HomeMy WebLinkAbout#5345_2021_0217_TLH_FINAL On-Site Inspection Report LABORATORY NAME: Town of Winton WWTP WATER QUALITY PERMIT #: WQ0001602 ADDRESS: 531 US 158 W Winton, NC 27986 CERTIFICATE #: 5345 DATE OF INSPECTION: February 17, 2021 TYPE OF INSPECTION: Field Municipal Maintenance AUDITOR: Tom Halvosa LOCAL PERSON(S) CONTACTED: Tyler Hughes, Chris Blackburn I. INTRODUCTION: This laboratory was inspected by a representative of the North Carolina Wastewater/Groundwater Laboratory Certification (NC WW/GW LC) program to verify its compliance with the requirements of 15A NCAC 02H .0800 for the analysis of compliance monitoring samples. II. GENERAL COMMENTS: The facility is neat and well organized and has all the equipment necessary to perform the analyses. The staff were forthcoming and responded well to suggestions from the auditor. All required Proficiency Testing (PT) Samples for the 2021 Calendar Year have not yet been analyzed. The laboratory is reminded that results must be received by this office directly from the vendor by September 30, 2021. Contracted analyses are performed by Environment 1, Inc. (Certification #10). Approved Procedure documents for the analysis of the facility’s currently certified Field Parameters were provided at the time of the inspection. The laboratory submitted their Standard Operating Procedure (SOP) documents in advance of the inspection. These documents were reviewed and deemed acceptable. The laboratory is reminded that any time changes are made to laboratory procedures, the laboratory must update the QA/SOP document(s) and inform relevant staff. Any changes made in response to the pre-audit review or to Findings, Recommendations or Comments listed in this report must be incorporated to insure the method is being performed as stated, references to methods are accurate, and the QA and/or SOP document(s) is in agreement with each approved practice, test, analysis, measurement, monitoring procedure or regulatory requirement being used in the laboratory. In some Page 2 #5345 Town of Winton WWTP instances, the laboratory may need to create an SOP to document how new functions or policies will be implemented. The laboratory is also reminded that SOPs are required to be reviewed at least every two years and are intended to describe procedures exactly as they are to be performed. Use of the word “should” is not appropriate when describing requirements (e.g., Quality Control (QC) frequency, acceptance criteria, etc.). Evaluate all SOPs for the proper use of the word “should”. III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS: Documentation A. Finding: The laboratory benchsheet for Total Residual Chlorine (TRC) and pH are lacking required documentation: Units of measure. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the proper units of measure. Each item shall be recorded each time samples are analyzed. Analyses shall conform to methodologies found in Subparagraph (a)(1) of this Rule. Ref: 15A NCAC 02H .0805 (g) (2) (L). Comment: The units of measure for TRC must be documented and reported as specified in the permit. B. Finding: Error corrections are not properly performed. Requirement: All documentation errors shall be corrected by drawing a single line through the error so that the original entry remains legible. Entries shall not be obliterated by erasures or markings. Wite-Out®, correction tape, or similar products designed to obliterate documentation are not to be used; instead the correction shall be written adjacent to the error. The correction shall be initialed by the responsible individual and the date of change documented. Ref: 15A NCAC 02H .0805 (g) (1). Comment: The laboratory is using Wite-Out® to perform error corrections. The laboratory was instructed not use Wite-Out® and to perform error corrections by striking out error with one line then dating and initialing. IV. PAPER TRAIL INVESTIGATION: The paper trail consisted of comparing original records (e.g., laboratory benchsheets, logbooks, etc.) and contract laboratory reports to DMRs submitted to the North Carolina Division of Water Resources. Data were reviewed for Town of Winton WWTP (Water Quality permit # WQ0001602) for January, June and December 2020. No transcription errors were observed. The facility appears to be doing a good job of accurately transcribing data. V. CONCLUSIONS: Correcting the above-cited Findings will help this laboratory to produce quality data and meet Certification requirements. The inspector would like to thank the staff for their assistance during the inspection and data review process. Please respond to all Findings and include supporting documentation, implementation dates and steps taken to prevent recurrence for each correcti ve action. Report prepared by: Tom Halvosa Date: February 23, 2021 Page 3 #5345 Town of Winton WWTP Report reviewed by: Tonja Springer Date: February 24, 2021