INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 5634
Laboratory Name: Chad Watkins Company
Inspection Type: Field Commercial Initial
Inspector Name(s): Todd Crawford
Inspection Date: 11/5/2014
Date Report Completed: 11/6/2014
Date Forwarded to Reviewer: 11/6/2014
Reviewed by: Nick Jones
Date Review Completed: 11/6/2014
Cover Letter to use: Insp. Initial Insp. Reg.
Insp. No Finding Insp. CP
Corrected Insp. Reg. Delay
Unit Supervisor/Chemist III: Dana Satterwhite
Date Received: 11/6/2014
Date Forwarded to Linda: 11/12/2014
Date Mailed: 11/12/2014
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Chad Watkins Company
ADDRESS: 3200 Wilmington Hwy
Jacksonville, NC 28540
CERTIFICATE #: 5634
DATE OF INSPECTION: November 5, 2014
TYPE OF INSPECTION: Field Commercial Initial
AUDITOR(S): Todd Crawford
LOCAL PERSON(S) CONTACTED: Chad Watkins
I. INTRODUCTION:
This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater
Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A
NCAC 2H .0800 for the analysis of environmental samples.
II. GENERAL COMMENTS:
This was the laboratory’s first inspection. The facility has all the equipment necessary to perform the
analyses. A system for traceability of reagents was in place but did not include all the reagents in use.
Proficiency Testing (PT) samples have been analyzed for all certified parameters for the 2014 proficiency
testing calendar year and the graded results were 100% acceptable.
Contracted analyses are performed by Beacham Laboratory (Certification #1).
Current quality assurance policies for Field Laboratories and approved procedures for the analysis of the
facility’s currently certified parameters were provided at the time of the inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Documentation
Recommendation: The facility is identified on the benchsheet with the NDPES permit number. It is
recommended that the facility name (i.e., Cabin Creek WWTP) also be added for ease of identification.
A. Finding: Error corrections are not performed properly.
Requirement: All documentation errors must be corrected by drawing a single line through the
error so that the original entry remains legible. Entries shall not be obliterated by erasures or
markings. Wite-Out®, correction tape or similar products designed to obliterate documentation are
not to be used. Write the correction adjacent to the error. The correction must be initialed by the
responsible individual and the date of change documented. All data and log entries must be
written in indelible ink. Pencil entries are not acceptable. Ref: Quality Assurance Policies for Field
Laboratories.
Page 2
#5634 Chad Watkins Company
Comment: Corrections were made by obliterating the error.
B. Finding: The laboratory needs to increase the documentation of purchased reagents.
Requirement: All chemicals, reagents, standards and consumables used by the laboratory must
have the following information documented: Date Received, Date Opened (in use), Vendor, Lot
Number, and Expiration Date. A system (e.g., traceable identifiers) must be in place that links
standard/reagent preparation information to analytical batches in which the solutions are used.
Documentation of solution preparation must include the analyst’s initials, date of preparation, the
volume or weight of standard(s) used, the solvent and final volume of the solution. This
information as well as the vendor and/or manufacturer, lot number, and expiration date must be
retained for chemicals, reagents, standards and consumables used for a period of five years.
Consumable materials such as pH buffers and lots of pre-made standards are included in this
requirement. Ref: Quality Assurance Policies for Field Laboratories.
Comment: pH buffers and Total Residual Chlorine (TRC) gel standard information was
documented but TRC buffer and DPD color reagent information was not.
Proficiency Testing
C. Finding: The preparation of TRC Proficiency Testing (PT) samples is not documented.
Requirement: PT samples received as ampules must be diluted according to the PT provider’s
instructions. The preparation of PT samples must be documented in a traceable log or other
traceable format. The diluted PT sample becomes a routine environmental sample and is added
to a routine sample batch for analysis. Ref: Proficiency Testing Requirements, February 20,
2012, Revision 1.2.
Comment: Dating and initialing the instruction sheet for the preparation of the TRC PT would
satisfy the documentation requirement.
D. Finding: The laboratory is not documenting PT sample analyses in the same manner as
environmental samples.
Requirement: All PT sample analyses must be recorded in the daily analysis records as for any
environmental sample. This serves as the permanent laboratory record. Ref: Proficiency Testing
Requirements, February 20, 2012, Revision 1.2.
Comment: The analysis of PT samples is designed to evaluate the entire process used to
routinely report environmental analytical results; therefore, PT samples must be analyzed and the
process documented in the same manner as environmental samples.
pH – Standard Methods, 4500 H+ B-2000
Temperature – Standard Methods, 2550 B-2000
Dissolved Oxygen – Standard Methods, 4500 O G-2001
Total Residual Chlorine – Hach 10014 (ULR)
E. Finding: The laboratory benchsheet was lacking pertinent data: Instrument identification.
Requirement: The following must be documented in indelible ink whenever sample analysis is
performed: Instrument Identification. Ref: NC WW/GW LC Approved Procedure for the Analysis
of pH, NC WW/GW LC Approved Procedure for the Analysis of Temperature, NC WW/GW LC
Page 3
#5634 Chad Watkins Company
Approved Procedure for the Analysis of Dissolved Oxygen, NC WW/GW LC Approved Procedure
for the Analysis of Total Residual Chlorine.
F. Finding: The analyst is not signing or initialing laboratory benchsheets.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data
must consist of date collected, time collected, sample site, sample collector, and sample analysis
time. The field benchsheets must provide a space for the signature or initials of the analyst, and
proper units of measure for all analytes. Ref: 15A NCAC 2H .0805 (g) (1).
Comment: The benchsheet had the analyst’s name printed on it but it was not actually signed or
initialed. A handwritten signature or initial is required to verify the data produced.
pH – Standard Methods, 4500 H+ B-2000
G. Finding: The laboratory benchsheet was lacking pertinent data: Units of measure.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified
Data must consist of date collected, time collected, sample site, sample collector, and sample
analysis time. The field bench sheets must provide a space for the signature or initials of the
analyst, and proper units of measure for all analyses. Ref: 15A NCAC 02H .0805 (g) (1).
Comment: The unit of measure for pH is Standard Units, abbreviated as S.U.
H. Finding: Values were reported that exceed the method specified accuracy of 0.1 units.
Requirement: By careful use of a laboratory pH meter with good electrodes, a precision of
±0.02 unit and an accuracy of ±0.05 unit can be achieved. However, ± 0.1 pH unit represents
the limit of accuracy under normal conditions, especially for measurement of water and poorly
buffered solutions. For this reason, report pH values to the nearest 0.1 pH unit. Ref: Standard
Methods, 4500 H+ B-2000, (6).
Total Residual Chlorine – HACH 10014 (ULR)
Comment: The DPD and buffer reagents were both expired. The Quality Assurance Policies for Field
Laboratories document states: Adherence to manufacturer expiration dates is required. Chemicals,
reagents, standards, consumables exceeding the expiration date can no longer be considered reliable.
If the expiration is only listed as a month and year (with no specific day of the month), the last day of
the month will be considered the actual date of expiration. Monitor materials for changes in appearance
or consistency. Any changes may indicate potential contamination and the item should be discarded,
even if the expiration date is not exceeded. If no expiration date is given, the laboratory must have a
policy for assigning an expiration date. If no date received or expiration date can be determined, the
item should be discarded. Demonstration of acceptable corrective action (i.e., all expired reagents were
thrown away and reagents that were within their expiration dates brought out and put into use) was
performed during the inspection. No further response is necessary for this finding.
I. Finding: The laboratory benchsheet was lacking pertinent data: Meter calibration time.
Requirement: The following must be documented in indelible ink whenever sample analysis is
performed: Meter calibration and meter calibration time(s). Ref: NC WW/GW LC Approved
Procedure for the Analysis of Total residual Chlorine.
Page 4
#5634 Chad Watkins Company
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing field testing records and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Resources. Data were
reviewed for Cabin Creek Campground & RV Park WWTP (NC0034339) for July, August and
September, 2014. No transcription errors were detected. The facility appears to be doing a good job of
accurately transcribing data.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendation will help this lab to produce quality
data and meet certification requirements. The inspector would like to thank the staff for its assistance during the
inspection and data review process. Please respond to all findings.
Report prepared by: Todd Crawford Date: November 6, 2014
Report reviewed by: Nick Jones Date: November 6, 2014