Loading...
HomeMy WebLinkAbout#65_0722_FINALNorth Carolina Department of Environmental Quality Pat McCrory Governor September 21, 2015 65 Ms. Alicia Goots TZ Osborne WRF Laboratory P.O. Box 3136 Greensboro, NC 27402-3136 Donald R. van der Vaart Secretary Subject: North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) Maintenance Inspection Dear Ms. Goots: Enclosed is a report for the inspection performed on July 22, 2015 by staff of the NC WW/GW LC program. I apologize for the delay in getting this report to you. Where finding(s) are cited in this report, a response is required. Within thirty days of receipt, please supply this office with a written item for item description of how these finding(s) were corrected and include an implementation date for each corrective action. If the finding(s) cited in the enclosed report are not corrected, enforcement actions may be recommended. For certification maintenance, your laboratory must continue to carry out the requirements set forth in 15A NCAC 2H .0800. Copies of the checklists completed during the inspection may be requested from this office. Thank you for your cooperation during the inspection. If you wish to obtain an electronic copy of this report by email or if you have questions or need additional information, please contact us at (828) 296-4677. Attachment cc: Beth Swanson Master File Sincerely, COY c�sa-t, Gary Francies, Technical Assistance/Compliance Specialist Division of Water Resources Water Sciences Section NC Wastewater/Groundwater Laboratory Certification Branch 1623 [fail Service Center, Raleigh, North Carolina 27699-1623 Location: 4405 Reedy Creek Road, Raleigh, North Carolina 27607 Phone: 919-733-3908 l FAX: 919-733-6241 Internet: www.dwglab.orq An Equal Opportunity t Affirmative Action Employer INSPECTION REPORT ROUTING SHEET To be attached to all inspection reports in-house only. Laboratory Cert. #: 65 Laboratory Name: TZ Osborne WRF Laboratory Inspection Type: Municipal Maintenance Inspector Name(s): Beth Swanson Gary Francies, Dana Satterwhite Nick Jones, Todd Crawford Jason Smith Inspection Date: Date Report Completed: Date Forwarded to Reviewer: Reviewed by: Date Review Completed: Cover Letter to use: Unit Supervisor/Chemist III: Date Received: Date Forwarded to Linda: Date Mailed: July 22 2015 August 20 2015 August 20 2015 Nick Jones September 2 2015 ❑ Insp. Initial ❑ Insp. No Finding ❑ Corrected Gary Francies ® Insp. Reg. ❑ Insp. CP ❑ Insp. Reg. Delay 9/4/2015 9/21 /2015 On -Site Inspection Report LABORATORY NAME: NPDES PERMIT #: ADDRESS: CERTIFICATE #: DATE OF INSPECTION: TYPE OF INSPECTION: AUDITOR(S): LOCAL PERSON(S) CONTACTED: I. INTRODUCTION: TZ Osborne WRF Laboratory NC0047384, NCO024325 P.O. Box 3136 Greensboro, NC 27402-3136 65 July 22, 2015 Municipal Maintenance Beth Swanson, Gary Francies, Dana Satterwhite, Nick Jones, Todd Crawford, and Jason Smith Alicia Goots and Martie Groome This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A NCAC 2H .0800 for the analysis of environmental samples. II. GENERAL COMMENTS: The laboratory was very clean and well organized. Data was readily available and analysts were knowledgeable and forthcoming. The organizational culture of this laboratory is one that focuses on quality of data. Everyone we interacted with was interested and willing to adapt their processes to the required and recommended changes. As shown in the report, the laboratory was proactive in responding to several Findings. Proficiency Testing (PT) samples have been analyzed for all certified parameters for the 2015 proficiency testing calendar year and the graded results were 100% acceptable. The laboratory is reminded that any time changes are made to laboratory operations; the laboratory must update the Quality Assurance (QA)/Standard Operating Procedures (SOP) document(s). Any changes made in response to the Findings, Recommendations or Comments listed in this report must be incorporated to insure the method is being performed as stated, references to methods are accurate, and the QA and/or SOP document(s) is in agreement with approved practice and regulatory requirements. In some instances, the laboratory may need to create a SOP to document how new functions or policy will be implemented. Contracted analyses are performed by Meritech, Inc. (Certification #165). The requirements associated with Findings B, E, M, N, P, and U have been implemented by our program since the last inspection. Page 2 #65 TZ Osborne WRF Laboratory III. FINDINGS REQUIREMENTS COMMENTS AND RECOMMENDATIONS: General Recommendation: It is recommended that the sample preservation and storage temperature requirements be changed in accordance with 40 CFR Part 136.3 Table II in all pertinent SOPs and in practice. The current practice is to store "at 4°C". 40 CFR Part 136.3 Table II allows 56°C for most inorganic and metals parameters and <10°C for bacterial tests, without evidence of freezing. Documentation Recommendation: It is recommended that all records of preventative maintenance performed on analytical instrumentation be documented in an individual equipment maintenance log book maintained by the laboratory. All records of corrective action taken or minor troubleshooting and repair performed should also be documented in the equipment maintenance log. The records may include the following, as appropriate: a. The identity of the item of equipment and its software; b. The manufacturer's name, type identification, and serial number or other unique identifier; C. The current location; d. The dates, results and copies of reports and certificates or all calibrations, adjustments, acceptance criteria, and the due date of the next calibration. e. The maintenance plan, where appropriate, and maintenance completed date. f. Documentation of any damage, malfunction, modification, or repair to the equipment. g. Signature or initials of the person performing the maintenance or troubleshooting. These records provide an historical reconstruction of the maintenance performed, a reference for effective troubleshooting in the future, and can also serve as a valuable training tool for new analysts. Comment: Improper error correction was noted that included the use of correction fluid and crossing out data with no initial and/or date. NC WW/GW LC policy states: All documentation errors must be corrected by drawing a single line through the error so that the original entry remains legible. Entries shall not be obliterated by erasures or markings Wite-OutO correction tape or similar products desianed to obliterate documentation are not to be used. Write the correction adjacent to the error. The correction must be initialed by the responsible individual and the date of change documented. All data and log entries must be written in indelible ink. Pencil entries are not acceptable. Notification of acceptable corrective action (i.e., on June 24, 2015, staff participated in a meeting detailing the proper procedure for error correction), was received by email on July 30, 2015 stating that all changes were implemented at that time. No further response is necessary for this Finding. Comment: Units of measure were not included on the temperature charts throughout the lab. This is considered pertinent information. The North Carolina Administrative Code, 15 NCAC 2H .0805 (a) (7) (A) states: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Notification of acceptable corrective action (i.e., a note was added to the top of all temperature charts in use stating that all temperatures are in °C) was received by email on July 30, 2015 and implemented immediately. No further response is necessary for this Finding. Comment: The date received for chemicals and consumables was not documented. NC WW/GW LC policy states: All chemicals, reagents, standards and consumables used by the laboratory must have the following information documented: Date received, Date Opened (in use) Vendor Lot Number, and Expiration Date (where specified). This information as well as the vendor and/or manufacturer, lot number, and expiration date must be retained for primary standards chemicals, reagents, and materials Page 3 #65 TZ Osborne WRF Laboratory used for a period of five years. Consumable materials such as pH buffers, lots of pre -made standards and/or media, solids and bacteria filters, etc. are included in this requirement. Notification of acceptable corrective action (i.e., traceability logs for consumables and chemicals received by the laboratory were created and implemented on August 3, 2015) was received by email on August 3, 2015. No further response is necessary for this Finding. Comment: The benchsheet for the analysis of DO, pH, and temperature did not contain instrument identification. The NC WW/GW LC Approved Procedure for the Analysis of Dissolved Oxygen, NC WW/GW LC Approved Procedure for the Analysis of pH and NC WW/GW LC Approved Procedure for the Analysis of Temperature documents state: The following must be documented in indelible ink whenever sample analysis is performed: Thermometer/Instrument Identification. This requirement is a new policy that has been implemented by our program since the last inspection. Notification of acceptable corrective action (i.e., instrument/thermometer ID was added to the benchsheet and implemented on August 3, 2015) was received by email on August 3, 2015. No further response is necessary for this Finding. A. Finding: The temperature correction factor is not always posted on thermometers for corrections of 0°C. Requirement: Also make any corrections to the data and document any correction that applies (even if zero) on both the thermometer/meter and on a separate sheet to be filed. Ref: NC WW/GW LC Policy. Comment: This was noted on the digital thermometer used for sample receipt temperature checks. All other thermometers had corrections (both zero and non -zero) posted. Quality Control B. Finding: Auto-pipettors are not being calibrated. Requirement: Mechanical volumetric liquid -dispensing devices (e.g., fixed and adjustable auto- pipettors, bottle -top dispensers, etc.), used for critical measurements, must be calibrated at least twice per year, approximately six months apart and documented. Each liquid -dispensing device must meet the manufacturer's statement of accuracy. Ref: NC WW/GW LC Policy. C. Finding: The annual verification of the digital thermometer used for sample receipt temperature checks cannot be confirmed. Requirement: All thermometers must meet NIST specifications for accuracy or be checked, at a minimum annually, against a NIST traceable thermometer and proper corrections made. Ref: 15A NCAC 2H .0805 (a) (7) (0). Comment: The thermometer is a NIST traceable thermometer but the expiration date was not legible on the thermometer and no paperwork was found to determine the expiration date. Either the expiration date information needs to be requested from the manufacturer or the thermometer needs to be verified against an NIST traceable thermometer. D. Finding: The thermal preservation of samples from North Buffalo is not being maintained during transportation. Requirement: A 15-minute limit to thermally or chemically preserve samples is allowed. The temperature of a temperature blank or of a representative sample from each cooler must be Page 4 #65 TZ Osborne WRF Laboratory recorded upon receipt in the laboratory. Occasionally, samples that are collected and delivered to the lab within a short period of time may not have time to cool to less than or equal to 6°C (less than 10' C for Bacteriological analyses) before they arrive at the laboratory. If these samples are placed in ice immediately after collection and are shipped on ice or in ice slurry (NOTE: frozen or blue ice packs are not acceptable) the sample collectors have complied with these requirements to the best of their ability and the samples will be accepted. Documentation of the actual sample temperature at the time of collection and upon receipt at the laboratory must exhibit a downward trend and will complete the preservation documentation requirements. Ref: NC WW/GW LC Policy Comment: The samples from North Buffalo are composite samples which are placed into a refrigerator for several hours until they are transported to the TZ Osborne facility. However, the samples are not transported on ice, and the temperature upon receipt exhibits an upward trend from the temperature taken at the time of collection. E. Finding: Chemical preservation is not being conducted within 15 minutes of sample collection. Requirement: Except where noted in this Table II and the method for the parameter, preserve each grab sample within 15 minutes of collection. For a composite sample collected with an automated sampler (e.g., using a 24-hour composite sampler; see 40 CFR 122.21 (g) (7) (i) or 40 CFR Part 403, Appendix E), refrigerate the sample at :56 °C during collection unless specified otherwise in this Table II or in the method(s). For a composite sample to be split into separate aliquots for preservation and/or analysis, maintain the sample at :56 °C, unless specified otherwise in this Table II or in the method(s), until collection, splitting, and preservation is completed. Add the preservative to the sample container prior to sample collection when the preservative will not compromise the integrity of a grab sample, a composite sample, or an aliquot split from a composite sample; otherwise, preserve the grab sample, composite sample, or aliquot split from a composite sample within 15 minutes of collection. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 72, No. 47, March 12, 2007; Table II. SOP Recommendation: It is recommended that all SOPs be reviewed and evaluated for use of the word "should". SOPs are intended to describe procedures exactly as they are to be performed. While some uses of the word "should" versus "must" are noted in Finding I, not all have been listed in this report. Recommendation: It is recommended that SOPs include a revision and review history with relevant dates (e.g., effective, review and revision dates) and a brief description of the change(s) made. Recommendation: It is recommended that acronyms and terms for Quality Assurance/Quality Control (QA/QC) elements (e.g., LRL or PQL, LRB, CAL BILK, FRB, etc.) be defined in the Metals SOP and used consistently across the data and SOP. Recommendation: It is recommended that the Metals SOP state the minimum purity of the Argon used. Recommendation: It is recommended that section C of the Metals SOP state examples of when the reporting limit would be adjusted (e.g., dilution and dropping the lowest standard). Recommendation: It is recommended that the quality control section of the Total Residual Chlorine (TRC) SOP only include items that apply to the low range procedure. For instance, the DPD is only used with the LaMotte 1200 colorimeter that is used for TRC screening in ammonia and CBOD samples. Comment: An inconsistency and/or omission was noted between the Standard Operating Procedure (SOP) and laboratory practice as follows: personnel were not following procedures as stated in the Page 5 #65 TZ Osborne WRF Laboratory Laboratory's SOP, the method reference in the SOP is not correct, the SOP is in direct conflict with method and/or regulatory requirements, and the SOP does not describe in detail how the method is performed (observations are listed below). The North Carolina Administrative Code, 15 NCAC 2H .0805 (a) (7) states: Each laboratory shall develop and maintain a document outlining the analytical quality control practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Demonstration of acceptable corrective action (i.e., updated SOPs were submitted) was received by email on July 30, 2015. No further response is necessary for this finding. Residue, Suspended (SM 2540 D-1997) o The SOP did not state that wide bore pipettes were being used as required by the method. o The SOP referenced the 18th Edition of Standard Methods rather than Standard Methods Committee approval year. o The SOP did not state that the volume of sample is chosen to yield between 2.5 and 200 mg of dried residue. o The SOP did not state that a dry filter blank was analyzed daily. • Proficiency Testing o The laboratory did not have an SOP to detail the analysis of PT samples. Bacteria- Coliform Fecal (SM 9222 D-1997) o The SOP referenced the 18th Edition of Standard Methods rather than Standard Methods Committee approval year. o The SOP included preservation and hold time language from Standard Methods, but this would not apply as the requirements in 40 CFR Part 136.3, Table II takes precedent. o The SOP incorrectly included the word "should" versus "must" in various sections. o The SOP stated that the buffered dilution water could be kept indefinitely instead of 6 months. o The SOP did not clearly describe the acceptance criterion for duplicate analyses. F. Finding: An inconsistency and/or omission was noted between the Standard Operating Procedure (SOP) and laboratory practice as follows: The method reference in the SOP is not correct. o Dissolved Oxygen (SM 4500 G-2001) — The SOP references SM 4500 O G- 1998 18th Edition. o Phosphorous, Total (SM 4500 P E-1997) —The SOP references the 18th Edition of Standard Methods rather than Standard Methods Committee approval year. o COD (SM 6220 D-1997) — The SOP references the 18th Edition of Standard Methods rather than Standard Methods Committee approval year. o BOD/CBOD (SM 5210 B-2001) —The SOP references the 18th Edition of Standard Methods rather than Standard Methods Committee approval year. o Nitrogen, Ammonia (SM 4500 NH3 D-1997) —The SOP references the 18th Edition of Standard Methods rather than Standard Methods Committee approval year. Personnel were not following procedures as stated in the Laboratory's SOP. o Metals (EPA 200.8, Rev 5.4, 1994) —The SOP states that the duplicate acceptance criterion is 10% RPD, but the analyst uses 20% RPD. o Metals (EPA 200.8, Rev 5.4, 1994) —The SOP states that a sample duplicate is being analyzed instead of a Laboratory Fortified Matrix Duplicate (LFMD). The SOP does not describe in detail how the method is performed. Page 6 #65 TZ Osborne WRF Laboratory o Dissolved Oxygen (SM 4500 G-2001) — The SOP contains use of the word "should" instead of "must" in the sections of Sampling and Holding Time. o pH (SM 4500 H+ B-2000) — The SOP contains use of the word "should" instead of "must" in the Holding Time section. o pH (SM 4500 H+ B-2000) — The SOP does not state the check standard buffer acceptance criterion (i.e., ±0.1 SU). o Phosphorous, Total (SM 4500 P E-1997) — The cell path length and wavelength are not specified in the SOP. o Phosphorous, Total (SM 4500 P E-1997) — The SOP does not include the sample digestion procedure. o Phosphorous, Total (SM 4500 P E-1997) — The spectrophotometer set-up section lacks detail about instrument warm up. o Phosphorous, Total (SM 4500 P E-1997) — Sections 3.a.1 and 3.a.2 of the SOP are lacking units (i.e., 5.0 Second Source Standard). o Phosphorous, Total (SM 4500 P E-1997) — The SOP does not include the calibration blank acceptance criterion. o Phosphorous, Total (SM 4500 P E-1997) — The SOP does not include the acceptance criterion for LFM recovery. o Phosphorous, Total (SM 4500 P E-1997) — The SOP does not include the addition of 30 mL of ascorbic acid to the single acid reagent. o BOD/CBOD (SM 5210 B-2001) — Section D.5. of the SOP does not detail how the residual chlorine of samples is checked. o BOD/CBOD (SM 5210 B-2001) — Section D.9. of the SOP does not include directions to seed the sample if it is greater than 6 hours past collection. o BOD/CBOD (SM 6210 B-2001) — The SOP does not specify the CBOD acceptance criterion. This may be set at 164 ± 30.7 mg/L [allowed by EPA for CBOD] or 198 ± 30.5 mg/L [specified in the method]; however, one range must be chosen and consistently used. o BOD/CBOD (SM 5210 B-2001) — The SOP does not state that nitrification inhibitor is added to the CBOD GGA standards. o BOD/CBOD (SM 6210 B-2001) — Section EA. specifies the number of dilutions for influent and effluent samples but not pretreatment/industry samples. o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not specify what type of bottle (i.e., polypropylene) is used for sample collection. o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not describe the manufacturer recommended operating conditions of the ICP/MS instrument. o Metals (EPA 200.8, Rev 5.4, 1994) — Section E.7 of the SOP has instances where "should" is used in place of "must". o Metals (EPA 200.8, Rev 6.4, 1994) — Section C of the SOP does not state the strength of HNO3 used (i.e., 1+1). o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not include reagent and standard preparation instructions (e.g., 1+1 HNO3, LFM, etc). o Metals (EPA 200.8, Rev 6.4, 1994) — The SOP does not state the manufacturer's conditions for an acceptable tune to evaluate against the requirements in Section 10.2.2 of the method. o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not state which specific internal standards are used. o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not state that samples are preserved with HNO3 to pH <2 at least 24 hours prior to analysis. o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not include the 18 om water rinse in the glassware instructions. Page 7 #65 TZ Osborne WRF Laboratory o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not state the masses being monitored. o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not describe the sample results reporting scheme (e.g., For all sample analyses, a minimum of three replicate integrations are reported for data acquisition. Use the average of the integrations for data reporting.). o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not state the frequency (e.g., quarterly) and acceptance criterion (i.e., ±10%) for the Quality Control Standard (QCS). o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not describe how to calculate and report sample results (e.g., use of significant figures, dilution factors, units, etc.). o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not state that an initial demonstration of performance, consisting of a determination of the linear calibration range, analysis of quality control samples and determination of method detection limits, must be performed prior to analyses conducted and any time a change in instrument hardware or operating conditions are made. o Metals (EPA 200.8, Rev 5.4, 1994) — The SOP does not state the acceptance criteria for blanks (e.g., continuing calibration blanks, laboratory reagent blank) o Nitrogen, Ammonia (SM 4500 NH3 D-1997) — The SOP does not state how the sample is checked for total residual chlorine. o Nitrogen, Ammonia (SM 4500 NH3 D-1997) — The SOP does not include the calculation needed when differing amounts of NaOH are added to calibration standards and samples. o Nitrogen, Ammonia (SM 4500 NH3 D-1997) — The SOP does not explicitly state that a LFMD is being analyzed although it does give an acceptance criterion for precision (i.e., relative percent difference). o Chlorine, Total Residual (SM 4500 CI G- 2000) — The SOP does not state the acceptance criterion for the annual calibration curve verification. o Chlorine, Total Residual (SM 4500 Cl G- 2000) — The SOP does not state what standard concentrations are used to verify the factory -set calibration. o Chlorine, Total Residual (SM 4500 CI G- 2000) — The calculation section of the SOP does not state that the blanking reagent value is subtracted from the sample result. o Chlorine, Total Residual (SM 4500 CI G- 2000) — The SOP does not state to record the calibration time. o COD (SM 5220 D-1997) — The SOP for COD held little to no description of the quality control elements for the parameter. This includes but may not be limited to; the reporting limit, the 5-point calibration curve, LCS, duplicates, blanks, QC acceptance criteria, and the spiking procedure. The SOP is in direct conflict with the method and/or a regulatory requirement. o pH (SM 4500 H+ B-2000) — The SOP does not state that the 4 SU check buffer is analyzed after calibration with 7 SU and 10 SU buffers. This means the sample values would not be bracketed with calibration and/or check buffers.. o Phosphorous, Total (SM 4500 P E-1997) — The SOP states that the single reaction reagent can be stored at 40C and is stable for one week, but the method states it is only stable for 4 hours. o BOD/CBOD (SM 6210 B-2001) — Section C of the SOP includes holding time information from Standards Methods, but this would not apply as 40 CFR Part 136.3, Table II takes precedent. o BOD/CBOD (SM 5210 B-2001) — Section D.6. does not include instructions to titrate the sample aliquot to the end point and then add the appropriate amount of sodium sulfite to the sample. Page 8 #65 TZ Osborne WRF Laboratory o BOD/CBOD (SM 5210 B-2001) — The SOP states that only 2 glucose glutamic acid (GGA) checks are analyzed. o BOD/CBOD (SM 5210 B-2001) — The SOP states that data is qualified when the dilution water blank depletes >0.2 mg/L, but the requirement is >0.20 mg/L. o BOD/CBOD (SM 5210 B-2001) — Section B of the SOP states that 2 dilutions are analyzed per sample, but the method requires at least three dilutions per sample. o BOD/CBOD (SM 5210 B-2001) — The SOP states that the meter is recalibrated after every 10 samples. This is only required if the drift check is unacceptable (i.e., not within ±0.2 mg/L). o Bacteria- Coliform Fecal (SM 9222 D-1997) — The SOP does not state that reagent water testing is performed yearly as required by Standard Methods, 9020 B-2005. Table 9020:11. NC WW/GW LC Policy only requires the testing of certain analytes in the referenced table. These include conductivity, TOC, Cadmium, Chromium, Nickel, Lead, and Zinc. o Nitrogen, Ammonia (SM 4500 NH3 D-1997) — The SOP states that the meter is performing a blank correction, which is not allowed by the method. o Nitrogen, Ammonia (SM 4500 NH3 D-1997) — The SOP states that the acceptance criterion for the second source laboratory control standard (LCS) is ±20% recovery, but the method states ±15%. The checklist completed by the laboratory states that ±10% is used. o Nitrogen, Ammonia (SM 4500 NH3 D-1997) — The SOP only lists one concentration for the LCS (i.e., 0.05 mg/L). An additional 0.5 mg/L LCS is required and is analyzed on the high range curve. The laboratory does not have an SOP for the certified parameter. o Temperature (SM 2550 B-2000) Requirement: Each laboratory shall develop and maintain a document outlining the analytical quality control practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A NCAC 2H .0805 (a) (7). Please submit updated SOPs with the items listed above addressed by December 1, 2015. Residue, Suspended — Standard Methods, 2540 D-1997 G. Finding: The verified thermometer reading varied more than 1 °C from the NIST traceable thermometer reading. Requirement: All thermometers and temperature measuring devices must be checked every 12 months against a NIST certified or NIST traceable thermometer and the process documented. To check a thermometer or the temperature sensor of a meter, read the temperature of the thermometer/meter against a NIST certified or NIST traceable thermometer and record the two temperatures. The calibration must be performed at a temperature that corresponds to the temperature used by the incubator, refrigerator, freezer, etc. In the case of temperature measuring devices used to perform variable temperature readings the calibration must be performed at a temperature range that approximates the range of the samples. The thermometer/meter readings must be less than or equal to 1°C from the NIST certified or NIST traceable thermometer reading. Ref: NC WW/GW LC Policy. Comment: The thermometer used in the suspended residue oven has a correction of + 2°C. Page 9 #65 TZ Osborne WRF Laboratory Phosphorous, Total — Standard Methods, 4500 P E-1999 Comment: The laboratory performs an acid wash on the glassware once a month. The method implies glassware is acid washed before each use. Once the analysis of a method blank is implemented (see Finding H), monitor the blank values on randomly selected glassware to determine if more frequent acid washing is required. Recommendation: The laboratory is adding 0.5 mg of ammonium persulfate during digestion but section 5.c of Standard Methods, 4500 P B-1997 states to use 0.4 mg. Since the ammonium persulfate is used as a catalyst, it is unlikely that this negligible difference will affect the results. No issues were evident in the recovery of the digested standards for the data reviewed. Despite this, we recommend the laboratory start using 0.4 mg of ammonium persulfate as stated in the method. Comment: The laboratory was not analyzing 6 non -zero standard concentrations in the calibration curve. The instrument (i.e., Thermo Scientific Genesys 20) was not able to store a calibration curve, so each day the analyst was unknowingly performing a one standard calibration instead of analyzing a check standard against a stored calibration curve. Standard Methods, 4500 P E-1999. (4) (c) states: Prepare individual calibration curves from a series of six standards within the phosphate ranges indicated in 4500 P.E.1 c. Notification of acceptable corrective action (i.e., the laboratory is analyzing samples using a Hach DR3900 meter with a 6 non -zero point curve that the analyst constructed on August 3, 2015 and is stored in the meter) was received by email on September 21, 2015, "with implementation of sample analysis on August 20, 2015. No further response is necessary for this Finding. Comment: The laboratory submitted notification on July 30, 2015, that all samples for total phosphorous analysis are being sent to a contract lab until calibration issues have been satisfactorily resolved. H. Finding: The laboratory is not analyzing a method blank. Requirement: Method Blank (MB): Include at least one MB daily or with each batch of 20 or fewer samples, whichever is more frequent. Ref: Standard Methods, 4020 B-2009, Rev. 2011. (2) (d). Requirement: A reagent blank (method blank) consists of reagent water and all reagents (including preservatives) that normally are in contact with a sample during the entire analytical procedure. Ref: Standard Methods, 1020 B-2011. (5). Finding: A calibration blank is not analyzed initially after calibration. Requirement: The calibration blank and calibration verification standard (mid -range) must be analyzed initially (i.e., prior to sample analysis), after every tenth sample and at the end of each sample group to check for carry over and calibration drift. Ref: NC WW/GW LC Policy. J. Finding: A calibration blank and mid -range standard are not analyzed at the end of analysis. Requirement: The calibration blank and calibration verification standard (mid -range) must be analyzed initially (i.e., prior to sample analysis), after every tenth sample and at the end of each sample group to check for carry over and calibration drift. Ref: NC WW/GW LC Policy. K. Finding: Documentation of the time samples are placed into the autoclave and autoclave maximum temperature and pressure are not linked to the applicable batch of samples. This is considered pertinent information. Page 10 #65 TZ Osborne WRF Laboratory Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15 NCAC 2H .0805 (a) (7) (A). Requirement: Each day an incubator, oven, waterbath or refrigerator is used, the temperature must be checked recorded, and initialed. During each use the autoclave maximum temperature and pressure must be checked recorded, and initialed. Ref: 15 NCAC 2H .0805 (a) (7) Requirement: Alternately, heat for 30 min in an autoclave or pressure cooker at 98 to 137 kPa. Ref: Standard Methods, 4500 P B-1999. (5) (c). Recommendation: The laboratory's autoclave prints a log of time, temperature and pressure during each use. It is recommended that the log printed each time samples are autoclaved be initialed and attached to the total phosphorous benchsheet to satisfy the documentation requirement. L. Finding: The laboratory is using the wrong spike concentration when calculating the LFM recovery. Requirement: LFM recovery: (CSXfj — C X 100 = %Recovery LFM or LFMD S where: CS = LFM concentration determined experimentally, f= spike dilution correction, C = concentration of sample before spiking, and S = concentration of spike. NOTE: f should be greater than 0.95. More than 5% dilution due to spiking changes the matrix significantly. Ideally keep f to above 0.99 (equivalent to 1 % dilution of sample due to spike addition), in which case f can be ignored and the equation simplified to eliminate f. Ref: Standard Methods, 4020 B-2009, Rev. 2011. (3) (a). Comment: Instead of using the true concentration of the spike that is added to the LFM, the laboratory is using the experimentally determined concentration obtained for the LFB. Comment: The note above from Standard Methods states that the volume of spike should be <5% of the total, but as a reminder, NC WW/GW LC policy allows up to 10% as long as spike volumes above 1 % are corrected through calculation. M. Finding: The laboratory is not analyzing a LFMD. Requirement: Include at least one LFM/LFMD daily or with each batch of 20 or fewer samples. Ref: SM 4020 B-2009, Rev. 2011, Table 4020:1 and (2) (g). Comment: LFM and LFMD are also known a Matrix Spike (MS) and Matrix Spike Duplicate (MSD). Comment: Analysis of the LFMD (or MSD) is acceptable to satisfy the sample duplicate requirement listed in North Carolina Administrative Code, 15A NCAC 2H .0805 (a) (7) (C). Page 11 #65 TZ Osborne WRF Laboratory N. Finding: The laboratory is exceeding 1 % sample volume with spike solution when preparing matrix spikes and is not using the proper correction factor when calculating the final result. Requirement: The volume of spike solution used in matrix spike preparation must in all cases be <_ 10% of the total matrix spike volume. It is preferable that the spike solution constitutes <_ 1 % of the total matrix spike volume so that the matrix spike can be considered a whole volume sample with no adjustment by calculation necessary. If the spike solution volume constitutes >1% of the total sample volume the sample concentration or spike concentration must be adjusted by calculation. Ref: NC WW/GW LC Policy. Recommendation: It is recommended that the spike solution constitutes :51 % of the total LFM volume. The laboratory is currently preparing LFMs with 4% spike volume. BOD/CBOD — Standard Methods, 5210 B-2001 Comment: The laboratory may choose to adopt the less stringent duplicate acceptance criterion of 30% RPD as prescribed in the method. Comment: The laboratory is reminded that data does not need to be flagged for seed control factors (SCF) above 1.0 mg/L as long as there is evidence to show that a higher SCF is necessary to obtain an acceptable GGA recovery. This may also be removed from the benchsheet. Comment: The laboratory is reminded that the meter does not need to be calibrated at the end of analysis. Recommendation: It is recommended that a seeded blank be analyzed with industrial samples as is being done with the facility's influent and effluent samples. Comment: The laboratory was only setting up one GGA standard with industrial samples. Standard Methods, 5210 B-2001. (6) (b) states: Add sufficient amounts of standard glucose-glutamic acid solution (¶ 3h) to give 3.0 mg glucose/L and 3.0 mg glutamic acid/L in each of three test bottles (20 mL GGA solution/L seeded dilution water or 6.0 mU300-mL bottle). Demonstration of acceptable corrective action (i.e., an industrial waste COC/analysis benchsheet with 3 GGA standards typed permanently on the benchsheet was received by email on July 30, 2015 and implemented on August 3, 2015. No further response is necessary for this Finding. O. Finding: The laboratory is only duplicating the influent sample. Requirement: Duplicates: When appropriate (Table 5020:1), randomly select routine samples to be analyzed twice. Ref: Standard Methods, 5020 B- 2010. (2) (f). Residue, Suspended — Standard Methods, 2540 D-1997 Comment: The laboratory was not seating the filter with water prior to filtering the sample. Standard Methods, 2540 D-1997. (3) (c) states: Assemble filtering apparatus and filter and begin suction. Wet filter with a small volume of reagent -grade water to seat it. Demonstration of acceptable corrective action (i.e., an updated SOP with the filter seating step added) was received by email on July 30, 2015, with a stated implementation date of August 3, 2015. No further response is necessary for this Finding. Comment: The laboratory was filtering longer than the method allowed maximum of 10 minutes. Laboratory practice was to filter between 15 and 20 minutes. Standard Methods, 2540 D-1997. (3) (b) states: If complete filtration takes more than 10 minutes increase filter diameter or decrease sample Page 12 #65 TZ Osborne WRF Laboratory volume. Standard Methods, 2540 D-1997. (1) (b) states: Prolonged filtration times resulting from filter clogging may produce high results owing to increased colloidal materials captured on the clogged filter. Demonstration of acceptable corrective action (i.e., an updated SOP with directions to decrease sample volume if filtration takes longer than 10 minutes) was received by email on July 30, 2015, with a stated implementation date of August 3, 2015. No further response is necessary for this Finding. Comment: The laboratory was not adjusting the reporting limit for the amount of sample analyzed when less than 2.5 mg of dried residue was obtained on volumes less than 1000mL. NC WW/GW LC Policy states: The minimum weight gain allowed by any approved method is 2.5 mg. Choose sample volume to yield between 2.5 and 200 mg dried residue. This establishes a minimum reporting value of 2.5 mg/L when 1000 mL of sample is analyzed. If complete filtration takes more than 10 minutes increase filter diameter or decrease sample volume. In instances where the weight pain is less than the required 2.5 mg the value must be reported as less than the appropriate value based upon the volume used. Demonstration of acceptable corrective action (i.e., an updated SOP stating the proper adjustment of the reporting limit and a statement that it was implemented in practice as of August 3, 2015) was received by emails on July 30, 2015 and August 6, 2015. No further response is necessary for this Finding. Bacteria- Coliform Fecal — Standard Methods, 9222 D-1997 Comment: The fecal water bath temperature was only being checked once per day. Standard Methods, 9020 B-2005. (4) (n) states: When incubator is in use, monitor and record calibration -corrected temperature twice daily. This is a new method requirement since the last inspection. Notification of acceptable corrective action (i.e., an updated SOP stating temperatures are checked twice daily and a statement that it was implemented in practice as of August 3, 2015) was received by emails on July 30, 2015 and August 6, 2015. No further response is necessary for this Finding. Comment: A total immersion thermometer was being used in the fecal water bath improperly. The Environmental Protection Agency's (EPA) guidance website titled "Selecting Alternatives to Mercury -Filled Thermometers" states: Thermometers with no indicated depth are the total immersion type. When a partial -immersion thermometer is used, the bottom of the thermometer up to the immersion line should be exposed to the temperature being measured, with the remainder of the thermometer exposed to ambient conditions. When a total immersion thermometer is used the bulb and the entire portion of the stem containing liquid except for the last 1 cm are exposed to the temperature being measured. If the thermometer is not used in this manner, the thermometer immersion is incorrect. Notification of acceptable corrective action (i.e., a statement that a NIST traceable partial immersion thermometer with corrections of 0.06°C at 40°C and 0.00°C at 500C was ordered and received and is being used in the fecal water bath immersed to the specified immersion line) was received by email on July 30, 2015 and September 18, 2015 with implementation on July 29, 2015. No further response is necessary for this Finding. Comment: The laboratory was not testing a culture positive weekly for purchased MFC broth. Standard Methods, 9020 B-2005. (9) (b) states: Use certified reference cultures. For each lot of medium received, each laboratory prepared batch of medium, and each lot of purchased prepared medium, verify appropriate response by testing with known positive and negative control cultures for the organism(s) under test. Record results. This is a new method requirement since the last inspection. Demonstration of acceptable corrective action (i.e., an updated SOP stating that a culture positive is tested weekly with the purchased MFC broth and an updated benchsheet with a space for the weekly culture positive check) was received by email on July 30, 2015, with implementation of the practice on August 3, 2015. No further response is necessary for this Finding. Comment: The laboratory was not sterilizing the magnesium chloride solution. Standard Methods, 9050 C-2006. (1) (a) (2) states: Magnesium chloride stock solution: Add magnesium chloride (38 g/L MgC12 or 81.1 9 MgC12 - 6H2O) to 1 L reagent grade water. Sterilize and store stock solution under refrigerated Page 13 #65 TZ Osborne WRF Laboratory conditions, discarding if solution becomes turbid. Demonstration of acceptable corrective action (i.e., an updated SOP stating the magnesium chloride solution is sterilized) was received by email on July 30, 2015, with implementation of the practice on August 3, 2015. No further response is necessary for this Finding. Comment: The filtration start and end times were not being documented. This is considered pertinent information to assure plates are placed into the incubator within 30 minutes of filtration. The North Carolina Administrative Code, 15 NCAC 2H .0805 (a) (7) (A) states: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Standard Methods 9222 D-1997. (2) (d) states: Place all prepared cultures in the water bath within 30 min after filtration. Demonstration of acceptable corrective action (i.e., an updated SOP and updated benchsheet with spaces to enter the filtration start and end times) was received by email on July 30, 2015, with implementation of the practice on August 3, 2015. No further response is necessary for this Finding. Comment: The laboratory was not conducting comparison testing on new lots of consumables (i.e., membrane filters, pads, and media). NC WW/GW LC Policy states: When a new lot of culture medium, pads, or membrane filters is to be used, a comparison of the current lot in use (reference lot) against the new lot (test lot), be made. As a minimum, make single analyses on five positive samples. Demonstration of acceptable corrective action (i.e., an updated SOP including the comparison testing on new lots of consumables) was received by email on July 30, 2015, with implementation of the practice on August 3, 2015. Please submit the first comparison test that is completed. Comment: The laboratory was not conducting a monthly recount of a positive plate by one analyst and comparison counts between analysts. Standard Methods, 9020 B-2005. (9) (a) states: For routine performance evaluation, repeat counts on one or more positive samples at least monthly, record results, and compare the counts with those of other analysts testing the same samples. Replicate counts for the same analyst should agree within 5% (within analyst repeatability of counting) and those between analysts should agree within 10% (between analysts reproducibility of counting). If they do not agree, initiate investigation and any necessary corrective action. Demonstration of acceptable corrective action (i.e., an updated SOP including monthly recounts of a positive plate by one analyst and comparison counts between analysts) was received by email on July 30, 2015, with implementation of the practice on August 3, 2015. No further response is necessary for this Finding. P. Finding: Heat indicating autoclave tape is not used on all materials with each sterilizing cycle. Requirement: Use heat -indicating tape to identify supplies and materials that have been sterilized. Ref: Standard Methods 9020 B-2005. (4) (h). Q. Finding: The pH of the stock phosphate buffer solution is not documented. This is considered pertinent information to show that the pH is 7.2 ± 0.5 s.u. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15 NCAC 2H .0805 (a) (7) (A). Requirement: Stock phosphate buffer solution - Dissolve 34.0 g potassium dihydrogen phosphate (KH2PO4) in 500 mL reagent grade water, adjust to pH 7.2 ± 0.5 with 1N NaOH and dilute to 1 L with reagent grade water. Ref: Standard Methods, 9050 C-2006. (1) (a) (1). Page 14 #65 TZ Osborne WRF Laboratory Nitroaen. Ammonia — Standard Methods, 4500 NH3 D-1997 Comment: The SOP only lists one acceptance criterion for the calibration blank (i.e., <_ 0.05 mg/Q. This is correct for the low range curve, but the calibration blank acceptance criterion for the high range curve can be up to <_ 0.5 mg/L. Comment: Calibration time for the low range and high range curve was being documented, but the analysis time of the low range samples was not. This is considered pertinent information to show that analysis is conducted after calibration. The North Carolina Administrative Code, 15 NCAC 2H .0805 (a) (7) (A) states: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Demonstration of acceptable corrective action (i.e., a benchsheet including a space for low range sample analysis time) was received by email on July 30, 2015, with implementation on August 3, 2015. No further response is necessary for this Finding. R. Finding: Samples are not being checked for total residual chlorine (TRC) and treated, if necessary, at time of collection. Requirement: Residual Chlorine reacts with ammonia; remove by sample pretreatment. If a sample is likely to contain residual chlorine, immediately upon collection, treat with dechlorinating agent as in 4500-NH3.B.3d. Ref: Standard Methods, 4500 NH3 A-1997. (2). Comment: Samples are checked and treated for TRC immediately prior to sample analysis. Once the Finding is corrected, the laboratory is reminded that dechlorinating agents used at the time of sampling must be documented to have been effective upon receipt in the laboratory. S. Finding: Samples are not allowed to come to room temperature before analysis. Requirement: Place 100 mL sample in 150 mL beaker and follow procedure in ¶ 4b above. Ref: Standard Methods, 4500 NH3 D-1997. (4) (e). Requirement: Maintain the same stirring rate and a temperature of about 250C through calibration and testing procedures. Ref: Standard Methods, 4500 NH3 D-1997. (4) (b). T. Finding: The true value of the LFM is not documented. Requirement: Each laboratory shall develop and maintain a document outlining the analytical quality control practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A NCAC 2H .0805 (a) (7). Comment: The true value of quality control samples and percent recovery obtained must be documented on the benchsheet and evaluated against established acceptance criteria to demonstrate that the analyst was aware of any out -of -control situation. U. Finding: Preservation acid is not being added to the method blank and LFB. Requirement: A reagent blank (method blank) consists of reagent water (see Section 1080) and all reagents (including preservative) that normally are in contact with a sample during the entire analytical procedure. Ref: Standard Methods, 1020 B-2011. (5). Requirement: A laboratory -fortified blank [laboratory control standard (LCS)] is a reagent water sample (with associated preservatives) to which a known concentration of the analytes(s) of interest has been added. Ref: Standard Methods, 1020 B-2011. (6). Page 15 #65 TZ Osborne WRF Laboratory V. Finding: The LFM is only performed on the TZO Primary (process control) sample. Requirement: To prepare an LFM, add a known concentration of analytes (ideally from a second source) to a randomly selected routine sample without decreasing its volume by more than 5%. Ref: Standard Methods 4020 B- 2009, Rev. 2011. (2) (g). Comment: The LFM needs to be rotated through all types of samples, including industry, to demonstrate there is no bias contributed to the analytical results by any matrix encountered for routine samples. W. Finding: The volume of NaOH (ISA solution) used to raise the pH > 11 is not documented. This is considered pertinent information. Requirement: Add a sufficient volume of 1 ON NaOH solution (1 mL usually is sufficient) to raise pH above 11. If the presence of silver or mercury is possible, use NaOH/EDTA solution in place of NaOH solution. If it is necessary to add more than 1 mL of either NaOH or NaOH/EDTA solution, note volume used, because it is required for subsequent calculations. Ref: Standard Methods 4500 NH3 D-1997. (4) (b). Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15 NCAC 2H .0805 (a) (7) (A). Metals — EPA Method 200.8, Rev. 5.4, 1994 Comment: When the use of Kinetic Energy Discrimination (KED) is implemented for chloride and other halide ion interference reduction, perform a study to demonstrate that it does not cause magnification of other interferences (which may happen in some effluents) and to demonstrate that sensitivity is not reduced. Include instruction for KED use and a validation study in the SOP. This validation study may either be attached to the SOP or kept on file. The study must consist of analyzing a minimum of 10 samples and standards side -by -side with and without the KED. It is recommended that the KED be used for all analyses to reduce the number of curves that would have to be run. If not used routinely, you must analyze a curve that reflects this addition, when in use. This study is a new requirement by EPA Region 4. Recommendation: It is recommended that the sequential run log be printed daily. Recommendation: It is recommended that the laboratory periodically prepare and analyze a field blank that is placed in a sample container in the laboratory and treated as a sample in all respects, including shipment to the sampling site, exposure to the sampling site conditions, storage, preservation, and all analytical procedures. Recommendation: It is recommended that instructions for saving electronic batch files be added to the SOP. X. Finding: The laboratory was not documenting sample pH and ID, acid lot numbers, or rotating cell temperature checks. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15 NCAC 2H .0805 (a) (7) (A). Page 16 #65 TZ Osborne WRF Laboratory Comment: The sample pH and ID, acid lot numbers, and rotating cell temperature checks can be added to the newly developed digestion log. Y. Finding: The laboratory is only spiking the effluent sample. Requirement: Unless otherwise specified by the data user, laboratory or program, the following laboratory fortified matrix (LFM) procedure (Section 9.4.2) is required. Ref: Method 200.8, Rev. 5.4 (1994), Section 9.4.1. Requirement: Over time, samples from all routine sample sources should be fortified. Ref: EPA Method 200.8, Rev. 5.4 (1994), Section 9.4.2. Z. Finding: Raw data printed from ICP/MS analyses is not initialed by the analyst. Requirement: All laboratories must use printed bench worksheets that include a space to enter the signature or initials of the analyst, date of analyses, sample identification, volume of samples analyzed, value from the measurement system, factor and final value to be reported and each item must be recorded each time samples are analyzed. The date and time BOD and coliform samples are removed from the incubator must be included on the laboratory worksheet. Ref: 15A NCAC 2H .0805 (a) (7) (H). Comment: The analyst must initial the raw data printed from ICP/MS analyses to verify the data produced. AA. Finding: The analyst does not consistently document sample dilution factors. Requirement: All laboratories must use printed bench worksheets that include a space to enter the signature or initials of the analyst, date of analyses, sample identification, volume of samples analyzed, value from the measurement system, factor and final value to be reported and each item must be recorded each time samples are analyzed. The date and time BOD and coliform samples are removed from the incubator must be included on the laboratory worksheet. Ref: 15A NCAC 2H .0805 (a) (7) (H). BB. Finding: The temperature of the hot block during acid digestion is not documented. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15 NCAC 2H .0805 (a) (7) (A). Requirement: To the beaker add 4 mL of (1+1) HNO and 10 mL of (1+4) HCl. Cover 3 the lip of the beaker with a watch glass. Place the beaker on a hot plate for reflux extraction of the analytes. The hot plate should be located in a fume hood and previously adjusted to provide a reflux temperature of approximately 950C. (See the following note.) Note: For proper heating adjust the temperature control of the hot plate such that an uncovered Griffin beaker containing 50 mL of water placed in the center of the hot plate can be maintained at a temperature approximately but no higher than 85°C. (Once the beaker is covered with a watch glass the temperature of the water will rise to approximately 200.8-26 95°C.) Also, a block digester capable of maintaining a temperature of 95°C and equipped with 250 mL constricted volumetric digestion tubes may be substituted for the hot plate and conical beakers in the extraction step. Ref: EPA Method 200.8, Rev. 5.4 (1994), Section 11.3.3. Page 17 #65 TZ Osborne WRF Laboratory CC. Finding: The laboratory is not applying an acceptance criterion (i.e., :550% of the reporting limit) to the blanks. Requirement: For analyses requiring a calibration curve, the concentration of reagent, method and calibration blanks must not exceed 50% of the reporting limit or as otherwise specified by the reference method. Ref: NC WW/GW LC Policy. Comment: The analyst noted that he does monitor the blank values. DD. Finding: The laboratory has not completed all components of the method required Initial Demonstration of Performance. Requirement: The initial demonstration of performance is used to characterize instrument performance (determination of linear calibration ranges and analysis of quality control samples) and laboratory performance (determination of method detection limits) prior to analyses conducted by this method. Comment: The instrument is currently shut down while the air handling system in that unit is being upgraded. An Initial Demonstration of Performance will be required prior to analyzing samples. The NC WW/GW LC linear calibration/dynamic range policy is attached to the end of this report. EE. Finding: The laboratory lacks traceability for the preparation of calibration standards, quality control standards, consumables and reagents. Requirement: All chemicals, reagents, standards and consumables used by the laboratory must have the following information documented: Date received, Date Opened (in use), Vendor, Lot Number, and Expiration Date (where specified). A system (e.g., traceable identifiers) must be in place that links standard/reagent preparation information to analytical batches in which the solutions are used. Documentation of solution preparation must include the analyst's initials, date of preparation, the volume or weight of standard(s) used, the solvent and final volume of the solution. This information as well as the vendor and/or manufacturer, lot number, and expiration date must be retained for primary standards, chemicals, reagents, and materials used for a period of five years. Consumable materials such as pH buffers, lots of pre -made standards and/or media, solids and bacteria filters, etc. are included in this requirement. Ref: NC WW/GW LC Policy. FF. Finding: When checking the sample pH to verify pH <2, the value, time, and date are not being documented. This is considered pertinent information. Requirement: All analytical data pertinent to each certified analysis must be filed in an orderly manner so as to be readily available for inspection upon request. Ref: 15 NCAC 2H .0805 (a) (7) (A). Comment: This information is needed to verify that 24 hours have passed between time of proper preservation and analysis. GG. Finding: The laboratory reagent blank (LRB) and LFB are not filtered through the digifilter before analysis. Requirement: Laboratory Fortified Blank (LFB) - An aliquot of LRB to which known quantities of the method analytes are added in the laboratory. The LFB is analyzed exactly like a sample, and its purpose is to determine whether the methodology is in control and whether the laboratory is capable of making accurate and precise measurements. Ref: EPA Method 200.8, Rev. 5.4 (1994), Section 3.8. Page 18 #65 TZ Osborne WRF Laboratory Requirement: Laboratory Reagent Blank (LRB) - An aliquot of reagent water or other blank matrices that are treated exactly as a sample including exposure to all glassware, equipment, solvents reagents and internal standards that are used with other samples. Ref: EPA Method 200.8, Rev. 5.4 (1994), Section 3.10. HH. Finding: The laboratory is not performing the method defined tune procedure. Requirement: Precalibration routine - The following precalibration routine must be completed prior to calibrating the instrument until such time it can be documented with periodic performance data that the instrument meets the criteria listed below without daily tuning. 10.2.1 Initiate proper operating configuration of instrument and data system. Allow a period of not less than 30 minutes for the instrument to warm up. During this process conduct mass calibration and resolution checks using the tuning solution. Resolution at low mass is indicated by magnesium isotopes 24, 25, and 26. Resolution at high mass is indicated by lead isotopes 206, 207, and 208. For good performance adjust spectrometer resolution to produce a peak width of approximately 0.75 amu at 5% peak height. Adjust mass calibration if it has shifted by more than 0.1 amu from unit mass. 10.2.2 Instrument stability must be demonstrated by running the tuning solution (Section 7.7) a minimum of five times with resulting relative standard deviations of absolute signals for all analytes of less than 5%. Ref: EPA Method 200.8, Rev. 5.4 (1994), Section 10.2. Comment: The laboratory was conducting the Perkin Elmer (PE) Performance Check; however, this performance check was not always repeated if a target failure occurred. II. Finding: The recovery and precision for quality control elements is not consistently documented. Requirement: Each laboratory shall develop and maintain a document outlining the analytical quality control practices used for the parameters included in their certification. Supporting records shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A NCAC 21-1.0805 (a) (7). Comment: The true value of quality control samples and percent recovery obtained must be documented on the benchsheet and evaluated against established acceptance criteria to demonstrate that the analyst was aware of any out -of -control situation. Chlorine, Total Residual — Standard Methods, 4500 Cl G- 2000 Comment: There does not seem to be an official copy of the benchsheet in use. Reviewing past data showed that the date of the calibration curve verification was not consistent. One of the most recent benchsheets listed a calibration curve verification date of 7/29/2011. Comment: The TRC benchsheet did not contain chemical, reagent, or consumable traceability information, instrument identification, or meter calibration verification time. The Approved Procedure for the Analysis of Total Residual Chlorine document states: The following must be documented in indelible ink whenever sample analysis is performed: Meter calibration and meter calibration time(s), Traceability for chemicals, reagents, standards, and consumables, and Instrument identification. Demonstration of acceptable corrective action (i.e., an updated benchsheet with spaces to enter traceability information, instrument identification and meter calibration verification time) was received by email on July 30, 2015 and implemented on August 3, 2015. No further response is necessary for this finding. JJ. Finding: The annual calibration curve verification does not bracket the concentration of the annual Proficiency Testing (PT) samples. Page 19 #65 TZ Osborne WRF Laboratory Requirement: For analytical procedures requiring analysis of a series of standards, the concentrations of those standards must bracket the concentration of the samples analyzed. One of the standards must have a concentration equal to the laboratory's lower reporting concentration for the parameter involved. Ref: 15A NCAC 02H .0805 (a) (7) (1). Recommendation: The concentrations used to verify the most recent curve are 14, 50, 100, 150 and 200 pg/L. The assigned value of the 2015 PT sample was 224 pg/L. It is recommended that the laboratory verify the internal curve with a standard concentration between the lowest standard and 50 pg/L and up to 400 pg/L. This will verify the analytical range used to measure Proficiency Testing (PT) samples as well as environmental samples. KK. Finding: Results were not qualified for a sample analyzed outside of hold time. Requirement: At any time a laboratory receives samples which do not meet sample collection, holding time, or preservation requirements, the laboratory must notify the sample collector or client and secure another sample if possible. If another sample cannot be secured, the original sample may be analyzed but the results reported must be qualified with the nature of the infraction(s) and the laboratory must notify the State Laboratory about the infraction(s). The notification must include a statement indicating corrective actions taken to prevent the problem for future samples. Ref: 15A NCAC 2H .0805 (a) (7) (N). Comment: On January 9, 2015, the North Buffalo effluent was collected at 8:50 am and analyzed at 9:15 am, which exceeded the holding time of 15 minutes. ILL. Finding: The laboratory is not analyzing a reagent blank with the standard curve verification nor the daily sample analyses. Requirement: Reagent Blank: A reagent blank (sometimes also referred to as a method blank) is only required when laboratory water is used to make quality control and/or calibration standards. If you are using a sealed standard (e.g., gel) for your daily check standard, a reagent blank would only be analyzed when preparing the annual 5-point calibration curve or 5 annual calibration curve verification standards. Ref: NC WW/GW LC Approved Procedure for the Analysis of Total Residual Chlorine. Comment: A reagent blank is made from the same laboratory water source used to make quality control and/or calibration standards with DPD. The concentration of reagent blanks must not exceed 50% of the reporting limit (i.e., the lowest calibration or calibration verification standard concentration), unless otherwise specified by the reference method, or corrective action must be taken. MM. Finding: The laboratory is filtering samples for TRC analysis. Requirement: Compensate for color and turbidity by using sample to zero photometer. Ref: Standard Methods, 4500 Cl G-2000. (1) (b). Comment: The laboratory is currently following the procedure as described in the Hach 10014 ULR method. Standard Methods, 4500 CI G- 2000, the method for which the laboratory is certified, does not allow for filtering of samples. However, if the laboratory chooses to update their certificate to cite Hach 10014 ULR, they would be required to continue filtering the sample. Page 20 #65 TZ Osborne WRF Laboratory PH — Standard Methods, 4500 H+ B-2000 Comment: Values were reported that exceed the method specified accuracy of 0.1 units. Standard Methods, 4500 H+ B-2000, (6) states in part: However, ± 0.1 pH unit represents the limit of accuracy under normal conditions, especially for measurement of water and poorly buffered solutions. For this reason, report pH values to the nearest 0.1 PH unit. Notification of acceptable corrective action (i.e., a statement that pH results would be reported to 0.1 units) was received by e-mail on July 30, 2015 with a stated implementation date of August 3, 2015. No further response is necessary for this Finding. Temperature — Standard Methods, 2550 B-2000 Comment: The benchsheet did not document the time of sample analysis. The North Carolina Administrative Code, 15 NCAC 2H .0805 (g) (1) states: Data pertinent to each analysis must be maintained for five years. Certified Data must consist of date collected, time collected, sample site, sample collector, and sample analysis time. The field bench sheets must provide a space for the signature or initials of the analyst, and proper units of measure for all analyses. Demonstration of acceptable corrective action (i.e., an updated benchsheet with a space to enter analysis time) was received by email on July 30, 2015 and implemented on August 3, 2015. No further response is necessary for this Finding. Dissolved Oxycien — Standard Methods, 4500 O G-2001 Comment: The benchsheet did not include units of measure for the DO meter calibration. This is considered pertinent information. The North Carolina Administrative Code, 15A NCAC 02H .0805 (g) (1) states: Data pertinent to each analysis must be maintained for five years. Certified Data must consist of date collected, time collected, sample site, sample collector, and sample analysis time. The field bench sheets must provide a space for the signature or initials of the analyst, and proper units of measure for all analyses. Demonstration of acceptable corrective action (i.e., an updated benchsheet with units of measure for the DO meter calibration) was received by email on July 30, 2015 and implemented on August 3, 2015. No further response is necessary for this Finding. COD —Standard Methods, 5220 D-1997 Comment: The laboratory does not analyze COD samples except for the yearly PT sample. The laboratory had indicated that they may request to drop the parameter from their certificate, but it has not been requested at this time. The SOP for COD did not contain any mention of quality control elements including: the reporting limit, the calibration curve, LCS, duplicates, blanks, acceptance criteria, or matrix spiking. If the laboratory decides to maintain certification for COD the SOP will have to be revised and submitted. Please respond in writing as to whether COD certification will be dropped or maintained. NN. Finding: No QC standards are analyzed. Requirement: The following are requirements for certification and must be included in each certified laboratory's quality control program. (D) Any quality control procedures required by a particular approved method shall be considered as required for certification for that analysis. (E) All quality control requirements in these Rules as set forth by the State Laboratory. Ref: 15A NCAC 2H .0805 (a) (7) (D) and (E). Comment: The only data available to check was the analysis of the PT. Page 21 #65 TZ Osborne WRF Laboratory 00. Finding: The laboratory is only documenting the final calculated result of the PT sample and no other supporting data to allow historical reconstruction of the final result. Requirement: All laboratories must use printed bench worksheets that include a space to enter the signature or initials of the analyst, date of analyses, sample identification, volume of samples analyzed, value from the measurement system, factor and final value to be reported and each item must be recorded each time samples are analyzed. The date and time BOD and coliform samples are removed from the incubator must be included on the laboratory worksheet. Ref: 15A NCAC 2H .0805 (a) (7) (H). PP. Finding: The PT sample was analyzed with a factory curve that has not been verified. Requirement: For colorimetric analyses, a series of five standards for a curve prepared annually or three standards for curves established each day or standards as set forth in the analytical procedure must be analyzed to establish a standard curve. The curve must be updated as set forth in the standard procedures, each time the slope changes more than ten percent at mid -range, each time a new stock standard is prepared, or at least every twelve months. Ref: 15A NCAC 2H .0805 (7) (1). IV. PAPER TRAIL INVESTIGATION: A paper trail was not conducted during this inspection. On April 23, 2015, Mike Mickey of the Winston- Salem Regional Office performed a Compliance Evaluation Inspection of the NPDES wastewater discharge from the T.Z. Osborne facility. Part of this inspection included reviewing Discharge Monitoring Reports (DMRs) for the period March, 2014 through February, 2015. V. CONCLUSIONS: Correcting the above -cited Findings and implementing the recommendations will help this lab to produce quality data and meet certification requirements. The inspector would like to thank the staff for its assistance during the inspection. Please respond to all Findings and include an implementation date for each corrective action. Report prepared by: Beth Swanson Date: August 20, 2015 Report reviewed by: Nick Jones Date: September 2, 2015