HomeMy WebLinkAbout6301_Moore_MSWLF_CDLF_April20LFG_FID1395726_202004014WMIAlln
C DENR " Environmental Monitoring
Division of Waste Management - Solid Waste Reporting Form
Notice: This form and any information attached to it are "Public Records" as defined in NC General Statute 132-1. As such, these documents are
available for inspection and examination by any person upon request (NC General Statute 132-6).
Instructions:
• Prepare one form for each individually monitored unlL
• Please type or print legibly.
Attach a notification table with values that attain or exceed NC 2L groundwater standards or NC 2B surface water standards. The notification
must include a preliminary analysis of the cause and significance of each value. (e.g. naturally occurring, off -site source, preexisting
condition, etc.).
• Attach a notification table of any groundwater or surface water values that equal or exceed the reporting limits.
• Attach a notification table of any methane gas values that attain or exceed explosive gas levels. This Includes any structures on or nearby the
facility (NCAC 138 .1629 (4)(a)(i).
Send the original signed and sealed form, any tables, and Electronic Data Deliverable to: Compliance Unit, NCDENR-DWM, Solid Waste
Section, 1646 Mail Service Center, Raleigh, NC 27699-1646.
owner):
Golder Associates INC, Inc. (on behalf of Moore County)
Contact for questions about date formatting. Include data preparers name, telephone number and E-mail address:
Name: Benjamin S. Draper, PG Phone: 336-862-1903
E-mail: bdreper@golder.cem
NC Landfill Rute: Actual sampling dates (e.g.,
Facility name: Facility Address: Facility Permit# (.0500 or.1600) October 20-24, 2006)
456 Moore County Landfill TurningLeaf 63-pt .0500
Aberdeen,en,f WFApnI72020
NC of
Environmental Status: (Check all that apply)
❑ Ini iallBadiground Monitoring ❑ Detection Monitoring
of data submitted: (Check all that apply)
Groundwater monitoring data from monitoring wells
Groundwater monitoring data from private water supply wells
Leachate monitoring data
Surface water monitoring data
❑ Assessment Monitoring XX Corrective Action
QX
Methane gas monitoring data
Corrective action data (specify)
Other(specify)
NotMeation attached?
eNo. No groundwater or surface water standards were exceeded.
Yes, a notification of values exceeding a groundwater or surface water standard is attached. It includes a list of groundwater and surface water
monitoring points, dales, analytical values, NC 2L groundwater standard, NC 28 surface water standard or NC Solid Waste GWPS and
preliminary analysis of the cause and significance of any concentration.
Yes, a notification of values exceeding an explosive methane gas limit is attached. It includes the methane monitoring points, dates, sample
values and explosive methane gas limits.
To the best of my knowledge, the information reported and statements made on this data submittal and attachments are true and COrrGCL
Furthermore, I have attached complete notification of any sampling values meeting or exceeding groundwater standards or explosive gas
levels, and a preliminary analysis of the cause and significance of concentrations exceeding groundwater standards. I am aware that there
are significant penalties for making any false statement, representation, or certification including the possibility of a fine and imprisonment
Benjamin S. Draper, PG
Facility Repres a
Senior Project Geologist
eY
Date
Golder Associates NC, Inc., 5B Oak Branch Drive, Greensboro, NC 27407
Facility Representative Address
PE C-2862; PG C-399
NC PE Firm License Number (d applicable effective May 1, 2009)
336-8524903
(Area Code) Telepha
Affix NC
t1
G A j�Q Geologist Seal
0
GA1".
L
s
231 tz
p�OG1g z
Revised 6/2009
NC Division of Waste Management - Solid Waste Section
Landfill Gas Monitoring Data Form
Notice: Thisform andany information attached to it are "Public Records"as dined in NC Genera! Statute 132-1. As such, these documents are
available for inspection mid examination by am' person upon request (SC General Statute 132-6).
FacilityName: Moore County Landfill
Sampling Date: 04/07/2020 N
Sample Collector Name & Position: Nicolas T
Gas Meter Type & Serial Number:
Field Calibration Date & Time:
GEM 5000
Number: 63-01
Landfill Rule (.0500 or.1600):
eda-Torres
04/07/2020 11:35
0500
Gas Meter Calibration Date:
03/31 /2020
Field Calibration Gas Type (115/15 5/I5 or 35150): Field Calibration Gas Canister Expiration Date: 01/06/2021
Gas Meter Pump Rate:
300 cc/min
Ambient Air Temperature: 77 F Barometric Pressure (in. or nun 29,79 Hg): Weather Conditions: Fair, winds SW at 5mph
Instructions: Under "Location or LFG Well", list monitoring well # or describe monitoring location (e.g., inside field office).
Attach a test location may or drawing. Report methane readings as both % LEL and % CH4 by volume.
Convert % CH4 (by volume) to % LEL as follows: % methane (by volume)/20 = % LEL.
"Hydrogen Sulfide (H2S) gas monitoring may be required for Construction & Demolition Landfills (CDLFs). See
individual permit conditions and/or Facility LFG monitoring plan.
Location or
LFG Well iD
Sample
Tube
Purge
Time of
Day
Time
Pum
(sac)
,:)
Initial
% LEL
Stabilized
%LEL
%CH4
(volume)
op
o 02
(volume)
v CO_
(volume)
oa
v lum
(volume)
NOTES
GP-9
60
13:29
60
0
0
0.0
20.5
0.2
0
1 Inch Well
GP-10
60
13:22
60
374
374
32.2
3.9
11.1
16
1 Inch Well
GP-11
60 1
13:36
60
0
0
0.0
11.1
6.1
0
1 Inch Well
GP-12
60
13:44
60
0
0
0.0
20.3
0.3
0
1 1 Inch Well
GP-13
60
12:13
60
0
0
0.0
19.1
0.6
0
1 Inch Well
GP-13R
60
12:17
60
0
0
0.0
19.0
0.9
0
1 Inch Well
GP-14
60
12:26
60
0
0
0.0
14.6
5.1
0
1 Inch Well
GP-14R
60
12:21
60
0
0
0.0
18.9
0.4
0
1 Inch Well
GP-15
60
12:45
60
0
0
0.0
9.1
5.5
0
1 Inch Well
GP-15R
60
12:33
60
0
0
0.0
18.7
0.5
0
1 Inch Well
GP-16
60
12:51
60
0
0
0.0
15.9
4.1
0
1 Inch Well
GP-17
60
13:50
60
0
0
0.4
0.9
13.3
0
1 Inch Well
,VOTE: !f needed, attach addiliowl dam forms to include additional LFG monitoring data lowitors jar the facility.
V UION Lt:v ELti: Methane: I._>",b by volume (inside structures) AND ",, by volume (at facility boundary)
Hydrogen Sulfide: - by volume (inside structures) AND _- by volume (at facility boundary)
Certification
To the best of my knowledge, the information reported and statements made on this data submittal and attachments
are true and correct. I am aware that there are significant penalties for making any false statement, representation,
or Bert' catiog.'mc •ng the possibility of a fine and imprisonment.
Geologist
SIGNATURE Njcot�f•s T6Ta0w-rattc5 TITLE
Revised -March 6, 2017
NC Division of Waste Management - Solid Waste Section
Landfill Gas Monitoring Data Form
Notice: This form mid mry infornmlion mtac)wd io it are "Public Remrds" as dined in NC General Statute 132-1. As such these documents are
available for inspecifon and exanrinarion by any person upon request (NC General.Statute 132-6).
Facility Name: Moore County Landfill
Number: 63-01
Sampling Date: 4/07/2020 NC Landfill Rule (.0500 or.1600):
Sample Collector Name & Position: Nicolas Tejeda-Torres
Gas Meter Type & Serial Number: GEM 5000
Field Calibration Date & Time:
04/07/2020 11:35
0500
Gas Meter Calibration Date:
03/31 /2020
Field Calibration Gas Type (15/15 or 35/50): 15/15 Field Calibration Gas Canister Expiration Date: 2/21 /21
Gas Meter Pump Rate:
300 cc/min
Ambient Air Temperature: 77 F Barometric Pressure (in. or rum 29.79 Hg): Weather Conditions: Fair, winds SW at 5mph
Instructions: Under "Location or LFG Well", list monitoring well # or describe monitoring location (e.g., inside field office).
Attach a test location map or drawing. Report methane readings as both % LEL and % CH4 by volume.
Convert % CH4 (by volume) to % LEL as follows: % methane (by volume)/20 = % LEL.
"Hydrogen Suy7de (H2S) gas monitoring may be required.for Construction &. Demolition Landfills (CDLFs). See
individual permit conditions and/or Facility LFG monitoring plan.
Location or
LFG Well ID
Sample
Tube
Purge
Time of
Day
Time
Pumped
(sec)
Initial
% �-
Stabilized
% LEL
% CH4
(volume)
r o
4o O_
(volume)
v Cme
(volume)
o, o •�
/ (volume)
(volume)
NOTES
MW-6
60
16:36
60
238
238
23.6
10.2
11.7
0
2 Inch Groundwater Well
MW-15
60
13:04
60
562
562
59.9
0.2
33.3
0
cvaa fees vam)
60
13:09
60
718
718
42.0
0.9
33.5
1
Gas Vent
ov-saiweam,v"m,
60
13:11
60
0
0
9.0
15.4
6.9
1
Gas Vent
SCALEHOUSE
60
12:00
90
0
0
0.0
19.8
0.0
0
Building
Mariumwce Shop
60
12:08
90
0
0
0.0
19.7
0.0
0
Building
Transfer station
60
12:04
90
0
0
0.0
19.7
0.0
0
Building
60
16:36
90
0
0
0.0
20.8
0.0
0
Building
The Landtec GEM 5000 has a
CH4 accuracy range of+/- 0.3 %
NOTE: Ifneeded,, attach additional dala forms is include additional LFG monitoring data lowllonsfor thefacilin'.
• +� .! v I Methane: r .]> by volume (insidc structures) AND ->` .by volume (at facility boundary)
Hydrogen Sulfide: by volume (inside structures) AND by volume (at facility boundary)
Certification
To the best of my knowledge, the information reported and statements made on this data submittal and attachments
are true and correct. I am aware that there are significant penalties for making any false statement, representation,
or certific 'on d't mcluhe possibility of a fine and imprisonment.
2 Geologist
p re
GNATUAE r/d,eeL.t•f TITLE
Revised —March 6, 2017
LEGEND
SW-2 [']�
l�\ DENSE GETATION
,I
DENSE VEGETATION
I
\ DENSE VEGETATION
I`
DENSE GETATION
GROUND OBSCURED
DENSE VEGETATION
SEE ACCURACY NOTE
\ DE VEGETATION
GROUND OBSCURED
SEE ACCURACY NOTE
SITE LOCATION MAP
NOT TO SCALE
V LILIV 1
MOORE COUNTY
wall vuvjon■ 444SIIO1G91W,
rr«rmmEu
GO L D E R DESIGN
REVIEW
DENSE
�
�`VEGETATI
DENSE
C—
STREAM, POND, AND WETLAND LIMITS
EXISTING 10 FT GROUND SURFACE CONTOUR
EXISTING 2 FT GROUND SURFACE CONTOUR
PROPERTY LINE
APPROXIMATE LIMITS OF WASTE
EXISTING ROAD
385 POTENTIOMETRIC GROUNDWATER CONTOURS (5 FT. INTERVAL)
GROUNDWATER FLOW SEGMENTS
MW-1 MONITORING WELL AND IDENTIFICATION
'380.94
GP-10 METHANE PROBE AND IDENTIFICATION
❑ SW-1 SURFACE WATER MONITORING POINT AND IDENTIFICATION
LANDFILL GAS TRENCH VENTS AND IDENTIFICATION
m PASSIVE LANDFILL GAS VENTS
0 PW-1 PRIVATE WELL LOCATION AND IDENTIFICATION
4 WETLAND AREA
SB-2 SOIL SAMPLE LOCATIONS
PZ-2 PIEZOMETERS
NOTES
1. TOPOGRAPHIC CONTOUR INTERVAL = 2 FEET
2. GROUNDWATER SURFACE CONTOUR INTERVAL = 5 FEET
3. GROUNDWATER ELEVATIONS MEASURED ON NOVEMBER 19, 2019.
4. GROUNDWATER CONTOURS BASED ON LINEAR INTERPOLATION BETWEEN AND
EXTRAPOLATION FROM KNOWN DATA, TOPOGRAPHIC CONTOURS, AND KNOWN FIELD
CONDITIONS. THEREFORE, GROUNDWATER CONTOURS MAY NOT REFLECT ACTUAL
CONDITIONS.
5. GROUNDWATER CONTOUR LINES SHOW THE WATER TABLE SHAPE AND ELEVATION. THESE
CONTOURS ARE INFERRED LINES FOLLOWING THE GROUNDWATER SURFACE AT A
CONSTANT ELEVATION ABOVE SEA LEVEL. THE GROUNDWATER FLOW DIRECTION IS
GENERALLY PERPENDICULAR TO THE GROUNDWATER SURFACE CONTOURS, SIMILAR TO THE
RELATIONSHIP BETWEEN SURFACE WATER FLOW AND TOPOGRAPHIC CONTOURS.
6. PROPERTY BOUNDARY SURVEY BY JAMES L. WRIGHT DATED NOVEMBER 1984. ITS LOCATION
IS RELATIVE TO TOPOGRAPHY APPROXIMATE BY HDR ENGINEERING, INC.
7. EXISTING TOPOGRAPHY WITHIN ACTIVE AND PROPOSED LANDFILL PROVIDED BY MATTHEWS
LAND SURVEYING & MAPPING, PLLC DATED APRIL 17, 2015 AND JUNE 2018.
8. MONITORING WELLS MW-1 THROUGH MW-5 SURVEYED JUNE 1987 BY S&ME, INC. MONITORING
WELLS MW-6 THROUGH MW-16 SURVEYED APRIL 1995, SEPTEMBER 1996, AND APRIL 2003 BY
HDR ENGINEERING INC./ ED BUCKNER RLS APRIL 17, 2015.
9. THE LOCATIONS OF LANDFILL GAS TRENCH VENTS ARE APPROXIMATE.
10. GROUNDWATER WELLS MW-11 DR, MW-13D, AND MW-17D ARE NOT SCREENED IN THE
SHALLOW AQUIFER AND ELEVATION DATA WERE NOT USED IN THE CONSTRUCTION OF
GROUNDWATER CONTOURS.
0 200 400
SCALE FEET
PROJECT
MOORE COUNTY LANDFILL
NC SOLID WASTE PERMIT NO. 63-01
ABERDEEN, MOORE COUNTY, NORTH CAROLINA
TITLE
GROUNDWATER SURFACE CONTOUR MAP
NOVEMBER 19, 2019
PROJECT No.
PHASE
Rev.
DRAWING
19127453
100
0
it
II'
11
APPROVED RPK
ENVIRO—EQUIPMENT, INC..
LANDTEC GEM 5000 Plus CALIBRATION DATA SHEET
Serial Number (�SQ-1637 Date 3J& O Inspector
Time/Date Battery Level Memory Cleared '� Pump Faults
Gas Alarms: CHa = 100.0% CO2 = 100.0% OZ = 22.0%
Zero Transducers v Pump Flow Rate _ 7 Vacuum Tested To 80" H2O !�
Calibration Check (Performed at Room Temperature)
Calibration Gas
Calibration Gas
Lot Number
Expiration
Reading
Accuracy*,
Concentration
Pure NZ
% (�
O,. c�
Q(� /
1 I• !�
High CHy
So/
l l9�'68S
_�
Se yo
t2
High CO,
3 $°l�
u 49 b$S
9
�
f 2%%
Low CHI
L s y�
►`-� ti a
! s lb
2%
Low CO2
t 2%
�T^
t 2%
o
High OZ
20.9%
N/A ,
N/A
a o .9 l�_
t 10%
co
5
19aga7
10%
H2S
'Accuracy depends upon calibration range; 15%-FS range for CHa & CO2,
0-5% volume for O2, 0%-FS for H2S, and 10- of reading
for CO.
INSPECTOR'S MAINTENANCE NOTES
EEI GEM 2000 Plus Cal Form 01/12