes must be within 20%. __ METHOD BLANK -(Present) -must be analyzed daily __ MATRIX SPIKE/ MATRIX SPIKE DUPLICATE (Present and Method criteria is met) -analyze a MS and a MSD every 10 samples. -Concentration = 50ug/L __ LABORATORY CONTROL SAMPLE (Present and method criteria is met) -analyze one LCS every l O carn~loo. UC\.~ . -Concentration = 20ug/L SURROGATE RECOVERY LIMITS: -Dibromofluoromethane -Toluene-dB -4-Bromofluoromethane 86-118% 88-110% 86-115% QC Outliers (Footnoted in report): ______________________________ _ I(.·_------------- 1 Reviewed by: ________ _ Date reviewed: -------- D E m I I I I I I I I I I I I I I ··• I SUBSET: from previous SAMPLE ID BATCH ID batch# QC Samples Blank Sample Dup MS MSD LCS date LCSD proiect due SAMPLES PROJECT# Date Entered:. ___ _ By Analyst:, __ MATRIX: Batch 92---------- DATAFILE(S) DATE(S) ANALVZED DATAFILE(S) DATE(S) ANALYZED ' Date Reviewed & Validated:. ___ _ By Analyst: __ D I I I I I I I I I I I I I I I :I ; I 1. I Pace Analytical Pace Analytical Services, Inc. 1700 Elm Street • Suite 200 Minneapolis, MN 55414 Tel: 612-617:6400 Fax: 612-617-6444 STANDARD OPERATING PROCEDUREr.::rci\/ED APR 2 , 1997 Methane in Water by GC FID SOP NUMBER MN-O-526-A AUTHOR Bob Schnobrich EFFECTIVE DA TE April 24, 1997 SUPERSEDES First Issue Control Copy Number: APPROVAL . ' . , W1JL .<}ha (JO.Ip/) ~ Labratory Manager . S\%vvv1-N ~.y-Lf~ Quality Assurance Officer L/1)._Lf /ci1- Date G2l CC:: :S'..U' cc:.;co~~· ..... METHANE IN WATER BY GCFID MN-O-526-A FILENAME: DATE·: PAGE: MNO526A.DOC April 23, 1997 I of 11 I. II. PURPOSE 1bis method is used for the determination of methane in water by headspace techniques, utilizing a GSQ column for separation of target compound with subsequent quantitation by a flame ionization detector (FID). SCOPE AND APPLICATION A. CONCENTRATION RANGES B. C. The range of methane starts at its practical quantitation limit (PQL) and ends at the concentration of the high level standard in the initial calibration curve. Method detection limits (MDL) (as determined by 40 CFR136, Appendix B, July l, 1987 and Chapter One, SW-846 Final Update n will be on file to support the PQL's for this analyte on each instrument used to report data. METI-IOD DETECTION LIMITS The method detection limit (MDL) is the minimum concentration of a compound that can be measured and reported with 99% confidence that the value is above zero. The MDL actually achieved in a given analysis will vary depending on instrument sensitivity and matrix effects; therefore, quantitation limits have been set. The quantitation limits are higher than the MDL and have been chosen to account for contamination from the sample preparation area and for some inherent instrument and matrix effects. HAZARDS AND PRECAUTIONS The toxicity or carcinogenicity of each reagent used in this method has not been precisely defined; however, each chemical compound should be treated as a potential health hazard. Exposure to these chemicals must be reduced to the lowest possible level by whatever means available (i.e., gloves, lab coats, goggles, and masks). Reference files of OSHA regulations and MSDS's are available to all personnel involved in the analysis. Additional references to laboratory safety have been identified and are available for inspection by the analyst. I D D I I I I I I I I I I I I I I I I I I I I I I I I I I I I METHANE IN WATER BYGCFID MN-O-526-A FILE NAME: DATE: PAGE: MNO526A.DOC April 2:3, 1997 2 of 11 III. SUMMARY OF METHOD IV. V. A. MATRIX I. This method is applicable to aqueous matrices. B. DESCRJPTION I. Methane is determined by heating a I OmL sample volume and injecting 3 cc of the volatilized gas into a gas chromatograph equipped with a column which allows for the separation of methane from other gaseous constituents. Methane is then detected using a flame ionization detection system. INTERFERENCES A. Impurities in the purge gas and organic compounds outgassing from the plumbing ahead of the trap may cause contamination problems. The analytical system must be demonstrated to be free from contamination by running method blanks under the condition of analysis. B. Samples can be contaminated by diffusion of volatile organics through the sample container septum during shipment and. Trip blanks, cooler blanks, and method blanks are prepared with organic-free water and used to check for this contamination. C. Contamination by carry-over can occur whenever high level and low level samples are sequentially analyzed. A DI blank should be run after highly contaminated samples. RESPONSIBILITIES A. B. PERSONNEL 1. All personnel involved with sample preparation and analysis are responsible for adherence to this SOP. 2. Personnel are responsible for ensuring that any deviations to this SOP are reported. 3. All personnel are responsible for notifying the laboratory manager and section supervisor of any required revisions to the SOP. LABORATORY MANAGER/SECTION SUPERVISOR METHANE IN WATER BYGCFID MN-0-526-A FILENAME: DATE: PAGE: MN0526A.DOC April 23, 1997 3 of 11 VI. C. I. The laboratory manager and section supervisor are responsible for ensuring adherence to this SOP. 2. The laboratory manager and section supervisor are responsible for performing an annual review of the SOP and reporting any required revisions to the Quality Assurance Office. QUALITY ASSURANCE OFFICE (QAO) I. The QAO is responsible for conducting laboratory audits to monitor adherence to this and other SOPs. Results of the audit will be reported to Management. 2. The QAO is responsible for ensuring that all revisions to the SOP are implemented. 3. The QAO is responsible for determining distribution of and maintaining document control for this SOP. REVIEWS/REVISIONS A. This SOP will be reviewed on an annual basis at a minimum. B. At the time of review, any required revision will be incorporated. C. The revised SOP will be distributed to all appropriate personnel and the superseded version replaced. I E I I I I I I I I I I I I I I I I I E I I I I I I I I I I I I I METHANE IN WATER BY GCFID MN-0-526-A VII. DISTRIBUTION FILE NAME: DATE: PAGE: MN0526A.DOC April 23, 1997 4 of 11 A. A controlled copy of this SOP will be available to all laboratory staff. The original copy will be maintained in a centralized laboratory location. B. The original of this SOP will be retained by the QAO. C. Distribution records will be retained by the QAO. VIII. APPARATUS AND MATERIALS A. GLASSWARE AND EQUIPMENT B. I. Sampling Containers: a) Water -20-mL vial ,vith Teflon-lined septum and crimp top cap. These vials are purchased pre-cleaned from a vendor. 2. Syringes: 5-mL gas tigbt glass hypodermic with Luerlock tip. 3. Microsyringes: I 0, 25, 50, I 00, 250, 500 and l000 u, gas tigbt 4. Volumetric flasks, Class A -10 mL 5. Graduated cylinder INSTRUMENTS I. Gas chromatographs a) b) Hewlett Packard 5890 Series II GC (or equivalent) Hewlett Packard 3396A integrator (or equivalent) METHANE IN WATER BYGCFID MN-0-526-A FILENAME: DATE: PAGE: MN0526A.DOC April 23, 1997 5 of 11 IX. X. 2. Headspace Sampler a) Hewlett Packard 19395A head space analyzer (or equivalent) 3 . Detectors 4. a) OI Flame Ionization Detector (or equivalent) Detector temperature is 250°C. All detector settings are typical. Analysts may make adjustments as required to optimize overall detector performance. Columns a) GSQ fused silica capillary column -30m x 0.53 mm ID, 0 urn film thickness (or equivalent). REAGENTS A. Reagent water: water in which there are no interferences at or above the reporting limit of the parameters of interest. A Culligan water pretreatment system is used to deionize the tap water. It is then sent through a polishing carbon filter to remove any organic contamination. B. Calibration Standard Solutions I. Standards are purchased at certified amounts in nitrogen. SAMPLE HANDLING AND STORAGE A. Aqueous Samples B. C. I. Samples are collected in 20 mL glass vials. The vials are filled with approximately I 0 mL of sample. The vial is inunediately sealed with a teflon-lined septum and a crimp top. The samples must be refrigerated at 4°C (+,-2°C) from the time of collection. All samples must be analyzed within 14 days from the date of collection. I I R I D D I I I I I I I I I I I I I E I I I I I I I 'I I I I I I METHANE IN WATER BYGCFID MN-O-526-A FILENAME: DATE·: PAGE: MNO526A.DOC April 23, 1997 6 of 11 XI. XII. GENERAL POLICIES AND PROCEDURES A. Compliant initial calibration for the analytes of concern is required before sample analysis may begin. B. Continuing calibration standards must meet established method criteria before proceeding with sample analysis. Using any calibrations not specifically meeting the established method criteria must be technically supported, and justification included in a narrative to the client. C. Quality control results must meet established method criteria before proceeding with sample analysis. Reporting any quality control results not specifically meeting the established method criteria must be technically supported, and justification included in a narrative to the client. CALIBRATION A. B. METIIOD START-UP AND VALIDATION I. To demonstrate the capability of the laboratory to generate valid data, the following steps need to be performed. a) b) c) d) e) Calibration standards are analyzed at a minimum of 3 concentrations. A calibration curve is established for each compound. The method detection limit is calculated by analyzing seven replicates prepared in blank water at 1 to 5 times higher than the estimated detection limit. Method detection levels are calculated according to 40 CFR 136, Appendix B (July I, 1987) and Chapter One, SW-846 Final Update I. The data are evaluated and, if acceptable, the method can be utilized on a routine basis. Any changes in laboratory preparation or chromatography that may effect the detection of the compounds requires that the MDL study be repeated. INITIAL CALIBRATION METHANE IN WATER BYGCFID MN-0-526-A FILENAME: DATE: PAGE: MN0526A.DOC April 23, 1997 7 of 11 I. 2. 3. Standard Preparation (See Section IX.B.) Instrument Calibration a) Calibration standards are prepared from the calibration gas standard in organic-free reagent water. One of the concentration levels should be at a concentration at or below the Pace Reporting Limit (PRL). The remaining concentrations will determine the working range of the system. The calibration levels used are 20, 100, l000 and l0000 ppm. The calibration range can be extended, but the high standard will determine the upper end of the calibration before dilution and reanalysis are required. The initial calibration is submitted for a technical review. b) The response factor for each compound is calculated by the data system for each level of calibration utilizing Equation I. The average response factor from these calibration standards is used to quantitate target analyte concentrations. Analysis of Calibration Data a) Tabulate peak area responses against concentration for each compound and internal standard and calculate response factors (RF) for each compound by using Equation I. Equation 1: Response Factor where: = = response for analyte of concern concentration of analyte of concern The percent relative standard deviation (¾RSD) for the response factors must be <20% to determine that the calibration is linear I I H D I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I METHANE IN WATER BY GCFID MN-0-526-A FILENAME: DATE: PAGE: MNOS26A.DOC April 23, 1997 8 of 11 C. DAILY (CONTINUING) CALIBRATION l. Standard Preparation a) Daily standards (midpoint concentrations of the calibration curve) are to be prepared in the same manner that the ICAL standards were prepared. b) The analysis of this standard must be repeated after each set of I 0 environmental samples. Method blanks and Laboratory control samples are not considered environmental samples. 2. Instrument Calibration a) b) c) d) The calibration factor for each analyte to be monitored must not exceed 15% difference when compared to the average response factor from the initial calibration. !fa check standard fails to meet the 15%D criteria, all associated samples ,vill require reanalysis .. Repeated analyte failures require an inspection of the GC system to determine the cause. Perform necessary maintenance before recalibration and resuming sample analysis. After the standards have been run, and are in control, a laboratory blank is analyzed to check the analytical system for interferences. The blank should contain less than the quantitation level of each analyte of interest before sample analysis can start. After calibration has been completed and the system is free of interferences, sample analysis can start. METHANE IN WATER BYGCFID MN-0-526-A FILE !'/AME: DATE: PAGE: MN0526A.DOC April 23, 1997 9 of 11 XIII. PROCEDURE A. B. C. GAS CHROMATOGRAPHY I. Detectors a) Flame Ionization Detector Detector temperature ; 25 0°C 2. Column • Restek GSQ fused silica capillary column 30m x 0.53 mm ID, 0 um film thickness • Column flow: IO mIJminute of Helium • Initial temperature: 40°C; hold for 5 minutes • Rate: 24°C/minute to 200°C • Final Hold: 15 minutes Note: All instrument conditions listed are examples. Analysts are expected to optimize instrument perfomiance and make changes as needed. All standards, blanks, samples and QC will be analyzed with the same conditions for the entire batch of samples. TROUBLESHOOTING 1. Preventative maintenance that is performed does not correct all problems associated with volatile organic analyses. Equipment malfunctions and samples with very high concentrations can cause a variety of problems that are difficult to diagnose. Each problem may require a combination of correction actions before acceptable data may be generated. 2. All corrective actions should be documented in the instrument specific maintenance logbook. n I g I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I METHANE IN WATER BY GCFID MN-O-526-A FILE J','AME: DATE: PAGE: MNOS26A.DOC April 23, 1997 IO of 11 XIV. QUALITY CONTROL A. B. Method Blank I. Before processing any samples, the analyst should demonstrate through the analysis of a method blank, that all glassware and reagents are interference free. 2. The method blank should be reanalyzed if these criteria are not met. All samples that contain a positive hit for analytes also found in the blank will be footnoted with a "B" flag. Laboratory Control Sample I. 2. The continuing calibration verification standards will be reported as the laboratory control sample. If statistically generated control limits are not being utilized, the acceptance criteria of ±20% from' the true value will be followed. 3. !fan LCS recovery falls outside of the limits, perform the following steps: a) b) c) Check to be sure there are no errors in calculations, standards preparation and spiking of the LCS solution, or problems with the instrument performance. For cases where the LCS recovery is above the QC limits, the decision can be made to accept the data if affected target analytes is not detected in the associated sample(s). Reanalyze the LCS once if the above steps failed to reveal or correct a problem. If the LCS recovery is within limits in the reanalysis, accept the associated sample data. METHANE IN WATER BYGCFID MN-0-526-A d) XV. REFERENCES FILE )',AME: DATE: PAGE: MN0526A.DOC April 23, 1997 11 of 11 If the recovery is outside of the limits after reanalysis, reprepare and reanalyze the LCS· and all associated samples when raw sample is available. If the LCS recovery is within limits in this analysis, accept the second set of data. When sufficient sample is not available to reanalyze samples after an LCS failure, the associated samples must be flagged in the final report. A. Test Methods for Evaluating Solid Waste, PhysicaVChemical Methods, SW-846, 3rd Edition, Final Update I, Method 3810. B. 40 CFR Part 136, Appendix B, July I, 1987. I D I u I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Pace Analytical Pace Analytical Services, Inc. 9800 Kincey Avenue, Suite 100 Huntersville. NC 28078 Tel: 704-875-9092 fax: 704-875-9091 STANDARD OPERATING PROCEDURE INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY Reference Methods: EPA 200.7 / SW846 6010A SOP NUMBER: EFFECTIVE DATE SUPERSEDES: APPROVAL ~r0 ,...,1¢='.i ch l \ ~ ::::.i Operations Manager ~()__II\. fl11lcfl'-- QuaJity Assurance Offic Fc,Q_General Manager -= NCI-1-002 April I 0, 1997 MN-1-313-B CONTROLILED COP{c. :\ COPY NO.. 'f Ge:I) t)/10/~1 Date 1-J/JO /er; Date INDUCTIVELY COUPLED PLASMA A TO MIC EMISSION SPECfROSCOPY File Name: NC11002 Date: April IO, 1997 NCI-1-002 Page: ·· 2 of 15 I. PURPOSE u. The; purpose of this Standard Operating Procedure (SOP) is to establish a procedure for the determination of metals by inductively coupled plasma atomic emissions spectroscopy. SCOPE/APPLICATION A. B. Scope: Inductively coupled plasma atomic emission spectroscopy (ICP-AES) is utilized for the determination of metals in solution. The method is applicable to a large number of matrices. All matrices, including ground water, aqueous samples, leachates, industrial wastes, soils, sludges, sediments, and other solid wastes, require digestion prior to analysis. Analytes: Elements for which this method is applicable are listed in Table I. C. Hazards and Precautions Concentrated acids are corrosive and should be used in a laboratory hood when possible. Protective clothing and safety glasses must be worn when working with concentrated acids.Each reagent and chemical used in this method should be treated as a potential health hazard. Reduce exposure by the use of gloves, lab coats, safety glasses and ventilation hoods. Material Safety Data Sheets(MSDSs) are on file in the library and available to all personnel. ID. RESPONSIBILITY A. Analysts · I. Analysts are responsible for adherence to the SOP. 2. All laboratory personnel are responsible for notifying the section supervisor/manager of any required revisions to the SOP. B. Operations Manager I. The operations manager is responsible for ensuring adherence to this SOP. 2. The operations manager is responsible for performing an annual review of the SOP. D H H a I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY File Name: NC11002 Date: April 10, 1997 NCl-I-002 Page: 3 of 15 C. Quality Assurance Officer (QAO) I. The QAO is responsible for conducting laboratory audits to monitor adherence to this SOP. Results of the audit are reported to Laboratory Management and Corporate Quality. 2. The QAO is responsible for coordinating annual reviews of this SOP with the general manager and the operations manager. More frequent reviews may be required. 3. The QAO is responsible for ensuring that all revisions to the SOP are implemented. 4. The QAO is responsible for determining distribution of and maintaining document control of this SOP. D. General Manager (GM) I. The GM is responsible for the overall implementation of and adherence to this SOP. 2. The GM is responsible for reviewing of this SOP with the QAO and the operations manager. IV. REVISIONS/REVIEWS A. This SOP will be reviewed on an annual basis at a minimum, by the operations manager, general manager and quality assurance officer. B. At the time of review, any required revision will be incorporated. C. The revised SOP will be distributed to all appropriate personnel and the superseded version replaced. V. DISTRIBUTION A. Distribution of this SOP will be determined by the QAO. INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY File Name: NCII002 Date: April IO, I 997 4 of 15 NCI-1-002 Page: VL VII. SUMMARY OF METHOD A. B. C. Prior to analysis, samples must be solubilized or digested using appropriate sample preparation methods. This method describes the simultaneous multielemental detennination of elements by ICP. The method measures element-emitted light by optical spectrometry. Samples are nebulized and the resulting aerosol is transported to the plasma torch. Element- specific atomic-line emission spectra are produced by a radio-frequency inductively coupled plasma. The spectra are dispersed by a grating spectrometer, and the intensities of the lines are monitored by photomultiplier tubes. Background correction may be required. Background is measured adjacent to analyte lines on samples during analysis. The position selected for the background-intensity measurement, on either or both sides of the analytical line, will be detennined by the complexity of the spectrum adjacent to the analyte line. The position used should be free of spectral interference and reflect the same change in background intensity as occurs at the analyte wavelength measured. Background correction is not required in cases ofline broadening where a background correction measurement would actually degrade the analytical result. The possibility of additional interferences named in Section VIII should also be recognized and appropriate corrections made as necessary. INTERFERENCES A. SPECIAL INTERFERENCES are caused by: I. Overlap of a spectral line from another element; 2. 3. 4. Unresolved overlap of molecular band spectra; Background contribution from continuous or recombination phenomena. Stray light from the line emission of high-concentration elements. Spectral overlap can be compensated for by computer-correcting the raw data after monitoring and measuring the interfering element. Background contribution and stray light can usually be compensated for by a background correction adjacent to the analyte line. Interelement correction factors are used on the simultaneous ICP. I n I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I CNDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY File Name: NCII002 Date: April 10, 1997 5 of 15 NCl-I-002 B. Page: PHYSICAL rNTERFERENCES I. These are effects associated with the sample nebulization and transport processes. Changes in viscosity and surface tension can cause significant inaccuracies, especially in samples containing high levels of dissolved solids or high acid concentrations. If physical interferences are present, they may be reduced by diluting the sample. Another problem that can occur with high dissolved solids is salt buildup at the tip of the nebulizer, which ,effects aerosol flow rate and causes instrument drift. The problem can be controlled by wetting the argon prior to nebulization, or diluting the sample. vm. APPARATUS AND MATERIALS A. INSTRUMENTATION B. C. D. E. F. G. I. 2. Thermo Jarrell Ash - T J A 61 E. Liquid argon BALANCE 1. Analytical balance, accurate to at least IO mg. BEAKERS 1. 150 mL or other appropriate vessel with watch glass covers. FILTER PAPER I. Whatman No. 41 or equivalent. VOLUMETRIC FLASKS · I. Assorted Class A volumetric flasks. PIPETS I. 2. Assorted Class A. Automatic pipets with disposable tips. HOTPLATES INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY File Name: NC11002 Date: April 10, 1997 NCl-1-002 Page: 6 of 15 IX. REAGENTS A B. C. D. E. F. G. Hydrogen peroxide, 30% Hydrochloric acid, Cone., trace metals grade. Hydrochloric acid, (1 :I). Add 500mL cone. HCI to 400mL DI water and dilute to 1 L. Concentrated Nitric Acid (HN03), trace metals grade, or equivalent. Nitric Acid (I: I) Add 500mL cone. HN03 to 400mL DI water and dilute to IL. Deionized Water (DI) Standard Stock Solutions, purchased (NIST Traceable). I. 2. 3. Mixed Calibration Standard Solutions: a. Prepare mixed calibration standard solutions by combining appropriate volumes of the stock solutions in volumetric flasks (see Table II for appropriate concentrations and element compatibility). Add 5 mL concentrated HN03 and dilute to I 00 mL with Type II water. Care should be taken when preparing the mixed standards to ensure that the elements are compatible and stable together. Fresh mixed standards should be prepared as needed, or if older than 180 days. Verify calibration standards using a second source standard. Some typical calibration standard combinations are listed in Table II. Spiking Solutions: a. A suitable spiking solution can be prepared from stock solutions by utilizing specified volumes and concentrations outlined in Table ill. Calibration Verification Standard: a. If the instrument was calibrated using the standards in Table n, a suitable verification standard would contain 1.0 mg/L of all listed constituents (10.0 mg/L of Potassium). D H I I g I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECfROSCOPY File Name: NCll002 Date: April 10, 1997 7 of 15 NCl-I-002 Page: X. SAMPLE PRESERVATION AND HOLDING TIMES A. SAMPLE PRESERVATION B. I. Water Sample Preservation 2. a. Measurement Parameter: Metals or dissolved metals. Samples are filtered through a 0.45 micron filter immediately on-site by the sampler before adding preservatives. b. Container: polyethylene or glass. c. Preservation: Sample preservation is performed by the sampler immediately upon sample collection. Use HN03 to bring the pH to <2. Soil and Sediment Preparation a. Soils/sediment will be maintained at 4° C ± 2 until analysis. HOLDfNG TIMES FOR WATER AND SOII.JSEDIMENT SAMPLES !. The maximum sample holding time for metals is 180 days from sample receipt. XL SAMPLEPREPARATION A. Refer to SW-846, 3000 series, sample preparation procedures and the corresponding PACE SOP. XII. INSTRUMENT AL ANALYSIS A. B. C. Consult instrument manufacturer's user's manuals for specific operational instructions. See Table IV for an example run sequence. fNSTRUMENT CALIBRATION I. Instrumental calibration is to be performed in accordance with the manufacturer's specifications. INDUCTIVELY COUPLED PLASMA A TO MIC EMISSION SPECfROSCOPY File Name: NC1I002 Date: April 10, 1997 8 of 15 NCl-1-002 2. Page: Instruments must be calibrated once every 24 hours and each time the instrument is set up. The instrument standardization date and time must be included in the raw data. xm. QA/QC REQUIREMENTS A. The QA/QC requirements for the analysis are listed below: B. C. 1. 2. 3. 4. 5. 6. Instrument Calibration Analysis of calibration standards Initial Calibration Verification (!CV) and Continuing Calibration Verification (CCV) Initial Calibration Blank (ICB), Continuing Calibration Blank (CCB) and Method Blank (MB) Laboratory Control Sample (LCS) Matrix Spike Sample (MS) and Matrix Spike Duplicate (MSD) 7. Interelement Corrections for ICP (ICSA, ICSAB) 8. Serial Dilution Analysis (L) ANALYSIS OF CALIBRATION ST AND ARDS 1. Each calibration standard should be analyzed after calibration. The results must agree with 5% of the accepted value. If the results are not within 5%, the analysis should be terminated, the problem corrected and the instrument recalibrated. INITIAL CALIBRATION VERIFICATION (ICV) AND CONTINUING CALIBRATION VERIFICATION (CCV) 1. Initial Calibration Verification (!CV) a. The Initial Calibration Verification Solution(s) should be obtained from a different source than the calibration standards. I I u I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I CNDUCTTVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY File Name: NC11002 · · Date: . Page: April IO, 1997 9 of 15 NCl-1-002 C. D. b. Immediately after the calibration standards have been analyzed, the accuracy of the initial calibration shall be verified and documented for every analyte by the analysis ofan Initial Calibration Verification Solution(s) at each wavelength used for analysis. When measurements exceed the control limits of 5% the analysis should be terminated, the problem corrected, the instrument recalibrated and the calibration reverified. CONTINUING CALIBRATION VERIFICATION (CCV) I. 2. To ensure calibration accuracy during each analysis run, a continuing calibration verification must be analyzed for each analyte, at a frequency of 10% during an analytical run. The standard must also be analyzed after the last analytical sample. The analyte concentrations in the continuing calibration standard should be at or near the mid-range levels of the calibration curve. The ICV solution can be utilized as the CCV. If the deviation of the continuing calibration verification is greater than the control limits of± 10% (± 5% for Method 200.7-Wastewater), the instrument must be recalibrated and the preceding analytical samples reanalyzed since the last acceptable calibration verification must be reanalyzed. INITIAL CALIBRATION BLANK (ICB), CONTINUING CALIBRATION BLANK (CCB) AND METHOD BLANK (MB) I. 2. Initial Calibration Blank (ICB) and Continuing Calibration Blank (CCB) a. A calibration blank must be analyzed at each wavelength used for analysis immediately after every initial and continuing calibration verification, at a frequency of I 0%. Method Blank (MB) Analysis a. At least one method blank (or reagent blank), consisting of DI water must be prepared and analyzed with each group of samples digested. If the concentration in the MB is greater than 3X the MDL, the samples associated with that MB must be reprepped. No elements should be present at a concentration greater than the Pace Reporting Limit (PRL). INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY File Name: NC11002 Date: April 10, 1997 NCl-1-002 E. F. Page: 10 of 15 LABORATORY CONTROL SAMPLE (LCS) I. Laboratory Control Samples (LCS) must be analyzed for each analyte using the same sample preparations, analytical methods and QA/QC procedures employed for the samples received. One aqueous LCS must be prepared and analyzed for every batch of samples digested. 2. If the percent recovery for the LCS falls outside the control limits of 90-110% ( exception: Ag and Sb), the analyses should be terminated, the problem corrected, and the samples associated with that LCS redigested and reanalyzed. MATRIX SPIKE SAMPLE (MS) The matrix spike sample analysis is designed to provide information about the effect of the sample matrix on the digestion and measurement methodology. · The spike is added before the digestion (i.e., prior to the addition of other reagents). At least one spiked sample must be analyzed for each batch of samples of a similar matrix spike type (i.e., water, soil) at a minimum frequency of 5%. Spiking levels are listed in Table ill. D H I I I I I I I I The percent recovery of the spike is calculated from the following equation: I G % Recovery= (SSR-SR) • 100 I ST Where: SSR = SR = ST = Spike sample result, mg/L or mg/kg dry Sample result, mg/L or mg/kg dry Spike target, mg/L or mg/kg dry 2. When sample concentration is less than the instrument detection limit, let SR= 0 only for calculating percent recovery. · MATRIX SPIKE DUPLICATE ANALYSIS l. 2. One matrix spike duplicate sample must be analyzed from each batch of samples of a similar matrix type (i.e., water, soil). This analysis will be performed at a minimum frequency of 5%. The relative percent differences can be calculated as follows: I I I I I I I I I I I I I I I I I I I I I I I I INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECfROSCOPY File Name: NC!.1002 Date: April 10, I 997 11 of 15 NCl-1-002 H. RPD= Where: RPD s D (S-D) • (100) (S+D)/2 = Relative Percent Difference Page: = Original Spiked Sample Value, mg/Lor mg/kg dry = Second Spiked Sample Value, mg/Lor mg/kg dry INTERELEMENT CORRECTIONS FOR SIMULTANEOUS ICP I. Interelement correction factors must be determined annually. Correction factors for spectral interference due to Al, Ca, Fe, and.Mg must be determined for all ICP instruments at all wavelengths used for each analyte reported by ICP. Correction factors for spectral interference due to analytes other than Al, Ca, Fe, and Mg must be reported if they were applied. 2. If the instrument was adjusted in any way that may affect the ICP interelement · correction factors, the factors must be redetermined. XIV. DOCUMENTATION A. B. RAW DAT A FILES/ MAINTENANCE I. Each daily folder must contain the sample sequence, all calibration data and raw data for samples and QC samples. Maintenance should also be recorded as performed and logged into the appropriate maintenance logbook. · ST AND ARD PREPARATION LOG BOOK I. Record the necessary information (volumes, manufacturer, lot number, etc.) in the standard logbooks. XV. REFERENCES A. B. Test Methods for Evaluating Water and Solid Waste, SW-846 3rd Edition, final Update L Method 6010A USEPA Methods for Chemical Analysis of Water and Wastes, EPA-600/4-79-020, March 1983. D INDUCTIVELY COUPLED PLASMA ATOMIC File Name: NC11002 I EMISSION SPECfROSCOPY Date: April 10, 1997 NCl-1-002 Page: 12 of 15 I TABLE I I TARGET ANALYTE LIST Wavelength' MDLb PRLC PRLC I Element (run) (mg/L) (mg/L) (mg/kg) Waters Soils Aluminum Al 308.216 0.0052 0. IO 10 I Antimony Sb 206.838 0.020 0.20 20 , -Krsenic:. As 189.042 0.026 0.50 so I . Baiium· Ba 493.409 0.0066 0.01 1.0 Beryllium Be 313.042 0.0001 0.01 1.0 Boron B 182.640 0.070 0.10 JO I < Cadrniull_l. Cd 228.802 0.0022 0.01 1.0 Calcium Ca 317.933 0.022 0.10 JO I Chromium Cr 267,716 0.0016 0.01 1.0 Cobalt Co 228.616 0.0024 0.01 1.0 Copper Cu 324.754 0.0014 0.01 1.0 I Iron Fe 259.940 0.0016 0.05 5.0 Lead-Pb 220.353 0.017 0.10 10 --. ··-- Magnesium Mg 279.079 0.033 0.20 20 I Manganese Mn 257,610 0.0006 0.01 1.0 Molybdenum Mo 202.030 0.0074 0.05 5.0 Nickel Ni 231.604 0.0097 0.05 5.0 I Potassium K 766.491 0.27 1.0 100 SeleniuJll Se 196.026 0.037 0,5 50 .. ··sijye( Ag 328.068 0.0019 0.05 5.0 I Sodium Na 588.995 0.027 0.20 20 Thallium TI 190.864 0.18 0.50 so Tin Sn 283.999 0.048 0.20 20 I Titanium Ti · 334.941 0.0009 0.01 1.0 Vanadium V 292.402 0.0023 0.01 1.0 Zinc Zn 213.856 0.0026 0.01 1.0 I Footnotes: I • The wavelengths listed are recommended because of their sensitivity and overall acceptance . b The method detection limits (MDL) shown are approximate. Actual detection limits are instrument specific and matrix dependent. The above limits were generated from replicate I analysis of spiked DI-water. C The Pace Reporting Limits (PRLs) shown are based on method requirements, regulatory requirements and client specifications. I I E I INDUCTIVELY COUPLED PLASMA ATOMIC File Name: NCH002 EMISSION SPECTROSCOPY Date: April 10, 1997 I NCl-I-002 Page: 13 of 15 I TABLE II CALIBRATION STANDARD CONCENTRATIONS I Calibration Standard -Mix A Calibration Standard -Mix B I Blank Level 1 Level 2 Level 3 Level 1 • Level 2 Level 3 Element ~ ~ ~ !!!filk ~ ~ mg/L I I As 0 2 5 Al 0 2 5 Ba 0 I 2 5 I Be 0 2 5 B 0 I 2 5 Cd 0 I 2 5 I Ca 0 I 2 5 Cr 0 2 5 Co 0 I 2 5 I Cu 0 I 2 5 Fe 0 I 2 5 Pb 0 I 2 5 I Mg 0 I 2 5 Mn 0 2 5 Mo 0 I 2 5 I Ni 0 I 2 5 K 0 10 20 50 Sb 0 I 2 5 I Se 0 I 2 5 Ag 0 2 5 I Na 0 I 2 5 Tl 0 I 2 5 Sn 0 I 2 5 I Ti 0 I 2 5 V 0. 2 5 Zn 0 I 2 5 I ii I INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY File Name: NC1I002 Date: April 10, 1997 14 of 15 NCl-l-002 Page: TABLE ID ICP SPIKING SOLUTION mLsof Cone. of Spike Target Element Stock Solution ( I) Spiking Solution (ppm)(]) (ppm) (2) Al l.0 5000 50 Ag 1.0 500 5 As l.O 500 5 Ba 1.0 500 5 Be 1.0 100 B l.O 100 Ca l.O 10,000 100 Cd 1.0 100 Co 1.0 100 Cr l.O 100 Cu l.0 100 Fe l.O 5000 50 Mg l.O 5000 50 Mn l.O 100 Mo 1.0 100 Ni 1.0 100 I K 1.0 10,000 100 Pb 1.0 500 5 Se l.O 500 5 Na 1.0 5000 50 Sb l.O 500 5 Tl 1.0 500 5 Sn 1.0 100 Ti 1.0 100 V 1.0 100 Zn l.O 100 (1) The spiking solution is made to a final volume of 1000 mL in 5% cone HNOJ. (2) Spike targets are the result of adding 1.0 mLs of spiking solution to I 00 mLs of sample. D I I I I I I I I I I I I I I I I I I m I I I I I I I I I I I I I I I I I I INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROSCOPY File Name: NC11002 Date: April 10, 1997 15 of 15 NCl-I-002 l 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 . 24 25 26 27 28 . 29 Page: TABLE IV TYPICAL SAMPLE RUN SEQUENCE Calibration Standards ICY ICB MDL Check Standard Independent Check Standard ICSA-Initial ICSA-Initial Nickel Std@0.02mg/L (Only for Drinking Water Samples) MB LCS/LCSD Sample 1 Sample 2 Sample 2 -Matrix Spike Sample 2 -Matrix Spike Duplicate Sample3 CCVI CCBl Sample 4 Sample 4 -Duplicate Sample 5 Sample6 Sample 7 Sample 8 Sample 9 Sample 10 ICSA-Final ICSAB-Final CCV2 CCB2 I I I I I I I I I I I I I I I I I I I Pace Analytical Pace Analytical Services, Inc. 9800 Kincey Avenue. Suite I 00 Huntersville. NC 28078 Tel: 704-875-9092 Fax: 704-875-9091 STANDARD OPERATING PROCEDURE Mercury (Hg) -Waters Reference Methods: EPA 245.l / SW846 -7470A vr111 .C // i, SOP NUMBER: EFFECTIVE DATE: NCl-1-031 April 24, I 997 Draft SOP SUPERSEDES APPROVAL Quality Assurance Officer General Manager CONTROLL.ED COPY COPY NO._.-=~-- 1,.-f-Z'<-91-- Date Date Date Mercury-Hg Waters File Name: NCII028 Apr. 24, 1997 2 of7 EPA 245.1 / SW846 7470A NCI-I-028 Date: Page: L II. m. PURPOSE The purpose of this standard operating procedure (SOP) is to determine the concentration of mercury in drinking, surface, saline and ground waters, domestic and industrial wastes. SCOPE/APPLICATION A. Summary -Samples are prepared for analysis by an acid digestion procedure. Digested samples are analyzed by a cold-vapor atomic absorption technique based on the absorption of radiation at 253. 7 run by mercury vapor. B. Hazards and Precautions Each reagent and chemical used in this method should be treated as a potential health hazard. Reduce exposure by the use of gloves, lab coats, safety glasses and ventilation hoods. Material Safety Data Sheets(MSDSs) are on file in the library . and available to all personnel. RESPONSIBILITY A. B. C. Analysts I. Analysts are responsible for adherence to the SOP. 2. All laboratory personnel are responsible for notifying the section supervisor/manager of any required revisions to the SOP .. Operations Manager 2. The operations manager is responsible for ensuring adherence to this SOP. The operations manager is responsible for performing an annual review of the SOP. Quality Assurance Officer (QAO) 1. The QAO is responsible for conducting laboratory audits to monitor adherence to this SOP. Results of the audit are reported to Laboratory Management and Corporate Quality. D H I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Mercury-Hg Waters File Name: NC!I028 Apr. 24, 1997 3 of 7, EPA 245.1 / SW846 7470A NC!-I-028 Date: Page: IV. V. D. 2. The QAO is responsible for coordinating annual reviews of this SOP with the general manager and the operations manager. More frequent reviews may be required. 3. The QAO is responsible for ensuring that all revisions to the SOP are implemented. 4. The QAO is responsible for determining distribution of and maintaining document control of this SOP. General Manager (GM) 1. 2. The GM is responsible for the overall implementation of and adherence to this SOP. The GM is responsible for reviewing of this SOP with the QAO and the operations manager. REVISIONS/REVIEWS A. This SOP will be reviewed on an annual basis at a minimum, by the operations manager, general manager and quality assurance officer. B. At the time of review, any required revision will be incorporated. C. The revised SOP will be distributed to all appropriate personnel and the superseded version replaced. DISTRIBUTION A. Distribution of this SOP will be determined by the QAO. Mercury-Hg Waters File Name: NCII028 Apr. 24, 1997 4of7 EPA 245.1 / SW846 7470A NCI-I-028 Date: Page: VI. APPARATUS AND MATERIALS A. B. Apparatus / Equipment I. 2. 3. 4. 5. Water Bath, capable of maintaining 95 degree Celsius. Pipettes: 100, 1000, 2500 and 10,000 uL Test Tubes: 13 x 100 with caps 25 ml volumetric flask, class A (Clean flask with I: I HNO3) Mercury Analyzer -Leeman PS200 or equivalent Reagents / Standards I. 2. 3. 4. 5. 6. 7. 8. 9. IO. II. 12. Deionized water (DI-H2O) Hg Stock Solution ( 1000 mg(L). Reagents Inc. or equivalent. Alternate Source -Hg Stock Solution ( 1000 mg(L). to be used for LCS and matrix spike. Primary Hg Solution (I 0mg(L): Fill a 25 mL volumetric flask half way with DI-H2O. Pipette 37.5 uL HNO3 and 250 uL of the Hg Stock Solution into flask and dilute to mark. Store at 4°C. Alternate Source -Primary Hg Solution (I 0mg(L): Fill a 25 mL volumetric flask halfway with DI-H2O. Pipette 37.5 uL HNO3 and 250 uL of the Alternate Source -Hg Stock Solution into flask and dilute to mark. Store at 4°C. Working Hg Solution (0.1 mg(L or I00ug(L): Fill a 25 mL volumetric flask halfway with reagent water. Pipette 37.5 uL HNO3 and 250 uL Primary Mercury Solution into flask and dilute to mark. Prepare daily. LCS -Working Hg Solution (0.1 mg(L or I00ug/L): Fill a 25 mL volumetric flask halfway with reagent water. Pipette 37.5 uL HNO3 and 250 uL Alternate Source -Primary Hg Solution into flask and dilute to mark. Prepare daily. Nitric Acid; HNO3 cone. Sulfuric Acid; H2SO, cone. Potassium Permanganate Solution: Dissolve 50 g KMnO, in 800 mL reagent water and dilute to I liter. Potassium Persulfate Solution (K2S2O8): Dissolve 50 g K2S2Os in 800 mL reagent water and dilute to I liter. Note: You may need to heat solution to dissolve completely. Sodium Chloride-Hydroxylarnine Solution: Dissolve 120 g NaCl and 120 g (NH2OH2)*H2SO4 in 800 mL reagent water and dilute to I liter. I u I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Mercury-Hg Waters File Name: NC1I028 Apr. 24, 1997 5 of7 EPA 245.l / SW846 7470A NCl-I-028 Date: Page: VII. PROCEDURE A. Preparation of Calibration Standards: I. 2. Make Working Hg Solution@ 100 ug/L. Using 25 mL volumetric flasks, prepare calibration standards by diluting aliquots of the. working mercury solution as indicated in the chart below. Calibration Volume of Cone. of Final Final Standard Hg Solution Hg Solution Volume (mL) Cone. Blank I 2 3 4 5 B. C. (mL) (ug/L) (ug/L) 0 -25 0 0.05 100 25 0.2 0.25 100 25 1.0 0.50 100 25 2.0 1.25 100 25 5.0 2.5 100 25 10.0 3. Transfer 8 mLs of each standard to separate test tubes to be digested. Preparation of LCS and Matrix Spike: I. Make LCS -Working Hg Solution@ 100 ug/L. 2. Using 25 mL volumetric flasks, prepare the QC samples as follows: a. LCS -Fill a 25mL volumetric flask halfway with DI-H2O. Add 0.5 mL of the LCS -Working Hg Solution. Dilute to mark. b. Matrix Spike -Fill a 25mL volumetric flask half way with a sample designated to be used as the matrix spike. Add l.25mL of the LCS -Working Hg Solution. Dilute to mark. 3.. Transfer 8 mLs of each QC sample to separate test tubes to be digested. Digestion: I. Turn on and fill water bath with reagent water. Adjust temperature to 95•c. 2. 3. 4. 5. Pipette 8 mLs ofDI-H2O into a test tube to be used as a method blank. Pipette 8 mLs of each sample into separate test tubes. Gather test tubes for calibration standards and QC samples prepared in steps A and B above. Add 400 uL H2SO., 200 uL HNO3, and 1200 uL KMnO. to each test tube (Including the calibration standards, method blank, and QC samples Mercury -Hg Waters File Name: NC1I028 Apr. 24, 1997 6of7 EPA 245.1 / SW846 7470A NCI-I-028 Date: Page: D. 6. 5. 6. 7. 8. Let stand for 15 minutes. If purple color does not persist, add more KMnO4 or dilute sample. (Note: If you add more KMnO4 to sample, make sure to add the same amount to the calibration standards, QC samples and all other samples.) After I 5 minutes pipette 650 uL K2S2O 8 into sample, cap and invert twice. Place in water bath for 2 hours. After samples have cooled to room temperature pipette 480 uL of (NH2OH2)*H2SO4 and mix until purple color turns clear. Samples are ready for analysis by the Cold Vapor method. Analysis: 1. Transfer the test tubes to the Mercury Analyzer autosampler. 2. Foil ow the manufacturers instructions for operation and calibration of the Mercury Analyzer. 3. The instrument must be calibrated every day. The correlation coefficient of _ the calibration curve must be~ 0.995. 4. Verify the reporting limit by reanalyzing calibration standard 1 @ 0.2ug/L after the instrument is calibrated. The results must be within +/-25%. 5. The instrument will print out a report of the concentration of Hg in each sample. VIlI. QUALITY CONTROL A. B. C. Method Blank: To be run daily with each batch of samples. Mercury should not be detected at or above the reporting limit of0.2 ug/L. Duplicate: NC requires a duplicate every IO samples. A matrix spike duplicate may count as a sample duplicate. The relative percent difference(RPD) must be '.520%. Special consideration must be made for concentrations less than ten times the reporting limit. RPD = Difference x I 00 Mean = (Cone. 1 -Cone. 2) x 100 (Cone. 1 + Cone. 2) / 2 Spike: A matrix spike must be analyzed every batch of20 samples for wastewater and every 10 samples for drinking water. Spike so that the resulting concentration is equal to 5.0ug/L. The percent recovery must be within+/-20% for wastewater and +/-10% for drinking water. % Recovery = (Spiked Sample Value -Sample Value) x 100 True Value I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Mercury -Hg Waters File Name: NC!I028 Apr. 24, 1997 7 of7 EPA 245.1 / SW846 7470A NCI-I-028 Date: Page: D. E. F. G. Laboratory Control Sample(LCS): To be analyzed with each batch of20 samples. Spike so that the resulting concentration is equal to 2.0ug/L. The percent recovery of the LCS must be within 10% of the expected value. % Recovery = Reported Value x I 00 True Value Pace Reporting Limit (PRL): 0.2 ug/L Sample Preservation: HNOi to pH <2. Holding Time: 28 days. E I I I I I I I I I I I I I I I I I I STANDARD OPERATING PROCEDURE Total Organic Carbon by Persulfate-Ultraviolet Oxidation Method Standard Method 5310 C SOP Number Author Effective Date Supersedes Quality Assurance Officer General Manager Approval L ASV-I-105 Robin K Blankenbaker April 2, 1997 All previous editions L.f-(D-i7 Date Date I I I I I I I ·• I I I I I I I I I TOC 1. Scope/Application ASV-I-105 April 2, 1997 Page 1 of 4 1.1. The persulfate-ultraviolet method is a rapid, precise method for the measurement of trace levels of organic carbon in water. This is important for industries where even trace levels of organic compounds may be detrimental to the process for which the water is being utilized. 1.2 Concentrations of 0.05mg carbon/L can be measured if close attention is given to minimizing sample contamination and method background. The combustion-infared method should be used for high concentrations of TOC. 2. Summary of Method 2.1 Organic carbon is oxidized to CO2 by persulfate in the presence of ultraviolet light. The CO2 produced can be measured directly by a nondispersive infrared analyzer, be reduced to methane and measured by a flame ionization detector, or be chemically titrated. 2.2 Samples are introduced into a reactor by autosampler or injected manually. The resulting CO2 is sparged continuously from the solution and carried in the gas stream to an infrared analyzer specifically tuned to the absorptive wavelength of CO2. 2.3 The instrument's microprocessor calculates the area of the peaks produced by the analyzer, compares them to the peak area of the calibration standard stored in it's memory, and prints out a calibrated organic carbon value in mg/L. 3. Interferences 3.1 Excessive acidification of sample, producing a reduction in pH of the persulfate solution to 1 or less, can result in slow or incomplete oxidation of organic carbon. 3.2 Highly turbid samples may not be completely oxidized due to the UV's inability to penetrate the sample. Also, very large or complex organic compounds may be oxidized slowly. 3.3 If the sample has a chloride concentration greater than 0.1 %, oxidation of organic matter may be inhibited completely. Mercuric nitrate should be added to the persulfate solution to eliminate the chloride. 3.4 Sample contamination during handling, sampling, or analysis is a potential problem in any organic carbon measurement. Take extreme care with the samples. TOC 4. Sample Handling and Preservation ASV-I-105 April 2, 1997 Page 2 of 4 Samples are collected and stored in 250ml plastic bottles and preserved at 4C with minimal exposure to light. Samples may be acidified to pH of~ with sulfuric acid. 5. Safety Safety glasses, lab coats, and gloves should be worn at all times. As sample constituents are not known, all samples should be treated as hazardous. 6. Apparatus a. A total organic carbon analyzer (Xertex-Dohrmann DC-80 or equivalent). b. Syringes: 0 to S0µL, 0 to 250µL, and 0-1 ml capacity fitted with blunt tipped needle. 7. Reagents a. Reagent water: Prepare blanks and standard solutions from carbon-free water; preferably use carbon filtered, redistilled water. b. Phosphoric acid, cone. Alternatively use sulfuric acid, but never use hydrochloric acid c. Organic carbon stock solution: A pre-made standard is purchased for the TOC analysis and diluted for use as working standards. A separate standard with a different lot number is diluted for use as check standard. e. Carrier gas: Purified oxygen or air, CO2-free and containing less than 1ppm hydrocarbon (as methane) f. Purging gas: any gas free of CO2 and hydrocarbons. I g I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I TOG 8. Procedure 8.1 Instrument Operation: ASV-I-105 April 2, 1997 Page 3 of 4 Follow manufacturer's instructions for operation of analyzer. 8.2 Sample Preparation: If a sample contains gross particles or inorganic matter, homogenize until a representative portion can be drawn through a syringe needle or autosampler tubing. 8.4 Standard curve preparation: Prepare an organic carbon standard series over the range of organic carbon concentrations in the samples. Inject standards and blanks and record analyzer's response. Before injecting samples, replicate injections of a midrange standard should be injected. If their measured concentrations differ by more than 10%, take corrective action, and reinject. 8.3 Sample injection: Withdraw a portion of prepared sample using a syringe fitted with a blunt-tipped needle. Select sample volume according to manufacturer's direction. Stir samples containing particulates with a magnetic stirrer. Select needle size consistent with sample particulate size. Inject samples and standards into analyzer according to manufacturer's directions and record response. 9. Calculations 9.1 Instrument printout will provide the analyst with concentrations. To these concentrations, a dilution factor, if any, should be applied. TOG 10. Quality Assurance/Quality Control ASV-I-105 April 2, 1997 Page 4 of 4 10.1 At least one sample in every 10 must be analyzed in duplicate. If less than 10 samples are analyzed per month, at least one sample each month should be analyzed in duplicate. For concentrations of 1-10mg/L, the duplicate acceptance range is± 0.39mg/L. For concentrations of 11-40mg/L, the range is ±4.2mg/L, and for concentrations greater than 40mg/L, the range is ±10% based on three times the standard deviation for the method. Any duplicates not meeting this criteria must be reanalyzed or a specific reason determined why the duplicate did not work. This reason must be written on the benchsheet. 10.2 A mid range reference standard must be analyzed every day samples are run and at least once for every 10 samples analyzed. It's value must be within ±10% of the true value. 10.3 A spike must be run with every ten samples, it's value should be within ±25% of it's actual value. 10.3 All data are reviewed by the inorganics department group leader, lab supervisor, and project manager prior to release. 11. References Standard Methods, 18th Edition, Method 5310-C. I I I I I I I I I I I I I I I I I I m m I I I I I I I I I I I I I I I I I Pace Analytical Pace Analytical Services, Inc. 9800 Kincey Avenue, Suite 100 Huntersville, NC 28078 Tel: 704-875-9092 Fax: 704-875-9091 STANDARD OPERA TING PROCEDURE Chloride Reference Method: EPA 325.3 SOP NUMBER: AUTHOR: NCI-I-003 Barney Kesler CONTROLLEt copv .. ,-\ COPY NO.~, G-(;_L) EFFECTIVE DATE: Februrary 28, I 997 Other chloride SOPs SUPERSEDES: APPROVAL Qu~ty Assurance Office a.c_ General Manager Date Chloride EPA325.3 NCI-1-003 File Name: Date: Page: NCII003 ·Feb. 28, 1997 2 of6 I. II. PURPOSE .The purpose of this Standard Operating Procedure (SOP) is to measure the concentration of chlorides in drinking, surface, and saline waters, domestic and industrial wastes .. SCOPE/APPLICATION A. B. C. Summary -An acidified sample is titrated with mercuric nitrate in the presence of mixed diphenylcarbozonebromophenol blue indicator. The end point of the titration is the formation of the blue-violet mercury diphenylcarbozone complex. Interferences I. Anions and cations at concentrations normally found in surface waters do not interfere. 2. Sulfite interference can be eliminated by oxidizing the l 00mLs of sample solution with 1.0 to 2.0 mL of hydrogen peroxide. Hazards and Precautions Each reagent and chemical compound used in this method should be treated as a potential health hazard. Reduce exposure by the use of gloves, lab coats, safety glasses and ventillation hoods. Material Safety Data Sheets(MSDSs) are on file in the library and available to all personnel. III. RESPONSIBILITY A. B. Analysts 2. Analysts are responsible for adherence to the SOP. All laboratory personnel are responsible for notifying the section supervisor/manager of any required revisions to the SOP. Department Manager 1. Managers are responsible for ensuring adherence to this SOP. 2. Managers are responsible for performing an annual review of the SOP. 6 I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Chloride EPA325.3 NCI-I-003 C. D. File Name: Date: Page: NCII003 Ft!b.28, 1997 3 of6 Quality assurance officer (QAO) I. The QAO is responsible for conducting laboratory audits to monitor adherence to this SOP. Results of the audit are rep·orted to Laboratory Management and Corporate Quality. 2. The QAO is responsible for coordinating annual reviews of this SOP with the regional director and the organics department manager. More frequent reviews may be required. 3. 4. The QAO is responsible for ensuring that all revisions to the SOP are implemented. The QAO is responsible for determining distribution of and maintaining document control of this SOP. General Manager (GM) 1. 2. The GM is responsible for the overall implementation of and adherence to this SOP. The GM is responsible for reviewing of this SOP with the QAO and the department manager. IV. REVISIONS/REVIEWS V. A. B. C. This SOP will be reviewed on an annual basis at a minimum, by the department manager, general manager and quality assurance officer. At the time of review, any required revision will be incorporated. The revised SOP will be distributed to all appropriate personnel and the superseded version replaced. · DISTRIBUTION A. Distribution of this SOP will be determined by the QAO. Chloride EPA325.3 NCI-I-003 File Name: Date: Page: NCII003 Feb. 28, 1997 4 of6 VI. APPARATUS AND MATERIALS VII. A. B. Apparatus I. !0mL Buret with 0.05mL increments 2. 250mL Erlenrnyer flasks 3. Magnetic Stirrer and magnetic stir bars 4. Transfer pipets Reagents / Standards I. Deionized Water {DI-H,O) 2. Standard Sodium Chloride (Prernade) -Lab Chern Inc. Cat. #LC13010-l. 3. LCS Spike Solution (Prernade) -I000pprn Chloride Standard, lmL = 1mg Cr, Lab Chern Inc. Cat. #LC13000-l. 4. Matrix Spike Solution (Prernade)-0.0141N Chloride Standard, lmL = 500pprn, RICCA Chemical Co. Cat. #1950-16. 5. 0. IN Nitric Acid {HNO,) Solution -3mLs HNO3 to 997mLs Dl-H2O. 6. 0. IN Sodium Hydroxide (NaOI:D Solution -!0g NaOH to I Liter Dl-H2O. · (Discard after 6 months) 7. Hydrogen Peroxide (30%), H2O2 -Fisher Cat. #MK-V340-500. 8. Mercuric Nitrate Titrant (0.14 IN) -Fisher Cat. #LC 16660-4, Standardize before each use. See section VII ~ A 9. Mercuric Nitrate Titrant (0.025N) -Fisher Cat. #LC!6657-2; Standardize before each use. See section VII -A 10. Mercuric Nitrate Titrant (0.014 IN) -Fisher Cat. #LC16650-4, Standardize before each use. See section VII -A. 11. Diphenylcarbazone Brornophenol Blue Mixed Indicator Solution - Lab Chern Inc. Cat. #LC13680-l PROCEDURE A. Standardization of Mercuric Nitrate Titrant 1. Add SmLs of0.0141N Standard Sodium Chloride Solution to 95mLs of DI-H2O in a 250mL Erlenmeyer flask 2. · Titrate following steps B, 1-7, omitting step B6. 3. Record mLs of titrant used calculate the normality of the titrant as follows: mLs of Standard N~CI (5) x Normality of NaCl {0.0141) = Normality ofHg(NO3)2 mLs ofHg(NO3)2 titrant used I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Chloride EPA325.3 NCI-I-003 B. C. File Name: Date: Page: Titration NCII003 Feb. 28, I 997 5 of6 I. Fill a buret with 0.0141N mercuric nitrate solution and zero the buret. 2. Measure I 00mLs of sample in a graduated cylander and transfer to a 250mL erlenmeyer flask. a. Method Blank -Use !00mLs ofDI-H2O. b. Matrix Spike -Dilute lmL of Matrix Spike solution to I00mLs of sample. Final concentration = 5 mg/L c. Laboratory Control Sample -Dilute 2 mLs of LCS solution to 100 mLs ofDI-H2O .. Final concentration= 20 mg/L d. Dilutions -Measure an aliquot of sample and dilute to IO0mLs with DI-H2O. 3. Add a magnetic stir bar and place on a magnetic stirrer place under the buret. Tum on the stir plate to a moderate speed. 4. Using a transfer pi pet add 5 to IO drops of mixed indicator solution to the sample. a. If a blue violet or red color appears on the addition of the mixed indicator, add 0. IN nitric acid one drop at a time until the color changes to yellow. b. If a yellow or orange color forms immediately on the addition of the mixed indicator, add 0. IN sodium hydroxide solution one drop at a time until the color changes to blue-violet: then add 0.1 N nitric acid solution one drop at a time until the color changes to yellow. 5. Add l.0mL excess 0. IN nitric acid. 6. To eliminate the interference of sulfite ions, add l.0mL of hydrogen peroxide (3 0%) solution to the sample and mix for I minute. 7. Titrate with the appropriate mercuric nitrate titrant (usually 0.0141N). Record in the logbook the mis in the buret before you start the titration. Add titrant to the sample until you get a dark purple color. Continue adding titrant one drop at a time until the color does not change. Record the mis in the buret at the endpoint. Record the total mLs used to titrate the sample and the adjusted mis (blank corrected). Calculation: where: mg chloride/ L = (A-B) N x 35450 mLs of sample A= ml of titrant used for sample B = ml of titrant used for blank N = normality of the mercuric nitrate titrant used Chloride EPA325.3 NCl-1-003 File Name: Date: Page: NC11003 Feb. 28, 1997 6of6 VIlI. QUALITY CONTROL A. B. C. D. E. F. Method Blank: To be run with each batch, Duplicate: A duplicate sample must be analyzed with each batch of samples. In addition, NC requires a duplicate every IO samples. The relative percent difference(RPD) must be ~20%. Special consideration must be made for concentrations less than ten times the reporting limit. RPO = Difference x I 00 Mean = {Cone. 1 -Cone. 2) x 100 (Cone. 1 + Cone. 2) / 2 Spike: A matrix spike must be analyzed with each batch of samples. The percent recovery must be within+/-15%. % Recovery = {Spiked Sample Value -Sample Value) x 100 True Value Laboratory Control Sample(LCS): To be analyzed with each batch of samples. The percent recovery of the LCS must be within 10% of the expected value. % Recovery = Reported Value x 100 True Value Sample Preservation: None required. Holding Time: 28 days D n I I I I I I I I I I I I I I I I I • I I I I I I I I I I I I I I I I I I Pace Analytical Pace Analytical Services. Inc. 9800 Kincey Avenue. Suite 100 Huntersville. NC 28078 Tel: 704-875-9092 Fax: 704-875-9091 STANDARD OPERATING PROCEDURE Sulfate -S04 Reference Method: EPA 375.4 SOP NUMBER: NCI-1-025 AUTHOR: Barney Kesler EFFECTIVE DATE: April I 5, 1997 Bench SOP SUPERSEDES: APPROVAL -cc,~ c.b. kl ~~ Operations Manager Quality Assurance-Officer C0NTROLILED COPY COPY N0 .. __,;;;;3;;:____ Date °'~dfC !~fr ~7 Oat Sulfate -S04 EPA375.4 NCI-I-025 File Name: Date: Page: NCII025 Apr. I 5, 1997 2 ofS I. Il. PURPOSE The purpose of this Standard Operating Procedure (SOP) is to measure the concentration of sulfate in groundwater, drinking water, surface waters, domestic and industrial wastes. SCOPE/APPLICATION A. B. C. D. Summary: Sulfate ion is converted to a barium sulfate suspension under controlled conditions. The resulting turbidity is by a nephelometer (Turbidimeter), filter photometer cir spectrophotometer and compared to a curve prepared from standard sulfate solutions. Concentration Ranges: This method is suitable for all concentration ranges of sulfate (SO4·2); however, in order to obtain reliable readings, use a sample aliquot containing not more than 40 mg SOJL. The minimum detectable limit is approximately I mg/L sulfate. Interferences: I. Suspended matter and color interfere. Correct by running blanks from which the barium chloride has been omitted. 2. Silica in concentrations over 500mg/L will interfere. Hazards and Precautions Each reagent and chemical used in this method should be treated as a potential health hazard. Reduce exposure by the use of gloves, lab coats, safety glasses and ventilation hoods. Material Safety Data Sheets(MSDSs) are on file in the library and available to all personnel. Ill. RESPONSIBILITY A. B. Analysts I. Analysts are responsible for adherence to the SOP. 2. All laboratory personnel are responsible for notifying the section supervisor/manager of any required revisions to the SOP. Operations Manager I. The operations manager is responsible for ensuring adherence to this SOP. D H I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Sulfate -S04 EPA375.4 NCI-I-025 File Name: Date: Page: NCII025 Apr. I 5, 1997 3 of5 IV. V. C. D. 2. The operations manager is responsible for performing an annual review of the SOP. Quality Assurance Officer (QAO) I. 2. 3. The QAO is responsible for conducting laboratory audits to monitor adherence to this SOP. Results of the audit are reported to Laboratory Management and Corporate Quality. The QAO is responsible for coordinating annual reviews of this SOP with the general manager and the operations manager. More freque:nt reviews may be required. The QAO is responsible for ensuring that all revisions to the SOP are implemented. 4. The QAO is responsible for determining distribution of and maintaining document control of this SOP. General Manager (GM) I. 2. The GM is responsible for the overall implementation of and adherence to this SOP. The GM is responsible for reviewing of this SOP with the QAO and the operations manager. REVISIONS/REVIEWS A. This SOP will be reviewed on an annual basis at a minimum, by the operations manager, general manager and quality assurance officer. B. At the time of review, any required revision will be incorporated. C. The revised SOP will be distributed to all appropriate personnel and the: superseded version replaced. DISTRIBUTION A. Distribution of this SOP will be determined by the QAO. Sulfate -SO4 EPA375.4 NCI-I-025 File Name: Date: Page: NCII025 Apr. 15, 1997 4 of5 VI. APP ARA TVS AND MATERIALS VII. A. . Apparatus/ Equipment I. Magnetic Stirrer, variable speed, use identical shape and size stir bars. 2. Photometer: one of the following which are given in the order of preference. a. Turbidimeter (nephelometer) b. Spectrophotometer for use at 420nm with light path of 4-5cm. c. Filter photometer with a violet filter having a miximum near 420 nm and a light path of 4 to 5cm. 3. Stopwatch 4. Beakers B. Reagents I. ASTM Type II Water (Deionized Water) 2. Hach Sulfaver Powder (Barium Chloride Mix) Proprietary reagents, such as Hach Sulfaver or equivalent, are acceptable. 3. Sulfate Calibration Standard @ I 000ppm, Ricca Cat.# 8 I 12° I 6 or equivalent. 4. Sulfate LCS Standard@ I00ppm (Alternate Source), Lab Chem Cat. # LC25 500-1 or equivalent. 5. Sulfate Spiking Standard@ I00ppm, Ricca Cat.# 8110-16 or equivalent. PROCEDURE A. B. _Standardize the trurbidity meter according to manufacturers specifications. Preparation of calibration standards: The calibration curve must be composed of a minimum of a blank and 3 standards. Standards should be in the range of 0-40 mg/L at 5 mg/L increments. (Note: Above 50mg/L, accuracy decreases and the suspensions lose stability.) Prepare calibration standards in DI-Water using the Ricca Sulfate solution at I 000ppm. Transfer 25 mLs of each standard to a beaker and add a small magnetic stirring bar. I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Sulfate -S04 EPA 375.4 NCl-I-025 File Name: Date: Page: NCII025 Apr. 15, 1997 5 of5 Calibration Volume of Cone. of Final Final Standard so. Solution SO. Solution Volume (mL) Cone. (ppm) C. D. E. F. G. H. I. J. K. 1 2 3 4 5 6 7 8 (mL) (oom) 0.25 1000 so 5 0.50 1000 so 10 0.75 1000 50 15 1.0 1000 so 20 1.25 1000 so 25 I.SO 1000 so 30 I. 75 1000 50 35 2.0 1000 so 40 Open pack of Hach Sulfaver Powder and add to each beaker. Begin timing for 1.0 min. while stirring at a constant speed. Transfer standards to a turbidity cell if using Turbidimeter or large absorbance cell if using spectrophotometer. Calibrate the instrument according to manufacturers specifications. Preparation of Samples: Place 25 mLs of sample into beaker and add a. small magnetic stir bar. Prepare QC samples as follows and add a stir bar.: -Method Blank -25 mL DI-Water -LCS -25 mL DI-Water+ 0.S0mL SO4 LCS Standard@ l00ppm (Final Cone.= 2 mg/L) -Matrix Spike -25 mL Sample+ lmL SO4 Spiking Standard@ I00ppm (Final Cone.= 4 mg/L) Open pack of Hach Sulfaver Powder and add to each beaker. Begin timing for I. 0 min. while stirring at a constant speed. Transfer samples to a turbidity cell if using Turbidimeter or large absorbance cell if using spectrophotometer. Measure turbidity at 30 second intervals for 4 minutes. Record the maximum reading obtained in the 4 minutes. Dilute samples if they are more concentrated than the highest standard Correction for sample color and turbidity. I. Run a sample blank using 6.3 without the addition of barium chloride mix (Hach powder) Calculation: I. Read from linear curve NTUS or absorbance vs. mg/L x dilution factor. Sulfate -SO4 EPA 375.4 NCl-I-025 File Name: Date: Page: NC1I025 Apr. 15, 1997 6 ofS VIII. QUALITY CONTROL A. Method Blank: To be run daily with each batch of samples. B. Duplicate: A duplicate sample must be analyzed with each batch of samples. In addition, NC requires a duplicate every 10 samples. The relative percent difference(RPD) must be ::,20%. Special consideration must be made for concentrations less than ten times the reporting limit. C. D. E. F. RPO = Difference x I 00 Mean = (Cone. I -Cone. 2) x I 00 (Cone. I + Cone. 2) / 2 Spike: A matrix spike must be analyzed every 20 samples for wastewater and every 10 samples for drinking water. The percent recovery must be within+/- 20% for wastewater and +/-I 0% for drinking water. % Recovery = (Spiked Sample Value -Sample Value) x 100 True Value Laboratory Control Sample(LCS): To be analyzed with each batch of samples. The percent recovery of the LCS must be within 10% of the expected value. % Recovery = Reported Value x 100 True Value Sample Preservation: Store cool@ 4°C. Holding Time: 28 days. n I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Pace Analytical Pace Analytical Services. Inc. 9800 Kincey Avenue. Suite 100 Huntersville. NC 28078 Tel: 704•875•9092 . Fax: 704-875-9091 STANDARD OPERATING PROCEDURE Nitrate -Cadmium Reduction (Manual and Automated) Reference Methods: 18th Edition SM 4500-NO3 E. & F. / EPA 353.3 & 353.2 SOP NUMBER: AlITHOR: EFFECTIVE DATE: SUPERSEDES: NCl-1-016 D.C. Coffey April I 5, I 997 Draft SOP APPROVAL -1'1~\ij /~~ Operations Manager J · ty Assurance Officer i a, c ______ General Manager CONTROLLEB, COPY COPY NO.__,-j-'---- Nitrate -NO3 File Name: NCII0!6 Apr. 15, 1997 2 of7 SM 4500-NO3 E. & F. / 353.3 & 353.2 NCI-I-016 Date: Page: I. II. ill. PURPOSE The purpose of this Standard Operating Procedure (SOP) is to measure the amount of nitrite and nitrate combined in drinking waters, surface waters, domestic and industrial wastes. SCOPE/APPLICATION A. Summary -A filtered sample is passed through a column containing granulated copper-cadmium to reduce nitrate (NO3") to nitrite (NO 2"). The combination of nitrite and reduced nitrate is determined by diazotizing with sulfanilamide and coupling with N-(1-naphthyl)-ethylenediamine dihydrochloride to form a highly colored azo dye which is measured colorimetrically. A correction may be made for any nitrite present in the sample by analyzing the sample without the reduction step. (See SOP for Nitrite.) B. Interferences: I. Suspended matter in the column will restrict sample flow. Samples may be filtered prior to analysis. 2. Concentrations of iron, copper and other metals above several milligrams per liter lower reduction efficiency. Add EDT A to samples to eliminate the interference from metals present in the sample. 3. Oil and grease will coat the cadmium surface. Remove by pre-extraction with an organic solvent. 4. Sample color that absorbs at ~540 nm interferes. Diluting the sample may resolve the interference. C. Hazards and Precautions Each reagent and chemical used in this procedure should be treated as a potential health hazard. Reduce exposure by the use of gloves, lab coats, safety glasses and ventilation hoods. Material Safety Data Sheets(MSDSs) are ori file in the library and available to all personnel. RESPONSIBILITY A. Analysts I. Analysts are responsible for adherence to the SOP. D u I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Nitrate -N03 .File Name: NC1I016 Apr. 15, 1997 3 of7 SM 4500-NO3 E. & F. / 353.3 & 353.2 NCI-I-016 Date: Page: IV. V. 2. All laboratory personnel are responsible for notifying the section supervisor or operations manager of any required revisions to the SOP. B. Operations Manager C. D. I. 2. The Operations manager is responsible for ensuring adherence to this SOP. The Operations manager is responsible for performing an annual review of the SOP. Quality assurance officer (QAO) I. The Q AO is responsible for conducting laboratory audits to monitor adherence to this SOP. Results of the audit are reported to Laboratory Management and Corporate Quality. 2. The QAO is responsible for coordinating annual reviews of this SOP with the general manager and the operations manager. More frequent reviews may be required. 3. The QAO is responsible for ensuring that all revisions to the SOP are implemented. 4. The QAO is responsible for determining distribution of and maintaining document control of this SOP. General Manager (GM) 2. The GM is responsible for the overall implementation of and adherence to this SOP. The GM is responsible for reviewing of this SOP with the QAO and the department manager. REVISIONS/REVIEWS A This SOP will be reviewed on an annual basis at a minimum, by the operations manager, general manager and quality assurance officer. B. At the time of review, any required revision will be incorporated. C. The revised SOP will be distributed to all appropriate personnel and the superseded version replaced. DISTRIBUTION A. Distribution of this SOP will be determined by the QAO. Nitrate -NO3 File Name: NCII016 Apr. 15, 1997 4 of7 SM 4500-NO3 E. & F. / 353.3 & 353.2 NCI-I-016 Date: Page: VI. VII. APPARATUS AND MATERIALS A. Equipment I. Cadmium Reduction column: Prepurchased 2. Colorimetric Equipment: a. Manual Method -Spectrophotometer for use at 543nm, providing a light path of I cm or longer. b. Aut_omated Method -Quick Chem AE LACHAT Automated Analyzer for use at 520 nm., with computer, printer, autosampler, and pump with tubing. B. Reagents I. Nitrate-free DI-Water: The absorbance ofa reagent blank prepared with this water should not exceed 0.01. Use for all solutions and dilutions. 2. Copper-cadmium granules: Wash 25 g Cd granules with 6N HCl and rinse with water. Swirl Cd with I 00 mL 2% CuSO. solution for 5 min or until blue color partially fades. Decant and repeat with fresh CuSO. until a brown precipitate begins to develop. Gently flush with water to remove all precipitated Cu. 3. Color reagent: To 800 mL of DI-Water add 100 mL 85% phosphoric acid and IO g sulfanilamide. After dissolving sulfanilamide completely, add g N-(I -naphthyl) -ethylenediamine dihydrochloride (N.E.D.). Mix to dissolve and dilute to I liter with DI-Water. Solution is stable for about a 4. 5. 6. 7. 8. 9. 10. month when stored in amber bottle in refrigerator. Ammonia chloride (NfuCl} -EDTA solution: Dissolve 13 g NH.Cl and 1.7 g disodium ethlenediamine tetraacetate in 900 mL DI-Water. Adjust to pH 8.5 with cone. NH.OH and dilute to I liter. Dilute NH.Cl -EDT A solution: Dilute 300 mL NH.CL-EDT A solution to 500 mL with DI-Water. Hydrochloric Acid, HCl -6N: Carefully dilute 500 mL cone. HCl to 500 mL with DI-Water. Copper sulfate solution, 2%: Dissolve 20 g CuSO,.5H20 in 500 mL DI- Water and dilute to I liter. · Stock Nitrate Solutions, JOO & 1000 mg/L. Calibration solution, 5 mg/L : Dilute 0. 5 mL ( 1000 mg/L) stock to I 00 mL with DI-Water. LCS/Spike Solution, 5 mg/L: Dilute 5 mL (100mg/L) stock to 100 mL with DI-Water. PROCEDURE A. Preparation of reduction column: n I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Nitrate -NO3 file Name: NC1IOl6 SM 4500-NO3 E. & F. / 353.3 & 353.2 NCl-I-016 Date: Page: Apr. 15, I 997 . 5 of7 B. C. I. 2. Manual Method: Insert a glass wool plug into bottom of reducl:ion column and fill with water. Add sufficient Cu-Cd granules to produce a. column 18.5 cm long for the manual method. Maintain water level above Cu-Cd granules to prevent entrapment of air. Wash column with 200 mLs dilute NH4Cl-EDTA solution. Activate column by passing at least 100 mL ofa solution composed of25% 1.0 mg/L nitrate solution and 75% NH4CJ - EDT A solution. Automated Method: Insert a glass wool plug into the bottom of the LACHAT reduction column and fill with water. Add sufficient Cu Cd granules to fill the column. Fill the reagent reservoirs on the LACHAT with appropriate reagent solutions. Set up the LACHAT system to automatically wash and activate the column. Preparation of Calibration standards: Prepare a series of calibration standards by using the calibration solution @ 5 mg/L. Std. Cone. 0.05 mg/L 0.10 mg/L 0.20 mg/L 0.50 mg/L 1.0 mg/L Final Volume 25 mL 25 mL 25 mL 25 mL 25 mL mLs of Cal. Solution 0.25 mL (250 uL) 0.5 mL (500 uL) 1.0 mL (1000 uL) 2.5 mL (2500 uL) 5.0 mL (5000 uL) Treatment of sample: I. 2. 3. Turbidity Removal -Colorimetric methods require an optically clear sample. Filter turbid samples through 0.45 um pore diameter membrane filter. pH Adjustment -Adjust pH of the samples to between 7 and 9, as necessary, using dilute HCL or NaOH. Sample Volume: Method Blank -25 mLs DI-Water LCS -25 mLs DI-Water+ 0.5 mLs ofLCS/Spike Solution Spike -25mLs Sample+ 0.5 mLs ofLCS/Spike Solution Samples -25 mLs Sample D. Sample Reduction and Color Development: I. Manual Method -To a 25 mL portion of sample or a portion diluted to 25 mL, add 75 mL NH.Cl -EDTA solution and mix. Pour mixed sample through column discarding first 25 mL. Collect the rest in original sample Nitrate -NO3 File Name: SM 4500-NO3 E. & F. / 353.3 & 353.2 NCI-I-016 Date: Page: NCII016 Apr. 15, 1997 6 of7 E. 2. flask. There is no need to rinse column between samples, but if column is not to be reused for several hours or longer, pour 50 mL dilute NH4C I - EDTA solution on top and let pass through the system. Store Cu -Cd column in this solution and never let it dry. After the sample has passed through column, add 2 mLs color reagent to 50 mLs of sample and mix. After IO min measure absorbance Automated Method -Transfer a portion of standards and samples, as measured in step 3, to a LACHAT sample cell. The LACHAT system will automatically perform the sample reduction and color development procedures. Measurement: Obtain a standard curve by plotting the absorbance vs. concentration. VIII. QUALITY CONTROL A. Method Blank: Analyze one daily. B. C. D. E. Spike: A matrix spike must be analyzed every 20 samples for wastewater and every IO samples for drinking water. The percent recovery must be within +/- 20% for wastewater and +/-10% for drinking water. % Recovery = (Spiked Sample Value -Sample Value) x 100 True Value Laboratory Control Sample(LCS): The percent recovery of the LCS must be within I 0% of the expected value. Analyze one LCS daily and then every 20 samples. % Recovery = Reported Value x 100 True Value Duplicate: Analyze a sample duplicate or matrix spike duplicate every I 0 samples. The relative percent difference(RPD) must be .:::20%. Special consideration must be made for concentrations less than ten times the reporting limit. RPO = Difference x I 00 Mean = (Cone. I -Cone. 2) x I 00 (Cone. I + Cone. 2) I 2 Sample Preservation: Cool, 4°C, H2SO, to pH<2 I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Nitrate -NO3 .File Name: SM 4500-NO3 E. & F. / 353.3 & 353.2 NCl-1-016 Date: Page: NC11016 Apr. l 5, l 997 7 of7 F. G. Holding Time: 48 hours if unpreserved, 28 days if preserved with H2S0,. Column Efficiency: Analyze a 0.1 mg/L NO2-N Standard and a 0.1 mg/L NO3-N Standard that have passed through the reduction column. Calculate the column efficiency as indicated below. The column needs repacking if the effici,:ncy falls below 75%. Column Efficiency = NOi-N Results x 100 NOi-N Results I I I I I I I I I I I I I I I I I I I Pace Analytical Pace Analytical Se1Vices. Im:. 54 Ravensao~ Drive AsneviUe. NC 28801 STANDARD OPERATING PROCEDURE Tel: 704-254-7176 Fax: 704-2S2-4618 Alkalinity Reference Methods: EPA 310.1 I SM 2320 B. SOP Number Author Effective Date Supersedes Quality Assurance Officer General Manager Approval . ASV-I-106 Robin K Blankenbaker April 18, 1997 All previous editions Date Date Alkalinity I. PURPOSE 04-24-97 09:lSA P,03 ASV-I-106 April 18, 1997 Page 1 of 5 The purpose of this Standard Operating Procedure (SOP) is to measure the alkalinity of groundwater, drinking water, surface waters, domestic and industrial wastes. II. SCOPE/APPLICATION A. Summary: An unaltered sample is titrated to an electrometrically determined endpoint of pH 4.5. The sample must not be filtered. diluted. concentrated, or altered in any way. 8. Concentration Ranges: This method is suitable for all concentration ranges of alkalinity; however, appropriate aliquots should be used to avoid a titration volume greater than 50ml. C. Interferences: 1. Salts of weak organic and inorganic acids. if present in large amounts, may cause interference in the pH measurement. 2. Oil and grease, by coating the pH electrode, may also interfere, causing sluggish response. D. Safety: Safety glasses, lab coats, and gloves should be worn at all limes. As sample constituents are not known, all samples should be treated as hazardous. MSDSs are on file and available to all personnel. Ill. RESPONSIBILITY A. Analysts 1. Analysts are responsible for adherence to the SOP. 2. All laboratory personnel are responsible for notifying the section supervisor/ manager of any required revisions to the SOP. B. Operations Manager 1. The operations manager is responsible for ensuring adherence to this SOP. g I I I I I I I I I I I I I I I I 04-24-97 09:1aA P.04 ASV-I-106 April 18, 1997 I -A=lk=a=l=in=it,;..y====-==~=~-=~~=====P=a=g"-e~2_o_f_:i~~~ I I I I I I I I I I I I I I I 2. The operations manager is responsible for performing an annual review of the SOP. C. Quality Assurance Officer (CAO) 1. The QAO is responsible for conducting laboratory audits to mo11itor adherence to this SOP. Results of the audit are reported to Laboratory Management and Corporate Quality. 2. The QAO is responsible for coordinating annual reviews of this SOP with the general manager and the operations manager. More frequent reviews may be required. 3. The QAO is responsible for determining distribution of and maintaining document control of this SOP. 4. The QAO is responsible for determining distribution of and maintaining document control of this SOP. D. General Manager (GM) 1. The GM is responsible for the overall implementation of and adherence to this SOP. 2. The GM is responsible for reviewing this SOP with the QAO and the operations manager. IV, REVISIONS/REVIEWS A. This SOP will be reviewed on an annual basis at a minimum, by the operations manager, general manager and quality assurance officer. B. At the time of review, any required revision will be incorporated. C. The revised SOP will be distributed to all appropriate personnel and the superseded version replaced. V. DISTRIBUTION A. Distribution of this SOP will be determined by the QAO. Alkalinity VI. Apparatus ASV-I-106 April 18, 1997 Page 3 of 5 04-24-97 09:t8A P,05 A. A pH meter that can be read lo 0.05 pH units. Standardize and calibrate according lo manufacturer's instructions. If automatic temperature compensation is not provided, make titration at 25 ±2C. B. A vessel of appropriate size to keep the air space above the solution at a minimum. C. Magnetic stirrer (Teflon coated), pipettes, flasks and other standard laboratory equipment. 0. Burets. Pyrex 50, 25, and 10ml VII. Reagents A. Sodium carbonate solution, approximately 0.0SN. B. Sulfuric acid 0.1 N. C. Standard acid (sulfuric or hydrochloric}, 0.02N. I I g I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I I Alkalinity VIII. Procedure A. Sample Size: ASV-I-106 April 18, 1997 Page 4 of 5 a. Use a sufficiently large volume of titrant (>20ml in a 50ml buret) to obtain good precision while keeping volume low enough to permit sharp end point. b. For< 1000mg CaC03/L use 0.02N titrant c. A preliminary titration is helpful. B. Place 100ml of sample in flask by pouring near bottom of flask. C. Measure pH of sample D. Titrate H2SO• into sample until pH is 4.5. Record volume of titrant. IX. Calculations Alkalinity, mg/L CaCOJ =Ax N x 50,000 ml of sample Where: A = ml standard acid N = normality standard acid X. QUALITY CONTROL A. Method Blank: To be run daily with each batch of samples. B. Duplicate: A duplicate sample must be analyzed with each batch of samples. In addition, NC requires a duplicate every 10 samples. The relative percent difference (RPO) must be ~20%. Special consideration must be made for concentrations less than ten times the reporting limit. RPO = Difference x 100 = (Cone. 1 -Cone. 2) x 100 Mean (Cone 1 + Cone. 2) / 2 Alkalinity ASV-I-106 April 18, 1997 Page 5 of 5 04-24-97 09;19A P,07 C. Laboratory Control Sample (LCS): To be analyzed with each batch of samples. The percent recovery of the LCS must be within 10% of the expected value. %Recovery = Reported Value x 100 True Value D. Sample Preservation: Store cool @ 4C. E. Holding Time: 14 days. F. References: EPA 310.1 and SM 2320 B I n I u I I I I I I I I I I I I I I I