HomeMy WebLinkAbout#618 inspection-final DL2007 On-Site Inspection Report LABORATORY NAME: Gulf Coast Analytical Laboratories, Inc. ADDRESS: 7979 GSRI Avenue Baton Rouge, LA 70820 CERTIFICATE NO: #618 DATE OF INSPECTION: April 12-13, 2007 TYPE OF INSPECTION: Maintenance Inspection EVALUATOR: Mr. David G. Livingston LOCAL PERSON(S) CONTACTED: Ms. Allison Naquin, Mr. Mark Peterman, Ms. Julia Robertson, Ms. Karen Melerine, Ms. Teresa Shealey, Ms. Michelle Rabon, Ms. Betsy Adams, Ms. Stacey Haniford, Mr. Robert Wilson, and Ms. Katherine Brown I. INTRODUCTION: This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for the analysis of environmental samples. II. GENERAL COMMENTS: The laboratory is spacious and well equipped. All facilities and equipment are well maintained. Records are well kept. Some further quality control procedures need to be implemented. Below are Policy changes or reminders for the laboratory since the last inspection. North Carolina Laboratory Certification (NCLC) POLICIES: 1) All North Carolina regulatory data for Waste Water must be reported on a dry weight basis. 2) Each sample collected in North Carolina must clearly indicate the State of North Carolina as the collection site on all record transcripts. 3) Whenever a certified commercial laboratory refers or subcontracts samples to another certified laboratory for analyses, the referring laboratory must supply the date and time samples were collected to insure holding times are met. Subcontracted samples must clearly indicate the State of North Carolina as the collection site on all record transcriptions. Laboratories may subcontract sample fractions, extracts, leachates and other sample preparation products provided that all Rules and requirements of 15A NCAC 02H .0800 are documented. The initial client requesting the analyses must receive the original or a copy of the report made by the laboratory that performs the analyses. Ref: 15A NCAC 2H .08005 (e) (2) 4) Certified analytical weights, Class S, 1 or 2, must be recalibrated and recertified a minimum of every 5 years. 5) NIST certified thermometers must be recalibrated, at a minimum, every five years or the NIST traceable thermometer within expiration date. A new certificate must be issued and maintained for inspection upon request. 6) The blank must not exceed 50% of the PQL/LRL. The lab must obtain a signal in which the constituent concentration in reagent water is 2(1.645)s above the mean of blank analyses (Lower Limit of Detection). The constituent concentration that produces a signal sufficiently greater than the blank, Page 2 Gulf Coast Analytical laboratories, Inc. that it can be detected within specified limits, by good laboratories, during routine operating conditions, is the Limit of Quantitation. Typically it is the concentration that produces a signal 10s above the reagent water blank signal. In most cases, this requires an absorbance value greater than 0.005 units, preferably greater than 0.010 units. Ref: Standard Methods, 20th Edition - Method 1010 C. and 1030 C. 7) For metals analysis of North Carolina Groundwater monitoring well samples, method 3030C must be used. 8) NCLC Requires 5% of samples to be analyzed in duplicate unless the method specifies a higher frequency. 9) NCLC Temperature requirements for sample handling and preservation are the same as stated in 40 CFR Part 136 Table II 10) Each titrant must be standardized monthly. All certified titrants which are purchased, may be used initially without standardization. They must be standardized monthly, thereafter, for as long as they are used. If the normality changes, a new titrant must be purchased or the results must be calculated by a factor derived from the new adjusted normality. 11) NCLC follows the EPA policy of 5% sample duplication unless otherwise stated in a method. Note: I recommend that your laboratory regularly visit our NCLC web site for our must up to date policies. You can visit our website at www.dwqlab.org III. DEFICIENCIES, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS: GENERAL: (A). COMMENT: The laboratory was not checking NC samples for Residual Chlorine. REQUIREMENT: For any sample that requires the removal of residual chlorine, (8260, 8270, 8021, TKN, Ammonia, BOD, etc.) the laboratory must check to ensure that the chlorine has been removed. The residual chlorine check must be performed to verify that the residual chlorine level is below 0.5 mg/L. Commercially available starch iodide paper can be used to perform this determination. WALK THROUGH: (B). COMMENT: The laboratory was not aware of the certification rule requiring the Laboratory to notify the NCLC program when NC samples are received which do not meet sample collection, holding time, or preservation requirements. REQUIREMENT: At any time a laboratory receives samples which do not meet sample collection, holding time, or preservation requirements, the laboratory must notify the sample collector or client and secure another sample if possible. If another sample cannot be secured the original sample may be analyzed, but the results reported must be qualified with the nature of the infraction(s) and the laboratory must notify the State Laboratory about the infraction(s) . The notification must include a statement indicating corrective action taken to prevent the problem for future samples. Ref. 15A NCAC 2H .0805 (a) (7) (N) (C). COMMENT: During the walk through a thermometer, in the metals digestion room, was observed to have a correction factor of 2° C. REQUIREMENT: The thermometer/meter readings must be less than or equal to 1ºC from the NIST thermometer reading. REQUIRED for certified data. Ref: NCLC Policy Page 3 Gulf Coast Analytical laboratories, Inc. TKN: (D). COMMENT: The laboratory does not analyze an organic standard in conjunction with the TKN sample analysis. REQUIREMENT: It is strongly recommended that the laboratory analyze an organic standard with each TKN analysis. A recommended standard for this purpose is Glutamic acid. COD: (E). COMMENT: the laboratory is not documenting the time samples are placed on and the time samples are taken off of the COD reactor. RECOMMENDATION: It is strongly recommended that the laboratory record the time that the samples were placed into the reactor and removed from the reactor in order to verify that the samples were digested for the appropriate amount of time. (F). COMMENT: The laboratory is not documenting the temperature of the COD reactor during digestion. RECOMMENDATION: Document the temperature of the block heater/reactor to meet method requirements. Block heater or oven to operate at 150 +/- 2°C. Ref: Standard Methods 18th Edition 5520 C. 2. c. IV. PAPER TRAIL INVESTIGATION: This consisted of comparing data with the values obtained on laboratory bench worksheets. V. CONCLUSIONS: Correcting the above cited and comments and implementing the requirements/recommendations will help this lab to produce quality data and meet certification requirements. Please respond to all lettered comments. Report prepared by: David G. Livingston Date: 4/23/2007 Report reviewed by: Jeffrey R. Adams Date: 4/27/2007