INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 5589
Laboratory Name: Catlin Engineers and Scientists
Inspection Type: Field Initial
Inspector Name(s): Todd Crawford
Inspection Date: July 24, 2012
Date Report Completed: August 10, 2012
Date Forwarded to Reviewer: August 10, 2012
Reviewed by: Jeff Adams
Date Review Completed: August 10, 2012
Cover Letter to use: X Insp. Initial Insp. Reg. Insp. No Finding Insp. CP ___Corrected
Unit Supervisor: Dana Satterwhite
Date Received: August 10, 2012
Date Forwarded to Linda: August 14, 2012
Date Mailed: August 14, 2012
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Catlin Engineers and Scientists
ADDRESS: P.O. Box 10279
W ilmington, NC 28405-0279
CERTIFICATE #: 5589
DATE OF INSPECTION: July 24, 2012
TYPE OF INSPECTION: Field Initial
AUDITOR(S): Todd Crawford
LOCAL PERSON(S) CONTACTED: Michael D. Mason and Dimitri Talbert
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 f or
the analysis of environmental samples.
II. GENERAL COMMENTS:
The facility has all the equipment necessary to perform the analyses. The system for traceability of
consumables was in place but needed more information. Proficiency Testing (PT) samples have been
analyzed for all certified parameters for the 2011 proficiency testing calendar year and the graded
results were 100% acceptable.
Please be reminded that acceptable PT results must be received in our office before September 30,
2012 in order to satisfy the 2012 PT requirements. Be sure to report the correct method reference as
well as your EPA Lab Code number and NC Certification number when submitting results.
Contracted analyses are performed by Environmental Chemists, Inc. (Certification #94).
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Documentation
Comment: Purchased standard and reagent traceability documentation did not indicate who the
vendor/manufacturer was. The Quality Assurance Policies for Field Laboratories document
states: All chemicals, reagents, standards and consumables used by the laboratory must have
the following information documented: Date Received, Date Opened (in use), Vendor, Lot
Number, and Expiration Date. A system (e.g., traceable identifiers) must be in place that links
standard/reagent preparation information to analytical batches in which the solutions are used.
Documentation of solution preparation must include the analyst’s initials, date of preparation,
the volume or weight of standard(s) used, the solvent and final volume of the solution. This
information as well as the vendor and/or manufacturer, lot number, and expiration date must
be retained for chemicals, reagents, standards and consumables used for a period of five
years. Consumable materials such as pH buffers and lots of pre -made standards are included
in this requirement. Notification of acceptable corrective action (i.e., an updated Calibration Solution
Receipt Log) was received by email on 08/03/12. No further response is necessary for this finding.
Page 2
#5589 Catlin Engineers and Scientists
A. Finding: The laboratory is not documenting Proficiency Testing (PT) sample analyses in the
same manner as environmental samples.
Requirement: All PT sample analyses must be recorded in the daily analysis records as for any
environmental sample. This serves as the permanent laboratory record. Ref: Proficiency Testing
Requirements, February 20, 2012, Revision 1.2.
Comment: The analysis of PT samples is designed to evaluate the entire process used to
routinely report environmental analytical results; therefore, PT samples must be analyzed and the
process documented in the same manner as environmental samples.
pH – Standard Methods, 21st Edition, 4500 H+ B
B. Finding: Values were reported that exceed the method specified accuracy (0.1 units) for this
analysis.
Requirement: By careful use of a laboratory pH meter with good electrodes, a precision of
±0.02 unit and an accuracy of ±0.05 unit can be achieved. However, ±0.1 pH unit represents
the limit of accuracy under normal conditions, especially for measurement of water and poorly
buffered solutions. For this reason, report pH values to the nearest 0.1 pH unit. Ref: Standard
Methods, 4500 H+ B-2000, (6).
Conductivity – EPA Method 120.1
Comment: To date, no samples have been analyzed for conductivity. Certification is being maintained
in anticipation of future work. There was no environmental sample data to review.
Comment: The benchsheet for this parameter did not contain a space to document the analysis of a
calibration check standard. The North Carolina W astewater/Groundwater Laboratory Certification
Approved Procedure for Field Analysis of Specific Conductance (Conductivity) states: Analyze and
document the value of a daily calibration standard. It is recommended that the calibration standard be in
the range of 1413 µmhos/cm. Rinse the conductivity cell with conductivity-free water between standards.
Analyze and document a calibration verification check standard. It is recommended that this standard be
at a lower range. The value obtained from the calibration check standard must read within 10% of the
true value of the check standard. If the obtained value is outside of the ±10% range, corrective action
must be taken. Notification of acceptable corrective action (i.e., an updated benchsheet which includes
space for the check standard known concentration and analytical result) was received by email on
08/02/12. No further response is necessary for this finding.
C. Finding: The Automatic Temperature Compensator (ATC) has not been verified.
Requirement: The ATC must be verified annually (i.e., every twelve months) and the process
documented. The ATC must be verified by analyzing a standard at 25 °C (the temperature that
conductivity values are compensated to) and a temperature(s) that brackets the temperature
ranges of the samples to be analyzed. This may require the analysis of a third temperature
reading that is > 25 °C. Ref: North Carolina W astewater/Groundwater Laboratory Certification
Approved Procedure for Field Analysis of Specific Conductance (Conductivity). Please submit
supporting documentation with your response.
Temperature – Standard Methods, 21st Edition, 2550 B
D. Finding: The temperature sensor check of the pH meter used to obtain reported temperatures
was not documented.
Page 3
#5589 Catlin Engineers and Scientists
Requirement: Thermometers and temperature measuring devices, used to measure
temperature for compliance monitoring, must be checked every 12 months against a NIST
traceable thermometer. The thermometer/meter readings must be less than or equal to 1ºC
from the NIST traceable thermometer reading. The documentation must include the serial
number of the NIST traceable thermometer that was used in the comparison. Document any
correction that applies on both the thermometer/meter and on a separate sheet to be filed. Ref:
North Carolina Wastewater/Groundwater Laboratory Certification Approved Procedure for Field
Analysis of Temperature. Please submit supporting documentation with your response.
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing field testing records and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for Port of Wilmington (Stormwater permit # NCS000174) and Port of Morehead City
(Stormwater permit # NCS000173). Data were reviewed for June, October and November, 2011, and
June, 2012. No transcription errors were detected. The facility appears to be doing a good job of
accurately transcribing data.
V. CONCLUSIONS:
Correcting the above-cited findings will help this lab to produce quality data and meet certification
requirements. The inspector would like to thank the staff for its assistance during the inspection and
data review process. Please respond to all findings.
Report prepared by: Todd Crawford Date: August 10, 2012
Report reviewed by: Jeff Adams Date: August 10, 2012