Loading...
HomeMy WebLinkAbout#5606 - 09 - 2011 - FINALINSPECTION REPORT ROUTING SHEET To be attached to all inspection reports in-house only. Laboratory Cert. #. Laboratory Name: Inspection Type: Inspector Name(s): Inspection Date: Date Report Completed: Date Forwarded to Reviewe Reviewed by: Date Review Completed: Cover Letter to use: X Unit Supervisor: Date Received: Date Forwarded to Linda: Date Mailed: 5606 Town of Dallas WTP Field Initial Chet Whiting, Gary Francies September 9, 2011 September 18, 2011 r: September 19, 2011 David Livingston 9/19/2011 Insp. Initial _ Insp. Reg. Insp. No Finding _ Insp. CID _Corrected Gary Francies 9/20/2011 9/30/2011 r _99 CERTIFICATE #: �_ 1:to] A1k61»01IQk'I fl-kT;M81M• r wl LOCAL PERSON(S) CONTACTED: oms Town of Dallas WTP 210 North Holland Street Dallas, NC 28034 5606 September 9, 2011 Field Initial Chet Whiting, Gary Francies Amelia Hamerick This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for the analysis of environmental samples. GENERAL COMMENTS: The facility has submitted an application to obtain Field Parameter certification. Compliance monitoring analyses are currently being analyzed by a North Carolina Certified Laboratory. The laboratory was analyzing process control samples, and using practice (training) analyses, to fine tune laboratory systems in preparation of generating certified data. The facility has all the equipment necessary to perform the analyses. fll'�91!IQIW�� IZ1=IIT, I*', I&-!L•T•7IT, l�il �i _ ► �' a ►� ► = r A. Finding: Copies of analytical methods were not available to each analyst. Requirement: A copy of each approved analytical procedure must be available to each analyst. Ref: 15A NCAC 2H .0805 (g) (3). Comment: Laboratory personnel indicated they intended to purchase a copy of Standard Methods 21st Edition. pH — Standard Methods, 215t Edition, 4500 H+ B Comment: The temperature sensing devices on the pH meter had been calibrated against an NIST certified thermometer; however the corrections were not posted. Technical Assistance for Field Analysis of pH states: All temperature sensing devices on meters must be calibrated against a NIST certified or NIST traceable thermometer annually (every 12 months) and proper corrections made and documented. The meter reading must be less than 1°C from the NIST certified reading to be acceptable. Document the serial number and manufacturer of the NIST thermometer that was used in the comparison. Document any correction that applies (even if zero) on both the meter and on a 0 Page 2 #5603 Town of Dallas WTP separate sheet to be filed. Acceptable corrective action was implemented during the audit. The proper corrections were posted on the pH meter. No further response is necessary for this finding B. Finding: pH values are reported that exceed the method specified accuracy (0.1 s.u) for this analysis. Requirement: A variance of ± 0.1 pH unit represents the limit of accuracy under normal conditions, especially for measurement of water and poorly buffered solutions. For this reason, report pH values to the nearest 0.1 pH unit. Ref: Standard Methods, 20th Edition, 4500 H+ B. (6). C. Finding: Proper units of measure are not being documented on bench and calibration sheets. Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data must consist of date collected, time collected, sample site, sample collector, and sample analysis time. The field bench sheets must provide a space for the signature or initials of the analyst, and proper units of measure for all analyses. Ref: 15A NCAC 2H .0805 (g) (1). D. Finding: Instances were observed where the pH calibration check exceeded ± 0.1 s.u. and no corrective action was taken. Requirement: For routine work, use a pH meter accurate and reproducible to 0.1 pH unit with a range of 0 to 14, equipped with a temperature compensation device. Follow all manufacturers' recommendations for the calibration of the meter each analysis day. In all cases, the meter must be calibrated with at least two buffers. Calibrating with the pH 4 and pH 10 buffers meets standard methods requirements and brackets the pH range for most monitoring (with the exception of sludge) scenarios. The calibration; however, must bracket the range of the samples being analyzed. In addition to the calibration standards, the meter must be verified with a third calibration standard (e.g., pH 7 buffer). In order to meet the above -specified criteria, the standard must read within a range of pH 6.9 to 7.1 to be acceptable. Ref: Technical Assistance for Field Analysis of pH. Comment: The meter is calibrated with a 4 and 7 buffer and checked with a 10. Recommended: It is recommended that the manufacturer's manual be referred to for troubleshooting guidance. No paper trail performed. V. CONCLUSIONS: Correcting the above -cited findings and implementing the recommendation will help this lab to produce quality data and meet certification requirements. The inspectors would like to thank the staff for its assistance during the inspection and data review process. Please respond to all findings. Report prepared by: Chet Whiting Report reviewed by: David Livingston Date: September 18, 2011 Date: September 19, 2011 Governor September 30, 2011 5606 Ms. Amelia Hamerick Town of Dallas WTP 210 North Holland Street Dallas, NC 28034 SUBJECT: North Carolina Wastewater/Groundwater Laboratory Certification (NC VWV/GW LC) Initial Inspection Dear Ms. Hamerick: Enclosed is a report for the inspection performed on September 9, 2011 by Chet Whiting and myself. Where finding(s) are cited in this report, a response is required. Within thirty days of receipt, please supply this office with a written item for item description of how these finding(s) were corrected. If the finding(s) cited in the enclosed report are not corrected, enforcement actions may be recommended. For certification maintenance, your laboratory must continue to carry out the requirements set forth in 15A NCAC 2H .0800. Copies of the checklists completed during the inspection may be requested from this office. Thank you for your cooperation during the inspection. If you wish to obtain an electronic copy of this report by email, or if you have questions or need additional information please contact me at 828-296-4677. Sincerely, Gary Francies Unit Supervisor Laboratory Section CC: Chet Whiting Gary Francies Master File DENR DWQ Laboratory Section NC Wastewater/Groundwater Laboratory Certification Branch One 1623 Mail Service Center, Raleigh, North Carolina 27699-1623 NorthCarohna Location: 4405 Reedy Creek Road. Raleigh, North Carolina 27607-6445 Phone: 919-733-3908 \ FAX: 919-733-6241 vvatumilff Internet: www.dwglab.org An Equal Opportunity\ Affirmative Action Employer Customer Service:1-877-623-6748 www.ncwaterqual4.org