Loading...
HomeMy WebLinkAbout#402_2022_0321_MC_FINAL NC Department of Environmental Quality | Division of Water Resources | Laboratory Certification Branch 4405 Reedy Creek Road | 1623 Mail Service Center | Raleigh, North Carolina 27699-1623 919-733-3908 July 1, 2022 402 Mr. Kyle Collins Waypoint Analytical - Charlotte P.O. Box 240543 Charlotte, NC 28224-0543 Subject: North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC WW/GW LCB) Maintenance Inspection Dear Mr. Collins: Enclosed is a report for the inspection performed on March 21, 2022 by Michael Cumbus, Beth Swanson, Anna Ostendorff, Todd Crawford, Tom Halvosa, Jill Puff, Tonja Springer and Jason Smith. I apologize for the delay in getting this report to you. Where Finding(s) are cited in this report, a response is required. Within thirty days, please supply this office with a written item for item description of how these Finding(s) were corrected. Please describe the steps taken to prevent recurrence and include an implementation date for each corrective action. If the Finding(s) cited in the enclosed report are not corrected, enforcement actions may be recommended. For Certification maintenance, your laboratory must continue to carry out the requirements set forth in 15A NCAC 02H .0800. A copy of the laboratory’s Certified Parameter List at the time of the audit is attached. This list will reflect any changes made during the audit. Copies of the checklists completed during the inspection may be requested from this office. Thank you for your cooperation during the inspection. If you have questions or need additional information, please contact me at (919) 733- 3908 Ext. 259. Sincerely, Beth Swanson Technical Assistance & Compliance Specialist Division of Water Resources Attachment cc: Todd Crawford, Michael Cumbus, #402 On-Site Inspection Report LABORATORY NAME: Waypoint Analytical - Charlotte ADDRESS: 449 Springbrook Road Charlotte, NC 28217 CERTIFICATE #: 402 DATE OF INSPECTION: March 21, 2022 – March 23, 2022 TYPE OF INSPECTION: Commercial Maintenance AUDITOR(S): Michael Cumbus, Beth Swanson, Anna Ostendorff, Todd Crawford, Tom Halvosa, Jill Puff, Tonja Springer, Jason Smith LOCAL PERSON(S) CONTACTED: Kyle Collins, McLayne Reese, Jeff Vogel, Jack Cattron, Jenelle Kamienski, Kelly LaTour, Tyrese Williams, Elizabeth Vrapi, Joseph Blydenburgh, Andrew Guptill, Taylor Lowe, Emily McComiskey, Paula Blakeborough, Sarah Locklear, Zoe Vigilant, Christopher Garcia, Stacy Ortiz, Sharol Wiley, Caroline Baratta, David Morris I. INTRODUCTION: This laboratory was inspected by representatives of the North Carolina Wastewater/Groundwater Laboratory Certification Branch (NC WW/GW LCB) to verify its compliance with the requirements of 15A NCAC 02H .0800 for the analysis of compliance monitoring samples. II. GENERAL COMMENTS: The facility is neat, well maintained and has all the equipment necessary to perform the analyses. Staff were forthcoming and responded well to suggestions from the auditors. The laboratory is highly organized, and staff were able to rapidly supply additional information requests. The laboratory has undergone significant personnel and organizational changes within the last 24 months and has recently transitioned to a new Laboratory Information Management System (LIMS). All required Proficiency Testing (PT) Samples for the 2022 PT Calendar Year have not yet been analyzed. The laboratory is reminded that results must be received by this office directly from the vendor by September 30, 2022. The laboratory is reminded that any time changes are made to laboratory procedures, QA/SOP document(s) must be updated and relevant staff retrained. Staff must acknowledge that they have read and understand the changes as part of the documented training program. The same requirements apply when changes are made in response to Findings, Recommendations or Comments listed in this report, to ensure the methods are being performed as stated, references to methods are accurate, and the QA and/or SOP document(s) is in agreement with each approved practice, test, analysis, measurement, monitoring procedure or regulatory requirement being used in the laboratory. In some instances, the laboratory may need to create an SOP to document how new functions or policies will be implemented. Revisions to the SOPs, based on the Findings, Comments and Recommendations within this report must be submitted to this office by July 1, 2023. Page 2 #402 Waypoint Analytical - Charlotte The laboratory is also reminded that SOPs are required to be reviewed at least every two years and are intended to describe procedures exactly as they are to be performed. Use of the word “should” is not appropriate when describing requirements (e.g., Quality Control (QC) frequency, acceptance criteria, etc.). Evaluate all SOPs for the proper use of the word “should”. Laboratory Fortified Matrix (LFM) and Laboratory Fortified Matrix Duplicate (LFMD) are also known as Matrix Spike (MS) and Matrix Spike Duplicate (MSD) and may be used interchangeably in this report. Contracted analyses are performed by Waypoint Analytical LLC (Certification #415), Waypoint Analytical Mississippi (Certification #694), Analytical Environmental Services, Inc. (Certification #562), Pace West Columbia (Certification #329) and Environment 1 (Certification # 10). Approved Procedure documents for the analysis of the facility’s currently certified Field Parameters were provided at the time of the inspection. III. FINDINGS, REQUIREMENTS, COMMENTS AND RECOMMENDATIONS: Documentation A. Finding: Error corrections are not always properly performed. Requirement: All documentation errors shall be corrected by drawing a single line through the error so that the original entry remains legible. Entries shall not be obliterated by erasures or markings. Wite-Out®, correction tape, or similar products designed to obliterate documentation shall not to be used; instead, the correction shall be written adjacent to the error. The correction shall be initialed by the responsible individual and the date of change documented. Ref: 15A NCAC 02H .0805 (a) (7) (E). Comment: Multiple instances were noted throughout the laboratory where error corrections were lacking the date and/or the initials of the person making the change. B. Finding: SOPs have not been updated for all of the methods included on the laboratory’s Certified Parameters Listing (CPL). Requirement: Each laboratory shall have a documented analytical quality assurance and quality control program. Each laboratory shall have a copy of each approved test, analysis, measurement, or monitoring procedure being used in the laboratory. Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. The quality assurance, quality control, and Standard Operating Procedure documents shall be available for inspection by the State Laboratory. Ref: 15A NCAC 02H .0805 (a) (7). Comment: Many instances were noted where the effective dates of the SOPs on file exceeded the two year requirement. The laboratory submitted drafts of upcoming revisions for many of the SOPs. C. Finding: The laboratory does not document SOP revisions. Page 3 #402 Waypoint Analytical - Charlotte Requirement: Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. Ref: 15A NCAC 02H .0805 (a) (7). Comment: There is not a summary sheet detailing dates and specific revisions completed. D. Finding: Sample preservation verification is not always clearly documented in the laboratory records. Requirement: Sample preservation shall be verified and documented. If a laboratory receives a sample subject to G.S. 143-215.1 and 143-215.63 that does not meet sample collection, holding time, or preservation requirements, the laboratory shall document the incident, notify the sample collector or client, and secure another sample that meets the regulatory requirements, if possible. If another viable sample cannot be secured, the original sample may be analyzed but the results reported shall be qualified with the nature of the sample collection, holding time, or preservation infractions and the laboratory shall notify the State Laboratory of the infractions. The notification shall include a statement indicating corrective action taken to prevent future infractions. Ref: 15A NCAC 02H .0805 (a) (7) (M). Comment: Documentation of sample preservation for each batch of client samples consisted of a single checkmark on the chain of custody form. Sample preservation was documented on the sample bottles, but this does not create a permanent (or lasting) record once the bottles are discarded. E. Finding: Units of measure are not consistently documented on the laboratory benchsheets. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the proper units of measure. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (xii). Comment: The laboratory benchsheet for Alkalinity was lacking the units of measure for the “initial pH”, “pH prior to low level 0.3 unit titr.” and “results” columns. Comment: The laboratory benchsheet for Cyanide by SM 4500 CN E-2011 was lacking the units of measure for the “pH >12” column. Comment: The columns on the Inorganic Phenols benchsheet by EPA 420.1, labeled “pH 4” and “pH 10” are lacking units of measure. Comment: The laboratory benchsheet for Hexavalent Chromium by SW-846 3060A/7196A was lacking units of measure for pH. Comment: The Quantitative Analysis Report benchsheets for Hexavalent Chromium (all methods) do not have the correct units on the "Concentration" columns. The value in the column is only "ug". On most of the results, the analyst then divides the value by 100 (mL of sample reacted) and documents that value, which is correctly ug/mL at that point (i.e., mg/L as reported). Comment: The prep benchsheet for Total Phosphorus by SM 4500 P F-2011, has a column labeled "Sample Amt (mL or g)" and nothing was circled or indicated which unit is correct. Page 4 #402 Waypoint Analytical - Charlotte Comment: The laboratory benchsheet for pH is lacking units of measure for Temperature (°C). Recommendation: This list may not be exhaustive. It is recommended that the laboratory review all benchsheets to ensure that they include the proper units of measure. F. Finding: The COD benchsheet and the laboratory calibration log for the pH meter used in BOD/CBOD analyses are lacking required documentation: instrument identification. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the instrument identification. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (iii). Comment: The laboratory uses two COD reactors for an analytical batch. It is not clear which samples were digested in each reactor. G. Finding: The laboratory benchsheet for pH and Temperature is lacking required documentation: the method or Standard Operating Procedure reference; the instrument identification and the proper units of measure. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the method or Standard Operating Procedure; the instrument identification; the proper units of measure. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (i) (iii) (xii). Comment: This Finding applies to Field pH and Temperature analyses. Comment: Temperature and pH results are documented in a notebook in the field and transferred to the Chain of Custody. The notebook is missing units of measure for Temperature, and method reference and instrument identification for both parameters. There is a log that includes all the required information, but it is no longer being used. H. Finding: The laboratory benchsheet is lacking required documentation: the signature or initials of the analyst. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the signature or initials of the analyst. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (v). Comment: This Finding applies to the laboratory Field Benchsheet used to document temperature and pH analyses. I. Finding: The analysis end time of Paint Filter Liquids is not documented. Requirement: Supporting records shall be maintained as evidence that these practices are being effectively carried out. Ref: 15A NCAC 02H .0805 (a) (7). Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the time of sample analyses (when required to document a required holding time or when time-critical steps are imposed by the method, a federal regulation, or this Rule). Each item shall be recorded Page 5 #402 Waypoint Analytical - Charlotte each time that samples are analyzed. The date and time that samples are placed into and removed from ovens, water baths, incubators and other equipment shall be documented if a time limit is required by the method. Ref: 15A NCAC 02H .0805 (a) (7) (F) (viii). Requirement: Allow sample to drain for 5 min into the graduated cylinder. Ref: EPA SW- 846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 9095 B. Rev. 2, November 2004, Section 7.5. J. Finding: The laboratory is not documenting all traceability information for purchased materials, reagents and standards. Requirement: 15A NCAC 02H .0805 (a) (7) (K) and (g) (7) requires laboratories to have a documented system of traceability for the purchase, preparation, and use of all chemicals, reagents, standards, and consumables. That system must include documentation of the following information: Date received, Date Opened (in use), Vendor, Lot Number, and Expiration Date (where specified). A system (e.g., traceable identifiers) must be in place that links standard/reagent preparation information to analytical batches in which the solutions are used. Documentation of solution preparation must include the analyst’s initials, date of preparation, the volume or weight of standard(s) used, the solvent and final volume of the solution. This information as well as the vendor and/or manufacturer, lot number, and expiration date must be retained for primary standards, chemicals, reagents, and materials used for a period of five years. Consumable materials such as pH buffers, lots of pre-made standards and/or media, solids and bacteria filters, etc. are included in this requirement. Ref: NC WW/GW LCB Policy. Comment: Traceability information for the filters used for Color, Oil and Grease, Total Suspended Residue and Total Dissolved Residue is not documented. K. Finding: Some benchsheets and/or report summaries reference unapproved methods and/or do not fully reference the approved methods. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the method or Standard Operating Procedure. Each item shall be recorded each time samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (i). Requirement: Analytical methods, sample preservation, sample containers, and sample holding times shall conform to the requirements found in: 40 CFR Part 136. Ref: 15A NCAC 02H .0805 (a) (1) (A). Comment: Many laboratory benchsheets do not reference the method year. This may be due to space constraints in transferring data from the LIMS. Comment: The laboratory benchsheets for BOD/CBOD cite the revision year as 2011 instead of 2015. Comment: One Seal autoanalyzer printout (i.e., AA090121A) reviewed references SM 4500 NO3- I, which is not an approved method. Comment: The client reports submitted in advance of the inspection did not always transfer the correct method year from the LIMS. Client report 21-175-0010 dated June 24, 2021 omitted the method year reference for Ammonia Nitrogen (SM4500 NH3-G 2011) and Total Phosphorus (SM4500 P-F 2011). Client report 21-302-0033 dated November 10, 2021 Page 6 #402 Waypoint Analytical - Charlotte omitted the method year for Ammonia Nitrogen (SM4500 NH3-G 2011) and listed the method year for Total Dissolved Residue (SM2540 C-2015) as 2011. L. Finding: The laboratory is not documenting the preparation of the combined reagent. Requirement: 15A NCAC 02H .0805 (a) (7) (K) and (g) (7) requires laboratories to have a documented system of traceability for the purchase, preparation, and use of all chemicals, reagents, standards, and consumables. That system must include documentation of the following information: Date received, Date Opened (in use), Vendor, Lot Number, and Expiration Date (where specified). A system (e.g., traceable identifiers) must be in place that links standard/reagent preparation information to analytical batches in which the solutions are used. Documentation of solution preparation must include the analyst’s initials, date of preparation, the volume or weight of standard(s) used, the solvent and final volume of the solution. This information as well as the vendor and/or manufacturer, lot number, and expiration date must be retained for primary standards, chemicals, reagents, and materials used for a period of five years. Consumable materials such as pH buffers, lots of pre-made standards and/or media, solids and bacteria filters, etc. are included in this requirement. Ref: NC WW/GW LCB Traceability Documentation Requirements for Chemicals, Reagents, Standards and Consumables Policy. Comment: All components of the combined reagent are documented, but the separate preparation time and directions for preparing (i.e., volume of each reagent used) the combined reagent are not documented. The combined reagent has a very short expiration period (4 hours) so the preparation time is especially important. Comment: This Finding applies to Total and Ortho-Phosphorus by SM 4500 P E-2011 (Aqueous) and Total Phosphorus by SM 4500 P F-2011 (Aqueous) Equipment M. Finding: The analytical balance level indicator bubble was outside the circle of accepted range. Requirement: Always check that the air bubble is in the center of the level indicator before each weighing. Use the leveling feet to make adjustments. Once leveled, always perform an internal adjustment. The exact procedure is described in the operating instructions for the balance. Ref: Mettler Toledo, Weighing the Right Way, January 2015, Section 3. Comment: The air bubble of the analytical balances in the wet chemistry weighing room and the semivolatiles analysis room were outside of the level indicator. General Laboratory Comment: All currently approved methods are not readily accessible to all analysts. The wet chemistry laboratory has online access to the 22nd Edition of Standard Methods as a reference, which does not include all methods or QC chapters currently approved and in use. Proficiency Testing Comment: The PT sample for Organochlorine Pesticides had several dilutions run. While running multiple dilutions on samples suspected or known to contain high levels of target analytes is considered acceptable practice, the laboratory QA Program and/or SOPs must include this information as part of the PT program and laboratory practice. Page 7 #402 Waypoint Analytical - Charlotte QA/QC Recommendation: The laboratory SOP for extraction by separatory funnel (SOP ORG-PRP007-02) mentions that the procedure may be used for Polychlorinated Biphenyls (PCBs) and Organochlorine Pesticides. However, the SOP does not provide any option for solvent exchange or the final volume of 10 mL. It is recommended that the laboratory eliminate the references to the above parameters or refer the reader to the appropriate analytical SOPs to complete the necessary steps. Recommendation: Client reports use the phrase “Method Quantitation Limit” but the SOPs do not define this term or explain how it is derived. It is recommended that the SOPs incorporate these terms into the next revision in order to improve data review. N. Finding: The laboratory QA Manual does not state the frequency of updates to all parameter QC items that use control charts. Requirement: Each laboratory shall have a documented analytical quality assurance and quality control program. Each laboratory shall have a copy of each approved test, analysis, measurement, or monitoring procedure being used in the laboratory. Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Ref: 15A NCAC 02H .0805 (a) (7). Comment: Currently, the QA manual only addresses the frequency of control chart updates for MS/MSD pairs. O. Finding: Thermometers in the Fecal Coliform and E. Coli incubators are not consistently verified against a Reference Temperature-Measuring Device at least every three months. Requirement: Digital temperature-measuring devices and temperature-measuring devices used in incubators shall be verified at the temperature of use every three months against a Reference Temperature-Measuring Device and their accuracy shall be corrected. Ref: 15A NCAC 02H .0805 (a) (7) (N) (iii). Comment: The thermometer in Bath #3 (T142) was verified on April 20, 2021, August 2, 2021, November 15, 2021, and February 22, 2022. The Micro (T140) thermometer was verified on March 19, 2021, June 30, 2021, October 13, 2021, and February 22, 2022. P. Finding: The laboratory is not correctly verifying the infrared (IR) temperature-measuring device against a Reference Temperature-Measuring Device. Requirement: Infrared temperature-measuring devices shall be verified every three months at three different temperatures over the temperature range of use against a Reference Temperature-Measuring Device and their accuracy shall be corrected. Each day of use, infrared temperature-measuring devices shall be verified against a non-Reference Temperature-Measuring Device that meets NIST specifications for accuracy. If the infrared temperature-measuring device does not agree within 0.5 degrees Celsius during the daily verification, the laboratory shall take corrective action. Ref: 15A NCAC 02H .0805 (a) (7) (N) (iv). Comment: The laboratory is verifying at two temperatures. Three temperatures are required. Q. Finding: The laboratory is not verifying the infrared (IR) temperature-measuring device each day of use. Page 8 #402 Waypoint Analytical - Charlotte Requirement: Infrared temperature-measuring devices shall be verified every three months at three different temperatures over the temperature range of use against a Reference Temperature-Measuring Device and their accuracy shall be corrected. Each day of use, infrared temperature-measuring devices shall be verified against a non-Reference Temperature-Measuring Device that meets NIST specifications for accuracy. If the infrared temperature-measuring device does not agree within 0.5 degrees Celsius during the daily verification, the laboratory shall take corrective action. Ref: 15A NCAC 02H .0805 (a) (7) (N) (iv). R. Finding: The State Laboratory is not notified when samples are received out of holding time or do not meet thermal and/or chemical preservation requirements. Cited Previously on July 21, 2016. Requirement: Sample preservation shall be verified and documented. If a laboratory receives a sample subject to G.S. 143-215.1 and 143-215.63 that does not meet sample collection, holding time, or preservation requirements, the laboratory shall document the incident, notify the sample collector or client, and secure another sample that meets the regulatory requirements, if possible. If another viable sample cannot be secured, the original sample may be analyzed but the results reported shall be qualified with the nature of the sample collection, holding time, or preservation infractions and the laboratory shall notify the State Laboratory of the infractions. The notification shall include a statement indicating corrective action taken to prevent future infractions. Ref: 15A NCAC 02H .0805 (a) (7) (M). S. Finding: The laboratory is not including both the original and manually integrated chromatograms in the data package when manual integrations are performed. Requirement: When manual integration is employed, the laboratory must clearly identify manually integrated compounds, document the reason the manual integration was performed, the date performed and who completed the work. A flag or qualifier code may suffice for simple manual integrations. In addition, a hardcopy printout of the data displaying the manual integration shall be included in the raw data package (i.e., both the original and manually integrated chromatograms, of similar scale, must be present in the data package). All information necessary for the historical reconstruction of data must be maintained by the lab. Additionally, the laboratory must employ a systematic data validation procedure to check manual integrations to assure integrations are technically sound and representative of the response. Ref: NC WW/GW LCB Manual Integration Policy. Comment: This Finding includes, but is not limited to, chromatograms generated for the following parameter methods: • Anions – EPA 300.0, Rev. 2.1, 1993 (Aqueous & Non-Aqueous) • Anions – SW-846 9056 A (Aqueous & Non-Aqueous) • Base Neutral/Acid, Organics – EPA 625.1, December 2016 (Aqueous) • Base Neutral/Acid, Organics – SW-846 8270 E (Aqueous & Non-Aqueous) • Extractable Petroleum Hydrocarbons – MADEP, December 2019, Rev. 2.1 (Aqueous & Non-Aqueous) • Pesticides, Organochlorine – EPA 608.3, December 2016 (Aqueous) • Pesticides, Organochlorine – SW-846 8081 B (Aqueous & Non-Aqueous) • Polychlorinated Biphenyls (PCBs) – EPA 608.3, December 2016 (Aqueous) • Polychlorinated Biphenyls (PCBs) – SW-846 8082 A (Aqueous, Non-Aqueous & Oil) • Purgeable, Organics – EPA 624.1, December 2016 (Aqueous) • Purgeable, Organics – SW-846 8260 D (Aqueous & Non-Aqueous) • TPH Diesel Range, Organics – SW-846 8015 C (Aqueous & Non-Aqueous) • TPH Gasoline Range, Organics – SW-846 8015 C (Aqueous & Non-Aqueous) Page 9 #402 Waypoint Analytical - Charlotte • Volatile Petroleum Hydrocarbons – MADEP, Feb 2018, Rev. 2.1 (Aqueous & Non- Aqueous) Comment: The laboratory Manual Integration SOP (QAU-GEN003-02) states in Section 6.0 that “If the QDEL function is used, the deletion should be marked as a “d” next to the compound. The ion profile or peak must be printed, and an explanation of the deletion should be documented.” The data reviewed did not include the peak that was deleted, nor an explanation of the deletion. T. Finding: The laboratory is not always examining computerized data reduction properly to ensure that peaks are integrated correctly. Requirement: All computerized data reduction must be reviewed carefully by the analyst to determine the accuracy and appropriateness of the quantitation performed by the data system. Any errors must be corrected using this document as a guideline to define appropriate integration. The document must describe procedures for completing and documenting corrections to analytical results. The failure of the software to appropriately integrate a peak is usually obvious from visual inspection of the chromatogram (i.e., at an appropriate scale - enlarged chromatograms must be used to discern baseline noise). Ref: NC WW/GW LCB Manual Integration Policy. Comment: Multiple instances were noted where the Chemstation software had improperly integrated peaks. These errors were often caused by examining the chromatogram at an inappropriate scale. U. Finding: The laboratory manually integrates chromatograms without documenting the reason why the manual integration was performed. Cited Previously on July 21, 2016. Requirement: All data must be integrated consistently in standards, samples and QC samples. Integration parameters (both automated and manual) must adhere to valid scientific chromatographic principles. Manual integration is employed to correct an improper integration performed by the data system and must always include documentation clearly stating the reason the manual integration was performed and who completed the work. Under no circumstances should manual integration be performed solely for the purpose of meeting quality control criteria. Ref: NC WW/GW LCB Manual Integration Policy. V. Finding: The laboratory was not able to demonstrate compliance with current ongoing Method Detection Limit (MDL) requirements for all applicable parameter methods, where required. Requirement: During any quarter in which samples are being analyzed, prepare and analyze a minimum of two spiked samples on each instrument, in separate batches, using the same spiking concentration used in Section 2. If any analytes are repeatedly not detected in the quarterly spiked sample analyses, or do not meet the qualitative identification criteria of the method (see Section 2(c) of this procedure), then this is an indication that the spiking level is not high enough and should be adjusted upward. Note that it is not necessary to analyze additional method blanks together with the spiked samples, the method blank population should include all of the routine method blanks analyzed with each batch during the course of sample analysis. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 82, No. 165, August 28, 2017; Appendix B–Definition and Procedure for the Determination of the Method Detection Limit–Revision 2. (3) (a). Requirement: Ensure that at least seven spiked samples and seven method blanks are completed for the annual verification. If only one instrument is in use, a minimum of seven Page 10 #402 Waypoint Analytical - Charlotte spikes are still required, but they may be drawn from the last two years of data collection. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 82, No. 165, August 28, 2017; Appendix B–Definition and Procedure for the Determination of the Method Detection Limit–Revision 2. (3) (b). Requirement: At least once every thirteen months, re-calculate MDLs and MDLb from the collected spiked samples and method blank results using the equations in Section 2. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 82, No. 165, August 28, 2017; Appendix B–Definition and Procedure for the Determination of the Method Detection Limit–Revision 2. (4) (a). W. Finding: The laboratory has not established an acceptance criterion for the back-calculation of the standard at the lowest reporting concentration. Requirement: For analytical procedures requiring analysis of a series of standards, the concentrations of these standards shall bracket the range of the sample concentrations measured. One of the standards shall have a concentration equal to or less than the laboratory's lowest reporting concentration for the parameter involved. All data sets shall reference the corresponding calibration. Laboratories shall analyze or back-calculate a standard at the same concentration as the lowest reporting concentration each day samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (H). Requirement: Unless specified by the method or this Rule, each laboratory shall establish performance acceptance criteria for all quality control analyses. Each laboratory shall calculate and document the precision and accuracy of all quality control analyses with each sample set. When the method of choice specifies performance acceptance criteria for precision and accuracy, and the laboratory chooses to develop laboratory-specific limits, the laboratory-specific limits shall not be less stringent than the criteria stated in the approved method. Ref: 15A NCAC 02H .0805 (a) (7) (A). Comment: This Finding applies to TKN by EPA Method 351.2, Rev. 2.0. X. Finding: The laboratory is not adjusting the sample concentration in the percent recovery calculation when the spike volume is >1% of the total sample volume. Requirement: The volume of spike solution used in MS preparation must in all cases be ≤ 5% of the total MS volume. It is preferable that the spike solution constitutes ≤ 1% of the total MS volume so that the MS can be considered a whole volume sample with no adjustment (i.e., volume correction) by calculation necessary. If the spike solution volume constitutes >1% of the total sample volume, the sample concentration must be adjusted by calculation. Ref: NC WW/GW LCB Matrix Spiking Policy. Comment: This Finding applies to TKN by EPA Method 351.2, Rev. 2.0. Comment: The spike volume is 5%. Y. Finding: Preservation acid is not being added to reagent (method) blanks. Requirement: Laboratory Reagent Blank (LRB) - An aliquot of reagent water or other blank matrices that are treated exactly as a sample including exposure to all glassware, equipment, solvents, reagents, internal standards, and surrogates that are used with other samples. Ref: EPA Method 353.2, Rev. 2.0 (1993). Section 3.6. Page 11 #402 Waypoint Analytical - Charlotte Requirement: A reagent blank (method blank) consists of reagent water (see Section 1080) and all reagents (including preservatives) that typically are in contact with a sample during the entire analytical procedure. The reagent blank is used to determine whether and how much reagents and the preparative analytical steps contribute to measurement uncertainty. Ref: Standard Methods 4020 B-2014. (5). Requirement: A reagent blank (method blank) consists of reagent water (see Section 1080) and all reagents (including preservatives) that normally` are in contact with a sample during the entire analytical procedure. The reagent blank is used to determine whether and how much reagents and the preparative analytical steps contribute to measurement uncertainty. Ref: Standard Methods 5020 B-2017. (5). Comment: This Finding applies to samples analyzed for COD and NO2+NO3 Nitrogen. Comment: The laboratory needs to ensure that chemical preservatives are added to all reagent blanks for all parameters where chemical preservation is required. Reporting Z. Finding: The laboratory is not properly qualifying all results on the client report. Cited Previously on July 21, 2016. Requirement: Reported data associated with quality control failures, improper sample collection, holding time exceedances, or improper preservation shall be qualified as such. Ref: 15A NCAC 02H .0805 (e) (5). Comment: Current laboratory practice for Base Neutral/Acid Organics samples that do not have detections is that when a Continuing Calibration Verification (CCV) standard is biased low, a reporting limit standard is analyzed and data is qualified on the client report. If a CCV standard is biased high without sample detections, data is reported without qualification. Even if the impact of a QC failure is minimal, the associated data must be qualified. This situation must also be addressed in the next laboratory QA manual revision. Comment: Samples for pH are sometimes analyzed at the bench out of holding time. The client report includes a paragraph on the cover page stating the report contains pH sample results that were analyzed out of holding time, but this is the only place the qualifier is documented in the report. It is a general statement and does not specifically list the individual sample that is out of holding time. Recommendation: It is recommended that the laboratory qualifier code be placed adjacent to the actual result. AA. Finding: Data is not always qualified when quality control failures occur. Requirement: If quality control results fall outside established limits or show an analytical problem, the laboratory shall identify the Root Cause of the failure. The problem shall be resolved through corrective action, the corrective action process documented, and any samples involved shall be reanalyzed, if possible. If the sample cannot be reanalyzed, or if the quality control results continue to fall outside established limits or show an analytical problem, the results shall be qualified as such. Ref: 15A NCAC 02H .0805 (a) (7) (B). Comment: Multiple samples analyzed on June 21, 2021 for Organochlorine Pesticides by EPA 608.3 were affected by an ending CCV that failed low for a significant number of analytes. Surrogates were also low. The analyst was aware of the failure and made a comment in the Page 12 #402 Waypoint Analytical - Charlotte "Additional Text" field but failed to check a checkbox that would have flagged the associated data. Consequently, the data was reported to clients without qualification. Revised reports to the affected clients will be required as part of the Corrective Action. BB. Finding: Values less than the established reporting limit are being reported on the client report. Requirement: For analytical procedures requiring analysis of a series of standards, the concentrations of these standards shall bracket the range of the sample concentrations measured. One of the standards shall have a concentration equal to or less than the laboratory's lowest reporting concentration for the parameter involved. Ref: 15A NCAC 02H .0805 (a) (7) (H). Comment: The MDL is being used as the reporting limit for Base Neutral/Acid Organics and ICP-MS metals, Mercury and Purgeable Organics, rather than the concentration of the lowest standard. No qualifiers were noted in the case narratives regarding the MDL being used as the reporting limit. The laboratory SOP and QA manual both require the reporting limit be set at or above the concentration of the lowest standard. Comment: A reporting limit of 1.14 mg/L for Total Hardness was observed for Client Report 21-307-0012. The SOP for Hardness (WET-GEN009-12) defines the reporting limit as 5 mg/L. Comment: Client report 21-244-0017 dated September 20, 2021 reported an MDL for Turbidity of 0.49 NTU and an MQL of 1.0 NTU. The laboratory SOP for Turbidity (WET- GEN003-05) defines the reporting limit as 1.0 NTU. Sample results were reported as < 0.49 NTU. CC. Finding: The laboratory does not report results of all tests on the characteristics of the effluent when duplicate sample analyses are performed. Requirement: The results of all tests on the characteristics of the effluent, including but not limited to NPDES Permit Monitoring Requirements, shall be reported on monthly report forms. Ref: 15A NCAC 2B .0506 (b) (3) (J). Requirement: Any average for Fecal Coliform is to be calculated as a geometric mean. Ref: Directions for Completing Monthly Discharge Monitoring Reports, Section II, (8). Comment: When samples are duplicated for Fecal Coliform and E. Coli analyses, only the value of the first sample is in the client report. The client report does not include the duplicate result or the geometric mean of the results. Comment: This Finding applies to Fecal Coliform and E. Coli samples analyzed by Standard Methods, 9222 D-2015 (MF) (Aqueous), IDEXX Colilert-18® (Aqueous) and IDEXX Colilert® (24 hr) (Aqueous). Bacteria, Coliform Fecal – Standard Methods, 9221 E-2014 (MPN) (Aqueous & Biosolids) Bacteria, Coliform Fecal – Standard Methods, 9222 D-2015 (MF) (Aqueous & Biosolids) DD. Finding: The autoclave log does not include the time that media are withdrawn from the autoclave to demonstrate that media are in the autoclave for 45 minutes or less after the sterilization period has concluded. This is considered pertinent information. Requirement: All analytical records, including original observations and information necessary to facilitate historical reconstruction of the calculated results, shall be maintained Page 13 #402 Waypoint Analytical - Charlotte for five years. All analytical data and records pertinent to each certified analysis shall be available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (E). Requirement: After each run cycle, record the items sterilized, sterilization temperature, total run time (heat exposure), programmed/preset sterilization period, actual pressure readings, and analyst initials. Ensure that the autoclave maintains 121°C with minimal temperature variation at ≥15 lb/in.2 (≥103 kPa) for 15 min during the media sterilization cycle and that media are withdrawn from the autoclave in 45 min or less. Ref: Standard Methods 9020 B-2015. (4) (h). Bacteria, Coliform Fecal – Standard Methods, 9221 E-2014 (MPN) (Biosolids) Bacteria, Coliform Fecal – Standard Methods, 9222 D-2015 (MF) (Biosolids) EE. Finding: An inconsistency was noted between the SOP and laboratory practice as follows: The SOP states that samples analyzed by Standard Methods, 9221 E-2014 (MPN) (Biosolids) and Standard Methods, 9222 D-2015 (MF) (Biosolids) have a 24-hour holding time. Laboratory practice is to analyze them within 8 hours. Requirement: Sample analysis should begin as soon as possible after receipt; sample incubation must be started no later than 8 hours from time of collection. Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 86, No. 95, May 19, 2021; Table II, Footnote 22. Requirement: Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. Ref: 15A NCAC 02H .0805 (a) (7). Comment: The 24-hour holding time only applies to EPA Methods 1680 and 1681 and the laboratory is not certified for these methods. FF. Finding: The laboratory is not heating the evaporating dishes in a 103–105°C oven for ≥1 hour prior to sample analysis for percent solids. Requirement: Preparation of evaporating dish - If only measuring total solids, heat dish for ≥1 hour in a 103–105°C oven. Cool in desiccator to ambient temperature and weigh. Store in desiccator or 103–105°C oven until needed. Ref: Standard Methods, 2540 G-2015. (3) (a) (1). GG. Finding: The drying cycle is not repeated until a weight change of <50 mg is achieved. Requirement: Repeat cycle (drying, cooling, desiccating, and weighing) until weight change is <50 mg. Ref: Standard Methods, 2540 G-2015. (2) (b). Alkalinity – Standard Methods, 2320 B-2011 (Aqueous) HH. Finding: The volume analyzed for low-alkalinity samples is not between 100 and 200 mL. Page 14 #402 Waypoint Analytical - Charlotte Requirement: Report alkalinity less than 20 mg CaCO3/L only if it has been determined by the low-alkalinity method of 2320B.4d. Ref: Standard Methods, 2320 B-2011. (1) (d). Requirement: Potentiometric titration of low alkalinity: For alkalinities less than 20 mg/L titrate 100 to 200 mL according to the procedure of ¶ c above, using a 10 mL microburet and 0.02N standard acid solution. Ref: Standard Methods, 2320 B-2011. (4) (d). Comment: Alkalinity is considered a method-defined parameter per the definition in the Code of Federal Regulations, Part 136.6, Section (a) (5). This means the method must be followed as written. Comment: The laboratory analyzes 50 mL for low-alkalinity samples. BOD – Standard Methods, 5210 B-2016 (Aqueous) CBOD – Standard Methods, 5210 B-2016 (Aqueous) II. Finding: The laboratory benchsheet for BOD/CBOD is lacking required documentation: quality control assessments. Requirement: All analytical records, including original observations and information necessary to facilitate historical reconstruction of the calculated results, shall be maintained for five years. All analytical data and records pertinent to each certified analysis shall be available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (E). Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: all quality control assessments. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (xv). Comment: Calibration buffer true values are documented in the columns for the true and observed values of the check buffer, which are not documented elsewhere.. JJ. Finding: Calculated results of sample dilutions are not being evaluated to ensure ≤ 30% difference between valid high and low values. Requirement: Identify results in the test reports when any of the following QC conditions occur: test replicates show >30% difference between highest and lowest values. Ref: Standard Methods, 5210 B-2016. (7) (b). KK. Finding: The laboratory is not analyzing a check standard buffer after calibration and prior to sample analysis. Requirement: Immerse in a third buffer below pH 10, approximately 3 pH units different from the second; the reading should be within 0.1 unit for the pH of the third buffer. If the meter response shows a difference greater than 0.1 pH unit from expected value, look for trouble with the electrodes or potentiometer. Ref: Standard Methods, 4500 H+ B-2011. (4). Comment: This Finding applies to the pH meter used in BOD/CBOD analyses. Chromium, Hexavalent – Standard Methods, 3500 Cr- B-2011 (Aqueous) Comment: On the benchsheet the column "drops Acid added" is not required, just the documented verification of sample pH 2.0 ± 0.5 S.U. However, if the column is retained, the type of acid must be Page 15 #402 Waypoint Analytical - Charlotte documented since it could be either H3PO4 or 0.2N H2SO4 depending which point in the procedure this represents. Chromium, Hexavalent – SW-846 7196 A (Non-Aqueous) LL. Finding: The laboratory does not document that each sample pH has been adjusted to 7.5 ± 0.5 S.U. following digestion. This is considered pertinent information. Requirement: All analytical records, including original observations and information necessary to facilitate historical reconstruction of the calculated results, shall be maintained for five years. All analytical data and records pertinent to each certified analysis shall be available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (E). Requirement: Adjust the pH of the solution to 7.5 ± 0.5 if the sample is to be analyzed using Method 7196 (adjust the pH accordingly if an alternate analytical method is to be used; i.e. 9.0 ± 0.5 if Method 7199 is to be used) and monitor the pH with a pH meter. If the pH of the digest should deviate from the desired range, discard the solution and redigest. Ref: EPA SW- 846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 3060A, Rev. 1, December 1996, Section 7.7. COD – Standard Methods, 5220 D-2011 (Aqueous) COD – Hach 8000 (Aqueous) MM. Finding: The laboratory does not analyze low range samples correctly. Requirement: For COD values between 100 and 900 mg/L, increase in Cr3+ in the 600-nm region is determined. Higher values can be obtained by sample dilution. COD values of 90 mg/L or less can be determined by following the decrease in Cr2O72- at 420 nm. The corresponding generation Cr3+ gives a small absorption increase at 420 nm, but this is compensated for in the calibration procedure. Ref: Standard Methods 5220 D-2011 (1). Comment: COD is considered a method-defined parameter per the definition in the Code of Federal Regulations, Part 136.6, Section (a) (5). This means the method must be followed as written. Comment: The laboratory has a reporting limit of 50 mg/L and analyzes a reporting limit standard with each batch. Comment: The laboratory may either obtain the low-level reagent kit and have a separate calibration curve for low range samples or raise the reporting limit to 100 mg/L. NN. Finding: An inconsistency was noted between the SOP and laboratory practice as follows: Personnel were not following procedures as stated in the Laboratory’s SOP. The SOP is in direct conflict with method and/or regulatory requirements. Requirement: Each laboratory shall have a documented analytical quality assurance and quality control program. Each laboratory shall have a copy of each approved test, analysis, measurement, or monitoring procedure being used in the laboratory. Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality Page 16 #402 Waypoint Analytical - Charlotte assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. The quality assurance, quality control, and Standard Operating Procedure documents shall be available for inspection by the State Laboratory. Ref: 15A NCAC 02H .0805 (a) (7). Comment: The laboratory is drying KHP at 100°C. The laboratory SOP states the drying temperature is 120°C, however, Standard Methods states to dry at 110 °C. COD is a method- defined parameter, and therefore must be followed as written. Color, PC – Standard Methods, 2120 B-2011 (Pt-Co) (Aqueous) OO. Finding: The laboratory benchsheet is lacking required documentation: the proper units of measure, all quality control assessments and any other data needed to reconstruct the final result. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the proper units of measure, all quality control assessments and any other data needed to reconstruct the final result. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (xii), (xv) and (xviii). Comment: The pH meter used (ID#: pH0016) holds an internal calibration, but this calibration information is not transferred to the LIMS or documented. Comment: The benchsheet does not contain the units of measure. PP. Finding: The laboratory needs to increase traceability of consumables used in the analysis of Color. Requirement: The laboratory shall have a documented system of traceability for the purchase, preparation, and use of all chemicals, reagents, standards, and consumables. Ref: 15A NCAC 02H .0805 (a) (7) (K). Comment: The laboratory benchsheet is lacking the comparison standard reference ID and the filter lot number. Conductivity – EPA 120.1, Rev. 1982 (Aqueous) Conductivity – Standard Methods, 2510 B-2011 (Aqueous) Conductivity – SW-846 9050 A (Aqueous) QQ. Finding: The Automatic Temperature Compensator (ATC) check is not being performed. Requirement: The State Laboratory may develop Approved Procedures for Field Parameters based upon the methods in any of the sources referenced in Parts(a)(1)(A) through (F) of this Rule. Ref: 15A NCAC 02H .0805 (a) (1) (F). Requirement: The Automatic Temperature Compensator (ATC) must be verified prior to initial use and annually (i.e., 12 months) thereafter at two temperatures by analyzing a standard or sample at 25°C (i.e., the temperature to which conductivity values are reported) and a temperature(s) that brackets the temperature ranges of the environmental samples routinely analyzed. This may require the analysis of a third temperature reading that is > 25°C. As the temperature increases or decreases, the value of the conductivity standard or sample must be within ±10% of the true value of the standard or ±10% of the value of the Page 17 #402 Waypoint Analytical - Charlotte sample at 25°C. If not, corrective action must be taken. Ref: NC WW/GW LCB Approved Procedure for the Analysis of Specific Conductance (Conductivity). Cyanide – Standard Methods, 4500 CN E-2016 (Aqueous) RR. Finding: Sample preservation verification checks performed at the laboratory benchtop are not consistently documented. Requirement: Sample preservation shall be verified and documented. Ref: 15A NCAC 02H .0805 (a) (7) (M). Comment: A column on the laboratory benchsheet is provided to document that the sample pH is > 12 S.U. but a check mark is not always recorded in this space. Data submitted prior to the inspection for Project #V11119 and V804 were lacking the check mark to indicate that the preservation requirement had been verified. Comment: The preservation requirements listed in Code of Federal Register, Title 40, Part 136, Federal Register Vol. 86 No. 95 May 19, 2021, Table II, state that the pH be >10 S.U. The laboratory may therefore choose to preserve samples to pH >10 S.U., rather than >12 S.U. SS. Finding: The laboratory benchsheet occasionally lacks required documentation: the volume of sample analyzed. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the volume of sample analyzed, where applicable. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (xi). Comment: The laboratory digestion worksheet has a column to record the “aliquot sample” final volume, but this column is not always filled out. Data reviewed prior to the inspection lacked the final volume for Project #V11119 and V804. pH – Standard Methods, 4500 H+B-2011 (Aqueous) Comment: Duplicates are being analyzed for pH. Duplicates are not required for Field parameters. This applies to bench and field analyses. Comment: The method only requires that one buffer be analyzed as a Check Standard Buffer. The laboratory’s current practice is to read back each calibration buffer as well as an outside source check standard buffer. Recommendation: The calibration check buffer is not clearly stated on the benchsheet. It is recommended that the laboratory change "Readback" to "Check Buffer". This applies to bench and Field analyses. TT. Finding: The laboratory is not analyzing a post-analysis check standard buffer when analyses are performed at multiple sample sites on a single day. Requirement: The State Laboratory may develop Approved Procedures for Field Parameters based upon the methods in any of the sources referenced in Parts(a)(1)(A) through (F) of this Rule. Ref: 15A NCAC 02H .0805 (a) (1) (F). Page 18 #402 Waypoint Analytical - Charlotte Requirement: When performing analyses at multiple sample sites, a post-analysis calibration verification using the check standard buffer must be analyzed at the end of the run. It is recommended that a mid-day check standard buffer be analyzed when samples are analyzed over an extended period of time. The post-analysis check standard buffer(s) must read within ±0.1 S.U. or corrective actions must be taken. If recalibration is necessary, all samples analyzed since the last acceptable calibration verification must be reanalyzed, if possible. If samples cannot be reanalyzed, the data must be qualified. Ref: NC WW/GW LCB Approved Procedure for the Analysis of pH. Comment: This Finding applies to pH analyses performed in the field. UU. Finding: The laboratory is not using fresh aliquots of standard buffers to calibrate the pH meter each day. Requirement: The State Laboratory may develop Approved Procedures for Field Parameters based upon the methods in any of the sources referenced in Parts(a)(1)(A) through (F) of this Rule. Ref: 15A NCAC 02H .0805 (a) (1) (F). Requirement: The calibration and check standard buffers must bracket the range of the samples being analyzed. A portion of the standard buffer is not to be used for more than one calibration. Discard any used buffer portions. Do not pour unused portions back into the original bottle. Ref: NC WW/GW LCB Approved Procedure for the Analysis of pH. Comment: The buffers are poured into a plastic tube from the stock buffers in the main lab. These are poured up at the beginning of the week and replaced weekly. Comment: This Finding applies to pH analysis performed in the field. VV. Finding: Values were reported that exceed the method specified accuracy of 0.1 units. Requirement: By careful use of a laboratory pH meter with good electrodes, a precision of ±0.02 unit and an accuracy of ±0.05 unit can be achieved. However, ± 0.1 pH unit represents the limit of accuracy under normal conditions, especially for measurement of water and poorly buffered solutions. For this reason, report pH values to the nearest 0.1 pH unit. Ref: Standard Methods, 4500 H+ B-2011. (6). Comment: The laboratory currently reports pH sample results to two decimal places on client reports. Comment: This Finding applies to pH analyses performed in the laboratory and in the field. Hardness – Standard Methods, 2340 C-2011 (Aqueous) WW. Finding: Documentation does not demonstrate that the samples were titrated within five minutes of the addition of the buffer. This is considered pertinent information. Requirement: All analytical records, including original observations and information necessary to facilitate historical reconstruction of the calculated results, shall be maintained for five years. All analytical data and records pertinent to each certified analysis shall be available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (E). Requirement: Select a sample volume that requires less than 15 mL EDTA and complete titration within 5 min, measured from time of buffer addition. Ref: Standard Methods, 2340 C-2011. (3) (b). Page 19 #402 Waypoint Analytical - Charlotte Ignitability – SW-846 1030 (Non-Aqueous) XX. Finding: A description of the material, the sample pre-test results, particle size, temperature of the test material (°C), air velocity through the fume hood (m/s) and burn rate are not documented. This is considered pertinent information. Requirement: All analytical records, including original observations and information necessary to facilitate historical reconstruction of the calculated results, shall be maintained for five years. All analytical data and records pertinent to each certified analysis shall be available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (E). Requirement: The NC WW/GW Laboratory Certification Branch must confirm that analytical data is valid by reviewing the QA/QC data generated during the sampling and analysis procedures when implementing SW-846 methods in order to be assured that scientifically sound decisions are made which will be protective of human health and the environment. To promote consistency with the use of SW-846 methods and to assure generation of data of known quality, the minimum recommended quality control benchmarks in the methods will be considered the minimum QA/QC requirements. For example, where a method states, “Documenting the effect of the matrix on target analyte measurements should include the analysis of at least one matrix spike and one duplicate unspiked samples or one matrix spike/matrix spike duplicate pair.”, the laboratory must analyze at least one matrix spike and one duplicate unspiked sample or one matrix spike/matrix spike duplicate pair. Laboratories may adopt more stringent QC acceptance criteria for method performance but may not omit or use less stringent criteria than that stated in SW-846 methods. Ref: NC WW/GW LCB SW-846 Method Implementation Policy. Comment: The laboratory only documents the preliminary screening for “Negative” or “positive” and the burn time in seconds. Results of the pre-test to determine if the sample is explosive or extremely flammable are not documented. YY. Finding: An inconsistency was noted between the SOP and laboratory practice as follows: the laboratory is not reporting results in the units of mm/sec and applying the “Positive” or “Negative” qualifiers as required by the SOP. Requirement: Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. Ref: 15A NCAC 02H .0805 (a) (7). Comment: The SOP (WET-GEN035-04) Section 9.1.6 states that if the result of the preliminary screening test is negative, results are to be reported as 0 mm/sec and a qualifier of “Negative” added. Section 9.3 states to calculate the average of the three results of the burn rate portion of the test and add a qualifier to report the results as “Positive” or “Negative”. The laboratory is reporting results as “Negative” without a numeric value for the burn rate. The method only requires reporting the burn rate. The addition of “positive” or “negative” is not required. Page 20 #402 Waypoint Analytical - Charlotte MBAS – Standard Methods, 5540 C-2011 (Aqueous) Comment: The reporting limit documented at the top left corner of the laboratory sample preparation log for MBAS does not match the reporting limit on the client report. There is a column header on the log captioned "Reporting Limit". In the column, the reporting limit is written in as 0.2 mg/L, which does not match the lowest standard in the calibration curve and the client report, which is 0.1 mg/L. This appears to be a copy and paste error. Nitrogen, Ammonia – Standard Methods, 4500 NH3 G-2011 (Aqueous) Nitrogen, Ammonia – Standard Methods, 4500 NH3 D-2011 (Aqueous) Nitrogen, Ammonia – Standard Methods, 4500 NH3 C-2011 (Aqueous & Non-Aqueous) ZZ. Finding: Samples are not checked for Total Residual Chlorine in the laboratory prior to analysis. Requirement: Residual Chlorine reacts with ammonia; remove by sample pretreatment. If a sample is likely to contain residual chlorine, immediately upon collection, treat with dechlorinating agent as in 4500-NH3.B.3d. Ref: Standard Methods, 4500 NH3 A-2011. (2). Chromium Hexavalent - Standard Methods, 3500 Cr B-2011 (Aqueous) Nitrogen, Ammonia - Standard Methods, 4500 NH3 G-2011 (Aqueous) Nitrogen, Nitrite - Standard Methods, 4500 NO2- B-2011 (Aqueous) Nitrogen, NO3 + NO2 - Standard Methods, 4500 NO3- F-2016 (Aqueous) Phosphate, Ortho - Standard Methods, 4500 P E-2011 (Aqueous) Phosphorus, Total - Standard Methods, 4500 P E-2011 (Aqueous) Phosphorus, Total - Standard Methods 4500 P F-2011 (Aqueous) AAA. Finding: The laboratory is not evaluating the recoveries of the back-calculated standards. Requirement: Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. The quality assurance, quality control, and Standard Operating Procedure documents shall be available for inspection by the State Laboratory. Ref: 15A NCAC 02H .0805 (a) (7). Requirement: Compare each calibration point to the curve and recalculate its concentration. If any recalculated values are not within the method's acceptance criteria -- up to twice the MRL ±50%; between 3 and 5 times the MRL ±20%; or greater than 5 times the MRL ±10% -- unless otherwise specified in individual methods, identify the source of any outlier(s) and correct before sample quantitation. Ref: Standard Methods, 3020 B-2017. (1) (b). Requirement: Compare each calibration point to the curve and recalculate its concentration. If any recalculated values are not within the method's acceptance criteria -- up to twice the MRL ±50%; between 3 and 5 times the MRL ±20%; or greater than 5 times the MRL ±10% -- unless otherwise specified in individual methods, identify the source of any outlier(s) and correct before sample quantitation. Ref: Standard Methods, 4020 B-2014. (1) (b). Page 21 #402 Waypoint Analytical - Charlotte Nitrogen, Nitrite – Standard Methods, 4500 NO2- B-2011 (Aqueous) BBB. Finding: The laboratory does not document that the sample pH is between 5-9 S.U. before color reagent is added. This is considered pertinent information. Requirement: If sample pH is not between 5 and 9, adjust to that range with 1N HCl or NH4OH as required. Ref: Standard Methods, 4500 NO2- B-2011. (4) (b). Requirement: All analytical records, including original observations and information necessary to facilitate historical reconstruction of the calculated results, shall be maintained for five years. All analytical data and records pertinent to each certified analysis shall be available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (E). Comment: The exact pH does not need to be recorded, but acknowledgment that the pH was between 5 and 9 S.U. must be documented. CCC. Finding: The MS/MSD are prepared with greater than 5% spike solution by volume. Requirement: The volume of spike solution used in MS preparation must in all cases be ≤ 5% of the total MS volume. It is preferable that the spike solution constitutes ≤ 1% of the total MS volume so that the MS can be considered a whole volume sample with no adjustment (i.e., volume correction) by calculation necessary. If the spike solution volume constitutes >1% of the total sample volume, the sample concentration must be adjusted by calculation. Ref: NC WW/GW LCB Matrix Spiking Policy. Comment: The benchsheet states that the laboratory uses 5mL of spike solution and 45mL of sample. The SOP states that 0.5 mL of spike solution is used in a total volume of 50mL, which is an acceptable volume and would not require sample concentration adjustment. Nitrogen, NO3 + NO2 - Standard Methods, 4500 NO3- F-2016 (Aqueous) Nitrogen, NO3 + NO2 - EPA Method 353.2, Rev. 2.0, 1993. (Aqueous) DDD. Finding: Samples are not consistently chemically preserved within fifteen minutes. Requirement: Except where noted in this Table II and the method for the parameter, preserve each grab sample within 15 minutes of collection. For a composite sample collected with an automated sample (e.g., using a 24-hour composite sample; see 40 CFR 122.21 (g)(7)(i) or 40 CFR Part 403, Appendix E), refrigerate the sample at ≤6°C during collection unless specified otherwise in this Table II or in the method(s). For a composite sample to be split into separate aliquots for preservation and/or analysis, maintain the sample at ≤6°C, unless specified otherwise in this Table II or in the method(s), until collection, splitting, and preservation is completed. Add the preservative to the sample container prior to sample collection when the preservative will not compromise the integrity of a grab sample, a composite sample, or aliquot split from a composite sample within 15 minutes of collection. If a composite measurement is required but a composite sample would compromise sample integrity, individual grab samples must be collected at prescribed time intervals (e.g., 4 samples over the course of a day, at 6-hour intervals). Ref: Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 86, No. 95, May 19, 2021; Table II, Footnote 2. Comment: Unpreserved samples are occasionally preserved at the bench within 48 hours of collection and analyzed. Page 22 #402 Waypoint Analytical - Charlotte EEE. Finding: The ICV standard is not properly identified on the benchsheet. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: sample identification. Ref: 15A NCAC 02H .0805 (a) (7) (F) (ix). Comment: The standard preparation section of the benchsheet lists both an "ICV" and a "CCV" standard preparation, however there is no standard labeled as "ICV" included in the analysis run log. The analyst stated that the first standard labeled as "CCV" in the analysis run log is the "ICV" standard. Nitrogen, Total Kjeldahl – EPA 351.2, Rev. 2.0, 1993 (Aqueous) FFF. Finding: The temperature of the block digestor is not checked and documented to be 160°C prior to digestion. This is considered pertinent data. Requirement: Place tubes in block digestor preheated to 160°C and maintain temperature for one hour. Ref: EPA Method 351.2, Rev. 2.0 (1993), Section 11.4. Requirement: All analytical records, including original observations and information necessary to facilitate historical reconstruction of the calculated results, shall be maintained for five years. All analytical data and records pertinent to each certified analysis shall be available for inspection upon request. Ref: 15A NCAC 02H .0805 (a) (7) (E). GGG. Finding: The laboratory is not properly preparing matrix spike samples that are above the calibration range. Requirement: The concentration of the spiked samples must be bracketed by the calibration range. If the spiked sample result is over the calibration range, the spiked sample must be diluted and re-analyzed. It is not acceptable to dilute the sample first and then add the spike solution so as not to affect bias attributed to matrix. Ref: NC WW/GW LCB Matrix Spiking Policy. Phosphorus, Ortho – Standard Methods, 4500 P E-2011 (Aqueous) HHH. Finding: The laboratory is not consistently analyzing the LFM and LFMD at the required frequency. Requirement: The LFM is used to evaluate analyte recovery in a sample matrix. If an LFM is feasible and the method does not specify LFM frequency requirements, then include at least one LFM with each sample set (batch) or on a 5% basis, whichever is more frequent. Ref: Standard Methods, 4020 B-2014. (7) and Table 4020:I. Requirement: As a minimum, include one duplicate sample or one LFM duplicate with each sample set (batch) or on a 5% basis, whichever is more frequent, and process it independently through the entire sample preparation and analysis. Ref: Standard Methods, 4020 B-2014. (8) and Table 4020:I. Comment: No LFM or LFMD were included in batches V8393 and V11225. Page 23 #402 Waypoint Analytical - Charlotte III. Finding: The laboratory is not qualifying results when samples are not filtered within 15 minutes of collection. Requirement: Basic documentation requirements for the sample collector include: Chemical and/or physical preservation/treatment(s) used where required (e.g., name of preservative, pH<2, pH>9, field filtration, TRC neutralization, etc.). Ref: NC WW/GW LCB Sample Collection, Preservation and Receipt Requirements Policy. Comment: Benchsheets and Chain of Custody documents reviewed did not contain any documentation that samples were filtered, so it cannot be assumed that field filtration is being performed as required per Code of Federal Regulations, Title 40, Part 136; Federal Register Vol. 82, No. 165, August 28, 2017; Table II. Phosphorus, Total – Standard Methods, 4500 P F-2011 (Aqueous) JJJ. Finding: The laboratory is not checking the pH of the digested sample and adjusting if necessary. Requirement: Cool, add 0.05 mL (1 drop) phenolphthalein indicator solution, and neutralize to a faint pink color with NaOH. Ref: Standard Methods 4500 P B-2011. (5) (c). Requirement: Add 0.05 mL (1 drop) phenolphthalein indicator solution to approximately 50 mL sample. If a red color develops, add H2SO4 (4500-P.F.3e) dropwise to just discharge the color. Standard Methods 4500 P F-2011. (4). Comment: Based on the two Requirements, it is understood that prior to sample analysis, the pH of the sample is to be adjusted within a neutral pH. Phosphorus, Total – EPA Method 365.1, Rev. 2.0, 1993 (Aqueous) KKK. Finding: The laboratory does not analyze matrix spike samples at the required frequency. Requirement: The laboratory must add a known amount of analyte to a minimum of 10% of the routine samples. Ref: EPA Method 365.1, Rev. 2.0 (1993), Section 9.4.1 Comment: The SOP (WET-AA003-05) combines EPA Method 365.1, Rev. 2.0, 1993 and SM 4500 PF-2011 and only requires an MS/MSD pair per analytical batch of 20 samples, which is a frequency of 5%. To combine the methods, the laboratory must use the strictest requirement between the two methods. LLL. Finding: The laboratory does not rerun samples whose computed value is less than 5% of its predecessor. Requirement: Any sample whose computed value is less than 5% of its immediate predecessor must be rerun. Ref: EPA Method 365.1, Rev. 2.0 (1993), Section 12.2. Comment: There were no instances of samples with concentrations <5% of the immediate predecessor in the three data packs reviewed. Oil & Grease – EPA 1664, Rev. B (Aqueous) Comment: The laboratory analyzes the parameter using solid phase extraction. Page 24 #402 Waypoint Analytical - Charlotte MMM. Finding: The laboratory benchsheet is lacking required documentation: sample preparation; the proper units of measure; all manual calculations; all quality control assessments. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: sample preparation, the proper units of measure, all manual calculations and all quality control assessments. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (x) (xii) (xiv) (xv). Comment: The data sheets do not include a post-analysis balance verification. Comment: The data sheets were lacking the volume of acid used in the analysis, acceptance criteria for the weight verification and units of measure for the temperature of the Speed Vap. Residue, Dissolved 180 C – Standard Methods, 2540 C-2015 (Aqueous) Residue, Suspended – Standard Methods, 2540 D-2015 (Aqueous) Residue, Total - Standard Methods, 2540 B-2015 (Aqueous) NNN. Finding: The laboratory benchsheet is lacking required documentation: the instrument identification; the proper units of measure; all manual calculations; all quality control assessments; and any other data needed to reconstruct the final calculated result. Requirement: All laboratories shall use printable laboratory benchsheets. Certified Data shall be traceable to the associated sample analyses and shall consist of: the instrument identification, the proper units of measure, all manual calculations, all quality control assessments and any other data needed to reconstruct the final calculated result. Each item shall be recorded each time that samples are analyzed. Ref: 15A NCAC 02H .0805 (a) (7) (F) (iii) (xi) (xiv) (xv) (xviii). Comment: The laboratory is not documenting that the weight change between drying cycles is < 0.5 mg. No data was found that did not meet the weight change criterion. Comment: The Dissolved Residue benchsheet for July 6, 2021 did not have the LCS ID#, the balance ID and the filter lot ID documented. The Dissolved Residue benchsheet for November 2, 2021 did not have the units of measure documented. Comment: The Total Residue benchsheet for September 30, 2021 did not have the oven ID# or the balance ID documented. Residue, Dissolved 180 C – Standard Methods, 2540 C-2015 (Aqueous) OOO. Finding: The laboratory is not properly evaluating the Method Blank. Requirement: Using the same containers and glassware normally in contact with samples, put 30 ml of DI water through the sample filter and proceed through the entire analytical process. Acceptance criterion is <0.5 mg weight gain. Ref: NC WW/GW LCB Policy. Comment: The laboratory was using a Method Blank acceptance criterion of < 25 mg/l. Mercury – EPA 245.1, Rev. 3.0, 1994 (Aqueous) Comment: It was noted during data review that the laboratory is diluting all digested samples and standards to a final volume of 30 mL. This makes data reconstruction difficult without access to the SOP or LIMS to view the preparation of individual standards, since standard preparation is recorded Page 25 #402 Waypoint Analytical - Charlotte electronically. Section 8.2.3.1 of the SOP (MET-CVA001-12) lists concentrations of the calibration standards at both a 20 mL volume (prior to digestion) and the 30 mL final volume. Calculations are based on the concentration of the calibration curve standards at the 30 mL final volume. The SOP does not include the concentration of calibration verification standards at the 30 mL final volume, and this made data reproduction difficult. The analyst did not know why this was current practice but did say it had been in place since before they started. Comment: The laboratory is preparing the LFB and LFM by adding 2500 µL of a 0.075 mg/L standard solution to 20 mL of reagent water and sample prior to digestion. After digestion, the sample is diluted to a final volume of 30 mL. This yields a theoretical value of 6.25 µg/L. This is well above the midpoint of the calibration curve. While this does not impact the final calculation and data quality, such a large volume in the LFM may impact the identification of matrix interferences in the original sample. Adding a smaller volume of a more concentrated standard solution with a theoretical value less than the midpoint of the calibration curve would better evaluate potential interferences while maintaining compliance with the analytical method. Recommendation: It is recommended that the laboratory revise their digestion logs to eliminate unnecessary columns and update preprinted headings to avoid strikethroughs. PPP. Finding: The laboratory digests the calibration standards. Requirement: Prepare calibration standards by transferring 0.5, 1.0, 2.0, 5.0, and 10 mL aliquots of the 0.1 µg/mL CAL (Section 7.6) to a series of sample containers (Section 6.5.2). Dilute the standard aliquots to 100 mL with reagent water (Section 7.2) and process as described in Sections 11.1.2, 11.1.3 (without heating), and 11.1.5. Ref: EPA Method 245.1, Rev. 3.0 (1994), Section 11.2.2. Comment: Section 11.2.2 states to process calibration standards as in 11.1.3 without heating. EPA Region IV has given guidance that the reason this method does not include heating is because the standards only include the inorganic form of Hg. Heating is meant to dissociate organic or bound Hg and is therefore unnecessary. As such, EPA did not recommend combining EPA 245.1 with SW-846 7470A since 7470A states that the standards are heated like the samples. Laboratories may combine methods but would have to follow the EPA 245.1 requirements. QQQ. Finding: The laboratory is not preparing the Laboratory Fortified Blank (LFB) at the correct concentration. Requirement: The LFB is prepared by fortifying a sample size volume of laboratory reagent blank solution with mercury to a suitable concentration of >10X the MDL but less than the midpoint concentration of the calibration curve. Ref: EPA Method 245.1, Rev. 3.0 (1994), Section 7.11. Comment: The laboratory is currently preparing the LFB at a concentration of 6.25 µg/L. which is well above the midpoint of the calibration curve (4.6875 µg/L). Metals – EPA 200.8, Rev. 5.4, 1994 (Aqueous) Metals – SW-846 6020 B-2011 (Aqueous) Recommendation: The laboratory is currently analyzing samples using both “Helium” and “No Gas” modes. It is unclear from the data package review which elements are reported with which mode. It is recommended that the data package document which mode is used for which elements. Page 26 #402 Waypoint Analytical - Charlotte Metals – EPA 200.8, Rev. 5.4, 1994 (Aqueous) RRR. Finding: An inconsistency was noted between the SOP and laboratory practice as follows: the SOP does not mention filtering samples and standards post digestion. Requirement: Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. Ref: 15A NCAC 02H .0805 (a) (7). Comment: When there are samples that need to be filtered post-digestion to eliminate high solids content, the MS/MSD and an additional blank are filtered. The method does allow filtration of samples after digestion. However, the process is not described in the SOP. Metals – EPA 200.7, Rev. 4.4, 1994 (Aqueous) Metals – SW-846 6010 D (Aqueous & Non-Aqueous) SSS. Finding: An inconsistency was noted between the SOP and laboratory practice as follows: The laboratory is not analyzing an ICS-AB at the end of the run. Requirement: Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. Ref: 15A NCAC 02H .0805 (a) (7). Comment: Section 10.2.4 of the SOP states “At the end of the analytical run, a CCV, CCB, an ICSA and ICS-AB are run and must pass the recovery requirements.” Comment: Neither method require the spectral interference checks to be analyzed at the end of the analytical sequence. 1,2-Dibromoethane (EDB) – SW-846 8011 (Includes DBCP) (Aqueous) TTT. Finding: The laboratory is not using an acceptance criterion of less than 10% for the RSD of the average calibration factor for the calibration curve. Requirement: Analyze each calibration standard and tabulate peak height or area response versus the concentration in the standard. Prepare a calibration curve for each compound. Alternatively, if the ratio of response to concentration (calibration factor) is a constant over the working range (< 10% relative standard deviation), linearity can be assumed and the average ratio or calibration factor can be used in place of a calibration curve. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8011 A. Rev. 0, July 1992, Section 7.2.2. Page 27 #402 Waypoint Analytical - Charlotte Comment: The laboratory is using an acceptance criterion of less than or equal to 10%. Base Neutral/Acid, Organics – SW-846 8270 E (Aqueous & Non-Aqueous) Base Neutral/Acid, Organics – EPA 625.1, December 2016 (Aqueous) TPH Diesel Range, Organics – SW-846 8015 C (Aqueous & Non-Aqueous) UUU. Finding: The laboratory is not consistently calculating, evaluating and reporting detections in the LRB. Requirement: If quality control results fall outside established limits or show an analytical problem, the laboratory shall identify the Root Cause of the failure. The problem shall be resolved through corrective action, the corrective action process documented, and any samples involved shall be reanalyzed, if possible. If the sample cannot be reanalyzed, or if the quality control results continue to fall outside established limits or show an analytical problem, the results shall be qualified as such. Ref: 15A NCAC 02H .0805 (a) (7) (B). Requirement: The blank must meet the requirements in Section 8.5 before sample extracts may be analyzed. Ref: EPA Method 625.1, December 2016, Section 13.6. Requirement: If an analyte of interest is found in the blank: 1) at a concentration greater than the MDL for the analyte, 2) at a concentration greater than one-third the regulatory compliance limit, or 3) at a concentration greater than one-tenth the concentration in a sample in the extraction batch, whichever is greater, analysis of samples must be halted, and the problem corrected. Ref: EPA Method 625.1, December 2016, Section 8.5.2. Requirement: Quality control procedures necessary to evaluate the GC system operation are found in Method 8000 and include evaluation of retention time windows, calibration verification and chromatographic analysis of samples. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8015 C, February 2007, Section 9.3. Requirement: Results of the method blank should be less than the lower limit of quantitation (LLOQ) (Sec. 9.7) for the analyte or less than the level of acceptable blank contamination specified in the approved QAPP or other appropriate systematic planning document. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 D, March 2018, Section 9.2.6.5. Requirement: If the method blank results do not meet the acceptance criteria above, the laboratory should take corrective action to locate and reduce the source of the contamination and re-extract and reanalyze any samples associated with the contaminated method blank. If the method blank results still do not meet the acceptance criteria in 9.2.6.5 and re-analysis is not practical, then the data user should be provided with the sample results, the method blank results, and a discussion of the corrective actions undertaken by the laboratory. Qualification of the samples may be needed. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 D, March 2018, Section 9.2.6.6. Requirement: The concentration of reagent, method, and calibration blanks shall not exceed 50 percent of the lowest reporting concentration or as otherwise specified by the reference method. Ref: 15A NCAC 02H .0805 (a) (7) (H) (i). Comment: During data review, it was noted that some LRBs for Diesel Range Organics (DRO) had values that exceeded the requirement of ≤ ½ the concentration of the reporting limit but were not qualified. The analysts noted that a linear regression was used as the Page 28 #402 Waypoint Analytical - Charlotte calibration model. DRO results for the LRBs and samples were “zeroed out” if the response from the instrument was less than the response of the lowest calibration standard, without qualification or justification on the client report. The analyst also noted that a similar approach was used for Benzoic Acid using EPA Method 625.1 and SW-846 8270 E. While chromatographic analyses are heavily dependent on the judgement of the analyst, this is not an acceptable justification. The auditor explained that bias on the low end of a calibration curve is extremely likely with this calibration model, and that forcing the curve through zero is permitted by the methods, which eliminates low bias, without using zero as a calibration point. Comment: Client reports associated with data that had been “zeroed out” lacked qualifiers or discussion in the case narratives. Comment: The analytical methods and NC WW/GW LCB Rules require corrective action to be taken when a blank does not meet the acceptance criterion listed. Zeroing out a blank is not listed as a possible corrective action in either. The blank acceptance criterion to follow and corrective actions may be chosen from either the analytical method or NC WW/GW LCB Rules but the SOP must clearly address what the acceptance criterion and corrective action will be. Comment: The analyst noted that after major equipment maintenance to both instruments, calibration curves using a different model (the average calibration factor model) have been successfully achieved and the practice has been discontinued. Recommendation: It is recommended that this situation be addressed in the next revision of the laboratory QA manual, as well as analyst training programs. Base Neutral/Acid, Organics – SW-846 8270 E (Aqueous & Non-Aqueous) Base Neutral/Acid, Organics – EPA 625.1, December 2016 (Aqueous) VVV. Finding: Acceptance criteria for the surrogate compounds have not been updated. Requirement: Once established, control limits should be reviewed regularly and updated on a routine basis as established by the laboratory’s quality management plan. Reviewing appropriateness of these criteria with respect to generated data is especially important for newly implemented procedures or those not in continual use. The laboratory should monitor trends in both analyte recovery performance and also in the control limits. Control limits used to evaluate sample results should be those in place at the time of sample analysis. Once control limits are updated, they should apply to all subsequent analyses of new samples. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 D. Rev. 5, March 2018, Section 9.6.5. Requirement: The NC WW/GW Laboratory Certification Branch must confirm that analytical data is valid by reviewing the QA/QC data generated during the sampling and analysis procedures when implementing SW-846 methods in order to be assured that scientifically sound decisions are made which will be protective of human health and the environment. To promote consistency with the use of SW-846 methods and to assure generation of data of known quality, the minimum recommended quality control benchmarks in the methods will be considered the minimum QA/QC requirements. For example, where a method states, “Documenting the effect of the matrix on target analyte measurements should include the analysis of at least one matrix spike and one duplicate unspiked samples or one matrix spike/matrix spike duplicate pair.”, the laboratory must analyze at least one matrix spike and one duplicate unspiked sample or one matrix spike/matrix spike duplicate pair. Laboratories may adopt more stringent QC acceptance Page 29 #402 Waypoint Analytical - Charlotte criteria for method performance but may not omit or use less stringent criteria than that stated in SW-846 methods. Ref: NC WW/GW LCB SW-846 Method Implementation Policy. Comment: The low and high control limits for each surrogate compound in the Chemstation analytical software have not been updated to the most recent limits. The analyst did not know when the last updates occurred but noted that the limits printed in the data packages were not the current criteria. Without access to the laboratory LIMS, the data package cannot be properly reviewed by a third party to determine if QC criteria have been met. Pesticides, Organochlorine – EPA 608.3, December 2016 (Aqueous) Polychlorinated Biphenyls (PCBs) – EPA 608.3, December 2016 (Aqueous) Comment: Sample 89261 was collected as a composite sample and is supposed to be a grab sample, per EPA 608.3. There was nothing on the Shipment Receipt Form mentioning that the sample was improperly collected. The sample is for a pre-treatment agreement between Total Clean of North Carolina, LLC and CMUD. The pretreatment agreement specifies the sample type as "Grab/Comp". The project manager stated that they have always received composite samples from them for organochlorine pesticides/PCB analyses. She stated that she would check with CMUD to ensure that a composite sample is what they really want for those parameters. Recommendation: It is recommended that the client report comment on the fact the sample type was a composite instead of the method specified grab. Pesticides, Organochlorine – EPA Method 608.3, December 2016 (Aqueous) Pesticides, Organochlorine – SW-846 8081 B (Aqueous) WWW. Finding: The laboratory is not documenting that samples are being checked and/or mitigated for residual chlorine when samples are being tested for Aldrin. Requirement: If aldrin is to be determined and residual chlorine is present, add 80 mg/L of sodium thiosulfate but do not add excess. Any method suitable for field use may be employed to test for residual chlorine (Reference 14). If sodium thiosulfate interferes in the determination of the analytes, an alternative preservative (e.g., ascorbic acid or sodium sulfite) may be used. Ref: EPA Method 608.3, Section 9.2. Comment: Extractions personnel stated they always check samples for TRC but do not document the result. They also stated that they have never seen a positive hit. Pesticides, Organochlorine – EPA 608.3, December 2016 (Aqueous) Pesticides, Organochlorine – SW-846 8081 B (Aqueous) Polychlorinated Biphenyls (PCBs) – EPA 608.3, December 2016 (Aqueous) Polychlorinated Biphenyls (PCBs) – SW-846 8082 A (Aqueous) Recommendation: It is recommended that the laboratory begin including client instructions with sample bottles shipped to clients stating not to pre-rinse bottles with sample prior to filling for Organochlorine Pesticides and PCBs. XXX. Finding: Samples are improperly transferred from the sample bottles to the separatory funnel. Requirement: Mark the water meniscus on the side of the sample bottle for later determination of sample volume. Pour the entire sample into the separatory funnel. Pipet the surrogate standard spiking solution (Section 6.8.6) into the separatory funnel. Ref: EPA Method 608.3, Section 10.2.2. Page 30 #402 Waypoint Analytical - Charlotte Requirement: Using a 1-liter graduated cylinder, measure 1 liter (nominal) of sample. Alternatively, if the entire contents of the sample bottle are to be extracted, mark the level of sample on the outside of the bottle. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 3510 C. Rev. 3, December 1996, Section 7.1. Requirement: Add 60 mL of methylene chloride to the sample bottle, seal, and shake for 30 seconds to rinse the inner surface. Transfer the solvent to the separatory funnel and extract the sample by shaking the funnel for two minutes with periodic venting to release excess pressure. Ref: EPA Method 608.3, Section 10.2.3. Requirement: Quantitatively transfer the sample from the graduated cylinder (or sample bottle) to the separatory funnel. Use 60 mL of methylene chloride to rinse the cylinder (or bottle) and transfer this rinse solvent to the separatory funnel. If the sample was transferred directly from the sample bottle, refill the bottle to the mark made in Sec. 7.1 with water and then measure the volume of sample that was in the bottle. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 3510 C. Rev. 3, December 1996, Section 7.4. Comment: SOP and in-house practice is to pour samples into a 1-liter graduated cylinder and then into the separatory funnel. The graduated cylinder is rinsed with methylene chloride but the residue left in the original sample bottle is not being captured. The SOP and laboratory practice must be revised to analyze the entire volume from the sample bottle, rinse the sample bottle with methylene chloride, and eliminate the transfer of the sample to the graduated cylinder. Pesticides, Organochlorine – EPA Method 608.3, December 2016 (Aqueous) Pesticides, Organochlorine – SW-846 8081 B (Aqueous & Non-Aqueous) YYY. Finding: An inconsistency was noted between the SOP and laboratory practice as follows: The laboratory extraction SOP requires an MS/MSD pair be prepared for Chlordane and Toxaphene, but this is not being done. Requirement: Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. Ref: 15A NCAC 02H .0805 (a) (7). Comment: The laboratory SOP (ORG-PRP006-03) for EPA 608.3 states in Section 7.2.2 to prepare MS/MSD pairs for pesticides Toxaphene and Chlordane. This does not appear to be a requirement of the analytical methods. Comment: The laboratory SOP (ORG-GCS012-02) for EPA 608.3 states to analyze a MS/MSD pair for Chlordane and Toxaphene at least once every two years. The laboratory SOPs (ORG-GCS007-09 and ORG-GCS006-10) for SW-846 8081B lack the 2-year wording, so it is understood that these be prepared with each analytical batch. Page 31 #402 Waypoint Analytical - Charlotte Polychlorinated Biphenyls (PCBs) – SW-846 8082 A (Oil) ZZZ. Finding: An inconsistency was noted between the SOP and laboratory practice as follows: Section 10.1.3 of the laboratory SOP says to use Florisil column cleanup after sulfuric acid cleanup but the Florisil column is not being used. Requirement: Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. Ref: 15A NCAC 02H .0805 (a) (7). Comment: The title of the SOP (ORG-GCS010-01) is "ANALYSIS OF POLYCHLORINATED BIPHENYLS (PCBs) IN OIL BY METHOD SW846-8082, WITH FLORISIL COLUMN CLEANUP". Purgeable, Organics – SW-846 8260 D (Aqueous & Non-Aqueous) AAAA. Finding: The laboratory is not re-calculating all calibration points using the final calibration curve. Requirement: All calibration points, especially those equivalent to the LLOQ, should be recalculated (not reanalyzed) using the final calibration curve in which this standard is used (i.e., re-fitting the response from the calibration standard back into the curve). See Method 8000 for additional details. The recalculated concentration of the calibration standard corresponding to the LLOQ, especially where linear regression fits are used, should be within ±50% of the standard's true concentration if it is the lowest point, and within ±30% for all others (i.e. above the low standard). No refit criteria need be passed for calibration levels below the LLOQ. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8260 D. Rev. 4, June 2018, Section 11.3.5.4. Requirement: The NC WW/GW Laboratory Certification Branch must confirm that analytical data is valid by reviewing the QA/QC data generated during the sampling and analysis procedures when implementing SW-846 methods in order to be assured that scientifically sound decisions are made which will be protective of human health and the environment. To promote consistency with the use of SW-846 methods and to assure generation of data of known quality, the minimum recommended quality control benchmarks in the methods will be considered the minimum QA/QC requirements. For example, where a method states, “Documenting the effect of the matrix on target analyte measurements should include the analysis of at least one matrix spike and one duplicate unspiked samples or one matrix spike/matrix spike duplicate pair.”, the laboratory must analyze at least one matrix spike and one duplicate unspiked sample or one matrix spike/matrix spike duplicate pair. Laboratories may adopt more stringent QC acceptance criteria for method performance but may not omit or use less stringent criteria than that stated in SW-846 methods. Ref: NC WW/GW LCB SW- 846 Method Implementation Policy. Comment: The analyst indicated that this had previously been routine practice but had been discontinued at some point prior to the inspection. Page 32 #402 Waypoint Analytical - Charlotte Purgeable, Organics – EPA 624.1, December 2016 (Aqueous) BBBB. Finding: An inconsistency was noted between the SOP and laboratory practice as follows: The laboratory is not evaluating all the parameter results for the BFB m/z Abundance Criteria. Requirement: Each laboratory shall have a documented analytical quality assurance and quality control program. Each laboratory shall have a copy of each approved test, analysis, measurement, or monitoring procedure being used in the laboratory. Each laboratory shall develop documentation outlining the analytical quality control practices used for the Parameter Methods included in its Certification, including Standard Operating Procedures for each certified Parameter Method. Quality assurance, quality control, and Standard Operating Procedure documentation shall indicate the effective date of the document and be reviewed every two years and updated if changes in procedures are made. Each laboratory shall have a formal process to track and document review dates and any revisions made in all quality assurance, quality control, and Standard Operating Procedure documents. Supporting Records shall be maintained as evidence that these practices are implemented. The quality assurance, quality control, and Standard Operating Procedure documents shall be available for inspection by the State Laboratory. Ref: 15A NCAC 02H .0805 (a) (7). Comment: The laboratory is not documenting and evaluating the results of the m/z 50 and m/z 75 abundance criteria. EPA Method 624.1, Table 4, Footnote 1 states: “Alternative tuning criteria from other published EPA reference methods may be used, provided method performance is not adversely affected.” The criteria used are those listed in SW-846 8260 D. While this does not impact data quality, the SOP must document the criteria used. CCCC. Finding: The laboratory is not always analyzing a reagent blank immediately after the LCS. Requirement: Spike the internal standards and surrogates into the blank. Analyze the blank immediately after analysis of the LCS (Section 8.4) and prior to analysis of the MS/MSD and samples to demonstrate freedom from contamination. Ref: EPA Method 624.1, Section 8.5.1. Comment: The laboratory sometimes analyzes a MS/MSD after the LCS but before the reagent blank. TPH Diesel Range, Organics – SW-846 8015 C (Aqueous & Non-Aqueous) DDDD. Finding: The retention time check standard is not being properly evaluated. Requirement: All target analytes, surrogates, and/or n-alkanes in the calibration verification analyses need to fall within previously established retention time windows. If the retention time of any analyte does not fall within the established window, then corrective action must be taken to restore the system or a new calibration curve must be prepared for that compound. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8015 C. Rev 3, February 2007, Section 11.5.3. Requirement: For the analysis of DRO, sum the area of all peaks eluting between C10 and C28. This area is generated by projecting a horizontal baseline between the retention times of C10 and C28. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8015 C. Rev 3, February 2007, Section 11.11.1. Comment: Multiple instances were noted where the established retention time window did not include the entire peak for Decane. After talking with the Semivolatiles Supervisor, it was explained that during a previous audit, analysts were informed that the only requirement was Page 33 #402 Waypoint Analytical - Charlotte that the mean retention time be at the apex of the peak. This led to some early portions of the peak not being included in the area response. EEEE. Finding: The retention time study is not performed over the correct time frame. Requirement: Make three injections of all single component standard mixtures and multi- component analytes (such as PCBs) over the course of a 72-hour period. Serial injections or injections over a period of less than 72 hours may result in retention time windows that are too tight. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 D. Rev 6, March 2018, Section 11.6.1. Comment: The analytical run on 6/21/21 lists current and past studies for the DRO retention time. In 2020, injections were made over a 72-hour period (3/31-4/3). Later studies have been performed on the same day in the same analytical batch. FFFF. Finding: The retention time study is not being performed for the DRO surrogate, o-terphenyl (OTP). Requirement: Make three injections of all single component standard mixtures and multi- component analytes (such as PCBs) over the course of a 72-hour period. Serial injections or injections over a period of less than 72 hours may result in retention time windows that are too tight. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 D. Rev 6, March 2018, Section 11.6.1. Comment: The laboratory includes OTP in the retention time check standard but is not performing the study for OTP. GGGG. Finding: The retention time window width is not properly evaluated. Requirement: Width of the retention time window for each analyte, surrogate and major constituent in multi-component analytes is defined as ± 3 times the standard deviation of the mean absolute retention time established during the 72-hour period or 0.03 minutes, whichever is greater. Ref: EPA SW-846 Test Methods for Evaluating Solid Waste, Physical/Chemical Methods; 3rd Edition, Method 8000 D. Rev 6, March 2018, Section 11.6.4 Comment: The laboratory is currently using 3 times the standard deviation, which is often less than the 0.03 minute requirement. Too narrow a retention time window may result in low bias for early and late eluting components. Extractable Petroleum Hydrocarbons – MADEP, December 2019, Rev. 2.1 (Aqueous & Non- Aqueous) Comment: The laboratory analyzes samples utilizing dual autosamplers, columns and detectors. Each autosampler, column, detector and calibration curve is specific to a particular hydrocarbon fraction. The next revision of the method SOP needs to include this information. Recommendation: The laboratory is performing multiple CCV analyses from the same vial, rather than using fresh aliquots from non-punctured vials. This can cause bias due to loss of solvent and/or more volatile compounds. It is recommended that fresh aliquots be used to avoid potential QC failures. HHHH. Finding: The laboratory is not consistently preparing a Laboratory Control Standard Duplicate (LCSD) with each batch. Page 34 #402 Waypoint Analytical - Charlotte Requirement: At a minimum, for each extraction batch (up to 20 samples of similar matrix), an LMB, LCS and an LCS Duplicate must also be prepared and results analyzed as part of the laboratory’s continuing QC program. Ref: MADEP, Method for the Determination of Extractable Petroleum Hydrocarbons (EPH), December 2019, Rev. 2.1, Section 10.1.3.2. Comment: The laboratory is preparing and analyzing an MS/MSD pair as part of the routine QC. These are optional QC elements for this parameter. IIII. Finding: Samples are not being spiked with the recommended surrogates at the correct concentration during extraction. Requirement: Prepare a surrogate spiking solution which contains the surrogate standards at a concentration of 40 ng/µL in acetone or methanol. Each sample, LMB, LCS, and matrix spike is fortified with a specified volume of the surrogate spiking solution in order to yield a final concentration of 40 µg/mL (on column, prior to correction for preparation factors). The use of higher concentrations is permissible and advisable when spiking highly contaminated samples. Ref: MADEP, Method for the Determination of Extractable Petroleum Hydrocarbons (EPH), December 2019, Rev. 2.1, Section 7.5.5. Comment: The laboratory is currently fortifying samples and extracted QC samples with the surrogate solution at a final concentration of 20 µg/mL (on column, prior to correction for preparation factors). JJJJ. Finding: The laboratory is not analyzing a Continuing Calibration Standard (called CCV by the laboratory) for the breakthrough compounds at the proper frequency. Requirement: At a minimum, for each analytical batch (up to 20 samples) or every 24 hours, whichever comes first, a beginning and ending Continuing Calibration Standard must be analyzed. Ref: MADEP, Method for the Determination of Extractable Petroleum Hydrocarbons (EPH), December 2019, Rev. 2.1, Section 10.1.3.1. Comment: The laboratory is only analyzing a CCV for naphthalene and 2-methylnaphthalene at the beginning of the analytical run. KKKK. Finding: The laboratory is not properly preparing and evaluating the fractionation efficiency of each lot of silica gel cartridges. Requirement: For each analyte included in the Fractionation Check Solution (FCS), excluding n-nonane, the mean percent recovery for four (4) replicate samples, expressed as a percentage of the true value, must be between 40 and 140%. Lower recoveries are permissible for n-nonane. If recovery of n-nonane is < 30%, the source of the problem should be found and the fractionation check repeated. Ref: MADEP, Method for the Determination of Extractable Petroleum Hydrocarbons (EPH), December 2019, Rev. 2.1, Appendix 5, Section 5.2. Comment: Current laboratory practice is to prepare a single silica gel cartridge with the FCS and proceed with the fractionation process. The extract is then analyzed. The hydrocarbon ranges and breakthrough compounds are then evaluated, but the individual hydrocarbons are not evaluated. Insufficient hexane volumes will cause low bias in the aliphatic fraction, while excess hexane will cause breakthrough of the aromatic compounds into the aliphatic fraction. Comment: It will be necessary to include each hydrocarbon analyte in the calibration curve for the associated hydrocarbon range in order to properly evaluate the fractionation efficiency. Page 35 #402 Waypoint Analytical - Charlotte The laboratory may include naphthalene and 2-methylnaphthalene in the aliphatic curve rather than establishing a separate calibration curve for those analytes. LLLL. Finding: Surrogate compounds are not always integrated properly. Requirement: In samples, collective peak area integration for the hydrocarbon ranges, or TPH, must be from baseline (i.e., must include the unresolved complex mixture “hump” areas). For the integration of individual Target PAH analytes, surrogate compounds, and internal standards, a valley-to-valley approach should typically be used, though this approach may be modified on a case-by-case basis by an experienced analyst. In any case, the unresolved complex mixture “hump” areas must not be included in the integration of individual Target PAH Analytes, surrogate compounds, and internal standards. Ref: MADEP, Method for the Determination of Extractable Petroleum Hydrocarbons (EPH), December 2019, Rev. 2.1, Section 9.8.6. Comment: Current laboratory practice is to integrate surrogate compounds by projecting a baseline horizontally from the first valley to the end of the peak. With samples that have high concentrations due to the unresolved complex mixture areas, this practice includes area under the peak that is not part of the actual surrogate peak. Manually integrating the surrogate peaks from valley to valley will ensure that surrogate recovery will not be improperly biased high due to the matrix of the sample. Volatile Petroleum Hydrocarbons – MADEP, Feb 2018, Rev. 2.1 (Aqueous & Non-Aqueous) MMMM. Finding: The laboratory analyzes and evaluates four (4) LCSs when performing Initial Demonstrations of Capability (IDOC). Requirement: Prepare and analyze seven (7) replicate Laboratory Control Samples (LCSs) fortified at a concentration of 50% of the highest calibration curve standard (100 µg/L for aqueous samples and 5 mg/kg for soil/sediment samples). Ref: MADEP, Method for the Determination of Volatile Petroleum Hydrocarbons (VPH), February 2018, Rev. 2.1, Appendix 7, Section 3.0. NNNN. Finding: The laboratory is not within the acceptance criteria for some of the hydrocarbon range retention time (RT) window markers. Requirement: The RT window of the C5-C8 aliphatic hydrocarbons is defined as beginning 0.1 minutes before the elution of n-pentane and ending 0.01 minutes before the elution of nonane. The C9-C12 aliphatic hydrocarbon range begins 0.01 minutes before the elution of nonane; therefore there is no overlap of the two ranges and the nonane peak is only included in the C9-C12 aliphatic hydrocarbon range. The C9-C12 aliphatic hydrocarbon range ends 0.1 minutes before the elution of naphthalene. The Rt window for the C9-C10 aromatic hydrocarbons is defined as beginning 0.1 minutes after the elution of o-xylene and ending 0.1 before the elution of naphthalene. Ref: MADEP, Method for the Determination of Volatile Petroleum Hydrocarbons (VPH), February 2018, Rev. 2.1, Section 9.3.6. Comment: One of the recent chromatograms examined during the inspection had the C5 - C8 Aliphatic RT marker greater than 0.01 minutes before the n-Nonane peak. Also, the RT marker for the end of the C9 - C10 Aromatic range was greater than 0.1 minutes from the Naphthalene peak. Page 36 #402 Waypoint Analytical - Charlotte IV. PAPER TRAIL INVESTIGATION: The paper trail consisted of comparing original records (e.g., laboratory benchsheets, logbooks, etc.), chain of custodies and client reports. Representative data and associated client reports were submitted prior to the inspection for all parameters analyzed during the months of July, September and November 2021. No transcription errors were observed. The facility appears to be doing a good job of accurately transcribing data. V. CONCLUSIONS: We are concerned with the Findings that were cited previously and not corrected. Correcting the above- cited Findings and implementing the Recommendations will help this laboratory to produce quality data and meet Certification requirements. The inspector would like to thank the staff for their assistance during the inspection and data review process. Please respond to all Findings and include supporting documentation, implementation dates and steps taken to prevent recurrence for each corrective action. Report prepared by: Michael Cumbus Date: April 14, 2022 Report reviewed by: Jill Puff Date: April 18, 2022 Certificate Number:402 Effective Date:1/1/2022 Expiration Date:12/31/2022 Lab Name:Waypoint Analytical - Charlotte Address:449 Springbrook Road Charlotte, NC 28217 North Carolina Wastewater/Groundwater Laboratory Certification Certified Parameters Listing Date of Last Amendment:2/2/2022 The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below. CERTIFIED PARAMETERS INORGANIC ACIDITY SM 2310 B-2011(Aqueous) ALKALINITY SM 2320 B-2011 (Aqueous) BACTERIA - COLIFORM FECAL IDEXX Colilert ®18 (MPN) (Aqueous) SM 9221 E-2014 (MPN) (Aqueous) SM 9221 E-2014 (MPN) (Biosolids) SM 9222 D-2015 (MF) (Aqueous) SM 9222 D-2015 (MF) (Biosolids) BACTERIA - E. coli IDEXX Colilert® (24 hr) (MPN) (Aqueous) IDEXX Colilert-18® (MPN) (Aqueous) BOD SM 5210 B-2016 (Aqueous) BROMIDE EPA 300.0, Rev. 2.1, 1993 (Aqueous) CBOD SM 5210 B-2016 (Aqueous) CHLORIDE EPA 300.0, Rev. 2.1, 1993 (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Non-Aqueous) SW-846 9056 A (Aqueous) SW-846 9056 A (Non-Aqueous) CHLORINE, TOTAL RESIDUAL SM 4500 Cl G-2011 (Aqueous) COD SM 5220 D-2011 (Aqueous) Hach 8000 (Aqueous) COLOR, PC SM 2120 B-2011 (PtCo) (Aqueous) CONDUCTIVITY EPA 120.1, Rev. 1982 (Aqueous) SM 2510 B-2011 (Aqueous) SW-846 9050 A (Aqueous) CYANIDE SW-846 9014 (Total) (Colorimetric) (Aqueous) SW-846 9014 (Total) (Colorimetric) (Non-Aqueous) SM 4500 CNˉ E-2016 (Total) (Aqueous) DISSOLVED OXYGEN Hach 10360-2011, Rev. 1.2 (Aqueous) SM 4500 O G-2016 (Aqueous) FLUORIDE EPA 300.0, Rev. 2.1, 1993 (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Non-Aqueous) SW-846 9056 A (Aqueous) SW-846 9056 A (Non-Aqueous) HARDNESS TOTAL - METALS SM 2340 B-2011 (Ca EPA 200.7, Rev. 4.4, 1994 + Mg EPA 200.7, Rev. 4.4, 1994) (Aqueous) HARDNESS TOTAL - WET CHEM SM 2340 C-2011 (Aqueous) IGNITABILITY SW-846 1030 (Non-Aqueous) INORGANIC PHENOLS EPA 420.1, Rev. 1978 (Aqueous) SW-846 9065 (Aqueous) MBAS SM 5540 C-2011 (Aqueous) NITROGEN, AMMONIA SM 4500 NH3 G-2011 (Aqueous) SM 4500 NH3 D-2011 (Aqueous) SM 4500 NH3 C-2011 (Aqueous) SM 4500 NH3 C-2011 (Non-Aqueous) NITROGEN, NITRATE (NO3 + NO2 EPA 353.2, Rev. 2.0, 1993) - (NO2 SM 4500-NO2ˉ B-2011) (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Non-Aqueous) This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions as set forth in 15A NCAC 2H.0807. Certificate Number:402 Effective Date:1/1/2022 Expiration Date:12/31/2022 Lab Name:Waypoint Analytical - Charlotte Address:449 Springbrook Road Charlotte, NC 28217 North Carolina Wastewater/Groundwater Laboratory Certification Certified Parameters Listing Date of Last Amendment:2/2/2022 The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below. CERTIFIED PARAMETERS SW-846 9056 A (Aqueous) SW-846 9056 A (Non-Aqueous) NITROGEN, NITRITE SM 4500-NO2ˉ B-2011 (Automated) (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Non-Aqueous) SW-846 9056 A (Aqueous) SW-846 9056 A (Non-Aqueous) NITROGEN, NO3 + NO2 SM 4500 NO3ˉ F-2016 (Aqueous) EPA 353.2, Rev. 2.0, 1993 (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Non-Aqueous) SW-846 9056 A (Aqueous) SW-846 9056 A (Non-Aqueous) NITROGEN, TOTAL KJELDAHL EPA 351.2, Rev. 2.0, 1993 (Aqueous) OIL & GREASE EPA 1664 Rev. B (Aqueous) SW-846 9071 B (Non-Aqueous) PAINT FILTER LIQUIDS SW-846 9095B pH SM 4500 H+B-2011 (Aqueous) SW-846 9040 C (Aqueous) SW-846 9045 D (Non-Aqueous) PHOSPHATE, ORTHO SM 4500 P E-2011 (Aqueous) PHOSPHORUS, TOTAL SM 4500 P F-2011 (Aqueous) SM 4500 P E-2011 (Aqueous) EPA 365.1, Rev. 2.0, 1993 (Aqueous) RESIDUE, DISSOLVED 180 C SM 2540 C-2015 (Aqueous) RESIDUE, SETTLEABLE SM 2540 F-2015 (Aqueous) RESIDUE, SUSPENDED SM 2540 D-2015 (Aqueous) RESIDUE, TOTAL SM 2540 B-2015 (Aqueous) SILICA - METALS EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SULFATE EPA 300.0, Rev. 2.1, 1993 (Aqueous) EPA 300.0, Rev. 2.1, 1993 (Non-Aqueous) SW-846 9056 A (Aqueous) SW-846 9056 A (Non-Aqueous) SULFIDE SM 4500 S²ˉ F-2011 (Aqueous) TEMPERATURE SM 2550 B-2010 (Aqueous) TURBIDITY EPA 180.1, Rev. 2.0, 1993 (Aqueous) SM 2130 B-2011 (Aqueous) METAL ALUMINUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) ANTIMONY EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions as set forth in 15A NCAC 2H.0807. Certificate Number:402 Effective Date:1/1/2022 Expiration Date:12/31/2022 Lab Name:Waypoint Analytical - Charlotte Address:449 Springbrook Road Charlotte, NC 28217 North Carolina Wastewater/Groundwater Laboratory Certification Certified Parameters Listing Date of Last Amendment:2/2/2022 The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below. CERTIFIED PARAMETERS SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) ARSENIC EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) BARIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) BERYLLIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) BORON EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) CADMIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) CALCIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) CHROMIUM HEXAVALENT SM 3500-Cr-B-2011 (Aqueous) SW-846 7196 A (Aqueous) SW-846 7196 A (Non-Aqueous) CHROMIUM TOTAL EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) COBALT EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) COPPER EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) IRON EPA 200.7, Rev. 4.4, 1994 (Aqueous) This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions as set forth in 15A NCAC 2H.0807. Certificate Number:402 Effective Date:1/1/2022 Expiration Date:12/31/2022 Lab Name:Waypoint Analytical - Charlotte Address:449 Springbrook Road Charlotte, NC 28217 North Carolina Wastewater/Groundwater Laboratory Certification Certified Parameters Listing Date of Last Amendment:2/2/2022 The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below. CERTIFIED PARAMETERS SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) LEAD EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) MAGNESIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) MANGANESE EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) MERCURY EPA 245.1, Rev. 3.0, 1994 (Aqueous) SW-846 7470 A (Aqueous) SW-846 7470 A (Non-Aqueous) SW-846 7471 B (Non-Aqueous) MOLYBDENUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) NICKEL EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) POTASSIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) SELENIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) SILVER EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) SODIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions as set forth in 15A NCAC 2H.0807. Certificate Number:402 Effective Date:1/1/2022 Expiration Date:12/31/2022 Lab Name:Waypoint Analytical - Charlotte Address:449 Springbrook Road Charlotte, NC 28217 North Carolina Wastewater/Groundwater Laboratory Certification Certified Parameters Listing Date of Last Amendment:2/2/2022 The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below. CERTIFIED PARAMETERS SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) STRONTIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) THALLIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) TIN EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) TITANIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) VANADIUM EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) ZINC EPA 200.7, Rev. 4.4, 1994 (Aqueous) SW-846 6010 D (Aqueous) SW-846 6010 D (Non-Aqueous) EPA 200.8, Rev. 5.4, 1994 (Aqueous) SW-846 6020 B (Aqueous) SW-846 6020 B (Non-Aqueous) ORGANIC 1,2, DIBROMOETHANE (EDB) SW-846 8011 (Includes DBCP) (Aqueous) BASE NEUTRAL/ACID, ORGANICS EPA 625.1 (Aqueous) SW-846 8270 E (Aqueous) SW-846 8270 E (Non-Aqueous) EXTRACTABLE PETROLEUM HYDROCARBONS MADEP, December 2019, Rev. 2.1 (Aqueous) MADEP, December 2019, Rev. 2.1 (Non-Aqueous) NONHALOGENATED VOLATILE, ORGANICS SW-846 8015 C (Aqueous) SW-846 8015 C (Non-Aqueous) PESTICIDES, ORGANOCHLORINE EPA 608.3 (Aqueous) SW-846 8081 B (Aqueous) SW-846 8081 B (Non-Aqueous) POLYCHLORINATED BIPHENYLS (PCBs) EPA 608.3 (Aqueous) SW-846 8082 A (Aqueous) SW-846 8082 A (Non-Aqueous) SW-846 8082 A (Oil) This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions as set forth in 15A NCAC 2H.0807. Certificate Number:402 Effective Date:1/1/2022 Expiration Date:12/31/2022 Lab Name:Waypoint Analytical - Charlotte Address:449 Springbrook Road Charlotte, NC 28217 North Carolina Wastewater/Groundwater Laboratory Certification Certified Parameters Listing Date of Last Amendment:2/2/2022 The above named laboratory, having duly met the requirements of 15A NCAC 2H.0800, is hereby certified for the measurement of the parameters listed below. CERTIFIED PARAMETERS PURGEABLE, ORGANICS EPA 624.1 (Aqueous) SM 6200 B-2011 (Aqueous) SW-846 8260 D (Aqueous) SW-846 8260 D (Non-Aqueous) TPH DIESEL RANGE, ORGANICS SW-846 8015 C (Aqueous) SW-846 8015 C (Non-Aqueous) TPH GASOLINE RANGE, ORGANICS SW-846 8015 C (Aqueous) SW-846 8015 C (Non-Aqueous) VOLATILE PETROLEUM HYDROCARBONS MADEP, Feb 2018, Rev. 2.1 (Aqueous) MADEP, Feb 2018, Rev. 2.1 (Non-Aqueous) This certification requires maintance of an acceptable quality assurance program, use of approved methodology, and satisfactory performance on evaluation samples. Laboratories are subject to civil penalties and/or decertification for infractions as set forth in 15A NCAC 2H.0807.