Loading...
HomeMy WebLinkAboutNC0081825_Cl2 Meter Calibration_20170711WN July 11, 2017 ANS . NVILLE Chad Turlington Environmental Specialist Division of Water Resources 225 Green Street, Suite 714 Fayetteville, NC 28301 RECEIVED DEQ/DWR JUL 1 8 2017 S FAYETTEVILLE�REGIONAL OFFICE Re: Chlorine Meter Calibration Curve Verification Dear Mr. Turlington: Enclosed please find the Chlorine Meter Calibration Curve Verification for the Town of Ansonville. If you have any questions, please contact Sincerely, Dianna McLaughlin Town Clerk Town of Ansonville me at 704-826-8404. PO 10OX 437 ANSONVILLE, NORTH CAROLINA © 28007 a (704) 826-8404 Chlorine Meter Calibration Curve Verification Date: 7/5/2017 Meter SIN: Analyst: Diana McLaughlin Meter Program #: Hach 86 ULR Lab Name: ANSONVILLE NC WW/GW Certificiation # 5242 Instructions: Fill in the information above. Enter the actual concentrations of the standards in your curve in cells A19 through A23. Enter the values that you observe on the meter in cells B19 through B23. The % Recovery and the Correlation Coefficient will be automatically calulated. For the curve to pass... The Correlation Coefficient must be >_ 0.995 and... Standard concentrations Tess than 50 pg/L must have a % Recovery between 75 -125 %. and... Standard concentrations of 50 pg/L or higher must have a % Recovery between 90 -110 %. Actual, pg/L Observed,-pg/L % Recovery Std Pass/Fail slope intercept 400 404 101 Passed 1.014 -5.125 200 190 95 Passed 100 95 95 Passed 50 48 96 Passed 25 23 92 Passed Corr. Coefficient = 1.000 Corr. Coefficient Gel/Sealed Liquid Standard Observed Concentration = Gel/Sealed Liquid Standard Manufacturer: HACH Meter Program: 86 Passed Lot #: p.g/L A0349