Loading...
HomeMy WebLinkAbout#10_08_2015_FINAL INSPECTION REPORT ROUTING SHEET To be attached to all inspection reports in-house only. Laboratory Cert. #: 10 Laboratory Name: Environment 1, Inc. Inspection Type: Commercial Maintenance Inspector Name(s): Nick Jones, Dana Satterwhite, Todd Crawford Tonja Springer, Jason Smith, Beth Swanson Inspection Date: August 11 and 12, 2015 Date Report Completed: September 30, 2015 Date Forwarded to Reviewer: October 12, 2015 Reviewed by: Jason Smith Date Review Completed: 10/29/2015 Cover Letter to use: Insp. Initial Insp. Reg Insp. No Finding Insp. CP Corrected Unit Supervisor/Chemist III: Gary Francies Date Received: 11/17/2015 Date Forwarded to Linda: 12/16/2015 Date Mailed: 12/16/2015 cc: Jay Sauber, Carrie Ruhlman, Chris Goforth On-Site Inspection Report LABORATORY NAME: Environment 1, Inc. ADDRESS: PO Box 7085 Greenville, NC 27835 CERTIFICATE #: 10 DATE OF INSPECTION: August 11 and 12, 2015 TYPE OF INSPECTION: Commercial Maintenance AUDITOR(S): Nick Jones, Dana Satterwhite, Todd Crawford Tonja Springer, Jason Smith, and Beth Swanson LOCAL PERSON(S) CONTACTED: Mark Oliveira and Steve Jones I. INTRODUCTION: This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A NCAC 2H .0800 for the analysis of environmental samples. II. GENERAL COMMENTS: The facility has all the equipment necessary to perform the analyses. Laboratory personnel communicate well with sample collectors and coordinate sample analyses effectively to manage workload and holding times. Proficiency Testing (PT) samples have been analyzed for all certified parameters for the 2015 proficiency testing calendar year. The laboratory has fulfilled its PT requirements for the 2015 calendar year. Contracted analyses are performed by Meritech, Inc. (Certification #165) and Eurofins Lancaster Laboratories Environmental, LLC (Certification #521). Current quality assurance policies and Field Parameter Approved Procedures for the analysis of the facility’s currently certified parameters were provided at the time of the inspection. Laboratory Fortified Matrix (LFM) and Laboratory Fortified Matrix Duplicate (LFMD) are also known a Matrix Spike (MS) and Matrix Spike Duplicate (MSD) and may be used interchangeably in this report. The laboratory is reminded that any time changes are made to laboratory operations; the laboratory must update the Quality Assurance (QA) / Standard Operating Procedures (SOP) documents. Any changes made in response to the Findings, Recommendations or Comments listed in this report must be incorporated to insure the Page 2 #10 Environment 1, Inc. Method is being performed as stated, references to Methods are accurate, and the QA and/or SOP documents are in agreement with approved practice and regulatory requirements. In some instances, the laboratory may need to create a SOP to document how new functions or policy will be implemented. III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS: General Laboratory Comment: The laboratory deleted the following parameter methods from their Certified Parameters Listing on November 3, 2015:  Chlorine, Total Residual – Standard Methods, 4500 Cl G-2000.  Phosphorus, Total – EPA Method, 365.1, Rev. 2.0 (1993).  Phosphate, Ortho – EPA Method, 365.3 (1978).  Purgeable, Aromatics & Halocarbons – SW -846 8021 B (Aqueous)  Chlorinated Acid Herbicides – Standard Methods, 6640 B-2006 (Aqueous) Comment: The sample preservation and storage temperature requirements may be changed in accordance with 40 CFR Part 136.3 Table II in all pertinent SOPs and in practice. The current practice is to store “at 4°C”. 40 CFR Part 136.3 Table II allows ≤6°C for most inorganic parameters and <10°C for bacterial tests, without evidence of freezing. Recommendation: It is recommended that the lab implement a temperature grid check of all block digesters by alternating the well location of the thermometer each time samples are digested. This will document heating uniformity and consistency of all sample wells in the block. A. Finding: The laboratory is reporting uncertified data without qualifying it as such. Requirement: Commercial laboratories are required to obtain certification for parameters which will be reported by the client to comply with State surface water monitoring, groundwater, and pretreatment Rules. Ref: 15A NCAC 2H .0804 (a). Comment: The lab was under the impression that being certified for the parameter method included both aqueous and non-aqueous matrices. In the Fall of 2012, the NC WW/GW LC program implemented certification by parameter method matrix. Effective August 11, 2015, the following Parameter Methods have been added to the labs certificate for Non-Aqueous matrix:  Cyanide – Standard Methods, 4500 CN- E-1999  Aluminum, Calcium, Chromium, Copper, Magnesium, Nickel, Potassium, Sodium and Zinc - EPA 200.7, Rev. 4.4 (1994)  Arsenic, Lead and Selenium – Standard Methods, 3113 B-2004 Data reported by the following Parameter Methods must be flagged as uncertified for Non-Aqueous matrix. Certification is not offered for these parameter methods in non-aqueous matrix because the methods only include procedures for preparation and analysis of aqueous samples. Page 3 #10 Environment 1, Inc.  Chloride – Standard Methods, 4500 Cl- B-1997  Fluoride – Standard Methods, 4500 F- C-1997  Phenols, Inorganic – EPA Method, 420.1 Rev. (1978)  Nitrogen, Ammonia – EPA 350.1, Rev. 2.0, 1993  Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0 (1993)  Nitrogen, Total Kjeldahl – EPA Method, 351.2, Rev. 2.0 (1993)  Total Phosphorus – EPA Method, 365.4 (1974)  Sulfate - Standard Methods, 4500 SO42 E-1997  Mercury – EPA Method, 245.1, Rev. 3.0 (1994) Documentation Recommendation: It is recommended that all records of maintenance performed on analytical instrumentation be documented in an individual equipment maintenance log book. All records of corrective action taken or minor troubleshooting and repair performed should also be documented in the equipment maintenance log. The records may include the following, as appropriate:  The identity of the item of equipment and its software;  The manufacturer’s name, type identification, and serial number or other unique identifier;  The current location;  The dates, results and copies of reports and certificates or all calibrations, adjustments, acceptance criteria, and the due date of the next calibration.  The maintenance plan, where appropriate, and maintenance completed date.  Documentation of any damage, malfunction, modification, or repair to the equipment.  Signature or initials of the person performing the maintenance or troubleshooting. These records provide an historical reconstruction of the maintenance performed, a reference for effective troubleshooting in the future, and can also serve as a valuable training tool for new analysts. Recommendation: It is recommended that labels be added to boxes on the benchsheets to designate their content. Some benchsheets have boxes which are not labeled but are used for data entry codes. Recommendation: It is recommended to include calculation formulas on benchsheets. Recommendation: It is recommended to include Quality Control acceptance criteria on benchsheets. Comment: There were several instances on benchsheets (i.e., benchsheets for Chemical Oxygen Demand, Inorganic Phenols and Sulfate) where the analyst’s handwriting was difficult to distinguish. This could pose issues with data defensibility or transcription errors. B. Finding: The laboratory needs to increase the documentation of standards and reagents prepared in the laboratory. Page 4 #10 Environment 1, Inc. Requirement: All chemicals, reagents, standards and consumables used by the laboratory must have the following information documented: Date received, Date Opened (in use), Vendor, Lot Number, and Expiration Date (where specified). A system (e.g., traceable identifiers) must be in place that links standard/reagent preparation information to analytical batches in which the solutions are used. Documentation of solution preparation must include the analyst’s initials, date of preparation, the volume or weight of standard(s) used, the solvent and final volume of the solution. This information as well as the vendor and/or manufacturer, lot number, and expiration date must be retained for primary standards, chemicals, reagents, and materials used for a period of five years. Ref: NC WW/GW LC Policy. C. Finding: The analyst is not signing/initialing or dating electronically generated laboratory benchsheets or data printouts. This is considered pertinent information Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (A). Standard Operating Procedures Recommendation: It is recommended that flow diagrams be included where applicable to improve SOP utilization. Recommendation: It is recommended that the SOPs for Nitrogen, Ammonia by EPA Method, 350.1, Rev. 2.0 (1993), Phosphorus, Total by EPA Method, 365.4 (1974) and Nitrogen, Total Kjeldahl by EPA Method, 351.2, Rev. 2.0 (1993) be revised to indicate that the analyst actually recalibrates when curve standards fail QC criteria. It currently states to only use curve points that are +10%. Comment: The Total Coliform, MF SOP states that the pH of M-Endo Liquid Media must be between 7.1 and 7.3 S.U. The reference method, Standard Methods, 9222 B-1997, allows 7.2 + 0.2 S.U. Comment: The Total Coliform, MF and MPN SOPs require preservation temperature of <4°C and hold time requirements of 6 and 8 hours, respectively. 40 CFR Part 136, Table II allows <10°C and 8 hours. Comment: The laboratory did not have a Standard Operating Procedure (SOP) for Thallium by EPA Method, 279.2 (Aqueous). The North Carolina Administrative Code, 15 NCAC 2H .0805 (a) (7) states: Each laboratory shall develop and maintain a document outlining the analytical QC practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Acceptable corrective action [i.e., a SOP for Thallium by EPA Method, 279.2 (Aqueous)] was received via email on 11/28/2015. No further response is necessary for this Finding. Comment: An inconsistency and/or omission was noted between the Standard Operating Procedure (SOP) and laboratory practice as follows: personnel were not Page 5 #10 Environment 1, Inc. following procedures as stated in the Laboratory’s SOP, the Method reference in the SOP was not correct, the SOP was in direct conflict with Method and/or regulatory requirements, and the SOP did not describe in detail how the Method is performed (observations are listed below ). The North Carolina Administrative Code, 15 NCAC 2H .0805 (a) (7) states: Each laboratory shall develop and maintain a document outlining the analytical QC practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Demonstration of acceptable corrective action (i.e., SOPs were revised) was observed during the second day of inspection (i.e., August 12, 2015). No further response is necessary for this Finding.  Biochemical Oxygen Demand -SM 5210 B-2001 o The SOP did not state that the incubation temperature is 20+1°C.  Phosphate, Ortho – Standard Methods, 4500 P E-1999 o The SOP did not describe the process of back–calculating each calibration standard. D. Finding: The laboratory does not have an Analytical QC Document or SOP for the following:  Pesticides, Organochlorine - EPA Method 608 (Aqueous)  Purgeable, Organics - EPA Method 624 (Aqueous)  Metals - EPA SW -846 Method 6020 B(Aqueous)  Mercury - EPA SW -846 Method 7470 A (Aqueous)  Mercury - EPA SW -846 Method 7471 B (Aqueous)  Purgeable, Aromatics & Halocarbons - EPA SW -846 Method 8021 B (Aqueous)  Proficiency Testing Samples Requirement: Each laboratory shall develop and maintain a document outlining the analytical QC practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A NCAC 2H .0805 (a) (7). Requirement: Laboratories must have a documented plan (this is usually detailed in the laboratory’s Quality Assurance Manual) of how they intend to cover the applicable program requirements for proficiency testing per their scope of accreditation. This plan shall cover any commercially available proficiency testing and any inter-laboratory organized studies, as applicable. The laboratory must also be able to explain when proficiency testing is not possible for certain parameters and provide a description of what the laboratory is doing in lieu of proficiency testing. This shall be detailed in the plan. The plan must also address the laboratory’s process for submission of proficiency testing results and related corrective action responses. Laboratory Standard Operating Procedures (SOPs) must address how low level samples will be analyzed, including concentration of the sample or adjustment of the normality of a titrant. These instructions shall be followed when the concentration of a PT sample falls below the range of their routine analytical Method. Instructions shall also be included in the laboratory’s SOP for how high level samples will be analyzed, including preparation of multiple Page 6 #10 Environment 1, Inc. dilutions of the sample. These instructions will be followed when the concentration of a PT falls above the range of their routine analytical Method. Ref: Proficiency Testing Requirements, February 20, 2012, Revision 1.2. E. Finding: An inconsistency and/or omission was noted between the Standard Operating Procedure (SOP) and laboratory practice as follows:  General o The SOPs do not describe in sufficient detail QC acceptance criteria and/or corrective actions for unacceptable QC measures for all QC elements. This was noted for the following parameters; however, this list may not be all inclusive: Acidity, Nitrogen (Ammonia, Nitrite, Nitrate and Nitrate + Nitrite), Total Phosphorus, Inorganic Phenols, Salinity, Organochlorine Pesticides, 1,2-Dibromoethane, Alkalinity, Hardness, Mercury, Purgeable Organics, Residues (Dissolved, Suspended and Total), Metals, Chloride, Chlorinated Acid Herbicides, Purgeables (Halocarbons & Aromatics), Fecal Coliform MF, Total MPN &Total MPN Sludge, Cyanide, and Fluoride. o SOPs are intended to describe procedures exactly as they are to be performed. Evaluate all SOPS for the proper use of the word “should”.  Acidity – Standard Methods, 2310 B-1997 o The SOP does not state preservation / hold time requirements. o . o The SOP does not state that the sample is brought to room temperature before analysis. o The SOP does not state that the titrant is added in increments of 0.5 mL or less, such that a change of less than 0.2 S.U. occur per increment.  Alkalinity – Standard Methods, 2320 B-1997 o The SOP does not define the reporting limit.  Biochemical Oxygen Demand – Standard Methods, 5210 B-2001 (Hach 10360-2011, Rev. 1.2) (LDO) o The SOP does not state that samples and dilution water are warmed to 20+3°C (it states “approximately 20 °C”). o The SOP does not state that the pH adjustment must not change the sample volume by more than 5%. o The SOP does not state the initial Dissolved Oxygen (DO) of the blank must be 7.5-9.0 mg/L. o The SOP does not state that the initial DO must be measured within 30 minutes of preparation. o The SOP does not state how sample results are calculated and which results are reported. o The SOP does not state details of calculating and applying the seed correction factor. o The SOP does not define the reporting limit (i.e., 2 mg/L). Page 7 #10 Environment 1, Inc.  Chemical Oxygen Demand – Hach Method 8000 o The SOP does not state preservation / hold time requirements. o The SOP does not state to wait 20 min for the vials to cool to 120°C or less before inverting. o The SOP does not state to invert the vials several times while still warm. o The SOP does not state the reporting limit for the high range curve. o The SOP does not include calculations for QC.  Chloride - Standard Methods, 4500 Cl- B-1997 o The SOP section for preparation of the Aluminum Hydroxide Solution states: Follow procedure in Standard Methods on page 4-67. The correct reference is 4500 Cl- B-1997, (3) (d) (1). o . o The SOP does not state how the LFM/LFMD is prepared.  Chlorophyll - EPA Method 445.0, Rev. 1.2 (Fluorometric) o The SOP does not include the revision in the Method reference. o The SOP does not include the units of the vacuum pressure in section 3.b.1. o The SOP does not include the spectrophotometer calibration sheet. o The SOP contains a section (i.e., section 1.h) which appears to be irrelevant to the analysis.  Coliform, Fecal - Standard Methods, 9222 D-1997 (MF) o The SOP does not state a range for reagent storage temperature.  Coliform, Total - Standard Methods, 9222 B-1997 (MF) o The SOP states that the lab is to set two dilutions. Lab practice is to set three dilutions. o The SOP does not state the volume of sample to be analyzed. o The SOP states that it is attached, but does not include the attachment regarding the preparation of phosphate dilution water. o The SOP does not state the specifications of the light source utilized for counting colonies.  Coliform, Fecal - Standard Methods, 9222 D-1997 (MF)  Coliform, Fecal - Standard Methods, 9221 C E-2006 (MPN) o The SOPs do not state to document the water bath temperature twice daily. o The SOPs do not describe the known culture positive.  Coliform, Total - Standard Methods, 9221 B-2006 (MPN) o The SOP states to shake the samples and observe for gas formation. The method says to swirl the sample gently after incubation. o The SOP does not state the size of the sterile loop or how much sample to transfer to the confirmation phase. o The SOP does not state the pH of the sterilized broth must be 6.8 + 0.2 S.U. o The SOP does not state what is used to adjust the pH of the brilliant green lactose bile broth. Page 8 #10 Environment 1, Inc. o The SOP does not state instructions for determining a reportable result. o The MPN table is not included with the SOP.  Color, PC – Standard Methods, 2120 B-2001 o The SOP does not state to document the pH adjustment. o The SOP does not state that the pH of samples must be reported as well.  Conductance, Specific – Standard Methods, 2510 B-1997 o The lab currently has three SOPs. Since this is a Field Parameter, not all of the analyses are subject to the same QC requirements. Ensure that QC is performed as required per individual SOP. (a) Lab and Field SOP (i) The SOP states to adjust meter to the proper scale, but does not state how to determine which scale is proper. (ii) Duplicates are not required, but if the lab continues to analyze them, then the SOP needs to include acceptance criteria and corrective actions for failures. (iii) The SOP does not state that the Automatic Temperature Compensator (ATC) check must be performed every twelve months. (b) Lab SOP (i) The SOP does not state the proper units for the 1412 µmhos/cm standard (i.e., was only listed as µmhos). (ii) The SOP does not state preservation and hold time requirements (i.e., cool to ≤6°C and 28 days). (iii) The SOP does not describe in sufficient detail how the standards are prepared. (iv) The SOP does not state the frequency, acceptance criteria or corrective actions regarding Laboratory Fortified Blanks/Second Source Standards. (c) Field SOP (i) The SOP does not state QC requirements for second source laboratory fortified blank, associated acceptance criteria and corrective actions for failures. (d) River Basin Coalition SOP (i) The SOP does not state directions for ATC verification.  Cyanide – Standard Methods, 4500 CN- E-1999 o The SOP does not state that the calibration standards are back-calculated against the curve and how acceptance criteria are applied. o The SOP does not state the frequency that IDOCs or MDLs are performed. o The SOP does not state how the titrant is standardized. o The SOP does not state that the blanks (i.e., calibration and method) and calibration standards are analyzed at the beginning of the analytical sequence.  Dissolved, Oxygen – Standard Methods, 4500 O G-2001 o The SOP does not state sample hold time (i.e., 15 minutes). Page 9 #10 Environment 1, Inc. o The SOP does not describe a detailed operating procedure. It refers the analyst to the manufacturer’s instructions for guidance, but instead the lab has a document titled “YSI 556 MPS Calibration Instructions” which needs to be the referenced document. o The SOP does not state the amount of time needed for the instrument to warm up. o The SOP does not state that a post analysis calibration verification is required. o The SOP does not state acceptance limits (i.e., ±0.5 mg/L) for the post analysis calibration verification. o The SOP does not state the analysis of a mid-day check and acceptance criteria.  Enterococci – Enterolert® IDEXX o The SOP does not include the IDEXX MPN table. o The SOP does not include instructions for opening the Quanti-Tray® without contaminating the interior of the foil or tray. o The SOP does not state the source of the culture positive control. o The SOP does not state the source of the culture sterility control. o The SOP does not state to allow any foam to settle before sealing.  Fluoride - Standard Methods, 4500 F- C-1997 o The SOP does not specify how sample volumes are measured. o The SOP states to wipe the electrode with a paper towel. The Method states to blot dry and cautions that this may poison the electrode if not done gently. o The SOP does not state to verify the calibration after every ten samples and at the end of an analysis run. o The SOP does not state to bring all of the samples and standards to the same temperature prior to buffer addition. o The SOP instructions for preparing working standards produce concentrations that are half of what is stated. o The SOP does not include the distillation requirement or reference the distillation SOP. o The SOP does not include QC calculations. o The laboratory is combining the standard preparation step with one of the other procedural steps causing confusion regarding the concentration of the standards. The method says add a volume of buffer equal to the volume of standard or sample being analyzed prior to analysis. Each working standard final volume is 100ml, however the stated concentration of the standard is based on a final volume of 50 ml. Since only 50 ml of sample is used for analysis and the adding of 50 ml of buffer to the sample gives the sample volume proportion as that of the standard (i.e. 50 ml of standard plus 50 ml of buffer), the final sample results, based on the stated standard concentrations, is correct. However, the description of the standards preparation, by itself, is misleading in that it appears to incorrectly state the final concentrations when based on the stated final volume of 100 ml.  Hardness – Standard Methods, 2340 C-1997 Page 10 #10 Environment 1, Inc. o The SOP does not describe in sufficient detail how the Matrix Spike (MS) is prepared.  Nitrogen, Ammonia – EPA Method, 350.1, Rev. 2.0, 1993 o The SOP does not state EPA 350.1, Rev. 2.0, 1993 as the distillation Method. It states SM 4500 NH3 B-1997.  Nitrogen, NO2 – EPA Method, 353.2, Rev. 2.0, 1993 o The SOP does not state that NO2 is determined by bypassing the Cd column. o The SOP does not describe how to adjust samples with pH values outside of 5-9 S.U. prior to analysis. o The SOP does not state that matrix spikes must be analyzed at 10% frequency with +10% recovery, as required by the method. The SOP states matrix spikes are to be analyzed at a 5% frequency with +15% recovery or 95% confidence interval.  Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0, 1993 o The SOP does not describe the preparation of all reagents used during analysis (e.g., Na2S2O3, 2% CuSO4, 50% H2SO4, 0.5 N HCl).  Nitrogen, NO2 – EPA Method, 353.2, Rev. 2.0, 1993  Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0, 1993 o The SOP contains typos in the color reagent section (i.e., the spellings of naphthylethylenediamine and sulfanilamide).  Nitrogen, Total Kjeldahl – EPA Method, 351.2, Rev. 2.0, 1993 o The SOP does not state the true value of the second source standard. o The SOP does not include an equipment / supplies section. o The SOP does not address interferences.  Phosphorus, Ortho – Standard Methods, 4500 P E-1999 o The SOP does not state that MS/MSDs are analyzed daily. o The SOP does not describe the analysis of MSDs.  Phosphorus, Total – EPA Method, 365.4, 1974 o The SOP references a version of the Method which is not approved (i.e., revision 2.0, 1993). o The SOP does not address all required QC elements (e.g., IDC MDL, MS/MSD, trend analysis). Refer to 40 CFR 136.7 for all 12 QC elements.  Phosphate, Ortho – Standard Methods, 4500 P E-1999 o The SOP does not describe actions taken if turbidity forms when making the combined reagent.  Nitrogen, Ammonia – EPA Method, 350.1, Rev. 2.0, 1993  Nitrogen, NO2 – EPA Method, 353.2, Rev. 2.0, 1993  Nitrogen, NO3 - (NO3 + NO2 EPA 353.2, Rev. 2.0, 1993) - (NO2 EPA 353.2, Rev. 2.0, 1993) Page 11 #10 Environment 1, Inc.  Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0, 1993  Phosphorus, Total – EPA Method, 365.4, 1974 o The SOPs do not state the lab name. o The SOPs do not state interferences. o The SOPs do not state utilized equipment. o The SOPs do not state the source of the reagent water. o The SOPs do not document all calculations (e.g., results, percent recovery, software regression and relative percent difference). o The SOPs do not state reporting requirements (e.g., significant figures, units, rounding, etc.). o The SOPs do not state calibration curve acceptance criteria. o The SOPs do not state preservation and hold time requirements. o The SOPs do not describe the chemical preservation of samples. o The SOPs do not state that blanks must be analyzed initially, after every 10 samples and at the end of an analysis batch. o The SOPs do not describe matrix spike preparation. o The SOPs do not describe glassware cleaning protocols. o The SOPS do not describe what is done for over range samples.  Nitrogen, Ammonia – EPA Method, 350.1, Rev. 2.0, 1993  Nitrogen, NO2 – EPA Method, 353.2, Rev. 2.0, 1993  Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0, 1993  Nitrogen, NO3 - (NO3 + NO2 EPA 353.2, Rev. 2.0, 1993) - (NO2 EPA 353.2, Rev. 2.0, 1993)  Nitrogen, Total Kjeldahl – EPA Method, 351.2, Rev. 2.0, 1993  Phosphorus, Total – EPA Method, 365.4, 1974 o The SOPs do not state that the 2nd source standard acceptance criterion must be ±10%. o The SOPs do not state that the calibration standards be analyzed in order of decreasing concentration.  Organic Carbon, Dissolved – Standard Methods, 5310 C-2000  Organic Carbon, Total – Standard Methods, 5310 C-2000 o The SOPs do not state the true value of the second source standard. o The SOPs do not include the calculations for QC.  Phenols, Inorganic – EPA Method, 420.1, Rev. 1978 o The SOP does not state that when the distillate is turbid, filter through a prewashed membrane filter. o In step five of the SOP, it does not state that the 50 mLs of deionized water added be warm. o In step twelve of the SOP, it does not state that after the additional ten shakes to let chloroform settle again, then vent chloroform fumes into hood.  Residue, Suspended – Standard Methods, 2540 D-1997 o The SOP does not state the total maximum filtration time allowed for samples (i.e., 10 minutes). o The SOP does not state that the drying time utilized daily is the drying time established from the annual study. Page 12 #10 Environment 1, Inc. o The SOP does not state QC acceptance criterion and corrective actions for the dry blank.  Residue, Dissolved – Standard Methods, 2540 C-1997  Residue, Suspended – Standard Methods, 2540 D-1997  Residue, Total – Standard Methods, 2540 B-1997 o The SOPs do not describe the annual drying study. o The SOPs do not include all necessary equipment to perform the analysis. o The SOPs do not state that weight 1 and weight 2 are in grams in the utilized calculation. o The SOP does not state how different sample volumes are measured (i.e., wide bore pipettes vs. graduated cylinders).  Residue, Suspended – Standard Methods, 2540 D-1997  Residue, Total – Standard Methods, 2540 B-1997 o The SOP does not state explicit QC acceptance criteria for duplicate analyses. The total residue SOP stated the duplicates “should be ±5%” and there was no criterion in the suspended residue SOP.  Residue, Dissolved – Standard Methods, 2540 C-1997  Residue, Total – Standard Methods, 2540 B-1997 o The SOP does not state that the given reporting limits are only applicable when using sample volumes of 250 mL for Dissolved Solids and 100 mL for Total Solids. o The SOP does not describe how to adjust the PQL if different volumes of sample are used and the minimum weight gain is not achieved.  Residue, Dissolved – Standard Methods, 2540 C-1997 o The SOP does not describe how the filters are washed. o The SOP does not state that filters must be wrinkled side up on the filtration apparatus. o Step 9 states to repeat steps 7 and 8. The lab is performing a drying time study instead of repeating steps 7 and 8.  Residue, Settleable – Standard Methods, 2540 F-1997 o The SOP does not state to clean and dry the Imhoff cone. o The SOP does not include a section for reporting results (i.e., units and reporting limit).  Salinity – Standard Methods, 2520 B-2010 o The SOP does not state the calculation that is utilized in the Microsoft Excel® calculation spreadsheet to determine results o The SOP does not describe the preparation and concentration of calibration standards.  Sulfate – Standard Methods, 4500 SO42- E-1997 o The SOP does not state the requirement to nor the instructions for back- calculating the calibration standards.  Sulfide – Standard Methods, 4500 S2- D-2000 Page 13 #10 Environment 1, Inc. o The SOP states to document “required sample information”, but does not specifically state what that includes. o At the end of step two, the SOP does not state that it is inverted slowly and only once. o The SOP does not state to standardize the methylene blue solution each day it is used. o The SOP does not state to analyze a MS/MSD daily.  Temperature – Standard Methods, 2550 B-2000 o The SOP does not state that the analysis must be in situ or immediate.  Mercury – EPA Method, 1631 E o The SOP does not state that the samples must be preserved within 48 hours (SOP states within 72 hours). o The SOP states that 20 mL of standard is added to 20 mL of sample. In practice, the laboratory adds 2 mL of standard to 38 mL of sample.  Metals – EPA Method, 200.7, Rev. 4.4,1994 o The SOPs (i.e., digestion and analysis) do not describe preparation and concentrations of all standards and QC. o The SOP does not define all standard nomenclature (i.e., CCB, ICB, LRB, CCV, etc.)  Base Neutral / Acid Organics – SW846 Method, 8270 D o The SOP skips sections three and four. o The SOP states in section 2, subsections 1, 2 and 3 to continue to step 3C. There is no step 3C.  1,2-Dibromoethane – EPA Method, 504.1 o The SOP does not include the second source confirmation. o The SOP does not include the DBCM retention time standard.  Pesticides, Organochlorine – SW846 Method 8081 B o The SOP does not cite the complete method reference, only 8081 B. o The SOP does not cite the extraction method. o The SOP does not state that samples with pH values outside of 5-9 S.U. will be adjusted and how.  Pesticides, Organochlorine – SW846 Method, 8081 B  1,2-Dibromoethane – EPA Method, 504.1 o The SOPs do not state the lab name. o The SOPs do not state interferences and their mitigations. o The SOPs do not state retention time windows. o The SOPs do not state confirmation techniques. o The SOPs do not state reagent preparation. o The SOPs do not state utilized equipment (including columns). o The SOPs do not state calculations (e.g., software regression, results, percent recovery and relative percent difference). o The SOPs do not state reporting requirements (e.g., significant figures, units, rounding, etc.). Page 14 #10 Environment 1, Inc. o The SOPs do not state calibration curve acceptance criteria. o The SOPs do not define all QC frequencies. o The SOPs do not state preservation and hold time requirements.  Purgeable, Aromatics – Standard Methods, 6200 C-1997 o The SOP does not state sample preservation requirements.  Purgeable, Halocarbons – EPA Method, 601  Purgeable, Aromatics – EPA Method, 602 o The SOPs do not include the complete method reference. It only references 601 and 602. o The SOPs do not state sample preservation requirements. o The SOPs do not state that the prepared aqueous calibration standards may be stored for up to 24 hours, if held in sealed vials with zero headspace. If not stored, they must be discarded after one hour. o The SOPs do not state the frequency of MS/MSDs. Requirement: Each laboratory shall develop and maintain a document outlining the analytical QC practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A NCAC 2H .0805 (a) (7). Please respond in writing stating that the SOPs will be revised to address the items in this Finding within one year. Comment: The inspectors left an example SOP review checklist for the lab’s internal use during the audit. Acidity – Standard Methods, 2310 B-1997 F. Finding: The lab has not established a reporting limit. Requirement: Calculation Acidity, as mg CaCo3/L = [(A x B) – (C x D)] x 50 000 mL sample where: A = mL NaOH titrant used, B = normality of NaOH, C = mL H2SO4 used (2310B.4a), and D = normality of H2SO4 Ref: Standard Methods, 2310 B-1997. (5). Comment: To determine the reporting limit with the above equation, “A” equals the smallest measurement increment of the buret utilized. G. Finding: The lab is not documenting the endpoint pH. This is considered pertinent information. Page 15 #10 Environment 1, Inc. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Requirement: Consequently, use fixed end points of pH 3.7 and pH 8.3 for standard acidity determinations via potentiometric titration in wastewaters and natural waters where the simple carbonate equilibria discussed above cannot be assumed. Ref: Standard Methods, 2310 B-1997. (1) (b). Alkalinity – Standard Methods, 2320 B-1997 H. Finding: The lab is not standardizing the titrant monthly. Requirement: Verify standardized titration reagents by periodically re- standardizing. Method parameters in Part 2000 that are determined using standardized titration reagents are acidity, alkalinity, and hardness. Typically, the standardized reagents are stable for several months when sealed to avoid evaporation and stored properly. Re-standardize reagents once a month or when improper storage occurs. Ref: Standard Methods, 2020 B-2010. (2) (b). I. Finding: The lab is not reporting results as “The alkalinity to pH ____=____ mg CaCO3/L.” Requirement: Report pH of end point used as follows: “The alkalinity to pH____=____ mg CaCO3/L” and indicate clearly if this pH corresponds to an inflection point of the titration curve. Ref: Standard Methods, 2320 B-1997. (5) (a). Biochemical Oxygen Demand – Standard Methods, 5210 B-2001 Recommendation: It is recommended to not change the sample volume by more than 5% when neutralizing Total Residual Chlorine with sodium sulfite. J. Finding: The laboratory was not seeding samples over six hours old. Requirement: Some samples (for example, some untreated industrial wastes, disinfected wastes, high-temperature wastes, wastes having pH values less than 6 or greater than 8, or wastes stored more than 6 h after collection) do not contain a sufficient microbial population. Seed such samples by adding a population of suitable microorganisms. Ref: Standard Methods 5210 B-2001. (4) (d). K. Finding: Samples were not properly adjusted for pH. Requirement: Check pH; if it is not between 6.0 and 8.0, adjust sample temperature to 20 ± 3 °C, then adjust pH to 7.0 to 7.2 using a solution of sulfuric acid or sodium hydroxide of such strength that the quantity of reagent does not dilute the sample by more than 0.5%. The pH of dilution water should not be affected by the lowest sample dilution. Always seed samples that have been pH adjusted. Ref: Standard Methods, 5210 B-2001. (4) (b) (1). Page 16 #10 Environment 1, Inc. Comment: The laboratory adjusts all samples to a neutral pH if the sample pH is less than 6.0 or greater than 8.5 units. Samples are adjusted to pH 6.5 to 7.5 units if needed. L. Finding: Incubation time is not always documented. This is considered pertinent information. Requirement: After 5 d + 6 h of incubation, determine DO in all sample dilutions, and in all blanks and checks as in ¶ 6b–d, using the azide modification of the titrimetric method or the membrane electrode method. Ref: Standard Methods, 5210 B-2001. (5) (i). Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Comment: All incubation start times are not documented on the Biochemical Oxygen Demand (BOD) benchsheets. Total incubation time cannot be verified without start and end times. Samples are occasionally set up in both the morning and afternoon. When they are, the time samples are incubated is documented in the morning, but is not documented for the samples set up in the afternoon. M. Finding: The units of measure for Dissolved Oxygen are not documented. Requirement: All laboratories must use printed laboratory bench worksheets that include a space to enter the signature or initials of the analyst, date of analyses, sample identification, volume of sample analyzed, value from the measurement system, factor and final value to be reported and each item must be recorded each time samples are analyzed. Ref: 15A NCAC 2H .0805 (a) (7) (H). Chloride - Standard Methods, 4500 Cl- B-1997 Recommendation: Method Blank titrant volumes are typically 0.7-0.8 mLs. The method states that 0.2 to 0.3 mLs is typical. It is recommended that the laboratory verify the purity of the water source utilized for the analysis with regard to the high blank titrant volumes. Chlorophyll  - EPA Method 445.0, Rev. 1.2 (Fluorometric) N. Finding: The lab is not taking appropriate corrective action when QC failures occur. Requirement: When quality control (QC) failures occur, the laboratory must attempt to determine the source of the problem and must apply corrective action. Part of the corrective action is notification to the end user. If data qualifiers are used to qualify samples not meeting QC requirements, the data may not be useable for the intended purposes. It is the responsibility of the laboratory to provide the client or end-user of the data with sufficient information to determine the usability of the qualified data. Ref: NC WW/GW LC Policy. Page 17 #10 Environment 1, Inc. Comment: Duplicate RPDs on 5/16/2015 and QCS standard comparisons in June and July, 2015, failed acceptance criteria. The associated data was not qualified on the client reports. Coliform, Fecal – Standard Methods, 9221 C E-2006 24hr 503 (MPN) O. Finding: The lab is not drying samples to a constant weight. Requirement: The solids content of treated domestic sludge can be highly variable. Therefore, an aliquot of each sample must be dried and the solids content determined in accordance with procedure 2540 G. of the 18th edition of Standard Methods for the Examination of Water and Wastewater (SM). Ref: Control of Pathogens and Vector Attraction in Sewage Sludge, EPA/625/R-92/013, July 2003, Appendix F, Section 1 (1.1). Coliform, Fecal – Standard Methods, 9222 D-1997 (MF) Comment: If the media is purchased pre-made, then NC requires only weekly known culture positive verifications. P. Finding: The lab is not always placing prepared cultures in the water bath within 30 min after filtration. Requirement: Place all prepared cultures in the water bath within 30 min after filtration. Ref: Standard Methods, 9222 D-1997. (d) (2) (d). Comment: For the data reviewed during the inspection, the requirement was not met on Fecal Coliform sample sets prepared on 5/4/2015, 5/5/2015, 6/2/2015 and 6/4/2015. Coliform, Fecal – Standard Methods, 9221 C E-2006 (MPN) (Aqueous) & 24hr 503 (Non-Aqueous) Q. Finding: Positive sample counts are not repeated by a different analyst on a monthly basis. Requirement: For routine performance evaluation, repeat counts on one or more positive samples at least monthly, record results, and compare the counts with those of other analysts testing the same samples. Replicate counts … should agree within 10%. Ref: Standard Methods, 9020 B-2005. (9) (a). R. Finding: The lab has not established duplicate acceptance criteria. Requirement: Each laboratory shall develop and maintain a document outlining the analytical quality control practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A NCAC 2H .0805 (a) (7). Page 18 #10 Environment 1, Inc. Comment: Acceptable options for determining acceptance criteria can be found in Standard Methods, 9020 B-2005. (9) (e), Standard Methods, 9221 C-2006. Table 9221:IV or establish your own. If the lab chooses Table 9221:IV to determine duplicate acceptance, then the 95% confidence windows of each sample must overlap to be acceptable. Fluoride – Standard Methods, 4500 F- C-1997 S. Finding: The lab is not analyzing a blank and calibration verification standard after every ten samples. Requirement: Verify calibration by periodically analyzing a calibration standard and calibration blank during a run—typically, after each batch of ten samples and at the end of the run. Ref: Standard Methods, 4020 B-2009. (2) (b). Requirement: The calibration blank and calibration verification standard (mid- range) must be analyzed initially (i.e., prior to sample analysis), after every tenth sample and at the end of each sample group to check for carry over and calibration drift. If either fall outside established quality control acceptance criteria, corrective action must be taken (e.g., repeating sample determinations since the last acceptable calibration verification, repeating the initial calibration, etc.). Ref: NC WW/GW LC Policy. T. Finding: The lab is not analyzing laboratory fortified matrix duplicates. Requirement: W hen appropriate for the analyte (Table 4020:I), include at least one LFM/LFMD daily or with each batch of 20 or fewer samples. Ref: Standard Methods, 4020 B-2009. (2) (g). Comment: The laboratory analyzes unspiked environmental sample duplicates. U. Finding: The data does not accurately reflect the QC practices which are being carried out. Requirement: Each laboratory shall develop and maintain a document outlining the analytical quality control practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A NCAC 2H .0805 (a) (7). Recommendation: It is recommended that the duplicate result column on the benchsheet be labeled as the average result, since the data in this column is the average of the results. Hardness – Standard Methods, 2340 C-1997 V. Finding: The lab is not diluting 25 mL of sample to 50 mL before titrating. Requirement: Titration precautions: Conduct titrations at or near normal room temperature. The color change becomes impractically slow as the sample approaches freezing temperature. Indicator decomposition becomes a problem in Page 19 #10 Environment 1, Inc. hot water. The specified pH may produce an environment conducive to CaCO3 precipitation. Although the titrant slowly re-dissolves such precipitates, a drifting end point often yields low results. Completion of the titration within 5 min minimizes the tendency for CaCO3 to precipitate. The following three methods also reduce precipitation loss. Dilute sample with distilled water to reduce CaCO3 concentration. This simple expedient has been incorporated in the procedure. If precipitation occurs at this dilution of 1 + 1 use modification 2) or 3). Using too small a sample contributes a systematic error due to the buret-reading error. Ref: SM 2340 C-1997. (1) (c) and SM 2340 C-1997. (1) (c) (1). Requirement: Dilute 25.0 mL sample to about 50 mL with distilled water in a porcelain casserole or other suitable vessel. Ref: SM 2340 C-1997. (3) (b). Nitrogen, Ammonia – EPA Method, 350.1, Rev. 2.0, 1993 Recommendation: It is recommended to develop and implement a distillation log. W. Finding: The lab is using ammonium sulfate as the stock solution. Requirement: Stock solution: Dissolve 3.819 g of anhydrous ammonium chloride, NH Cl 4 (CASRN 12125-02-9), dried at 105°C, in reagent water, and dilute to 1 L. 1.0 mL = 1.0 mg NH -N. Ref: EPA Method, 350.1, Rev. 2.0 (1993), Section 7.11. Comment: The lab has not performed a validation study to show that ammonium sulfate is equivalent to ammonium chloride. The lab is not reporting results as having been acquired via a modified method. X. Finding: The lab is not using the method prescribed sodium hypochlorite solution. Requirement: Dilute 250 mL of a bleach solution containing 5.25% NaOCl (CASRN 7681-52-9) (such as "Clorox") to 500 mL with reagent water. Available chlorine level should approximate 2-3%. Since "Clorox" is a proprietary product, its formulation is subject to change. The analyst must remain alert to detecting any variation in this product significant to its use in this procedure. Due to the instability of this product, storage over an extended period should be avoided. Ref: EPA Method 350.1, Rev. 2.0, 1993, Section 7.8. Comment: The lab uses a different formulation (i.e., not Clorox). Y. Finding: The lab is not analyzing both MS/MSD distilled and un-distilled for the distillation study. Requirement: Commercial Laboratories shall analyze initially at least six samples distilled and un-distilled, but at least two samples from all waste types. Analyze one sample distilled and un-distilled yearly from all wastes types. Ref: NC WW/GW Policy. Requirement: Manual distillation is not required if comparability data on representative effluent samples are on file to show that this preliminary distillation step is not necessary: however, manual distillation will be required to resolve any Page 20 #10 Environment 1, Inc. controversies. In general, the analytical method should be consulted regarding the need for distillation. If the method is not clear, the laboratory may compare a minimum of 9 different sample matrices to evaluate the need for distillation. For each matrix, a matrix spike and matrix spike duplicate are analyzed both with and without the distillation step. (A total of 36 samples, assuming 9 matrices). If results are comparable, the laboratory may dispense with the distillation step for future analysis. Comparable is defined as < 20% RPD for all tested matrices). Alternatively the two populations of spike recovery percentages may be compared using a recognized statistical test. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 77, No. 97, May 18, 2012; Table IB Notes (6). Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0, 1993 Recommendation: It is recommended to decrease the curve range to alleviate the possibility of having results which may be biased low on the upper 10% of the curve. The curve is beginning to plateau on the upper end. Recommendation: It is recommended to establish accuracy criteria for triplicate measurements of calibration standards. Phosphorus, Total – EPA Method, 365.4, 1974 Comment: The only data available for review was from PT samples. Z. Finding: The lab was not sufficiently documenting spike preparation. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon req uest. Ref: 15A NCAC 2H .0805 (a) (7) (A). Recommendation: The lab was not completely filling out the spike preparation worksheet (i.e., ___ mLs of spike up to ___ mLs with sample). It was recommended that the spike preparation worksheet be revised to read “_ml of _ mg/L spike solution added to the digestion tube” to more accurately describe the laboratory’s spike preparation scheme. AA. Finding: The lab is not documenting digestion times and temperatures. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Requirement: Place tubes in digester and switch to automatic mode. Set low temperature timer for 1 hour. Reset high temperature to 380°C and set timer for 2 1/2 hours. Ref: EPA Method, 365.4 (1974), Section 6.3. Nitrogen, Ammonia – EPA Method, 350.1, Rev. 2.0, 1993 Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0, 1993 Page 21 #10 Environment 1, Inc. BB. Finding: The lab is not using +10% recovery as the acceptance criterion for matrix spiking. Requirement: Calculate the percent recovery for each analyte, corrected for concentrations measured in the unfortified sample, and compare these values to the designated LFM recovery range 90-110%. Ref: EPA Method, 350.1, Rev. 2.0 (1993), Section 9.4.2 and EPA Method, 353.2, Rev. 2.0 (1993), Section 9.4.2. Nitrogen, Total Kjeldahl – EPA Method, 351.2, Rev. 2.0, 1993 Recommendation: It is recommended that the lab reanalyze samples known to have high nitrate concentrations. High nitrate concentrations (10X or more than the TKN level) result in low TKN values. If interference is suspected, samples should be diluted and reanalyzed. Ref: EPA Method, 351.2, Rev. 2.0 (1993), Section 4.1. Nitrogen, Ammonia – EPA Method, 350.1, Rev. 2.0, 1993 Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0, 1993 Nitrogen, Total Kjeldahl – EPA Method, 351.2, Rev. 2.0, 1993 CC. Finding: The lab is not using +10% recovery as the acceptance criterion for the QC Sample (QCS). Requirement: If the determined concentrations are not within ±10% of the stated values, performance of the determinative step of the method is unacceptable. Ref: EPA Method, 350.1, Rev. 2.0 (1993), Section 9.2.3, EPA Method, 351.2, Rev. 2.0 (1993), Section 9.2.3 and EPA Method, 353.2, Rev. 2.0 (1993), Section 9.2.3. Oil & Grease – EPA Method, 1664 Rev. B DD. Finding: The method defined balance weights are not checked. Requirement: Calibrate the analytical balance at 2 mg and 1000 mg, using class “S” or ASTM E 617-1997 Class 1 weights. It is recommended that the balance should also be calibrated with an additional class “S” or ASTM E 617-1997 Class 1 weight that will bracket the final expected weighing value. Ref: EPA Method 1664 Rev. B, Section 10.1. Comment: The balance is only checked with a 100 mg weight daily. EE. Finding: The Practical Quantitation Limit (PQL) is not being adjusted when less than 1L of sample is being analyzed. Requirement: Determine the original sample volume (Vs) in liters by filling the sample bottle to the mark with water and measuring the volume of water in a 1- to 2-L graduated cylinder. If the sample weight was used (Section 11.1.4), weigh the empty bottle and cap and determine Vs by difference, assuming a sample density of 1.00. Ref: EPA Method 1664 Rev. B, Section 11.4.5. Page 22 #10 Environment 1, Inc. Comment: Three instances of incorrect PQLs were observed in the data reviewed on 6/30/2015, 7/2/2015 and 7/7/2015 on samples from Carolina Seafood, CRSWMA Interim Regional and Captain Neill’s Seafood, respectively. Recommendation: It is recommended that the lab evaluate reported data to determine whether amended reports are needed to demonstrate compliance. pH – Standard Methods, 4500 H+ B-2000 – Field Comment: Only one buffer is required for the mid-day and post analysis calibration verification. Both the 4 S.U. and 7 S.U. buffers are currently being analyzed. Specific Conductance – Standard Methods, 2510 B-1997 – Field and Lab pH – Standard Methods, 4500 H+ B-2000 – Field and Lab Comment: Duplicates are not required. Specific Conductance – Standard Methods, 2510 B-1997 – Field and Lab FF. Finding: The benchsheet for Specific Conductance does not include meter calibration time. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Requirement: The following must be documented in indelible ink whenever sample analysis is performed: Meter Calibration Time. Ref: NC WW/GW LC Approved Procedure for the Analysis of Specific Conductance (Conductivity). Dissolved Oxygen - Standard Methods, 4500 O G-2001 Specific Conductance – Standard Methods, 2510 B-1997 – Field and Lab pH – Standard Methods, 4500 H+ B-2000 – Field and Lab GG. Finding: The units of measure are not all documented on the calibration logs. Requirement: All laboratories must use printed laboratory bench worksheets that include a space to enter the signature or initials of the analyst, date of analyses, sample identification, volume of sample analyzed, value from the measurement system, factor and final value to be reported and each item must be recorded each time samples are analyzed. Ref: 15A NCAC 2H .0805 (a) (7) (H). HH. Finding: The laboratory benchsheets and field log notebooks were lacking pertinent data: Laboratory Name, Instrument identification. Requirement: The following must be documented in indelible ink whenever sample analysis is performed: Sample site including facility name and location, ID, etc., Instrument Identification. Ref: NC WW/GW LC Approved Procedure for the Analysis of Dissolved Oxygen, NC WW/GW LC Approved Procedure for the Page 23 #10 Environment 1, Inc. Analysis of pH, NC W W/GW LC Approved Procedure for the Analysis of Specific Conductance (Conductivity). Phenols, Inorganic – EPA Method, 420.1, Rev. 1978 II. Finding: The benchsheet does not have a space for the adjusted pH value or the initial and final sample volumes. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Requirement: Measure 500 mL sample into a beaker. Lower the pH to approximately 4 with 1 + 9 H PO (7.1), add 5 mL CuSO solution (7.2) and transfer to the distillation apparatus. Ref: EPA Method, 420.1, Rev. (1978). Section 8.1.1. Requirement: Distill 450 mL of sample, stop the distillation, and when boiling ceases add 50 mL of warm distilled water to the flask and resume distillation until 500 mL have been collected. Ref: EPA Method, 420.1, Rev. (1978). Section 8.1.2. Residue, Dissolved – Standard Methods, 2540 C-1997 Recommendation: It is recommended to rinse the filter with a small portion of deionized water just before sample introduction to ensure the filter is properly seated. Residue, Total – Standard Methods, 2540 B-1997 Residue, Dissolved – Standard Methods, 2540 C-1997 Comment: Blanks are not required. JJ. Finding: Sample drying times and temperatures cannot be verified. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Requirement: Dry evaporated sample for at least 1 h in an oven at 103 to 105°C, cool dish in desiccator to balance temperature, and weigh. Ref: Standard Methods, 2540 B-1997. (3) (b). Requirement: Dry evaporated sample for at least 1 h in an oven at 180 + 2°C, cool in a desiccator to balance temperature, and weigh. Ref: Standard Methods, 2540 C- 1997. (3) (d). Comment: After the evaporation cycles at 98°C, the temperatures of the ovens are increased to 105 and 180°C for the drying cycles. The lab is documenting the time at which the ovens reach the drying cycle temperatures on a document which is not retained. Page 24 #10 Environment 1, Inc. KK. Finding: The length of time samples are dried during the drying time study is not documented. Requirement: Constant weights must be documented. The approved Methods require the following: “Repeat the cycle of drying, cooling, desiccating, and weighing until a constant weight is obtained or until the weight change is less than 4% of the previous weight or 0.5 mg, whichever is less.” In lieu of this, an annual study documenting the time required to dry representative samples to a constant weight may be performed. Verify minimum daily drying time is greater than or equal to the time used for the initial verification study drying cycle. Drying cycles must be a minimum 1 hour for verification. Ref: NC WW/GW LC Policy. Comment: The document which contains the drying time study does not include the length of time samples were in the oven at the proper temperature. This makes it impossible to determine the minimum amount of time verified by the drying time study. Comment: North Carolina allows for an annual drying study in lieu of the requirement above to repeat the drying cycle for every sample. A random full set of samples should be used for the drying study. The repeated drying time in the oven should be at least 1 hour long. The time used for the annual drying study is the minimum time that samples are to be dried until a new drying study is performed. Residue, Suspended – Standard Methods, 2540 D-1997 Residue, Dissolved – Standard Methods, 2540 C-1997 Recommendation: It is recommended that duplicate samples agree within 5% of their average weight. Residue, Dissolved – Standard Methods, 2540 C-1997 Residue, Suspended – Standard Methods, 2540 D-1997 Residue, Total – Standard Methods, 2540 B-1997 Comment: The laboratory was analyzing duplicates at 5% frequency. Standard Methods, 2540 C-1997 (3) (d), 2540 B-1997 (3) (b), and 2540 D-1997 (3) (c) state: Analyze at least 10% of all samples in duplicate. Demonstration of acceptable corrective action (i.e., the LIMS system was updated to require a duplicate analysis with every 10 samples with an implementation date of 8/13/2015) was observed by the inspector on 8/12/2015. No further response is necessary for this Finding but the laboratory is reminded to make any applicable SOP revisions. Comment: Incorrectly calculated results were observed on benchsheets / client reports for Suspended Residue on 5/11/2015 [i.e., reported 107 mg/L (due to calculating with an incorrect sample volume) but the correct result was 81.5 mg/L] and for Dissolved Residue on 6/8/2015 (i.e., reported 290 mg/L (due to miscalculating the initial average weight) but the correct result was 261 mg/L) and 7/7/2015 [i.e., reported 261 mg/L (due to calculating with an incorrect sample volume] but the correct result was 522 mg/L). Standard Methods, 2540 D-1997. (4) and Standard Methods, 2540 C-1997. (4) state: Calculations: Page 25 #10 Environment 1, Inc. mg total dissolved solids/L = (A - B) x 100 Sample volume, mL Where: A= weight of dried residue + dish, mg, and B= weight of dish, mg. mg total suspended solids/L = (A - B) x 100 Sample volume, mL Where: A= weight of filter + dried residue, mg, and B= weight of filter, mg. Demonstration of acceptable corrective action (i.e., results were recalculated and corrected reports were submitted to clients on 8/12/2015) was observed by the inspector on 8/12/2015. No further response is necessary for this Finding. Sulfate – Standard Methods, SO42 E-1997 LL. Finding: The lab is not analyzing a standard every three or four samples. Requirement: Check reliability of calibration curve by running a standard with every three or four samples. Ref: Standard Methods, SO42- E-1997. (4) (c). MM. Finding: The lab is not analyzing a MSD. Requirement: W hen appropriate for the analyte (Table 4020:I), include at least one LFM/LFMD daily or with each batch of 20 or fewer samples. Ref: Standard Methods, 4020 B-2009. (2) (g). Sulfide – Standard Methods, 4500 S²ˉ D-2000 Recommendation: It is recommended to document the wavelength used to analyze the samples on the benchsheets (i.e., 664 nm). Turbidity – Standard Methods, 2130 B-2001 NN. Finding: The lab is not retaining the documentation of the gel standard verification. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Requirement: Not all secondary standards have to be discarded when comparison with a primary standard shows that their turbidity value has changed. In some cases, the secondary standard should be simply relabeled with the new turbidity value. Ref: Standard Methods, 2130 B-2001. (3) (d). Page 26 #10 Environment 1, Inc. Comment: The verification of the Gel standard against prepared standards is documented on an index card that is taped to the wall and the concentration is written on the cap of the gel standard. Mercury – EPA Method, 1631 E OO. Finding: The lab is not recalibrating the instrument when more than 12 hours have elapsed since verification of system performance. Requirement: If system performance is verified at the end of an analytical batch using the OPR, analysis of samples and blanks may proceed without recalibration, unless more than 12 hours has elapsed since verification of system performance. Ref: EPA 1631 E, Section 10.1. Mercury – EPA Method, 245.1, Rev. 3.0, 1994 (Aqueous) PP. Finding: The lab is not analyzing laboratory reagent blanks. Requirement: Laboratory reagent blank (LRB) - The laboratory must analyze at least one LRB (Section 7.11.2) with each batch of 20 or fewer samples of the same matrix. Ref: EPA Method, 245.1, Rev. 3.0 (1994), Section 9.3.1. Hardness – Standard Methods, 2340 C-1997 Mercury – EPA Method, 1631 E Mercury – EPA Method, 245.1, Rev. 3.0, 1994 (Aqueous) Metals – EPA Method, 200.7, Rev. 4.4, 1994 QQ. Finding: The laboratory is exceeding 1% sample volume with spiking solution when preparing matrix spikes and is not using the proper correction factor when calculating the final result. Requirement: The volume of spike solution used in matrix spike preparation must in all cases be ≤ 10% of the total matrix spike volume. It is preferable that the spike solution constitutes ≤ 1% of the total matrix spike volume so that the matrix spike can be considered a whole volume sample with no adjustment by calculation necessary. If the spike solution volume constitutes >1% of the total sample volume, the sample concentration or spike concentration must be adjusted by calculation. Ref: NC WW/GW LC Policy. Comment: See the NC WW/GW LC Matrix Spiking Technical Assistance Document for further information, attached. Metals – EPA Method, 200.7, Rev. 4.4, 1994 RR. Finding: The laboratory is using an acceptance criterion of ≤ the reporting limit for the reagent, method and calibration blanks. Page 27 #10 Environment 1, Inc. Requirement: For analyses requiring a calibration curve, the concentration of reagent, method and calibration blanks must not exceed 50% of the reporting limit or as otherwise specified by the reference Method. Ref: NC WW/GW LC Policy. SS. Finding: The lab is not consistently analyzing matrix spikes at a frequency of 10%. Requirement: The laboratory must add a known amount of each analyte to a minimum of 10% of the routine samples. Ref: EPA Method, 200.7, Rev. 4.4 (1994), Section 9.4.2. Comment: For example, the data from April 6, 2015 contained 2 matrix spikes for 45 samples. TT. Finding: The lab is not always waiting 24 hours after chemical preservation to digest samples. Requirement: However, acid must be added at least 24 hours before analysis to dissolve any metals that adsorb to the container walls. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 77, No. 97, May 18, 2012; Table II. Footnote 19. UU. Finding: The lab is not using 5% as the acceptance criterion for the daily Instrument Performance Check (IPC). Requirement: Analysis of the IPC solution immediately following calibration must verify that the instrument is within ±5% of calibration with a relative standard deviation <3% from replicate integrations ≥4. Ref: EPA Method, 200.7, Rev. 4.4 (1994), Section 9.3.4. Comment: The laboratory is using an acceptance criterion of ± 10%. The lab refers to the IPC as ICV. VV. Finding: The lab is not using 5% as the acceptance criterion for the quarterly QCS. Requirement: To verify the calibration standards the determined mean concentrations from three analyses of the QCS must be within ±5% of the stated values. Ref: EPA Method, 200.7, Rev. 4.4 (1994), Section 9.2.3. Comment: The laboratory is using an acceptance criterion of ± 10%. WW. Finding: The lab is not using the method-defined acceptance criterion for Spectral Interference Check (SIC) analyses. Requirement: If the SIC solution confirms an operative interference that is ≥10% of the analyte concentration, the analyte must be determined using a wavelength and background correction location free of the interference or by another approved test procedure. Ref: EPA Method, 200.7, Rev. 4.4 (1994), Section 7.14. Page 28 #10 Environment 1, Inc. XX. Finding: The lab is not adding the correct volume of (1+1) hydrochloric acid (i.e., 1mL/100 mL of sample) to the sample. Requirement: For the determination of total recoverable analytes in aqueous samples (other than drinking water with <1 NTU turbidity), transfer a 100 mL (±1 mL) aliquot from a well-mixed, acid preserved sample to a 250 mL Griffin beaker (Sections 1.2, 1.3, 1.6, 1.7, 1.8, and 1.9). (When necessary, smaller sample aliquot volumes may be used.) Ref: EPA Method, 200.7, Rev. 4.4 (1994), Section 11.2.2. Requirement: Add 2 mL (1+1) nitric acid and 1.0 mL of (1+1) hydrochloric acid to the beaker containing the measured volume of sample. Ref: EPA Method, 200.7, Rev. 4.4 (1994), Section 11.2.3. Comment: 5 mL of (1+1) hydrochloric acid is added to 50 mL of sample. YY. Finding: Watch glasses are not being utilized during evaporation or reflux. Requirement: The beaker should be covered with an elevated watch glass or other necessary steps should be taken to prevent sample contamination from the fume hood environment. Ref: EPA Method, 200.7, Rev. 4.4 (1994), Section 11.2.3. Requirement: Cover the lip of the beaker with a watch glass to reduce additional evaporation and gently reflux the sample for 30 minutes. Ref: EPA Method, 200.7, Rev. 4.4 (1994), Section 11.2.5. ZZ. Finding: The lab benchsheet does not document digestion start time. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Requirement: However, acid must be added at least 24 hours before analysis to dissolve any metals that adsorb to the container walls. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 77, No. 97, May 18, 2012; Table II. Footnote 19. AAA. Finding: The lab benchsheet does not document digestion end time. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Requirement: Heat the sample and gently reflux for 30 minutes. Ref: EPA Method, 200.7, Rev. 4.4 (1994). Section 11.3.4. Metals - Standard Methods, 3111 B-1999 Page 29 #10 Environment 1, Inc. Recommendation: It is recommended that the lab periodically analyze field blanks. BBB. Finding: The lab is not carrying standards through the same digestion procedure used for samples. Requirement: If digestion is used, carry standards through the same digestion procedure used for samples. Ref: Standard Methods, 3111 A-1999. (5). CCC. Finding: The lab is not adding a known amount of the metal of interest approximately equal to the amount found in the sample after initial analysis and reanalyzing to confirm recovery to one sample out of every ten. Requirement: To one sample out of every ten (or one sample from each group of samples if less than ten are being analyzed) add a known amount of the metal of interest and reanalyze to confirm recovery. The amount of metal added should be approximately equal to the amount found. If little metal is present add an amount close to the middle of the linear range of the test. Recovery of added metal should be between 85 and 115%. Ref: Standard Methods, 3111 A-1999. (7). Mercury – EPA Method, 1631 E DDD. Finding: The lab is not analyzing 10% of samples as MS/MSDs. Requirement: To assess the performance of the Method on a given sample matrix, the laboratory must spike, in duplicate, a minimum of 10% (1 sample in 10) from a given sampling site or, if for compliance monitoring, from a given discharge. Therefore, an analytical batch of 20 samples would require two pairs of MS/MSD samples (four spiked samples total). Ref: EPA 1631 E, Section 9.3. Metals – EPA Method, 200.8, Rev. 5.4, 1994 Nitrogen, Ammonia – EPA Method, 350.1, Rev. 2.0, 1993 Nitrogen, NO2 + NO3 – EPA Method, 353.2, Rev. 2.0, 1993 EEE. Finding: The lab is not analyzing 10% of samples as MSs. Requirement: The laboratory must add a known amount of analyte to a minimum of 10% of the routine samples. Ref: EPA Method, 200.8, Rev. 5.4, (1994), Section 9.4.2. Requirement: The laboratory must add a known amount of analyte to a minimum of 10% of the routine samples. Ref: EPA Method, 350.1, Rev. 2.0 (1993), Section 9.2.3 and EPA Method, 353.2, Rev. 2.0 (1993), Section 9.2.3. Comment: The lab was analyzing 5% of samples as MSs. Phosphorus, Total – EPA Method, 365.4, 1974 Chloride – Standard Methods, 4500 Cl- B-1997 Fluoride – Standard Methods, 4500 F- C-1997 Purgeable, Aromatics – Standard Methods, 6200 C-1997 Page 30 #10 Environment 1, Inc. FFF. Finding: The lab is not analyzing MSDs. Requirement: Incorporate the following twelve quality control elements, where applicable, into the laboratory’s documented standard operating procedure (SOP) for performing compliance analyses when using an approved part 136 method when the method lacks such QA/QC procedures. One or more of the twelve QC elements may not apply to a given method and may be omitted if a written rationale is provided indicating why the element(s) is/are inappropriate for a specific method. Ref: Code of Federal Regulations, Title 40, Part 136.7; Federal Register Vol. 77, No. 97, May 18, 2012. (c) (1). Requirement: Matrix spike (MS) and matrix spike duplicate (MSD), or laboratory fortified matrix (LFM) and LFM duplicate, may be used for suspected matrix interference problems to assess precision. Ref: Code of Federal Regulations, Title 40, Part 136.7; Federal Register Vol. 77, No. 97, May 18, 2012. (c) (v). Requirement: Include at least one LFM/LFMD daily or with each batch of 20 or fewer samples. Ref: Standard Methods, 4020 B-2009, Table 4020:I and (2) (g). Requirement: Prepare at least one LFM/LFMD each day samples are prepared or with each preparation batch of 20 or fewer samples. Ref Standard Methods, 6020 B-2011, Table 6020:I and (2) (f). 1,2-Dibromoethane – EPA Method, 504.1 Recommendation: TCMX surrogate calculation is based on a single point estimation. Although surrogate analysis is not required by the method, it is recommended that surrogate recovery be based upon a multi-level calibration curve. Recommendation: It is recommended to optimize the temperature program of the oven. This may alleviate the co-elution of 1,2,3-TCP with other analytes. The current resolution requires either a qualifier to accompany the data or subcontracting of the analysis. There are currently no laboratory clients under our program that require the reporting of this analyte. GGG. Finding: The lab is not weighing standard preparation flasks or sample bottles for actual volume determination for standard preparation or samples. Requirement: Allow the flask to stand, un-stoppered, for about 10 minutes and weigh to the nearest 0.1 mg. Calculate the concentration in micrograms per microliter from the net gain in weight. Replace the container cap and weigh the container with contents to the nearest 0.1 g and record this weight for subsequent sample volume determination (Section 11.3). Ref: EPA Method, 504.1, Sections 7.3.1, 7.3.3 and 11.1.2. Pesticides, Organochlorine – SW846, 8081 B Comment: The only data available for review was from PT samples. Page 31 #10 Environment 1, Inc. HHH. Finding: The lab is not preparing a surrogate calibration curve. Requirement: For each analyte and surrogate of interest, prepare calibration standards at a minimum of five different concentrations by adding volumes of one or more stock standards to volumetric flasks and diluting to volume with an appropriate solvent. Ref: EPA SW -846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 B. Rev. 2, December 1996, Section 7.4.1.1. III. Finding: The use of manual integration is not properly documented. Requirement: When manual integration is employed, the laboratory must clearly identify manually integrated compounds, document the reason the manual integration was performed, the date performed and who completed the work. A flag or qualifier code may suffice for simple manual integrations. Ref: NC WW/GW Policy. Comment: The lab does not document the following when manual integrations are performed for Organochlorine Pesticides by SW846 8081 B: the reason the manual integration was performed, the date performed and who completed the work. JJJ. Finding: The laboratory is referencing an unapproved method on the extraction log. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15A NCAC 2H .0805 (a) (7) (A). Comment: The extraction log references SW -846 Method 8080. 1,2-Dibromoethane – EPA Method 504.1 Purgeable, Halocarbons – EPA Method 601 & SW-846 8021 B Purgeable, Aromatics – EPA Method 602 & SW-846 8021 B Purgeable, Organics – EPA Method 624 & SW-846 8260 B Base Neutral/Acid, Organics – EPA Method 625 & SW-846 8270 D Pesticides, Organochlorine – SW-846 8081 B KKK. Finding: Quality Control standards do not always contain all of the analytes of interest. Requirement: Quality Control Sample (QCS) -- A solution of method analytes of known concentrations that is used to fortify an aliquot of LRB or sample matrix. Ref: EPA Method 504.1, Rev. 1.1 (1995), Section 3.11. Requirement: A quality control (QC) check sample concentrate is required containing each parameter of interest at a concentration of 10 μg/mL in methanol. Ref: EPA Method 601, Section 8.2.1 and EPA Method 602, Section 8.2.1. Page 32 #10 Environment 1, Inc. Requirement: Prepare the QC check sample as described in Section 8.2.2. EPA Method 624, Section 7.4.1. Requirement: A quality control (QC) check sample concentrate is required containing each parameter of interest... Ref: EPA Method 625, Section 8.2.1. Requirement: The solution used to fortify a sample and/or an LCS should contain all of the target analytes and their concentration levels should be determined as described in Secs. 9.4.1 and 9.4.2. Ref: EPA SW -846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 D. Rev. 4, July 2014, Section 9.4.2. Requirement: Except where the determinative method contains alternative calibration verification criteria, if the % Drift or % Difference is >± 20%, the initial calibration relationship may no longer be valid. Ref: EPA SW -846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 D. Rev. 4, July 2014, Section 11.7. Requirement: It is the intent of the Environmental Protection Agency (EPA or the Agency) that all target analytes for a particular analysis be included in the initial calibration and calibration verification standard(s). These target analytes may not include the entire list of analytes (Sec. 1.1) for which the method has been demonstrated. However, the laboratory must not report a quantitative result for a target analyte that was not included in the calibration standard(s). Ref: EPA SW - 846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8021 B. Rev. 3, July 2014, Section 7.7.3. Requirement: The initial calibration curve (Sec. 7.3) for each compound of interest should be verified once every 12 hours prior to sample analysis, using the introduction technique used for samples. This is accomplished by analyzing a calibration standard at a concentration near the midpoint concentration for the calibrating range of the GC/MS. Ref: EPA SW -846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8260 B. Rev. 2, December 1996, Section 7.4.2. Requirement: The initial calibration (Sec. 11.3) for each compound of interest should be verified once every twelve hours prior to sample analysis, using the introduction technique and conditions used for samples. This is accomplished by analyzing a calibration standard (containing all the compounds for quantitation) at a concentration either near the midpoint concentration for the calibrating range of the GC/MS or near the action level for the project. The results must be compared against the most recent initial calibration curve and should meet the verification acceptance criteria provided in Secs. 11.4.5 through 11.4.7. Ref: EPA SW -846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8270 D. Rev. 5, July 2014, Section 11.4.3. Comment: Proficiency Testing samples do not contain all target analytes. The lab is utilizing left over Proficiency Testing samples as Quality Control standards (i.e., calibration verification standards, second source standards, etc.) Page 33 #10 Environment 1, Inc. PAPER TRAIL INVESTIGATION: The paper trail consisted of comparing original records (e.g., laboratory benchsheets, logbooks, chain of custodies, etc.) and client reports. Data were reviewed for the following project numbers: Client #s 1400 and 1001 for the parameters of Color, PC and Inorganic Phenol on 5/6/2015 and 6/9/2015, respectively. No transcription errors were detected. The facility appears to be doing a good job of accurately transcribing data. V. CONCLUSIONS: Correcting the above-cited Findings and implementing the Recommendations will help this lab to produce quality data and meet certification requirements. The inspector would like to thank the staff for its assistance during the inspection and data review process. Please respond to all Findings and include an implementation date for each corrective action. Report prepared by: Nick Jones Date: September 30, 2015 Report reviewed by: Jason Smith Date: October 29, 2015 Page 34 #10 Environment 1, Inc. Matrix Spiking (NC WW/GW LC Policy 05/11/2012) Unless the referenced method states a greater frequency, spike 5% of samples on a monthly basis. Laboratories analyzing less than 20 samples per month must analyze at least one matrix spike (MS) each month samples are analyzed. If MS results are out of control, the results must be qualified or the laboratory must take corrective action to rectify the effect, use another method, or employ the method of standard additions. When the method of choice specifies MS performance acceptance criteria for accuracy, and the laboratory chooses to develop statistically valid, laboratory-specific limits, the laboratory-generated limits cannot be less stringent than the criteria stated in the approved method. When spiking with multi-component standards, if the method does not specify the spiking components, the Laboratory Control Spike (LCS) and MS must contain all analytes that are reported. If the unspiked sample result is in the top 40% of the calibration range, the sample should be diluted and the MS prepared using the diluted sample. The recovery of the MS samples must be bracketed by the calibration range. The volume of spike solution used in MS preparation must in all cases be ≤ 10% of the total MS volume. It is preferable that the spike solution constitutes ≤ 1% of the total MS volume so that the MS can be considered a whole volume sample with no adjustment (i.e., volume correction) by calculation necessary. If the spike solution volume constitutes >1% of the total sample volume, the sample concentration or spike concentration must be adjusted by calculation. Parameters Excluded from MS Requirements Acidity Alkalinity BOD/CBOD Aquatic Humic Substances Chlorophyll All Bacteriological Parameters Color – ADMI Color - PtCo Conductivity Dissolved Oxygen Ignitability All Residues Paint Filter Test Turbidity pH Temperature Salinity Sulfite Total Residual Chlorine Vector Attraction Reduction (All Options) (Field Laboratories and Field Setting analyses are exempt.)