Loading...
HomeMy WebLinkAbout#5590_2012_0823_JA_FINALBeverly Eaves Perdue Charles Wakild, P. E. Dee Freeman Governor Director Secretary October 2, 2012 5590 Mr. Johnny Somers DyStar LP 209 Watlington Industrial Dr. Reidsville, NC 27320 Subject: North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) Initial Inspection Dear Mr. Somers: Enclosed is a report for the inspection performed on August 23, 2012 by Mr. Jeff Adams. Where finding(s) are cited in this report, a response is required. Within thirty days of receipt, please supply this office with a written item for item description of how these finding(s) were corrected. If the finding(s) cited in the enclosed report are not corrected, enforcement actions may be recommended. For certification maintenance, your laboratory must continue to carry out the requirements set forth in 15A NCAC 2H .0800. Copies of the checklists completed during the inspection may be requested from this office. Thank you for your cooperation during the inspection. If you wish to obtain an electronic copy of this report by email or if you have questions or need additional information, please contact us at 919-733-3908. —'1'&nr erely, Dana ''atterwhite Certification Unit Supervisor Laboratory Section Enclosure cc: Jeff Adams Gary Francies DENR DWQ Laboratory Section NC Wastewater/Groundwater Laboratory Certification Branch 1623 Mail Service Center, Raleigh, North Carolina 27699-1623 Location: 4405 Reedy Creek Road. Raleigh, North Carolina 27607-6445 Phone: 919-733-3908 � FAX: 919-733-6241 Internet: www,dwglab.org An Equal Opportunity l Affirmative Action Employer INSPECTION REPORT ROUTING SHEET To be attached to all inspection reports in-house only. Laboratory Cert. #: Laboratory Name: Inspection Type: Inspector Name(s): Inspection Date: Date Report Completed: Date Forwarded to Reviewer: Reviewed by: Date Review Completed: #5590 Dystar LP Field Initial Jeffrey R. Adams August 23, 2012 August 29, 2012 August 29, 2012 Jason Smith September 10, 2012 Cover Letter to use: ® Insp. Initial ❑ Insp. Reg. ❑ Insp. No Finding ❑ Insp. CP ❑ Corrected Unit Supervisor Dana Satterwhite'�� Date Received: September 18, 2012 Date Forwarded to Linda: October 2 2012 Date Mailed: October 2 2012 LABORATORY NAME: NPDES PERMIT #: ADDRESS: CERTIFICATE #: DATE OF INSPECTION: TYPE OF INSPECTION: AUDITOR(S): LOCAL PERSON(S) CONTACTED: lit kI7��C•P►>G Dystar LP NCG060000 209 Watlington Industrial Dr. Reidsville, NC 27320 5590 August 23, 2012 Field Initial Jeffrey R. Adams Johnny Somers and Whitney Rogers This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A NCAC 2H .0800 for the analysis of environmental samples. GENERAL COMMENTS: The laboratory was clean and equipment appeared well maintained. Records were well organized. This facility is operating a pre-treatment program under the administration of the City of Reidsville. The laboratory was given a packet containing North Carolina Laboratory Certification quality control requirements and approved procedures during the inspection. Contracted analyses are performed by Pace Analytical Services, Inc. - Huntersville (Certification #12) and Meritech, Inc. (Certification #165). Traceability A. Finding: The laboratory needs to increase the documentation of purchased materials and reagents (i.e., pH buffers). Requirement: All chemicals, reagents, standards and consumables used by the laboratory must have the following information documented: Date Received, Date Opened (in use), Vendor, Lot Number, and Expiration Date. A system (e.g., traceable identifiers) must be in place that links standard/reagent preparation information to analytical batches in which the solutions are used. Documentation of solution preparation must include the analyst's initials, date of preparation, the volume or weight of standard(s) used, the solvent and final volume of the solution. This information as well as the vendor and/or manufacturer, lot number, and expiration date must be retained for chemicals, reagents, standards and consumables used for a period of five years. Consumable materials such as pH buffers and lots of pre -made In Page 2 #5590 Dystar LP standards are included in this requirement. Ref: Quality Assurance Policies for Field Laboratories. pH — Standard Methods, 18th Edition, 4500 H+ B B. Finding: The pH meter is not equipped with Automatic Temperature Compensation (ATC). Requirement: The laboratory must have a pH meter consisting of potentiometer, a glass electrode, a reference electrode or combination electrode and a temperature compensating device. Ref: Standard Methods, H B-2000, 4500 H + B. (2) (a). A copy of the manufacturer's specifications showing that the meter contains a temperature compensating device, documentation showing the purchase of an ATC probe compatible to the meter currently used, or a new meter equipped with a temperature compensating device, must be submitted with the response to this report. Requirement: Each facility must have glassware, chemicals, supplies, equipment, and a source of distilled or de -ionized water that will meet the minimum criteria of the approved methodologies. Ref: 15A NCAC 02H .0805 (g) (4). Temperature — Standard Methods, 18th Edition, 2550 B C. Finding: The laboratory is checking the temperature on the mercury glass thermometer semi- annually against a National Institute of Standards and Technology (NIST) traceable thermometer, but is not posting the corrections on the thermometer. Requirement: The temperature correction (even if it is zero) must be posted on the temperature sensing device, as well as in hard copy format (to be retained for 5 years). Ref: Approved Procedure for Field Analysis of Temperature. The paper trail consisted of comparing field testing records and contract lab reports to semi-annual and daily Industrial User Pretreatment Permit (IUP) reports submitted to the City of Reidsville Pretreatment Division. Data were reviewed for Dystar LP (NPDES permit #NCG060000) City of Reidsville (Industrial User Pretreatment Permit #0007) for March and November, 2011 and March, April, May and June, 2012. The following transcription error was detected: The contract laboratory is reporting Oil & Grease data as "ND" (which stands for non -detect), but lists the reporting limit as <5 mg/L. The laboratory is interpreting the ND as "0", thus Oil & Grease values are being reported on the IUP as "0" when they should be reporting these "ND" values as "<5 mg/L". V. CONCLUSIONS: Correcting the above -cited findings will help this la requirements. The inspector would like to thank the data review process. Please respond to all findings. Report prepared by: Jeffrey R. Adams Report reviewed by: Jason Smith b to produce quality data and meet certification staff for its assistance during the inspection and Date: August 29, 2012 Date: September 10, 2012