INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 5585
Laboratory Name: Envision Plastics
Inspection Type: Field Initial
Inspector Name(s): Jeffrey R. Adams
Inspection Date: August 23, 2012
Date Report Completed: August 28, 2012
Date Forwarded to Reviewer: August 28, 2012
Reviewed by: Todd Crawford
Date Review Completed: August 31, 2012
Cover Letter to use: X Insp. Initial Insp. Reg. Insp. No Finding Insp. CP __ Corrected
Unit Supervisor: Dana Satterwhite
Date Received: September 11, 2012
Date Forwarded to Linda: October 5, 2012
Date Mailed: October 8, 2012
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Envision Plastics
NPDES PERMIT #: NC0024881
ADDRESS: 606B Walters Street
Reidsville, NC 27320
CERTIFICATE #: 5585
DATE OF INSPECTION: August 23, 2012
TYPE OF INSPECTION: Field Initial
AUDITOR(S): Jeffrey R. Adams
LOCAL PERSON(S) CONTACTED: Melvernia Holly
I. INTRODUCTION:
This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater
Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A
NCAC 2H .0800 for the analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was clean and records were well organized. Some quality control procedures need to be
implemented. The laboratory is currently analyzing samples only for pH for pretreatment monitoring. The
facility reports data to the City of Reidsville pretreatment program.
The laboratory was given a packet containing North Carolina Laboratory Certification quality control
policies and approved procedures for field analyses during the inspection.
Contracted analyses are performed by Meritech, Inc. (Certification #165).
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Traceability
A. Finding: The laboratory needs to increase the documentation of purchased materials and
reagents (i.e., pH buffers).
Requirement: : All chemicals, reagents, standards and consumables used by the laboratory
must have the following information documented: Date Received, Date Opened (in use),
Vendor, Lot Number, and Expiration Date. A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the
solutions are used. Documentation of solution preparation must include the analyst’s initials,
date of preparation, the volume or weight of standard(s) used, the solvent and final volume of
the solution. This information as well as the vendor and/or manufacturer, lot number, and
expiration date must be retained for chemicals, reagents, standards and consumables used for
a period of five years. Consumable materials such as pH buffers and lots of pre-made
Page 2
#5585 Envision Plastics
standards are included in this requirement. Ref: Quality Assurance Policies for Field
Laboratories.
Proficiency Testing
B. Finding: The laboratory is not documenting Proficiency Testing (PT) sample analyses in the
same manner as environmental samples (i.e., on a laboratory benchsheet or field notebook).
Requirement: All PT sample analyses must be recorded in the daily analysis records as for any
environmental sample. This serves as the permanent laboratory record. Ref: Proficiency Testing
Requirements, February 20, 2012, Revision 1.2.
pH – Standard Methods, 18th Edition, 4500 H+ B
C. Finding: The proper units of measure (i.e., S.U. for pH) are not documented on the
benchsheet.
Requirement: Data pertinent to each analysis must be maintained for five years. Certified Data
must consist of date collected, time collected, sample site, sample collector, and sample
analysis time. The field bench sheets must provide a space for the signature or initials of the
analyst, and proper units of measure for all analyses. Ref: 15A NCAC 02H .0805 (g) (1).
D. Finding: The pH meter calibration does not bracket the range of samples being analyzed.
Requirement: For routine work, use a pH meter accurate and reproducible to 0.1 pH unit with a
range of 0 to 14, equipped with a temperature compensation device. Follow all manufacturers’
recommendations for the calibration of the meter each analysis day. The meter must be
calibrated with at least two buffers. In addition to the calibration standards, the meter must be
verified with a third calibration standard. The calibration and check standard buffer must bracket
the range of the samples being analyzed. Ref: NC WW/GW LC Approved Procedure for Field
Analysis of pH.
Comment: Pretreatment effluent sample values were observed to sometimes exceed 10 S. U.
(i.e., 10.39 and 10.54), which exceeds the highest calibration buffer.
Recommendation: It is recommended that the laboratory calibrate its instrument with the 4.0,
7.0 and 10 buffers (or per manufacturer’s instructions) and analyze the 11.0 or 12.0 buffer as a
check standard in order to bracket the range of the samples being analyzed.
E. Finding: The laboratory is not analyzing a check buffer on each analysis day.
Requirement: The meter must be calibrated with at least two buffers. In addition to the
calibration standards, the meter must be verified with a third calibration standard. The
calibration and check standard buffer must bracket the range of the samples being analyzed.
The check standard buffer must read within a range of ± 0.1 pH units to be acceptable. Ref: NC
WW/GW LC Approved Procedure for Field Analysis of pH.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing field testing records and contract lab reports to semi-annual and
daily Industrial User Pretreatment Permit (IUP) Reports submitted to the City of Reidsville Pretreatment
Division. Data were reviewed for Envision Plastics (NPDES permit #NC0024881) and City of Reidsville
Page 3
#5585 Envision Plastics
(Industrial User Pretreatment Permit #0011) for May, June and July, 2012. No transcription errors were
detected. It appears the laboratory is doing a good job of accurately transcribing data.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendation will help this lab to produce
quality data and meet certification requirements. The inspector would like to thank the staff for its
assistance during the inspection and data review process. Please respond to all findings.
Report prepared by: Jeffrey R. Adams Date: August 28, 2012
Report reviewed by: Todd Crawford Date: August 31, 2012