HomeMy WebLinkAbout#300 - 2012 Insp - Final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 300
Laboratory Name: Town of Beaufort
Inspection Type: Municipal Maintenance
Inspector Name(s): Todd Crawford
Inspection Date: March 13, 2012
Date Report Completed: April 11, 2012
Date Forwarded to Reviewer: April 11, 2102
Reviewed by: Chet Whiting
Date Review Completed: April 16, 2012
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding __ Insp. CP___ Corrected
Unit Supervisor: Dana Satterwhite
Date Received: April 18, 2012
Date Forwarded to Linda: April 23, 2012
Date Mailed: April 23, 2012
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Town of Beaufort
NPDES PERMIT # NC00021831
ADDRESS: 412 Hedrick St
Beaufort, NC 28516
CERTIFICATE #: 300
DATE OF INSPECTION: March 13, 2012
TYPE OF INSPECTION: Municipal Maintenance
AUDITOR(S): Todd Crawford
LOCAL PERSON(S) CONTACTED: Larry Third
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for the
analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was clean and well organized. The facility has all the equipment necessary to perform the
analyses.
The laboratory is reminded that any time changes are made to laboratory operations; the laboratory
must update the Quality Assurance (QA)/Standard Operating Procedures (SOP) document(s). Any
changes made in response to the Findings, Recommendations or Comments listed in this report must
be incorporated to insure the method is being performed as stated, references to methods are
accurate, and the QA and/or SOP document(s) is in agreement with approved practice and regulatory
requirements. In some instances, the laboratory may need to create a SOP to document how new
functions or policy will be implemented.
The requirements associated with Findings C, D, F, G and H are new policies that have been implemented
by our program since the last inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Quality Control
A. Finding: The facility has not updated established Standard Operating Procedures (SOPs) since
2000.
Requirement: Each laboratory shall develop and maintain a document outlining the analytical
quality control practices used for the parameters included in their certification. Supporting records
shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A
NCAC 2H .0805 (a) (7).
Recommendation: It is recommended that SOPs include a review date history, to show
maintenance, and a revision date history with a brief description of the change(s) made.
Page 2
#300 Town of Beaufort
B. Finding: The thermometer in the sample storage refrigerator has not been calibrated against a
NIST certified or traceable thermometer in the past 12 months.
Requirement: All thermometers and temperature measuring devices must be checked every
12 months against a NIST certified or NIST traceable thermometer and the process
documented. To check a thermometer or the temperature sensor of a meter, read the
temperature of the thermometer/meter against a NIST certified or NIST traceable thermometer
and record the two temperatures. The calibration must be performed at a temperature that
corresponds to the temperature used by the incubator, refrigerator, freezer, etc. In the case of
temperature measuring devices used to perform variable temperature readings the calibration
must be performed at a temperature range that approximates the range of the samples. The
thermometer/meter readings must be less than or equal to 1ºC from the NIST certified or NIST
traceable thermometer reading. The documentation must include the serial number of the NIST
certified thermometer or NIST traceable thermometer that was used in the comparison. Also
make any corrections to the data and document any correction that applies (even if zero) on
both the thermometer/meter and on a separate sheet to be filed.
• NIST traceable thermometers used for temperature measurement must be recalibrated
in accordance with the manufacturer’s recalibration date. If no recalibration date is
given, the NIST traceable thermometer must be recalibrated annually.
• NIST certified thermometers must be recalibrated, at a minimum, every five years. A
new certificate must be issued and maintained for inspection upon request.
Ref: North Carolina Wastewater/Groundwater Laboratory Certification Policy.
C. Finding: Auto-pipettors have not been calibrated.
Requirement: Mechanical volumetric liquid-dispensing devices (e.g., fixed and adjustable auto-
pipettors, bottle-top dispensers, etc.) must be calibrated at least twice per year, approximately
six months apart and documented. Each liquid-dispensing device must meet the manufacturer’s
statement of accuracy. Ref: North Carolina Wastewater/Groundwater Laboratory Certification
Policy.
Documentation
D. Finding: The laboratory needs to increase the documentation of purchased materials and
reagents, as well as documentation of standards and reagents prepared in the laboratory.
Requirement: All chemicals, reagents, standards and consumables used by the laboratory must
have the following information documented: Date received, Date Opened (in use), Vendor, Lot
Number, and Expiration Date (where specified). A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the
solutions are used. Documentation of solution preparation must include the analyst’s initials, date
of preparation, the volume or weight of standard(s) used, the solvent and final volume of the
solution. This information as well as the vendor and/or manufacturer, lot number, and expiration
date must be retained for primary standards, chemicals, reagents, and materials used for a period
of five years. Consumable materials such as pH buffers, lots of pre-made standards and/or media,
solids and bacteria filters, etc. are included in this requirement. Ref: North Carolina
Wastewater/Groundwater Laboratory Certification Policy.
Page 3
#300 Town of Beaufort
Comment: A standard/reagent prep log was in place; however, there was not an adequate link to
the specific analyses in which they were used.
Temperature - Standard Methods, 18th Edition, 2550 B
E. Finding: The temperature sensor on the Dissolved Oxygen meter used to obtain reported
temperature values has not been checked against a NIST thermometer.
Requirement: All thermometers, or temperature sensing devices on field meters used to
measure temperature for compliance monitoring must meet NIST specifications for accuracy or
must be calibrated against a NIST certified or NIST traceable thermometer annually (every 12
months) and proper corrections made and documented. A correction factor must be posted on
the meter even if that correction factor is 0ºC. Ref: North Carolina Wastewater/Groundwater
Laboratory Certification Policy.
Total Suspended Solids - Standard Methods, 18th Edition, 2540 D
Comment: Samples were not weighed to a constant weight, nor had an annual multiple weighing
study, to verify the adequacy of the drying time, been performed. North Carolina
Wastewater/Groundwater Laboratory Certification Policy states: Constant weights must be
documented. The approved methods require the following: “Repeat the cycle of drying, cooling,
desiccating, and weighing until a constant weight is obtained or until the weight change is less than 4%
of the previous weight or 0.5 mg, whichever is less.” In lieu of this, an annual study documenting the
time required to dry representative samples to a constant weight may be performed. Verify minimum
daily drying time is greater than or equal to the time used for the initial verification study drying cycle.
Drying cycles must be a minimum 1 hour for verification. Demonstration of acceptable corrective action
(i.e., documentation of an acceptable drying study) was received by email on 3/26/12. No further
response is necessary for this finding.
F. Finding: Filters are not weighed to constant weight prior to sample analysis, nor is a dry filter blank
analyzed with each set of samples.
Requirement: If pre-prepared filters are not used, the method requires that filters must be weighed
to a constant weight after washing. Repeat cycle of drying, cooling, desiccating, and weighing until
a constant weight is obtained or until weight change is less than 4% of the previous weighing or 0.5
mg, whichever is less. In lieu of this process, it is acceptable to analyze a single daily dry filter
blank to fulfill the method requirement of drying all filters to a constant weight prior to analysis. Ref:
North Carolina Wastewater/Groundwater Laboratory Certification Policy based upon Standard
Methods 20th Edition 2540 D. (3) (a).
G. Finding: The laboratory is not basing the reporting limit on the minimum weight gain required by
the method.
Requirement: The minimum weight gain allowed by any approved method is 2.5 mg. Choose
sample volume to yield between 2.5 and 200 mg dried residue. This establishes a minimum
reporting value of 2.5 mg/L when 1000 mL of sample is analyzed. If complete filtration takes
more than 10 minutes increase filter diameter or decrease sample volume. In instances where
the weight gain is less than the required 2.5 mg, the value must be reported as less than the
appropriate value based upon the volume used. Ref: North Carolina Wastewater/Groundwater
Laboratory Certification Policy based upon Standard Methods, 20th and 21st Editions, 2540 D.
(3) (b).
Page 4
#300 Town of Beaufort
Enterococci – Enterolert IDEXX
H. Finding: The Quanti-Tray® sealer is not checked monthly for leaks.
Requirement: If the Quanti-Tray® or Quanti-Tray®/2000 test is used, the sealer must be
checked monthly by adding a dye (e.g., bromcresol purple) to a water blank. If dye is observed
outside the wells, either perform maintenance or use another sealer. Ref: North Carolina
Wastewater/Groundwater Laboratory Certification Policy.
Coliform, Fecal – Standard Methods, 18th Edition, 9222 D (MF)
I. Finding: Consumable materials used for the Fecal Coliform MF method are not tested prior to use
to ensure that they are reliable.
Requirement: When a new lot of culture medium, pads, or membrane filters is to be used, a
comparison of the current lot in use (reference lot) against the new lot (test lot), be made. As a
minimum, make single analyses on five positive samples. Ref: Standard Methods, 18th Edition,
9020 B. (3) (d). Include all supporting documentation with your response to this report.
Recommendation: At this time the laboratory will not be able to conduct consumables testing
against materials that have been previously tested. In effect the baseline for the evaluation has
been lost. It is recommended that the laboratory obtain materials from a laboratory or facility that
have been tested in order to reestablish a baseline evaluation against material of proven quality.
Comment: Guidance for conducting the comparison was provided during the inspection. Since
there is no baseline for determining the acceptability of materials currently in use, acceptable
consumables from another lab must be used for the initial testing of materials currently in use.
Another option would be to use a contract laboratory with acceptable consumables to perform this
testing.
J. Finding: Forceps are not being flamed before handling each sample filter.
Requirement: Sterilize before use by dipping in 95% ethyl or absolute methyl alcohol and flaming.
Ref: Standard Methods, 18th Edition, 9222 B. (1) (i).
Residual Chlorine – Standard Methods, 18th Edition, 4500 Cl G
K. Finding: The HACH meter’s internal curve has not been verified in last 12 months.
Requirement: For colorimetric analyses, a series of five standards for a curve prepared annually
or three standards for curves established each day or standards as set forth in the analytical
procedure must be analyzed to establish a standard curve. Ref: 15A NCAC 02H .0805 (a) (7) (I).
Recommendation: It is recommended that the laboratory verify the internal calibration using
the concentrations: 10, 25, 50, 150 and 300 µg/L. This will verify the analytical range used to
measure Proficiency Testing (PT) samples as well as environmental samples.
L. Finding: A daily mid-range calibration check standard is not being analyzed.
Requirement: The curve must be updated as set forth in the standard procedures, each time the
slope changes by more than 10 percent at midrange, each time a new stock standard is prepared,
or at least every twelve months. Ref: 15A NCAC 02H .0805 (a) (7) (I).
Page 5
#300 Town of Beaufort
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing laboratory benchsheets and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for October, November and December, 2011. Data for metals were reviewed for January,
October, November and December, 2011. No transcription errors were detected, however, values with
a “less than” sign were reported in the “average” field on the DMR. When calculating an arithmetic
mean, you may consider a "less than" value as equal to zero. Therefore, if all monthly values are “less
than” values, the monthly arithmetic average would be “zero”. The documents, NC DWQ NPDES
Permitting Guidance for DMR Calculations and Directions for Completing Monthly Discharge Monitoring
Reports were provided during the inspection for additional guidance.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendations will help this lab to produce
quality data and meet certification requirements. The inspector would like to thank the staff for its
assistance during the inspection and data review process. Please respond to all findings.
Report prepared by: Todd Crawford Date: April 11, 2012
Report reviewed by: Chet Whiting Date: April 16, 2012