INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: #125
Laboratory Name: Tar River Regional WWTP Laboratory
Inspection Type: Municipal Maintenance
Inspector Name(s): Jeffrey R. Adams
Inspection Date: March 29, 2012
Date Report Completed: April 10, 2012
Date Forwarded to Reviewer: April 10, 2012
Reviewed by: Jason Smith
Date Review Completed: April 11, 2012
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding Insp. CP ___ Corrected
Unit Supervisor: Dana Satterwhite
Date Received: April 11, 2012
Date Forwarded to Linda Chavis April 20, 2012
Date Mailed: April 20, 2012
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Tar River Regional WWTP Laboratory
NPDES PERMIT #: NC0063096
ADDRESS: P.O. Box 1180
Rocky Mount, NC 27802-1180
CERTIFICATE #: 125
DATE OF INSPECTION: March 29, 2012
TYPE OF INSPECTION: Municipal Maintenance
AUDITOR(S): Jeffrey R. Adams
LOCAL PERSON(S) CONTACTED: Angela Boswell, David Gardner, Stacy Parks, Amanda James and
Andy Hicks
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for
the analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was clean and well organized. The laboratory had already implemented the matrix spiking
policy for its Ammonia, TKN, Nitrate-Nitrite and Total Phosphorous analyses, the new lower reporting
limits for all residue analyses, and the new nutrient addition requirement for BOD prior to the inspection.
The facility has a dedicated temperature-controlled BOD laboratory, which appears to contribute to the
consistency of quality control results for BOD testing (e.g., Glucose - Glutamic Acid standard recoveries).
The facility has all the equipment necessary to perform the analyses, however, some quality control
procedures need to be implemented.
The laboratory is reminded that any time changes are made to laboratory operations; the laboratory
must update the Quality Assurance (QA)/Standard Operating Procedures (SOP) document(s). Any
changes made in response to the Findings, Recommendations or Comments listed in this report must
be incorporated to insure the method is being performed as stated, references to methods are
accurate, and the QA and/or SOP document(s) is in agreement with approved practice and regulatory
requirements. In some instances, the laboratory may need to create a SOP to document how new
functions or policy will be implemented.
The laboratory was given a packet containing North Carolina Laboratory Certification quality control
requirements and policies during the inspection.
The requirements associated with Findings A, and B are new policies that have been implemented by our
program since the last inspection.
Page 2
#125 Tar River Regional WWTP Laboratory
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
General Laboratory
A. Finding: The laboratory needs to increase the documentation of materials and reagents used or
made in the laboratory.
Requirement: All chemicals, reagents, standards and consumables used by the laboratory must
have the following information documented: Date received, Date Opened (in use), Vendor, Lot
Number, and Expiration Date (where specified). A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the solutions
are used. Documentation of solution preparation must include the analyst’s initials, date of
preparation, the volume or weight of standard(s) used, the solvent and final volume of the solution .
This information as well as the vendor and/or manufacturer, lot number, and expiration date must be
retained for primary standards, chemicals, reagents, and materials used for a period of five years.
Consumable materials such as pH buffers, lots of pre-made standards and/or media, solids and
bacteria filters, etc. are included in this requirement. Ref: North Carolina Wastewater/Groundwater
Laboratory Certification Policy.
Chloride – Standard Methods, 18th Edition, 4500 Cl C
B. Finding: The laboratory is not analyzing matrix spikes for chloride analyses.
Requirement: Unless the referenced method states a greater frequency, spike 5% of samples
on a monthly basis. Laboratories analyzing less than 20 samples per month must analyze at
least one matrix spike (MS) each month samples are analyzed. Ref: North Carolina
Wastewater/Groundwater Laboratory Certification Policy.
Total Residual Chlorine – Standard Methods, 18th Edition, 4500 Cl G
C. Finding: The laboratory’s calibration curve verification does not bracket Proficiency Testing
(PT) sample concentrations.
Requirement: For analytical procedures requiring analysis of a series of standards, the
concentrations of these standards must bracket the concentration of the samples analyzed. One
of the standards must have a concentration equal to the laboratory's lower reporting concentration
for the parameter involved. For colorimetric analyses, a series of five standards for a curve
prepared annually or three standards for curves established each day or standards as set forth in
the analytical procedure must be analyzed to establish a standard curve. Ref : 15A NCAC 2H
.0805 (a) (7) (I). A copy of the new annual 5-point Total Residual Chlorine curve verification
documentation must be submitted with the response to this report.
Recommendation: The laboratory’s National Pollutant Discharge Elimination System (NPDES)
Permit limit is 28 µg/L for Total Residual Chlorine. It is recommended that the laboratory verify
the internal calibration using the concentrations: 20, 30, 50, 200 and 400 µg/L. This will verify
the analytical range used to measure Proficiency Testing (PT) samples as well as
environmental samples.
Recommendation: It is recommended that the values obtained not vary by more than 10% of the
known value for standard concentrations greater than or equal to 50 g/L and not vary by more than
25% of the known value for standard concentrations less than 50 g/L. The overall correlation
coefficient of the curve must be ≥0.995.
Page 3
#125 Tar River Regional WWTP Laboratory
IV. PAPER TRAIL INVESTIGATION:
The paper trail consisted of comparing laboratory benchsheets and contract lab reports to Discharge
Monitoring Reports (DMRs) submitted to the North Carolina Division of Water Quality. Data were
reviewed for the Tar River Regional WWTP Laboratory (NPDES permit #NC0030317) for November
and December, 2011 and January, 2012. No transcription errors were detected. The facility appears to
be doing a good job of accurately transcribing data.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendations will help this lab to
produce quality data and meet certification requirements. The inspector would like to thank the staff for
its assistance during the inspection and data review process. Please respond to all findings.
Report prepared by: Jeffrey R. Adams Date: April 10, 2012
Report reviewed by: Jason Smith Date: April 11, 2012