Loading...
HomeMy WebLinkAboutNCD003200383_19930917_Koppers Co. Inc._FRBCERCLA SAP QAPP_QAPP Manual - Replacement Pages Appendix A-2-OCRg ·-R I I I I I I ·-I I I I I I I • I I KIJPPERS COMPANY. INC. NPL SITE LAB QUALITY ASSURANCE MANUAL -ATT. A MARDI 1993 I •• I I I I I I I -I I I I I Vj = = DF = The aliquot size (such as injection amount) of the - - prepared sample, taken for analysis, in units of mL The total final v_olume of the prepared sample ( extract volume, digest volume, etc.) in units of mL The amount of sa:f'e taken for preparation. For liquid samples, use ; for solid samples use the weight in kg. If the results are to be determined on the basis of dry weight, use the following to determine the sample size: As (dry) = As (wet) x %solids 100 Dilution factor. The dilution factor is 1 for samples that are prepared exactly as prescribed in the protocol. If the extract or digestate require dilution, then the dilution factor differ from unity. For example, if an extract is diluted from 1 mL to 100 mL the dilution factor becomes 100. If an extract is concentrated from 10 mL to 1 mL the dilution factor becomes 0.1. In many methods, the data reduction is computerized, alleviating the need for extensive manual reduction of the results. , However, the analyst must review the data, relating it back to the fundamental measurements on which the analyses are based. Thus, in chromatographic techniques, the analyst will compare area counts of peaks with those of the corresponding standard, and verify that the data were in fact correctly reduced. It is also the responsibility of the analyst to verify the identification of parameters. 6.3 Data Validation Data validation within each analytical group has been discussed previously. It is not the responsibility of the QA/QC personnel to check and verify every value generated and reported by the laboratory; however, the QA/QC Section will function in a reviewing capacity to assure that the data validation process remains reliable and is performed according to laboratory policy. \W3J237-aJ/9'! 6-7 I •• I I I I I I I • I I I I I I I •• I These items must be verified during the data validation process: a. Is the batch complete? b. c. d. e. f. g. h. i. J. Have all the analyses been performed within holding times, and if not is there an acceptable explanation for deviations from the holding times? Is there a valid continuing calibration associated with the runs for each parameter of the individual samples within the batch? Is the sample sequence proper for the method (i.e., are there regular blanks run when necessary, are there standards run for methods requiring periodic standards, are there laboratory blank spikes, mati-ix spikes, and laboratory duplicates)? If surrogates are required by the method, are recoveries within the control limits in the spiked blank? Are surrogate recoveries within control limits in the samples? Is the recovery of spiked compounds in the laboratory spiked blank acceptable? Is the recovery of spiked compounds in the spiked sample acceptable? Do duplicate analyses in the run sequence exhibit precision within the control limits? Is the documentation in order? If the answers to all the questions above are "yes", the data will be released and reported . W83J237 -OJ /93 6-8 I I I I I I I - I I I I I I I 1• I If the· answer to··· any of · the questions is '· negative, ··corrective · action will be undertaken and reported to the QA/QC Section and the Project Manager. A Corrective Action Notice (Figure 5-2) will ·be generated and submitted to the QA/QC Section by the Group Leader, thus documenting the source of the problem and its resolution. Data Validation is a check of the batch for completeness, ac=acy and precision. If all criteria are met, ,the batch is released. If some are not met, the batch may still be released, depending upon what criteria are not met and what is the explanation for not meeting the criteria. There are certain circumstances when the release of the data is not allowed regardless of the rationale for the lack of acceptability. These are: a. b. C. Continuing calibration was either not performed as required, I Laboratory control standards (spiked blanks) were not run or did not meet acceptance criteria Data sets are not complete Corrective action for these situations will be discussed in Section 7.0. If certain aspects of the data do not meet acceptance criteria, but consultation with the project manager or the Group Leader res\ilted in a decision to release the data, the Quality Assurance Manager will annotate on the Corrective Action (Figure 5-2) the shortcomings_ of the data, the decision to :proceed with reporting the data, and the person with whom that decision had been:reached. The QA Manager will then submit the form to the Information Services Section so that the final report can be prepared. If the decision has been made not to release the data, the form will be marked "DO NOT RELEASE DATA" and submitted to the Information Services Section. In addition to validating the data, it is also necessary to validate specific jobs before they are released. Specifically, releasing a job consists of insuring that all required parameters have been analyzed and are reported. The result for each parameter W&1,1Z17. aJ/93 6-9 I ~-I I I I I I I .. I I I I I I ·· · · must have been acceptably validated through approval of the individual batches that· make up all of the parameters associated with the job. It is the responsibility of the Information Services .Section in conjunction with the Project Manager to assure that all data are validated and entered and that the documentation is complete before the final report is generated. The QA/QC Section will only spot check the final reports. 6.4 Data Compilation Information Services personnel will review the job file to confirm that all the data are in .the file and references are made to those items which are associated with the job but are present in other files. The data will then be entered in standardized forms, and the report printed out . 6.5 Final Review When the final report has been printed, the Project Manager will proof the report against the raw data in the file. If all the data are properly entered, the report will be reviewed by the Lab Director and approved before it is sent to the client. A copy of the report will be placed in the job file, and the file will be transferred to the ' central file storage for completed jobs. Wll.12J7. OJ/93 6-10 I I I I I I I .. I I I I I I I 7.0 · CORRECTIVE ACTION . There are many areas of the laboratory functions which may require corrective action. The decision to undertake corrective action, and the ensuing action must be documented so that traceability can be maintained. The point or originating the corrective action varies, depending upon the mode of detecting that such action is necessary. It is, however, frequently the role of the QA/QC Section to initiate such action, simply because it is this section whic~ is most exposed to the malfunctions of the laboratory as they reflect upon the data produced. Those actions that affect the quality of the data will be recorded and the record maintained by the QA/QC Section. 7.1 Identification of Potential Problem Identification of sources of problems is not always an easy matter. It is not expected that the QA/QC Section will be able to trace the source of every problem identified. However, the QA/QC Section will be responsible for informing the Section Manager or Group Leader and the Laboratory Director when a problem exists in a particular analysis. In this case, data for that type of analysis will not be accepted until the problem is isolated and corrected. It will be the responsibility of the Group Leader to address the identification of the source of the problem, and the responsibility of the Laboratory Director to assure that the Group Leader is acting upon the need for corrective action. In some situations, the need to correct an operation is apparent to the analyst. For example, instrument failures are detected by 'the analyst and the corrective action is taken in the form of repairing the instrument either through a service call or using laboratory personnel. Such action must be recorded in the instrument maintenance log and close scrutiny will be paid to the analyses just preceding the instrument failure. If these analyses meet the required acceptance criteria, no further action will be taken relative to that data. WKJJ231-aJfll 7-1 I I I I I I I - I I I I I I I f' I In other situations, the time of data validation is the more logical time to isolate a potential problem. For example, if a systematic drift occurs due to chromatographic column contamination, it is easier to identify at the time of validating the data and comparing the results with historically available information. 7.2 Problems and Actions 7.2.1 Continuing Calibration Outside Acceptance Range When the continuing calibration is outside the acceptance range, the problem should be identified by the analyst and corrected before further sample processing is undertaken. However, the acceptability of the continuing calibration is also subject to review at all later checks in the data validation process. The data on all samples that have been analyzed following the last time that the calibration was within specification will be rejected. (If data have been released to Information Services, the QA/QC Section will be notified i=ediately using the Form shown in Figure 5-2.) The affected samples will be reanalyzed after the new initial calibration curve has been constructed. 7.2.2 Calibration Standards Exceeding the Permitted Holding Time If calibrations s~andards have been continuously used beyond their permitted shelf- life, the group responsible for the analysis will be notified, and a Corrective Action Notice (Figure 5-2) will be generated. The group will then be responsible for preparing fresh calibration standards, and the instrument calibration will be checked against the new standards. If the previous runs, performed with the expired standards meet the acceptance criteria based on the new standard, the data generated will be considered valid, in spite of the use of an expired calibration standard. W83.1Da • OJ/93 7-2 I ~-I I I I I I I .. I I I I I I I ~ If the calibrations performed with the expiriid standard do.not meedhe acceptance - criteria when measured against the new standards, the_ samples that have been analyzed against the expired standard will be reanalyzed and quantified using the freshly prepared standards. 7.2.3 Laboratory Method Blanks Exceed Method Detection Limit but are Below Quantitation Limit When laboratory blanks exhibit the presence: of target analytes at a level exceeding the method detection limit, but still below the quantitation limit, the responsible group will be notified. The.responsible section will check the reagent blanks that have been retained at the time o_f use of the reagents in order to determine if contamination or interferences are due to impurities in the reagents. If this' is the case, the reagent batch will be discarded, and new reagents will be used . If the reagents appear to be sufficiently 'pure, the cleanliness in laboratory procedures will be reviewed to establish if the source of problems may have been contamination of the apparatus. 7.2.4 Laboratory Method Blank Exceeds the Quantitation Limit When the laboratory method blank exceeds the quantitation limit, the Group responsible for the analysis will be notified. As discussed in Section 7.2.3, the analysts will check for potential contamination of reagents and apparatus. If the reagents are contaminated, the existing reagent batch will be discarded and a fresh batch from new containers will be prepared. If the problem arose from the apparatus or glassware, the problem will be corrected within the analytical group, and the correction documented before any further analyses can be undertaken. Notification of the corrective action will be submitted by the Group Leader to the QA/QC Section. WIJJ.231-(CJ rn 7-3 H I I I I I I .. I I I I I I I ,, I Toe data associated with the failed method blank will not be accepted. Toe samples will be reanalyzed to produce acceptable data. 7.2.5 Laboratory Control Standard Exhibits Recoveries Outside the Acceptance Criteria When the laboratory control standard (spiked blank) does not meet the acceptance criteria, the batch of samples associated with the laboratory control standard will be reanalyzed if determinative methods call for this as corrective action. Where, as in many organic methods, a check sample is analyzed only for failed spiked sample components, then reanalysis of the batch will also be necessary, but only for the failed-components. In methods that require the use of a laboratory control standard at frequent intervals through the day, only those samples analyzed since the last acceptable laboratory control standard analysis will require repeating. Data originating from batch analyses preceding the failed QC check will be rejected . Before repeating a whole set of preparations of samples, the calibration of the instrument or method shall be checked. If the instrument is within calibration, the samples will require repreparation. If the instrument calibration has drifted, the prepared samples from the initial preparation can be reanalyzed after the instrument has been recalibrated. 7.2.6 Surrogates and Sample Spikes Exhibit Recoveries Outside the Acceptance Limits When recoveries from spiked samples are outside the acceptance limits, but the laboratory spiked blank is within the acceptance criteria, the poor recovery or enhanced apparent recovery may be due to matrix effect. One such sample from the batch will be reprepared and reanalyzed. If, the same phenomenon is observed, it will be assumed that the failure to meet recovery criteria was, in fact, a matrix effect. This information will be included in the report to the client; however, the original data will be accepted. W&1J23a -IJ3/9'l 7-4 I I I I I I I .. I I I I I I I '9 I 7.2. 7 Control Chart Exhibits a Regular Trend By their very nature,. the individual points that make up the control chart for any analyte vary randomly about a mean value. The control chart is used to assess the . ' · acceptability of recovery data on the basis of historical data. The control chart is also used to warn the analyst that some consistent problem or deviation in the method may be occurring. When five successive points on the control chart form a steady pattern, either regularly increasing or regularly decreasing, they imply that some change is occurring in the analytical scheme. Even if all the points are within the control limits_, a warning will be issued by the QA/QC Section to the responsible group to investigate the cause of the pattern. If, in fact, a change has occurred in the method, and if ·the change indicates an improvement in recoveries (an improvement is defined as approaching complete recovery, not necessarily an upward trend), then a new control chart will be established, and sub'sequent data will be compared to the new control chart limits. If a change has occurred that worsens the recovery, it will be the responsibility of the Group Leader to assure that a return to the previously used technique is made in his section. 7.2.8 Internal and External Audits and Corrective Actions Internal system ~udits are quarterly samples issued by the QA/QC Section, using known spiked solutions to determine the acceptability of performance in every group of the laboratory. External performance evaluation is either done through a contract required performance audit, or though voluntary participation in interlaboratory studies. 7.2.8.1 Evaluation of System Audits When the achieved results on these audits fall below acceptable standards, as defined on the basis of historical recovery data, a thorough review of the system will be initiated. The QA/QC Section is responsible for the initiation of the process. W!3JZ31 • CD/9] 7-5 r I ~-I . I I I I I I - I I I I I I I . The first steiwill consist of complete review of the correctness of the documenting of the job and the calculations of the results. This process will be performed by the personnel of the QA/QC Section. When the results of the review are complete, .a memo will be issued to the responsible group of the laboratory, itemizing the deficiencies that have been identified. If no deficiencies have been identified, the difficulty may be with the performance of the analyst, the analytical method, or incorrect preparation of the audit sample. 7.2.8.2 Corrective Action and Feedback The gr.oup leaders will be responsible for investigating problem areas identified by the QA/QC Section. A written report of the findings and corrective action taken will be submitted to the QA/QC Section within one week of the beginning of the Group· Leader's investigation into the problem. If the source of the problem has not been located at this point, the QA/QC Section will issue a repeat performance evaluation audit sample. It will be assumed that the problem was with the original audit sample itself if the repeat evaluation sample is successfully analyzed within acceptance limits. Otherwise, the process outlined. above will continue until the manager of the QA/QC Section determines •that the problem has been corrected and notifies the Group Leader of the successful resolution. 7.2.8.3 Documentation of Corrective Action When the cause for failed internal or external audit samples has been identified and corrective action implemented, the QA/QC Section will document the resolution of the problem in a brief report to laboratory management. The QA/QC Section will also maintain a file of corrective actions which were necessary to remedy performance deficiencies which have arisen. Wl],1231-aJ/93 7-6 ... --,-•····. - I '• I I I I I I I • I I R I I I I •• I 5.11 Development of New or Modified Methods There are occasions when it is necessary in the laboratory either to devise a new approach to analysis or modify an existing method. The former may be needed for analytes for which proven methods do not exist, or if they do, they are not applicable to the matrix being handled. The latter case is frequently the· situation that arises because of unusual matrix interferences. It is not the intent of this manual to prescribe analytical approaches to analytes for which methods do not yet exist. . However, in order that the laboratory may use the methods with any reliance on the data, the new method must be subjected to the repetitive analyses of a spiked matrix in order to ascertain method precision and accuracy. No method will be employed by the laboratory without the establishment of precision, accuracy, and detection limits. When a method is being modified, or a new method is being devised and tested for anaiytes for which an existing approved method is available, a new or modified method will be subjected to equivalency testing. In performing the equivalency testing, the analytes will be subjected to seven replicate analyses in parallel by the existing method and the new method. To be acceptable, the new method must produce results that are statistically equivalent or are better than the old method. In addition, sufficient replicate analyses will be performed to assure that acceptance criteria may be established. The new or modified method will then be written, the method with the supporting data and documentation will be reviewed by the Section Manager for the section in which the analyses will be performed and by the QA/QC Manager. When approved, the method may become part of the repertoire of the laboratory for clients whose work does not require method approval by EPA or other regulatory agencies. W!J, 1236 • 03/93 5 -30 I I H R I I I - I I I I I I I •• I 6.0 DATA HAi-..DLING The data produced by the various groups of the laboratory are ultimately the product which the laboratory offers. Hence, not only is it necessary to produce results with accuracy and precision, but it is also necessary to be able to maintain the traceability of the data and the association of sets of data with each other. The responsibilities to the production of accurate and precise data are with the individual analysts. They are, after all, the producers of the data. No amount of supervision and validation can correct for mistakes or omissions occurring at the bench. At best, supervision and validation can cull the unsupported data. Tracea~ility of the data, however, pertains to the manner in which the records are kept within the laboratory as a whole. Because of this, record keeping in the laboratory, will be addressed first. This will be followed by a brief discussion of data reduction, data validation, data review and release, and reporting. 6.1 Data Recording The modes of maintaining data in the various groups will differ with the methods of analysis. Certain aspects of record keeping have been addressed in the section pertaining to the sample log in and distribution of the information through the laboratory. The traveller and the sample transfer forms constitute the first step in the generation of data for any set of samples. Samples in the laboratory, however, are analyzed in batches. On large jobs, a batch may constitute just a portion of the total number of samples in the job. For small job, a batch may consist of samples from several travellers handled together. Thus, the maintenance or records in the laboratory must also provide for cross-referencing batches and travellers. 6.1.1 Notebooks Certain records are maintained in notebooks by the analysts. These records may pertain to methods that are entirely manual, such as many of the methods in wet chemistry, or they may be used to record unusual observations during the performance of preparation and analyses of instrumental techniques. These W&].1217 • <D/93 6-1 u .. I I I I I I I • I I I I I I I •• I notebooks become a permanent record of laboratory work, and they must be traceable. Bound hardcover notebooks, with prenumbered pages, will be numbered and issued by the QA/QC section. The QA/QC section will number the notebook, and record in its own notebook log the date of issuance and the laboratory section to which the · notebook has been issued. When the notebook has been filled, it will be maintained in the analytical section for permanent archiving. I Users of the notebooks will maintain good laboratory practices in their use. No I pages will be tom out of the notebook; corrections will be done by single line-out erasure·s, followed by initialling and dating the correction; and each page of the notebook will be signed and dated by the analyst at the time entries are made on the page. The use of correction tape, liquid paper, erasures, or other means to make corrections is not permitted . 6.1.2 Record Keeping in Sample Preparation· All records for routine sample preparation wiU be kept on standardized forms. The forms will be filled by the person preparing the samples. The forms will be filled in ink, with no erasures. If an error occurs, the preparer will line out the erroneous entry once, and enter the correct entry. The preparer will then initial and date the correction. The following information must be entered: a. b. C. W&JZ37. 03(9J The analysis type for which the samples are prepared (usually the heading of the sheet) The date of preparation, and the name of the preparer The QC batch identification 6-2 I I I I I I I - I I I I I I I -d. ----A listing of the samples being prepared, using the laboratory's sample . identification e. f. g. h. -I. l· · For each sample listed, the quantity of sample taken, including the units of measurement (for solid samples, use the wet weight) If the method calls for the addition of surrogates, then for each sample the surrogate mixture identification and the quantity, including units For any sample that is spiked, the identification of the spiking mixture and the quantity, including units The identification of the medium of the prepared sample (i.e., the solvent for organic extraction, the acid for metal preparation, etc.) and the lot numbers of the reagent media The final volume of each prepared sample, including units Any unusual observations. If these are recorded in a notebook, the notebook and page numbers must be entered on the form. In addition to filling the form, the preparer must also label the prepared samples. As a minimum, the label will identify the sample and the QC batch in which the sample has been prepared. When the preparation is complete, the form will be checked for completeness by the Group Leader .and initialed and dated. The preparer will make as many copies of the filled form as there are different jobs in the batch. The original of the form will be maintained in the file of the preparing section. The preparer will then transfer the prepared samples with the accompanying copies of the preparation records to the section responsible for the analysis of the prepared samples. Examples of standard forms for maintaining the records in the sample preparation areas are shown in Figures 6-1 and 6-2. WKJ.1ZJ7 • 03/93 6-3 I FIGURE6-1 CONTINUOUS EXTRACTION BENCH SHEET l=xtracted by: ~eter _________ _ IJate Extraction Began. _______ _ Date Completed. _______ _ length of Extraction ~olvent, _____ _ Lot# __________ _ Manufacturer ___ --'------ Surrogate Matrix Spike Lot# Lot# Sample Sample Extract Work Order Number Source pH Volume Amount Added Amount Added ' tnalyst: ____________ _ Date: •• wed by: ___________ _ Date: I 6-3a I --.--- - a, I w O" Case Number (CLP only) SDG Number (CLP only) Work Order Sample Number Number - Lot Numbers: ------- ------- Sample Source H202 --I --FIGURE 6-2 CHESTER LABNET-MONROEVILLE AQUEOUS SAMPLE METALS DIGESTION Date of Dlgesllon ----- Time of Dlgesllon ----- lnllial Vol. Final Vol. Color Color (mis) (mis) Before After . HCI HNO3 ---------- Page N ----- Technician __________ _ Dlgesllon for (ICP/GFAA) ______ _ Clarlly Clarlly Before After Comments i i ' i I i i I ., n I I I I I I • I I I I I I I •• I - 6.1.3 Record Keeping for Instrument Analysis Many instrument m~thods of analysis produce printed traces, charts, or tables. Some of the information specified below is entered through a data system before samples are analyzed. Clearly, those items of information need not be reentered manually. For every technique, however, a run log must be maintained. The following information must be entered in the run log by the analyst. a. Identification of the analysis (usually this will be the heading of the log) b. Identification of the instrument c. Date of analysis, and identification of the analyst d. e. f. g. Identification of the QC batch Sequential listing of samples (including tuning, initial calibration or continuing calibration, blanks, spiked blanks, samples, duplicate samples, spiked samples, etc). This listing must be accurate and sequential, and the run number (where applicable) must be included. Aliquot of prepared sample that is taken for analysis, including units Any unusual observations. If these are entered in a notebook, they need only be summarized on the form, and the notebook and page numbers entered for further identification of the comments. ' Where printouts for each sample are obtained separately, the printout must include the identification of the method of analysis, date of analysis, identification of the ' analyst, identification of the run, identification of the QC batch, and the aliquot of sample that was used. W8JJ237. fl.3/93 6-4 I ~-I I I I I I I - I I I I I I I {' I Before·transferring the data further, the analyst must"confirm that··the results,-as---- they appear on the printout, are correct. If the results are incorrect, the analyst will line them out once, enter the correct val1;1e manually, and initial and date the change. These corrections will be done in ink. When the analyst is finished with the batch of samples, including the review of the data, he will initial the run log and transfer the run log with the data to the Group Leader. The Group Leader will review the run log to verify that the sample sequence has been properly followed as required by the specific method. Next, the Group Leader will check the QC data pertaining to the batch and verify that the instrument was performing within the required specificatiotjs. The Group Leader also reviews any dilutions to make sure that they are appropriate, assures that calculations are accurate, and verifies that reporting units are proper. The QA/QC Section will spot check ·data and provide guidance for data validation to the Group Leader. When the Group Leader is satisfied that the data are valid (see Section 6.3, Data Validation), he will initial the run log and date it. He will have as many copies made of the run log as there are jobs represented in the batch that has been analyzed. He will then transfer the sample data to Information Services. Together with the data, the Group Leader will transfer the copies of the records from the preparation group to the Information Servic,es Section. The original run log will be maintained in the files and records of the section responsible for the analysis. For instrument methods which do not produce the identifying code of the run on the permanent records, for example, methods that employ a strip chart as the instrumental output, the individual traces will be identified manually by the run number, and the identification of the parameters associated with the run will be recorded manually on the run log. W8.1J237. (JJf9l 6-5 I •• I I I I I I I - I I I I I I I •• I 6.1.4 . Record Keeping· in Noninstrument Analyses-. -... ' Raw data from these analyses will be kept in permanently bound, sequentially paginated notebooks and will include all the information discussed above in reference to automated printouts. Copies of this data will be submitted to the Information Services Section, and the original data will remain in the laboratory notebooks. 6.2 Data Reduction ReduciJ;lg the raw data to a presentable forin is the responsibility of the analyst performing the analysis. In reducing data,, the analyst must take into account whether the sample is aqueous or solid, the sample size, and whether or not the data are to be presented in the form of wet weight or dry weight. All final values must be accompanied with the units of the value. The following equation applies to the calculation for concentration of an analyte in most analytical techniques employed in the laboratory. Where: WB3.17n. 03('13 C I RF = = = Concentration of the analyte in the sample, in appropriate units [ng/L (ppt), ug/L (ppb), mg/L (ppm); ng/kg (ppt), ug/kg (ppb), mg/kg (ppm)] Signal size, in units appropriate to the method The response factor, as defined in Section 5.6.1, in units of signal size per unit weight of the analyte. This response factor is essentially a mean response factor, determined through regression of the initial calibration data. If is not possible to use the mean response factor ( usually because of instrumental software limitations), then the response factor from the run that corresponds to the midpoint of the linear range is used, as long as this response factor does not differ from the mean by more than 10%. · The response factor to be used is always taken from the initial calibration . 6-6 D a n B I I I .. I I I I I I I •• I 5.8.2 Spiked Blank The spiked blank, or laboratory control sample, serves as a measure of accuracy of the analytical procedure in the laboratory. ni,e spiked blank is prepared by adding prescribed amounts of specific analytes to laboratory reagent water prior to preparation of the water for analysis. For inorganic parameters, a spiked blank is prepared for each batch of twenty or fewer samples that are prepared at the same time. Water is spiked for certain analytes of interest for the inorganic parameters. For organic parameters, a spiked blank is pre~ared for every batch of 20 samples that is subjected to sample preparation. The spike contains only selected analytes which are specified in the pertinent methods of analysis in the Standard Operating Procedures. Preparation of the spiking mixture is done in the same manner as the preparation of standard solutions for calibration. The spiking mixture is also assigned an identifying code, which is recorded at the time of preparil).g the spiked blank. The spiked blank is carried through the entire analytical procedure, and the concentrations of the spiked analytes determined. These values are accumulated by the QA/QC Section to update acceptance criteria! and to validate data. The spiked blank forms the backbone of the determination of the reproducibility of data in the laboratory, I since it is based on a well-characterized matrix (water) and is designed to be essentially free of matrix effects. If the spiked blank does not meet established acceptance I criteria, the requirements of the analytical °;lethod will be relied upon to determine whether the sample failing the acceptance criteria will be reprepared and reanalyzed. 5.8.3 Spiked Sample Spiked samples serve to confirm that the sample matrix is causing certain effects which I preclude the ability to recover analytes using the prescribed method. Thus, the spiked sample is used only to determine matrix effects. One sample per batch of twenty or fewer samples, of the same apparent matrix, will be spiked with a spiking solution in the same manner as the spiked blank. The spiked W83,i236 • 03193 5 -22 I I I I I I I le I I I I I I I sample will be processed through the entire analytical scheme, and the recovery of the spiked analytes will be determined. In utilizing spiked sample information, great care should be taken because deviations from acceptability may be due either to procedure or to matrix effects. Generally, if the spiked blank associated with the batch exhibits acceptable recoveries, it can be assumed that the sample preparation and analysis have been performed correctly. Since sample spiking is performed before the sample is initially analyzed, it is possible that for some parameters the analyte level in the sample is so high that the spiked amount is insignificant. Under those circumstances, the analyte spike recovery will be meaningless and need not be calculated. In selecting a sample for spiking, every effort ,will be made to choose a field sample, and not one of the field or trip blanks. This ca~ be accomplished only if the identity of the field and trip blanks is known in advance. If those samples are submitted entirely ' as blind samples, then the selection of the sample to be spiked will be random. 5.8.4 Sample Duplicate Either sample or spiked sample duplicates are run to assess precision of the laboratory work. One in a batch of 20 or fewer samples of the same matrix will be prepared and analyzed as a sample or spiked sample duplicate. The precision exhibited by the analyses of sample or spiked sample duplicates may vary due to the effect of the matrix on the analyses. If that is the case, the related batch samples will not be reanalyzed unless the approved method calls for such action. As in the case of the spiked sample analysis, the choice of sample for sample or spiked sample duplicate analysis will be limited to actual field samples, excluding blanks, unless the identity of the blanks is not known. In that case, the selection will be random. W&3,1236-0J/9l 5 -23 I I I I I I I .. I I I I I I I ~ I 5.8.5 External Quality Control Audit An external quality control audit will be performed periodically by submitting for analysis known standard materials from a source other than that from which the calibration standards are prepared. These quality control audit samples will be I submitted for analysis by the QNQC Section, and they will be analyzed by the same procedures as are used for the analyses of samples. The external quality control audit will be performed on a quarterly basis. If there are projects in the laboratory which have these audit samples submitted by the project, and where the results are made known to the laboratory, then the QNQC Section will not submit the external quality control audit samples for those parameters which are includ_ed in project audit samples. 5.8.6 Record Keeping on Analysis of QC Samples The results of the analyses of the QC samples may pertain to more than one project, since the sample preparation area batches samples from potentially several sources to ' provide for efficient operation. Hence, the retrieval of the QC data may be necessary for several different projects. The QC data is identified by the batching identification provided upon the preparation of the samples. Hence, the retrieval of the data can be readily achieved by identification of the preparation batch. For every batch of samples, the identification of the batch will be entered at the preparation stage on the appropriate forms. Sufficient number of copies of the forms will be made to file with every project associated with the batch. The results of the analysis of the QC samples will be filed sequenti_ally by the QC batch number and will be retrieved through the batch number system whenever it is necessary to retrieve these data. The QC approval of an analyzed batch will be. based on the acceptance of the QC samples associated with the batch. Once the QC samples are determined to be within acceptable limits, the entire batch of samples will be released and the analytical data for each sample will be recorded with the appropriate project. W83, 1236 • 03193 5 -24 I I {' I I I I I I .. I I I I I I I r8 I 5.9 Use of Surrogates The use of surrogates in orgarnc analysis serves as an additional measure of the acceptability of the results. The significant advantage of the use of surrogates is in measuring recovery against a historically established acceptance range in the performance of each analysis. Thus, the data do not depend solely on the spiked blank to assess the quality of each run. Surrogates are compounds that are expected to behave analytically in a manner similar to the target analytes. The surrogates are added into the sample before the preparation stage is initiated, and their recovery is a measure of the efficiency of the extraction. The fqllowing surrogates have been used at CHESTER LabNet and are recommended for use in the approved methodology. W33.1236-0Jl93 5 -25 I D I I I I I II I I I I I I I ,. n TABLES-3 SURROGATE COMPOUNDS AND TYPICAL RECOVERIES EPA METHOD 501/601/8010 502/602/8020 604/8040 605/8050 606/8060 608/8080 609/8090 610/8310 611/8110 612/8120 615/8150 COMPOUND Bromochloromethane 2-Bromo-1-chloropropane 1,4-Dichlorobutane Benzotrifluoride 2-Fluorophenol 2,4,6-Tribromophenol 1-Naphthylamine Dibutylchlorendate Dibutylchlorendate 2, 4 ,5 ,6-Tetrachloro-m-xylene 2-Fluorobiphenyl Benzo( e )pyrene 2-Fluorobiphenyl Isodrin Dibutylchlorendate 2,4-DB RECOVERY RANGE 70 · 120 70 · 120 70 · 120 . 80. 120 21 • 100 10 · 123 10 · 94 24 • 154 24 · 154 25 · 125 43 -116 15 -113 40 · 116 33 · 141 24 · 154 40 -140 For GC/MS, standard EPA surrogate compounds will be used for all analyses. 5.10 Establishment of Acceptance Criteria Establishment of acceptance criteria is necessary in order to be able to determine whether or not quantitative data generated by the laboratory are within the control limits. The principal criteria that are used to measure the quality of the data are accuracy and precision. The initial determination of acceptance criteria hinges upon repetitive measurements of prepared spiked solutions and determination of spike recovery. Twenty samples of W&3,1236-03J9l 5 -26 I I I I I I I -I I I I I I I laboratory reagent water are spiked with the analytes of interest at a concentration of approximately twice the quantitation limit. Where applicable, the water is also spiked ' with surrogates and internal standards. The samples are then prepared for analysis following the appropriate protocol, and the concentrations of the analytes are determined. From these values, the mean and the standard deviation for the recovery of each analyte are determined. The deviation of the mean from the known spiked amount is a measure of the accuracy of the method. The standard deviation of the series of measurements is a measure of the precision of the method. The percent recovery is calculated as follows: R = 100 X Where: R = Percent recovery of the ana!yte. Cm = The measured concentration of the analyte C1 = The native concentration of the analyte in the sample (C1 = 0 for blank spike). Cs = The amount of analyte spiked into the sample. The accepted recoveries of the analytes must be, within three standard deviations of the mean. It should be emphasized that recoveries are dependent upon both the method of sample preparation and the sample matrix. Thus, recoveries from soil are not expected to be within the acceptance limits as determined for water. Similarly, extraction by sonication may ·not show the same recovery as would an extraction via a Soxhlet extractor. Thus, acceptance criteria must be determined matrix by matrix, and method by method. Frequently, samples are spiked at the time of sample preparation, without knowing if the analytes that are being spiked into the sample are present or not, and without knowing if these analytes are at levels that would make the spike amount insignificant. The analyst is cautioned that recoveries of spikes should not be calculated if the amount in the sample is greater than 5 times the spike_. For most applications, if the ratio WSJ.1236-03193 5 -27 I I I I I I I • I I I I I I I C1/Cs is equal to or greater than 5, the spike recovery should not be calculated, since the uncertainty in the native concentration is sufficient to cause greater uncertainties in the spike recovery. The acceptable prec_ision range is defined in a similar manner. The precision is a measure of the deviations from the mean of repetitive measurements. Thus, standard deviation can be used as a measure of precision. More frequently, the relative percent deviation will be used because, at best, measurements are performed in duplicate. The relative percent deviation is determined by the equation: Xl -x2 %RPD= 100 X ' X m · Where: x1 = High value for the_ analyte x2 = Low value for the analyte Xm = mean value for the. analyte x1 + x2 = 2 The results of the determination of recoveries and relative percent deviations will be ' plotted, indicating the upper and lower limits of acceptance, and the upper and lower warning limits. Future data will be considered acceptable if recoveries of spikes and relative deviations of duplicates fall within the acceptance criteria. The control charts will be updated regularly, generally requiring 20 data points to obtain a· valid control chart. The QC Section will maintain the control charts and update them regularly, when sufficient data are collected. Whenever the control charts are updated and the acceptance limits modified, the QC Section will issue the new limits to the analytical section of the laboratory in which the analyses are performed. It is the responsibility of the Group Leader to assure that data within their section are within the acceptance limits. If they are not, corrective action must be initiated immediately by the Group Leader of the appropriate section. If five successive measurements, while falling within the control limits, appear on the same side of the mean, the analyst will stop to investigate if the trend indicates that a W!J,1236 • 00/93 5 -28 I ~-I I I I I I I .. I I I I I I I change in methodology has occurred. Such· successive points may indicate a pattern, and it would be necessary to institute a return to preexisting conditions to avoid the possibility that out of controfsituations may arise. It is not feasible, in organic analysis, to obtain control charts for every analyte. Thus, while initial control charts are constructed for all the parameters that are being analyzed, continuous verification of the control is obtained through the use of surrogate compounds and the spiked blank analysis. All other analytes in the organic analysis will be assumed to be within control if their relative response to the internal standards and surrogate have remained constant. While accuracy and precision form the backbone of the acceptability of quantitative data, _9ualitative identification is more difficult to cast into quantitative measures. In organic analysis, the principal criterion for chromatographic analysis is the retention time, or relative retention time. Relative retention time is used with those methods employing internal standards. It is the more reliable measure because it is less dependent on such physical parameters as the length of the column. In all cases, the retention time for each analyte, or the relative retention time for the analyte, will be based on the data obtained from the nearest standard. To determine the acceptance windows for retention times, the continuing calibration data will be employed. For each compound, the retention times obtained in performing the continuing calibrations during the period of one week will be averaged and their standard deviation determined. The acceptance window will consist of three standard deviations from. the mean retention time for each compound. The retention time acceptance window will be redetermined whenever the chromatographic column is changed or the chromatographic conditions altered. It is the responsibility of the analyst to maintain the records for retention time criteria. Copies of the established acceptance windows will be available to the QC Section for reference in reviewing work. In mass spectrometric analysis, in addition to the retention time, the mass spectral match of the compound to the standard will be used to verify its identity. WI0,1236 · OJ/93 5 -29 I I {' I I I I I I le I I I I I I I •• I g. The name of the analyst preparing the standard A sample logbook page is shown in Figure 5-3, however, Group Leaders may use another format as long as the listed information is recorded. In preparing the diluted working standards, it is the analyst's responsibility to make sure that the stock standard is of valid vintage. The preparation of all working standards is also recorded in the logbook. The following information is recorded in the logbook: a. Date of preparation and expiration date b. Identification number of the stock standard solution C. Volume of stock standard solution taken d. Final volume of the diluted standard solution e. Concentration of each parameter in the diluted standard f. Identification number for the diluted standard g. Name of the analyst preparing the diluted standards All stock and working standard solutions will be labelled with ID number, composition and concentration, date of preparation, initials· of the responsible individual, and expiration date. The ID number assigned in the logbook is used for labelling. It consists of: logbook number, page number, and item or line number identifier (e.g., 1 - 2 - 3 is the ID number assigned to the solution found in logbook 1 on page 2 as item 3). The working standards for organic analysis will be stored in a freezer in the work area. No other samples or extracts will be stored in the same freezer. The working standards W83,1236. 0319) 5-7 I - u, I ___, "' -.- SOLUTION ID. NO. --- PREP. DATE BY 1-UMEl - --• -- STOCK IDENTIFICATION COMPOUND PLfE STOCI CONG.STOCK EXP. COM'. S01."N SOL'N. uglml DATE . -- . FIGURE 5-3 STANDARD PREPARATION LOG -- ---• QUANTITY TAKEN FINAL FIIIAL VOL. CONC. ml ug/ml AMOUNl 9 mg ul mL I I I' I I I I I I .. I I I I I I I •• I • for metals will be stored in the work area in a cabinet or shelf designated for standards only. These do not need to be refrigerated. For other parameters, the working standards will be maintained in either a refrigerator or at room temperature in the work area on specifically designated cabinets or shelves where no other materials are being stored. 5.5 Determination of Detection and Quantitation Limits For many projects, knowledge of detection limits is essential in order to be able to bracket analytical results that are obtained. Several such limits exist, and different experts define these limits differently. For this reason the definition and determination used by CHESTER LabNet is given below. 5.5.1 Instrument Detection Limit In simple terms, the instrument detection limit is the smallest quantity of material that the instrument can detect. It has been defined in the past as a certain value of the signal-to-noise ratio. Many modern instruments, however, are designed to self compensate for noise, so that the measurement of the signal-to-noise ratio is not a simple matter. For the purpose of work at CHESTER LabNet, the instrument detection limit is defined as three times the standard deviation from the mean of seven replicate measurements of a low concentration standard that produces a definite, measurable signal. The signal may be an area count, a peak height, an absorbance reading, or electric measures (such as voltage, current, resistance). The nature of the signal is W!l,1236-03193 5 -8 I •• I I I I I I I - I I I I I I I ~ I dictated by the instrument and detector that are used. The instrument detection limit is calculated from the following equation. Where: IDL = s = RF = The calculation of the procedures are discussed. Where: x· 1 = Xm = n = = IDL 3S RF Instrument detection limit, in weight units (ng, mg) for those parameters where the signal depends on an absolute quantity (such as chromatographic methods), and in concentration units for those parameters that are concentration dependent Standard deviation of the seven replicate readings, in units of the reading (i.e., area count, peak height, etc.) Response factor, in units of signal reading/unit weight, or concentration, depending upon the units used for ID L response factor is shown in Section 5 .6 where calibration The calculation of th!! standard deviation is shown below. s = n ( -Xm)Z . E xi I = 1 (n -1) The value of the i'th reading of the set of replicates The mean value ofthe replicates The number of replicate measurements The mean, Xm, is determined as follows: WSJ,1236. 03/93 5-9 I I •• I I I I I I .. I I I I I I I •• I In order for the results to be useful, the standard chosen to obtain the detection limit should be such that the mean of its readings, Xm, is slightly greater than 3S. This may require some trial and error initially when a new instrument is installed. Records of the instrument detection limit determination will be kept in the instrument log, and the values of the instrument detection limit will be updated in the working SOP' s at each time that the instrument detection limits are determined. 5.5.2 Method Detection Limit The method detection limit is obtained in a manner very similar to that of the instrument detection limit. The principal difference is that in determining the method detection limit (MDL), the analytes are subjected to the entire analytical protocol for the specific method that is being employed. Ideally, the method detection limit should be determined for every matrix that is being analyzed. Unfortunately, obtaining reproducible, well characterized matrices for media other than water is not yet feasible. Hence, method detection limits will be determined for water only . To determine the method detection limit, seven replicates of laboratory pure water are each spiked with a known amount of the analyte. The amount that is being added is the same for all seven replicates, and should be 2 - 3 times greater than the previously determined instrument detection limit. The seven replicates are subjected to the same analytical procedures as a sample would be, and the concentrations of the analytes of interest are measured. The method detection limit, as was the instrument detection limit, is definec! as three times the standard deviation of the seven readings. The calculation of the method detection limit should be done in units of weight of the analyte. In this fashion, such variables as injection volume in chromatographic techniques or pathlength in spectrophotometric t,echniques are eliminated. W&J. !236 -03193 5 -10 I I •• I I I I I I le I I I I I I I •" I The equations that apply to the calculations of method detection limits are identical to those used for the instrument detection limit. Where: MDL= = RF = MDL = RF Method detection limit, in units of weight (ng,ug) for those methods that depend upon an absolute quantity, and in concentration units for those methods that depend upon concentration. The standard deviation of the seven readings from the mean, in units of signal size (area, height, etc.) The response factor of the instrument to the analyte, in units of signal size/unit weight or concentration, depending upon the MDL units. The standard deviation, sm,is determined just as before from the equation: Where: Yi = n = n Yn/ E (Y; -i = 1 Sm = (n -1) The instrument reading the i' th sample that has gone through the entire preparation procedure. The mean value of the replicate readings. The number of replicates that have been run. The mean value is determined as follows: n E - i = 1 n The method detection limit will be determined· for all analytes associated with the method on an annual basis. W!IJ.1236,0J/93 5 -11 I •• I I I I I I I - I I I I I I I •• I A method detection limit srudy will . also be performed whenever the instrument undergoes major repair or modification, if the measurement of the instrument detection limit shows a significant departure from the previously determined instrument detection limit. If the instrum_ent detection limit has remained substantially unchanged after the repair or the modification, there is no need to run the method detection limit again. Additionally, a method detection limit srudy must be performed whenever the sample preparation mode is modified. 5.5.3 Quantitation Limits The q1:1antitation limit is determined at the same time as the method detection limit and from the same runs. The quantitation limit (QL) is defined as five times the standard devfation that has been measured in determining the method detection limit. Thus: QL = RF ' where the symbols have the same meaning as above. 5.5.4 Conversion of Detection Limits to J\,linimum Detectable Concentration The conversion. of the detection limit (MDL and QL) to a minimum detectable concentration in a sample is done as follows: WIO, 1236 • 03/93 5 -12 I I n I I le I I I I I I I •• I Where: DLS = DL V· I V· J s = = = = DLS = DL Vj V· I X s Detection limit in sample in units of weight per unit weight or per unit volume. Either the MDL or the QL as defined in the preceding sections. Volume of prepared sample taken for analysis (such as the volume of extract injected into a GC), in mL. The volume of the prepared sample (such as the final volume of an extract), in mL. The sample size that was taken to produce the prepared sample of volume Vj. Sample size is normally measured in liters for aqueous samples and in grams, dry weight, for solid samples. 5.5.5 Documentation or Detection Limits Whenever instrument detection limits, method detection limits, and quantitation limits are determined, the results will be copied to the QA/QC Section. The results must identify the type of detection limits and include both the value in terms of weight (absolute quantity) and concentration. In reporting the concentration units, standard sample sizes and aliquots will be reported. 5.5.6 Application or Limits in Data Reporting In reporting data, the following rules pertaining to detection limits shall apply: a. Results which are equal to or greater than the value of DLS that has been calculated on the basis of Q L will be reported with the concentrations found, without any qualifiers. W83,1236-0Jf9'J 5 -13 I •• I I I I I I I • I I I I I I I •• I b. Results below the value of D LS, calculated on the basis of the MDL, will be reported as not detected at the DLS value. Exceptions to these rules will be applied only for contracts and projects which specify their detection limits, and whose detection limits exceed the value of QL. Also, exceptions will be applied when a prepared sample requires very high dilution in order to have the analyte of the highest concentration within the linear range of the method. 5.6 Instrument and Equipment Calibration Instrument and equipment calibration must be rigorously and routinely performed m order to provide reasonable assurance that the data generated are valid and acceptable. Two principal types of calibration are performed. The first is an initial calibration, which consist of determining the linear range of the instrument and its response factor. The second is a verification calibration, which serves, during the course of running samples, to ascertain that the instrument calibration has not drifted unacceptably. The frequencies of performing the different types of calibrations are shown in Table 5-2 . 5.6.1 Initial Calibration All instrumental methods of analysis are subjected to an initial calibration, consisting of the measurements of responses to various concentration levels of the analytes of interest. The standard solution of the lowest concentration should have a concentration of the analytes of interest approximately 2-5 times the concentration that corresponds to the Instrument Detection Limit; and the standard solution of the highest concentration should have a concentration of the analytes of interest at or near the upper end of the linear range of the method. In performing the analyses of the standards to determine the response factor and the linear range, the standard solution is prepared. as discussed in Section 5.4, and surrogates and internal standards are added as required. The identification of the working standard solution and the date of performing the calibration are entered on the records of the run by the analyst. W&J,1236·03193 5 -14 I I I I I I I .. I I I I I I I •• I - When the calibration is completed, the responses are fitted to a straight line of the form: y=ax+b Where: y = The measured response X = The known amount of the analyte a = The response factor b = They-intercept In addition to determining the values of a and b, the correlation coefficient is determined. The latter is a measure of how closely the five points were to the straight line. The correlation coefficient is determined from the following equation: n n n n I: X;Y; -I: X, I: Y; i = 1 ' i = l i = l T = J n -2 [i ~ I X;J 2 J n [~ I Y; J 2 n I: XI -n I: Yl -i = l i = 1 Where: r = Correlation coefficient x· I = The known amount.of the analyte in the i'th standard run Yi = The measured response to the i' th standard n = The number of standards run to obtain the calibration curve In order for the calibration curve to be valid, the correlation coefficient must be 0.995 or higher, and the ratio b/a must be no greater than the method detection limit (MDL). If the correlation coefficient is not met, it usually implies that either the lowest or the WSJ, 1236 • 0319) 5 -15 I I I I I I I le I I I I I I I •• I highest concentration of standard is outside the linear range. To correct for this, the analyst should rerun the highest standard, and also run a high standard somewhat more dilute than the initially used highest concentration. Similarly, the analyst should examine the effect of increasing slightly the concentration of the lowest standard. If the ratio b/a criterion is not met, the problem may be with contamination in the system or change in the noise level of the instrument. To correct for this, the instrument detection limit should be first checked. If it has in fact changed, the ratio should be compared to the newly determined noise level, in order to see if the criterion is met. While, as much as possible, certified standards are used in the preparation of solutions for ca!ibration, it is always possible that the manufacturer has made a mistake. To circumvent the possibility of error due to a mistake in the manufactured primary standard, a calibration check sample will be analyzed whenever an initial calibration curve is constructed. The calibration check sample will consist of a solution of the analytes of interest, and at known concentration, but obtained and prepared by a different source than the manufacturer of the calibration standards. When the analyte concentrations in the check sample are calculated, they should differ by no more than 15 % from the known concentration. If the discrepancy is greater than 15 % , a determination of the source of inaccuracy will be performed. Once the initial calibration curve has been determined and verified, a table is prepared with the response factors for all the analytes. The table also includes the identification of all the standards used in generating the data, and the date of running the initial calibration. A copy of the calibration curve is: maintained in the work area for ready reference on a daily basis and is available for review by the QA/QC Section. Another copy will be submitted with the sample data to the Information Services Section. 5.6.2 Continuing Calibration Continuing calibrations, sometimes also called verification calibrations, serve to insure that the instrument, during the course of running samples, is remaining sufficiently stable so that the response factor calculated in the initial calibration remains valid. W83,1236. OJ/93 5 -16 I I •• I I I I I I .. I I I I I I I •• I In performing a continuing calibration, the analyst analyzes a midrange standard containing all the analytes of interest and internal standards and surrogate compounds if applicable. The response factor is determined for each analyte by dividing the signal by the known concentration of the analyte. If the response factor is sufficiently representative of the originally determined response factor as specified by the acceptance criteria for the relevant method, then the instrument is considered to be within calibration. Analysis may continue without performing the initial calibration procedure again. If the response factor is determined to be outside the acceptance range, then the instrument will be recalibrated using the initial calibration process. Samples that have been analyzed since the last acceptable calibration will also require reanalysis after the instrument has been recalibrated. In recording the information on continuing calibrations, the analyst will enter the identification of the initial calibration to which .the cortinuing calibration relates. Copy or copies will be submitted to the Information Services Section for inclusion in the file of the project from which samples have been analyzed . At no time should the response factor be corrected on the basis of the continuing calibration. Until such time as it is necessary ·to reestablish the initial calibration, the response factors determined in the initial calibration will be adhered to. 5.6.3 Calibration Frequency Instruments have widely variable stabilities, requiring variable frequencies of calibration. Table 5-2 summarizes the frequencies of such calibrations. It should be emphasized that these frequencies are based. on instruments that are performing normally. It is obvious that after change in instrument parameters or after repair the initial calibration must be repeated and the cycle started over again. Thus, the frequencies included in the table are minimum requirements. W8J, 12:36 • 0319'3 5 -17 I •• I I I I I I I .. I I I I I I I •• I INSTRUMENT GC/MS GC/MS GC GC HPLC IC Autoanalyzer W83, 1236 -03193 TABLE 5-2 CALIBRATIONS FREQUENCIES APPLICATION Volatiles Semi-Volatiles Volatiles by Purge- and-Trap Extracts Extracts Solutions Solutions INITIAL CALIBRATION As Needed As Needed Every Three Weeks Every Three Weeks Every Four Weeks ' Every Three Days Daily 5 -18 CONTINUING CALIBRATION Every twelve hours and at the end of a sample sequence, if the instrument is about to be idle. Every twelve hours and at the end of a sample sequence, if the instrument is about to be idle. After every ten samples, and at the end of a sample sequence. After every ten samples, and at the end of a sample sequence. After every ten samples, and at the end of a sample sequence. After every ten samples, and at the end of a sample sequence. After every ten samples, and at the end of a sample sequence. I I I I I I I .. I I I I I I I ~ I INSTRUMENT ICP AA Mercury Analyzer TOC Analyzer TOX Analyzer IR UV/Vis Spectrophoto- meter pH Meter Analytical Balance Thermometers TABLE 5-2 CALIBRATIONS FREQUENCIES (CONTINUED) APPLICATION Digests Digests Digests Solutions Solutions Solutions for O&GorTPH Solutions Aqueous Solutions Solids Temperature INITIAL CALIBRATION Daily Daily Daily Daily Daily Daily Daily Daily1 Annual2 Semi-Annual3 CONTINUING CALIBRATION After every ten samples, and at the end of a sample sequence. After every ten samples, and at the end of a sample sequence. After every ten samples, and at the end of a sample sequence. After every ten samples, and at the end of a sample sequence. After every ten samples, and at the end of a sample sequence. Every ten samples, and at the end of a sample sequence. Every ten samples, and at the end of a sample sequence. Every ten samples, and at the end of a sample sequence. Daily2 Not Applicable I) 2) pH calibration will require the use of three standards. 3) W83, 1236. 03/9J Analytical balances are calibrated and serviced annually by a contractor. Calibration is checked daily using Class S weights. Deviations from the true weight are recorded and applied to weighings when appropriate. Thermometers are calibrated semiannually against an NIST certified thermometer and any required corrections are applied to measurements. 5 -19 I I I I I I I .. I I I I I I I •• I - 5.7 Monitoring Laboratory Reagent Water Multipurpose high purity water is prepared on-site. The laboratory measures the specific conductance. and pH of this water on a daily basis. The records are kept in chronological sequence in a bound laboratory notebook with the date of each analysis and responsible analyst. The filters and ion beds of the system are changed as required to assure that high quality water is constantly available. 5.8 Analysis of Quality Control Samples Routine quality control samples are analyzed to assure that the operation is within control as established for the laboratory on the basis of historical data. The routine qual_ity control consists of blanks, spiked blanks,' spiked samples, duplicate samples and external check samples analyses. These are discussed separately in the following sections . 5.8.1 Blanks Four types of blanks may be associated with any batch of samples. These are: reagent blank, method blank, trip blank, and field blank. The latter two types are treated by the laboratory as ordinary samples and are not part of the internal laboratory QC, although they are very much part of the program QC. Thus, trip and field blanks will not be discussed here. 5.8.1.1 Reagent Blank Reagent blanks are set aside whenever the reagents are used for preparation of samples. The reagent blank is taken from the reagent or group of reagents that is normally used for sample preparation and that does not go through any of the preparation steps. This reagent blank is normally not analyzed unless the method blank, discussed in Section 5.8.1.2, shows the presence of contamination' which may have arisen from the reagents. The reagent blank will be labelled with the QC batch identification, followed by the letter R. It will be set aside until all the, samples of the QC batch have been WBJ.1236 -03193 5 -20 I I I I le I I I I I I I •• I analyzed. At that point, if the method blank was acceptable, the reagent blank may be discarded. 5.8.1.2 Method Blank The method blank is a preparation carried through all preparatory steps, except that the reagents do not come in contact with a sample. Rather, laboratory reagent water is used in lieu of a real sample. Ideally, the method blank would be prepared using a matrix that is similar to the matrix of the samples that are being prepared. Since a reference matrix is not available for matrices other than water, water is the only matrix used as a method blank. In preparing the method blank, water is spiked with surrogates and internal standards, if appropriate for the method, and the water sample is carried through the entire analytical procedure. The method blank is prepared with every batch of samples that is being prepared at the same time, provided the batch is no greater than twenty samples. For batches of greater than twenty samples, a method blank will be prepared for every sub-batch of twenty samples or part thereof. The preparation lot will then indicate which samples are associated with the new lot number of reagents. The method blank is analyzed before any samples are analyzed, and the data of the ' analysis are reviewed. If no analytes are found above the quantitation limit, analyses of the prepared samples may be undertaken. If analyte concentrations are found above the quantitation limit, analyses of the associated samples will not be undertaken until; the contamination source is identified and isolated. At this stage the reagent blank wjll be analyzed. If it is found that the reagent blank shows the contamination, the samples will be reprepared1 using a new lot of reagents. WID,1236 • 03/93 5 -21 I I I I I I I .. I I I I I I I 5.0 ANALYTICAL PROCEDURES In order to produce meaningful results, both sampling and analytical procedures must be sound and complimentary to each other. While close coordination of activities between the laboratory and field services is highly advisable in order to produce a high quality product, laboratory services may involve analyses of samples collected through organizations other than CHESTER Environmental. Thus, the application of strict procedures regarding the transfer of samples from the field in some cases may involve conditions beyond the control of the laboratory. 5.1 Coordination of Activities Analytical services are requested through several sources. The project manager acts as liaison between the client and the laboratory on all matters pertaining to the project. However, requests for analytical services may originate with the. laboratory sales representatives or other key technical persons in the laboratory. The person receiving the request will enter a description of the requested work on an Analytical Request Form, shown in Figure 5-1, and the information will be entered into the Laboratory Information System as proposed work. When the proposed project becomes a definite task for the laboratory, the Laboratory Director will assign a Project Manager who will activate the project and distribute the information together with the anticipated schedule. In that way, the laboratory sections and groups that will be involved in the analysis are made aware of the upcoming work. All further coordination and scheduling of the work will be through the assigned Project Manager. 5 .2 Preparation of Sample Containers For clients and projects that require the laboratory to supply containers for samples, the Manager of Sample Receiving will be responsible for preparing the containers, labeling them, and shipping them to the client. It is the responsibility of the Project Manager to inform the Manager of Sample Receiving of the need for containers. Such notification WBJ, 1236. OJ/93 5 -1 I I •• I I I I I I I I I I I I FlGURES-1 ANALYTICAL REQUEST FORM Client Contact: Date: ----------- Client: __________ _ ' Telephone: Address: __________ _ Project Manager: ------------ ------------ ------------- #SAMPLES MATRIX PARAMETERS TURNAROUND .. I I Special Instructions: ------------------------------- ~ I 5-1a .. I I I I I I .. I I I I I I I ~ I must be received by the Sample Receiving Section at least three working days prior to the time for which the sample containers are required. For aqueous field samples, new pre-cleaned :sample bottles with Teflon lined screw caps will be used. The containers will be prepared with the appropriate preservatives and will be labelled ·with information regarding the analytes that are to be determined on the sample. For solid samples, a variety of containers may be used. For soils, glass containers with Teflon liners are employed. When the samples are to be collected in polyethylene, appropriate new and pre-cleaned containers will be used. Table 4-1 lists the appropriate containers and preservatives for aqueous samples. 5.3 Instrument Maintenance Maintenance of the instruments in the laboratory consists of two major aspects. The first is routine preventative maintenance and the second is repair due to malfunction . Acceptable procedures and specifications in either case confirm to those recommended by the manufacturer of the specific equipment' as well as any criteria provided in the applicable approved methods. Written equipment maintenance documentation is part of the laboratory records for both types of maintenance as described below. 5.3.1 Preventative Maintenance The manufacturer's recommended schedule 9f preventative maintenance will be followed to reduce occurrences of instrument failure and to help maintain the reliability of analytical results. If more frequent routine maintenance is necessary to assure proper operation, then these practices will be added to the maintenance scheduled, and will be performed and documented in laboratory records. WD.1236-031'13 5-2 I I I I I I I .. I I I I I I I It is the responsibility of the Group Leader to provide the time required for preventative maintenance. Preventative maintenance records must be kept and include procedures performed by laboratory personnel and any preventative maintenance contracted for using outside repair technicians. The Gro~p Leader will insure that preventative maintenance records are complete, accurate, and comply with the Standard Operating Procedures dealing · with these laboratory: records. Verification of instrument performance and recalibration as necessary must take place inunediately after maintenance has been completed and before resuming sample analys~s. 5.3.2 Repair After Instrument Failure Unscheduled repairs which result from instrument malfunction are arranged for inunediately after the failure is detected. Deteriorating equipment performance is detected both directly through observations during analyses and indirectly through the routine use of verification samples throughout the course of an analytical run. If instrument response and performance on verification samples is not within acceptable criteria, analyses are halted, preceding analytical results to the last acceptable verification sample of the run are marked "VOID", and repair is undertaken. The Group Leader will make the decision as to whether the scope of the repair is within the capability of laboratory personnel or whether a service call to a repair contractor is necessary. Data from voided analyses will not be entered into the LIMS. Notice of invalid data will be given by the Group Leader to the Quality Assurance Deparnnent if data from the voided analyses have been released by the analytical section to the Information Services Group. A Corrective Action Notice 'Form, as shown in Figure 5-2, will be initiated by the Group Leader to track such data so that the appropriate action can be taken. Other laboratory documentation which results from instrument repairs are the equipment maintenance records that are described in the previous section on preventative maintenance. As indicated in the maintenance recordkeeping Standard Operating Procedures, repairs are documented in logbooks along with related information which will include the identification of any sample results which have been affected by the malfunction, and a subsequent verification of performance within acceptable criteria prior to resuming sample analyses. W8J, 12J6-03193 5-3 I FIGURE 5-2 CORRECTIVE ACTION NOTICE I _ Work Order: ______________ _ •• Client: Dept.: I I I I ACTION I. State the analysis and/or reporting problem recognized: (Analyst or Manager) II. State the action taken: 111. Invalid data in UMS? Y/N Individual completing the necessary corrective action: APPROVALS I I. MANAGER OF ORIGINATING DEPARTMENT The problem appears laboratory (e.g. procedure, instrument) related. ----I ____ sample (e.g. matrix, volume) relaled. II. PROJECT MANAGER Ill. OC MANAGER .. The client was contacted. ----Additional follow-up I I I I I I Signature: Date: was not contacted. ---- Signature: Date: COMMENTS 1. HOW TO USE: Use sheet to track and document one workorder. Date: -------- Signature: -------- Date: -------- 1s necessary. ---- ____ is not necessary. 2. WHEN TO USE: Any time a problem cannot be immediately isolated and corrected. 3. OISTRIBUTION: Complete and submit to QC. I R3320 5-3a I I I I I I .. I I I I I I I (' I 5.4 Preparation of Standards A calibration standard is made by the appropriate dilution of a pure substance, the purity of which is traceable to an approved (e.g., NIST or equivalent) standard. Because of the high sensitivity of many analytical instruments, the calibration standard is an extremely dilute version of the pure compound. Because of the high dilution required, in order to be within the linear range of the instrument, the preparation of the calibration standard is frequently made by serial dilution rather than in a single step. In order to provide standard solutions at sufficiently low concentration, a miniscule I amount of the pure substance will be required,' the measurement of which is subject to extreme error. Thus, it is preferable to deal with potential dilution errors, rather than with the large error associated with the measurement of a very small amount of the pure substance. The initial pure standard is obtained either as · a pure material or already in solution prepared as a certified solution of a given concentration of the pure compound or compounds. In preparing the stock solution of the calibration standard:, great care must be exercised in measuring weights and volumes as accurately as possible since all the analyses following the calibration will be based ·on the accuracy of the calibration, and the accuracy of the ultimate data cannot be any !Jetter than that of the calibration curve. Table 5-1 summarizes the valid lifetime of primary and secondary standards used in many tests. These lifetimes should be taken 'as a guide only. It is the analyst's responsibility to assure that all standards used by him are within the standard solution holding time, and to prepare fresh standard solutions whenever necessary. In preparing working solutions, or using working solutions, the analyst must check for signs of deterioration, such as the formation of cloudiness, precipitation, or discoloration. The standard must also be periodically compared with previous runs of standards, and with independently prepared standards to assure that response factors · fall within an historically accepted range. W&:I. 1236 • 03m 5-4 I I I I I I .. I I I I I I I {" I TABLE 5-1 STANDARDS AND SOLUTIONS HOLDING TIMES HOLDING TIME PURE STOCK WORKING MATERIAL COMPOUND SOLUTION SOLUTION Volatile organic compounds I Yr@ -IO'C 2 Mo@ -IO'C I Wk@ -I0'C for GC or GC/MS analysis Semivolatile organic compounds for GC or GC/MS analysis I Yr@4'C I Yr@4'C 6 Mo@ -IO'C Semivolatile organic compounds for HPLC analysis I Yr@4'C I Yr@4'C ' 6 Mo@4'C Pesticides (Cl, P, N ,) and herbicides I Yr@4'C I Yr@4'C 6 Mo@ -I0'C Polychlorinated dioxins and furans I Yr@4'C I Yr@4'C 6 Mo@4'C Metals for ICP analysis lndef@ RT I I Yr@RT 6Mo@RT Metals for GFAA analysis Indef@ RT I Yr@RT 6Mo@RT Mercury Indef@ RT 6 Mo@RT I D@RT Hexavalent chromium Indef@ RT I Yr@RT I Yr@RT For anions and other parameters, the holding times of standard solutions should be checked in the appropriate method. All standards and standard solutions of organic compounds will be maintained in glass containers, protected from light, and stored under the conditions specified in the particular method. The position of the meniscus in each container will be marked after each time that the container is opened so that changes due to evaporation can be detected. WD,1236-03193 5-5 I I I I I I I -I I I I I I I Metals working solutions and stock solutions will be kept in polyethylene containers at room temperature. The position of the meniscus will be marked each time a solution is used to insure that concentration changes due to evaporation are detected. Before using any standard solution, the analyst will examine· it for signs of precipitation and changes in color. If precipitation has occurred, the solution will be discarded and a new standard prepared. Discoloration frequently is only a warning sign, but will not affect I the results. If a solution is discolored, the analyst will compare the results with historically established response factors, to assure that the solution is still within the operating range of the method, and within experimental error, of its original concentration. The preparation of standards is very exacting. 'To facilitate the operation of preparing standai:ds, a separate area is set in the laboratory equipped with a small hood and analytical balance. A freezer in the same room is used to store all primary standards and no other samples or extracts. In this fashion, the contamination of standards by samples, and vice versa, is minimized. For each stock standard solution that is prepared, accurate records will be kept in a special logbook used only the maintenance. of standards data. The following information will be entered in the logbook at the time of stock standard preparation: a. b. C. d. e. f. W83, 1236. 03193 Date of preparation and expiration date Application for which the standard is being prepared (i.e., identification of. the method) For each compound, the supplier of the primary standard, the batch number, and the amount taken The solvent identification (compound, supplier, batch nu~ber) The final volume of the stock standard The identification number assigned to the stock standard preparation 5-6 I •• I I I I I I I - I I I I I I I These operations are the responsibility of Sample Receiving. The individual steps are itemized below: 4.2.1 Sample Receipt in the Laboratory Samples will be received in the laboratory either by commercial carrier, the postal service, or hand-carried. Personnel in Sample Receiving sign for the receipt of each shipment of samples and retain a copy of the shipping documents. The individual receiving the sample shipment will open a sample shipment checklist (see Figure 4- 2) at the time of receiving the shipment. If, for any reason, the shipping container is not expected to be opened immediately, then the seals on the container must remain intact. 4.2.2 Shipment Inspection It is expected that a shipment received in the laboratory will be opened and inspected immediately upon receipt. Prior to opening the shipping container, the custody seals will be inspected to assure that no tampering has been done with the sample containers. The state of the custody seals will be noted on the sample shipment checklist by the person inspecting th~ shipment. After the seals are inspected, the ice chest is opened. The temperature of the interior of the ice chest is measured and recorded. To measure the temperature, the lid is quickly opened and a thermometer is inserted into the ice chest, and the lid is closed again for five minutes. At the end of five minutes, the lid is quickly opened and the temperature read rapidly. The temperature is recorded on the sample shipment checklist. The ice chest is opened, the chain-of-custody document is removed from the inside of the lid, and the individual sample containers are removed from the ice chest. The sample or sub-sample containers are counted and reconciled with the number of W83,Ill5. fJJ/93 4-6 I •• I I I I I I I .. I I I I I I I f' I FIGURE4--2 CHESTER LABNET-MONROEVll.LE SAMPLE SHIPMENT CHECKLIST ( ) MOllJOCVillc 108 NO.: _______ _ ( ) Howton DATE SHIPPED: CUSTOMER: DATE RECEIVED: -------- CUSTOMER CONT ACT: SH IP PED VIA: -------- TELEPHONE NO.: AIR BILL NO.: -------- CHECKED BY: SM OF ORMS: -------- NUMBER OF SHIPPING CONTAINERS (COOLERS, BOXES, ETC.): ___ _ -------- -------- -------- -------- -------- CUSTODY TAPE #OF CHAIN-OF-CUSTODY PRESENT? INTACT? TEMPERATURE SAMPLE PRESENT/ RECORD AGREE WITH CONTAINER I.D. (YIN) (YIN) (C) I CONTAINERS ABSENT NO. SAMPLES ... IRREGULARITIES __ ~NO ___ YES (IF YES, PLEASE EXPLAIN) SAMPLE I.D. SUB SAMPLE I.D. IRREGULARITY CLIENT REPRESENTATIVE: TELEPHONE NO.: -------------- CONT A CT ED BY: --------------DATE: ____ TIME: ____ _ WRITTEN FOLLOWUP (YIN)'-: ____________ _ DATE: ---- DECISION: --------------------------------- SIGNATURE: I •• I I I I I I I .. I I I I I I I such containers indicated on the chain-of-custody. If the number of retrieved sample containers is less than that indicated on the chain-of-custody, the packing materials inside the. ice chest are further checked to make sure that no sample container has been accidently left in the ice chest. Each individual sample or sub-sample container is visually inspected to determine ' that no breakage, cracking, external corrosion, or leakage has occurred. If none have occurred, the individual sample containers may be removed from the fume hood and placed on a workbench to complete the inspection. If, on the other hand, there is indication that breakage, cracking, corrosion, or leakage has occurred, the inspecfton of the sample containers will be completed while the containers are kept in the fume hood. The integrity of the individual sample or sub-sample containers is recorded on the sample shipment checklist . 4.2.3 Reconciliation with Chain-of-Custody Document Once the integrity of the sample containers has been determined, the sample containers are reconciled against the records on the chain-of-custody. This is done by checking the sample identification on the chain-of-custody and on the sample container label. In addition, the analyte identifications are checked to ensure that they are correct. Any discrepancy is noted on the chain-of-custody, signed, and dated. The discrepancies are also entered on the sample shipment checklist. At this point, the sample shipment inspection is complete. The sample shipment checklist is signed ' and dated by the person performing the inspection. If there are no discrepancies, and the shipment is complete as evidenced by the inspection, the shipment of samples is ready to be logged in. If there are discrepancies or inconsistencies, the sample shipment checklist is submitted to the Project Manager and the logging in process is delayed until the discrepancies are resolved. W83,l:D5. aJ/f/3 4-7 I I I I I I I .. I I I I I I I ft I 4.2.4 Resolution of Shipment Irregularities If any irregularities are noted during the ,sample shipment inspection, they are recorded on the sample shipment checklist, and the checklist is submitted to the Project Manager. The Project Manager will contact the client's representative to determine the fate of the sample shipment.' The records of the conversation with the client's representative are entered on the sample shipment checklist, including name of contact, time and date of the conversation and the resolution of the irregularities. There are several possibilities for the resolution of the irregularities. These are: a. Return the sample shipment to sender. b. Destroy the entire shipment of samples . c. Log in and process those samples that are intact. The sample shipment checklist containing comments regarding resolution of any irregularities. is returned to Information Services, the personnel of which will act according to the annotated agreement with the client. To maintain the custody of the samples, the entire sample shipment during this period is either locked up in a secure area or is in view of Sample Receiving personnel. 4.2.5 Sample Log In Once the_ sample shipment has been inspected and any irregularities resolved, the sample shipment is ready to be logged in. For laboratory purposes, a single sample shipment from a specific client constitutes a single job. The shipment may contain one or many samples and may have arrived in a single shipping container or in many shipping containers. Wl0,1235 • 03/93 4-8 I ~-I I I I I I I .. I I I I I I I •• I For the log-in process, Sample Receiving staff performing the logging-in will use the field chain-of-custody and the sample shipment checklist, both with whatever corrections were required to be made during the inspection and resolution steps. The log-in process consists of the four steps described below: ' ' 4.2.5.1 Entry in Master Log The Master Log is a hardbound book in which all jobs received in the laboratory are chronologically recorded. The following information is entered in the Master Log: (a) Job Number (b) Date of Receipt (c) Date of Logging-in (d) Name of Client (e) Number of Samples (not sub-samples) (f) Due Date (g) Sample Matrix The job number consists of a letter followed by a seven digit number. The initial letter code identifies the laboratory (H -Houston, P -Portland, NI -Neville Island, M -Monroeville). The four digits following the letter identify the year and the month of the sample shipment logging-in, and the last three digits are chronological within the month. Thus, M9309005 is a job number issued by the Monroeville Laboratory for the fifth job logged in during September 1993. WlDJD!i: • 03/9] 4-9 I •• I I I I I I I .. I I I I I I I ~ I - A sample page of the Master Log is shown in Figure 4-3. At the time of entering the job in the Master Log, the job number is manually written on the Field Chain-of-Custody and on the Sample Shipment Checklist. The job number remains the identification of the job in the laboratory and on the job records. 4.2.5.2 Job Traveller In addition to opening an entry in the Master If>g, a Job Traveller is issued for every job in the laboratory. The Job Traveller is a computer generated document that gives all the pertinent details regarding the individual samples in the job. For each sample in the job, a unique number is assigned. The unique number consists of the job number followed by three digits, which are sequehtial within the job. This number is entered on the traveller, as well as the original sample identification from the field chain-of-custody,' the date of sampling, the sample matrix, and the parameters for which the sample or sub-sample is to be analyzed. In addition, the traveller heading contains information such as client· identification, project contacts, date of collection, date of logging in, and due date. A sample of a Job Traveller is shown in Figure 4-;4. 4.2.5.3 Sample Labeling When the logging-in process is complete, the operator entering the information proofs the input and verifies that all the inforination is correct. The complete laboratory sample or sub-sample identifications are entered on labels. The operator then places the correct label on each container of each sample, and verifies once more that the information has been correctly recorded. The labelled sample containers are then placed m a secure storage area for preservation at the appropriate temperatures. WK3J2l:J. aJ/93 4-10 I --.----·- Job Date Date Number Received Logged - - -I - - FIGURE4-3 SAMPLE MASTER LOG. Client Number of Samples -----.-- Sample Matrix Due Date Soil Water Sludge Oil Other - I I I I I I I .. I I I I I I I FIGURE4-4 CHESTER LABNET-MONROEVILLE Date: ___________ _ Received By: ------------ Job Name: Date Received: ----------- Customer: -'------------Time Received: --------'-----P.O. Number: ___________ _ Project Number: ___________ _ Description: ------------Priority: Matrix: ------------ Sample No SOURCE/ SM!PLE TAG NO ..• · DATE COLLECTED NO. OF CONTAINERS PRESERVED CORRECTL • ' ' ft COMMENTS: I 4-10b - I ,. I I I I I I I .. I I I I I I I {" I - 4.2.5.4 Information Distribution and Job Filing The Job Traveller is the working document for each job. Sample Receiving personnel make a copy of each traveller for each group in the laboratory as well as for the relevant Project Manager and Information Services. The copies are • I distributed to the Group Leaders and Section Managers as required. The original traveller is used to open a file for the job in which the laboratory copy of the field chain-of-custody and the sample shipment checklist are placed. The file is identified by the job number. As the work on the job is completed, the file. will be used to store all laboratory records regarding the job. The file for the job is placed in the active laboratory central job filing system. The sample log-in process is extremely critical for the proper functioning of the laboratory. It must be performed rapidly and accurately, so that holding times will be met, and so that the correct analyses will be performed on the appropriate samples. 4.2.6 Custody Transfer Within the Laboratory Because the laboratory is considered a secure facility, samples will be considered as being in the custody of the laboratory from the time that sample receipt is recorded. All samples will be maintained at secure locations under supervision of Sample Receiving staff until the group responsible for the sample preparation for analysis is ready to start work on the samples. The group representative will then obtain the samples from their designated storage area. Tl)e person retrieving the samples will fill in a transfer form, a sample of which is shown in Figure 4-5, and submit the form to the Sample Receiving Section. 4-11 I I ~NEDOUTBY I I I I I I .. I I I I I I I , t\ DATE FIGURE4-S Chester LabN et-Monroeville Internal Transfer Forin SAMPLE NUMBERS ANALYSES 4-lla JOB: ______ _ RETURNED BY DATE I I I I I I It I I I I I I I ,. I - When the sample preparation or analysis is completed and residual samples are returned to the original, secure storage location, the analyst will indicate on the Internal Transfer Form which samples, if any, have been completely used up. The Sample Receiving Section will maintain a record of these forms. 4.2.7 Sample Disposal Sample Disposal is an issue requiring attention to the specific, and often changing, regulatory requirements. For these reasons, laboratory practices in this area are dealt with separately in an SOP written to address these concerns. WUl.m. 0'3(93 4-12 I •• I I I I I I I .. I I I I I I I •• I - 2.0 OBJECTIVES The Quality Assurance Program at CHESTER LabNet is principally aimed at producing results of verifiable high quality. Towards this goal, the Program addresses several areas: 1. 2. 3. 4 . 5. WaJ,IZ!:J. '1J/9l Detection of problems through statistical measures of acceptability and confidence. Implementation of corrective action. Documentation procedures designed to produce legally defensible results. Establishment of training programs to assure that each person is thoroughly familiar with the methods, procedures, and documentation ' ' of his area of activity. Development of a review and validation process to verify that all data produced by the Lab are within the guidelines defined in this Manual, and the associated Standard Operating Procedures. 2-1 I I I I I I I lie H I I I I I I ~ I 3.0 ORGANIZATION Ai'ffi RESPONSIBILITY The laboratories have been placed within the CHESTER LabNet framework according to the organizational structure described in this section. 3.1 CHESTER LahNet CHESTER LabNet-Monroeville is an integral part of a larger organization, CHESTER LabNet, whose organizational scheme is illustrated in Figure 3-1. The LabNet Operations Manager is responsible for assuring that the Laboratory Direct9rs are thoroughly familiar with the Quality Assurance Manual and good laboratory practices. The role of the Laboratory Directors will be discussed in the next section. 3.2 Laboratory 3.2.1 Organization Figure 3-2 illustrates generically the organization of each Laboratory. 3.2.2 Responsibility The Laboratorv. Director is responsible for assuring that all Section Managers and Group Leaders are thoroughly familiar with the Quality Assurance Manual and good laboratory practices and that all laboratory personnel meet the requisite qualifications for their positions within the laboratory. The Laboratory Director, or his designee, must review and approve all outgoing reports. The Laboratory Director is also responsible for effective daily management of the laboratory and its staff, and for co=unication and liaison with the client. \W3,1ZJ,4 • (JJ/'ll 3-1 I •• I I I n I I I .. I I I I I I FIGURE3-l CHESTER LABNET ORGANIZATION CHART Operations Manager Marketing/ Programs Sales Office Information Support Management Lab Director Lab Director Monroeville, PA Houston, TX 3-1a I --- - w I ' INFORMATION SERVICES I I . WET METALS CHEMISTRY ANALYSIS ANALYSIS GROUP GROUP - --· -- FIGURE 3-2 LABORATORY ORGANIZATION LABORATORY DIRECTOR I I SAMPLE PROJECT RECEIVING MANAGEMENT I I ORGANIC HPLC EXTRACTIONS ANALYSIS GROUP GROUP - - ---.-- I QA/QC SECTION I I GC GC/MS ANALYSIS ANALYSIS GROUP GROUP I •• I I I I I I I .. I I I I I I I •• I The Group Leader is responsible for the production of quality results within the group. To achieve this, the Group Leader must be thoroughly familiar with the Quality Assurance Manual and Standard Operating Procedures used by the group. He is also responsible for familiarizing personnel with the Quality Assurance Manual and Standard Operating Procedures, seeing that required protocols are followed, reviewing the results, and approving release of the data to the Information Services Section. The Group Leader, in coordination with Sample Receiving, Information Services, and the Project Manager is responsible for scheduling work and conforming to required holding times. The Group Leader, in conjunction with the Quality Assurance Manager, is responsible for providing the necessary training to laboratory personnel. The Analvsts are responsible for performing all analyses as required and noting the required QC analyses demanded by the analytical method or technique. In order to provide proper analysis, they must be familiar with the Quality Assurance Manual and the associated Standard Operating Procedures. They are also responsible for initiating system or method corrective action should they become aware of a malfunction. Initiation of corrective action requires appropriate notification, as discussed later in this manual. The Manager of Sample Receiving is responsible for the following functions: sample receipt,· storage, distribution of the information through the laboratory, sample custody, and sample disposal. It is his responsibility to notify the Group Leaders and Project Managers should there be any discrepancies or irregularities in the shipment of samples. The Manager of Information Services is responsible for maintaining the status of the work within the Laboratory, coordinating the compilation of the data, and preparation of reports for review and approval. WKl.12JA • l'.0/93 3-2 I •• I I I I I I I .. I I I I I I I •• I - The Quality Assurance Manager is responsible for assuring that the QA/QC requirements of the Quality Assurance Manual and its associated Standard Operating Procedures and addenda are strictly followed. He is responsible for reviewing data validation procedures, alerting the sections and groups should the need for corrective action exist, performing internal audits, introduction of performance evaluation samples on a periodic basis, and maintenance of the QC records. He is also responsible for preparing project-specific QA/QC plans. The QA/QC Manger functions independently of the laboratory staff. In order to achieve independence from the pressures of daily production in the laboratory and maintain the necessary objectivity, the QA/QC Manager reports both to the Laboratory Director and the Operations Manager. The Project Manager is a position assigned by the Laboratory Director for specific projects. Projects may require a specifically assigned manager because of the unusual nature of the project, complexity of the analytical techniques or reporting requirements, or the need for coordination of activities in several laboratories. The responsibility of the Project Manager to the specific project transcends that of the Laboratory Director. The Project Manager will alert the . Group Leaders if problems arise in meeting schedules or sample holding times. It is also the Project Manager's responsibility to assure that work on the project is performed in accordance wiih project specified protocols, following project specific QC requirements. Acceptance of results on analyses for the project is subject to approval by the Project Manager. W83,1D4 • O'J/'Tl 3-3 I •• I I I I I I I .. I I I I I I I •• I 4.0 SAi"\iPLE CUSTODY To provide for legal defensibility of all work performed at a given site, it is essential to be able to provide documentation tracing the samples from collection, to the laboratory, and through the analytical procedures. CHESTER Environmental performs both sampling functions and analytical functions: however, the laboratory can guarantee that this Manual is followed from the point of origin only for samples that are both collected and analyzed by CHESTER Environmental. Maintaining sample custody consists of two distinct aspects: maintenance of the samples in the field, and maintenance of the samples from the time of receipt in the laboratory. These two aspects are discussed separately in the following sections. Inasmuch as sampling is not necessarily performed by our personnel, the custody and documentation in the field are included here as a recommendation . 4.1 Custody and Documentation in the Field The field sample custodian, which, depending upon the project, may be the sampler or another person in the same sampling group is considered to have custody of the samples at all times during the field operations, until the samples are shipped to storage or to the laboratory. Upon collection of a sample in the field, the sampler tags the sample with its site and type (water, soil, sludge, etc.) identification. The sampler also indicates on the tag the date and time of sampling. After cleaning the exterior of the sample container, the field sampler transfers the container with the tag to the field sample custodian. Throughout this document, the term sample is used to indicate a quantity of one type of material collected at one time, at a single location. Thus, a water sample may be shipped to the laboratory in several containers, depending upon the required testing and sample preservation dictated by the project. Each container is identified as a sub-sample, but it is not classified as a unique sample. WKJJ2l!i. a:J/'13 4-1 I ~-I I I I I I I .. I I I I I I I {' I The field sample custodian compares the identification of the individual sample with the sampling plan, and enters all pertinent information on the chain-of-custody document and on the label of the sub-sample container. The information that must be included consists of the following: Project identification Sample identification (such as station number and location) Date of sampling Time of sampling Name of sampler Parameters for which the sample is to be analyzed Number of containers Sample matrix (groundwater, surface water, wastewater, soil, sediment, sludge, unknown waste, etc.) Added preservatives in each sample container Ice chest number Chain of custody number A sample field chain-of-custody is shown in Figure 4-1. WBl,1235 • 03/9'3 4-2 I - ... I N '" -.-------I -- - ---.-- -·------------------------·---·. ---------- CHAIN OF CUSTODY RECORD ·-=-~C~~PROJECTNAME NUMBER t: ··---; ~ SAMPLERS OF i:: tS,gnalutU} CONTAINERS J REMARKS OR p OBSEAVA 110115 ·--------C • -8 ~ DATE TIME • • I S IATION LOCATION Q SIA NO. • • I • • I ,_ --~------·---· ------·-- ---·· -----------· -·· ------- - - -------------·------·-· --·--·-----·--- --------~ -------------- ----------·------·-----------· - ------· ---------· -· -·-- -----------------------. ·------·- --------· ----------------- ----------------------------- ----------·------------------·----·- ·--·· ----------------···----· ----· -· -------·-----·-----------·--·- -------------------------------------- ------------------------· -----·---------------------- ------· --·--·-- Aolinquishod by: ISignafur•J Dale Time Aoceivod by: tSignalwo} Aeliquishod by: lSignalur•J Da10 Time Aocoivod by: ISignaluro) ------- Rolinquishod by: (SignafuroJ Dato Time necaivod by: (Sig11alu1t1J Aoliquishod by: {Signalur•J Oalo limo RocoivDd by: tS19na1u10J Roliquishod by: (Signalutll) Dalo Time Rocaivod lor laboralory by: (Signalure) Dalo limo lcti Chest Temp Ice Olest Chain of Cuuody oc # 'OISlRIBUTION: O1iginal accompanies shipment. Copy lo Coordina10, fiokt Filos. Tag # PAGE __ OF ___ ------. -.. ----------------------------------------------- J:lf;I IAF d-1 I I I I I I I .. I I I I D I I re I The field sample custodian is then responsible for packaging the sample(s) for shipping, adding ice if the samples require chilling, signing and dating the chain of custody document, placing the chain of custody document in a waterproof envelope and attaching the envelope to the inside of the ice chest lid. If the samples are to be shipped by a common carrier, then the field sample custodian must also place custody seal on the ice chest. Simultaneously with filling out the chain of custody document, the field sample custodian also records the information in the field logbook. In filing the chain of custody document and the field logbook, any corrections that need to be made must be done so that the original incorrect entry is legible. Hence, the incorrect entry is lined _out, and the change is initialed and dated by the field sample custodian. Table 4-1 lists the requir~d types of containers, preservatives, and holding times for each type of analyte. It is the responsibility of the field sample custodian to assure that each sample or sub-sample are packaged correctly. The samples are considered formally to be in the custody of the field sample custodian until they are officially transferred to the carrier, and the transfer is documented on shipping records, or until the samples are transferred to the laboratory in person. 4.2 Sample Custody in the Laboratory The laboratory operation, as it penains to the sample custody, consists of several functions. Specifically, these are: sample receipt, inspection of the samples, reconciliation of the information on the sample label and the chain-of-custody, alening the project manager of any inconsistencies in the shipment, logging in of the samples, placing the samples in appropriate storage areas, distribution of the information to the laboratory analytical sections, transferring of the custody of the samples to the analysts, and recovery of the samples at the completion of the analysis. WB3J235 • OJ/93 4-3 I •• I I TABLE 4-1 CONTAINERS, PRESERVATIVES, AND HOLDING TIMES I ANALYSIS CONTAINER PRESERVATIVE HOLDING TIME I Volatile organics G HCl, Cool to 4°C 14 d Semivolatile organics G Cool to 4°C 7 d to extraction, I 40 d for extract. Organochlorine pesticides G Cool to 4°c 7 d to extraction, I 40 d for extract. Herbicides G Cool to 4°C 7 d to extraction, 40 d for extract. I Organophosphorus pesticides G Cool to 4°c 7 d to extraction, 40 d for extract . .. Metals ( except mercury) p HNO3 6m Mercury p HNO3 28 d I Hexavalent chromium p Cool to 4°C 24 h I Acidity, Alkalinity P,G Cool to 4°C 14 d Ammonia, COD, total P,G H2SO4, 28 d Phosphorus, Nitrate-Nitrite Cool to 4°C I BOD P,G Cool to 4°c 48 h I Chloride, Fluoride p None 28 d Color P,G Cool to 4°C 48 h I Cyanide P,G NaOH, 14 d Cool to 4°C I Hardness p HNO3 6m pH P,G None Immediately I Nitrogen, Kjeldahl and G H2SO4, 28 d Total organic, Phenols Cool to 4°C 118 WKJ.lZlS · OJ/93 4-4 11 I ~-I TABLE 4-1 I CONTAINERS, PRESERVATIVES, A.t'!D HOLDING TIMES (continued) I ANALYSIS CONTAINER PRESERVATIVE HOLDING TIME I Nitrate & Nitrite P,G Cool to 4oc 48 h Oil and Grease G H2S04, 28 d Cool to 4°C I Ortho-phosphate P,G Cool to 4°C 48 h I Disso!ved oxygen G None Immediately Residue, Total, Filterable, P,G Cool to 4°C 7d N onfilterable, Volatile I Residue, Settleable P,G Cool to 4°C 48 h .. Silica p Cool to 4°C 28 d Sulfate, Specific P,G Cool to 4°C 28 d Conductance I Sulfide P,G Zinc acetate + 7d NaOH, Cool to 4°c I Sulfite P,G None Immediately Surfactants, Turbidity P,G Cool to 4°c 48h I Temperature P,G None Immediately I TOC G H2S04 28 d Cool to 4°C TOX G H2S04 8d I Cool to 4°c NOTES: I P = Polyethylene G = Glass I h = hours d = days m = months {i W8l,1Zl1 • r:JJ/93 4-5 I I •• I I I I I I I .. I I I I I I I •"' I ATTACHMENT A LABORATORY QUALITY ASSURANCE MANUAL CHESTER LABNET MONROEVILLE, PA I ~-I I I I I I I .. I I I I I I I ft I QUALI1Y ASSURANCE MANUAL ... CHESTER LABNET MONROEVILLE APPROVALS Approved by:._-1.:..µ.~!-!..I.Ls:i..l<lc,~~.i......: Approved by:__.._/_L-./ _ ___;;_f.._:._(L ___ _ Laboratory Director Date: '3 / q / q 3 I I ~-I I I I I I I .. I I I I 'I I I •• I TABLE OF CONTENTS Page 1.0 INTRODUCTION ............................................................................................................... 1-l 2.0 OBJECTIVES ...................................................................................................................... 2-l 3.0 ORGANIZATION AND RESPONSIBILl1Y .............................................................. .3-l 3.1 CHESTER LabNet. ................................................................................................ 3-l 3.2 Laboratory ................................................................................................................ 3-l 3.2.1 Organization ................................................................................................ 3-l 3.2.2 Responsibility .............................................................................................. 3-l 4.0 SAMPLE CUSTODY ...................... ; .................................................................................. 4-l 4.1 Custody and Documentation in the Field .......................................................... .4-1 4.2 Sample Custody in the Laboratory ...................................................................... .4-3 4.2.1 Sample Receipt in the Laboratory .......................................................... .4-6 4.2.2 Shipment Inspection ................................................................................. .4-6 4.2.3 Reconciliation with Chain-of-Custody Document ............................... .4-7 4.2.4 Resolution of Shipment Irregularities .................................................... .4-8 4.2.5 Sample Log In ............................................................................................ .4-8 4.2.5.1 Entry in Master Log ..................................................................... 4-9 4.2.5.2 Job Traveller ............................................................................... 4-10 4.2.5.3 Sample Labelling ....................................................................... 4-10 4.2.5.4 Information Distribution and Job Filing ................................ 4-11 4.2.6 Custody Transfer Within the Laboratory ............................................ 4-11 4.2.7 Sample Disposal ...................................................................................... 4-12 5.0 ANALYTICAL PROCEDURES ........................................................................................ 5-1 5.1 Coordination of Activities ..................................................................................... 5-1 5.2 Preparation of Sample Containers ....................................................................... 5-1 5.3 Instrument Maintenance ....................................................................................... .5-2 5.3.1 Preventative Maintenance ....................................................................... .5-2 5.3.2 Repair After Instrument Failure ............................................................ .5-3 5.4 Preparation of Standards ....................................................................................... 5-4 5.5 Determination of Detection and Quantitation Limits ..................................... .5-8 5.5.1 Instrument Detection Limit ...................................................................... 5-8 5.5.2 Method Detection Limit ........................................................................ 5-10 5.5.3 Quantitation Limits ................................................................................. 5-12 Page i I ~-I I I I I I I .. I I I I I I I f' I - 6.0 TABLE OF CONTENTS ( continued) 5.5.4 Conversion of Detection Limits to Minimum Page Detectable Concentration ...................................................................... 5-12 5.5.5 Documentation of Detection Limits .................................................... 5-13 5.5.6 Application of Limits in Data Reporting ............................................ 5-13 5.6 Instrument and Equipment Calibration ........................................................... 5-14 5.7 5.8 5.6.1 Initial Calibration .................................................................................... 5-14 5.6.2 Continuing Calibration ........................................................................... 5-16 5.6.3 Calibration Frequency ............................................................................ 5-17 Monitoring Laboratory Reagent Water ........................................................... 5-20 Analysis of Quality Control Samples ................................................................ 5-20 5 .8.1 Blanks ........................................................................................................ 5-20 5.8.1.1 Reagent Blank ............................................................................ 5-20 5.8.1.2 Method Blank ............................................................................. 5-21 5.8.2 Spiked Blank ............................................................................................ 5-22 5.8.3 Spiked Sample .......................................................................................... 5-22 5.8.4 Sample Duplicate .................................................................................... 5-23 5.8.5 External Quality Control Audit ............................................................ 5-24 5.8.6 Record Keeping on Analysis of QC Samples ..................................... 5-24 5.9 Use of Surrogates ................................................................................................. 5-25 5.10 Establishment of Acceptance Criteria .............................................................. 5-26 5.11 Development of New or Modified Methods ................................................... 5-30 DATA HANDLING ............................................................................................................ 6-1 6.1 Data Recording ....................................................................................................... 6-1 6.1.1 Notebooks .................................................................................................... 6-1 6.1.2 Record Keeping in Sample Prepartion ................................................... 6-2 6.1.3 Record Keeping for Instrument Analysis ............................................... 6-4 6.1.4 Record Keeping in Noninstrument Analyses ......................................... 6-6 6.2 Data Reduction ....................................................................................................... 6-6 6.3 Data V alidation ....................................................................................................... 6-7 6.4 Data Compilation ................................................................................................ 6-10 6.5 Final Review ......................................................................................................... 6-10 Page ii I ~-I I 7.0 I I I I I .. I I I I I I I •• I TABLE OF CONTENTS (continued) Page CORRECTIVE ACTION ................................................................................................... 7-1 7.1 Identification of Potential Problem ..................................................................... 7-1 7.2 Problems and Actions ............................................................................................ 7-2 7.2.1 7.2.2 7.2.3 7.2.4 7.2.5 7.2.6 7.2.7 7.2.8 Continuing Calibration Outside Acceptance Range ............................ 7-2 Calibration Standards Exceeding the Permitted Holding Time .................................................................... 7-2 Laboratory Method Blanks Exceed Method Detection Limit but are Below Quantitation Limit. ................................................ 7-3 Laboratory Method Blank Exceeds the Quantitaiton Limit. .............. 7-3 Laboratory Control Standard Exhibits Recoveries Outside the Acceptance Criteria ............................................................. 7-4 ~~Ya!t~/!~c~~~~e st~t~ri~!~'.~.~.~.~~.~~.~~·~········· ...................... 1-4 Control Chart Exhibits and Regular Trend ........................................... 7-5 Internal and External Audits and Corrective Actions .......................... 7-5 7.2.8.1 Evaluation of System Audits ....................................................... 7-5 7.2.8.2 Corrective Action and Feedback. ............................................... 7-6 7.2.8.3 Documentation of Corrective Action ........................................ 7-6 Page iii I I •• LIST OF TABLES I Page I Table 4-1 Containers, Preservative, and Holding Times ................................................... .4-4 Table 5-1 I Table 5-2 Standards and Solutions Holding Times ............................................................. 5-5 Calibrations Frequencies .................................................................................... 5-18 I Table 5-3 Surrogate Compounds and Typical Recoveries .............................................. 5-26 I I .. I I I I I I I ft Page iv I I ~-I LIST OF FIGURES I ·Page I Figure 3-1 Chester LabNet Organization Chart ................................................................ 3-la Figure 3-2 Laboratory Organization .................................................................................... 3-lb I Figure 4-1 Chain-of-Custody Record ... , ............................................................................... 4-2a I Figure 4-2 Figure 4-3 Sample Shipment Checklist ................................................................................ 4-6a Sample Master Log ............................................................................................ 4-l0a I Figure 4-4 Job Traveller ....................................................................................................... 4-l0b Figure 4-5 Internal Transfer Form ..................................................................................... 4-1 la I Figure 5-1 Analytical Request Form .................................................................................... 5-la .. Figure 5-2 Figure 5-3 Corrective Action Notice .................................................................................... 5-3a Standard Preparation Log .................................................................................. 5-7a I Figure 6-1 Figure 6-2 Continuous Extraction Bench Sheet. ................................................................ 6-3a Aqueous Sample Metals Digestion ................................................................... 6-3b I I I I I I ,, I Pagev ,, I I I I I I I .. I I I I I I I ~ I - 1.0 INTRODUCTION CHESTER LabNet is committed to excellence in analysis. All data generated by CHESTER LabNet must be technically sound, properly documented, legally defensible, and supported by defined and verified confidence limits. The program therefore, whenever possible, employs analytical methods derived from EPA, ASTM, AOAC, or Standard Methods. This document is designed to serve as a guideline to CHESTER LabNet as a whole. Specifically, it defines the laboratory objectives, organization, functional activities, and QA/QC programs that routinely apply to each laboratory. This document is also supplemented by sets of Standard Operating Procedures. The standard practices within CHESTER LabNet are set forth; however, for specific projects, addenda will be prepared corresponding to project needs . W8J,J232 • aJ/'JJ 1-1