HomeMy WebLinkAbout#118-final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 118
Laboratory Name: New Bern WWTP Lab
Inspection Type: Municipal Maintenance
Inspector Name(s): Todd Crawford, David Livingston
Inspection Date: March 24, 2009
Date Report Completed: April 1, 2009
Date Forwarded to Reviewer: April 1, 2009
Reviewed by: Ramon Cook
Date Review Completed: May 6, 2009
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding Insp. CP
Unit Supervisor: Dana Satterwhite
Date Received: May 6, 2009
Date Forwarded to Alberta: May 7, 2009
Date Mailed:
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: New Bern WWTP Lab
NPDES PERMIT #: NC0025348
ADDRESS: 307 Glenburnie Drive
New Bern, NC 28560
CERTIFICATE #: 118
DATE OF INSPECTION: March 24, 2009
TYPE OF INSPECTION: Municipal Maintenance
AUDITORS: Todd Crawford, David Livingston
LOCAL PERSONS CONTACTED: Cheryl King
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for the analysis
of environmental samples.
II. GENERAL COMMENTS:
The laboratory was spacious and well organized. All equipment appeared well maintained. Records are well kept and
data appeared accurate. Some further quality control procedures and documentation improvements need to be
implemented.
Findings B, C, and D are new policies that have been implemented by our program since the last inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
pH - Standard Methods, 20th Edition, 4500 H+ B
A. Finding: The column heading for the pH results did not indicate that the results were in Standard Units
(S.U.)
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data must
consist of date collected, time collected, sample site, sample collector, and sample analysis time. The
field bench sheets must provide a space for the signature or initials of the analyst, and proper units of
measure for all analyses. Ref: 15A NCAC 02H .0805 (g) (1).
Total Suspended Solids - Standard Methods, 18th Edition, 2540 D
B. Finding: Deionized water is being put through the filter blank instead of leaving it dry.
Requirement: If pre-prepared filters are not used, the method requires that filters must be weighed to a
constant weight after washing. Repeat cycle of drying, cooling, desiccating, and weighing until a constant
weight is obtained or until weight change is less than 4% of the previous weighing or 0.5 mg, whichever is
less. In lieu of this process, it is acceptable to analyze a single daily dry filter blank to fulfill the method
requirement of drying all filters to a constant weight prior to analysis. This is performed b y washing all filters
as required by the method, then drying and desiccating them. The process of repeatedly heating, cooling,
desiccating, and weighing until a constant weight is documented is not required. Instead, a single dry blank
filter (i.e., no additional rinsing during the analysis) is analyzed. The acceptance criterion for the blank is
a weight change of less than 4% of the filter’s initial weight or 0.5 mg, whichever is less. Ref: NC WW/GW
LC Policy based upon Standard Methods, 20th Edition, 2540 D. (3) (a).
Page 2
#118 New Bern WWTP Lab
Ammonia Nitrogen - Standard Methods, 20th Edition, 4500 NH3 D
C. Finding: A calibration blank is not being analyzed prior to sample analysis. Also, a blank and mid-range
standard are not being analyzed after every tenth sample.
Requirement: The calibration blank and mid-range standard must be analyzed initially (i.e., prior to
sample analysis), after every tenth sample and at the end of each sample group to check for carry over
and calibration drift. If either fall outside established quality control acceptance criteria, corrective action must
be taken (e.g., repeating sample determinations since the last acceptable calibration verification, repeating
the initial calibration, etc.). Ref: NC WW/GW LC Policy based upon Standard Methods, 20th Edition, 1020 B.
(10) (c), 3020 B. (2) (b), and 4020 B. (2).
Ammonia Nitrogen - Standard Methods, 20th Edition, 4500 NH3 D
Total Phosphorus – EPA Method 365.3
D. Finding: Matrix Spike analyses are not performed.
Requirement: Unless the referenced method states a greater frequency, include one Matrix Spike (MS) with
each sample set (batch – 20 or fewer samples) or on a 5% basis. Prepare the MS from a reference source
different from that used for calibration unless otherwise stated in the method. Laboratories analyzing less
than 20 samples per month must analyze at least one MS each month samples are analyzed. If MS results
are out of control, the results must be qualified or the laboratory must take corrective action to rectify the
effect, use another method, or employ the method of standard additions. Refer to the method of choice for
specific acceptance criteria for the MS until the laboratory develops or adopts statistically valid, laboratory-
specific performance criteria for accuracy. Ref: NC WW/GW LC Policy based upon Standard Methods, 20th
Edition, 1020 B. (6).
Dissolved Oxygen - Standard Methods, 18th Edition, 4500 O G
E. Finding: The calibration time is not being recorded. This is considered pertinent information.
Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so
as to be readily available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (A).
IV. PAPER TRAIL INVESTIGATION:
No paper trail performed.
V. CONCLUSIONS:
Correcting the above-cited findings will help this lab to produce quality data and meet certification requirements.
The inspectors would like to thank the staff for its assistance during the inspection and data review proce ss.
Please respond to all findings.
Report prepared by: Todd Crawford Date: April 1, 2009
Report reviewed by: Ramon Cook Date: May 4, 2009