METHOD 1312
SYNTHETIC PRECI.P.I.TATION LEACHING PROCEDURE
1.0 SCOPE AND APPLICATION
1.1 Method 1312 is designed to determine the mobility of both organic
and inorganic analytes present in liquids, soils, and wastes.
2.0 SUMMARY OF METHOD
2.1 For liquid samples (i.e., those containing less than 0.5 % dry
solid material), the sample, after filtration through a 0.6 to 0.8 um glass fiber
filter, is defined as the 1312 extract.
2.2 For samples containing greater than 0.5 % solids, the liquid phase,
if any, is separated from the solid phase and stored for later analysis; the
particle size of the solid phase is reduced, if necessary. The solid phase is
extracted with an amount of extraction fluid equal to 20 times the weight of the
solid phase. The extraction fluid employed is a function of the region of the
country where the sample site is located if the sample is a soil. If the sample
is a waste or wastewater, the extraction fluid employed is a pH 4.2 solution.
A special extractor vessel is used when testing for volatile analytes (see Table
1 for a list of volatile compounds). Following extraction, the liquid extract
is separated from the solid phase by filtration through a 0.6 to 0.8 um glass
fiber filter.
2.3 If compatible (i.e., multiple phases will not form on combination),
the initial liquid phase of the waste is added to the liquid extract, and these
are analyzed together. If incompatible, the liquids are analyzed separately and
the results are mathematically combined to yield a volume -weighted average
concentration.
3.0 INTERFERENCES
3.1 Potential interferences that may be encountered during analysis are
discussed in the individual analytical methods.
4.0 APPARATUS AND MATERIALS
4.1 Agitation apparatus: The agitation apparatus must be capable of
rotating the extraction vessel in an end -over -end fashion (see Figure 1) at 30
+ 2 rpm. Suitable devices known to EPA are identified in Table 2.
4.2 Extraction Vessels
4.2.1 Zero Headspace Extraction Vessel (ZHE). This device is for
use only when the sample is being tested for the mobility of volatile
analytes (i.e., those listed in Table 1). The ZHE (depicted in Figure 2)
allows for liquid/solid separation within the device and effectively
precludes headspace. This type of vessel allows for initial liquid/solid
CD-ROM 1312 - 1 Revision 0
September 1994
DEQ-CFW 00059923
separation, extraction, and final extract filtration without opening the
vessel (see Step 4.3.1). These vessels shall have an internal volume of
500-600 mL and be equipped to accommodate a 90-110 mm filter. The devices
contain VITON01 0-rings which should be replaced frequently. Suitable ZHE
devices known to EPA are identified in Table 3.
For the ZHE to be acceptable for use, the piston within the ZHE
should be able to be moved with approximately 15 psig or less. If it
takes more pressure to move the piston, the 0-rings in the device should
be replaced. If this does not solve the problem, the ZHE is unacceptable
for 1312 analyses and the manufacturer should be contacted.
The ZHE should be checked for leaks after every extraction. If the
device contains a built-in pressure gauge, pressurize the device to 50
psig, allow it to stand unattended for 1 hour, and recheck the pressure.
If the device does not have a built-in pressure gauge, pressurize the
device to 50 psig, submerge it in water, and check for the presence of air
bubbles escaping from any of the fittings. If pressure is lost, check all
fittings and inspect and replace 0-rings, if necessary. Retest the
device. If leakage problems cannot be solved, the manufacturer should be
contacted.
Some ZHEs use gas pressure to actuate the ZHE piston, while others
use mechanical pressure (see Table 3). Whereas the volatiles procedure
(see Step 7.3) refers to pounds -per -square -inch (psig), for the
mechanically actuated piston, the pressure applied is measured in torque -
inch -pounds. Refer to the manufacturer's instructions as to the proper
conversion.
4.2.2 Bottle Extraction Vessel. When the sample is being
evaluated using the nonvolatile extraction, a jar with sufficient capacity
to hold the sample and the extraction fluid is needed. Headspace is
allowed in this vessel.
The extraction bottles may be constructed from various materials,
depending on the analytes to be analyzed and the nature of the waste (see
Step 4.3.3). It is recommended that borosilicate glass bottles be used
instead of other types of glass, especially when inorganics are of
concern. Plastic bottles, other than polytetrafluoroethylene, shall not
be used if organics are to be investigated. Bottles are available from a
number of laboratory suppliers. When this type of extraction vessel is
used, the filtration device discussed in Step 4.3.2 is used for initial
liquid/solid separation and final extract filtration.
4.3 Filtration Devices: It is recommended that all filtrations be
performed in a hood.
CD - ROM
4.3.1 Zero-Headspace Extraction Vessel (ZHE): When the sample
is evaluated for volatiles, the zero-headspace extraction vessel described
1VITON® is a trademark of Du Pont.
1312 - 2
Revision 0
September 1994
DEQ-CFW 00059924
in Step 4.2.1 is used for filtration. The device shall be capable of
supporting and keeping in place the glass fiber filter and be able to
withstand the pressure needed to accomplish separation (50 psig).
NOTE: When it is suspected that the glass fiber filter has been
ruptured, an in -line glass fiber filter may be used to filter the
material within the ZHE.
4.3.2 Filter Holder: When the sample is evaluated for other than
volatile analytes, a filter holder capable of supporting a glass fiber
filter and able to withstand the pressure needed to accomplish separation
may be used. Suitable filter holders range from simple vacuum units to
relatively complex systems capable of exerting pressures of up to 50 psig
or more. The type of filter holder used depends on the properties of the
material to be filtered (see Step 4.3.3). These devices shall have a
minimum internal volume of 300 mL and be equipped to accommodate a minimum
filter size of 47 mm (filter holders having an internal capacity of 1.5 L
or greater, and equipped to accommodate a 142 mm diameter filter, are
recommended). Vacuum filtration can only be used for wastes with low
solids content (<10 %) and for highly granular, liquid -containing wastes.
All other types of wastes should be filtered using positive pressure
filtration. Suitable filter holders known to EPA are listed in Table 4.
4.3.3 Materials of Construction: Extraction vessels and
filtration devices shall be made of inert materials which will not leach
or absorb sample components of interest. Glass, polytetrafluoroethylene
(PTFE), or type 316 stainless steel equipment may be used when evaluating
the mobility of both organic and inorganic components. Devices made of
high -density polyethylene (HOPE), polypropylene (PP), or polyvinyl
chloride (PVC) may be used only when evaluating the mobility of metals.
Borosilicate glass bottles are recommended for use over other types of
glass bottles, especially when inorganics are analytes of concern.
4.4 Filters: Filters shall be made of borosilicate glass fiber, shall
contain no binder materials, and shall have an effective pore size of 0.6 to
0.8-um . Filters known to EPA which meet these specifications are identified in
Table 5. Pre -filters must not be used. When evaluating the mobility of metals,
filters shall be acid -washed prior to use by rinsing with IN nitric acid followed
by three consecutive rinses with reagent water (a minimum of 1-L per rinse is
recommended). Glass fiber filters are fragile and should be handled with care.
4.5 pH Meters: The meter should be accurate to + 0.05 units at 25°C.
4.6 ZHE Extract Collection Devices: TEDLAR02 bags or glass, stainless
steel or PTFE gas -tight syringes are used to collect the initial liquid phase and
the final extract when using the ZHE device. These devices listed are
recommended for use under the following conditions:
2TEDLAR11 is a registered trademark of Du Pont.
CD-ROM 1312 - 3 Revision 0
September 1994
DEQ-CFW 00059925
4.6.1 If a waste contains an aqueous liquid phase or if a waste
does not contain a significant amount of nonaqueous liquid (i.e., <1 % of
total waste), the TEDLAR® bag or a 600 mL syringe should be used to collect
and combine the initial liquid and solid extract.
4.6.2 If a waste contains a significant amount of nonaqueous
liquid in the initial liquid phase (i.e., >1 % of total waste), the
syringe or the TEDLAR® bag may be used for both the initial solid/liquid
separation and the final extract filtration. However, analysts should use
one or the other, not both.
4.6.3 If the waste contains no initial liquid phase (is 100
solid) or has no significant solid phase (is <0.5% solid) , either the
TEDLAR® bag or the syringe may be used. If the syringe is used, discard
the first 5 mL of liquid expressed from the device. The remaining
aliquots are used for analysis.
4.7 ZHE Extraction Fluid Transfer Devices: Any device capable of
transferring the extraction fluid into the ZHE without changing the nature of the
extraction fluid is acceptable (e.g., a positive displacement or peristaltic
pump, a gas -tight syringe, pressure filtration unit (see Step 4.3.2), or other
ZHE device).
4.8 Laboratory Balance: Any laboratory balance accurate to within +
0.01 grams may be used (all weight measurements are to be within + 0.1 grams)
flask.
4.9 Beaker or Erlenmeyer flask, glass, 500 mL.
4.10 Watchglass, appropriate diameter to cover beaker or Erlenmeyer
4.11 Magnetic stirrer.
5.0 REAGENTS
5.1 Reagent grade chemicals shall be used in all tests. Unless
otherwise indicated, it is intended that all reagents shall conform to the
specifications of the Committee on Analytical Reagents of the American Chemical
Society, where such specifications are available. Other grades may be used,
provided it is first ascertained that the reagent is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.
5.2 Reagent Water. Reagent water is defined as water in which an
i nterferant is not observed at or above the method's detection limit of the
analyte(s) of interest. For nonvolatile extractions, ASTM Type II water or
equivalent meets the definition of reagent water. For volatile extractions, it
is recommended that reagent water be generated by any of the following methods.
Reagent water should be monitored periodically for impurities.
5.2.1 Reagent water for volatile extractions may be generated
by passing tap water through a carbon filter bed containing about 500
grams of activated carbon (Calgon Corp., Filtrasorb-300 or equivalent).
CD-ROM 1312 - 4 Revision 0
September 1994
DEQ-CFW 00059926
5.2.2 A water purification system (Millipore Super-O or
equivalent) may also be used to generate reagent water for volatile
extractions.
5.2.3 Reagent water for volatile extractions may also be prepared
by boiling water for 15 minutes. Subsequently, while maintaining the
water temperature at 90 ± 5 degrees C, bubble a contaminant -free inert gas
(e.g. nitrogen) through the water for 1 hour. While still hot, transfer
the water to a narrow mouth screw -cap bottle under zero-headspace and seal
with a Teflon -lined septum and cap.
5.3 Sulfuric acid/nitric acid (60/40 weight percent mixture) HISO4/HNO3.
Cautiously mix 60 g of concentrated sulfuric acid with 40 g of concentrated
nitric acid. If preferred, a more dilute HZSO4/HNO3 acid mixture may be prepared
and used in steps 5.4.1 and 5.4.2 making it easier to adjust the pH of the
extraction fluids.
5.4 Extraction fluids.
5.4.1 Extraction fluid #1: This fluid is made by adding the
60/40 weight percent mixture of sulfuric and nitric acids (or a suitable
dilution) to reagent water (Step 5.2) until the pH is 4.20 + 0.05. The
fluid is used to determine the leachability of soil from a site that is
east of the Mississippi River, and the leachability of wastes and
wastewaters.
NOTE: Solutions are unbuffered and exact pH may not be attained.
5.4.2 Extraction fluid #2: This fluid is made by adding the
60/40 weight percent mixture of sulfuric and nitric acids (or a suitable
dilution) to reagent water (Step 5.2) until the pH is 5.00 ± 0.05. The
fluid is used to determine the leachability of soil from a site that is
west of the Mississippi River.
5.4.3 Extraction fluid #3: This fluid is reagent water (Step
5.2) and is used to determine cyanide and volatiles leachability.
NOTE: These extraction fluids should be monitored frequently for
impurities. The pH should be checked prior to use to ensure that
these fluids are made up accurately. If impurities are found or
the pH is not within the above specifications, the fluid shall be
discarded and fresh extraction fluid prepared.
5.5 Analytical standards shall be prepared according to the appropriate
analytical method.
6.0 SAMPLE COLLECTION, PRESERVATION, AND HANDLING
6.1 All samples shall be collected using an appropriate sampling plan.
6.2 There may be requirements on the minimal size of the field sample
depending upon the physical state or states of the waste and the analytes of
concern. An aliquot is needed for the preliminary evaluations of the percent
CD-ROM 1312 - 5 Revision 0
September 1994
DEQ-CFW 00059927
solids and the particle size. An aliquot may be needed to conduct the
nonvolatile analyte extraction procedure. If volatile organics are of concern,
another aliquot may be needed. Quality control measures may require additional
aliquots. Further, it is always wise to collect more sample just in case
something goes wrong with the initial attempt to conduct the test.
6.3 Preservatives shall not be added to samples before extraction.
6.4 Samples may be refrigerated unless refrigeration results in
irreversible physical change to the waste. If precipitation occurs, the entire
sample (including precipitate) should be extracted.
6.5 When the sample is to be evaluated for volatile analytes, care
shall be taken to minimize the loss of volatiles. Samples shall be collected and
stored in a manner intended to prevent the loss of volatile analytes (e.g.,
samples should be collected in Teflon -lined septum capped vials and stored at
VC. Samples should be opened only immediately prior to extraction).
6.6 1312 extracts should be prepared for analysis and analyzed as soon
as possible following extraction. Extracts or portions of extracts for metallic
analyte determinations must be acidified with nitric acid to a pH < 2, unless
precipitation occurs (see Step 7.2.14 if precipitation occurs). Extracts should
be preserved for other analytes according to the guidance given in the individual
analysis methods. Extracts or portions of extracts for organic analyte
determinations shall not be allowed to come into contact with the atmosphere
(i.e., no headspace) to prevent losses. See Step 8.0 (Quality Control) for
acceptable sample and extract holding times.
7.0 PROCEDURE
7.1 Preliminary Evaluations
Perform preliminary 1312 evaluations on a minimum 100 gram aliquot of
sample. This aliquot may not actually undergo 1312 extraction. These
preliminary evaluations include: (1) determination of the percent solids (Step
7.1.1); (2) determination of whether the waste contains insignificant solids and
is, therefore, its own extract after filtration (Step 7.1.2); and (3)
determination of whether the solid portion of the waste requires particle size
reduction (Step 7.1.3).
7.1.1 Preliminary determination of percent solids: Percent
solids is defined as that fraction of a waste sample (as a percentage of
the total sample) from which no liquid may be forced out by an applied
pressure, as described below.
7.1.1.1 If the sample will obviously yield no free
liquid when subjected to pressure filtration (i.e., is 100% solid),
weigh out a representative subsample (100 g minimum) and proceed
to Step 7.1.3.
7.1.1.2 If the sample is liquid or multiphasic,
liquid/solid separation to make a preliminary determination of
percent solids is required. This involves the filtration device
CD-ROM 1312 - 6 Revision 0
September 1994
DEQ-CFW 00059928
CD-ROM
discussed in Step 4.3.2, and is outlined in Steps 7.1.1.3 through
7.1.1.9.
7.1.1.3 Pre -weigh the filter and the container that will
receive the filtrate.
7.1.1.4 Assemble filter holder and filter following the
manufacturer's instructions. Place the filter on the support
screen and secure.
7.1.1.5 Weigh out a subsample of the waste (100 gram
minimum) and record the weight.
7.1.1.6 Allow slurries to stand to permit the solid phase
to settle. Samples that settle slowly may be centrifuged prior to
filtration. Centrifugation is to be used only as an aid to
filtration. If used, the liquid should be decanted and filtered
followed by filtration of the solid portion of the waste through
the same filtration system.
7.1.1.7 Quantitatively transfer the sample to the filter
holder (liquid and solid phases). Spread the sample evenly over
the surface of the filter. If filtration of the waste at 4°C
reduces the amount of expressed liquid over what would be expressed
at room temperature, then allow the sample to warm up to room
temperature in the device before filtering.
Gradually apply vacuum or gentle pressure of 1-10 psig,
until air or pressurizing gas moves through the filter. If this
point is not reached under 10 psig, and if no additional liquid has
passed through the filter in any 2-minute interval, slowly increase
the pressure in 10 psig increments to a maximum of 50 psig. After
each incremental increase of 10 psig, if the pressurizing gas has
not moved through the filter, and if no additional liquid has
passed through the filter in any 2-minute interval, proceed to the
next 10-psig increment. When the pressurizing gas begins to move
through the filter, or when liquid flow has ceased at 50 psig
(i.e., filtration does not result in any additional filtrate within
any 2-minute period), stop the filtration.
NOTE: If sample material (>1 % of original sample weight) has
obviously adhered to the container used to transfer the sample to
the filtration apparatus, determine the weight of this residue and
subtract it from the sample weight determined in Step 7.1.1.5 to
determine the weight of the sample that will be filtered.
NOTE: Instantaneous application of high pressure can degrade the
glass fiber filter and may cause premature plugging.
7.1.1.8 The material in the filter holder is defined as
the solid phase of the sample, and the filtrate is defined as the
liquid phase.
1312 - 7 Revision 0
September 1994
DEQ-CFW 00059929
NOTE: Some samples, such as oily wastes and some paint wastes,
will obviously contain some material that appears to be a liquid,
but even after applying vacuum or pressure filtration, as outlined
in Step 7.1.1.7, this material may not filter. If this is the
case, the material within the filtration device is defined as a
solid. Do not replace the original filter with a fresh filter
under any circumstances. Use only one filter.
7.1.1.9 Determine the weight of the liquid phase by
subtracting the weight of the filtrate container (see Step 7.1.1.3)
from the total weight of the filtrate -filled container. Determine
the weight of the solid phase of the sample by subtracting the
weight of the liquid phase from the weight of the total sample, as
determined in Step 7.1.1.5 or 7.1.1.7.
Record the weight of the liquid and solid phases.
Calculate the percent solids as follows:
Weight of solid (Step 7.1.1.9)
Percent solids = x 100
Total weight of waste (Step 7.1.1.5 or 7.1.1.7)
7.1.2 If the percent solids determined in Step 7.1.1.9 is equal
to or greater than 0.5%, then proceed either to Step 7.1.3 to determine
whether the solid material requires particle size reduction or to Step
7.1.2.1 if it is noticed that a small amount of the filtrate is entrained
in wetting of the filter. If the percent solids determined in Step
7.1.1.9 is less than 0.5%, then proceed to Step 7.2.9 if the nonvolatile
1312 analysis is to be performed, and to Step 7.3 with a fresh portion of
the waste if the volatile 1312 analysis is to be performed.
7.1.2.1 Remove the solid phase and filter from the
filtration apparatus.
7.1.2.2 Dry the filter and solid phase at 100 + 20°C
until two successive weighings yield the same value within ± 1 %.
Record the final weight.
Caution: The drying oven should be vented to a hood or other
appropriate device to eliminate the possibility of fumes from the
sample escaping into the laboratory. Care should be taken to
ensure that the sample will not flash or violently react upon
heating.
7.1.2.3 Calculate the percent dry solids as follows:
Percent (Weight of dry sample + filter) - tared weight of filter
dry solids = x 100
Initial weight of sample (Step 7.1.1.5 or 7.1.1.7)
CD-ROM 1312 - 8 Revision 0
September 1994
DEQ-CFW 00059930
CD-ROM
7.1.2.4 If the percent dry solids is less than 0.5%,
then proceed to Step 7.2.9 if the nonvolatile 1312 analysis is to
be performed, and to Step 7.3 if the volatile 1312 analysis is to
be performed. If the percent dry solids is greater than or equal
to 0.5%, and if the nonvolatile 1312 analysis is to be performed,
return to the beginning of this Step (7.1) and, with a fresh
portion of sample, determine whether particle size reduction is
necessary (Step 7.1.3).
7.1.3 Determination of whether the sample requires particle -size
reduction (particle -size is reduced during this step): Using the solid
portion of the sample, evaluate the solid for particle size. Particle -
size reduction is required, unless the solid has a surface area per gram
of material equal to or greater than 3.1 cm', or is smaller than 1 cm in
its narrowest dimension (i.e., is capable of passing through a 9.5 mm
(0.375 inch) standard sieve). If the surface area is smaller or the
particle size larger than described above, prepare the solid portion of
the sample for extraction by crushing, cutting, or grinding the waste to
a surface area or particle size as described above. If the solids are
prepared for organic volatiles extraction, special precautions must be
taken (see Step 7.3.6).
NOTE: Surface area criteria are meant for filamentous (e.a.,
paper, cloth, and similar) waste materials. Actual measurement of
surface area is not required, nor is it recommended. For materials
that do not obviously meet the criteria, sample -specific methods
would need to be developed and employed to measure the surface
area. Such methodology is currently not available.
7.1.4 Determination of appropriate extraction fluid:
7.1.4.1 For soils, if the sample is from a site that is
east of the Mississippi River, extraction fluid #1 should be used.
If the sample is from a site that is west of the Mississippi River,
extraction fluid #2 should be used.
7.1.4.2 For wastes and wastewater, extraction fluid #1
should be used.
7.1.4.3 For cyanide -containing wastes and/or soils,
extraction fluid #3 (reagent water) must be used because leaching
of cyanide -containing samples under acidic conditions may result
in the formation of hydrogen cyanide gas.
7.1.5 If the aliquot of the sample used for the preliminary
evaluation (Steps 7.1.1 - 7.1.4) was determined to be 100% solid at Step
7.1.1.1, then it can be used for the Step 7.2 extraction (assuming at
least 100 grams remain), and the Step 7.3 extraction (assuming at least 25
grams remain). If the aliquot was subjected to the procedure in Step
7.1.1.7, then another aliquot shall be used for the volatile extraction
procedure in Step 7.3. The aliquot of the waste subjected to the
procedure in Step 7.1.1.7 might be appropriate for use for the Step 7.2
extraction if an adequate amount of solid (as determined by Step 7.1.1.9)
1312 - 9 Revision 0
September 1994
DEQ-CFW 00059931
was obtained. The amount of solid necessary is dependent upon whether a
sufficient amount of extract will be produced to support the analyses. If
an adequate amount of solid remains, proceed to Step 7.2.10 of the
nonvolatile 1312 extraction.
7.2 Procedure When Volatiles Are Not Involved
A minimum sample size of 100 grams (solid and liquid phases) is
recommended. In some cases, a larger sample size may be appropriate, depending
on the solids content of the waste sample (percent solids, See Step 7.1.1),
whether the initial liquid phase of the waste will be miscible with the aqueous
extract of the solid, and whether inorganics, semivolatile organics, pesticides,
and herbicides are all analytes of concern. Enough solids should be generated
for extraction such that the volume of 1312 extract will be sufficient to support
all of the analyses required. If the amount of extract generated by a single
1312 extraction will not be sufficient to perform all of the analyses, more than
one extraction may be performed and the extracts from each combined and aliquoted
for analysis.
7.2.1 If the sample will obviously yield no liquid when subjected
to pressure filtration (i.e., is 100 % solid, see Step 7.1.1), weigh out
a subsample of the sample (100 gram minimum) and proceed to Step 7.2.9.
7.2.2 If the sample is liquid or multiphasic, liquid/solid
separation is required. This involves the filtration device described in
Step 4.3.2 and is outlined in Steps 7.2.3 to 7.2.8.
7.2.3 Pre -weigh the container that will receive the filtrate.
7.2.4 Assemble the filter holder and filter following the
manufacturer's instructions. Place the filter on the support screen and
secure. Acid wash the filter if evaluating the mobility of metals (see
Step 4.4).
NOTE: Acid washed filters may be used for all nonvolatile
extractions even when metals are not of concern.
7.2.5 Weigh out a subsample of the sample (100 gram minimum) and
record the weight. If the waste contains <0.5 % dry solids (Step 7.1.2),
the liquid portion of the waste, after filtration, is defined as the 1312
extract. Therefore, enough of the sample should be filtered so that the
amount of filtered liquid will support all of the analyses required of the
1312 extract. For wastes containing >0.5 % dry solids (Steps 7.1.1 or
7.1.2), use the percent solids information obtained in Step 7.1.1 to
determine the optimum sample size (100 gram minimum) for filtration.
Enough solids should be generated by filtration to support the analyses to
be performed on the 1312 extract.
7.2.6 Allow slurries to stand to permit the solid phase to settle.
Samples that settle slowly may be centrifuged prior to filtration. Use
centrifugation only as an aid to filtration. If the sample is
centrifuged, the liquid should be decanted and filtered followed by
CD-ROM 1312 - 10 Revision 0
September 1994
DEQ-CFW 00059932
CD-ROM
filtration of the solid portion of the waste through the same filtration
system.
7.2.7 Quantitatively transfer the sample (liquid and solid phases)
to the filter holder (see Step 4.3.2). Spread the waste sample evenly
over the surface of the filter. If filtration of the waste at 4°C reduces
the amount of expressed liquid over what would be expressed at room
temperature, then allow the sample to warm up to room temperature in the
device before filtering.
Gradually apply vacuum or gentle pressure of 1-10 psig, until air
or pressurizing gas moves through the filter. If this point if not
reached under 10 psig, and if no additional liquid has passed through the
filter in any 2-minute interval, slowly increase the pressure in 10-psig
increments to maximum of 50 psig. After each incremental increase of 10
psig, if the pressurizing gas has not moved through the filter, and if no
additional liquid has passed through the filter in any 2-minute interval,
proceed to the next 10-psig increment. When the pressurizing gas begins
to move through the filter, or when the liquid flow has ceased at 50 psig
(i.e., filtration does not result in any additional filtrate within a
2-minute period), stop the filtration.
NOTE: If waste material (>1 % of the original sample weight) has
obviously adhered to the container used to transfer the sample to
the filtration apparatus, determine the weight of this residue and
subtract it from the sample weight determined in Step 7.2.5, to
determine the weight of the waste sample that will be filtered.
NOTE: Instantaneous application of high pressure can degrade the
glass fiber filter and may cause premature plugging.
7.2.8 The material in the filter holder is defined as the solid
phase of the sample, and the filtrate is defined as the liquid phase.
Weigh the filtrate. The liquid phase may now be either analyzed (see Step
7.2.12) or stored at 4°C until time of analysis.
NOTE: Some wastes, such as oily wastes and some paint wastes, will
obviously contain some material which appears to be a liquid. Even
after applying vacuum or pressure filtration, as outlined in Step
7.2.7, this material may not filter. If this is the case, the
material within the filtration device is defined as a solid, and
is carried through the extraction as a solid. Do not replace the
original filter with a fresh filter under any circumstances. Use
only one filter.
7.2.9 If the sample contains <0.5% dry solids (see Step 7.1.2),
proceed to Step 7.2.13. If the sample contains >0.5 % dry solids (see
Step 7.1.1 or 7.1.2), and if particle -size reduction of the solid was
needed in Step 7.1.3, proceed to Step 7.2.10. If the sample as received
passes a 9.5 mm sieve, quantitatively transfer the solid material into the
extractor bottle along with the filter used to separate the initial liquid
from the solid phase, and proceed to Step 7.2.11.
1312 - 11 Revision 0
September 1994
DEQ-CFW 00059933
7.2.10 Prepare the solid portion of the sample for extraction by
crushing, cutting, or grinding the waste to a surface area or particle -
size as described in Step 7.1.3. When the surface area or particle -size
has been appropriately altered, quantitatively transfer the solid material
into an extractor bottle. Include the filter used to separate the initial
liquid from the solid phase.
NOTE: Sieving of the waste is not normally required. Surface area
requirements are meant for filamentous (e.g., paper, cloth) and
similar waste materials. Actual measurement of surface area is not
recommended. If sieving is necessary, a Teflon -coated sieve should
be used to avoid contamination of the sample.
7.2.11 Determine the amount of extraction fluid to add to the
extractor vessel as follows:
Weight of =
extraction fluid
20 x % solids (Step 7.1.1) x weight of waste
filtered (Step 7.2.5 or 7.2.7)
100
Slowly add this amount of appropriate extraction fluid (see Step
7.1.4) to the extractor vessel. Close the extractor bottle tightly (it is
recommended that Teflon tape be used to ensure a tight seal), secure in
rotary extractor device, and rotate at 30 + 2 rpm for 18 ± 2 hours.
Ambient temperature (i.e., temperature of room in which extraction takes
place) shall b.e maintained at 23 + 2°C during the extraction period.
NOTE: As agitation continues, pressure may build up within the
extractor bottle for some types of sample (e.g., limed or calcium
carbonate -containing sample may evolve gases such as carbon
dioxide). To relieve excess pressure, the extractor bottle may be
periodically opened (e.g., after 15 minutes, 30 minutes, and 1
hour) and vented into a hood.
7.2.12 Following the 18 ± 2 hour extraction, separate the material
in the extractor vessel into its component liquid and solid phases by
filtering through a new glass fiber filter, as outlined in Step 7.2.7.
For final filtration of the 1312 extract, the glass fiber filter may be
changed, if necessary, to facilitate filtration. Filter(s) shall be
acid -washed (see Step 4.4) if evaluating the mobility of metals.
7.2.13 Prepare the 1312 extract as follows:
7.2.13.1 If the sample contained no initial liquid phase,
the filtered liquid material obtained from Step 7.2.12 is defined
as the 1312 extract. Proceed to Step 7.2.14.
7.2.13.2 If compatible (e.g., multiple phases will not
result on combination), combine the filtered liquid resulting from
Step 7.2.12 with the initial liquid phase of the sample obtained
CD-ROM 1312 - 12 Revision 0
September 1994
DEQ-CFW 00059934
in Step 7.2.7. This combined liquid is defined as the 1312
extract. Proceed to Step 7.2.14.
7.2.13.3 If the initial liquid phase of the waste, as
obtained from Step 7.2.7, is not or may not be compatible with the
filtered liquid resulting from Step 7.2.12, do not combine these
liquids. Analyze these liquids, collectively defined as the 1312
extract, and combine the results mathematically, as described in
Step 7.2.14.
7.2.14 Following collection of the 1312 extract, the pH of the
extract should be recorded. Immediately aliquot and preserve the extract
for analysis. Metals aliquots must be acidified with nitric acid to pH <
2. If precipitation is observed upon addition of nitric acid to a small
aliquot of the extract, then the remaining portion of the extract for
metals analyses shall not be acidified and the extract shall be analyzed
as soon as possible. All other aliquots must be stored under
refrigeration WC) until analyzed. The 1312 extract shall be prepared
and analyzed according to appropriate analytical methods. 1312 extracts
to be analyzed for metals shall be acid digested except in those instances
where digestion causes loss of metallic analytes. If an analysis of the
undigested extract shows that the concentration of any regulated metallic
analyte exceeds the regulatory level, then the waste is hazardous and
digestion of the extract is not necessary. However, data on undigested
extracts alone cannot be used to demonstrate that the waste is not
hazardous. If the individual phases are to be analyzed separately,
determine the volume of the individual phases (to ± 0.5 %), conduct the
appropriate analyses, and combine the results mathematically by using a
simple volume -weighted average:
Final Analyte Concentration =
where:
V, = The volume of the
C, = The concentration
V, = The volume of the
CZ = The concentration
(mg /L).
(V,) (CO + (Vz) (C2)
V, + V2
first phase (L).
of the analyte of concern in the first phase (mg/L).
second phase (L).
of the analyte of concern in the second phase
7.2.15 Compare the analyte concentrations in the 1312 extract with
the levels identified in the appropriate regulations. Refer to Section
8.0 for quality assurance requirements.
7.3 Procedure When Volatiles Are Involved
Use the ZHE device to obtain 1312 extract for analysis of volatile
compounds only. Extract resulting from the use of the ZHE shall not be used to
evaluate the mobility of non-volatile analytes (e.g.,.metals, pesticides, etc.).
CD-ROM
1312 - 13
Revision 0
September 1994
DEQ-CFW 00059935
The ZHE device has approximately a 500 mL internal capacity. The ZHE can
thus accommodate a maximum of 25 grams of solid (defined as that fraction of a
sample from which no additional liquid may be forced out by an applied pressure
of 50 psig), due to the need to add an amount of extraction fluid equal to 20
times the weight of the solid phase.
Charge the ZHE with sample only once and do not open the device until the
final extract (of the solid) has been collected. Repeated filling of the ZHE to
obtain 25 grams of solid is not permitted.
Do not allow the sample, the initial liquid phase, or the extract to be
exposed to the atmosphere for any more time than is absolutely necessary. Any
manipulation of these materials should be done when cold WC) to minimize loss
of volatiles.
7.3.1 Pre -weigh the (evacuated) filtrate collection container
(see Step 4.6) and set aside. If using a TEDLAR® bag, express all liquid
from the ZHE device into the bag, whether for the initial or final
liquid/solid separation, and take an aliquot from the liquid in the bag
for analysis. The containers listed in Step 4.6 are recommended for use
under the conditions stated in Steps 4.6.1-4.6.3.
7.3.2 Place the ZHE piston within the body of the ZHE (it may be
helpful first to moisten the piston 0-rings slightly with extraction
fluid). Adjust the piston within the ZHE body to a height that will
minimize the distance the piston will have to move once the ZHE is charged
with sample (based upon sample size requirements determined from Step 7.3,
Step 7.1.1 and/or 7.1.2). Secure the gas inlet/outlet flange (bottom
flange) onto the ZHE body in accordance with the manufacturer's
instructions. Secure the glass fiber filter between the support screens
and set aside. Set liquid inlet/outlet flange (top flange) aside.
7.3.3 If the sample is 100% solid (see Step 7.1.1), weigh out
a subsample (25 gram maximum) of the waste, record weight, and proceed to
Step 7.3.5.
7.3.4 If the sample contains <0.5% dry solids (Step 7.1.2), the
liquid portion of waste, after filtration, is defined as the 1312 extract.
Filter enough of the sample so that the amount of filtered liquid will
support all of the volatile analyses required. For samples containing
>0.5% dry solids (Steps 7.1.1 and/or 7.1.2), use the percent solids
information obtained in Step 7.1.1 to determine the optimum sample size to
charge into the ZHE. The recommended sample size is as follows:
7.3.4.1 For samples containing <5% solids (see Step
7.1.1), weigh out a 500 gram subsample of waste and record the
weight.
7.3.4.2 For wastes containing >5% solids (see Step
7.1.1), determine the amount of waste to charge into the ZHE as
follows:
CD-ROM 1312 - 14 Revision 0
September 1994
DEQ-CFW 00059936
25
Weight of waste to charge ZHE = x 100
percent solids (Step 7.1.1)
Weigh out a subsampl e of the waste of the appropriate size and
record the weight.
7.3.5 If particle -size reduction of the solid portion of the
sample was required in Step 7.1.3, proceed to Step 7.3.6. If particle -
size reduction was not required in Step 7.1.3, proceed to Step 7.3.7.
7.3.6 Prepare the sample for extraction by crushing, cutting, or
grinding the solid portion of the waste to a surface area or particle size
as described in Step 7.1.3.1. Wastes and appropriate reduction equipment
should be refrigerated, if possible, to 4°C prior to particle -size
reduction. The means used to effect particle -size reduction must not
generate heat in and of itself. If reduction of the solid phase of the
waste is necessary, exposure of the waste to the atmosphere should be
avoided to the extent possible.
NOTE: Sieving of the waste is not recommended due to the
possibility that vol ati1es may be lost. The use of an
appropriately graduated ruler is recommended as an acceptable
alternative. Surface area requirements are meant for filamentous
(e.g., paper, cloth) and similar waste materials. Actual
measurement of surface area is not recommended.
When the surface area or particle -size has been appropriately
altered, proceed to Step 7.3.7.
7.3.7 Waste slurries need not be allowed to stand to permit the
solid phase to settle. Do not centrifuge samples prior to filtration.
7.3.8 Quantitatively transfer the entire sample (liquid and solid
phases) quickly to the ZHE. Secure the filter and support screens into
the top flange of the device and secure the top flange to the ZHE body in
accordance with the manufacturer's instructions. Tighten all ZHE fittings
and place the device in the vertical position (gas inlet/outlet flange on
the bottom). Do not attach the extraction collection device to the top
plate.
Note: If sample material M % of original sample weight) has
obviously adhered to the container used to transfer the sample to
the ZHE, determine the weight of this residue and subtract it from
the sample weight determined in Step 7.3.4 to determine the weight
of the waste sample that will be filtered.
Attach a gas line to the gas inlet/outlet valve (bottom flange)
and, with the liquid inlet/outlet valve (top flange) open, begin applying
gentle pressure of 1-10 psig (or more if necessary) to force all headspace
slowly out of the ZHE device into a hood. At the first appearance of
liquid from the liquid inlet/outlet valve, quickly close the valve and
discontinue pressure. If filtration of the waste at 4°C reduces the
CD-ROM 1312 - 15 Revision 0
September 1994
DEQ-CFW 00059937
amount of expressed liquid over what would be expressed at room
temperature, then allow the sample to warm up to room temperature in the
device before filtering. If the waste is 100 q solid (see Step 7.1.1),
slowly increase the pressure to a maximum of 50 psig to force most of the
headspace out of the device and proceed to Step 7.3.12.
7.3.9 Attach the evacuated pre -weighed filtrate collection
container to the liquid inlet/outlet valve and open the valve. Begin
applying gentle pressure of 1-10 psig to force the liquid phase of the
sample into the filtrate collection container. If no additional liquid
has passed through the filter in any 2-minute interval, slowly increase
the pressure in 10-psig increments to a maximum of 50 psig. After each
incremental increase of 10 psig, if no additional liquid has passed
through the filter in any 2-minute interval, proceed to the next 10-psig
increment. When liquid flow has ceased such that continued pressure
filtration at 50 psig does not result in any additional filtrate within a
2-minute period, stop the filtration. Close the liquid inlet/outlet
valve, discontinue pressure to the piston, and disconnect and weigh the
filtrate collection container.
NOTE: Instantaneous application of high pressure can degrade the
glass fiber filter and may cause premature plugging.
7.3.10 The material in the ZHE is defined as the solid phase of
the sample and the filtrate is defined as the liquid phase.
NOTE: Some samples, such as oily wastes and some paint wastes,
will obviously contain some material which appears to be a liquid.
Even after applying pressure filtration, this material will not
filter. If this is the case, the material within the filtration
device is defined as a solid, and is carried through the 1312
extraction as a solid.
If the original waste contained <0.5 % dry solids (see Step 7.1.2),
this filtrate is defined as the 1312 extract and is analyzed directly.
Proceed to Step 7.3.15.
7.3.11 The liquid phase may now be either analyzed immediately
(see Steps 7.3.13 through 7.3.15) or stored at 4°C under minimal headspace
conditions until time of analysis. Determine the weight of extraction
fluid #3 to add to the ZHE as follows:
Weight of extraction fluid
20 x % solids (Step 7.1.1) x weight
of waste filtered (Step 7.3.4 or 7.3.8)
100
7.3.12 The following steps detail how to add the appropriate
amount of extraction fluid to the solid material within the ZHE and
agitation of the ZHE vessel. Extraction fluid #3 is used in all cases
(see Step 5.4.3).
CD-ROM 1312 - 16 Revision 0
September 1994
DEQ-CFW 00059938
7.3.12.1 With the ZHE in the vertical position, attach a
line from the extraction fluid reservoir to the liquid inlet/outlet
valve. The line used shall contain fresh extraction fluid and
should be preflushed with fluid to eliminate any air pockets in the
line. Release gas pressure on the ZHE piston (from the gas
inlet/outlet valve), open the liquid inlet/outlet valve, and begin
transferring extraction fluid (by pumping or similar means) into
the ZHE. Continue pumping extraction fluid into the ZHE until the
appropriate amount of fluid has been introduced into the device.
7.3.12.2 After the extraction fluid has been added,
immediately close the liquid inlet/outlet valve and disconnect the
extraction fluid line. Check the ZHE to ensure that all valves are
in their closed positions. Manually rotate the device in an
end -over -end fashion 2 or 3 times. Reposition the ZHE in the
vertical position with the liquid inlet/outlet valve on top.
Pressurize the ZHE to 5-10 psig (if necessary) and slowly open the
liquid inlet/outlet valve to bleed out any headspace (into a hood)
that may have been introduced due to the addition of extraction
fluid. This bleeding shall be done quickly and shall be stopped
at the first appearance of liquid from the valve. Re -pressurize
the ZHE with 5-10 psig and check all ZHE fittings to ensure that
they are closed.
7.3.12.3 Place the ZHE in the rotary extractor apparatus
(if it is not already there) and rotate at 30 + 2 rpm for 18 ± 2
hours. Ambient temperature (i.e., temperature of room in which
extraction occurs) shall be maintained at 23 + 20C during
agitation.
7.3.13 Following the 18 + 2 hour agitation period, check the
pressure behind the ZHE piston by quickly opening and closing the gas
inlet/outlet valve and noting the escape of gas. If the pressure has not
been maintained (i.e., no gas release observed); the ZHE is leaking.
Check the ZHE for leaking as specified in Step 4.2.1, and perform the
extraction again with a new sample of waste. If the pressure within the
device has been maintained, the material in the extractor vessel is once
again separated into its component liquid and solid phases. If the waste
contained an initial liquid phase, the liquid may be filtered directly
into the same filtrate collection container (i.e., TEDLAR® bag) holding the
initial liquid phase of the waste. A separate filtrate collection
container must be used if combining would create multiple phases, or there
is not enough volume left within the filtrate collection container.
Filter through the glass fiber filter, using the ZHE device as discussed
in Step 7.3.9. All extracts shall be filtered and collected if the TEDLAR®
bag is used, if the extract is multiphasic, or if the waste contained an
initial liquid phase (see Steps 4.6 and 7.3.1).
NOTE: An in -line glass fiber filter may be used to filter the
material within the ZHE if it is suspected that the glass fiber
filter has been ruptured
CD-ROM 1312 - 17 Revision 0
September 1994
DEQ-CFW 00059939
7.3.14 If the original sample contained no initial liquid phase,
the filtered liquid material obtained from Step 7.3.13 is defined as the
1312 extract. If the sample contained an initial liquid phase, the
filtered liquid material obtained from Step 7.3.13 and the initial liquid
phase (Step 7.3.9) are collectively defined as the 1312 extract.
7.3.15 Following collection of the 1312 extract, immediately
prepare the extract for analysis and store with minimal headspace at 4°C
until analyzed. Analyze the 1312 extract according to the appropriate
analytical methods. If the individual phases are to be analyzed
separately (i.e., are not miscible), determine the volume of the
individual phases (to 0.5%), conduct the appropriate analyses, and combine
the results mathematically by using a simple volume- weighted average:
(V1) (CO + (V2) (C2)
Final Analyte =
Concentration V1 + V2
where:
V1 = The
volume of the
first phases (L).
C1 = The
concentration
of the analyte of concern in the first phase (mg/L).
V2 = The
volume of the
second phase (L).
C2 = The
concentration
of the analyte of concern in the second phase
(mg/L).
7.3.16 Compare the analyte concentrations in the 1312 extract with
the levels identified in the appropriate regulations. Refer to Step 8.0
for quality assurance requirements.
8.0 QUALITY CONTROL
8.1 A minimum of one blank (using the same extraction fluid as used for
the samples) for every 20 extractions that have been conducted in an extraction
vessel. Refer to Chapter One for additional quality control protocols.
8.2 A matrix spike shall be performed for each waste type (e.g.,
wastewater treatment sludge, contaminated soil, etc.) unless the result exceeds
the regulatory level and the data is being used solely to demonstrate that the
waste property exceeds the regulatory level. A minimum of one matrix spike must
be analyzed for each analytical batch. As a minimum, follow the matrix spike
addition guidance provided in each analytical method.
CD-ROM
8.2.1 Matrix spikes are to be added after filtration of the 1312
extract and before preservation. Matrix spikes should not be added prior
to 1312 extraction of the sample.
8.2.2 In most cases, matrix spike levels should be added at a
concentration equivalent to the corresponding regulatory level. If the
analyte concentration is less than one half the regulatory level, the
1312 - 18
Revision 0
September 1994
DEQ-CFW 00059940
spike concentration may be as low as one half of the analyte
concentration, but may not be less than five times the method detection
limit. In order to avoid differences in matrix effects, the matrix spikes
must be added to the same nominal volume of 1312 extract as that which was
analyzed for the unspiked sample.
8.2.3 The purpose of the matrix spike is to monitor the
performance of the analytical methods used, and to determine whether
matrix interferences exist. Use of other internal calibration methods,
modification of the analytical methods, or use of alternate analytical
methods may be needed to accurately measure the analyte concentration in
the 1312 extract when the recovery of the matrix spike is below the
expected analytical method performance.
8.2.4 Matrix spike recoveries are calculated by the following
formula:
1R (% Recovery) = 100 (Xs - %) 1 K
where:
XS = measured value for the spiked sample
X,, = measured value for the unspiked sample, and
K = known value of the spike in the sample.
8.3 All quality control measures described in the appropriate analytical
methods shall be followed.
8.4 The use of internal calibration quantitation methods shall be
employed for a metallic contaminant if: (1) Recovery of the contaminant from the
1312 extract is not at least 50% and the concentration does not exceed the
appropriate regulatory level, and (2) The concentration of the contaminant
measured in the extract is within 20% of the appropriate regulatory level.
8.4.1. The method of standard additions shall be employed as the
internal calibration quantitation method for each metallic contaminant.
8.4.2 The method of standard additions requires preparing
calibration standards in the sample matrix rather than reagent water or
blank solution. It requires taking four identical aliquots of the
solution and adding known amounts of standard to three of these aliquots.
The forth aliquot is the unknown. Preferably, the first addition should
be prepared so that the resulting concentration is approximately 50% of
the expected concentration of the sample. The second and third additions
should be prepared so that the concentrations are approximately 100% and
150% of the expected concentration of the sample. All four aliquots are
maintained at the same final volume by adding reagent water or a blank
solution, and may need dilution adjustment to maintain the signals in the
linear range of the instrument technique. All four aliquots are analyzed.
8.4.3 Prepare a plot, or subject data to linear regression, of
instrument signals or external -calibration -derived concentrations as the
dependant variable (y-axis) versus concentrations of the additions of
standards as the independent variable (x-axis). Solve for the intercept
CD-ROM 1312 - 19 Revision 0
September 1994
DEQ-CFW 00059941
of the abscissa (the independent variable, x-axis) which is the concentra-
tion in the unknown.
8.4.4 Alternately, subtract the instrumental signal or external -
calibration -derived concentration of the unknown (unspiked) sample from
the instrumental signals or external -calibration -derived concentrations of
the standard additions. Plot or subject to linear regression of the
corrected instrument signals or external -calibration -derived concentra-
tions as the dependant variable versus the independent variable. Derive
concentrations for the unknowns using the internal calibration curve as if
it were an external calibration curve.
8.5 Samples must undergo 1312 extraction within the following time
periods:
SAMPLE MAXIMUM HOLDING TIMES (days)
From: Field
From: 1312
From: Prepara-
Total
Collec-
extrac-
tive
Elapsed
tion
tion
extrac-
Time
tion
To: 1312
To: Prepara=
extrac-
tive
To: Determi-
tion
extrac-
native
tion
analysis
V o 1 a t i 1 e s
14
NA
14
28
Semi-
vol atiles
14
7
40
61
Mercury
28
NA
28
56
Metals,
except
180
NA
180
360
mercury
rNA = Not Applicable
If sample holding times are exceeded, the values obtained will be considered
minimal concentrations. Exceeding the holding time is not acceptable in
establishing that a waste does not exceed the regulatory level. Exceeding the
holding time will not invalidate characterization if the waste exceeds the
regulatory level.
9.0 METHOD PERFORMANCE
9.1 Precision results for semi-volatiles and metals: An eastern soil
with high organic content and a western soil with low organic content were used
for the semi -volatile and metal leaching experiments. Both types of soil were
analyzed prior to contaminant spiking. The results are shown in Table 6. The
concentration of contaminants leached from the soils were reproducible, as shown
CD-ROM 1312 - 20 Revision 0
September 1994
DEQ-CFW 00059942
by the moderate relative standard deviations (RSDs) of the recoveries (averaging
29% for the compounds and elements analyzed).
9.2 Precision results for volatiles: Four different soils were spiked
and tested for the extraction of volatiles. Soils One and Two were from western
and eastern Superfund sites. Soils Three and Four were mixtures of a western
soil with low organic content and two different municipal sludges. The results
are shown in Table 7. Extract concentrations of volatile organics from the
eastern soil were lower than from the western soil. Replicate teachings of Soils
Three and Four showed lower precision than the leachates from the Superfund
soils.
10.0 REFERENCES
1. Environmental Monitoring Systems Laboratory, "Performance Testing of
Method 1312; OA Support for RCRA Testing: Project Report". EPA/600/4-
89/022, EPA Contract 68-03-3249 to Lockheed Engineering and Sciences
Company, June 1989.
2. Research Triangle Institute, "Interlaboratory Comparison of Methods 1310,
1311, and 1312 for Lead in Soil". U.S. EPA Contract 68-01-7075, November
1988.
CD-ROM
1312 - 21 Revision 0
September 1994
DEQ-CFW 00059943
Table 1. Volatile Analytes'
Compound CAS No.
Acetone
67-64-1
Benzene
71-43-2
n-Butyl alcohol
71-36-3
Carbon disulfide
75-15-0
Carbon tetrachloride
56-23-5
Chlorobenzene
108-90-7
Chloroform
67-66-3
1,2-Dichloroethane
107-06-2
1,1-Dichloroethylene
75-35-4
Ethyl acetate
141-78-6
Ethyl benzene
100-41-4
Ethyl ether
60-29-7
Isobutanol
78-83-1
Methanol
67-56-1
Methylene chloride
75-09-2
Methyl ethyl ketone
78-93-3
Methyl isobutyl ketone
108-10-1
Tetrachloroethylene
127-18-4
Toluene
108-88-3
1,1,1,-Trichloroethane
71-55-6
Trichloroethylene
79-01-6
Trichlorofluoromethane
75-69-4
1,1,2-Trichloro-1,2,2-trifluoroethane
76-13-1
Vinyl chloride
75-01-4
Xylene
1330-20-7
' When testing for any or all of these analytes, the zero-headspace extractor
vessel shall be used instead of the bottle extractor.
CD-ROM
1312 - 22
Revision 0
September 1994
DEQ-CFW 00059944
Table 2. Suitable Rotary Agitation Apparatus'
Company Location Model No.
Analytical Testing and
Consulting Services,
Inc.
Associated Design and
Manufacturing Company
Environmental Machine and
Design, Inc.
IRA Machine Shop and
Laboratory
Lars Lande Manufacturing
Millipore Corp.
Warrington, PA
4-vessel
extractor (DC20S);
(215) 343-4490
8-vessel
extractor (DC20);
12-vessel
extractor (DC20B)
Alexandria, VA
2-vessel
(3740-2);
(703) 549-5999
4-vessel
(3740-4);
6-vessel
(3740-6);
8-vessel
(3740-8);
12-vessel
(3740-12);
24-vessel
(3740-24)
Lynchburg, VA
8-vessel
(08-00-00)
(804) 845-6424
4-vessel
(04-00-00)
Santurce, PR
8-vessel
(011001)
(809) 752-4004
Whitmore Lake, MI 10-vessel (10VRE)
(313) 449-4116 5-vessel (5VRE)
Bedford, MA 4-ZHE or
(80`0) 225-3384 4 1-liter
bottle extractor
(YT300RAHW)
1 Any device that rotates the extraction vessel in an end -over -end fashion at 30
±2 rpm is acceptable.
CD-ROM
1312 - 23
Revision 0
September 1994
DEQ-CFW 00059945
Table 3. Suitable Zero-Headspace Extractor Vessels'
Company Location Model No.
Analytical Testing &
Consulting Services, Inc
Associated Design and
Manufacturing Company
Lars Lande Manufacturing'
Millipore Corporation
Environmental Machine
and Design, Inc.
Warrington, PA C102, Mechanical
(215) 343-4490 Pressure Device
Alexandria, VA 3745-ZHE, Gas
(703) 549-5999 Pressure Device
Whitmore Lake, MI
ZHE-11, Gas
(313) 449-4116
Pressure Device
Bedford, MA
YT30090HW, Gas
(800) 225-3384
Pressure Device
Lynchburg, VA
VOLA-TOX1, Gas
(804) 845-6424
Pressure Device
' Any device that meets the specifications listed in Step 4.2.1 of the method is
suitable.
' This device uses a 110 mm filter.
CD-ROM
1312 - 24
Revision 0
September 1994
DEQ-CFW 00059946
Table 4. Suitable Filter Holders'
Model/
Company Location Catalogue # Size
Nucleopore Corporation
Pleasanton, CA
425910
142
mm
(800) 882-7711
410400
47
mm
Micro Filtration
Dublin, CA
302400
142
mm
Systems
(800) 334-7132
311400
47
mm
(415) 828-6010
Millipore Corporation
Bedford, MA
YT30142HW
142
mm
(800) 225-3384
XX1004700
47
mm
' Any device capable of separating the liquid from the solid phase of the waste
is suitable, providing that it is chemically compatible with the waste and the
constituents to be analyzed. Plastic devices (not listed above) may be used when
only inorganic analytes are of concern. The 142 mm size filter holder is
recommended.
Table 5. Suitable Filter Media'
Pore
Si ze
Company Location Model (um)
Millipore Corporation
Bedford, MA
AP40
0.7
(800) 225-3384
Nucleopore Corporation
Pleasanton, CA
211625
0.7
(415) 463-2530
W'hatman Laboratory
Clifton, NJ
GFF
0.7
Products, Inc.
(201) 773-5800
Micro Filtration
Dublin, CA
GF75
0.7
Systems
(800) 334-7132
(415) 828-6010
' Any filter that meets the sp-ecifications in Step 4.4 of the Method is suitable.
CD-ROM 1312 - 25 Revision 0
September 1994
DEQ-CFW 00059947
TABLE 6 - METHOD 1312 PRECISION RESULTS FOR SEMI-VOLATILES AND METALS
FORTIFIED ANALYTES
bis(2-chloroethyl)-
ether
2-Chlorophenol
1,4-Dichlorobenzene
1,2-Dichlorobenzene
2-Methylphenol
Nitrobenzene
2,4-Dimethylphenol
Hexachlorobutadiene
Acenaphthene
2,4-Dinitrophenol
2,4-Dinitrotoluene
Hexachlorobenzene
gamma BHC (Lindane)
beta BHC
METALS
Lead
Cadmium
Eastern Soil (pH 4.2) Western Soil (pH 5.0
Amount Amount Amount
Spiked Recovered* % RSD Recovered* % RSD
(ug) (ug) (ug)
1040
834
12.5
616
14.2
1620
1010
6.8
525
54.9
2000
344
12.3
272
34.6
8920
1010
8.0
1520
28.4
3940
1860
7.7
1130
32.6
1010
812
10.0
457
21.3
1460
200
18.4
18
87.6
6300
95
12.9
280
22.8
3640
210
8.1
310**
7.7
1300
896**
6.1
23**
15.7
1900
1150
5.4
585
54.4
1840
3.7
12.0
10
173.2
7440
230
16.3
1240
55.2
640
35
13.3
65.3
51.7
5000
70
4.3
10
51.7
1000
387
2.3
91
71.3
* = Triplicate analyses.
** = Duplicate analyses; one value was rejected as an outlier at the 90%
confidence level using the Dixon Q test.
CD-ROM
1312 - 26
Revision 0
September 1994
DEQ-CFW 00059948
TABLE 7 - METHOD 1312 PRECISION RESULTS FOR VOLATILES
Compound Name
Acetone
Acrylonitrile
Benzene
n-Butyl Alcohol
(1-Butanol)
Carbon disulfide
Carbon tetrachloride
Chlorobenzene
Chloroform
1,2-Dichloroethane
1,1-Dichloroethane
Ethyl acetate
Ethylbenzene
Ethyl ether
Isobutanol (4-Methyl
-1-propanol)
Methylene chloride
Methyl ethyl ketone
(2-Butanone)
Methyl isobutyl
ketone
1,1,1,2-Tetrachloro-
ethane
1,1,2,2-Tetrachloro-
ethane
Tetrachloroethene
Toluene
1,1,1-Trichloro-
ethane
1,1,2-Trichloro-
ethane
Trichloroethene
Trichloro-
fluoromethane
1,1,2-Trichloro-
trifluoroethane
Vinyl chloride
Soil No. 1
(Western)
Avg.
%Rec.* %RSD
44.0 12.4
52.5 68.4
47.8 8.29
55.5 2.91
21.4 16.4
40.6 18.6
64.4 6.76
61.3 8.04
73.4 4.59
31.4 14.5
76.4 9.65
56.2 9.22
48.0 16.4
0.0 ND
47.5 30.3
Soil No. 2
(Eastern)
Avg.
%Rec.* %RSD
43.8
2.25
50.5
70.0
34.8
16.3
49.2
14.6
12.9
49.5
22.3
29.1
41.5
13.1
54.8
16.4
68.7
11.3
22.9
39.3
75.4
4.02
23.2
11.5
55.1
9.72
0.0
ND
42.2
42.9
Soil No. 3
(Western and
Sludge)
Avg.
%Rec.** %RSD
116.0 11.5
49.3 44.9
49.8 36.7
65.5 37.2
36.5 51.5
36.2 41.4
44.2 32.0
61.8 29.1
58.3 33.3
32.0 54.4
23.0
119.8
37.5
36.1
37.3
31.2
61.8
37.7
52.0
37.4
Soil No. 4
(Western and
Sludge)
Avg.
%Rec.*** %RSD
21.3 71.4
51.8 4.6
33.4 41.1
73.0 13.9
21.3 31.5
24.0 34.0
33.0 24.9
45.8 38.6
41.2 37.8
16.8 26.4
11.0 115.5
27.2 28.6
42.0 17.6
76.0 12.2
37.3 16.6
56.7
5.94
61.9
3.94
73.7
31.3
40.6
39.0
81.1
10.3
88.9
2.99
58.3
32.6
39.8
40.3
69.0
6.73
41.1
11.3
50.8
31.5
36.8
23.8
85.3
7.04
58.9
4.15
64.0
25.7
53.6
15.8
45.1
12.7
15.2
17.4
26.2
44.0
18.6
24.2
59.2
8.06
49.3
10.5
45.7
35.2
31.4
37.2
47.2
16.0
33.8
22.8
40.7
40.6
26.2
38.8
76.2
5.72
67.3
8.43
61.7
28.0
46.4
25.4
54.5
11.1
39.4
19.5
38.8
40.9
25.6
34.1
20.7
24.5
12.6
60.1
28.5
34.0
19.8
33.9
18.1
26.7
6.95
58.0
21.5
67.8
15.3
24.8
10.2
20.3
7.17
72.8
25.0
61.0
11.8
25.4
* Triplicate analyses
** Six replicate analyses
*** Five replicate analyses
CD-ROM
1312 - 27
Revision 0
September 1994
DEQ-CFW 00059949
CD-ROM
Figure 1. Rotary Agitation Apparatus
Gas
Bottom Flange
Pressurized Gas —+►^— Z_
Inlet/Outlet Valve Pressure
Gauge
Figure 2. Zero-Headspace Extractor (ZHE)
1312 - 28 Revision 0
September 1994
DEQ-CFW 00059950
CD-ROM
METHOD 1312
SYNTHETIC PRECIPITATION LEACHING PROCEDURE
1312 - 29 Revision 0
September 1994
DEQ-CFW 00059951
CD-ROM
METHOD 1312
SYNTHETIC PRECIPITATION LEACHING PROCEDURE (continued)
e
Discard ISolids Separate liquids
Solids from solids.
Extract
Z
le
Prepare and analyze
N0
al
D_<phase?
each liquid
separately,
mathematically
combine results.
s
Combine extract I
L-LOP
with liquid phase
of waste.
Prepare extract
according to
appropriate
methods.
Analyze extract.
Stop
1312 - 30 Revision 0
September 1994
DEQ-CFW 00059952
Phase Separation
PHASE PROFILE
TIERING PHASE
PHASE DOCUMENTS
TECHNICAL ICAL UPDATE
TACT PHASE
TECHNICAL UPOATEJ
PHASE FACTS Is The TCLP 1311 Extraction The Correct Method For Your Project ?
Links to Related Sites What About SPLP 1312?
Phase Facts:
Sgeciation of#
Metals:
Chromium
New*
Approach to
Soil Handlinq
for Volatiles
SamplingrVolatiles#
In Soil by 6035
MTBE - The*
Next
Challenge
Utilizing TPH*
Methods
Effectively
ILH Method* TP
Overla
Specific*
Petroleum
Compound
Testin
TPH 1664 vs.*
413.1 and
418.1
TCLP vs.*
SPLP
Extraction
Procedures
Anatyziny for#
Metals by
IMMS
TCLP extractions are often misapplied. TCLP is the acronym for Toxicity
Characteristics Leaching Procedure. This methodology was developed to estimate
the mobility of specific inorganic and organic contaminates that are destined for
disposal in municipal landfills.
The TCLP extraction is performed by subjecting the material in question to a
simulated landfill leachate. Acetic acid was chosen as the extraction fluid because it
is the major component of typical municipal landfill leachates. The pH of the acetic
acid buffer solution is maintained at 4.93. This sample/acetic acid mixture is
subjected to an 18 (( 2 hour) rotary extraction, designed to accelerate years of
material/landfill exposure in the shortest possible time. After extraction, the resulting
liquid is subjected to analysis utilizing a list of 39 contaminants that are contained in
the EPA's TCLP Final Rule. These contaminants include eight metals, eleven
volatile organic compounds, twelve semi -volatile organic compounds, six pesticides,
and two herbicides.
Consequently, the TCLP test should be limited to determining if a waste
should be considered a RCRA-regulated waste or, if it can be disposed of in a
municipal (sanitary) landfill.
The TCLP procedure may be misapplied in two ways. First, analysis of compounds
not included in the methods development or TCLP Final Rule may provide
technically invalid results. An example of this would be the utilization of the TCLP
extraction for cyanide. The extraction procedure itself will result in the loss of simple
cyanide complexes, representing the most toxic form of cyanide.
The second way the TCLP is misapplied is by use of the procedure for materials that
are not destined for landfills where interaction with acetic acid is likely. In these
cases an alternative procedure, the SPLP, (Synthetic Precipitation Leaching
Procedure) EPA SW-846 Method 1312 may be appropriate. This procedure is
applicable for materials where the leaching potential due to normal rainfall is to be
determined. Instead of the landfill leachate simulating acetic acid mixture, nitric and
sulfuric acids are utilized in an effort to simulate the acid rains resulting from
airborne nitric and sulfuric oxides.
DEQ-CFW 00059953
ICPJMS•
Qualitative
All Elements
Scan
(TCLP vs SPLP)
The SPLP procedure is useful in those situations where a determination must be made if a
potentially contaminated material, left in situ, will leach toxic substances when exposed to normal
weathering. PSS has experience in performing this procedure. Please call if you have any questions or
need additional information.
PHASE HOMEPAGE
PHASE PROFILE - UTILIZING PHASE - PHASE DOCUMENTS - TECHNICAL UPDATES - CONTACT PHASE
- LINKS TO RELATED SITES - COMPOUND LISTS -
- PHASE FACTS -
$peciation Of Metals: Chromium #Sampling Volatiles In Soil by 5035 * MTBE -The Next Challenge * Utilizing TPH Methods Effectively
0 TPH Method Overlap * Specific Petroleum Compound Testing 0 TPH 1664 vs. 413.1 and 418.1 * TCLP vs. SPLP Extraction Procedures
! Analyzing for Metals by ICPIMS i ICPIMS Qualitative All Elements Scan
This site created and hosted by e.magination network, LLC
DEQ-CFW 00059954