HomeMy WebLinkAbout#128 - 2012 Insp - Final
INSPECTION REPORT ROUTING SHEET
To be attached to all inspection reports in-house only.
Laboratory Cert. #: 128
Laboratory Name: Jacksonville Wilson Bay Laboratory
Inspection Type: Municipal Maintenance
Inspector Name(s): Todd Crawford
Inspection Date: March 6, 2012
Date Report Completed: March 20, 2012
Date Forwarded to Reviewer: March 22, 2102
Reviewed by: Jeff Adams
Date Review Completed: March 22, 2012
Cover Letter to use: Insp. Initial X Insp. Reg. Insp. No Finding __ Insp. CP___ Corrected
Unit Supervisor: Dana Satterwhite
Date Received: March 22, 2012
Date Forwarded to Linda: April 25, 2012
Date Mailed: April 25, 2012
_____________________________________________________________________
On-Site Inspection Report
LABORATORY NAME: Jacksonville Wilson Bay WWTP Laboratory
NPDES PERMIT # NC0088455, WQ0006735
ADDRESS: PO Box 128
Jacksonville, NC 28541
CERTIFICATE #: 128
DATE OF INSPECTION: March 6, 2012
TYPE OF INSPECTION: Municipal Maintenance
AUDITOR(S): Todd Crawford
LOCAL PERSON(S) CONTACTED: Jill Puff
I. INTRODUCTION:
This laboratory was inspected to verify its compliance with the requirements of 15A NCAC 2H .0800 for the
analysis of environmental samples.
II. GENERAL COMMENTS:
The laboratory was clean and well organized. The facility has all the equipment necessary to perform the
analyses. Benchsheets were easy to understand and records were well organized. A system of
traceability for consumables, prepared reagents and standards was in place, however, there was not a
clear link to individual analyses.
The laboratory is reminded that any time changes are made to laboratory operations; the laboratory
must update the Quality Assurance (QA)/Standard Operating Procedures (SOP) document(s). Any
changes made in response to the Findings, Recommendations or Comments listed in this report must
be incorporated to insure the method is being performed as stated, references to methods are
accurate, and the QA and/or SOP document(s) is in agreement with approved practice and regulatory
requirements. In some instances, the laboratory may need to create a SOP to document how new
functions or policy will be implemented.
The requirements associated with Findings C, D, F, H, L, M and O are new policies that have been
implemented by our program since the last inspection.
Ms. Jill Puff is also the supervisor of the laboratory at the Jacksonville Nano-Filtration Water Treatment
Plant. Analyses performed by staff at that facility are pH, Dissolved Oxygen (DO) and Conductivity. Data
and equipment at that facility were not included in this inspection.
III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS:
Quality Control
A. Finding: The National Institute of Standards and Technology (NIST) traceable thermometer used
to check all laboratory thermometers has not been calibrated in the past 12 months.
Page 2
#128 Jacksonville Wilson Bay WWTP Laboratory
Requirement: Once the NIST thermometer has been certified, repeat the calibration verification for
all thermometers in the lab. Include all supporting documentation with your response to this
report.
Requirement: All thermometers must meet NIST specifications for accuracy or checked, at a
minimum annually, against a NIST traceable thermometer and proper corrections made. Ref: 15A
NCAC 2H .0805 (a) (7) (O).
Requirement: All thermometers and temperature measuring devices must be checked every
12 months against a NIST certified or NIST traceable thermometer and the process
documented. To check a thermometer or the temperature sensor of a meter, read the
temperature of the thermometer/meter against a NIST certified or NIST traceable thermometer
and record the two temperatures. The calibration must be performed at a temperature that
corresponds to the temperature used by the incubator, refrigerator, freezer, etc. In the case of
temperature measuring devices used to perform variable temperature readings the calibration
must be performed at a temperature range that approximates the range of the samples. The
thermometer/meter readings must be less than or equal to 1ºC from the NIST certified or NIST
traceable thermometer reading. The documentation must include the serial number of the NIST
certified thermometer or NIST traceable thermometer that was used in the comparison. Also
make any corrections to the data and document any correction that applies (even if zero) on
both the thermometer/meter and on a separate sheet to be filed.
• NIST traceable thermometers used for temperature measurement must be recalibrated
in accordance with the manufacturer’s recalibration date. If no recalibration date is
given, the NIST traceable thermometer must be recalibrated annually.
• NIST certified thermometers must be recalibrated, at a minimum, every five years. A
new certificate must be issued and maintained for inspection upon request.
Ref: North Carolina Wastewater/Groundwater Laboratory Certification Policy.
B. Finding: Consumable materials used for the Fecal Coliform MF method are not tested prior to use
to ensure that they are reliable.
Requirement: When a new lot of culture medium, pads, or membrane filters is to be used, a
comparison of the current lot in use (reference lot) against the new lot (test lot), be made. As a
minimum, make single analyses on five positive samples. Ref: Standard Methods, 18th Edition,
9020 B. (3) (d). Include all supporting documentation with your response to this report.
Comment: Guidance for conducting the comparison was provided during the inspection. Since
there is no baseline for determining the acceptability of materials currently in use, acceptable
consumables from another lab must be used for the initial testing of materials currently in use.
Another option would be to use a contract laboratory with acceptable consumables to perform this
testing.
Documentation
C. Finding: The laboratory needs to increase the documentation that links purchased materials,
reagents and standards to individual analyses.
Requirement: All chemicals, reagents, standards and consumables used by the laboratory must
have the following information documented: Date received, Date Opened (in use), Vendor, Lot
Page 3
#128 Jacksonville Wilson Bay WWTP Laboratory
Number, and Expiration Date (where specified). A system (e.g., traceable identifiers) must be in
place that links standard/reagent preparation information to analytical batches in which the
solutions are used. Documentation of solution preparation must include the analyst’s initials, date
of preparation, the volume or weight of standard(s) used, the solvent and final volume of the
solution. This information as well as the vendor and/or manufacturer, lot number, and expiration
date must be retained for primary standards, chemicals, reagents, and materials used for a period
of five years. Consumable materials such as pH buffers, lots of pre-made standards and/or media,
solids and bacteria filters, etc. are included in this requirement. Ref: North Carolina
Wastewater/Groundwater Laboratory Certification Policy.
D. Finding: Data that does not meet all quality control requirements is not qualified on the
Discharge Monitoring Report (DMR).
Requirement: When quality control (QC) failures occur, the laboratory must attempt to
determine the source of the problem and must apply corrective action. If data qualifiers are used
to qualify samples not meeting QC requirements, the data may not be useable for the intended
purposes. A notation must be made on the Discharge Monitoring Report (DMR) form, in the
comment section or on a separate sheet attached to the DMR form, when any required sample
quality control does not meet specified criteria, and another sample cannot be obtained. Ref: North
Carolina Wastewater/Groundwater Laboratory Certification Policy based upon Standard
Methods, 20th Edition, 1020 B. (14).
Dissolved Oxygen - Standard Methods, 19th Edition, 4500 O G
E. Finding: The unit of measure is not documented on the calibration benchsheet.
Requirement: All laboratories must use printed laboratory bench worksheets that include a
space to enter the signature or initials of the analyst, date of analyses, sample identification,
volume of sample analyzed, value from the measurement system, factor and final value to be
reported and each item must be recorded each time samples are analyzed. Ref: 15A NCAC
2H .0805 (a) (7) (H).
Total Suspended Solids - Standard Methods, 19th Edition, 2540 D
F. Finding: Filters are not weighed to constant weight prior to sample analysis, nor is a dry filter blank
analyzed with each set of samples
Requirement: If pre-prepared filters are not used, the method requires that filters must be weighed
to a constant weight after washing. Repeat cycle of drying, cooling, desiccating, and weighing until
a constant weight is obtained or until weight change is less than 4% of the previous weighing or 0.5
mg, whichever is less. In lieu of this process, it is acceptable to analyze a single daily dry filter
blank to fulfill the method requirement of drying all filters to a constant weight prior to analysis. Ref:
North Carolina Wastewater/Groundwater Laboratory Certification Policy based upon Standard
Methods 20th Edition 2540 D. (3) (a).
Comment: The lab was analyzing a wet filter blank.
Biochemical Oxygen Demand - Standard Methods, 19th Edition, 5210 B
G. Finding: Samples that require a pH adjustment are not being seeded.
Requirement: Some samples do not contain a sufficient microbial population (for example, some
untreated industrial wastes, disinfected wastes, high-temperature wastes or wastes with extreme
Page 4
#128 Jacksonville Wilson Bay WWTP Laboratory
pH values). For such wastes, seed the dilution water by adding a population of microorganisms.
Ref: Standard Methods, 19th Edition, 5210 B. (4) (d).
Total Residual Chlorine - Standard Methods, 19th Edition, 4500 Cl G
H. Finding: A calibration verification standard (mid-range) is not analyzed prior to sample
analyses and at the end of the sample group.
Requirement: The calibration blank and calibration verification standard (mid-range) must be
analyzed initially (i.e., prior to sample analysis), after every tenth sample and at the end of each
sample group to check for carry over and calibration drift. If either fall outside established quality
control acceptance criteria, corrective action must be taken (e.g., repeating sample
determinations since the last acceptable calibration verification, repeating the initial calibration,
etc.). Ref: North Carolina Wastewater/Groundwater Laboratory Certification Policy based upon
Standard Methods, 20th Edition, 1020 B. (10) (c), 3020 B. (2) (b), and 4020 B. (2).
I. Finding: Duplicate analyses are not analyzed.
Requirement: Analyze five percent of all samples in duplicate to document precision. Laboratories
analyzing less than 20 samples per month must analyze at least one duplicate each month
samples are analyzed. Ref: 15A NCAC 2H .0805 (a) (7) (C).
Ammonia Nitrogen - Standard Methods, 18th Edition, 4500 NH3 F
Comment: The laboratory was not analyzing matrix spikes prior to receiving the checklist for the
inspection. North Carolina Wastewater/Groundwater Laboratory Certification Policy states: Unless the
referenced method states a greater frequency, spike 5% of samples on a monthly basis. Laboratories
analyzing less than 20 samples per month must analyze at least one matrix spike each month samples
are analyzed. Demonstration of acceptable corrective action (i.e., documentation of spikes performed
after receiving the checklist and prior to the inspection) was observed during the inspection. No further
response is necessary for this finding.
Comment: Prior to receiving the inspection checklist, the laboratory was not verifying that sample pHs
were being elevated above a pH of 11 Standard Units (S.U.) before reading results. Standard Methods,
18th Edition, 4500 NH3 F. (4) (b) states to add sufficient volume of 10 N NaOH solution to raise pH
above 11. Demonstration of acceptable corrective action (i.e., documentation of pH on the
benchsheets) was observed during the inspection. No further response is necessary for this finding.
J. Finding: The distillation apparatus is not being properly steamed out.
Requirement: Add 500 mL water and 20 mL borate buffer to a distillation flask and adjust pH
to 9.5 with 6N NaOH solution. Add a few glass beads or boiling chips and use this mixture to
steam out the distillation apparatus until distillate shows no traces of ammonia. Ref: Standard
Methods, 18th Edition, 4500 NH3 B. (4) (a).
Comment: The distillation apparatus was being steamed out with DI water only.
K. Finding: The laboratory is not analyzing a sufficient number of samples in the distillation study.
Requirement: The titration method requires all samples to be distilled. For all other methods,
manual distillation is not required if comparability data on representative effluent samples are on
file to show that this preliminary distillation step is not necessary; however, manual distillation will
be required to resolve any controversies. In the case of low concentration effluents, the samples
Page 5
#128 Jacksonville Wilson Bay WWTP Laboratory
must be spiked to allow for a meaningful statistical comparison. It is recommended that samples
with a concentration < 1.0 mg/L be spiked to yield a value in the 1 – 5 mg/L range. Comparisons
performed in this concentration range should check within 20% RPD. Additional samples must be
analyzed comparatively if the results do not indicate approximately the same values for distilled
and undistilled samples. It is recommended that both the distilled and undistilled portions of the
sample be analyzed by the same laboratory using the same methodology. The following
frequencies are required:
Municipal and Industrial Laboratories shall analyze initially four samples distilled and undistilled
from each discharge and; thereafter, analyze two samples a year from each effluent. Submit
analytical results for four samples, distilled and undistilled, from each discharge within 60
days of receipt of this report.
Ref: North Carolina Wastewater/Groundwater Laboratory Certification Policy based upon Code of
Federal Regulations, Title 40, Part 136; Federal Register Vol. 72, No. 47, March 12, 2007; Table IB
Footnote 6.
L. Finding: A calibration blank is not analyzed at the end of the sample group.
Requirement: The calibration blank and calibration verification standard (mid-range) must be
analyzed initially (i.e., prior to sample analysis), after every tenth sample and at the end of each
sample group to check for carry over and calibration drift. If either fall outside established quality
control acceptance criteria, corrective action must be taken (e.g., repeating sample
determinations since the last acceptable calibration verification, repeating the initial calibration,
etc.). Ref: North Carolina Wastewater/Groundwater Laboratory Certification Policy based upon
Standard Methods, 20th Edition, 1020 B. (10) (c), 3020 B. (2) (b), and 4020 B. (2).
Nitrite Nitrogen - Standard Methods, 18th Edition, 4500 NO2 B
M. Finding: The laboratory is not analyzing matrix spikes.
Requirement: Unless the referenced method states a greater frequency, spike 5% of samples
on a monthly basis. Laboratories analyzing less than 20 samples per month must analyze at
least one matrix spike each month samples are analyzed. Prepare the matrix spike from a
reference source different from that used for calibration unless otherwise stated in the method.
If matrix spike results are out of control, the results must be qualified or the laboratory must take
corrective action to rectify the effect, use another method, or employ the method of standard
additions. When the method of choice specifies matrix spike performance acceptance criteria for
accuracy, and the laboratory chooses to develop statistically valid, laboratory-specific limits, the
laboratory-generated limits cannot be less stringent than the criteria stated in the approved
method.
If the unspiked sample result is in the top 40% of the calibration range, the sample should be
diluted and the matrix spike prepared using the diluted sample. The recovery of the matrix spike
samples must be bracketed by the calibration range.
The volume of spike solution used in matrix spike preparation must in all cases be ≤ 10% of the
total matrix spike volume. It is preferable that the spike solution constitutes ≤ 1% of the total
matrix spike volume so that the matrix spike can be considered a whole volume sample with no
adjustment by calculation necessary. If the spike solution volume constitutes >1% of the total
sample volume, the sample concentration or spike concentration must be adjusted by
calculation.
Page 6
#128 Jacksonville Wilson Bay WWTP Laboratory
Ref: North Carolina Wastewater/Groundwater Laboratory Certification Policy.
Recommendation: It is recommended that the spike solution constitutes ≤1% of the total MS
volume.
Recommendation: If the duplicate sample value is near or below the reporting limit, it is
recommended that a Matrix Spike Duplicate (MSD) be analyzed. The MSD fulfills the duplicate
required by 15A NCAC 2H .0805 (a) (7) (C). This would result in better statistical results to
monitor method and analyst precision.
N. Finding: Sample pH is not being adjusted to between 5 and 9 S.U. as needed.
Requirement: If sample pH is not between 5 and 9, adjust to that range with 1N HCl or NH4OH
as required. Ref: Standard Methods, 18th Edition, 4500 NO2 B. (4) (b).
Comment: All routine sample data that was observed were within the proper pH range. Only
monitoring well samples, which are analyzed quarterly, were in need of pH adjustment.
Conductivity - Standard Methods, 19th Edition, 2510 B
O. Finding: The Automatic Temperature Compensator (ATC) has not been verified.
Requirement: The ATC must be verified annually (i.e., every twelve months) by analyzing a
conductivity standard at 25°C (the temperature that conductivity values are compensated to) and a
temperature(s) that bracket the temperature ranges of the samples to be analyzed. This may
require the analysis of a third temperature reading that is > 25°C. Ref: North Carolina
Wastewater/Groundwater Laboratory Certification Policy.
P. Finding: No acceptance criteria have been established for the calibration check standards.
Requirement: Each laboratory shall develop and maintain a document outlining the analytical
quality control practices used for the parameters included in their certification. Supporting
records shall be maintained as evidence that these practices are being effectively carried out.
Ref: 15A NCAC 2H .0805 (a) (7).
Comment: Acceptance criteria must be established in order to determine whether or not there are
quality control exceedances. Acceptance criteria for the check standards may be determined using
control charts or may be set using default limits such as a ± 10% recovery.
IV. PAPER TRAIL INVESTIGATION:
No paper trail performed.
V. CONCLUSIONS:
Correcting the above-cited findings and implementing the recommendations will help this lab to produce
quality data and meet certification requirements. The inspector would like to thank the staff for its
assistance during the inspection and data review process. Please respond to all findings.
Report prepared by: Todd Crawford Date: March 20, 2012
Report reviewed by: Jeff Adams Date: March 22, 2012