NC Department of Environmental Quality | Division of Water Resources | Laboratory Certification Branch
4405 Reedy Creek Road | 1623 Mail Service Center | Raleigh, North Carolina 27699-1623
919-733-3908
December 4, 2023
5778
Mr. M. Obie Moore
TVA Chickamauga Power Service Center
4601 North Access Road (Building E)
Chattanooga, TN 37415
Subject: North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC
WW/GW LCB) Maintenance Inspection
Dear Mr. Moore:
Enclosed is a report for the inspection performed on October 25, 2023, by Jason Smith. Where
Finding(s) are cited in this report, a response is required. Within thirty days, please supply this
office with a written item for item description of how these Finding(s) were corrected. Please
describe the steps taken to prevent recurrence and include an implementation date for each
corrective action. If the Finding(s) cited in the enclosed report are not corrected, enforcement
actions may be recommended. For Certification maintenance, your laboratory must continue to
carry out the requirements set forth in 15A NCAC 02H .0800.
A copy of the laboratory’s Certified Parameter List at the time of the audit is attached. This list will
not reflect any changes made during the audit. Copies of the checklists completed during the
inspection may be requested from this office. Thank you for your cooperation during the
inspection. If you have questions or need additional information, please contact me at (919) 733-
3908 Ext. 251.
Sincerely,
Anna Ostendorff
Technical Assistance & Compliance Specialist
Division of Water Resources
Attachment
cc: Todd Crawford, Jason Smith, Master File #5778
On-Site Inspection Report
LABORATORY NAME: TVA Chickamauga Power Service Center
NPDES PERMIT #: NC0027332 and NC0027359
ADDRESS: 600 Powerhouse Road
Murphy, NC 28906
CERTIFICATE #: 5778
DATE OF INSPECTION: October 25, 2023
TYPE OF INSPECTION: Field Initial
AUDITOR(S): Jason Smith
LOCAL PERSON(S) CONTACTED:
Obie Moore, Don McGee, Brad Love and Steve Keenum
I. INTRODUCTION:
This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory
Certification Branch (NC WW/GW LCB) to verify its compliance with the requirements of 15A NCAC 02H
.0800 for the analysis of compliance monitoring samples.
II. GENERAL COMMENTS:
The analyst was forthcoming and responded well to suggestions from the auditor. The laboratory has all
the equipment necessary to perform the analyses.
All required Proficiency Testing (PT) Samples have been analyzed for the 2023 PT Calendar Year and the
graded results were 100% acceptable.
Any time changes are made to laboratory procedures, Quality Assurance (QA) and/or Standard Operating
Procedure (SOP) document(s) must be updated and relevant staff retrained. Staff must acknowledge that
they have read and understand the changes as part of the documented training program. The same
requirements apply when changes are made in response to Findings, Recommendations or Comments
listed in this report, to ensure the methods are being performed as stated, references to methods are
accurate, and the QA and/or SOP document(s) is in agreement with each approved practice, test, analysis,
measurement, monitoring procedure or regulatory requirement being used in the laboratory. In some
instances, the laboratory may need to create an SOP to document how new functions or policies will be
implemented. Revisions to the SOPs, based on the Findings, Comments and Recommendations
within this report must be submitted to this office by March 1, 2023.
The laboratory is reminded that SOPs are intended to describe procedures exactly as they are to be
performed. Use of the word “should” is not appropriate when describing requirements (e.g., Quality Control
(QC) frequency, acceptance criteria, etc.). Evaluate all SOPs for the proper use of the word “should”.
Contracted analyses are performed by Earth Environmental Services (Certification #352).
Page 2
#5778 TVA Chickamauga Power Service Center
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Documentation
A. Finding: Chemical containers are not dated when received and when opened.
Requirement: Chemical containers shall be dated when received and when opened. Ref:
15A NCAC 02H .0805 (g) (7).
Comment: This information is recorded on the benchsheet and was documented on the pH
buffer containers during the inspection.
IV. CONCLUSIONS:
Correcting the above-cited Finding will help this laboratory to produce quality data and meet
Certification requirements. The inspector would like to thank the staff for their assistance during
the inspection and data review process. Please respond to the Finding and include supporting
documentation, implementation dates and steps taken to prevent recurrence for each
corrective action.
Report prepared by: Jason Smith Date: November 6, 2023
Report reviewed by: Michael Cumbus Date: November 6, 2023
Certificate Number:5778
Effective Date:6/22/2023
Expiration Date:12/31/2023
Lab Name:TVA Chickamauga Power Service Center
Address:4601 North Access Road (Building E)
Chattanooga, TN 37415
North Carolina Wastewater/Groundwater Laboratory Certification
Certified Parameters Listing
Date of Last Amendment:
The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below.
CERTIFIED PARAMETERS
INORGANIC
pH
SM 4500 H+B-2011 (Aqueous)
TEMPERATURE
SM 2550 B-2010 (Aqueous)
This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions
as set forth in 15A NCAC 2H.0807.