HomeMy WebLinkAbout#191 2011-final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: #191
Laboratory Name: Roxboro WWTP Laboratory
Inspection Type: Municipal Maintenance
Inspector Name(s): Jeffrey R. Adams
Inspection Date: July 28, 2011
Date Report Completed: August 18, 2011
Date Forwarded to Reviewer: August 18, 2011
Reviewed by: Todd Crawford
Date Review Completed: August 18, 2011
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding Insp. CP ___ Corrected
Unit Supervisor: Dana Satterwhite
Date Received: August 19, 2011
Date Forwarded to Linda: August 24, 2011
Date Mailed: August 24, 2011
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Roxboro WWTP Laboratory
NPDES PERMIT #: NC0021024 and NC0003042
ADDRESS: P.O. Box 128
Roxboro, NC 27573
CERTIFICATE #: 191
DATE OF INSPECTION: July 28, 2011
TYPE OF INSPECTION: Municipal Maintenance
AUDITOR(S): Jeffrey R. Adams
LOCAL PERSON(S) CONTACTED: Crystal Shotwell and Derek Clayton
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for
the analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was clean and well organized. The facility has all the equipment necessary to perform the
analyses. Records were well organized; however, some quality control procedures need to be
implemented.
As a reminder, the 2011 proficiency testing results must be submitted by your vendor and received in the
certification office no later than October 31, 2011.
The laboratory was given a packet containing North Carolina Laboratory Certification quality control
requirements and policies during the inspection.
Findings B and C are new policies that have been implemented since the last inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
General Laboratory
A. Finding: The analytical balance weights have not been verified against ASTM standard
weights.
Requirement: ASTM Class 1 and 2 weights must be verified at least every 5 years. ASTM
Class 1 weights (20 g to 25 kg) and ASTM Class 2 weights (10 g to 1 mg) are equivalent to the
NBS Class S weights specified in 15A NCAC 2H .0805 (a) (7) (K). Ref: North Carolina
Wastewater/Groundwater Laboratory Certification Policy. See attachment, titled Weight
Verification, for additional guidance.
Page 2
#191 Roxboro WWTP Laboratory
B. Finding: Auto-pipettors are not calibrated twice per year.
Requirement: Mechanical volumetric liquid-dispensing devices (e.g., fixed and adjustable auto-
pipettors, bottle-top dispensers, etc.) must be calibrated at least twice per year, approximately
six months apart and documented. Each liquid-dispensing device must meet the manufacturer’s
statement of accuracy. Ref: North Carolina Wastewater/Groundwater Laboratory Certification
Policy. (See attachment, titled Calibration of Mechanical Volumetric Liquid-Dispensing
Devices Procedure, for an example procedure for calibrating auto-pipettors. NOTE: Other
procedures may also be suitable or this may be accomplished by a commercial
calibration company.)
BOD – Standard Methods, 18th Edition, 5210 B
D. Finding: BOD blank values were >0.20 mg/L on several data sets.
Requirement: The DO uptake in 5 days must not be more than 0.20 mg/L and preferably not
more than 0.10 mg/L, before making seed corrections. This means readings should be made to
two decimal places. If the dilution water blank exceeds 0.20 mg/L, discard all data for tests
using this dilution water or clearly identify such sample in data records. Ref: North Carolina
Wastewater/Groundwater Laboratory Certification Policy based upon Standard Methods, 21st
Edition, 5210 B. (6) (c).
Recommendation: The BOD blank values have been consistently averaging >0.20 mg/L. It is
recommended that the laboratory seek an outside water source and set up and analyze a
comparison check of a series of water blanks, using the in-house water vs. the outside water.
Comment: The laboratory has been qualifying the data on the DMR reports as required;
however, steps must be taken to correct the problem.
pH – Standard Methods, 18th Edition, 4500 H+ B
pH – EPA SW–846 Method 9040 C
E. Finding: The laboratory is analyzing and reporting sludge by SW -846 Method 9040 C, but is
not certified for this method.
Requirement: Municipal and Industrial Laboratories are required to obtain certification for
parameters which will be reported to the State to comply with State surface water monitoring,
groundwater, and pretreatment Rules. Ref: 15A NCAC 02H .0804 (a). Please submit an
Amendment to Certification form requesting the addition of pH SW-846 Method 9040 C with
the response to this inspection report.
Comment: The laboratory is analyzing sludge samples with a solid content of 3% and a water
content of 97%.
pH – Standard Methods, 18th Edition, 4500 H+ B
Total Residual Chlorine – Standard Methods, 18th Edition, 4500 Cl G
Comment: The laboratory also analyses samples for the water treatment plant for Dissolved Oxygen,
Settleable Solids, pH and Total Residual Chlorine (TRC). After review of the pH and Total Residual
Chlorine data of the water treatment plant and observing the times being recorded on the benchsheets,
there is a concern that the distance between the water plant and the wastewater laboratory may be too
great to meet the 15 minute holding time for pH and TRC. The laboratory states the travel time
Page 3
#191 Roxboro WWTP Laboratory
between locations is approximately 8 to 10 minutes, therefore making it difficult to meet the 15 minute
sample hold time.
Recommendation: To ensure compliance with Clean Water Act requirements, it is recommended that
the water treatment plant obtain a separate field laboratory certification. The requirements for initial
certification of field laboratories are described on the Laboratory Certification website at
http://portal.ncdenr.org/web/wq/lab/cert/field. To begin the process, simply complete the Field
Parameter Environmental Laboratory Certification application and analyze the appropriate PT samples.
The inspector will be glad to assist the water treatment plant personnel with obtaining certification.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing laboratory benchsheets and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for the City of Roxboro WWTP (NPDES permit #NC0021024) and the City of Roxboro WTP
(NPDES permit #NC0003042) for February, March and April, 2011. The following error was noted for
permit #NC0021024:
Date Parameter Location Value on Benchsheet Value on DMR
5/2/11 Total Residual Chlorine Effluent 37 µg/L < 15 µg/L
In order to avoid questions of legality, it is recommended that you contact the appropriate Regional
Office for guidance as to whether an amended Discharge Monitoring Report will be required. A copy of
this report will be forwarded to the Regional Office.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendations will help this lab to
produce quality data and meet certification requirements. The inspector would like to thank the staff for
its assistance during the inspection and data review process. Please respond to all findings.
Report prepared by: Jeffrey R. Adams Date: August 18, 2011
Report reviewed by: Todd Crawford Date: August 18, 2011
Calibration of Mechanical Volumetric Liquid-Dispensing Devices Procedure
A laboratory may use the following method for the gravimetric determination of measurement error of
mechanical volumetric liquid-dispensing devices (e.g., fixed and adjustable auto-pipettors, bottle-top
dispensers, etc.) used to dispense volumes greater than or equal to 100 μl.
Apparatus and Equipment
Weighing vessel (of sufficient capacity to hold all volumes dispensed for the test and a 3:1 height-to-
diameter ratio is recommended to minimize evaporation)
Reagent water (e.g., distilled or deionized water)
Analytical balance (accurate to 0.0001 g for 100 μl to 10 ml, and accurate to 1 mg for 10 ml to 200 ml
dispensed volumes)
Calibrated thermometer (graduated, at a minimum, in 0.5 ºC increments)
Documentation Requirements
Date
Analyst's signature or initials
Device identification (if there is no unique identification such as serial number, assign a laboratory
identification)
Temperature of the reagent water
Nominal or test volume
Weights of dispensed volumes
Procedure
1. Allow apparatus and reagent water to equilibrate to room temperature.
2. Check the calibration of the balance.
3. Wet the liquid-dispensing device and lubricate the piston insuring smooth operation by pipetting
several volumes to waste.
4. Weigh the vessel and tare the balance.
5. Dispense a minimum of five aliquots of water into the vessel, weighing and recording the mass
for each replicate.
6. Calculate the following:
m = mean mass = total of individual weight measurements ÷
total number of measurements
Z = Z correction factor obtained from Table 1 using the temperature of the test liquid and
current barometric pressure.
V = volume equivalent (or corrected volume) of mass measured = m x Z
RE = relative error (%) = [Vm - Vo] x 100
Vo
where:
Vm = corrected mean volume = total of corrected volumes ÷
total number of dispensed volumes
Vo = nominal or test volume
S = standard deviation = For each replicate, subtract the mean volume from the corrected
volume, then square the value. Sum the squared values of all replicates, and then divide the sum by n-
1 (the number of measurements minus 1). Take the square root of this value to get the standard
deviation.
CV = coefficient of variation (%) = s x 100
Vm
To simplify this process, spreadsheets for calculating and documenting the relative error and
coefficient of variation can be found on the NC Wastewater/Groundwater Laboratory Certification (NC
WW/GW LC) website at http://h2o.enr.state.nc.us/lab/download.htm.
The Relative Error (a measure of accuracy) and Coefficient of Variation (a measure of precision) must
not be greater than the manufacturer's specifications. If the results are greater than the manufacturer's
specification, repeat the procedure. If the results remain greater, follow the manufacturer's instructions
for maintenance and adjustment and repeat the procedure. If repeated calibration attempts do not yield
results within the manufacturer's specifications, the liquid-dispensing device must be repaired or
replaced.
Note: Microliter pipettors may need to be tested under more rigorously controlled conditions or sent to
a calibration service for testing. Since most analytical balances used in environmental laboratories read
to only four decimal places, it may be necessary, when checking microliter pipettors (less than 1000
μl), to include the weight of the weighing vessel in each replicate and subtract it and subsequent
aliquots out manually in order to have a measurable mass for the purpose of this verification. Contact
your auditor if you have questions.
Table 1. Z Correction Factor Table.
Temperat
ure C
Air pressure, kPa (mbar)
80
(800)
85 (850) 90 (900) 95 (950) 100
(1000)
101.3
(1013)
105
(1050)
15.0 1.0017 1.0018 1.0019 1.0019 1.0020 1.0020 1.0020
15.5 1.0018 1.0019 1.0019 1.0020 1.0020 1.0020 1.0021
16.0 1.0019 1.0020 1.0020 1.0021 1.0021 1.0021 1.0022
16.5 1.0020 1.0020 1.0021 1.0021 1.0022 1.0022 1.0022
17.0 1.0021 1.0021 1.0022 1.0022 1.0023 1.0023 1.0023
17.5 1.0022 1.0022 1.0023 1.0023 1.0024 1.0024 1.0024
18.0 1.0022 1.0023 1.0023 1.0024 1.0025 1.0025 1.0025
18.5 1.0023 1.0024 1.0024 1.0025 1.0025 1.0026 1.0026
19.0 1.0024 1.0025 1.0025 1.0026 1.0026 1.0027 1.0027
19.5 1.0025 1.0026 1.0026 1.0027 1.0027 1.0028 1.0028
20.0 1.0026 1.0027 1.0027 1.0028 1.0028 1.0029 1.0029
20.5 1.0027 1.0028 1.0028 1.0029 1.0029 1.0030 1.0030
21.0 1.0028 1.0029 1.0029 1.0030 1.0031 1.0031 1.0031
21.5 1.0030 1.0030 1.0031 1.0031 1.0032 1.0032 1.0032
22.0 1.0031 1.0031 1.0032 1.0032 1.0033 1.0033 1.0033
22.5 1.0032 1.0032 1.0033 1.0033 1.0034 1.0034 1.0034
23.0 1.0033 1.0033 1.0034 1.0034 1.0035 1.0035 1.0036
23.5 1.0034 1.0035 1.0035 1.0036 1.0036 1.0036 1.0037
24.0 1.0035 1.0036 1.0036 1.0037 1.0037 1.0038 1.0038
24.5 1.0037 1.0037 1.0038 1.0038 1.0039 1.0039 1.0039
25.0 1.0038 1.0038 1.0039 1.0039 1.0040 1.0040 1.0040
25.5 1.0039 1.0040 1.0040 1.0041 1.0041 1.0041 1.0042
26.0 1.0040 1.0041 1.0041 1.0042 1.0042 1.0043 1.0043
26.5 1.0042 1.0042 1.0043 1.0043 1.0044 1.0044 1.0044
27.0 1.0043 1.0044 1.0044 1.0045 1.0045 1.0045 1.0046
27.5 1.0045 1.0045 1.0046 1.0046 1.0047 1.0047 1.0047
28.0 1.0046 1.0046 1.0047 1.0047 1.0048 1.0048 1.0048
28.5 1.0047 1.0048 1.0048 1.0049 1.0049 1.0050 1.0050
29.0 1.0049 1.0049 1.0050 1.0050 1.0051 1.0051 1.0051
29.5 1.0050 1.0051 1.0051 1.0052 1.0052 1.0052 1.0053
30.0 1.0052 1.0052 1.0053 1.0053 1.0054 1.0054 1.0054
Z correction factors are for distilled water as a function of liquid test temperature and air
pressure.
Z values are in microliters per milligram.
Reprinted from ISO 8655-6:2002(E).
If your laboratory does not have a mercury barometer, the true local barometric pressure for locations throughout
North Carolina may be obtained from the North Carolina State University, State Climate Office of North Carolina
website at http://www.nc-climate.ncsu.edu/.
Weight Verification (NC WW/GW LC Policy 05/23/2008)
ASTM Class 1 and 2 weights must be verified at least every 5 years. ASTM Class 1 weights (20 g to 25
kg) and ASTM Class 2 weights (10 g to 1 mg) are equivalent to the NBS Class S weights specified in
15A NCAC 2H .0805 (a) (7) (K). Verification may be accomplished by:
1. Sending laboratory weights back to the manufacturer for recertification - reference weights shall
be calibrated by a body that can provide traceability to ASTM specifications, or
2. Checking laboratory weights against certified reference weights (i.e., weights that have been
recertified as above) and found to be within ASTM Class 1 or Class 2 tolerances (see table
below) - often the balance service technician may provide this service.
Note: Although some manufacturers will assign a one-year calibration due date, 5 years is considered
an acceptable calibration interval due to the limited use of the reference weight set.
Documentation of weight verifications or recertification must be maintained for 5 years. I f the condition
of a weight(s) is in question at any time due to damage (e.g., corrosion, nicks, scratching, etc.), the
laboratory must have that weight(s) re-verified as described above.
Maximum tolerances (Ref. ASTM E 617-97, 2003)
Denomination Maximum tolerance for
ASTM Class 1 and 2
weights, (± mg)
500 g 1.2
300 g 0.75
200 g 0.50
100 g 0.25
50 g 0.12
30 g 0.074
20 g 0.074
10 g 0.074
5 g 0.054
3 g 0.054
2 g 0.054
1 g 0.054
500 mg 0.025
300 mg 0.025
200 mg 0.025
100 mg 0.025
50 mg 0.014
30 mg 0.014
20 mg 0.014
10 mg 0.014
5 mg 0.014
3 mg 0.014
2 mg 0.014
1 mg 0.014